Release Notes

Drivers and Tools

Version Overview
Release Date Drivers and Lowest possible PrismaPro2 Notes
(y-m-d) Tools firmware recommendation
version for full
functionality
2021-10-13 4.6.21286 HAx80 v2.11.0 PP 2.7.4.00127 Released, tested on
Windows 10 build
1909 and 2004
2021-10-05 4.6.21266 HAx80 v2.11.0 PP 2.7.3.65518 Released, tested on
Windows 10 build
1909 and 2004
2021-02-02 4.6.21013 HA480/HA680 PP 2.7.3.65518 Released, tested on
V2.06.4 Windows 10 build
1909 and 2004
2020-10-08 4.6.20282.1 HA480/HA680 PP 2.7.3.65518 Released, tested on
V2.06.4 Windows 10 build
1909 and 2004
2020-04-02 4.5.2.1178 HA480/HA680 PP 2.7.3.651 Released, tested on
V2.06.4 Windows 10 build
1903 and 1909
2019-11-29 4.5.2.1138 HA480/HA680 PP 2.7.3.651 Released, tested on
V2.06.4 Windows 10 build
1903
2019-09-20 4.5.2.1095 HAx80 PP 2.7.2.630 Released
v2.02

2019-04-19 4.5.0.829 HAx80 PP 2.7.2.630 Released


v2.02
X-PROTINT
v2.02

Fast & Fluid Management Page | 1 / 8


2019-02-13 4.5.0.602 HAx80 PP 2.7.2.625 Released
v2.02
X-PROTINT
v2.02
2018-12-19 4.5.0.470 HAx80 PP 2.7.2.625 Released
v2.02
Known issues:
X-PROTINT
In simulation
v2.02
dispensing is limited
to three colorants.
Switching from a
manual to a different
scale requires a
restart in CW4
2018-12-03 4.5.0.385 HAx80 PP 2.7.2.625 Released
v2.02
Known issue: in
X-PROTINT simulation dispensing
v2.02 is limited to three
colorants.
2018-09-17 4.5.0.123 HAx80 PP 2.7.2.614 Released
v2.02
When the 4 flush
X-PROTINT times functionality is
v2.02 required for HAx80,
(item 1894) the
minimum firmware
version is 2.03.4
2018-05-08 4.4.1.598 HAx80 PP 2.7.1 Released
v2.02
For PrismaPro 2 users:
when refilling in kg is
desired for HAx80 or
X-PROTINT you need
at least this Drivers
and Tools version
2018-01-25 4.4.0.574 HAx80 Released
v2.02
2017-03-30 4.3.0.219 Released
Does not support
HA680

Fast & Fluid Management Page | 2 / 8


Release notes per version

Drivers and Tools 4.6.21286 (2021-10-13)


14978 New default 29ml calibration and verification plans are supplied when there are no
plans defined in the .ide4 file.

Drivers and Tools 4.6.21266 (2021-10-05)


14765 Various items in relation to the launch of AutoDrive2. AutoDrive2 needs HAx80
firmware 2.11.0 or higher to work. Please refer to your Sales Manager for more
information.
14608 Added Radwag scale support.
13099 The gravimetric algorithm will now stop dispensing when the measured weight
differs more than 20 percent from the expected weight. This used to be 10 percent.
12939 Various changes to easily inspect gravimetric results in the timeline.
12938 Changed the density tolerance from 10 to 20 percent in the gravimetric algorithm.
12904 Added the pump type “29 ML” to the driver to support sample dispensing.
12882 Improved the gravimetric algorithm to better handle over- and under dispenses.
11994 The driver is now able to handle negative weights from the scale.
11591 Added WANT scale support.
11169 Improved robustness of scale taring procedure.

Drivers and Tools 4.6.21013 (2021-02-02)*


10503 Added support for the Kern PLE scale to the scale driver.
10312 Improved support for Sartorius scales.
10196 Improved translations of several languages including Spanish and Polish.
10081 We now distinguish valve, pump and piston travel in the Service Dashboard.
9929 Added the refill amount to the Refill event in the timeline.
9921 Added the option to create a custom scale.
9880 It’s now possible to filter on events of a specific channel in the timeline.
9873 Added initializations of the driver to the timeline in the Service Dashboard.
9826 Added driver and firmware updates to the timeline in the Service Dashboard.
9707 Fixed an issue with the display of levels in the Service Dashboard.
9665 Included the correct version of the Calibration Analyzer excel sheet in the installer.
9661 Improved the Service Dashboard so that it is more clear what scales are supported.
9487 Added Autodrive events to the timeline in the Service Dashboard.
Fast & Fluid Management Page | 3 / 8
3785 Introduced the ‘Service pump’ actions to the Maintenance tab of the Service
Dashboard. The Service pump action assists the service engineer in resetting the
correct dispenser parameters after servicing a pump.

Note: The maintenance actions are registered in the timeline.


2997 Added the KPI Power-on hours. This is a KPI that measures the percentage of hours
the dispenser has had power in the last 30 days. The dispenser needs power to
perform important autonomous processes like stirring.
* Known issues: The timeline feature in the standalone Service dashboard application is not usable
and may crash the application. The timeline works fine when the service dashboard is opened via
the IDD Test.net application.

Drivers and Tools 4.6.20282.1 (2020-10-08)*


8852 Improved the gravimetrical dispensing algorithm in IDD4. Please refer to the
Gravimetric Dispensing manual of Fast & Fluid Management for a full description.
8767 Canisters with a colorant level below the minimum level are now skipped during
purging.
7870 The Service Dashboard now contains a Timeline. The Timeline currently shows
dispenses, errors, refills and purges.
7828 The IDD2 driver now correctly considers the configured Agitation duration again.
7792 Made several improvements to the user interface of the Service Dashboard.
7664 Added X-SMART XL support in the driver.
7565 IDD Test now correctly considers the canister size when refilling.
5855 Improved the Scale Driver. It’s now possible to select a scale from the Service
Dashboard.
3418 IDD Test now handles grams correctly.
3386 Fixed FFMAssistent Collect Service data function.
3318 The Service Dashboard will now start in the language you choose during installation.
2908 Improved handling of decimal separators in IDD Test.
* This Drivers and Tools release includes and requires .NET Framework 4.8.

Drivers and Tools 4.5.2.1178 (2020-04-02)


5288 Fixed a bug that in rare occasions could cause turntable position errors on HAx80s
dispensers.
4517 Corrected an issue that prevented UDCP from running.
4451 The Service dashboard now immediately displays the Autodrive percentage changes
of an X-PROTINT dispenser.

Drivers and Tools 4.5.2.1138 (2019-11-29)


4494 Improved the calculation of the “Average number of tints between cleans” value in
the Service Dashboard.
4454 Improved the calculation of the “Colorant Dispensed” value in the Service
Dashboard.

Fast & Fluid Management Page | 4 / 8


4451 The Service Dashboard now correctly shows Autodrive speed adjustments for X-
PROTINT again.
4438 The Service Dashboard now opens a lot quicker when a lot of data is available.
4400 CW4 now correctly cleans the nozzles during calibration sessions of an X-PROTINT.
4359 The Service Dashboard will now open regardless if the dispenser is in error state.
4391 The Service Dashboard will now show channel and job information for machine
errors when it’s available for HA480/HA680 dispensers. This requires firmware
version 2.06.4.

The channel and job information are not available yet for the HA180 dispenser. This
will be made available in a future firmware update.
4312 Added a default Validation plan for the HA180 to CW4.
4261 Fixed a bug where the values entered in the 'Date installed' field and the 'Location'
field were not stored in the database.
4226 It’s now possible to dispense from IDDtest.NET using the IDD2 driver again.
4207 Improved the Italian translations of IDD2 and IDD4.
4056 Made an improvement to the firmware upload mechanism of FFM Maintenance.

Drivers and Tools 4.5.2.1095 (2019-09-20)


3927 Reduced the risks of .ide4 files getting corrupted by improving the way setting
changes are saved.
3905 Updated the driver translations of the French, German, Italian, Dutch, Polish, Czech
language.
3631 We implemented a mechanism which allows us to show machine errors (E02, E14,
etc.) in the Service dashboard. In this release, only a small number of machine
errors can be shown. This number will increase with future releases.
3547 Backups are now created in the same folder as the .ide4 file. This solves access
rights issues at customer sites with very strict security policies.
3261 CW4: the default calibration profile is now saved correctly.
3232 The Spanish translation file is now included in the installation of IDE4.
3056 CW4: The verification script now shows the absolute deviation in millilitres and the
deviation as a percentage.
3009 Fixed a couple of issues with English text strings that were not translatable.
2933 Improved the performance of the Set All functionality in the Channel window of the
Service Dashboard.
2853 IDD4 now shows the active language in the dispenser settings screen.
2836 It’s no longer necessary to have administrator rights to change the language of the
Service Dashboard.
2664 Fixed an issue in FFM Maintenance that prevented resetting mixers and shakers.
2606 When back-ups of .ide4 files are restored an event is logged in the Service
Dashboard.
2597 CW4: Switching from a manual scale to an actual scale no longer causes an error.
2394 CW4: When using a test script, you now receive a message when there is not
enough colorant.
1513 CW: The 'Active calibration' mark of a session is now saved correctly.

Fast & Fluid Management Page | 5 / 8


Drivers and Tools 4.5.0.829 (2019-04-19)
2985 Fixed an issue where the installer failed to register the scale driver on a Turkish
Windows.
2870 Fixed a bug where ‘Pump travel distance’ and ‘Piston travel distance’ did not update
after a reset.
2842 Changes in the fields ‘Location’ and ‘Date installed’ in the Service dashboard are
now saved correctly
2815 A Canister is now only disabled if all Channels are disabled
2804 Fixed a bug where the canister warning levels were not checked correctly when
dispensing in Kilograms

Drivers and Tools 4.5.0.602 (2019-02-13)


2704 Switching from a manual to a different scale no longer requires a restart in CW4
2680 A bug that caused a delay after each dispense with a small amount of colorant was
fixed
2593 In the Service dashboard the refresh button now functions like it did in versions
prior to 4.5.0.385
2578 In the Service dashboard the firmware is now fully visible
2562 The IDD4 driver now dispenses more than 3 colorants in Simulation
2561 The back-up procedure of the .ide4 file and .levels file has been made more robust
2431 In the Service dashboard the ‘Colorant dispensed’ and the ‘Recipes dispensed’ in
the last 365 days are now visible
2429 In the Service dashboard additional metrics are shown for Refill observations
2428 In the Service dashboard additional metrics are shown for Brush maintenance
2362 In the Service dashboard the design of the ‘Set for all channels’ option has been
improved
2164 The default Brush clean interval was set to 7 days (from 14 days)

Drivers and Tools 4.5.0.470 (2018-12-19)


CW4*
2590 Fixed an issue in CW4 that prevented calibrations.
* Known issues: In simulation dispensing is limited to three colorants and switching from a manual
to a different scale requires a restart in CW4.

Drivers and Tools 4.5.0.385 (2018-12-03)


IDD*
2446 Fixed an issue in the LocalDB where (in simulation mode) multiple identical entries
were added to the Devices table.
2415 Table slack is now displayed in the Service Dashboard.
2409 The LocalDB now correctly displays the firmware version and the driver version.
2370 In the maintenance tab of the Service dashboard it is now possible to swap two
colorants and their channel and component settings. This can be done with
colorants contained in canisters with a similar pump size.
* Known issue: in simulation dispensing is limited to three colorants.
Fast & Fluid Management Page | 6 / 8
Drivers and Tools 4.5.0.123 (2018-09-11)
IDD
2269 There are now 3 more possible flush moments enabled and the original flush
moment is preserved as the first flush moment. Channels will be configured to use
only the original flush time to ensure there is now impact on existing flush settings.
2186 Backups of the ide4 and levels files are now in a separate folder below dispenser
folder.
2184 A service engineer can now reset the meters the piston (and pump) travelled from
the channels settings by right clicking on a channel header.
Calibration Workshop 4
2208 The execution of a verification script now stops after the first error
2207 In CW4 It is now possible to abort execution of a verification script after it has
started.
2205 The installer now writes information tin the installation logs when the
CurrentLanguage is not found
2195 From CW4 it is no longer possible to set a canister level higher than the total size of
the canister.

Drivers and Tools 4.5.0.96 (2018-08-14)


IDD
2151 The input window for manual scales will no longer drop behind the main window
2119 Improved readability on the dashboard (flowrate view)
2106 A service engineer can now export the data on the dashboard to csv format
2032 The operator can see the actual flowrate when auto drive is enabled (for both X-
PROTINT and HAx80)
2005 Dashboard and Channel views improved. On the Channel view a reset button is
added for auto drive reset to 100%
1996 For third-party software the service technician or support employee can find
information in the log file when a IDD2 license is not valid.
1981 When a (Third Party) formula book program sets the language, it's now applied
1980 During installation the IDD3 installation Message is no longer displayed.
1958 Place and remove can are now translatable
1938 A message is given by the device driver when a dispenser has an incompatible
firmware version.
1894 It is now possible to configure flush settings per channel.
1856 Improved the flowrate prediction in L\min in the Service Dashboard to be more
accurate and match the actual amount in Liters that can be dispensed in 1 minute.
(calculations are triggered during each initialization)
1648 When the dispenser is flushing during an Initialization a correct error message is
given.
1637 A backup of the Dispenser file is created before saving
1605 Below minimum dispense amount does not show a second message when first
component was already cancelled.
1593 Added functionality to present a warning to the operator for dispensing below the
minimum amount.

Fast & Fluid Management Page | 7 / 8


1303 Drivers and tools can be used in simulation mode

IDD Test.Net
1970 Fixed a bug that could occur when refilling multiple Canisters.

FFM Maintenance 2
1965 FFM Maintenance crashed when the windows format is set to EN(nl). This has been
fixed.

Calibration Workshop 4
2159 The About window now displays correct information
2150 Dropdown menu properly hides when selecting Settings/Verification Calibration
plans
1957 Calibration Workshop 4 does no longer give an error when cancelling a "select
configuration file" action.
1700 The checkbox to make a calibration plan the Default calibration is now working
correctly.
1634 New default 100 and 300ml calibration and verification plans are supplied when
there are no plans defined in the .ide4 file.
14978 New default 29ml calibration and verification plans are supplied when there are no
plans defined in the .ide4 file.

IDD2 2.7.4.211 (2018-05-17)


IDD
1800 Fixed a possible issue where IDD2 was unable to write log data to the logfiles in the
IDD2\Logs folder.

Fast & Fluid Management Page | 8 / 8