Impact on the analyzer installed base : Affected Spare Parts / Kits: Estimated implementation labor time:
ALL Liaison® XL, Liaison® XL LAS, N.A. 2,5 hours.
Liaison® XL PMA.
ALL Countries except for CHINA. Note:
N.A.
Timeline:
Within 9 months from the release of Rev.
A of this TN (December 2018).
1. DISTRIBUTION ............................................................................................................................... 3
2. PURPOSE ......................................................................................................................................... 3
3. ADMINISTRATIVE NOTES ............................................................................................................... 3
3.1. SGAT ( FOR SUBS ONLY ) ..................................................................................................................... 3
3.1.1. Worksheet (WS) management ................................................................................................... 3
3.1.2. Additional information ............................................................................................................... 3
3.2. FOR DISTRIBUTORS .............................................................................................................................. 3
4. SPARE PARTS .................................................................................................................................. 4
5. PROCEDURE .................................................................................................................................... 4
5.1. INSTALLATION PROCEDURE...................................................................................................................... 4
Page 1 of 6
validity extended until next revision
5.1.1. Important remarks for the installation ........................................................................................ 4
5.1.2. Installation Check-List ............................................................................................................... 4
5.2. DOWNLOADS ....................................................................................................................................... 4
5.3. USER SOFTWARE CHANGES ...................................................................................................................... 5
5.4. USER DOCUMENTATION .......................................................................................................................... 5
5.5. UNSOLVED ANOMALIES ........................................................................................................................... 5
5.5.1. Important remarks .................................................................................................................... 5
5.6. CUSTOMER TRAINING ............................................................................................................................ 5
6. REQUALIFICATION ......................................................................................................................... 5
7. PPE .................................................................................................................................................. 6
Revision History
Revision Date Changed by Detail
DiaSorin Confidential
Page 2 of 6
validity extended until next revision
Subject: Release of software version 4.2.1.1
ID Number: 172 Revision: A Date: 08/03/2018 Author: D.BELLUTO
1. Distribution
All DiaSorin Subsidiaries (except CHINA) and Distributors.
2. Purpose
The aim of this Technical Note is to release the Liaison® XL SW version 4.2.1.1.
3. Administrative notes
The installation of Liaison® XL SW version 4.2.1.1 applies to Liaison® XL, Liaison® XL LAS and
Liaison® XL PMA.
During each service intervention, Field Service Engineer (FSE) and/or System Specialist (SS) has to
identify and check the root cause of the instrument failure (if any), and evaluate the related effect.
If the identified effect may lead to potential risks for the sample result and/or the end user, FSE /
SS is required to report the event to the reference Service Manager and to Corporate Service
department for the proper assessment.
Page 3 of 6
validity extended until next revision
Subject: Release of software version 4.2.1.1
ID Number: 172 Revision: A Date: 08/03/2018 Author: D.BELLUTO
4. Spare Parts
N.A.
5. Procedure
5.1. Installation procedure
The Liaison® XL SW installation procedure is described within the following document:
XL-SDI-N11-T (171218) (Rev.C) Liaison XL SW V4.2.1 Installation Procedure.pdf
5.1.1. Important remarks for the installation
The software upgrade will reset the file ‘BottleTypes.txt’ (the system will not maintain the data
from the pre-existing version). Right after the installation, in case of customized sample rack,
check and update the ‘BottleTypes.txt’.
In a near future, new manufactured analyzers will be equipped with a Sample Bay Top
Cover (to comply with new requirements of EN 61010-1:2010 - Safety requirements for
electrical equipment for measurement, control, and laboratory use. General requirements). For
compatibility reasons, such hardware change will not allow anymore the usage of racks “C” and
“D”. Therefore, Liaison XL SW V4.2.1.1 disables by default the recognition of C.rac and D.rac
file.
NOTE – Sample Bay Top cover is NOT intended to be installed on currently installed analyzers.
For all analyzers without the sample Bay Top cover, the usage of racks “C” and “D” is still
allowed. Once the upgrade to Liaison XL SW V4.2.1.1 is performed, as also indicated in the
Liaison XL SW V4.2.1 installation procedure, in order to restore the correct recognition:
- Copy the C.rac and D.rac files present in the folder “D:\LiaisonXL\Data\9-pos” and paste them
into the folder “D:\LiaisonXL\Data”.
5.1.2. Installation Check-List
The Liaison® XL SW installation Check-List is available within the following document:
XL-SDI-N33-T (171212) (Rev.A) LIAISON XL SW V4.2.1.1 Installation Check-List.pdf
This document is a guideline that could be used by FSE. It is not required to be attached to
SGAT WS record.
5.2. Downloads
The software package is available on FileZilla at the following path:
- Remote site: /DATA/LXL
- Filename: LXL_SW_Setup_4.2.1.1.zip
Page 4 of 6
validity extended until next revision
Subject: Release of software version 4.2.1.1
ID Number: 172 Revision: A Date: 08/03/2018 Author: D.BELLUTO
- For INTERNAL USERS (including DiaSorin’s personnel and Distributor’s personnel only)
XL-SDI-N47-M (171220) (Rev.A) LIAISON XL Introduction to SW V4.2.1 (internal
users).pdf
Local organizations may translate the document in local language but it is NOT to be
delivered to the end user.
6. Requalification
In order to re-qualify the analyzer after SW installation, perform the following tasks via User
Software logging in with the USER access rights:
Page 5 of 6
validity extended until next revision
Subject: Release of software version 4.2.1.1
ID Number: 172 Revision: A Date: 08/03/2018 Author: D.BELLUTO
7. PPE
Picture of the PPE: Description of the PPE: Activity:
N.A. N.A. N.A.
Page 6 of 6
validity extended until next revision