Software V4.2.1
LIAISON® XL
Introduction to Software V4.2.1
(all users)
Contents
1. MAIN PURPOSE OF SOFTWARE V4.2.1 .............................................................. 2
2. INCLUSIONS ....................................................................................................... 2
3. COMPATIBILITY ................................................................................................. 2
3.1. SYSTEMS .............................................................................................................. 2
3.2. ASSAY PROTOCOLS .................................................................................................. 2
4. NEW FEATURES AND IMPROVEMENTS (FOR ALL USERS) ................................. 3
4.1. INHIBIT THE USE OF INTEGRALS STIRRED FOR LESS THAN 15' .............................................. 3
4.2. OFFLINE INTEGRALS REMOVED FROM INVENTORY PAGE ...................................................... 4
4.3. NEW COMMAND TO TRIGGER RESUME VIA LAS ................................................................ 5
4.4. FINISHEDJOBS.TXT FILE POPULATED ON A SECONDARY PATH ............................................... 6
4.5. NEW SAMPLE BAY TOP COVER AND ANCILLARY RACK .......................................................... 6
4.6. CONFIGURATION UPDATES ......................................................................................... 7
4.7. IMPROVED SYSTEM RELIABILITY .................................................................................. 9
2. Inclusions
The changes included in SW V4.2.1 are listed in comparison to V4.2.0.2. DiaSorin released
additional Hot Fixes for V4.2.0.2, namely
- LXL_HotFix_201
This Hot Fix has been included in V4.2.1, and it is not reported here as previously
documented and notified separately.
3. Compatibility
3.1. Systems
The LIAISON® XL software V4.2.1 is compatible with all LIAISON® XL, LIAISON® XL LAS and
LIAISON® XL PMA systems. For the sake of brevity, the system may be referred to as
LIAISON® XL.
Update (V4.2.1)
In V4.2.1, the user is provided with an option to force the system to ignore kits for which the
15’ countdown was not elapsed yet.
The kits for which the 15’ countdown was not elapsed yet are not reported:
in the Status Inventory page.
to the LAS system within the Inventory message
As a result the system will not schedule any test on those integrals, until the countdown
reaches the end.
A dedicated message (see User Manual Event ID: ”262.004.00005”) is used to inform the
user in case the system is not able to start a test due to this option.
Important notes
! The new functionality is not intended to replace the needed instructions that shall be
followed as detailed in the assay Instructions For Use.
Update (V4.2.1)
In V4.2.1, system is removing from the Inventory page those integrals, which were set
offline due to no/not enough liquid detection.
The same integrals are no longer communicated as available to the LAS system within the
Inventory message.
Important notes
! As per explanation provided for Event ID “265.000.00045” and “265.003.00017” (see User
Manual), when an integrals is set offline it shall be checked to ensure the proper integral
conditions. The “Reset LLD” function provided in the Diagnostics page can be used to
reset offline conditions due to no/not enough liquid.
Update (V4.2.1)
With SW V4.2.1 the LAS system can make use of a new dedicated command to trigger the
resume of the LIAISON XL from remote.
If LIAISON XL is in a suitable status for executing the resume (i.e. Stand By), it will positively
answer to the received command and start the resume process.
The LAS system can get feedback that the resume successfully completed by polling the
LAISON XL system for receiving its current status.
Important notes
! Due to the LAS protocol architecture, the new feature has no impact on any current LAS
Interface/System.
Update (V4.2.1)
In V4.2.1, the FSE can configure a secondary path (that can be a network path), where the
system saves and keeps updated a file with the same contents of the FinishedJobs.txt. This
copy is also named FinishedJobs.txt.
The system will not automatically remove the copy of the FinishedJobs.txt file.
Configuration
This option shall be configured by DiaSorin representatives.
Important notes
! .The LIAISON XL shall be always granted of Read/Write access to the secondary path/file
when a new result is obtained or a dose is calculated.
The FinishedJobs.txt file only reports those results that are accompanied by an actual RLU
measurement (i.e. in case more than one replicate is calculated, only the single replicates are
reported and only the single son results are reported for combi tests).
Update (V4.2.1)
SW V4.2.1 installs the new control rack file (T.rac) that is compatible with the sample bay
cover.
The new T.rac is described in chapter §4.2.5 of the User Manual.
The new Ancillary rack is described in chapter §4.2.6 of the User Manual.
Important notes
! Analyzers provided with the sample bay top cover do not support “C” and “D” racks. In case
“C” and “D” racks are loaded on such analyzers will not be recognized and used.
Update (V4.2.1)
Software V4.2.1 includes a version of User Manual and Quick Guide updated to the latest
features, as well as the translation of the User Software itself in the following languages:
Notes:
(1) Distributed separately from the LIAISON XL Software
(2) Uses the English version
(3) Same as English USA
Language specific errors have been corrected in the Export/Printouts and User Interface
elements.
The virus definition has been updated.
A new cumulative security patch for Windows 10 based systems is applied. A new setting for
Windows Vista based systems is applied to improve cybersecurity.
Update (V4.2.1)
V4.2.1 addresses the identified improvement areas to introduce related countermeasures.
The Export file, generated form the Results –> Controls page now features all the
expected columns.