You are on page 1of 10

ClinicalArticle

Clinical Usefulness of the


EASI 12-Lead Continuous
Electrocardiographic
Monitoring System
Mary Jahrsdoerfer, RN, MHA
Karen Giuliano, RN, PhD
Dean Stephens, RN, MS

T he 12-lead electrocardiogram
(ECG) is the diagnostic reference stan-
dard for evaluating cardiac rhythm
tion, however, is not practical for
ambulatory patients in the telemetry
setting. Subsequently, over the years, we
of the heart, but rather to derive
continuous monitoring information
from the 1 or 2 most commonly
and myocardial ischemia. The tradi- have used a tool that captures limited used ECG leads. Recent research
tional 12-lead ECG system requires and partial ECG data with just 1 or 2 and national guidelines on monitor-
10 electrodes that are strategically views of the heart. This tool consists of a ing for myocardial ischemia2-6 rec-
placed on the chest and the extremi- portable monitor (commonly referred ommended continuous 12-lead
ties (Figure 1). This lead configura- to as a telemetry box), and 3 or 5 leads ST-segment monitoring for at-risk
(depending on the type of telemetry patients because silent ischemia
that the institution has installed). (without chest pain) is common and
Typically, 4 leads are attached to the may not always be detected with the
patient’s torso (modified extremity same leads. The busy environment of
* This article has been designated for CE credit. A leads), and the last lead acts as the areas devoted to patients’ care, cou-
closed-book, multiple-choice examination follows
this article, which tests your knowledge of the precordial or “V” lead (Figure 2). This pled with the need for frequent 12-
following objectives: method has been used successfully lead ECG analysis, makes using a
1. Identify the differences between standard
and continuous 12-lead systems for continuous cardiac monitoring.1 conventional, diagnostic 12-lead ECG
2. Identify 3 clinical conditions that are appropriate Telemetry monitoring was not device an impractical approach for
for continuous 12-lead monitoring
3. Describe the benefits of examining 12-lead designed, however, to obtain 12 views continuous interpretation of cardiac
electrocardiograms in the clinical setting
rhythms and myocardial ischemia.
Authors What is most needed in the clinical
monitoring environment is a contin-
Mary Jahrsdoerfer is a clinical consultant for Philips Medical Systems in the New
York–New Jersey metropolitan area. uous 12-lead ECG monitoring sys-
tem to replace the traditional 3- to
Karen Giuliano is a clinical research specialist for Philips Medical Systems in Andover, Mass.
5-lead telemetry monitoring system
Dean Stephens is the assistant director of nursing at North Shore University Hospital in now used in most hospitals for con-
Manhasset, NY. tinuous ECG monitoring.
To purchase electronic or print reprints, contact The InnoVision Group, 101 Columbia, Aliso Viejo, CA 92656. In this article, we describe the
Phone, (800) 809-2273 or (949) 362-2050 (ext 532); fax, (949) 362-2049; e-mail, reprints@aacn.org.
clinical outcomes that occurred after

28 CRITICALCARENURSE Vol 25, No. 5, OCTOBER 2005


ECG configura- Dower and colleagues7,8 developed in
tion and the 1980 a 12-lead ECG based on the
standard 12-lead vector ECG principles described by
ECG configura- Frank.9 Instead of the traditional 10
tion is that all electrodes, ECG signals are derived
12 views of the from 4 chest electrodes and 1 refer-
heart are avail- ence electrode (Figure 3). The elec-
V1-V6 able on a contin- trodes are placed on the upper part
uous basis with of the sternum (S), the lower part of
the derived con- the sternum (E) at the level of the fifth
figuration, intercostal space, and on the right and
RA LA rather than the left midaxillary lines (I and A) at the
periodic basis same level as the electrode on the
available with lower part of the sternum. A fifth
the traditional ground electrode can be placed any-
12-lead ECG where. This 5-lead configuration gives
system. us a 3-dimensional portrayal of the
On the basis electrical activity through the heart’s
of the clinical conduction system, rather than
need for contin- independent channels of unipolar
uous 12-lead and bipolar energy. The result is 12
ECG monitor- leads of information based on acqui-
RL LL
ing and the sition of simulatneous events in the
impracticality frontal, horizontal, and sagittal car-
of using a tradi- diac planes. A mathematical calcula-
Figure 1 Placement of leads for standard 12-lead electro- tional 12-lead tion using standardized and fixed
cardiography. configuration coefficients for each lead results in a
Abbreviations: LA, left arm; LL, left leg; RA, right arm; RL, right leg. for continuous linear transformation of the vectors,
Used with permission from Philips Medical Systems, Andover, Mass.
monitoring, creating a derived 12-lead ECG.

implementation of a new 12-lead


derived ECG system designed for
continuous, bedside monitoring as a
replacement for a traditional 3- to 5-
lead system.

Derived 12-Lead ECG for


Continuous Monitoring
Any time that the limb leads are
moved away from the extremities
and placed on a patient’s torso and
the precordial leads are modified
down to a single lead that is calcu-
lated for the unipolar precordial V
lead on the horizontal plane, the Figure 2 Typical placement of leads for 3-lead (left) or 5-lead (right) telemetry
with the Mason-Likar format.
result is termed a derived ECG.7 The
Used with permission from Philips Medical Systems, Andover, Mass.
main difference between a derived

CRITICALCARENURSE Vol 25, No. 5, OCTOBER 2005 29


versity Hospi- 12-lead system (Philips Medical Sys-
tal, one of the tems, Andover, Mass). The EASI 12-
leading medical lead monitoring system is integrated
centers on Long into patients’ monitors, allowing
Island in New display of up to 3 of the 12 continu-
York. The study ously monitored derived leads. The
was imple- EASI is the only system in the mar-
mented in 3 sep- ketplace that does continuous 12-lead
arate intensive monitoring with only 5 electrodes.
care telemetry By using the stored data (referred to
step-down as the full-disclosure feature), data
units, a total of from all 12 leads of the ECG can be
128 telemetry displayed and printed in either real
beds. Before the time or retrospectively for any period
Figure 3 Placement of leads for derived 12-lead electrocar- new system was in the preceding 24 to 96 hours.
diography. EASI lead placement: E, electrode placed on the
lower part of the sternum, level with the fifth intercostal installed, ECG Automatic documentation of derived
space; A, electrode placed at the level of the fifth intercostal monitoring 12-lead ECGs can also be triggered
space, on the left midaxillary line; S, electrode placed on the
upper part of the sternum; I, electrode placed at the fifth were accom- by heart-rate alarms or ST-segment
intercostal space, on the right midaxillary line. The fifth plished by using changes.7,8
ground electrode can be placed anywhere.
a traditional 3- All nursing and monitoring tech-
lead ECG sys- nical staff received extensive education
Importance of Continuous tem, with periodic standard 12-lead and training on the new monitoring
Monitoring of Multiple ECG Leads ECGs obtained as needed to detect system before use of the system was
One of the strongest arguments various dysrhythmias or myocardial implemented. The educational pro-
for continuous monitoring of all 12 damage. Nonlicensed personnel (mon- gram consisted of information about
leads of the ECG is the situation in itoring technicians) specially trained the benefits of examining 12 leads of
patients with suspected or confirmed for ECG monitoring do the continu- the ECG to improve diagnostic
myocardial damage. ECG monitoring ous observation of ECG rhythms in interpretation, how a 12-lead ECG is
is more sensitive than a patient’s self- the unit. derived from the 5 leads of the 12-lead
reporting of symptoms (eg, chest pain, The new telemetry system installed continuous monitoring system, and
chest discomfort) for detecting tran- at North Shore University Hospital is how and where to apply the 5 leads to
sient myocardial ischemia because currently being marketed as the EASI the chest. Staff had ample time for
70% to 90% of the episodes of myo-
cardial ischemia detected with ECG Table 1 Recommendations for clinical conditions for which continuous 12-lead
are clinically silent.2-5 A consensus electrocardiographic monitoring may be beneficial6
group of experts in ST-segment moni- Clinical conditions Recommended duration of monitoring
toring6 have suggested a variety of Unstable angina and ST-segment elevation Minimum of 24-48 hours or until event-free
clinical conditions and diagnoses for 12-24 hours
other than myocardial ischemia that Acute myocardial infarction without ST Minimum of 24-48 hours or until event-free
are appropriate for continuous 12- elevation for 12-24 hours

lead ECG monitoring (Table 1). Chest pain that prompts a visit to the 8-12 hours with determination of serum
emergency department biochemical markers of injury
Catheter-based interventions During procedure and for 6-12 hours after
Implementation of a New 12-Lead procedure in patients with unstable conditions
Continuous Monitoring System Cardiac surgery 24-48 hours postoperatively
In March 2002, we implemented
Noncardiac surgery for patients at risk During and immediately after surgery
a research study for a newly installed for myocardial ischemia
telemetry system at North Shore Uni-

30 CRITICALCARENURSE Vol 25, No. 5, OCTOBER 2005


Table 2 Evaluation questionnaire completed by personnel monitoring electrocardio- Table 3 Primary reasons for continu-
grams (ECGs) after implementation of a new continuous 12-lead ECG system ous electrocardiographic monitoring for
64 patients included in the evaluation
Questions Ratings
% of
1. How many times during the course of your J None J 1-2 times J 3-4 times Reason for monitoring patients*
shift did you go into full disclosure*? J 5-6 times J 7-8 times
J 9-10 times J >10 times Bradycardia 2

2. How many times during the course of your J None J 1-2 times J 3-4 times Hyperkalemia 2
shift have you accessed EASI 12-lead? J 5-6 times J 7-8 times Sinus arrhythmia 2
J 9-10 times J >10 times
Supraventricular
3. Did the physician request the EASI 12-lead? J No J Yes tachycardia 4
4. Was there any change in the interpretation J No J Yes (if yes, describe): Congestive heart failure 5
of a rhythm because of the retrospective
ECG information found in full disclosure*? Myocardial infarction 7

5. Has the patient’s treatment changed in any J No J Yes (if yes, describe): Ventricular tachycardia 7
way because of the information found in full Atrial fibrillation 16
disclosure?
Rule out myocardial
*Full disclosure is a decision support tool that provides clinicians with current and retrospective analysis of infarction 21
a patient’s condition and past events. At the information center, beat-to-beat analysis, waveforms, trends,
and alarms can be retrieved and reviewed wherever they are needed. It creates a continuous record of the Chest pain 30
patient’s electrocardiographic data that can be accessed no matter where the patient is transferred.
Missing information 11
*Percentages do not total 100 because some
patients had more than 1 reason for monitoring.
hands-on experience with the new sys- In response to the questions about
tem, and each staff member com- the use of the new 12-lead system,
pleted a competency checklist before most surveys
the new system was implemented. indicated use of
Table 4 Summary of results from the questionnaires completed
Four months after implementation the full-disclosure by monitor technicians on 64 patients
of the new monitoring system, data function of the % of
were collected for 4 months to evalu- system more Question Rating responses*
ate the system’s performance. Mem- than 10 times 1. How many times during the None 0
bers of the clinical staff were asked per shift, similar course of your shift did you go 1-2 times 3
into full disclosure? 3-4 times 3
to complete a voluntary question- to the frequency 5-6 times 3
naire whenever they were assigned of accessing the 7-8 times 9
9-10 times 0
to monitor ECG rhythms for the EASI 12-lead >10 times 81
unit (Table 2). ECG (Table 4). 2. How many times during the None 0
Most of the time course of your shift have you 1-2 times 5
Clinical Performance Outcomes (75%), the deci- accessed EASI 12-lead? 3-4 times 8
5-6 times 8
During the 4 months of data col- sion to use the 7-8 times 5
lection (May through August 2002) 12-lead compo- 9-10 times 0
>10 times 75
to evaluate the new 12-lead system, a nents of the sys-
total of 64 surveys were completed tem was 3. Did the physician request the No 75
EASI 12-lead? Yes 25
by staff members. Survey responses independently
4. Was there any change in the No 31
were based on patients who were made by the interpretation of a rhythm Yes 69
being monitored for a variety of rea- monitoring because of the retrospective
ECG information found in full
sons; the most common problems technician, as disclosure?
were actual or potential myocardial opposed to
5. Has the patient’s treatment No 50
damage (myocardial infarction, chest deciding to use changed in any way because of Yes 50
pain, rule out myocardial infarction, it because a the information found in full
disclosure?
58%), atrial fibrillation (16%), and physician
*Because of rounding, percentages may not all total 100.
ventricular tachycardia (7%; Table 3). requested it.

32 CRITICALCARENURSE Vol 25, No. 5, OCTOBER 2005


Of the surveys completed, 69% of
the staff said that they changed their A

interpretation of ECG rhythms


because of information obtained
after examining the 12-lead stored
information in the system (Table 4,
Figures 4-7). Survey results indicated
that these changes in ECG interpre-
tation led to a different therapeutic
regimen 50% of the time.
Staff described 49 anecdotes of
differences in rhythms observed on
the continuous 2-lead display and
when the full-disclosure 12-lead sys-
tem was accessed (Table 5). Of the
49 anecdotes, 15 described situa-
tions involving ventricular tachycar- B
dias or wide complex tachycardias
for which interpretations of rhythms I
1mV aVR V1 V4

were changed after review of the full-


disclosure function of the 12-lead
EASI. II aVL V5

Implications of Clinical
Performance Outcomes III
aVF V3 V6

Monitoring of Cardiac Rhythm


In previous surveys10 of critical
care clinicians, researchers found that II

the most common ECG lead used for


continuous ECG monitoring in a
single-channel monitor was lead II
Figure 4 Change in interpretation of cardiac rhythm and patient’s treatment after
(74%); the second most commonly stored 12-lead information is examined. A, monitor layout; B, traditional layout.
used was lead MCL1 (18%). Half of the Patient is monitored in leads I and V1 (A). Monitor technician, nurse, and physician
all agree that rhythm is “V-Tach” (A, B). By using full-disclosure tools, EASI 12-
survey respondents used a single- lead electrocardiogram is pulled up to the screen (A). The rhythm shows torsades
channel monitor, 48% used a dual- de pointes. Leads 1 and V1 are the only leads in which torsades de pointes is not
evident on the electrocardiogram (A). The patient’s treatment was changed.
channel monitor, and the remaining
2% used systems that allowed 3 or 4
leads to be displayed simultaneously. tachycardia and supraventricular complex are both better in lead V1 or
The clinical reality that most clini- tachycardia with aberrant conduction lead MCL1 than in lead II; thus leads
cians use lead II for continuous moni- clearly cannot be correctly differen- V1 and MCL1 are better than lead II
toring is in stark contrast to a growing tiated with lead II. Only a third of for differentiating supraventricular
body of ECG monitoring research10-24 these tachycardias were correctly tachycardia with aberrant conduction
that monitoring and diagnostic capa- identified when lead II of the ECG was from ventricular tachycardia. Because
bility are better when multiple leads examined; the best leads for correct treatment of supraventricular tachy-
and ST-segment monitoring are used. diagnosis were leads V1 and MCL1. cardia with aberrancy differs dramat-
As indicated in several studies by The visibility of the P wave and the ically from treatment of ventricular
Drew and colleagues,11-13 ventricular diagnostic morphology of the QRS tachycardia, misdiagnosis of these

CRITICALCARENURSE Vol 25, No. 5, OCTOBER 2005 33


Our experience with the imple-
Medical Record Number TEL 101 mentation of a continuous, derived
HR 82 PVC 3

I 1mV aVR V1 V4
12-lead ECG monitoring system
emphasizes the results of prior
research on the best leads for moni-
toring. The ability to check multiple
II aVL V2 V5
leads when evaluating a dysrhythmia
resulted in numerous anecdotal nota-
tions of changes in diagnoses from
III aVF V3 V6 initial interpretations, and many of
those changes resulted in different
therapies. Many of these changes in
interpretation were related to differ-
II
entiation of ventricular tachycardia
from supraventricular tachycardia
Figure 5 Change in interpretation of cardiac rhythm with no change in patient’s with aberrant conduction.
treatment after stored 12-lead information is examined. Patient monitored in leads
II and V1. V1 shows apparent 6-beat run of ventricular tachycardia. EASI 12-lead
electrocardiogram is put on the screen. V6 shows downward beats of atrioventric- Monitoring of Myocardial Ischemia
ular dissociation. EASI 12-lead electrocardiogram provided critical information
that would have otherwise been missed with the 2-lead monitoring system. Inter- The strongest case for the use of
pretation of the rhythm was changed, but the treatment remained the same. continuous 12-lead ECG monitoring
comes from clinical studies of patients
at risk for myocardial ischemia.17-25
Medical Record Number The value of continuous monitoring
HR 115 PVC 30

I 1mV aVR V1 V4
of ST segments in multiple leads of
the ECG has been shown in a variety
of patients. Continuous 12-lead ST-
segment monitoring was highly sen-
II aVL V2 V5
sitive and specific for identifying
reperfusion and vessel patency after
reperfusion therapy in patients with
III aVF V3 V6 acute coronary syndromes17-20 and
in patients being treated with anti-
ischemic drugs.22
Continuous ST-segment monitor-
II
ing is particularly beneficial in pre-
dicting vessel patency in patients with
Figure 6 Change in evaluation of pacemaker function. Patient was monitored in extensive initial elevation of the ST
leads II and V1, with no evidence of pacemaker malfunction. EASI full-disclosure
format was pulled up on the screen. The 12-lead electrocardiogram shows that the segment, patients who are at highest
pacemaker was undersensing. The patient was sent to the electrophysiology labo- risk for adverse cardiac events. Stud-
ratory for reprogramming of the pacemaker.
ies of patients with acute coronary
syndrome and no persistent elevation
cardiac rhythms could result in dele- with other leads, include atrial flutter, of the ST segment have shown that
terious clinical outcomes.14,15 Correct bundle branch blocks, and heart ST-segment monitoring in all 12 leads
identification of rhythm is critical and blocks.1,16 In addition, many abnormal significantly increases the detection
requires use of leads other than lead II. cardiac rhythms are identified cor- of ischemic episodes compared with
Other arrhythmias that may not rectly more often when more than a ST-segment monitoring in 3 leads.21,23
be easily identified in lead II, compared single lead is used for evaluation.1,11-13 Patients in the emergency department

34 CRITICALCARENURSE Vol 25, No. 5, OCTOBER 2005


rarely were involved in the ischemia
Medical Record Number: 00889402 patterns (V2 to V4).5
HR: 58 PVC:0 SINUS BRADY SVPBs 1 7/15/02 23
NBP: ?/ ?( ?) PULSE: ? An interesting finding of ST-
I aVR V1 V4
segment research is that monitoring
the ECG leads that have the greatest
ST-segment deviation during acute
II aVL V2 V5
myocardial infarction or balloon
inflation during cardiac catheteriza-
tion, also called the “fingerprint”
III aVF V3 V6
leads, is not sufficient to detect
future episodes of ischemia.10,21,22 On
the basis of those studies, it is rec-
II
ommended that monitoring of
patients with the potential for
ischemia include all 12 leads rather
Figure 7 Early warning of myocardial infarction. Patient complains of chest pain.
Monitor technician pulls up EASI 12-lead electrocardiogram (ECG) immediately, than just the fingerprint leads.6
which shows an evolving myocardial infarction. Minutes pass before staff can
get an order for and obtain a standard 12-lead ECG. Physician compares findings
of the 2 methods and finds identical ECG patterns of an evolving myocardial Conventional 12-Lead ECG Versus
infarction. Patient receives treatment for acute myocardial infarction. Derived 12-Lead Monitoring Systems
One aspect of the new monitor-
Recent stud- ing system that initially concerned
5,26,27
Table 5 Categorization of 49 anecdotal notations made by ies of clinicians was how well the EASI
monitoring technicians on the questionnaires
patients at high derived 12-lead monitoring system
risk for myocar- approximated the standard 12-lead
Categories of anecdotal notes No. of notes
dial ischemia ECG. Although research has shown
Ventricular tachycardia missed on the 2-lead
continuous monitoring system 2 after elective that use of the EASI 12-lead system
Ventricular tachycardia misinterpreted on the surgery indi- for continuous monitoring results in
2-lead continuous monitoring system 5 cated a 6% to the clinical detection of more ischemic
Ventricular tachycardia confirmed with EASI 12-lead 11% incidence episodes and the capture of more
system when conflicting interpretations obtained of myocardial arrhythmias (because data are col-
in each lead of the continuous monitoring system 1
ischemia after lected continuously rather than as a
Wide complex tachycardia of undetermined etiology
with the 2-lead continuous monitoring system vascular sur- snapshot in time),12,28-36 the conven-
but interpretation confirmed on EASI 12-lead gery. Of interest tional 12-lead ECG remains the refer-
electrocardiogram (ECG) 9 in these studies ence standard as a “diagnostic” tool.
Nonventricular tachycardia or wide-complex is the inability Having a continuous ECG system
tachycardia rhythm interpretation changed
after review of EASI 12-lead ECG 15 to detect these with all 12 leads stored in the com-
Capture of short-duration dysrhythmias with EASI episodes of puter memory was also better than
12-lead when unable to obtain with standard ischemia with standard 12-lead ECGs for verifying
12-lead ECG 3 traditional sin- changes in cardiac rhythm or myocar-
Findings with standard 12-lead ECG equivalent to gle- or dual-lead dial oxygenation. Cardiac rhythms
findings with EASI 12-lead ECG 8
monitoring. or ischemic episodes that were of
Miscellaneous comments 6
Even 2-lead ST- short duration could always be
segment moni- recalled by using the full-disclosure
with suspected acute coronary syn- toring had a low sensitivity (<5%) for system of the EASI system. Such
drome who have transient ischemic detecting myocardial ischemia post- recall not only allowed accurate doc-
episodes also benefit from continu- operatively, because the leads selected umentation of the changes in rhythm
ous 12-lead ECG monitoring.24,25 for standard monitoring (II and V5) or myocardial oxygenation but also

CRITICALCARENURSE Vol 25, No. 5, OCTOBER 2005 35


allowed the staff to treat the rhythm marker for early unfavorable outcomes in in acute myocardial infarction during
patients with unstable angina. N Engl J Med. recombinant tissue-type plasminogen acti-
or perfusion problem more quickly 1986;314:1214-1219. vator therapy: The TEAHAT Study Group.
3. Drew BJ, Pelter MM, Adams MG, Wung SF, Am J Cardiol. 1991;67:343-349.
because the clinical information was Chou TM, Wolfe CL. 12-lead ST-segment 20. Klootwijk P, Lannger A, Meij S, et al. Non-
readily available. monitoring vs single-lead maximum ST-seg- invasive prediction of reperfusion and
ment monitoring for detecting ongoing coronary artery patency by continuous ST-
ischemia in patients with unstable coronary segment monitoring in the GUSTO-I trial
syndromes. Am J Crit Care. 1998;7:355-363. [published correction appears in Eur Heart J.
Summary 4. Pelter M, Adams M, Drew B. Association of 1996;17:1451]. Eur Heart J. 1996;17:689-698.
Although the conventional 12- transient myocardial ischemia with adverse 21. Klootwijk P, Meij S, von Es G, et al. Com-
in-hospital outcomes for angina patients parison of usefulness of computer assisted
lead ECG is routinely used for diag- treated in a telemetry unit or coronary care continuous 48-h 3-lead with 12-lead ECG
nostic purposes, it is not practical unit. Am J Crit Care. 2002;11:318-324. ischaemia monitoring for detection and
5. Martinez E, Kim L, Faraday N, et al. Sensi- quantitation of ischaemia in patients with
for continuous monitoring. However, tivity of routine intensive care unit surveil- unstable angina. Eur Heart J. 1997;18:931-940.
lance for detecting myocardial ischemia. 22. Klootwijk P, Meij S, Melkert R, Lenderink T,
the standard telemetry placement Crit Care Med. 2003;31:2302-2308. Simoons ML. Reduction of recurrent ischemia
produces only some of the ECG 6. Drew B, Kruchoff M. Multilead ST-segment with abiciximab during continuous ECG-
monitoring in patients with acute coronary ischemia monitoring in patients with unsta-
information that is needed, highlight- syndromes: a consensus statement for ble angina refractory to standard treatment
healthcare professionals. Am J Crit Care. (CAPTURE). Circulation. 1998;98:1358-1364.
ing only 1 or 2 views of the heart. 1999;8:372-386. 23. Veldkamp RF, Green CL, Wilkins ML, et al.
Technological advancements now 7. Dower G, Machado H, Osborne J. On deriv- Comparison of continuous ST-segment
ing the electrocardiogram from vectorcar- recovery analysis with methods using static
enable us to obtain all 12 views of diographic leads. Clin Cardiol. 1980;3:87-95. electrocardiograms for noninvasive patency
8. Dower G, Yakush A, Nazzal S, Jtuzy R, Ruiz assessment during acute myocardial infarc-
the heart on a continuous basis by C. Deriving the 12-lead electrocardiogram tion: Thrombolysis and Angioplasty in
using just 5 electrodes. By shifting from four (EASI) electrodes. J Electrocardiol. Myocardial Infarction (TAMI) 7 Study
1988;21(suppl):S182-S187. Group. Am J Cardiol. 1994;73:1069-1074.
the axis in which the 5-lead ECG is 9. Frank E. An accurate, clinically practical sys- 24. Norgaard BL, Andersen K, Dellborg M,
tem for special vectorcardiography. Circula- Abrahamsson P, Ravkilde J, Thygesen K.
obtained, we can have important 12- tion. 1956;13:737-746. Admission risk assessment by cardiac tro-
lead information at our fingertips on 10. Drew B, Ide B, Sparacino P. Accuracy of ponin T in unstable coronary artery disease:
bedside electrocardiographic monitoring: a additional prognostic information from
a continuous basis. The clinical advan- report on current practices of critical care continuous ST segment monitoring. TRIM
nurses. Heart Lung. 1991;20:597-609. study group. Thrombin Inhibition in
tages of knowing that valuable 12-lead 11. Drew B, Scheinman M. Value of electrocar- Myocardial Ischemia. J Am Coll Cardiol.
information can be obtained at any diographic leads MCL1, MCL6, and other 1999;33:1519-1527.
selected leads in the diagnosis of wide QRS 25. Gomez MA, Anderson JL, Karagounis LA,
given time or at a designated time complex tachycardia. J Am Coll Cardiol. Muhlestein JB, Mooers FB. An emergency
during any shift is important for cor- 1991;18:1025-1033. department-based protocol for rapidly ruling
12. Drew B, Scheinman M, Evans G. Compari- out myocardial ischemia reduces hospital
rect diagnosis and treatment of ECG son of a vectorcardiographically derived 12- time and expense: results of a randomized
lead electrocardiogram with the conventional study (ROMIO). J Am Coll Cardiol. 1996;
abnormalities. electrocardiogram during wide QRS com- 28:25-33.
In this study, the ability to gener- plex tachycardia and its potential applica- 26. Landesberg G, Mosseri M, Zahger D, et al.
tion for continuous bedside monitoring. Am Myocardial infarction after vascular sur-
ate both real-time and retrospective J Cardiol. 1992;69:612-618. gery: the role of prolonged, stress-induced,
13. Drew B, Scheinman M. ECG criteria to dis- ST depression-type ischemia. J Am Coll
12-lead ECGs resulted in important tinguish between aberrantly conducted Cardiol. 2001;37:1839-1845.
differences in clinical management. supraventricular tachycardia and ventricu- 27. Ouyang P, Gerstenblith G, Furman WR,
lar tachycardia: practical aspects for the Golueke PJ, Gottlieb SO. Frequency and sig-
These results suggest that use of con- immediate care setting. Pacing Clin Electro- nificance of early postoperative silent
physiol. 1995;18:2194-2208. myocardial ischemia in patients having
tinuous 12-lead monitoring of the 14. Dancy M, Camm J, Ward D. Misdiagnosis peripheral vascular disease. Am J Cardiol.
ST segment in all patients in the car- of chronic recurrent ventricular tachycar- 1989;64:1113-1116.
dia. Lancet. 1985;8450:320-323. 28. Drew B, Koops R, Adams M. Dower G. The
diac step-down area would have a 15. Stewart R, Bardy G, Green H. Wide complex derived electrocardiogram: comparison
tachycardia: misdiagnosis and outcome after with the standard electrocardiogram during
clinical benefit. emergent therapy. Ann Intern Med. 1986; myocardial ischemia and its potential appli-
104:766-771. cation for continuous ST-segment monitor-
Acknowledgments 16. Paul S, Dangerfield L. Care of the Patient With ing. J Electrocardiol. 1994;27:249-255.
We acknowledge the staff in the cardiac telemetry an Arrhythmia. Aliso Viejo, Calif: American 29. Drew B, Adams M, Pelter M, Wung S, Cald-
units at North Shore University Hospital, Man- Association of Critical-Care Nurses; 2001. well M. Comparison of standard and derived
hasset, NY. Without their assistance and support, 17. Hohnloser S, Zabel M, Kasper W, Meinertz 12-lead electrocardiograms for diagnosis of
this project would not have been completed. T, Just H. Assessment of coronary artery coronary angioplasty-induced myocardial
Philips Medical Systems (Andover, Mass) pro- patency after thrombolytic therapy: accu- ischemia. Am J Cardiol. 1997;79:639-644.
vided partial funding for this project. rate prediction utilizing the combined 30. Drew B, Pelter M, Wung S, et al. Accuracy
analysis of three noninvasive markers. J Am of the EASI 12-lead electrocardiogram com-
Coll Cardiol. 1991;18:44-49. pared to the standard 12-lead electrocardio-
References 18. Krucoff MW, Green CE, Satler LF, et al. gram for diagnosing multiple cardiac
1. Drew BJ. Bedside electrocardiographic Noninvasive detection of coronary artery abnormalities. J Electrocardiol. 1999;32:38-47.
monitoring: state of the art for the 1990s. patency using continuous ST-segment mon- 31. Horacek B, Warren J, Stovicek P, Feldman C.
Heart Lung. 1991;20:610-623. itoring. Am J Cardiol. 1986;57:916-922. Diagnostic accuracy of derived versus stan-
2. Gottlieb SD, Wisfeldt ML, Ouyang P, Mel- 19. Dellborg M, Hiha M, Swedberg K. Dynamic dard 12-lead electrocardiograms. J Electro-
lits ED, Gerstenblith G. Silent ischemia as a QRS-complex and ST-segment monitoring cardiol. 2000;33:155-160.

36 CRITICALCARENURSE Vol 25, No. 5, OCTOBER 2005


32. Denes P. The importance of derived 12-lead
electrocardiography in the interpretation of
arrhythmias detected by Holter recording.
Am Heart J. 1992;124:905-908.
33. Denes P. Morphologic characteristics of
nonsustained ventricular tachycardia
detected during Holter monitoring associ-
ated with atherosclerotic heart disease. Am J
Cardiol. 1992;69:612-618.
34. Drew BJ, Adams MG, Pelter MM, Wung SF.
ST-segment monitoring with a derived 12-
lead electrocardiogram is superior to rou-
tine cardiac care unit monitoring. Am J Crit
Care. 1996;5:198-206.
35. Drew BJ, Pelter MM, Brodnick DE, Yadav
AV, Dempel D, Adams MG. Comparison of
a new reduced lead set ECG with the stan-
dard ECG for diagnosing cardiac arrhyth-
mias and myocardial ischemia. J
Electrocardiol. 2002;35(suppl):13-21.
36. Rautaharju PM, Zhou SH, Hancock EW, et
al. Comparability of 12-lead ECGs derived
from EASI leads with standard 12-lead
ECGs in the classification of acute myocar-
dial ischemia and old myocardial infarction.
J Electrocardiol. 2002;35(suppl):35-39.

CRITICALCARENURSE Vol 25, No. 5, OCTOBER 2005 37


CE Test Test ID C0552: Clinical Usefulness of the EASI 12-Lead Continuous ECG Monitoring System
Learning objectives: 1. Identify the differences between standard and continuous 12-lead systems 2. Identify 3 clinical conditions that are appropriate
for continuous 12-lead monitoring 3. Describe the benefits of examining 12-lead electrocardiograms in the clinical setting

1. Which of the following best describes the standard 12-lead 6. The EASI 12-lead system can display how many continuous derived leads?
electrocardiogram (ECG) system? a. 12 of 12 leads
a. A system that requires 12 electrodes placed in various locations of the chest and b. 6 of 12 leads
extremities c. 3 of 12 leads
b. A system that requires 10 electrodes that are strategically placed on the chest and d. 5 of 12 leads
extremities
c. A system that requires only 8 electrodes with 2 grounding electrodes placed on the 7. According to the studies by Drew and colleagues, how many
chest and extremities tachycardias were correctly identif ied using lead II?
d. A system that requires 10 electrodes that are strategically placed on the extremities a. One half
alone for a true 12-lead tracing b. One third
c. Three quarters
2. Which of the following best describes the 5-lead telemetry monitoring setup? d. Two thirds
a. Four leads are attached to the patient’s torso, and the last lead acts as the precordial
V lead. 8. Which are the 2 best leads for differentiating supraventricular
b. Three leads are attached to the patient’s torso, and the last 2 leads are the V and tachycardia with aberrant conduction from ventricular tachycardia?
MCL2 leads. a. V1 and MCL2
c. All 5 leads are attached to the patient’s torso, and the telemetry box is programmed b. V1 and MCL1
for which axis is wanted to be viewed. c. V2 and MCL6
d. Leads I, II, and III are attached to the patient’s torso and the last 2 leads are attached d. V3 and V4
to the upper extremities.
9. Which 2 arrhythmias may not be easily identif ied in lead II?
3. Which of the following best identif ies recent research and national a. Atrial fibrillation and fascicular blocks
guidelines on monitoring for myocardial ischemia? b. Atrial tachycardia and atrial flutter
a. Use the diagnostic 12-lead ECG monitoring devices to help detect ischemia in c. Junctional tachycardia and heart blocks
patients with chest pain. d. Bundle branch blocks and heart blocks
b. Focus on the emergency department as the primary area to help identify and
interpret cardiac dysrhythmias. 10. Which of the following best describes the benef it of continuous
c. Use standard 12-lead ECG monitoring in all busy critical care areas for the indemnifi- ST-segment monitoring?
cation and interpretation of cardiac rhythms. a. Predicting where ischemic changes will occur in the heart with elevation of the
d. Use continuous 12-lead ST-segment monitoring for at-risk patients, because silent T wave
ischemia is common and may not always be detected in the same leads. b. Identifying vessel occlusion in acute myocardial infarction
c. Predicting vessel patency in patients with extensive initial elevation of the ST
4. Which of the following best describes the difference between standard and segment
derived ECG monitoring? d. Predicting mortality of patients with significant ST-segment and T-wave abnor-
a. Only 8 views are available with standard ECG monitoring, and 10 continuous views malities
are available with derived ECG monitoring.
b. Only 10 views are available continuously with standard ECG monitoring, and all 12 11. What is the incidence of myocardial infarction after vascular surgery?
views are available with derived ECG monitoring. a. 6% to 11%
c. All 12 views are available periodically with standard ECG monitoring, and all 12 b. 10% 20%
views are available continuously with derived ECG monitoring. c. 5% to 12%
d. All 12 views of the heart are available continuously with standard ECG monitoring, d. 6% to 15%
and only 10 views are available with derived ECG monitoring.
12. Which of the following recommendations may be benef icial for
5. What is the percentage of clinically silent myocardial ischemia detected with continuous 12-lead ECG monitoring?
ECG monitoring? a. Unstable angina and no ST-segment elevation
a. 70% to 90% b. Acute myocardial infarction without ST-segment elevation and cardiac surgery
b. 60% to 95% c. Acute myocardial infarction with ST-segment elevation and noncardiac surgery
c. 50% to 80% d. Chest pain that prompts a visit to an urgent care center and cardiac
d. 50% to 70% catheterization

Test answers: Mark only one box for your answer to each question. You may photocopy this form.
1. K a 2. K a 3. K a 4. K a 5. K a 6. K a 7. K a 8. K a 9. K a 10. K a 11. K a 12. K a
Kb Kb Kb Kb Kb Kb Kb Kb Kb Kb Kb Kb
Kc Kc Kc Kc Kc Kc Kc Kc Kc Kc Kc Kc
Kd Kd Kd Kd Kd Kd Kd Kd Kd Kd Kd Kd
Test ID: C0552 Form expires: October 1, 2007 Contact hours: 2.5 Fee: $13 Passing score: 9 correct (75%) Category: A Test writer: Todd M. Grivetti, RN, BSN, CCRN
Program evaluation Name Member #
Yes No
Objective 1 was met K K Address
Objective 2 was met K K City State ZIP
Objective 3 was met K K
Content was relevant to my Country Phone
Mail this entire page to: nursing practice K K
My expectations were met K K E-mail
AACN This method of CE is effective RN Lic. 1/St RN Lic. 2/St
101 Columbia for this content K K
The level of difficulty of this test was: Payment by: K Visa K M/C K AMEX K Discover K Check
Aliso Viejo, CA 92656 K easy K medium K difficult
To complete this program, Card # Expiration Date
(800) 899-2226
it took me hours/minutes.
Signature
The American Association of Critical-Care Nurses is accredited as a provider of continuing nursing education by the American Nurses Credentialing Center’s Commission on Accreditation.

38 CRITICALCARENURSE Vol 25, No. 5, OCTOBER 2005

You might also like