CGD50/500
Calibration Instruction
Consilium Building Safety AB
Phone: +46 31 28 59 40
Fax: +46 31 28 85 95
E-mail: sales.cbs@consilium.se
service.cbs@consilium.se
The contents of this document are subject to revision without notice
due to continued progress in methodology, design and
manufacturing. Consilium Building Safety AB assumes no legal
responsibility for any error or damage resulting from the usage of
this document.
April 2019
Document no. 5103028_Salwico ST400 H2S_Calibration Instruction_I0_EN_2019_C
Part no. 5103028
1.1 Preliminaries
This instructions applies to the following products:
Figure 1. Salwico ST400 H2S, 0-30 ppm 4-20 mA. Part no. 4462
DANGER!
H2S – Hydrogen Sulphide – is toxic and dangerous also in rather low concentrations.
H2S is a sticky gas and smells very distinctly from rotten eggs.
Consilium Replacement article number for ST400 H2S detectors is no. 004462-S.
A new sensor is supplied without the metal housing and must be fitted in the
housing. This applies to both H2S and O2 sensors of ST400 type.
Calibration equipment
Normal measuring range is 0-30 ppm, our standard calibration bottle no.
5201005-00A contains 20 ppm of gas in air.
For SW2020 a mix-bottle is often used, with H2S 20 ppm as well as N2 99% and
C3H8 50% LEL, article no. 5201010-01A.
Test points (A) for the multi-meter as well as “SPAN” (B) and “ZERO” (C)
potentiometers are indicated below.
CAUTION!
Do not adjust the other potentiometers.
A Test points
B SPAN potentiometer
C ZERO potentiometer
1.2 Prerequisites
1 Use the arrow keys on the control panel to
manoeuvre through the different menus.
Press the right arrow key to access the
menu tree.
7 NOTE!
For the following steps in the
zero calibration instruction
proceed to the Salwico ST400
detector.
To verify which type is in use, make a span calibration as below and verify the
display reading in the central cabinet.
Span calibration
www.consilium.se