U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
2
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Introduction
This guide details installation, commissioning and maintenance of U Series dental comprehensive Treatment
unit. Before using the product, please read this manual carefully and keep it properly for future review. Any
installation, commissioning and maintenance operations must be performed in accordance with related
procedure.
Any warning and caution stated herein shall be read and complied with to avoid damage to equipment or
injury to operator or patient.
For any malfunction occurring during use, consult this guide for troubleshooting or contact your local dealer
or our headquarters.
● Safety Considerations
During using the equipment, be sure to comply with the following safety precautions to mitigate risks of
equipment damage, fire, electric shock and personal injury.
* Make sure to shut off the main power, water and air supplies before any maintenance and cleaning
operation on the equipment.
* Do not place the equipment on unstable ground or cart or rack to prevent fall of it and thereby damage.
* Do not locate the equipment beside any heat radiator or heater.
* Do not operate the equipment in any location exposed to any inflammable or an aesthetic gas.
* Do not place or hang any heavy object on the equipment.
* Do not use any power cable not complying with equipment requirements to prevent performance
degradation or risk of fire or electric shock.
* Do not stretch any object or splash any liquid into the equipment through the enclosure or any opening
therein to avoid risk of electric shock or fire.
* Improper assembly may cause electric shock during operation of the equipment.
* In case of damage to or wearing of any power cable, plug or connection, or entry of water into the
equipment, immediately shut off the power supply and contact the professional service personnel for
assistance.
* Adjustment can be conducted only with regard to the options as stated in the instructions. Improper
adjustment to any other options may cause damage to the equipment.
* Do not use any defective plug, and insecure connection of plug may cause electric shock or fire.
* Any plug and socket must be grounded properly to prevent electric shock or damage to the equipment.
● About This Manual
This manual presents instructions on installation and maintenance of U200 and U100 Dental
Comprehensive Treatment Unit, which are valid as of the date of publishing this manual. We reserve the
right to change equipment specifications or design without prior notification. Without Siger's prior
written approval, no modification, spread, reproduction or sale of this manual shall be done in any way.
Siger Medical reserves the right to interpret or alter any information in this document.
Caution: The warranty shall become invalid if any of the following requirements is not met:
1) Comply with any instruction in this guide.
2) The equipment must be installed, trial run and maintained in accordance with the instructions in this
guide.
3) A three-core power cable with cross section of core ≧ 1 sq. mm shall be provided for indoor
3
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
5) Any spare part for the equipment must be supplied by Siger Medical so as to ensure proper performance
and safety of the equipment.
4
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Contents
SECTION 1 - INSTALLATION AND COMMISSIONING...................................................................................6
SECTION 3 - TROUBLESHOOTING................................................................................................................... 34
5
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Note: This section present important parameters of the product, which shall be well understood by the user before
1 2 345 6 7 8 9 10 11
12 13 14 15 16 1718 19 20 21
Name of Key Component
11 Treatment Platform
6
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
U100 Pull-Up “S” Treatment Platform U100Hung “X” Treatment U200 Pull-Up “S” Treatment U200Hung “X” Treatment
It consists of: side case, treatment platform, assistant section, operation lamp, operating chair, pedal and therapist chair.
Optional components for treatment platform include: ejection gun, high-speed worm handheld sprayer, low-speed
handheld sprayer and scaler.
Optional components for assistant section include: strong/weak-suction saliva ejectors, ejection gun and light curing unit.
The system options include: wet/dry negative pressure inspiration, water treatment and endoscope systems.
Name plate: A name plate containing product parameters is attached at the inner side of the side case, including: product
model and name, power voltage, frequency and electrical safety classification, input power and operation mode, product
license number, product standard number, product registration number, product SN, factory delivery date and
manufacturer.
Operating Conditions:
7
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
1.1.4 Dental chair full range of movement (Max. length: 2,100mm; Max. width: 1,780mm)
Overall Dimensions of Treatment Unit: 190cm (L) * 85cn (W) * 180cm (H) Weight: 205 kg
1) Max. height from cushion surface to the ground ≥820mm; 2) Min. height from cushion surface to the ground ≤420mm;
5) Max. cushion retroversion angle ≥10°; 6) Instrument panel vertical movement range: 480 mm;
Caution: The equipment must be installed by our professionals or any personnel authorized by us.
This equipment shall be installed at a clean, dry, cool and well-ventilated location, and sufficient space shall be
reserved around the unit to facilitate access during routine maintenance.
The equipment shall not be installed or operated at any location
a) exposed to humidity, heavy dust, poor ventilation or direct sun; or
b) subject to substantial temperature or humidity change, such as place near air conditioner or heater.
Caution: Improper installation may cause damage to the equipment!
1.2.2 Requirements for Installation Field
The equipment shall be located based on general layout of the treatment room, lighting condition and operation
convenience, and should be installed on flat, horizontal and solid ground at a clean, dry, cool and well-ventilated
location.
Standard Layout for S Series Comprehensive Treatment Platform: The layout is for single-person operation and
excludes any auxiliary equipment. A is 500 mm and B is 800 mm. For double-person operation, A should be 1,000
mm to accommodate additional auxiliary equipment or larger based on actual requirement.
)
管
(
污
进
排
管)
气
进水
管
)
电源)
8
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
9
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
with wire diameter ≧ 1.5 mm². The underground hose line shall lain at the same elevation as that of the floor ground as possible,
or not above 100 mm higher than the floor. Connection between PC and endoscope (if any) shall be lain through the
underground hose line. The VGA cable and AV video cable shall be connected to the computer.
Air pressure ≥ 500-800Kpa, flow rate ≥50L/min Water pressure: 200~400Kpa, liquid flow rate ≥10 L/min;
Remove the front and bottom plate enclosures of the chair, and remove the four holts locking the chair to the bottom plate,
and mount the bottom of the chair onto the position as predetermined.
Mount foam material for chair backrest.
Insert the headrest into the backrest in place (as shown in Figure 1), and adjust looseness of headrest puller with the
jackscrew (as shown in Figure 2).
Fix the bottom plates of side case and operating chair onto the ground with four 8mm expansion bolts through the fixing
holes (as shown in Figure 3).
10
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Fix side case to the swiveling arm with four locking screws and then adjust the four jackscrews to level it and then lock it
with 8*20 screw, as shown in the figure. .
Insert lamp pillar into the hole in side case column, and thread the wire connecting to lamp arm into the pillar and then
connect and fix operation lamp head; then connect lamp arm wire plug to the connector on circuit board in side case in
place; Operation lamp connection terminal is as shown below.
U100 Oral Cavity Lamp Wiring Diagram U200 Oral Cavity Lamp Wiring Diagram
. Installation of Hose line and Cables on Ground Case
Connect G1/4*Ф8 straight-through connector to air inlet ball valve, and connect one end of blue Ф8PU hose to the
straight-through connector, and the other end to the main air inlet port on operating chair.
Connect G1/4*Ф8 straight-through connector to water inlet ball valve, and connect one end of blue Ф8PU hose to the
straight-through connector, and the other end to the main water inlet port on operating chair.
Insert operating chair drain hose and saliva ejector air outlet hose into the Ф50PVC drain hose in the ground hole.
Connect power cable and grounding wire securely, and connect all wire plugs into the corresponding sockets on circuit
board in ground case and arrange the hose line in good order.
1.5.1The packaging contains six sub-packages: operating chair, treatment platform, operation lamp, side case accessories,
lamp arm (rod)
and foam material (a set of five components).
Operating chair components include: operating chair bracket, side case connecting arm, balancing arm (front arm
connecting rod) assembly, curve plate, physician chair accessories, chair bottom plates (big and small) and front chair
enclosure, as shown below.
11
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Treatment platform includes: treatment platform assembly, treatment platform silica gel cushion, tray, tray silica gel pad,
tray bracket assembly, as shown below.
First unpack the operating chair, leave the accessories aside, and then handle the chair to the mounting position and fix and
level it as shown below (Figure 3).
Mount the foam material, curve plate, backrest enclosure, headrest and side case connecting arm as shown below, and
adjust looseness of headrest puller with the jackscrew (as shown in Figure 2).
①
②
⑤
1 Assemble foam material seat cushion to chair bracket with 10*35mm countersunk head socket screws.
12
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
2 Assembly of curve plate and rocker: Align holes in curve plate with those in rocker and lock them with M10
3 Assembly of side case connecting arm: Align holes in connecting arm with those in chair bracket and lock them
4 Assembly of curve plate and backrest foam material: Lock them with M10 stainless steel socket screws.
5 Mounting of U backrest enclosure: Align holes in enclosure with those in foam material and push them in with
Unpack and handle side case assembly to the connecting arm, and fix it with four 8*25 countersunk socket screws, and
level and lock it with screws (as shown in the figure).
a) Mount balancing arm to side case column and lock it with two 8*15 socket screws (as shown below).
1.6.4 U100 Side Case Hose line and Chair Communication and Power Cables Connection Instructions
Connect the five hose in balancing arm assembly respectively to the hose in side case assembly in the same color, and
connect J3/J7 on communication cable (SG117) respectively to the ports on side case control board.
Notes:
The five hose include: 8*5 blue air hose, 4*1 orange water hose, 4*1 blue HP air hose
4*1 light green LP air hose, 4*1 yellow air hose (for control of water)
Connect 8*5 blue air hose, 4*1 yellow air hose (for control of water) and 4*1 light green LP air hose respectively to the
13
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Connect 4*1 orange water hose and 4*1 blue HP air hose respectively to the hoses in the same color on distributor in side case,
Connect transformer power cable, chair power and communication cables and main operation key cable to side case
1.6.5 U200 Side Case Hose line and Chair Communication and Power Cables Connection Instructions
b) As shown below, connect the hose line from the balancing swiveling arm to the connector in side case.
14
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
J30 Chair
Communication Treatment
Cable
Platform
Communic
Chair Power and Communication Cables and Treatment Platform Power and
Communication Cables Connection Diagram
Notes:
1) In installation of lamp arm, be sure not to neglect installation of lamp rod ring.
2) In installation of lamp rod, tighten the rod by rotating it clockwise or counterclockwise till into the locking groove in
place.
2) 2)locking
Lamp rod 灯杆卡槽位groove
灯杆、灯臂安装图
Lamp Rod and Arm Installation Diagram
15
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
1) Unpack operation lamp, align lamp with two terminals of lamp arm wire, and use 5*12 socket head screws to fix lamp on
the arm (the screws must be tightened to prevent looseness).
2) Installation of operation lamp handle: Hold lamp in right hand, hold left handle in left hand with thumb pressing the
handle button down, insert it into lamp arm shaft locking groove, and then release handle button. Install the right handle in
the same way as with the left handle.
Removal of operation lamp handle: Hold lamp in right hand, hold left handle in left hand with thumb pressing the handle
button down, and pull the handle out with force. Remove the right handle in the same way as with the left handle.
1) Unpack operation lamp, align lamp with two terminals of lamp arm wire, and use 5*12 socket head screws to fix lamp on
the arm (the screws must be tightened to prevent looseness).
2) Installation of operation lamp handle - Hold lamp in right hand, hold left handle in left hand with thumb pressing the
handle button down, insert it into lamp arm shaft locking groove, and then release handle button. Install the right handle in
the same way as with the left handle.
3) Removal of operation lamp handle - Hold lamp in right hand, hold left handle in left hand with thumb pressing the handle
button down, and pull the handle out with force. Remove the right handle in the same way as with the left handle.
16
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Unpack spittoon gargling basin and accessories and install the basin, spittoon filter screen, U spittoon cover, outlet nozzle
1. U Spittoon Base
a) Unpack the treatment platform and fix it onto front arm connecting rod with four 6*20 socket head screws; Unpack tray
assembly and fix it onto front arm connecting rod with four 6*15 countersunk head screws, as shown below.
Notes: Ports in corresponding colors are provided in treatment platform for connection of hose line from side case.
Connection of hose line and communication and power cables is described below.
2
3
4
5
7
8
9
Hose line Description: 1) Treatment platform communication cable 2) Reserved AC 24V power cable 3)
17
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Ground cable
4) 8*5 Blue air hose 5) 4*1 Blue HP air hose
6) 4*1 Light green LP air hose 7) 4*1 orange water hose
8) 4*1 yellow air hose (for control of water) 9) 4*1 white air-braking hose
b) Connect six water/air hoses from balancing arm assembly respectively to the hoses in the same color in treatment platform,
as shown below.
c) Connect the communication cable to the port on the connecting board in treatment platform, and then mount the platform
cover with fix it with 4*10 pointed-end self-threading screws.
Treatment Platform Communication Cable & Main Operation Communication Cable Connection Diagram
a) Unpack the treatment platform and fix it onto front arm connecting rod with four socket head screws; unpack tray
assembly and fix it onto front arm connecting rod with four 6*15 cross head screws, as shown below.
Notes: Ports in corresponding colors are provided in treatment platform for connection of hose line from side case.
Connection of hose line and communication and power cables is described below.
1) 6*4 blue air hose
b) The hose line and communication and power cables as numbered as 1~8 above are connected to the ports in treatment
Combination valve
connecting 4*0.75
Figure Figure
The first RT
The first LT one in one in right
solenoid valve set solenoid valve
connecting to a short set connecting
blue 4*0.75 air hose Treatment Treatment Platform
to a long
(HP hose) on Platform Communication Cable
transparent blue
balancing arm hose on main Power Cable Socket J35
operation panel Socket J34
Figure 3 Figure
Blue 4*0.75
air-braking hose
connecting main
air inlet hose
Figure 5
After connection of treatment platform hose line and power and communication cables in place, fix the panel covers with
19
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
If scaler components have been mounted on the treatment platform, connect the wires with label on scaler control box, as shown
below.
1: Yellow
2: Green
3: White
4: Blue
5: Black
6: Brown
7: Red
8-12:Empty
喷枪 Ejection Gun
洁牙机 Scaler
分道阀 Separating Valve
六角压水阀 Hexagonal Pressure Valve
四孔三联膜片阀 4-Hole Triad Diaphragm Valve
洁牙机控制板 Scaler Control Board
电位器 Potentiometer
气控电阀 Air-Control Electrovalve
三通 T Joint
If no scaler component is mounted on treatment platform, the additional parts for installation of scalers (hexagonal valve, flow
regulating valve, water hose and air-control electrovalve) shall be mounted as shown in the figure.
Mounting potentiometer on
Air-Control treatment platform Air-Control Electrovalve (Constant Open) on Treatment
Hexagonal Valve
20
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Spare parts required for pre-installation of scaler: scaler control board, instrument panel control board, small-flow regulating valve,
Installation Procedure - Connect white water hose to small-flow regulating valve > connect it to scaler control board > connect
black power cable on scaler control board to instrument panel control board > output voltage: 24V AC > Connect white conductor
on scaler Control Board to switch J51JYJ—SW on instrument panel control board. This installation procedure is applicable to
Korean scaler and Woodpecker scaler. For installation of any other scaler, refer to the schematic diagram.
洁牙机 Scaler
小流量调节阀 Small-Flow Regulating Valve
微型电磁阀 Mini Solenoid Valve
三通 T Joint
六角压水阀 Hexagonal Pressure Valve
洁牙机控制板 Scaler Control Board
洁牙面控制盒 Scaler Panel Control Box
电位器 Potentiometer
治疗台控制板 Treatment Platform Control Board
Scaler-Instrument Panel Connection Instructions:
Power Supply connecting to J33 (AC24) port on Treatment platform Mounting potentiometer on treatment platform Locate
Fix scaler main unit to the locking hole in treatment platform, and mount power potentiometer to the locking hole in treatment
platform.
Installation of Scaler Hose line: Cut out the sheath to the proper length, snore then install it by following the installation procedure
for handheld sprayer unit.
After fixing in place, connect water hose to the flow regulating valve and connect output to the main unit.
Notes: For installation of scaler for electric control bottom-mounted model, first remove the GK1 short circuiting pin before
installation.
21
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Mount the light curing unit onto assistant section and cut it off to the proper length and thread the connecting line through the wire
hole in assistant section metal board and fix it in place! Connect red and green wires to the AC 24V port on side case control board
through assistant section arm. It is notable that, in case of DC powered light curing unit, a rectifier box needs to be mounted in side
U200 Light Curing Unit Installation Instructions (as shown at the right)
1) Fix light curing unit hose line onto assistant section welding bottom plate.
2) Connect light curing unit power cable to AC 24V port on assistant section control board.
① ④
②
③
(1) Lock the lamp arm with M5 internal hex screws. (2) Lock the display with four 4*35 cross head screws.
(3) Fix endoscope control box with two 4*15 cross head screws. (4) Assemble handle hanger.
22
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
U100 Electric Motor (with Carbon Brush) Installation & Connection Diagram
23
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Connect double-throw switch to Pins #3 and #4 of J2 terminal of electric motor control board.
Instructions on Installation of Bien Air Brushless Electric Motor with LED Light Guide
Installation & Connection Diagram for U100 with Bien Air Brushless Electric Motor with LED
Light Guide
24
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Connect the double-throw change-over switch to Pins #3 and #5 of J2 terminal of electric motor control board.
For setting jumpers for side case and treatment platform control board during installation of electric motor, refer to the
instructions under 3.10 and 3.11.
In installation of electric motor, the shift lever pedal shall be changed as below.
Mount an additional 30 pedal magnetic control board and direction control board to magnetic control wire SG041.
Add two small magnets to the toothed bar, take care to keep the opposite sides of the magnet in south and north poles.
For more details see instruction under 3.12.
25
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
U200 Electric Motor (with Carbon Brush) Installation & Connection Diagram
小流量调节阀 Small-Flow Regulating Valve
26
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Instructions on Installation of Bien Air Brushless Electric Motor with LED Light Guide
Installation & Connection Diagram for U200 with Bien Air Brushless Electric Motor with LED Light Guide
27
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
(3) Connect white air hose to T joint assembly, with the other two ends of the T joint assembly respectively to Ф3.2 nozzle
of diaphragm valve and air-control electrovalve.
(4) ~ (8) Connect blue, black, red, orange and brown conductors to J1 terminal of electric motor control board.
Connect Pins #9 and #10 of terminal J2 of electric motor to the terminals of 30 instrument panel control board. Output
voltage: DC3.2V.
Connect Pins #1 and #2 of terminal J2 of electric motor to the terminals of 30 instrument panel control board. Output
voltage: AC24V.
Connect Pin #1 of air-control electrovalve to the Pin #6 of J2 terminal of electric motor control board.
Connect Pin #5 of J2 terminal of electric motor control board and Pin #2 of air-control electrovalve to J36 terminal of
instrument panel control board.
Connect double-throw change-over switch to Pins #3 and #5 of J2 terminal of electric motor control board.
28
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Installation & Connection Diagram for U200 with TKD Brushless Electric Motor with Led Light Guide
29
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
a) Connect G1/4*ф8 straight-through connector to air inlet ball valve, and connect one end of blue ф8PU hose to the
straight-through connector, and the other end to the main air inlet port on operating chair.
b) Connect G1/4*Ф8 straight-through connector to water inlet ball valve, and connect one end of blue Ф8PU hose to the
straight-through connector, and the other end to the main water inlet port on operating chair.
c) Insert operating chair drain hose and Saliva Ejector air outlet hose into the Ф50PVC drain hose in the ground hole.
d) Connect power cable and grounding wire securely, and connect all wire plugs into the corresponding sockets on circuit
board in ground case and arrange the hose line in good order.
30
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Adjusting Screw
3
5 0
5 0
0 K
p
k a
P
a
High-Speed Handheld sprayer air pressure adjustment: Adjust the air pressure to 250 Kpa with the knob on the bottom of
instrument panel.
High-Speed Handheld sprayer air pressure adjustment (S60/S30): Adjust low-speed handheld sprayer pressure to
250Kpa-300Kpa by using the knob on the bottom of instrument panel.
31
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Unit
The filter shall be replaced by following the procedure as below: Open Side Case, remove the filter, remove the filter
enclosure by rotating it counterclockwise, and then take off the sealing ring and filter element and clean them before
re-mounting; ensure good airtightness of the filter.
a) Distilled water system consists of distilled water cylinder and selector switch at the front of side case. Water to supply all
instruments on the instrument panel is from the distilled water in the cylinder or filtered tap water, so the cylinder must be
b) The cylinder shall be refilled with distilled water in the following way:
Close the air switch on distilled water cylinder till the air in the cylinder is discharged off.
Remove the cylinder by rotating it clockwise and inject distilled water to the maximum level.
Re-mount the cylinder by rotating it counterclockwise, and then open the air switch and change-over switch on it.
Caution:
The distilled water cylinder must not be opened until the air in it has been discharged off. As water/air hoses are connected
to the cylinder cover, perform the operations by holding the cover with left hand and rotating the cylinder body with right hand
instead of directly rotating the cover. It is recommended to use distilled water instead of tap water during treatment to avoid
effect of any impurity in tap water on service life of the hose line and instruments.
32
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
When the cylinder is empty, open the air switch on it to discharge any humidity and dirt in the hose line with the air in the
cylinder.
Close the air switch and remove the distilled water cylinder and mount a red cylinder containing disinfectant.
Open the air switch, and manipulate the instruments one by one till disinfectant is sprayed out and then keep the unit idle
Close the air switch, remove the disinfectant cylinder, re-mount the distilled water cylinder and open the switch, and clean
the internal hose by using the aforesaid method to finish cleaning and disinfection.
Handheld sprayer, ejection gun and strong/weak suction heads shall be sterilized at a temperature of 121℃ (1 Bar) or
134℃ (2 Bar). For more details on disinfection of handheld sprayer, see the operating instructions for the sprayer.
Operating chair requires no special maintenance, but it is recommended to clean it after treatment of each patient.
Oral cavity lamp handle and any external surface of equipment, especially those in contact with the physician, nurse or
Caution: Do not use any abrasive or acid agent. Operate the treatment chair to check whether its all functions are available
before delivery.
33
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Section 3 - Troubleshooting
3.1 U100 Water/AIR Supply Schematic Diagram
34
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
35
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
喷枪 Ejection Gun
小流量调节阀 Small-Flow Regulating Valve
洁牙机 Scaler
Mini Solenoid Valve
微型电磁阀
Water/Air Distributor
水气分配器 Air-Braking Valve
气刹阀 Air-Braking Solenoid Valve
气刹电磁阀 Operating Air Pressure Adjustment
工作气压调节 Main Air Solenoid Valve
总气电磁阀 Main Water Inlet
总进水 5-Way Joint and T Joint Combination
五通一三通组合 Water Filter
Air Pressure Regulator
水过滤器
T Joint
气调压器 Main Air Inlet
三通 Air Pressure Regulator
总进气 Drain Hose
气调压器 Air Cylinder
排污管道 Door Air Switch
气缸 Water/Air Distributor
门气开关
水气分配器
电马达 Electric Motor
高速手机 High-Speed Handheld Sprayer
RT 水源转换开关 RT Water Source Change-Over Switch
200KPa Simple Reducing Valve
200KPa 简易减压阀
Switch with Air Return
带回气开关 Distilled Water Bottle
蒸馏水瓶 Solenoid Valve
电磁阀 Proportional Solenoid Valve
比例电磁阀 T Joint
三通 Strong Suction Handle
强吸手柄 Inspiration Generator
吸吸发生器 Weak Suction Handle
Air Return Cylinder
弱吸手柄
Air Pressure Meter
回气瓶 Separating Valve
气压表 Voltmeter
分道阀 Water Cup
电压表 Spittoon
水杯 Regulating Valve
痰盂 Water Cup
调节阀 Regulating Valve
Water Heater
水杯
调节阀
热水器
The control system of U100 Dental Comprehensive Treatment Unit consists of the following components:
Side case components include: 100CX-02-F5 side case control board, 100ZS-03-F5 assistant section control board
Treatment platform includes: 100-ZL-02/F5 treatment platform control board
Chair components include: ZY-01-F6 chair control board, 90JT-02-F5 pedal direction control board;
36
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
红外感应头 IR Sensor
LED 手术灯主控制板 LED Operation Lamp Main Control Board
手术灯 Operation Lamp
手术灯输出跳线 Operation Lamp Output Jumper
电源输出 Power Supply Output
加热 Heating
加热检测 Heating Detection
Assistant Section 24V
助手 24V
侧箱控制板 Side Case Control Board
冲盂 Spittoon Rinsing
总进气 Main Air Inlet
Cup Water
杯水
Distilled Water Pressure
蒸馏水压力 Main Water Inlet
总进水 Distilled Water
蒸馏水 Chair Communication
座椅通信 Treatment Platform Communication
治疗台通信 Solenoid Valve
电磁阀 Dirt Separation
污物分离 Endoscope
Communication
内窥镜
J6 Power Supply
通讯 Chair Ascending
J6 电源 Chair Descending
座椅上升 Power Supply
座椅下降 Chair Control Board
电源 Backrest Protection Switch
座椅控制板 Chair Protection Switch
椅背保护开关 Chair Motor
Backrest Motor
座椅保护开关
座椅电机
椅背电机
主操作触摸板 Main Operation Touch Panel
12864 液晶模块 12864 LCD Module
气刹板 Air-Braking Board
Main Operation Touch Control Board
主操作触摸控制板
Film Observation Lamp
观片灯 Upper Drawbar Switch
上拉杆开关 Power Supply Input
电源输入 Air-Braking Solenoid Valve
气刹电磁阀 Single-Spray Solenoid Valve
单喷气电磁阀 Big Film Observation Lamp
大观片灯 DSP Board
DSP 板 Electric Motor Control Board
Treatment Platform Control Board
电马达控制板
High-Speed Handheld Sprayer Control Board
治疗台控制板
高速手机控制板
Six-Purpose Gun
六用枪 Scaler
洁牙机 Water Valve
38
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
水阀 Electric Motor
电马达 Air Valve
气阀 High-Speed Handheld Sprayer
Proportional Valve
高速手机
Scaler Connection Board
比例阀
洁牙机连接板
侧箱板说明
1、 出厂默认不短接 JP03,将 JP01 左边两脚短接上。
2、 手术灯输出路线默认短接 2-5 3-6 4-1.
J11 输出:接 LED 手术灯时,A 为 OV、C 为 12V
接卤素灯时,A 为 0V、B 为 10V、C 为 12V、D 为 17V。
3、 LED3,LD4 常亮,LD1 闪烁,LD2 连接治疗台板正常为闪烁。
Transformer Input
Assistant Section
Communication
Main Operation Communication Cable Pedal Communication Cable
Cable
Operatio
n Lamp Heating Detection Saliva Rinsing
39
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Water Heater Power Supply Output: AC 24V Heating Detection Port: The port is
J9 J10 disconnected when water is heated to 40°C and
voltage output when heater is operating.
is connected during heating.
Spittoon rinsing water solenoid valve port: DC
Gargling water solenoid valve port: DC 24V when
J13 J12
in operation 24V when in operation.
Assistant Section Board Main Control Board
J3 Treatment Platform Communication Port J32
Communication Port
Pedal direction control board communication
J30 Chair Control Board Communication Port J31
port
3.4 ZY-01-F6 Chair Control Board Interface Function Description (applicable to all models)
Chair Motor
Descending J2 is backrest motor
Manual Button
40
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
41
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
The control system of U200 dental comprehensive treatment unit consists of the following components:
Side case components include: 30CX-01-F5 side case control board, 90ZS-01-F5 assistant section control board,
U500ZS-GK-F0 assistant section strong suction optic control board, weak suction optic control board and 900ZS-02-F5
assistant section touch control board.
Treatment platform includes: 30ZL-01-F5 instrument panel control board, 30ZL-03-F5 main operation touch control board.
42
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Chair components include: ZY-01-F6 chair control board, 90JT-02-F5 pedal direction control board.
43
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
内窥镜 Communication
通讯 J6 Power Supply
J6 电源 Chair Ascending
Chair Descending
座椅上升
Power Supply
座椅下降 Chair Control Board
电源 Backrest Protection Switch
座椅控制板 Chair Protection Switch
椅背保护开关 Chair Motor
座椅保护开关 Backrest Motor
座椅电机
椅背电机
主操作触摸板 Main Operation Touch Panel
12864 液晶模块 12864 LCD Module
气刹板 Air-Braking Board
Main Operation Touch Control Board
主操作触摸控制板
Film Observation Lamp
观片灯 Upper Drawbar Switch
上拉杆开关 Power Supply Input
电源输入 Air-Braking Solenoid Valve
气刹电磁阀 Single-Spray Solenoid Valve
单喷气电磁阀 Big Film Observation Lamp
大观片灯 DSP Board
DSP 板 Electric Motor Control Board
Treatment Platform Control Board
电马达控制板
High-Speed Handheld Sprayer Control Board
治疗台控制板
高速手机控制板
Six-Purpose Gun
六用枪 Scaler
洁牙机 Water Valve
水阀 Electric Motor
电马达 Air Valve
气阀 High-Speed Handheld Sprayer
Proportional Valve
高速手机
Scaler Connection Board
比例阀
洁牙机连接板
侧箱板说明
1、 出厂默认不短接 JP03,将 JP01 左边两脚短接上。
2、 手术灯输出路线默认短接 2-5 3-6 4-1.
J11 输出:接 LED 手术灯时,A 为 OV、C 为 12V
接卤素灯时,A 为 0V、B 为 10V、C 为 12V、D 为 17V。
3、 LED3,LD4 常亮,LD1 闪烁,LD2 连接治疗台板正常为闪烁。
44
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Assistant
Section 24V
Output
Transformer Input
Chair Power
Supply 24V
Output
Distilled
Water
Pressure
Heating
Heating Saliva Cup Distilled Weak Strong Dirt
Detection Rinsing Water Water Suction Suction Separation
45
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Handheld
Electric Sprayer Handheld Handheld Handheld
Reserved Instrument Sprayer
Motor Pull-Up Sprayer Pull-Up Sprayer Pull-Up
24V Panel Pull-Up
Switch Switch 1 Switch 3 Switch 4
Power Switch 2
Bottom-mounted
optical control
Power Supply Port
Communicat
Bottom-mounted
optical control
Communication
Port
Main Key
Operations
Function Description: Main operation panel key scanning and water/air control
46
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Function Interf
Interface Interface Description Interface Description Function Description
Description ace
Treatment platform Connect to J10 on side case
Reserved Reserved AC 24V J34
J33 power supply interface board (provided by side
case board)
Connect to Pins 4 and 5 on
Communication Connect to J29 on Motor power control
J35 J36 motor control board
interface side case board interface
J41 Optic control power DC3.3V (for
J37GK1
supply interface bottom-mounted model)
Optic control signal Input (for bottom-mounted
J38GK1 J42
Pull shift lever interface model)
Pull-up instrument
down to close the KEY-CONNECT connects
panel dynamic
switch, reset the to main operation touch
instrument switch Main key operation
J39GK1 shift lever to open J43 control board (shared by
interface signal interface
the switch S30) mainly for control
keyboard IO signal.
Film observation lamp
J40GK1 J44 DC12V
power supply port
Interface for control
J48 handheld sprayer water DC 24V when in operation
solenoid valve
Interface for control of
Triad solenoid valve DC 24V when in air sprayer and
J45 J49 DC 24V when in operation
group interface operation single-spray solenoid
J46
valve
J47
Interface for control of
J50 scaler water solenoid DC 24V when in operation
valve
DC 3.3V at the
DC 3.3V at the port when
J51 Scaler switch wire port port when relay J52 Light Guide 2 interface
relay operates.
operates.
DC 3.3V at the
DC 3.3V at the port when
J53 Light Guide 1 interface port when relay J54 Light Guide 3 interface
relay operates.
operates.
3.8.3 U200 Main Operation Touch Control Board (as shown at the right) J102 Socket
J43 KEY-CONNECT.
Interface Function Description: KEY-CONNECT connects to S60 main S30/60 Main Operation Touch Control
Board
operation touch control board (shared by S30) mainly for scanning of keyboard
IO signals.
47
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
U200 Assistant Section Main Control Board U200 Assistant Section Touch Control Board
Adjustment
Measure voltage at ADJ test point
when shift lever is not operated; make
rough adjustment to potentiometer if
when voltage is at much deviation to
-0.3V.
Measure voltage at test point when shift lever is not operated; make no adjustment if voltage is -0.3V, or make adjustment
to potentiometer when voltage is at much deviation to -0.3V (final adjustment to -0.3V).
48
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
49
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
of 90JT-02-F6 or V1000JT-01-F5.
Shift Lever
Treadle Pedal
Pedal
Whether 30JT-01-F5 is installed or not, this switch must be connected to J71 socket of 90JT-02-F6 or
V1000JT-01-F5.
Function Description: For key switch board and connection to side case board
3.14 Upper/Lower Motor and Air-Operated Spring Repair and Replacement Instructions
Replacement of upper motor: Remove treatment chair pad foam material > shrink upper motor range to the minimum >
use clamp ring nippers to remove four clamp rings and spring at both ends of motor > use screwdriver to remove the motor
shaft > replace upper motor.
50
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
Replacement of upper motor: Put an object with 30KG bearing capacity under the bottom of side case to support it >
expand lower motor range to the maximum > use cross-head screwdriver to remove the upper parallel cover > use clamp
ring nippers to remove the clamp ring off motor shaft > remove motor shaft > replace motor.
51
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
a) U200 Inspections:
1) Check whether fuse SSD F2 is broken, Operation Lamp Power Cable is in good connection to J12 interface - for 2P
wire connection of halogen/LED lamps, the voltage values for Interfaces A, B, C and D are AC17V-12V-10V-0V.
2) If fuse SSD F2 is in good condition, but no voltage is output at J12, measure the output voltage at transformer; if there
is no output voltage at the transformer port, replace the transformer; if the output voltage at transformer port is normal,
replace the side case control board and then perform inspection again.
52
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
b) U100 Inspections:
1) Check whether fuse SSD F2 is broken, Operation Lamp Power Cable is in good connection to J11 interface - for 2P
wire connection of halogen/LED lamps, the voltage values for Interfaces A, B, C and D are AC17V-12V-10V-0V.
2) If fuse SSD F2 is in good condition, but no voltage is output at J11, measure the output voltage at transformer; if there
is no output voltage at the transformer port, replace the transformer; if the output voltage at transformer port is normal,
replace the side case control board and then perform inspection again.
Inspections:
1) If water heater fails, check whether there is AC 23V voltage output at J13; if yes, the heater may be broken and replace
it; if no, the side case control board or the heater is damaged; if heater cannot be disconnected, and J14 is always closed,
the heater may be broken and replace it.
2) If there is no voltage output at J9, check whether fuse F1 is broken; if yes, replace it with a 10 fuse. (U100)
If there is no voltage output at J13, check whether fuse F4 is broken; if yes, replace it with a 6A fuse. (U200)
Failure 5: Handheld sprayer operates normally but no water or air is sprayed out.
Inspections:
1) First check whether the blower operates currently in dry grinding mode where no water or air is sprayed by operating
the change-over switch on it. For more details see the operating instructions.
2) Check the water/air hose line is in good connection in instrument panel. If the hose line is in good condition, then
Check whether there is DC 24V output at water/air circuit solenoid valves; if yes, the solenoid valve may be broken,
and replace solenoid valve; if not, measure voltage output at reserved port J33, if there is a voltage output, replace
treatment platform control board, if not, inspect the power supply in side case (U200).
3) Check whether the handheld sprayer hanging rack valve (hung type) on treatment platform and constant-on valve with
peak can reset properly, if not or in case of leakage, replace the relevant valve. (U100)
4) Foot-operated micro switch: Check whether the foot-operated micro switch resets normally; if not, replace the micro
switch!
5) Replace side case control board.
53
U Series Dental Comprehensive Treatment Unit - Installation & Maintenance Manual
2) If the measurement is normal, then measure and check whether the voltage on treatment platform control board
(30ZL-01-F5) meets the value as specified below:
3) If the measurement is normal, then measure and check whether the voltage on motor control board (90ZL-04-D3) meets
the value as specified below:
4) If the voltage value does not vary depending on pedal shift lever angle in one of the three steps as stated above, the
corresponding circuit board has been broken, and replace it.
54