U100/U200

Product Standard Code: YZB/粤0207-2006


Product Registration No.: 粤食药监械(准)字2008第2550198号

TECHNICAL MANUAL
FOR
SIGER DENTAL UNIT

ZHUHAI SIGER MEDICAL EQUIPMENT CO., LTD


Manufacturer: Zhuhai Siger Medical Equipment Co., Ltd
Manufacture & Registration Address: Building 2 No. 1 Chuangxin Yi Road, Tangjiawan Town, Zhuhai City,
Guangdong, China
Tel: 0756–3881018 Fax: 0756–3881028
Website: Http:.//www.siger.cn
E-mail: info@siger.cn
Post Code: 519000 Version: A1
Technical Manual for U Series Dental Unit

目 录
Chapter 1 Installation of U Series Dental Unit ........................................................................................... 2
1.1 Function and Technical Parameters....................................................................................................... 2
1.2 Installation Condition and Place Requirements .................................................................................. 4
1.3 Handling and Unpacking of the Unit .................................................................................................... 6
1.4 Installation of U200............................................................................................................................... 7
1.5 Installation of U100............................................................................................................................... 7
1.5.1 Installation of patient’s chair.............................................................................................................. 7
1.5.2 Installation of hydro-group................................................................................................................. 8
1.5.3 Installation of instrument table assembly (see above)........................................................................ 9
1.6 Installation of operating light ................................................................................................................ 9
1.7 Installation of water, air and power supply ........................................................................................... 9
1.8 Installation of Scaler ............................................................................................................................. 9
1.8.1 Installation of scaler on U100 ............................................................................................................ 9
1.8.2 Installation of scaler on U200 .......................................................................................................... 10
1.8.3 Connection between Scaler and Instrument Table Control Board.....................................................11
1.9 Installation of Curing Light..................................................................................................................11
1.9.1 Installation of curing light on U100 ..................................................................................................11
1.9.2 Installation of curing light on U200 ................................................................................................. 12
1.10 Installation of Intra-oral Camera (see the following illustration)...................................................... 12
1.11 Installation of Micromotor on U200 ................................................................................................. 12
Chapter 2 Adjustment of U Series Dental Unit ......................................................................................... 14
2.1 Adjustment .......................................................................................................................................... 14
2.2 Safety Function and Pipeline Disinfection.......................................................................................... 15
Chapter 3 Maintenance of U Series Dental Unit....................................................................................... 16
3.1 Replacement of Motors ....................................................................................................................... 16
3.2 Replacement of the Air Spring ............................................................................................................ 17
3.3 Replacement of the Upholstery ........................................................................................................... 17
3.4 Replacement of the Parts of LED Light .............................................................................................. 17
3.5 U100 Water-air Schematic Diagram ................................................................................................... 18
3.6 U200 Water-air Schematic Diagram ................................................................................................... 19
3.7 Fault Analysis of Air and Water System.............................................................................................. 20
3.8 Constitution of U100 Electric Control System.................................................................................... 21
3.12 Constitution of U200 Electric Control System.................................................................................. 23
3.13 U200 Hydro-group Control Board .................................................................................................... 25
3.14 Chair Control Board (the same with U100) ...................................................................................... 26
3.15 U200 Instrument Table Control Board.............................................................................................. 26
3.16 U200 Dentist’s Touch Control Board................................................................................................ 27
3.17 Foot Pedal Magnetic Control Board (optional for micromotor) (see 30JT-01/D6)........................... 27
3.18 Installation of Jumpers on the Hydro-group Control Board 30CX-01-F1 ........................................ 28
3.19 Installation, Adjustment and Fault Analysis of Foot Pedal ............................................................... 29
3.20 U200 Assistant’s Control Board........................................................................................................ 30
3.21 Fault Analysis of U200 Control Board.............................................................................................. 30

1
Technical Manual for U Series Dental Unit

Chapter 1 Installation of U Series Dental Unit


1.1 Function and Technical Parameters

NOTE: This chapter will provide the vital technical parameters of the unit. Please read this chapter

before any installation, adjustment or maintenance is performed.

1.1.1 Description of main parts

12345 6 7 8 9 10 11

12 13 14 15 16 1718 19 20 21

Names of main parts

1 curing light (optional) 12 headrest


2 HV suction tube 13 foot pedal
3 assistant’s control panel 14 patient’s chair
4 saliva ejector tube 15 left armrest
5 assistant’s syringe 16 dentist’s syringe
6 spittoon 17 scaler (optional)
7 display (optional) 18 motor
8 hydro-group 19 turbine
9 operating light 20 dentist’s control panel
10 instrument table arm 21 tray
11 instrument table

2
Technical Manual for U Series Dental Unit

1.1.2 Instrument table of swing type and hanging type

² Nameplate: the nameplate is located on the right side of the hydro-group. It contains the following data:

product name and type, power voltage, frequency, classification of electric safety, power consumption,

mode of operation, registration number, implementation of the standards, serial number, date of

manufacture and information of manufacturer.

² Technical parameters: The device complies with the following standards: YZB/Guangdong 0207-2008, EC

93/42. And it is classified as Class II a.

The technical parameters are shown as follows


safety
I main fuse tube 10A 250V
classification
rated illuminance
protection type Type B 8000-20000Lx
of operating light
LED operating
operation mode intermittent operation AC12V 8VA
light
ordinary equipment
protection degree without liquid input power 900VA
protection
voltage of power power supply
Ac.220±22V 50±1Hz
supply frequency
air supply
≥550 kPa air supply flow ≥50L/min
pressure
water supply
200 kPa~400 kPa water supply flow ≥10L/min
pressure
heater AC24V 120AV

² Ambient conditions:

Temperature: 5℃~40℃; relative humidity: ≤80%; atmospheric pressure: 86kPa ~106kPa;

Power supply: AC 220V±22V, 50Hz±1Hz; water supply pressure: 200kPa~400kPa; air supply pressure:

≥500kPa。

The dental unit should be fixed on flat ground, and the skewness should be <0.5° after installation.

3
Technical Manual for U Series Dental Unit

1.2 Installation Condition and Place Requirements


1.2.1 Installation condition

² NOTE: The installation must be performed by the professional personnel from Siger or trained and

authorized by Siger.

² The unit must be placed in clean, dry, airy, shady and cool place. There should be enough room around

the unit to facilitate routine maintenance.

² Don’t install the unit in places where:

A. It is humid, dusty, airless, or in direct sunshine;

B. Changes of high temperature or high humidity often occur, such as places near air-conditioner or

heaters, etc.

NOTE: Improper installation may lead to damages to the unit!

1.2.2 Place requirements for installation

² Choose the installation location according to the specific conditions of the dental treatment room such as

the general layout, lighting and easy access to instruments, etc. The unit should be installed in clean, dry,

airy and cool place. And the floor on which the dental chair is located should be flat, level, and solid.

² Standard layout: The layout below shows the size of single-person operation. It does not include size of

other auxiliary equipments in the treatment room. A is 500mm, and B is 800mm. In case of two-people

operation, A is 1000MM. With additional auxiliary equipments, the size should be increased accordingly.





管)

进水

电源)

4
Technical Manual for U Series Dental Unit

² Size chart of the unit

² Layout requirement of holes under the floor box

5
Technical Manual for U Series Dental Unit

U Series Dental Unit Floor-box Hole Position

Waste discharge
pipe

Air inlet pipe


Water inlet pipe

Power supply

In the ground holes, the drainage pipe should be Ф50MM PVC pipe, water supply pipe should be

Ф15MM, and air supply pipe should be Ф15MM. High-pressure plastic composite pipes can be used for water

supply pipe and air supply pipe. Valves should be connected to the ports. The power supply is 220v with earth

wire, the diameter of the power supply wire should be ≧1.5MM². It is preferred that the pipelines are not

above the ground, once it is, the vertical interval should be no more than 100mm. If there is intra-oral camera

connected to a computer, it must be added to the ground holes so that the VGA cable and the AV cable can be

connected to the computer.


voltage of power supply: 220±10% VAC frequency of power supply: 50±1Hz

air pressure≥500-800Kpa, air flow ≥ 50L/min

water pressure: 200—400Kpa, liquid flow ≥10 L/min

1.3 Handling and Unpacking of the Unit


² Carry the unit to the installation place in accordance with the identification

on the packing container.

² The product is packed in one container. Remove the outer screws to

demount the packings (U100 is packed in separate containers).

² Take out the packing list and check whether the unit and spare parts are

intact. If there is any problem, please contact Siger as soon as possible.

6
Technical Manual for U Series Dental Unit

1.4 Installation of U200


² Remove the front cover of the dental chair and the cover of the bottom plate, and then

remove the four screws fixed on the bottom plate. Put the bottom part of the seat that

can bear load on the predetermined place while maintaining a balanced state.

² Installation of backrest cover: insert the four buckling parts on the cover into the

corresponding holes on the upholstery scaleboard.

² Insert the headrest into the corresponding position on the backrest as shown in the

picture.

² Move out the hydro-group from the fixing plate to the connecting arm while retaining the balance.

² Insert the light post into corresponding holes of the hydro-group upright;

penetrate the wire connected to the light arm into the light post; then connect

and fix the head of the operating light; and then connect and fix the wire plug

of the light arm with corresponding connectors on the circuit board in the

hydro-group; the connection port for the operating light is shown in the
Connection port
picture:

² For the installation of the hydro-group, please see 1.5.2.

1.5 Installation of U100


1.5.1 Installation of patient’s chair

² First unpack the container and remove the accessories, then move the chair to the installation position and fix it as is

shown below, finally adjust its level.

² The installation of upholstery, bending plate, backrest cover, headrest and hydro-groupconnecting arm is

as follows:



① Fix the upholstery and the chair frame with black countersunk inner hexagon screw sized

7
Technical Manual for U Series Dental Unit

10*35mm;

② Installation of bending plate and swing link: make the hole on the bending plate aligning at that on

the swing link, then lock them together with M10 stainless steel inner hexagon screw;

③ Installation of the hydro-group connecting arm: align the connecting arm and the chair frame, lock

them together with black countersunk inner hexagon screw sized 10*35mm;

④ Installation of bending plate and backrest upholstery: lock with M10 stainless steel inner hexagon

screw;

⑤ Installation of backrest cover: insert the four buckling parts on the cover into the corresponding

holes on the upholstery scaleboard.

1.5.2 Installation of hydro-group

² Move the hydro-group assembly from the container to the connecting arm, adjust

the level and lock the screw.

² Installation of the balance arm: install to the hydro-group, the air and water tubes

go through the hole and are fixed with four screws.

² Connection of air and water tubes: the white water tube is connected to the water connector on the 3-5

way link, the blue air tube is connected to the air connector on the 3-5 way link, the other three

connectors are connected to the corresponding tubes and colors in the foot pedal:

White water Blue air tube

tube

the upper connector of the hexagon


² Connection of electric wires: the
valve is connected to orange, while the
the blue 6*4 is
AC24V power wire is connected to J7 lower one to white water tube
connected to
on the hydro-group control board, the the membrane
valve
button wires of dentist’s control panel
the blue 4*1 is
are connected to J3, the power input connected to
3-way link the blue 4*2.5 is
connection and chair communications
connected to the air
to the hydro-group are as follows: distributor
the hard tube is connected to the
release switch on the air brake

8
Technical Manual for U Series Dental Unit

1.5.3 Installation of instrument table assembly (see above).

1.6 Installation of operating light


² Insert the light post into corresponding holes of the hydro-group

upright; penetrate the wire connected to the light arm into the light
U100 接手
post; then connect and fix the head of the operating light; and then

connect and fix the wire plug of the light arm with corresponding

connectors on the circuit board in the hydro-group; the connection

port for the operating light is shown in the picture below:

Operating light
connector

1.7 Installation of water, air and power supply


² Connect a G1/4*Ф8 straight coupling with the air supply ball valve. Insert one end of a blue Ф8PA pipe into the

straight coupling, and insert the other end into the whole air supply port.

² Connect a G1/4*Ф8 straight coupling with the water supply ball valve. Insert one end of a blue Ф8PA pipe into the

straight coupling, and insert the other end into the whole water supply port.

² Insert the drainage pipe and vent pipe of saliva ejector into the Ф50PVC drainage pipe ground hole.

² Connect the power cable and earth wire. Be sure to have a good connection. Plug all the connectors in the

corresponding sockets on the circuit board inside the ground box. And pack up the lines.

1.8 Installation of Scaler


1.8.1 Installation of scaler on U100

² If the instrument table is already installed with preassembled accessories for scaler, install according to the

identification on the scaler control box, the power connection is a follows:

9
Technical Manual for U Series Dental Unit

Power supply
AC 24V
AC 24V

Solenoid valve
Foot
S/W

Potentiometer: 10k

1. Yellow 2. Green 3. White 4.Blue 5.Black 6. Brown 7. Red 8-12. Null

² If the instrument table is not pre-installed with scaler accessories, install those accessories first (hexagon valve,

flow adjusting valve, water tube, pneumatic control switch).

Pneumatic
Potentiometer connected to pneumatic switch
control
installed on on the instrument table board
switch
the (constantly open)

instrument
table board
connected
to
reserved
24V on the
Hexagon
instrumen
valve
t table
board Tapping position for the
potentiometer

1.8.2 Installation of scaler on U200

² Preassembled accessories: scaler control board, 30ZL instrument table control board, small flow adjusting valv

potentiometer, scaler tubing, and solenoid valve.

² Installation steps: connect the white water tube ① to the small flow adjusting valve => connect ②③ to the

scaler control board => connect the black power cable on the scaler control board to the instrument table control

board => output voltage: AC24V => connect the white wire on the scaler control board to the switch

10
Technical Manual for U Series Dental Unit

J51JYJ—SW on the instrument table control board => this installation applies to scalers from South Korea and

Woodpecker; for other scalers, please refer to the schematic diagram.

①white water
tube Ф1.5/2.5
②white wire
③red wire

1.8.3 Connection between Scaler and Instrument Table Control Board

power supply <=> port AC24 on the instrument table

foot pedal switch <=> pneumatic switch

potentiometer <=> instrument table scaler board<=> corresponding position

² Fix the scaler board on the fixing plate inside the instrument table, and install the potentiometer under the

instrument table.

² Installation of scaler tubing: cut the surface cover to a proper length and refer to the installation of handpiece

tubings.

² After fixation, connect the water tube to the flow adjusting valve, and the output wire to the scaler board.

² Note: for unit of hanging type, remove the GK1 short-circuit bolt before the installation of scaler.

1.9 Installation of Curing Light


1.9.1 Installation of curing light on U100

² The curing light is to be installed on the assistant’s unit. Cut the tubing to proper length, put it through the black

hole into the assistant’s unit and fix it. The red and green wires go through the assistant’s unit arm and are

connected to AC24V on the hydro-group board. Note: if the curing light is supported by DC power supply, a

rectifier box should be installed in the hydro-group before the installation of the curing light.

11
Technical Manual for U Series Dental Unit

1.9.2 Installation of curing light on U200

² Installation: fix the curing light tubings on the assistant's assembly welding plate => connect the power cable to

the assistant's control board => this installation applies to curing light with a input voltage of AC24V.

1.10 Installation of Intra-oral Camera (see the following illustration)

① Fasten the M5M hexagon screw onto the light arm

② Use four cross recess head screws sized 4*35 to lock the

display

③ Use two cross recess head screws sized 4*15 to hang


the camera control box
④ Handle holder

1.11 Installation of Micromotor on U200

12
Technical Manual for U Series Dental Unit

1. air-control electric valve


J36
2. micromotor control
①white air tube
board 3. instrument table
with A Ф1.5/2.5
control board 4. 4-hole ②green water

3-group membrane valve 5. tube Ф1.5/2.5


③white air tube
3-way link assembly 6.DT
Ф1.5/2.5
switch 7.small flow
④red wire
adjusting valve ⑤black wire
⑥brown wire
8.micromotor tubing
⑦blue wire
U200 Micromotor Installation and
Wiring Diagram

² The white air tube ① with A is connected to the port Ф2.6 on the membrane valve.

² The green water tube ② is connected to the small flow adjusting valve.

² The white air tube ③ is connected to the 3-way link assembly; the other two ends of the assembly are

respectively connected to the port Ф3.2 on the membrane valve and the air-control electric valve on the

micromotor. See the S30/S60 Micromotor Installation and Wiring Diagram above.

² The red and black wires ④⑤ are connected to the DT switch. The brown and blue wires ⑥⑦ are connected to

the terminal on the instrument table control board of U200. The output voltage is DC3.2V.

² The terminal 1 and 2 on the micromotor board are connected correspondingly to the terminals on the instrument

table control board of U200. The output voltage is AC24V.

² A on the air-control valve is connected to the terminal 6 on the micromotor control board.

² The terminal 5 on the micromotor control board and B on the air-control electric valve are connected to J36 on

the instrument table control board of U200

² The DT switch is connected to the terminal 3 and 4 on the micromotor control board.

13
Technical Manual for U Series Dental Unit

² Change of the foot pedal.

² Add a magnetic control board of S90 foot pedal on the original foot pedal.

² Add two small magnets on the rack. Note: the opposite sides of the magnets should be south and north poles.

² Add directional board to the magnetic control wire SG041.

Chapter 2 Adjustment of U Series Dental Unit


2.1 Adjustment
² Fix the bottom plate on the ground with four 8mm anchors (pic 3)

Fixing Fixing
hole hole

Fixing Fixing
hole hole

pic 3 pic 2

² Tightness adjustment of the headrest pulling plate: adjust through the jackscrew (pic 2).

² Level adjustment of the chair: Place a spirit level on the instrument table in lateral and longitudinal

direction. Then adjust the nine inner hexagon screws mounted on the bottom plate of the chaie to keep

the instrument table level with misalignment within 3 degrees’ deflection. (pic 3)

14
Technical Manual for U Series Dental Unit

pic 4 pic 5 pic 6

² Instrument table arm force adjustment: Turn the adjusting screw clockwise to increase the force, and turn it

anticlockwise to decrease the force (pic 4).

² As shown in the pictures above, turn the input air adjusting valve to let the two pressure
gauges display 0.55Mkpa and 0.3Mpa (pic 5 and 6).
² Screw the hydro-group onto the fixing plate with four fixing screws, adjust the level through four

jackscrews and lock it with screw sized 8*20 (pic 7).

pic 7 pic 8 pic 9

² Air pressure adjustment of high speed turbine: Turn the knob on the bottom of the instrument table to

adjust the air pressure to 250Kpa.

² Air pressure adjustment of low speed air motor: Turn the knob on the bottom of the instrument table to

adjust the air pressure to 250Kpa-300Kpa.

² Tightness adjustment of the light arm: open the adjusting cover, use inner hexagon adjusting nut to

choose the best working tightness (pic 10).

pic 10 pic 11

² Fixation of swing-type instrument table (pic 11)

2.2 Safety Function and Pipeline Disinfection


2.2.1 Operation of distilled water system

15
Technical Manual for U Series Dental Unit

² Note: the air pressure inside the distilled water bottle should not exceed 300Kpa.

2.2.2 Safety function checking

² The chair or backrest would stop moving down (U200) in the presence of obstacles and the chair would

move up slightly to allow the user to remove the obstruction.

² Any one of the dental chair control buttons can stop the chair from moving.

² When the dynamic instrument is in use, the dental chair will be under safety protection and none

movement is available.

² An interlocking device prevents the instruments from working at the same time. The first instrument

removed is operative. Those removed afterwards are inactivated by the interlocking device (U200).

Chapter 3 Maintenance of U Series Dental Unit


3.1 Replacement of Motors
² Replacement of the upper motor: unscrew the four cross recess head screws sized 6*25 from the

upholstery with a cross screwdriver => remove the foamed cushion => use a snap ring plier to remove

the four snap rings and springs at the two ends of the motor => pry out the motor shaft with a

screwdriver => replace the motor.

² Maintenance of the lower motor: put the hydro-group on an upholder which can bear a weight of 30kg

=> use a cross screwdriver to demount the upper parallel cover => remove the snap ring on the motor

with a snap ring plier => take out the motor => replace the motor.

16
Technical Manual for U Series Dental Unit

² Replacement of the lower motor of U200: put the hydro-group on an upholder which can bear a weight

of 30kg => use a cross screwdriver to demount the upper parallel cover => take out

the locking jackscrew with an inner haxegon wrench => tak out the motor shaft with

a screwdriver => replace the motor.

3.2 Replacement of the Air Spring


² Steps: remove the four screws from the upholstery => remove the upper parallel cover => demount the

clamp springs at the ends of the stationary shaft of the air spring => pry out the the stationary shaft with a

screwdriver and replace teh air spring.

3.3 Replacement of the Upholstery


² Replacement of the headrest upholstery: first pry out a space between the cover and

the headrest upholstery, then grab the cover with right hand, and tear apart the

upholstery with left hand.

² Replacement of the backrest upholstery: first pry out a space between the cover and

the backrest upholstery, then grab the cover with right hand, and tear apart the

upholstery with left hand.

² Replacement of the chair upholstery: loosen the four screws on the upholstery to replace.

3.4 Replacement of the Parts of LED Light


² Press the light covers hard to make a space between, then tear apart the covers to reolace the inside parts

as follows:

17
Technical Manual for U Series Dental Unit

3.5 U100 Water-air Schematic Diagram

18
Technical Manual for U Series Dental Unit

3.6 U200 Water-air Schematic Diagram

19
Technical Manual for U Series Dental Unit

3.7 Fault Analysis of Air and Water System


² Fault 1: There’s no air.

Maintenance: Check the air supply valve inside the floor box, check if there is any twist of PU pipeline,

check if there is leakage or amlfunction of pressure adjusting valve filter, and then replace related elements.

² Fault 2: There’s no water.

Maintenance: Check the water supply valve inside the floor box, check if there is any twist of PU

pipeline, check if there is leakage or amlfunction of filter, and then replace related elements.

² Fault 3: No water for bowl rinsing or cup filling.

Maintenance: Check if the input voltage of the related solenoid valve is DC24V. Check if the PU pipe

twists. Check the electrical control board. Replace the related elements.

² Fault 4: HV suction or saliva ejector does not work.

Maintenance: Check if the input voltage of HV suction solenoid valve and saliva ejector solenoid valve

is DC24V. Check the optic-control switch located at the holder and the corresponding control board. Replace

related elements.

² Fault 5: There is no water or air for the three-way syringe.

Maintenance: Check if there is any twist of PU pipeline or syringe connection tube in the hydro-group

or if the spray-head is blocked. Replace related elements.

² Fault 6: No working air for handpiece.

Maintenance: Check if there is any twist of PU pipeline. Check the control circuit. Check the pedal air

control valve, instrument table three-way membrane valve. Replace related elements.

² Fault 7: No working water for handpiece.

Maintenance: Check if there is any twist of PU pipeline. Check the control circuit. Check the pedal

micro-control switch, the three-way membrane valve inside the instrument table and water adjusting valve.

Check if the input voltage of air supply solenoid valve and water supply solenoid valve is DC24V. Replace

related elements.

² Fault 8: There is no working mist spray for the handpiece.

Maintenance: Check if there is any twist of PU pipeline. Check the control circuit. Check the pedal

micro-control switch, the three-way membrane valve inside the instrument table and mist adjusting valve.

Check if the input voltage of mist solenoid valve is DC24V. Replace related elements.

² Fault 9: The air brake malfunctions.

Maintenance: Check if there is any twist of PU pipeline. Check air brake valve, check if there is any

leakage or malfunction of the valve. Replace related elements.

² Fault 10: No hot water for mouthwash.

20
Technical Manual for U Series Dental Unit

Maintenance: Check the heating control circuit. Check the AC24V power output. Check the heater

temperature controller and the heating element. Replace the related elements.

² Fault 11: The spittoon has a slow flush.

Maintenance: Check the filter. Check if the pipeline is blocked or twists. Check if there is any obstruction inside the

drainage pipeline.

² Fault 12: There is water or oil inside the air pipeline.

Maintenance: Check the air supply and air filter. Clean the filter or replace related elements.

3.8 Constitution of U100 Electric Control System


3.8.1 System constitution

² Hydro-group: 100cx hydro-group control board, 100ZS-30-A0 assistant’s control board

² Instrument table: 100ZL-02/F0 dentist’s push-button control board.

² Patient’s chair: ZY-01/F1 chair control board, U100 foot pedal

3.8.2 Electric diagram (see 3.12.1)

3.9 Description of U100 Hydro-group Control Board


² U100 hydro-group control board

J14
transformer

J3 arm cable
J32 assistant’s unit J31foot pedal

J11 operating light J30 chair control board

J12 bowl rinsing


J7, J8 reserved AC24V

J10 heating
J13 cup filling
J9 heating detection

3.10 Connector Function of ZY-01/F1 Chair Control Board (for all units)

21
Technical Manual for U Series Dental Unit

Backrest Connected to Communications cable connected


safety switch chair angle to the hydro-group
sensor

Power input 24V


Chair safety
switch

Connected to
Chair motor
backrest angle
sensor
Chair motor ascending

Backrest
Chair motor descending
motor

3.10.1 Function of connectors


No. connector function No. connector function
backrest safety switch
1 J1 chair motor 5 J5
access
2 J2 backrest motor 6 J6 AC24 power input
communications with
3 J3 7 J7 chair angle sensor access
hydro-group J30
backrest angle sensor
4 J4 chair safety switch access 8 J8
access

3.11 Fault Analysis of U100 Dentist’s Push-button Control Board

² Fault 1: There’s no response of the system when

the dental unit is plugged in.


J69 X-ray film
Maintenance:
viewer power

1) Check the power supply. connector


Connected to J3 on
2) Check the 220V fuse. 100CX-02-F1

² Fault 2: The bowl rinsing and cup filling don’t

work when the dental unit is plugged in.

Maintenance:

1) Check the DC24V output of each connector, if ok, check the solenoid valve and its connection.

2) If the DC24V is abnormal, check if F1 has blown out, if it has, replace the 10A fuse and check the

system to make sure there is no short circuit between DC24V and the ground. Then turn on the
22
Technical Manual for U Series Dental Unit

power switch, if the fuse does not blow out this time and the AC17V is also normal, replace the

hydro-group control board; if there is no AC17V output, replace the transformer.

² Fault 3: The operating light does not work.

Maintenance:

1) Check if the on-off switch located at the light head is on.

2) Check the fuse F2.

3) Check the connection of J11 and the voltages. If the voltages are correct, then replace the operating

light.

4) Check the output voltage of J11. If it is not correct or it is zero, then check the corresponding

transformer output winding. If it is abnormal, then replace the transformer; otherwise replace

hydro-group control board.

² Fault 4: The heater does not work properly.

Maintenance:

1) When the heater is working, the normal output of J9 is DC24V, if it is, the heater itself may

malfunction, replace it.

2) If there is no output of J9, check the fuse F1, if it has blown out, replace the 10A fuse.

3) If the fuse is normal, detect the voltage of the corresponding windings of the transformer output, if

no voltage, replace the transformer; otherwise, replace the hydro-group control board.

² Fault 5: The buttons on the assistant’s control panel do not work.

Maintenance:

1) Check the connection between J32 on the hydro-group board and J57 on the assistant’s control

board; if ok, replace the hydro-group board.

² Fault 6: The buttons on the dentist’s control panel do not work, while those on the assistant’s control

panel are normal.

Maintenance:

1) Check the connection between J95-3B and J3 on the hydro-group board.

2) If the connection is ok, detect the voltage between terminal 13 and 14 of J95-3B, if it is 5V, then

replace the dentist’s push-button panel.

3.12 Constitution of U200 Electric Control System


² Hydro-group: 30CX hydro-group control board, S90 assistant’s control board, S90 assistant’s optic

control board and S90 assistant’s touch control board.

² Instrument table: 30ZL instrument table control board, S30 dentist’s touch control board.

23
Technical Manual for U Series Dental Unit

² Patient’s chair: ZY-01/F1 chair control board, U200 foot pedal direction control board

3.12.1 U200 Electric Diagram

24
Technical Manual for U Series Dental Unit

3.13 U200 Hydro-group Control Board


3.13.1 Illustration of U200 hydro-group control board

Foot pedal communications


Chair communications

nstrument table
communications

Chair communications

Air pressure
in the water Heating 24V
Saliva Heating Heating
bottle Cup Bowl HV
ejector detection Operating
filling rinsing suction
light
Assistant’s
unit 24V
Distilled
output
water

Spittoon
safety
switch

Chair Reserved
Intra-oral Waste Water Air
power AC24V Transformer input
camera separator supply supply
supply
24V
3.13.2 Function of connectors

² J9 is transformer input connector which has 7 output windings from left to right: AC23V, AC23, AC18V,

AC6V, AC_0V, AC_11V and AC_12/17V, mainly for the power supply to the operating light.

² J10 is instrument table power output connector: power supplies to the instrument table are AC18V and

AC23V from left to right.

² J12 is operating light power output connector: when connected to LED light, AB is switch, CD has an

output of AC12V when the light is on; when connected to halogen light, B is AC12V/17V, C is AC11V,

and D is AC0V.

25
Technical Manual for U Series Dental Unit

² J21 is intra-oral camera power supply connector.

² J20 is waste separator power supply connector: the working mode of the separator is controlled by the

relay which controls on/off of the connector.

² J15 is reserved AC24V.

² J16 is connected to assistant’s unit to supply power for the curing light.

² J13 is heater power output: AC24V.

² J14 is heating detection connector: when the mouthwash is heated to 35℃, this connector is disconnected;

when the mouthwash is being heated, it is closed.

² J23 is HV suction solenoid valve: working voltage is DC24V.

² J22 is saliva ejector solenoid valve: working voltage is DC24V.

² J24 is bowl rinsing solenoid valve: working voltage is DC24V.

² J25 is Mouthwash solenoid valve: working voltage is DC24V.

² J32 is connected to assistant’s control board.

² J31 is connected to the direction control board of foot pedal.

² J29 is connected to the instrument table communications.

² J28 is spittoon safety switch connector.


Whip arm
3.14 Chair Control Board (the same with U100) switch 2

3.15 U200 Instrument Table Control Board Whip arm Whip arm
Whip arm
switch 3 switch 4
预留24V
Reserved Power input Micromot switch 1
or switch
Optic control
power supply
Communicatio
ns with the
hydro-group Optic
control
signal

Dentist’s
control
penal

X-ray film
viewer

Water for
F.O. 3 F.O. 2 F.O. 1 Scaler Driving Driving
scaler 3-way link
power water air
26 solenoid valve
supply
Technical Manual for U Series Dental Unit

² Function description: scanning of dentist’s touch panel buttons and air and water controlling.

² Function of connectors
connector description function connector description function
J37 GK1 KEY-CONNECT is
J38 GK2 dynamic working connected to the
instruments dentist’s touch dentist’s touch control
J39 GK3 button no control board
J43 board (the same with
connectors for signal connector S30), mainly for
J40 GK4 swing type controlling of the
keyboard IO signals
optic control
X-ray film DC3.3V (for hanging
J44 DC12V J41 power supply
viewer connector type)
connector
Reserved optic control Input (for hanging
J33 reserved AC24V J42
connector signal connector type)
connected to J29 on the When detection is
communications Scaler switch
J35 hydro-group control J37 effective, the scaler is
connector connector
board powered on.
(supplied by the
Instrument table When relay acts, the
hydro-group board), Air solenoid valve
J34 power supply J48 voltage of this port
connected to J10 on the connector
connector should be DC24V.
hydro-group board
When relay acts, the when relay acts, the
Solenoid valve 1 Water solenoid
J38 voltage of this port J49 voltage of this port
connector valve connector
should be DC24V. should be DC24V.
when relay acts, the when relay acts, the
Solenoid valve 2 Scaler power
J39 voltage of this port J51 voltage of this port
connector output connector
should be DC24V. should be DC24V.
when relay acts, the when relay acts, the
J52 F.O. 2 connector voltage of this port J53 F.O. 1 connector voltage of this port
should be DC3.3V. should be DC3.3V.
when relay acts, the
J54 F.O. 3 connector voltage of this port
should be DC3.3V.

3.16 U200 Dentist’s Touch Control Board


² Function description: keyboard signal switchover.

² Function description of connectors:

1) KEY-CONNECT is connected to U200 dentist’s touch control

board (the same with S30) for the scanning of keyboard button

signal.

2) U200 dentist’s touch control board (the same with S30).

3) U200 chair control board (the same with U100).

4) Fault analysis: please refer to fault analysis of U100 chair control board.

3.17 Foot Pedal Magnetic Control Board (optional for micromotor) (see 30JT-01/D6)

27
Technical Manual for U Series Dental Unit

² Function description: adjust the rotation speed

and power of micromotor.


² Connector function: connected to J70 of
90JT-02/F0.
² Installation and adjustment: Turn on the main

power, make sure the red indicator light is

lighting; slide the drive rod, and the output

voltage of detecting point C (adj) should be

DC-0.6~3.3V. If not within this range, make some adjustment as follows:

rough tuning
When the drive rod is not moved and
the voltage of ADJ has a difference
fine-tuning more than -0.6V, rough tune the
When the drive rod is not moved, fine-tune the potentiometer to about -0.6V then fine
potentiometer if the voltage of the detecting point tune the Hall sensor within -0.7V~0V.
ADJ is within -0.7V~0V; the distance of Hall sensor
makes the voltage of ADJ -0.6V (within -0.5 to -0.7).

Measure the detecting point when the drive rod is not moved, if the voltage is within
-0.7V-0V, fine tune the Hall sensor; if the difference is more than -0.6V, first rough
tune the potentiometer, then fine-tune the Hall sensor (to -0.6V).

rough tuning
When the drive rod slides to the
maximum position, and the
fine tuning
voltage of ADJ is less than +2.5V
When the drive rod slides to the
or more than +4V, adjust the
maximum position, and the voltage
potentiometer to make the voltage
of ADJ is within +2.5V~+4V, adjust
within +2.5V~4V and then fine
the distance of Hall sensor and
tune it.
make the voltage of ADJ about
+3.3V.

When the drive rod slides to the maximum position, measure the detecting point, if the
voltage is within 2.5V-4V, fine tune the Hall sensor, otherwise first rough tune the
potentiometer to 2.5V-4.5V and then fine tune the Hall sensor (to 3.3V).

3.18 Installation of Jumpers on the Hydro-group Control Board 30CX-01-F1


² Two kinds of jumpers on the 30CX-01-F1 board

28
Technical Manual for U Series Dental Unit

Short circuited here


when micromotor is

Empty installed on S30

here

Short
circuit here

3.19 Installation, Adjustment and Fault Analysis of Foot Pedal


² The installation of foot pedal cableb is shown below:

J70 is empty

J71 is
This wire is empty
connected to J70
on 90JT-02/F0

The switch must be connected to J70 on 90JT-02/F0 if


the 30JT-01/D6 is nor installed.
The switch will not be installed if the
30JT-01/D6 micromotor is installed.

² Foot pedal direction control board (see 90JT-02/F0)

² Function description: button switch board and hydro-group connection

² Function of connectors

connector function connector function


chair ascending
J67 connected to hydro-group control board S4
button
connected to aero position/air blast chair descending
J68 S3
button button
connected to rinsing position/water swift backrest ascending
J90 S2
button button
connected to micromotor control board, backrest descending
J70 S1
or to J71 if there’s no micromotor button
29
Technical Manual for U Series Dental Unit

3.20 U200 Assistant’s Control Board

Connected to
hydro-group

HV suction

reserved vacuum
saliva ejector
connector

²ejector
Saliva U200 助手位光控板
Connected to
touch control
board

² Function of connectors
connector function connector function
connected to J22 on the
J55 J61 AC24 input
hydro-group control board
HV suction, saliva
vacuum suction system and J57
J58 J57 ejector and optic control
optic control connector J59
connector
Connected to curing light
(note: when connected to connected to S90
AC24V curing light, and J56 assistant’s unit touch
J60
not for this connector, a control board (J1)
power converter is needed).

3.21 Fault Analysis of U200 Control Board


² Fault 1: There’s no response of the system when the dental unit is plugged in.

Maintenance:

1) Check the power supply.

2) Check the 220V fuse.

² Fault 2: No air or water when the dental unit is plugged in.


30
Technical Manual for U Series Dental Unit

Maintenance:

1) Check the connection of 18 and J19.

2) Power output of J18 and J19 should be DC24V. If the voltage is 0V, check if Fuse F1 blows out. If it

does, replace a fuse of 10A, then check the system and make sure that there is no short circuit

between DC24V and the ground, AC24VL and N. Turn on the power switch; if the fuse does not

blow out this time, and the power output AC24V is also normal, then the hydro-group control board

should be replaced.

² Fault 3: The LED power indicator of the hydro-group board does not light.

Maintenance:

1) Check the power input DC3.3V. If it is abnormal, replace the hydro-group control board.

2) Check the LED indicator. If it has malfunctioned, the system can work well. Replace the LED or

ignore the fault.

² Fault 4: The running indicator of the hydro-group board system does not light.

Maintenance:

1) Check the indicator. If it has malfunctioned, the system can work well. Replace the LED or ignore the

fault.

2) Check the power input DC3.3V. If it is normal, replace the hydro-group control board; if it is
abnormal (much higher than DC3.3V or short circuit), replace the hydro-group control board.

² Fault 5: The operating light does not work.

Maintenance:

1) Check if the on-off switch located at the light head is on.

2) Check the fuse SSD F2.

3) Check the connection of J11 and the voltages. If the voltages are correct, then replace the operating

light (LED light 0V13, switch).

4) Check the output voltage of J11. If it is not correct or it is zero, then check the corresponding

transformer output winding. If it is abnormal, then replace the transformer; otherwise replace

hydro-group control board.

² Fault 6: The vacuum pump, intra-oral camera, dental curing light, or waste separator does not work.

Maintenance:

1) Check the fuse F1.

2) Check the related port voltage AC24V (DC24). If the input voltage is normal, then probably the
related device malfunctions, replace it.

² Fault 7: The heater does not work properly.

31
Technical Manual for U Series Dental Unit

Maintenance:

1) If there’s no heating, first check the output of J13, if it is 24V, replace the heater; if the heating does

not turn off and J14 is always closed, it may also be the problem of the heater, then replace it.

2) If the voltage of J13 is zero, check the fuse F1. If it blows out, replace a 10A fuse.

3) If the fuse is in good condition, check the voltage of the corresponding transformer output winding.
If the voltage is zero, replace the transformer; otherwise, replace the hydro-group control board.

² Fault 8: All or some of the distilled water valve, distilled pressure valve, bowl rinsing solenoid valve,

cup filling solenoid valve, HV suction solenoid valve, and saliva ejector solenoid valve do not work

properly.

Maintenance

1) Check the output voltage of each port. If it is DC24V, then check the connection. If it is good, the

corresponding solenoid valve may have malfunctioned.

2) If the DC24V is abnormal, then check the fuse F1. If it has blown out, replace it with a 10A one, and

then check the system to ensure there’s no short circuit between DC24V and the ground, AC24VL

and N. Turn on the power switch; if the fuse does not blow out this time, and the power output

AC24V is also normal, then the hydro-group control board should be replaced. If there is no AC24V

output, then replace the transformer.

² Fault 9: The pedal does not work or some buttons malfunction.

Maintenance:

1) Check the connection between J31 and J67.

2) Check each switch and button on the pedal.

3) If the above two are ok, then replace the hydro-group control board.

² Fault 10: The intra-oral camera can’t be controlled on-off. And meanwhile, all the dynamic instruments

do not work properly.

Maintenance:

1) Check if the input voltage of pedal AD board is DC3.3V.

2) Check if U1 and U2 on the pedal AD board are at the right place.

3.22 Fault Analysis of U200 Instrument Table Control Board

² Fault 1: The buttons on the dentist’s control panel do not work, while those on the assistant’s control

panel are normal.

Maintenance:

1) Check the connection between J35 and the hydro-group J28; check the connection between J43 and

32
Technical Manual for U Series Dental Unit

the touch-button panel.

2) If the connection is ok, then check if there is +5V power output of the detecting point, if no voltage,

then replace U200 instrument table control board.

3) If there is +5V power input, pick up or pull out any handpiece to hear if the corresponding relay is

working; if working, then replace the dentist’s touch control panel.

² Fault 2: The X-ray film viewer does not light.

Maintenance:

1) Check if the power connector of X-ray viewer is loose.

2) Press the X-ray viewer button to see if there is an output of 12V, if not, replace the instrument table

board; if there is, then replace the viewer.

² Fault 3: There is no light inside the optical fiber.

Maintenance:
1) Check the output voltage. It should be less than DC3.5V.
2) If there is no output at all, replace the instrument table control board.
² Fault 4: The turbine works with no air or water.

Maintenance:
1) First check if the turbine works in the dry grinding status. In this case, choose air spray or water
spray through the pedal. Refer to the user manual for detailed information.
2) Check the connections of air supply solenoid valve and water supply solenoid valve. If there is no
problem, then check the input voltage DC24V. If it is ok, replace the valves.
3) If there is no DC24V, then check AC24V. If it is ok, replace instrument table control board.

Otherwise, check the power supply inside the hydro-group.

² Fault 5: The beeper does not work.

Maintenance:

1) Check the voltage of VCC.

2) If voltage is normal, replace the instrument table board.

² Fault 6: Air blast solenoid valve does not work.

Maintenance:
1) Step on the air blast button repeatedly, and check the relay. If the relay does not work, then check
the communications.
2) If the relay works, check if the input voltage for the solenoid valve is DC24V. If it is and the air
system works properly, then replace the solenoid valve.
4) If there is no DC24V but AC24V is normal, then replace the instrument table control board.

33
Technical Manual for U Series Dental Unit

² Fault 7: Neither the HV suction nor saliva ejector works.

Maintenance: Adjust the distance between the potentiometer and the optic sensor.

² Fault 8: Buttons mulfunction.

Maintenance: Check the connection with the touch control board.