TECHNICAL MANUAL
FOR
SIGER DENTAL UNIT
目 录
Chapter 1 Installation of U Series Dental Unit ........................................................................................... 2
1.1 Function and Technical Parameters....................................................................................................... 2
1.2 Installation Condition and Place Requirements .................................................................................. 4
1.3 Handling and Unpacking of the Unit .................................................................................................... 6
1.4 Installation of U200............................................................................................................................... 7
1.5 Installation of U100............................................................................................................................... 7
1.5.1 Installation of patient’s chair.............................................................................................................. 7
1.5.2 Installation of hydro-group................................................................................................................. 8
1.5.3 Installation of instrument table assembly (see above)........................................................................ 9
1.6 Installation of operating light ................................................................................................................ 9
1.7 Installation of water, air and power supply ........................................................................................... 9
1.8 Installation of Scaler ............................................................................................................................. 9
1.8.1 Installation of scaler on U100 ............................................................................................................ 9
1.8.2 Installation of scaler on U200 .......................................................................................................... 10
1.8.3 Connection between Scaler and Instrument Table Control Board.....................................................11
1.9 Installation of Curing Light..................................................................................................................11
1.9.1 Installation of curing light on U100 ..................................................................................................11
1.9.2 Installation of curing light on U200 ................................................................................................. 12
1.10 Installation of Intra-oral Camera (see the following illustration)...................................................... 12
1.11 Installation of Micromotor on U200 ................................................................................................. 12
Chapter 2 Adjustment of U Series Dental Unit ......................................................................................... 14
2.1 Adjustment .......................................................................................................................................... 14
2.2 Safety Function and Pipeline Disinfection.......................................................................................... 15
Chapter 3 Maintenance of U Series Dental Unit....................................................................................... 16
3.1 Replacement of Motors ....................................................................................................................... 16
3.2 Replacement of the Air Spring ............................................................................................................ 17
3.3 Replacement of the Upholstery ........................................................................................................... 17
3.4 Replacement of the Parts of LED Light .............................................................................................. 17
3.5 U100 Water-air Schematic Diagram ................................................................................................... 18
3.6 U200 Water-air Schematic Diagram ................................................................................................... 19
3.7 Fault Analysis of Air and Water System.............................................................................................. 20
3.8 Constitution of U100 Electric Control System.................................................................................... 21
3.12 Constitution of U200 Electric Control System.................................................................................. 23
3.13 U200 Hydro-group Control Board .................................................................................................... 25
3.14 Chair Control Board (the same with U100) ...................................................................................... 26
3.15 U200 Instrument Table Control Board.............................................................................................. 26
3.16 U200 Dentist’s Touch Control Board................................................................................................ 27
3.17 Foot Pedal Magnetic Control Board (optional for micromotor) (see 30JT-01/D6)........................... 27
3.18 Installation of Jumpers on the Hydro-group Control Board 30CX-01-F1 ........................................ 28
3.19 Installation, Adjustment and Fault Analysis of Foot Pedal ............................................................... 29
3.20 U200 Assistant’s Control Board........................................................................................................ 30
3.21 Fault Analysis of U200 Control Board.............................................................................................. 30
1
Technical Manual for U Series Dental Unit
NOTE: This chapter will provide the vital technical parameters of the unit. Please read this chapter
12345 6 7 8 9 10 11
12 13 14 15 16 1718 19 20 21
2
Technical Manual for U Series Dental Unit
² Nameplate: the nameplate is located on the right side of the hydro-group. It contains the following data:
product name and type, power voltage, frequency, classification of electric safety, power consumption,
mode of operation, registration number, implementation of the standards, serial number, date of
² Technical parameters: The device complies with the following standards: YZB/Guangdong 0207-2008, EC
² Ambient conditions:
Power supply: AC 220V±22V, 50Hz±1Hz; water supply pressure: 200kPa~400kPa; air supply pressure:
≥500kPa。
The dental unit should be fixed on flat ground, and the skewness should be <0.5° after installation.
3
Technical Manual for U Series Dental Unit
² NOTE: The installation must be performed by the professional personnel from Siger or trained and
authorized by Siger.
² The unit must be placed in clean, dry, airy, shady and cool place. There should be enough room around
B. Changes of high temperature or high humidity often occur, such as places near air-conditioner or
heaters, etc.
² Choose the installation location according to the specific conditions of the dental treatment room such as
the general layout, lighting and easy access to instruments, etc. The unit should be installed in clean, dry,
airy and cool place. And the floor on which the dental chair is located should be flat, level, and solid.
² Standard layout: The layout below shows the size of single-person operation. It does not include size of
other auxiliary equipments in the treatment room. A is 500mm, and B is 800mm. In case of two-people
operation, A is 1000MM. With additional auxiliary equipments, the size should be increased accordingly.
)
管
(
污
进
排
管)
气
进水
管
)
电源)
4
Technical Manual for U Series Dental Unit
5
Technical Manual for U Series Dental Unit
Waste discharge
pipe
Power supply
In the ground holes, the drainage pipe should be Ф50MM PVC pipe, water supply pipe should be
Ф15MM, and air supply pipe should be Ф15MM. High-pressure plastic composite pipes can be used for water
supply pipe and air supply pipe. Valves should be connected to the ports. The power supply is 220v with earth
wire, the diameter of the power supply wire should be ≧1.5MM². It is preferred that the pipelines are not
above the ground, once it is, the vertical interval should be no more than 100mm. If there is intra-oral camera
connected to a computer, it must be added to the ground holes so that the VGA cable and the AV cable can be
² Take out the packing list and check whether the unit and spare parts are
6
Technical Manual for U Series Dental Unit
remove the four screws fixed on the bottom plate. Put the bottom part of the seat that
can bear load on the predetermined place while maintaining a balanced state.
² Installation of backrest cover: insert the four buckling parts on the cover into the
² Insert the headrest into the corresponding position on the backrest as shown in the
picture.
² Move out the hydro-group from the fixing plate to the connecting arm while retaining the balance.
² Insert the light post into corresponding holes of the hydro-group upright;
penetrate the wire connected to the light arm into the light post; then connect
and fix the head of the operating light; and then connect and fix the wire plug
of the light arm with corresponding connectors on the circuit board in the
hydro-group; the connection port for the operating light is shown in the
Connection port
picture:
² First unpack the container and remove the accessories, then move the chair to the installation position and fix it as is
² The installation of upholstery, bending plate, backrest cover, headrest and hydro-groupconnecting arm is
as follows:
①
②
⑤
① Fix the upholstery and the chair frame with black countersunk inner hexagon screw sized
7
Technical Manual for U Series Dental Unit
10*35mm;
② Installation of bending plate and swing link: make the hole on the bending plate aligning at that on
the swing link, then lock them together with M10 stainless steel inner hexagon screw;
③ Installation of the hydro-group connecting arm: align the connecting arm and the chair frame, lock
them together with black countersunk inner hexagon screw sized 10*35mm;
④ Installation of bending plate and backrest upholstery: lock with M10 stainless steel inner hexagon
screw;
⑤ Installation of backrest cover: insert the four buckling parts on the cover into the corresponding
² Move the hydro-group assembly from the container to the connecting arm, adjust
² Installation of the balance arm: install to the hydro-group, the air and water tubes
² Connection of air and water tubes: the white water tube is connected to the water connector on the 3-5
way link, the blue air tube is connected to the air connector on the 3-5 way link, the other three
connectors are connected to the corresponding tubes and colors in the foot pedal:
tube
8
Technical Manual for U Series Dental Unit
upright; penetrate the wire connected to the light arm into the light
U100 接手
post; then connect and fix the head of the operating light; and then
connect and fix the wire plug of the light arm with corresponding
Operating light
connector
straight coupling, and insert the other end into the whole air supply port.
² Connect a G1/4*Ф8 straight coupling with the water supply ball valve. Insert one end of a blue Ф8PA pipe into the
straight coupling, and insert the other end into the whole water supply port.
² Insert the drainage pipe and vent pipe of saliva ejector into the Ф50PVC drainage pipe ground hole.
² Connect the power cable and earth wire. Be sure to have a good connection. Plug all the connectors in the
corresponding sockets on the circuit board inside the ground box. And pack up the lines.
² If the instrument table is already installed with preassembled accessories for scaler, install according to the
9
Technical Manual for U Series Dental Unit
Power supply
AC 24V
AC 24V
Solenoid valve
Foot
S/W
Potentiometer: 10k
² If the instrument table is not pre-installed with scaler accessories, install those accessories first (hexagon valve,
Pneumatic
Potentiometer connected to pneumatic switch
control
installed on on the instrument table board
switch
the (constantly open)
instrument
table board
connected
to
reserved
24V on the
Hexagon
instrumen
valve
t table
board Tapping position for the
potentiometer
² Preassembled accessories: scaler control board, 30ZL instrument table control board, small flow adjusting valv
² Installation steps: connect the white water tube ① to the small flow adjusting valve => connect ②③ to the
scaler control board => connect the black power cable on the scaler control board to the instrument table control
board => output voltage: AC24V => connect the white wire on the scaler control board to the switch
10
Technical Manual for U Series Dental Unit
J51JYJ—SW on the instrument table control board => this installation applies to scalers from South Korea and
①white water
tube Ф1.5/2.5
②white wire
③red wire
² Fix the scaler board on the fixing plate inside the instrument table, and install the potentiometer under the
instrument table.
² Installation of scaler tubing: cut the surface cover to a proper length and refer to the installation of handpiece
tubings.
² After fixation, connect the water tube to the flow adjusting valve, and the output wire to the scaler board.
² Note: for unit of hanging type, remove the GK1 short-circuit bolt before the installation of scaler.
² The curing light is to be installed on the assistant’s unit. Cut the tubing to proper length, put it through the black
hole into the assistant’s unit and fix it. The red and green wires go through the assistant’s unit arm and are
connected to AC24V on the hydro-group board. Note: if the curing light is supported by DC power supply, a
rectifier box should be installed in the hydro-group before the installation of the curing light.
11
Technical Manual for U Series Dental Unit
² Installation: fix the curing light tubings on the assistant's assembly welding plate => connect the power cable to
the assistant's control board => this installation applies to curing light with a input voltage of AC24V.
② Use four cross recess head screws sized 4*35 to lock the
display
12
Technical Manual for U Series Dental Unit
² The white air tube ① with A is connected to the port Ф2.6 on the membrane valve.
² The green water tube ② is connected to the small flow adjusting valve.
² The white air tube ③ is connected to the 3-way link assembly; the other two ends of the assembly are
respectively connected to the port Ф3.2 on the membrane valve and the air-control electric valve on the
micromotor. See the S30/S60 Micromotor Installation and Wiring Diagram above.
² The red and black wires ④⑤ are connected to the DT switch. The brown and blue wires ⑥⑦ are connected to
the terminal on the instrument table control board of U200. The output voltage is DC3.2V.
² The terminal 1 and 2 on the micromotor board are connected correspondingly to the terminals on the instrument
² A on the air-control valve is connected to the terminal 6 on the micromotor control board.
² The terminal 5 on the micromotor control board and B on the air-control electric valve are connected to J36 on
² The DT switch is connected to the terminal 3 and 4 on the micromotor control board.
13
Technical Manual for U Series Dental Unit
² Add a magnetic control board of S90 foot pedal on the original foot pedal.
² Add two small magnets on the rack. Note: the opposite sides of the magnets should be south and north poles.
Fixing Fixing
hole hole
Fixing Fixing
hole hole
pic 3 pic 2
² Tightness adjustment of the headrest pulling plate: adjust through the jackscrew (pic 2).
² Level adjustment of the chair: Place a spirit level on the instrument table in lateral and longitudinal
direction. Then adjust the nine inner hexagon screws mounted on the bottom plate of the chaie to keep
the instrument table level with misalignment within 3 degrees’ deflection. (pic 3)
14
Technical Manual for U Series Dental Unit
² Instrument table arm force adjustment: Turn the adjusting screw clockwise to increase the force, and turn it
² As shown in the pictures above, turn the input air adjusting valve to let the two pressure
gauges display 0.55Mkpa and 0.3Mpa (pic 5 and 6).
² Screw the hydro-group onto the fixing plate with four fixing screws, adjust the level through four
² Air pressure adjustment of high speed turbine: Turn the knob on the bottom of the instrument table to
² Air pressure adjustment of low speed air motor: Turn the knob on the bottom of the instrument table to
² Tightness adjustment of the light arm: open the adjusting cover, use inner hexagon adjusting nut to
pic 10 pic 11
15
Technical Manual for U Series Dental Unit
² Note: the air pressure inside the distilled water bottle should not exceed 300Kpa.
² The chair or backrest would stop moving down (U200) in the presence of obstacles and the chair would
² Any one of the dental chair control buttons can stop the chair from moving.
² When the dynamic instrument is in use, the dental chair will be under safety protection and none
movement is available.
² An interlocking device prevents the instruments from working at the same time. The first instrument
removed is operative. Those removed afterwards are inactivated by the interlocking device (U200).
upholstery with a cross screwdriver => remove the foamed cushion => use a snap ring plier to remove
the four snap rings and springs at the two ends of the motor => pry out the motor shaft with a
² Maintenance of the lower motor: put the hydro-group on an upholder which can bear a weight of 30kg
=> use a cross screwdriver to demount the upper parallel cover => remove the snap ring on the motor
with a snap ring plier => take out the motor => replace the motor.
16
Technical Manual for U Series Dental Unit
² Replacement of the lower motor of U200: put the hydro-group on an upholder which can bear a weight
of 30kg => use a cross screwdriver to demount the upper parallel cover => take out
the locking jackscrew with an inner haxegon wrench => tak out the motor shaft with
clamp springs at the ends of the stationary shaft of the air spring => pry out the the stationary shaft with a
the headrest upholstery, then grab the cover with right hand, and tear apart the
² Replacement of the backrest upholstery: first pry out a space between the cover and
the backrest upholstery, then grab the cover with right hand, and tear apart the
² Replacement of the chair upholstery: loosen the four screws on the upholstery to replace.
as follows:
17
Technical Manual for U Series Dental Unit
18
Technical Manual for U Series Dental Unit
19
Technical Manual for U Series Dental Unit
Maintenance: Check the air supply valve inside the floor box, check if there is any twist of PU pipeline,
check if there is leakage or amlfunction of pressure adjusting valve filter, and then replace related elements.
Maintenance: Check the water supply valve inside the floor box, check if there is any twist of PU
pipeline, check if there is leakage or amlfunction of filter, and then replace related elements.
Maintenance: Check if the input voltage of the related solenoid valve is DC24V. Check if the PU pipe
twists. Check the electrical control board. Replace the related elements.
Maintenance: Check if the input voltage of HV suction solenoid valve and saliva ejector solenoid valve
is DC24V. Check the optic-control switch located at the holder and the corresponding control board. Replace
related elements.
Maintenance: Check if there is any twist of PU pipeline or syringe connection tube in the hydro-group
Maintenance: Check if there is any twist of PU pipeline. Check the control circuit. Check the pedal air
control valve, instrument table three-way membrane valve. Replace related elements.
Maintenance: Check if there is any twist of PU pipeline. Check the control circuit. Check the pedal
micro-control switch, the three-way membrane valve inside the instrument table and water adjusting valve.
Check if the input voltage of air supply solenoid valve and water supply solenoid valve is DC24V. Replace
related elements.
Maintenance: Check if there is any twist of PU pipeline. Check the control circuit. Check the pedal
micro-control switch, the three-way membrane valve inside the instrument table and mist adjusting valve.
Check if the input voltage of mist solenoid valve is DC24V. Replace related elements.
Maintenance: Check if there is any twist of PU pipeline. Check air brake valve, check if there is any
20
Technical Manual for U Series Dental Unit
Maintenance: Check the heating control circuit. Check the AC24V power output. Check the heater
temperature controller and the heating element. Replace the related elements.
Maintenance: Check the filter. Check if the pipeline is blocked or twists. Check if there is any obstruction inside the
drainage pipeline.
Maintenance: Check the air supply and air filter. Clean the filter or replace related elements.
J14
transformer
J3 arm cable
J32 assistant’s unit J31foot pedal
J10 heating
J13 cup filling
J9 heating detection
3.10 Connector Function of ZY-01/F1 Chair Control Board (for all units)
21
Technical Manual for U Series Dental Unit
Connected to
Chair motor
backrest angle
sensor
Chair motor ascending
Backrest
Chair motor descending
motor
Maintenance:
1) Check the DC24V output of each connector, if ok, check the solenoid valve and its connection.
2) If the DC24V is abnormal, check if F1 has blown out, if it has, replace the 10A fuse and check the
system to make sure there is no short circuit between DC24V and the ground. Then turn on the
22
Technical Manual for U Series Dental Unit
power switch, if the fuse does not blow out this time and the AC17V is also normal, replace the
Maintenance:
3) Check the connection of J11 and the voltages. If the voltages are correct, then replace the operating
light.
4) Check the output voltage of J11. If it is not correct or it is zero, then check the corresponding
transformer output winding. If it is abnormal, then replace the transformer; otherwise replace
Maintenance:
1) When the heater is working, the normal output of J9 is DC24V, if it is, the heater itself may
2) If there is no output of J9, check the fuse F1, if it has blown out, replace the 10A fuse.
3) If the fuse is normal, detect the voltage of the corresponding windings of the transformer output, if
no voltage, replace the transformer; otherwise, replace the hydro-group control board.
Maintenance:
1) Check the connection between J32 on the hydro-group board and J57 on the assistant’s control
² Fault 6: The buttons on the dentist’s control panel do not work, while those on the assistant’s control
Maintenance:
2) If the connection is ok, detect the voltage between terminal 13 and 14 of J95-3B, if it is 5V, then
² Instrument table: 30ZL instrument table control board, S30 dentist’s touch control board.
23
Technical Manual for U Series Dental Unit
² Patient’s chair: ZY-01/F1 chair control board, U200 foot pedal direction control board
24
Technical Manual for U Series Dental Unit
nstrument table
communications
Chair communications
Air pressure
in the water Heating 24V
Saliva Heating Heating
bottle Cup Bowl HV
ejector detection Operating
filling rinsing suction
light
Assistant’s
unit 24V
Distilled
output
water
Spittoon
safety
switch
Chair Reserved
Intra-oral Waste Water Air
power AC24V Transformer input
camera separator supply supply
supply
24V
3.13.2 Function of connectors
² J9 is transformer input connector which has 7 output windings from left to right: AC23V, AC23, AC18V,
AC6V, AC_0V, AC_11V and AC_12/17V, mainly for the power supply to the operating light.
² J10 is instrument table power output connector: power supplies to the instrument table are AC18V and
² J12 is operating light power output connector: when connected to LED light, AB is switch, CD has an
output of AC12V when the light is on; when connected to halogen light, B is AC12V/17V, C is AC11V,
and D is AC0V.
25
Technical Manual for U Series Dental Unit
² J20 is waste separator power supply connector: the working mode of the separator is controlled by the
² J16 is connected to assistant’s unit to supply power for the curing light.
² J14 is heating detection connector: when the mouthwash is heated to 35℃, this connector is disconnected;
3.15 U200 Instrument Table Control Board Whip arm Whip arm
Whip arm
switch 3 switch 4
预留24V
Reserved Power input Micromot switch 1
or switch
Optic control
power supply
Communicatio
ns with the
hydro-group Optic
control
signal
Dentist’s
control
penal
X-ray film
viewer
Water for
F.O. 3 F.O. 2 F.O. 1 Scaler Driving Driving
scaler 3-way link
power water air
26 solenoid valve
supply
Technical Manual for U Series Dental Unit
² Function description: scanning of dentist’s touch panel buttons and air and water controlling.
² Function of connectors
connector description function connector description function
J37 GK1 KEY-CONNECT is
J38 GK2 dynamic working connected to the
instruments dentist’s touch dentist’s touch control
J39 GK3 button no control board
J43 board (the same with
connectors for signal connector S30), mainly for
J40 GK4 swing type controlling of the
keyboard IO signals
optic control
X-ray film DC3.3V (for hanging
J44 DC12V J41 power supply
viewer connector type)
connector
Reserved optic control Input (for hanging
J33 reserved AC24V J42
connector signal connector type)
connected to J29 on the When detection is
communications Scaler switch
J35 hydro-group control J37 effective, the scaler is
connector connector
board powered on.
(supplied by the
Instrument table When relay acts, the
hydro-group board), Air solenoid valve
J34 power supply J48 voltage of this port
connected to J10 on the connector
connector should be DC24V.
hydro-group board
When relay acts, the when relay acts, the
Solenoid valve 1 Water solenoid
J38 voltage of this port J49 voltage of this port
connector valve connector
should be DC24V. should be DC24V.
when relay acts, the when relay acts, the
Solenoid valve 2 Scaler power
J39 voltage of this port J51 voltage of this port
connector output connector
should be DC24V. should be DC24V.
when relay acts, the when relay acts, the
J52 F.O. 2 connector voltage of this port J53 F.O. 1 connector voltage of this port
should be DC3.3V. should be DC3.3V.
when relay acts, the
J54 F.O. 3 connector voltage of this port
should be DC3.3V.
board (the same with S30) for the scanning of keyboard button
signal.
4) Fault analysis: please refer to fault analysis of U100 chair control board.
3.17 Foot Pedal Magnetic Control Board (optional for micromotor) (see 30JT-01/D6)
27
Technical Manual for U Series Dental Unit
rough tuning
When the drive rod is not moved and
the voltage of ADJ has a difference
fine-tuning more than -0.6V, rough tune the
When the drive rod is not moved, fine-tune the potentiometer to about -0.6V then fine
potentiometer if the voltage of the detecting point tune the Hall sensor within -0.7V~0V.
ADJ is within -0.7V~0V; the distance of Hall sensor
makes the voltage of ADJ -0.6V (within -0.5 to -0.7).
Measure the detecting point when the drive rod is not moved, if the voltage is within
-0.7V-0V, fine tune the Hall sensor; if the difference is more than -0.6V, first rough
tune the potentiometer, then fine-tune the Hall sensor (to -0.6V).
rough tuning
When the drive rod slides to the
maximum position, and the
fine tuning
voltage of ADJ is less than +2.5V
When the drive rod slides to the
or more than +4V, adjust the
maximum position, and the voltage
potentiometer to make the voltage
of ADJ is within +2.5V~+4V, adjust
within +2.5V~4V and then fine
the distance of Hall sensor and
tune it.
make the voltage of ADJ about
+3.3V.
When the drive rod slides to the maximum position, measure the detecting point, if the
voltage is within 2.5V-4V, fine tune the Hall sensor, otherwise first rough tune the
potentiometer to 2.5V-4.5V and then fine tune the Hall sensor (to 3.3V).
28
Technical Manual for U Series Dental Unit
here
Short
circuit here
J70 is empty
J71 is
This wire is empty
connected to J70
on 90JT-02/F0
² Function of connectors
Connected to
hydro-group
HV suction
reserved vacuum
saliva ejector
connector
²ejector
Saliva U200 助手位光控板
Connected to
touch control
board
² Function of connectors
connector function connector function
connected to J22 on the
J55 J61 AC24 input
hydro-group control board
HV suction, saliva
vacuum suction system and J57
J58 J57 ejector and optic control
optic control connector J59
connector
Connected to curing light
(note: when connected to connected to S90
AC24V curing light, and J56 assistant’s unit touch
J60
not for this connector, a control board (J1)
power converter is needed).
Maintenance:
Maintenance:
2) Power output of J18 and J19 should be DC24V. If the voltage is 0V, check if Fuse F1 blows out. If it
does, replace a fuse of 10A, then check the system and make sure that there is no short circuit
between DC24V and the ground, AC24VL and N. Turn on the power switch; if the fuse does not
blow out this time, and the power output AC24V is also normal, then the hydro-group control board
should be replaced.
² Fault 3: The LED power indicator of the hydro-group board does not light.
Maintenance:
1) Check the power input DC3.3V. If it is abnormal, replace the hydro-group control board.
2) Check the LED indicator. If it has malfunctioned, the system can work well. Replace the LED or
² Fault 4: The running indicator of the hydro-group board system does not light.
Maintenance:
1) Check the indicator. If it has malfunctioned, the system can work well. Replace the LED or ignore the
fault.
2) Check the power input DC3.3V. If it is normal, replace the hydro-group control board; if it is
abnormal (much higher than DC3.3V or short circuit), replace the hydro-group control board.
Maintenance:
3) Check the connection of J11 and the voltages. If the voltages are correct, then replace the operating
4) Check the output voltage of J11. If it is not correct or it is zero, then check the corresponding
transformer output winding. If it is abnormal, then replace the transformer; otherwise replace
² Fault 6: The vacuum pump, intra-oral camera, dental curing light, or waste separator does not work.
Maintenance:
2) Check the related port voltage AC24V (DC24). If the input voltage is normal, then probably the
related device malfunctions, replace it.
31
Technical Manual for U Series Dental Unit
Maintenance:
1) If there’s no heating, first check the output of J13, if it is 24V, replace the heater; if the heating does
not turn off and J14 is always closed, it may also be the problem of the heater, then replace it.
2) If the voltage of J13 is zero, check the fuse F1. If it blows out, replace a 10A fuse.
3) If the fuse is in good condition, check the voltage of the corresponding transformer output winding.
If the voltage is zero, replace the transformer; otherwise, replace the hydro-group control board.
² Fault 8: All or some of the distilled water valve, distilled pressure valve, bowl rinsing solenoid valve,
cup filling solenoid valve, HV suction solenoid valve, and saliva ejector solenoid valve do not work
properly.
Maintenance
1) Check the output voltage of each port. If it is DC24V, then check the connection. If it is good, the
2) If the DC24V is abnormal, then check the fuse F1. If it has blown out, replace it with a 10A one, and
then check the system to ensure there’s no short circuit between DC24V and the ground, AC24VL
and N. Turn on the power switch; if the fuse does not blow out this time, and the power output
AC24V is also normal, then the hydro-group control board should be replaced. If there is no AC24V
Maintenance:
3) If the above two are ok, then replace the hydro-group control board.
² Fault 10: The intra-oral camera can’t be controlled on-off. And meanwhile, all the dynamic instruments
Maintenance:
² Fault 1: The buttons on the dentist’s control panel do not work, while those on the assistant’s control
Maintenance:
1) Check the connection between J35 and the hydro-group J28; check the connection between J43 and
32
Technical Manual for U Series Dental Unit
2) If the connection is ok, then check if there is +5V power output of the detecting point, if no voltage,
3) If there is +5V power input, pick up or pull out any handpiece to hear if the corresponding relay is
Maintenance:
2) Press the X-ray viewer button to see if there is an output of 12V, if not, replace the instrument table
Maintenance:
1) Check the output voltage. It should be less than DC3.5V.
2) If there is no output at all, replace the instrument table control board.
² Fault 4: The turbine works with no air or water.
Maintenance:
1) First check if the turbine works in the dry grinding status. In this case, choose air spray or water
spray through the pedal. Refer to the user manual for detailed information.
2) Check the connections of air supply solenoid valve and water supply solenoid valve. If there is no
problem, then check the input voltage DC24V. If it is ok, replace the valves.
3) If there is no DC24V, then check AC24V. If it is ok, replace instrument table control board.
Maintenance:
Maintenance:
1) Step on the air blast button repeatedly, and check the relay. If the relay does not work, then check
the communications.
2) If the relay works, check if the input voltage for the solenoid valve is DC24V. If it is and the air
system works properly, then replace the solenoid valve.
4) If there is no DC24V but AC24V is normal, then replace the instrument table control board.
33
Technical Manual for U Series Dental Unit
Maintenance: Adjust the distance between the potentiometer and the optic sensor.
34