Professional Documents
Culture Documents
Precise results
A repeatability test was performed by measuring NIST 698
standard twenty times, to test the precision of the Epsilon 4. The
results are given in Table 3.
of this information and we shall not be liable for errors contained herein or for damages in connection with the use of this material. Malvern Panalytical reserves the right to change the content in this material at any time without notice.
Table 3. Results of the repeatability test
Disclaimer: Although diligent care has been used to ensure that the information in this material is accurate, nothing herein can be construed to imply any representation or warranty as to the accuracy, correctness or completeness
NIST 698 Certified Average measured RMS
(n = 20) concentration concentration (wt%)
(wt%) (wt%)
Al2O3 48.2 48.04 0.04
CaO 0.62 0.572 0.003
Cr2O3 0.08 0.076 0.001
Fe2O3 19.6 19.36 0.01
0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0
K2O 0.01 0.015 0.003
Fe2O3 concentration (wt%)
MgO 0.058 0.037 0.003
Figure 3. Calibration graph for Fe2O3 in bauxite
MnO 0.38 0.373 0.002
Copyright: © 2018 Malvern Panalytical. This publication or any portion thereof may not be copied or transmitted without our express written permission. 9498 707 62011 PN11199
P2O5 0.37 0.43 0.01
SO3 0.144 0.139 0.002
SiO2 0.69 0.69 0.01
TiO2 2.38 2.32 0.01
Corrected intensity
MALVERN PANALYTICAL
Grovewood Road, Malvern Lelyweg 1, 7602 EA Almelo,
Worcestershire, UK, WR14 1XZ the Netherlands
P.O. Box 13, 7600 AA Almelo,
Tel +44 1684 892456 The Netherlands
Fax +44 1684 892789
Tel +31 546 534 444
Fax +31 546 534 598
info@malvernpanalytical.com
www.malvernpanalytical.com