You are on page 1of 9

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/272469731

Permanent pacemakers: An overview

Article  in  British Journal of Cardiac Nursing · June 2009


DOI: 10.12968/bjca.2009.4.6.42422

CITATIONS READS

4 6,189

1 author:

Alan Davies
The University of Manchester
78 PUBLICATIONS   243 CITATIONS   

SEE PROFILE

Some of the authors of this publication are also working on these related projects:

GazeMod View project

Exploiting an understanding of human visual perception to facilitate human-machine electrocardiogram interpretation of drug-induced long QT syndrome View project

All content following this page was uploaded by Alan Davies on 02 May 2016.

The user has requested enhancement of the downloaded file.


Clinical

Permanent pacemakers:
An overview
Alan Davies is Staff Nurse, Cardiac Catheterization Labs, Cardiology,
Royal Devon and Exeter Hospital, Barrack Rd, Exeter, EX2 5DW. Email: alan6386@hotmail.com

T
he permanent pacemaker (PPM) has increased in were 603 pacemaker implants per million in the UK alone.
use over recent years, with around 25 000 PPMs Many nurses who work in a variety of clinical areas are
implanted each year in the UK alone and rising exposed to patients with a PPM in situ. Nurses who care
(National Institute for Health and Clinical Excellence for such patients should have a general awareness of
(NICE), 2005). Ector et al (2007) state that in 2005 there PPMs, as the presence of a PPM can impact on the
management of the patient in the clinical setting. This
article aims to provide an overview of the PPM, including
the insertion procedure, pre- and post-procedure care and
patient advice.

Left bundle
The intrinsic conduction system
Sinus branch Specialized conduction cells are distributed through the
node
heart in the form of an electrical pathway (Davies, 2007a)
Anterior
AV
fascicle
(Figure 1). These cells possess a quality known as automa-
node ticity, this is the ability of the cells in the heart to depolar-
ize spontaneously and initiate an action potential (impulse)
Right
atrium (Bowbrick and Borg, 2006).
Depolarization is a process whereby the myocardial cel-
Atrioventricular lular membrane switches from being more negatively-
region
Right bundle Posterior charged inside and positively-charged outside (polarized)
branch fascicle to the reverse (depolarized) (Khan, 2006). This is achieved
by the sudden influx of positively charged sodium (Na+)
and calcium ions (Ca2+) into the cell (Conover, 2003). The
sinoatrial node (SAN) activates the atria at a rate between
Figure 1. Conduction system of the heart 60 and 100 beats per minute in an adult patient. The wave
of depolarization, originating in the SAN spreads down
the conduction system to the atrioventricular (AV) node;
Abstract through the His bundle and into the left and right bundle
Many patients in many different clinical areas have a permanent branches, ending in the Purkinje fibers (Davies, 2007b). In
pacemaker (PPM) device implanted. Nurses are better equipped to some patients there is a problem with the intrinsic con-
deal with the unique needs of this patient group by understanding the duction system, leading to the need to insert a permanent
different types of pacemaker device available, how these devices pacemaker system (British Heart Foundation (BHF),
function, and how this impacts on the patient’s clinical care. An 2008).
understanding of the pacemaker device also helps to identify or
anticipate potential problems with the device, which may affect the Components of a permanent pacemaker
patient’s health. Certain clinical tests and equipment may also affect A PPM system is made up of a pulse generator (battery),
the functioning of a pacemaker device and it is important that nurses and one or more electrodes, known as leads (Johnson and
are aware of these effects so that steps can be taken to prevent or Rawlings-Anderson, 2007).
reduce these effects, and deliver the best possible care to patients All of the currently used pulse generators or pacemaker
with a permanent pacemaker. boxes are variations of lithium-iodine cells (Figure 2).
These cells work by generating electricity due to a chemi-
Key words cal reaction that takes place between the lithium and
w Pacemakers w PPM w Pulse generators w Leads iodine (Kenny, 2007). Pulse generators can last on average
Submitted for peer review 9 August 2008. Accepted for publication 18 May 2009. between 5 and 10 years (Kumar and Clark, 2003). Pulse
Conflict of interest: None
generators are usually about the size of a matchbox and
weigh around 20–50 grams (BHF, 2008).

262 British Journal of Cardiac Nursing June 2009 Vol 4 No 6


Clinical

Table 1.
Indications for implantation of
a permanent pacemaker
w Heart block, including
Sinoatrial block
Mobitz type II block
Complete heart block
Trifascicular block
w Symptomatic bradycardia
w Sick sinus syndrome (sinus node disease)
w Ventricular standstill
w Hypertrophic cardiomyopathy
Figure 2. Pacemaker pulse generator
w Dilated cardiomyopathy
The leads are insulated wires that are implanted into the w Long QT syndromes
heart transvenously, attached at one end to the pulse gen- w Drug-refractory paroxysmal atrial fibrillation
erator and the other to the myocardium via active or pas- w Bradycardia–tachycardia syndrome
sive fixation (Figure 3). This in turn creates a two-way path
w Pause-dependent ventricular tachycardia
through which an impulse can be delivered to the myocar-
dium and received via the pulse generator (Kenny, 2007). w Ventricular pauses >3 seconds in length
The lead delivers an electrical impulse exciting myocardial w Recurrent syncope
cells causing a subsequent contraction (Johnson and Johnson and Rawlings-Anderson, 2007; Vardas et al,
Rawlings-Anderson, 2007). Once the lead(s) are in place, 2007; Timperley et al, 2008
a fibrotic capsule forms over the top of the lead fixing
them firmly in place (Kenny, 2007).
The choice of pacemaker used depends on many factors,
Indications for insertion including the underlying pathophysiology, patient’s age,
The most basic function of a PPM is to act as a safety net, activity level and underlying rhythm (Kumar and Clark,
to protect patients who are at risk from symptomatic 2003; Vardas et al, 2007).
bradycardia (Houghton and Grey, 2003). Permanent pace-
maker insertion is also indicated for patients with any of Pacemaker codes
the conditions summarized in Table 1. A generic shorthand method of referring to pacemakers
came into being in 1974 and has been revised several
Types of permanent pacemaker
There are several types of PPM available that pace the dif-
a) b)
ferent chambers of the heart, commonly known as:
w Atrial pacing
w Ventricular pacing
w Dual chamber/sequential A-V pacing
w Bi-ventricular pacing

c) d)

Figure 4. Lead positions for different types of pacing:


atrial (a), ventricular (b), dual chamber/sequential A-V
Figure 3. A single lead positioned in the right ventricle (c) and bi-ventricular (d)

British Journal of Cardiac Nursing June 2009 Vol 4 No 6 263


Clinical

Table 2.
NBG pacing codes
1 2 3 4 5
Chamber(s) paced Chamber(s) sensed Sensing response Rate modulation Multisite pacing
O = None O = None O = None O = None O = None
A = Atrium A = Atrium T = Triggered R = Rate modulation A = Atrium
V = Ventricle V = Ventricle I = Inhibited V = Ventricle
D = Dual (A+V) D = Dual (A+V) D = Dual (T+I) D = Dual (A+V)
Adapted from Kenny, 2007

times since by different organizations until the current heart rhythm) (Mills, 2005). This type of pacemaker
codes devised by the North American Society of Pacing requires the patient to have an intact conduction system
and Electrophysiology and British Pacing and and functioning AV node (Mills, 2005).
Electrophysiology Group came into use (usually referred Single chamber pacemakers that pace the ventricle (e.g.
to as the NBG codes) (Kenny, 2007) (Table 2). VVI) work by pacing and sensing the ventricle. The pri-
The first letter in the code corresponds to the chamber mary benefits are for patients who require intermittent
being paced, this can be the atrium, ventricle or both. The pacing or those with complete heart block (Kenny, 2007).
second letter corresponds to the chamber from which the This type of pacemaker does not affect atrial activity, so
impulses are sensed. The third letter indicates the response the ‘atrial kick’ contribution to the cardiac output (an extra
to the sensed signal. The fourth and fifth columns are 15-30%) may be lost (Mills, 2005).
additional programmable features. For example a code This is seen in Figure 5, which shows a patient with
such as VVIR implies that the ventricle is paced, the ven- intrinsic rhythm and ventricular pacing. The line below
tricle is sensed and the pacemaker will not active if it the ECG is the pressure waveform, which is dampened
detects the patients own intrinsic rhythm and the rate will when the ventricles are paced, showing the effect of loss of
change depending on the needs of the patient (Mills, ‘atrial kick’ on the patients pressure.
2005). Rate response adjusts the rate of pacing based on Another problem with this lack of synchronization is
the activity level of the patient, pacemakers can sense this that mitral or tricuspid insufficiency could develop. This
increase in activity in a variety of ways, including; vibra- type of pacemaker is normally prescribed for patients who
tion, increase in respiration rate and changes in blood lead a more sedentary lifestyle (Mills, 2005).
temperature (Houghton and Grey, 2003).
Dual chamber
Single chamber Dual chamber pacing has been shown to produce small
Single chamber pacemakers will pace either the atrium or but potentially significant benefits for patients with sick
the ventricle. For example an on-demand atrial pacemaker sinus syndrome (SSS) and/or atrioventricular (AV) block,
(AAI) will pace and sense the atrium. Pacing is inhibited when compared with single chamber ventricular pace-
when it detects the patient’s intrinsic rhythm (natural makers (Castelnuovo et al, 2005). Dual chamber pacemak-
ers are often given to more active or younger patients
(Castelnuovo et al, 2005). A systematic review found a
significant statistical improvement in patients’ exercise
capacity when given dual chamber pacemakers as opposed
to single chamber pacemakers. (Castelnuovo et al, 2005).
Dual chamber pacemakers respond to each atrial stimu-
lation and activate the ventricle when this does not occur
naturally ensuring A-V synchrony (Mills, 2005).

Bi-ventricular pacing
Also known as cardiac resynchronization therapy (CRT) is
used for the treatment of heart failure when a patient ful-
fills the required criteria (Table 3). Underwood (2004)
describes heart failure as a relatively common condition
with a poor prognosis, characterized by reduced cardiac
output resulting in arterial under-filling, causing the body
Figure 5: Ventricular pacing with intrinsic rhythm showing loss of ‘atrial to retain fluid, thus increasing fluid volume. This can lead
kick’ when pacing. to systemic venous congestion and pulmonary oedema.

264 British Journal of Cardiac Nursing June 2009 Vol 4 No 6


Clinical

Table 3.
Patient criteria for cardiac
resynchronization therapy
w New York Heart Association (NYHA) symptoms,
class III to IV
w Patients in sinus rhythm
With QRS on ECG ≥ 150 ms
Or QRS on ECG of 120–149 ms with mechanical
dyssyncrony confirmed by echocardiography
w Left ventricular ejection fraction (LVEF) ≤ 35%
w Currently receiving optimal pharmacological
therapy
National Institute for Health and Clinical Excellence, 2007

CRT uses three leads placed in the right atrium, the right Figure 6. Fluoroscopic image of bi-ventricular lead
ventricle, and a third passed through the coronary sinus to positions
pace the left ventricle from behind (Johnson and Rawlings-
Anderson, 2007) (Figure 6). Leads are not passed directly complexes on the ECG as the electrical impulse does not
into the left ventricle as the risk of thromboembolism and/ travel down the normal fast conduction pathway
or impairment to the aortic valve, pose an unacceptable (Houghton and Grey, 2003).
risk to the patient (Timperley et al, 2008). CRT attempts to
improve the efficiency of the pumping action of the heart Bi-polar/uni-polar leads
by ‘re-synchronizing’ the pumping action of the hearts All electrical circuits have two poles: a positive pole called
chambers (Fox et al, 2007). This works by improving ven- an anode, and a negative pole called a cathode. Uni-polar
tricular contraction by stimulating the right and left ventri- leads have a pole on the lead itself whereas bi-polar leads
cle, improving cardiac output, diastolic left ventricle filling have two poles on the lead. The large circuit made by the
and ejection fraction (Timperley et al, 2008). uni-polar lead causes a large pacing spike on the surface
ECG, the smaller bi-polar circuit causes a smaller pacing
Recognizing pacemakers on the ECG spike on the ECG (Kenny, 2007).
The main indicator of a pacemaker on an ECG is the pres-
ence of pacing spikes, otherwise known as pacing artifact. Preparation of the patient
These spikes occur as the pacemaker sends an electrical The procedure should be fully explained to the patient and
impulse to the heart muscle (Mills, 2005). The location of any questions or concerns addressed. Timperley et al
the pacing spike(s) shows which chambers are being (2008) state that before insertion of a PPM, patients
paced (Figure 7). Ventricular pacing causes broad QRS should be prepared for the procedure by taking bloods,

Pacing spike

a. Single chamber ventricular pacing, pacing spike located before QRS complex. The underlying
rhythm is complete heart block.

b. Dual chamber pacemaker, pacing spike located both before P wave and before QRS complex

Figure 7. Pacing spikes on a rhythm strip

British Journal of Cardiac Nursing June 2009 Vol 4 No 6 265


Clinical

may vary depending on the individual patient’s risk


assessment.
The PPM is normally positioned in the non-dominant
side. For example, if the patient is right-handed then the
PPM is positioned in the left upper chest and vice versa
Subclavian (Johnson and Rawlings-Anderson, 2007).
vein
Cephalic vein
Where cardiac monitoring is used before implantation,
the sticky residue left behind from the electrodes can be
removed with acetone. Alternatively the electrodes can be
repositioned away from the implant area to avoid the need
to do this.

The implantation procedure


PPMs are normally inserted in theatre or a cardiac cathe-
terization laboratory, with the patient being fully con-
scious, and optimal sterility maintained throughout the
Figure 8. The subclavian and cephalic veins are used for procedure (Johnson and Rawlings, 2007). The implanta-
transvenous access tion procedure normally lasts between 60 and 90 minutes
depending on the type of pacemaker inserted and indi-
including full blood count, urea and electrolytes and an vidual patient factors (Kumar and Clark, 2003).
international normalized ratio (INR) if appropriate (i.e. if During the procedure the patient’s ECG, blood pressure
the patient is on anticoagulants?) and oxygen saturation are measured non-invasively (Johnson
The patient should also have a 12-lead ECG, a valid and Rawlings-Anderson, 2007). The patient’s non dominant
consent form, and intravenous access for delivery of pro- side is used for implantation. Transvenous access is nor-
phylactic antibiotics and any sedation that may be required. mally achieved via the subclavian or cephalic veins (Kenny,
Prophylactic antibiotics (i.e. flucloxacillin/vancomycin 1g) 2007), located just below the collar bone (Figure 8). The leads
aid in the reduction and prevention of PPM infections are positioned using fluoroscopy to guide them into the cor-
(Johnson and Rawlings-Anderson, 2007). rect position (Kumar and Clark, 2003) (Figure 9).
Swanton and Banerjee (2008) state that aspirin and/or The permanent pacemaker pocket is made for the pulse
clopidogrel are also stopped if possible for around three generator, usually located sub-cutaneously in front of the
to four days before the implantation procedure. Warfarin pectoral muscle (Kumar and Clark, 2003) (Figure 9).
can be stopped four to five days prior to procedure. The Sometimes for young and/or thin females, the pocket can
INR should be ≤1.5 to prevent a haematoma forming be located in the retro-mammary or axillary position
around the pacemaker box. Warfarin is normally restart- (Kenny, 2007).
ed the same night as the procedure (Swanton and Following the procedure patients should be monitored to
Banerjee, 2008). The decision to stop warfarin or aspirin establish underlying rhythm and to highlight any pacing
irregularities. In addition the patient should also receive
regular observations (blood pressure, SpO2, heart rate, tem-
perature and respiratory rate) and regular checks of the

Pacemaker pocket

Figure 9. Fluoroscopic image of dual chamber pacing


leads positioned in the right atrium and ventricle Figure 10. The permanent pacemaker pocker

266 British Journal of Cardiac Nursing June 2009 Vol 4 No 6


Clinical

wound to monitor for signs of infection or haematoma


(Johnson and Rawlings-Anderson, 2007). Pacing in ventricle
The implant wound can be closed in a variety of ways
including staples, glue, dissolvable or non-dissolvable
sutures. Local trust policy should always be followed. Any
non-dissolvable closure methods such as staples or non-
dissolvable sutures are usually removed at the local prac- P wave sensed Retrograde
tice surgery around 7-10 days after the procedure. in atrium conduction
Patients are usually followed up in pacing clinics once or
twice a year post-procedure (Stain, 2008). The frequency
of follow-up for each patient will depend on factors such
as his/her overall clinical condition and the type of pace-
maker implanted (more complex pacemakers may need Atrial contraction
more frequent visits to adjust the PPM’s programming)
(Vardas et al, 2007). Figure 11: Pacemaker-mediated tachycardia (adapted
from Bowbrick and Borg, 2006)
Post-insertion complications associated
with permanent pacemakers period for one cycle to interrupt the tachycardia (Bennett,
Pneumothorax and haemothorax 2006). PMT may cause haemodynamic instability.
There is a small risk of pneumothorax (1–2%) when the Appropriately-trained staff can use a magnet to stop atrial
subclavian vein is used for lead access (Timperley et al, sensing, or alternatively the pacemaker can be repro-
2008). Breathlessness and increased pain on inspiration grammed (Bowbrick and Borg, 2006).
may occur following pneumothorax/haemothorax. Other
symptoms can include an increase in respiratory rate and Pacing of the diaphragm
cyanosis (Leach, 2004). This can lead to the need to insert In some cases the pacemaker may stimulate the patient’s
a chest drain (Johnson and Rawlings-Anderson, 2007). diaphragm due to the proximity of the electrical signal to
The chest x-ray is used in addition to fluoroscopic imag- the phrenic nerve, causing hiccoughs (Mills, 2005).
ing—as it is hard to identify a pneumothorax/haemotho-
rax when the patient is lying down an upright x-ray should Cardiac tamponade
be used (O’Grady, 2007). Direct trauma inflicted during the implant procedure can
cause the pericardial sac to fill with fluid, causing elevated
Infection pressure, which in turn prevents the ventricles filling
Infection of the wound may occur. Many clinicians attempt properly and can reduce the stroke volume. This is a
to further reduce the risk of infection by adding antibiotics medical emergency and should be treated immediately,
to the pacemaker pocket, such as gentamicin 80 mg usually via pericardiocentesis (passing a needle into the
(Timperley et al, 2008). As a result of a low grade infection pericardium and aspirating the fluid). Signs and symp-
or other mechanical means, a pacemaker may start to pro- toms include persistent hiccoughs, reduced pulse strength
trude through the skin (termed erosion) (Kumar and Clark, during inspiration, reduced blood pressure, cyanosis, dis-
2003). Monitoring the patient’s vital signs including tem- tended jugular veins, reduced urinary output and distant-
perature accompanied by regular checks of the wound site sounding heart sounds (O’Grady, 2007).
for swelling, pain, redness, inflammation or weeping can
alert the nurse to signs of an infection. Pocket haematoma
Bleeding into the pacemaker pocket can occur. To reduce
Reentrant tachycardia the risk of this the patient’s INR should be less than 1.5
Pacemaker-mediated tachycardia (PMT) is a problem before implantation unless there are clinical reasons that
associated with dual-chamber pacemaker systems. A retro- require a higher INR (e.g. mechanical valve) (Kumar and
grade (from ventricle to atrium) conduction triggers a loop Clark, 2003).
of events resulting in a sustained tachycardia (Bowbrick
and Borg, 2006) (Figure 11). This form of tachycardia is not Lead displacement
caused by the pacemaker but owing to the retrograde con- Before the fibrotic capsule forms to better hold the lead in
duction (backward impulse) the pacemaker acts as a path place, the lead can become displaced, causing the pace-
for reentry of the electrical impulse, generating a never- maker not to function as intended. Most patients receive a
ending loop of tachycardia. This is usually triggered by a chest x-ray and pacing check to confirm lead position
premature ventricular contraction (Kenny, 2007). before discharge (Kumar and Clark, 2003).
PMT can be seen on the ECG by closely grouped paced
complexes and can usually be prevented by prolonging the Pacing problems
atrial refractory (recovery) period. Many pacemakers can There are also various problems associated with pace-
detect PMT and automatically prolong the atrial refractory maker malfunction (Table 4). To this end the patient is

British Journal of Cardiac Nursing June 2009 Vol 4 No 6 267


Clinical

Table 4.
Common pacemaker malfunctions

Problem Explanation Effect on patient Identification


Failure Pacemaker Can expose patient to Presence of a pacing spike on ECG without
to produces an the condition the appropriate atrial or ventricular response
capture impulse but pacemaker was inserted
fails to pace to treat
the myocardium

Failure No impulse is Can lead to reduced No pacing spike or subsequent P/QRS wave where
to pace generated by cardiac output and expected
the pacemaker subsequent drop in
blood pressure

Under- The pacemaker The patient may be Pacing spikes seen on the ECG at various points in
sensing fails to detect aware of palpitations or the cardiac cycle not before P waves or QRS waves
the intrinsic missed beats. If pacing as would be expected
cardiac activity spikes fall on T waves
and paces this can be dangerous
anyway and cause ventricular
tachycardia or fibrillation
(VT/VF) requiring
emergency treatment
Adapted from Mills, 2005 and Kenny, 2007

usually placed on a cardiac monitor overnight and receives that no driving is allowed for 1 week post-insertion (DVLA
regular observations (blood pressure, pulse etc) to ensure need not be informed (DVLA, 2009)). Patients with a group
the pacemaker is functioning as intended and identify any 2 licence (formerly LGV/PCV licence) should not drive for
potential haemodynamic compromise (O’Grady, 2007). 6 weeks following PPM insertion (the DVLA should be
When presented with any of these problems it is impor- informed (DVLA, 2009) (Timperly et al, 2008).
tant to monitor the patient’s vital signs. A 12-lead ECG Patients should also be advised to keep the wound dry
should also be obtained. for 5–7 days or until the wound is both dry and healed.
Placing a magnet over the pulse generator reverts the
pacemaker to asynchronous mode and can be used by Table 5.
appropriately-trained personnel to determine the battery
life of the pulse generator, to allow the use of emergency
Clinical equipment affecting
pacing if the pacemaker is malfunctioning, identify which permanent pacemakers
chambers are being paced and assess capture. Magnet
application can also be used to pace a patient despite the w Defibrillators (position defibrillator pads or
presence of electromagnetic interference (Mills, 2005). paddles away from pulse generator site, check
PPM post defibrillation)
Clinical considerations w Diathermy (can cause inhibition of the PPM)
Some clinical equipment and/or tests can affect the func- w Magnetic resonance imaging (MRI) (usually
tioning of PPMs. Health professionals caring for PPM contraindicated)
patients should be aware of these possible effects and take
w Radiotherapy (can cause damage to PPM circuits)
steps to reduce risks to PPM patients (Table 5).
w TENS (transcutaneous electrical nerve
Advice to patients stimulation)
Patients living with a PPM should receive health education Mills, 2005; Johnson and Rawlings-Anderson, 2007
pertaining to the pacemaker. Current UK driving law states

268 British Journal of Cardiac Nursing June 2009 Vol 4 No 6


Clinical

Patients should also be advised to see a relevant health


professional if there are any signs of fever, infection or
pain/swelling/heat at the wound site (Timperley et al,
Key Points
w Many nurses in a variety of clinical areas are exposed to patients
2008). Vigorous activity and contact sports should also be
with a permanent pacemaker
avoided following insertion of a PPM (Swanton and
Banerjee, 2008). w A working knowledge of the different types of PPM device can help
If non-dissolvable stitches or staples are used to close the nurse to care for a patient more effectively
the wound the patient should be advised to have them
w An awareness of what affects a PPM can help health professionals
removed (usually by a GP or practice nurse/district nurse)
to take steps to prevent or reduce risks to patients with a PPM, both
in 7–10 days (Timperley at al, 2008).
in and out of the clinical environment
Patients should carry their pacemaker card detailing the
type of pacemaker they have. This card should be pre- w ECG recognition of pacemakers can help to identify pacing problems
sented to doctors and/or dentists prior to any procedure
(Johnson and Rawlings-Anderson, 2007).
Bennet DH (2006) Cardiac Arrhythmias: Practical Notes on Interpretation
Sources of electromagnetic interference can cause the and Treatment. 7th edn. Hodder Arnold, London
pacemaker to revert to asynchronous pacing mode, prompt British Heart Foundation (2008) Living with a heart condition:
Pacemakers. http://tinyurl.com/qsh737(accessed 19 May 2009)
unnecessary pacing, or even mimic cardiac activity (Mills, Bowbrick S, Borg A (2006) ECG complete. Churchill Livingstone, Edinburgh
2005; Schoenfeld, 2007). Patients exposed to any sources Castelnuovo E, Stein K, Pitt M, Garside R, Payne E (2005) The effective-
of electromagnetic interference that could affect the func- ness and cost-effectiveness of dual-chamber pacemakers compared with
single-chamber pacemakers for bradycardia due to atrioventricular
tioning of the PPM should not linger near these sources or block or sick sinus syndrome: systematic review and economic evalua-
expose themselves without due reason (Table 6). Pacemaker tion. Health Technol Assess 9(43)
patients should also be advised to avoid areas displaying a Conover MB (2003) Understanding Electrocardiography. 8th edn. Mosby, St
Louis
‘no pacemakers’ sign. Davies A (2007a) Recognizing and reducing interference on 12-lead elec-
trocardiograms. Br J Nurs 16(13): 800-4
Davies A (2007b) Calculating cardiac axis on a 12-lead ECG. British
Conclusions Journal of Cardiac Nursing 2(11): 513–7
There are a number of arrhythmic conditions that may Driver and Vehicle Licensing Agency (2009) About Medical Standards for
mean a patient is prescribed a permanent pacemaker, and Drivers http://tinyurl.com/olvppq (accessed 19 May 2009)
Ector H and Vardas P for European Heart Rhythm Association, European
the specific type of PPM, functionality and position of the Society of Cardiology (2007) Current use of pacemakers, implantable
leads will vary. However, the cautions and considerations Cardioverter defibrillators and resynchronization devices: data from the
that apply to all patients with a PPM implanted should be registry of the European Heart Rhythm Association. Eur Heart J
9(suppl 1): 144–9
kept in mind in order to provide effective care. BJCardN Fox M, Mealing S, Anderson R et al (2007) The clinical effectiveness of
cardiac resynchronization (biventricular pacing) for heart failure: sys-
Acknowledgments tematic review and economic model. Health Technol Assess 11(47)
The author would like to thank Dr M Ghandi and Jenny Tagney, and Kenny T (2007) The Nuts and Bolts of Cardiac Pacing. Blackwell Futura,
Medtronic for the image of the pulse generator in Figure 2. Oxford
Khan E (2006) The pathological origins of arrhythmias. British Journal of
Cardiac Nursing 1(9): 408–17
Kumar P, Clark M (2003) Clinical Medicine. 5th edn. WB Saunders,
Table 6. Edinburgh
Leach R (2004) Critical Care Medicine at a Glance. Blackwell Publishing,
Potential sources of electromagnetic Oxford
National Institute for Health and Clinical Excellence (2005) Dual-chamber
interference to be avoided Pacemakers for Symptomatic Bradycardia due to Sick Sinus Syndrome and/
or Atrioventricular Block [Technology appraisal 88]. NICE, London
w Airport weapons detectors National Institute for Health and Clinical Excellence (2007) Cardiac
w CB radio-transmitting systems Resynchronization Therapy for the treatment of Heart Failure.
[Technology appraisal 120]. NICE, London
w Some (ultrasonic) dental drills O’Grady E (2007) A Nurse’s Guide to Caring for Cardiac Intervention
Patients. John Wiley & sons Ltd, Sussex
w Shop anti-theft systems Mills EJ (2005) ECG Interpretation: A 2-in-1 Reference for Nurses.
Lippincott, Williams and Wilkins, Philadelphia
w Mains-driven electric motors Schoenfeld MH (2007) Contemporary pacemaker and defibrillator device
(especially if sparking/faulty) therapy: Challenges confronting the general cardiologist. Circulation
115: 638-653
w Magnets Stain N (2008) Home monitoring for ICDs and Pacemakers. Coronary
w Vibrations (helicopters/hovercrafts) Heart 10: 24–7
Swanton RH, Banerjee S (2008) Swanton’s Cardiology. 6th edn. Blackwell
w High-powered radar stations Publishing, Massachusetts
Timperley J, Leeson P, Mitchell RJ, Betts T (2008) Pacemakers and ICDs.
w Mobile phones—patients normally advised to use Oxford University Press, Oxford
on opposite side to pulse generator (i.e. if Underwood JCE (2004) General and Systematic Pathology. 4th edn.
Churchill Livingstone, Edinburgh
implanted on right side would use left arm/ear) Vardas PE, Auricchio A, Blanc JJ et al for European Society of Cardiology;
and to keep phones at least 6 inches/15 cm European Heart Rhythm Association (2007) Guidelines for cardiac pac-
away from the pulse generator ing and cardiac resynchronization therapy: The Task Force for Cardiac
Pacing and Cardiac Resynchronization Therapy of the European Society
Mills, 2005; Schoenfeld, 2007 of Cardiology. Developed in collaboration with the European Heart
Rhythm Association. Eur Heart J 28(18): 2256-95

British Journal of Cardiac Nursing June 2009 Vol 4 No 6 269

View publication stats

You might also like