You are on page 1of 5

Laboratory Performance in Neonatal Bilirubin Testing

Using Commutable Specimens


A Progress Report on a College of American Pathologists Study
Stanley F. Lo, PhD; Bernadine Jendrzejczak, BA; Basil T. Doumas, PhD

● Context.—In 2003 the Chemistry Resource Committee of variation for the 4 major instrument groups (Dimension,
the College of American Pathologists introduced a com- Olympus, Synchron, and Vitros), which report 65% of all
mutable specimen in the Neonatal Bilirubin Surveys. This results, varied from 2% to 3%. College of American Pa-
specimen was intended to help evaluate all bilirubin meth- thologists All Data mean bilirubin values were within 0.46
ods. mg/dL (7.8 ␮mol/L) of the reference method mean in
Objective.—To evaluate the effect of commutable spec- 2003; in subsequent years these differences became larger,
imens on the performance of selected clinical analyzers in peaking at 1.87 mg/dL (32 ␮mol/L) in 2005.
measuring neonatal bilirubin from 2003 through 2006. Conclusions.—The large systematic error of bilirubin
Design.—A human serum–based specimen enriched measurements is due primarily to failure of instrument
with unconjugated bilirubin in human serum has been in- manufacturers to produce reliable bilirubin calibrators.
cluded since 2003 in the Neonatal Bilirubin Surveys. The Primary calibrators should consist of human serum en-
bilirubin values of these specimens were determined by the riched with unconjugated bilirubin. Bilirubin values must
reference method and used to evaluate results reported by be assigned by the reference method, the performance and
various chemistry analyzers. robustness of which are reported in this article. Secondary
Results.—Coefficients of variation for College of Amer- calibrators distributed to users must be traceable to pri-
ican Pathologists All Data ranged from 4.9% to 6.2% for mary calibrators.
the Neonatal Bilirubin Survey. However, coefficients of (Arch Pathol Lab Med. 2008;132:1781–1785)

mg/dL (428 ␮mol/L) cannot be reduced by photothera-


I n 2003, the Chemistry Resource Committee of the Col-
lege of American Pathologists (CAP), acting on a report
regarding inaccuracies in the measurement of bilirubin in
py.2 Because this is considered a critical bilirubin value,
the CAP Chemistry Resource Committee decided to adjust
clinical laboratories,1 introduced a new specimen in the the bilirubin level in the human serum–based specimen
Neonatal Bilirubin Surveys (NB-Surveys). This specimen, near 25 mg/dL (428 ␮mol/L).
human serum enriched with unconjugated bilirubin, was Results from the 2003 human serum–based specimen
expected to be commutable (free of matrix effects) because (NB-02) were very encouraging. The grand mean value for
it closely resembles specimens drawn from healthy neo- CAP All Data was only 2.4% higher than the reference
nates. Properly calibrated clinical analyzers were expected method mean and the coefficient of variation (CV) % of
to be accurate, providing total bilirubin (TBIL) values close 5.7.
to those obtained by the reference method. The noncom- This communication is a progress report on the perfor-
mutable or conventional NB-Survey specimens consisted mance of laboratories participating in the NB-Surveys for
of unconjugated bilirubin and ditaurobilirubin in bovine the years 2004 through 2006. The practice of including the
serum.1 special, commutable NB-Survey specimen in the Chemis-
A guideline issued by the American Academy of Pedi- try Survey (C-Survey), as specimen C-96 or C-97, contin-
atrics recommended a blood exchange transfusion for in- ued until the end of year 2006.
fants older than 48 hours when bilirubin levels near 25
MATERIALS AND METHODS
The preparation and distribution to participating laboratories
Accepted for publication April 2, 2008.
of the human serum–based specimens has been described pre-
From the Reference Standards Laboratory, Department of Pathology
and Laboratory Medicine, Children’s Hospital of Wisconsin, Milwau-
viously.3 All NB-Survey and C-Survey specimens (from 2003
kee (Drs Lo and Doumas and Ms Jendrzejczak); and the Department through 2006) were analyzed for TBIL by the reference method4
of Pathology, Medical College of Wisconsin, Milwaukee (Drs Lo and at the Reference Standards Laboratory of the Children’s Hospital
Doumas). of Wisconsin, Milwaukee. All other bilirubin results shown in this
The authors have no relevant financial interest in the products or article have been reported in the participant summary report.
companies described in this article. Instruments specifically considered in this report include the BR2
Reprints: Stanley F. Lo, PhD, Reference Standards Laboratory, Chil- (Advanced Instruments Inc, Norwood, Mass), Advia (Siemens
dren’s Hospital of Wisconsin, 9000 W Wisconsin Ave, Wauwatosa, WI Medical Solutions Diagnostics, Tarrytown, NY), Aeroset and Ar-
53226 (e-mail: slo@mcw.edu). chitect (Abbott Diagnostics, Abbott Park, Ill), Cobas Integra,
Arch Pathol Lab Med—Vol 132, November 2008 Laboratory Performance in Neonatal Bilirubin Testing—Lo et al 1781
Table 1. College of American Pathologists (CAP) All
Data and Reference Method Mean Bilirubin Values
and Coefficient of Variation (CV)% for the Neonatal
Bilirubin Survey and Chemistry Survey Specimens
CAP All Data Reference
Mean Bilirubin Method Value, No. of
Specimen Value, mg/dL CV% mg/dL Laboratories
2003
NB-02 19.90 5.7 19.44 1743
C-96 19.92 5.6 19.44 4982
2004
NB-06 23.69 5.2 22.06 1784
C-97 23.41 6.4 22.06 4757
2005
Figure 1. College of American Pathologists (CAP) Neonatal Bilirubin
NB-05 26.25 6.2 24.38 1848
and Chemistry Surveys 2003–2006. Differences between CAP All Data
C-96 26.35 7.0 24.38 4552
and reference method mean values in milligrams per deciliter (1 mg/
2006 dL bilirubin ⫽ 17.1 ␮mol/L bilirubin).
NB-01 22.96 5.2 21.19 1890
C-97 22.44 5.5 21.19 4775
NB-11 22.32 4.9 21.04 1718
tween the NB-01 and NB-11 CAP All Data bilirubin mean
values (Figure 2) is due to large shifts in the mean values
reported by the Cobas Integra, Dimension, Roche Modu-
Roche Modular/Hitachi (Roche Diagnostic Systems, Branchburg, lar/Hitachi, and Vitros, which constitute 60% of the par-
NJ), Dimension (Dade Behring, Newark, Del), Olympus (Olym-
pus America Inc, Melville, NY), Synchron (Beckman Coulter Inc,
ticipating laboratories. Discrepancies between NB-01 and
Fullerton, Calif), Unistat (Leica Microsystems Inc, Buffalo, NY), C-97 (Figure 2) are primarily due to the large percentage
and Vitros (Ortho-Clinical Diagnostics, Raritan, NJ). of Vitros participants in the C-Survey.
According to the participant summary report, methods based Figure 3 shows differences between bilirubin mean val-
on the coupling of bilirubin with a diazo compound are used in ues for selected field methods and the reference method.
the Advia, Aeroset, Architect, Cobas Integra, Dimension, Roche With 3 exceptions, Advanced BR2, Aeroset (2003–2004),
Modular/Hitachi, Olympus, Synchron, and Vitros (TBIL slide). and Architect (2006), mean bilirubin values for all other
Methods based on the oxidation of bilirubin are used in the Ad- instrument groups are higher than those of the reference
via (vanadate oxidation) and Synchron (bilirubin oxidase). Meth- method. Note that there were no data from the Abbott
ods based on direct spectrophotometry are used in the Aeroset,
instruments in 2005, and both Abbott instruments use the
Architect, Advanced BR2, Unistat, and Vitros neonatal bilirubin
(BuBc) slide, which measures the sum of unconjugated bilirubin same diazo and direct spectrophotometry methods. The
and conjugated bilirubins (bilirubin monoglucuronide and diglu- positive bias ranges from less than 1 mg/dL to 4 mg/dL
curonide). According to Abbott, both the Aeroset and Architect (17.1 to 68.4 ␮mol/L). The Synchron group consistently
methods use a diazo method (diazotized 2,4-dichloroaniline), but shows a small positive bias. The negative bias of the BR2
for the NB-Survey users prefer direct spectrophotometry. has decreased over time and is smallest in 2006.
Reference methods are considered the gold standard
RESULTS against which field methods are compared and evaluated.
Table 1 shows CAP All Data mean bilirubin values and It is important, therefore, to establish the long-term per-
CV% for the NB-Survey and C-Survey specimens consist- formance characteristics of the reference method for serum
ing of unconjugated bilirubin in pooled human sera sent TBIL. From September 2003 to April 2007, the Reference
to participating laboratories during 2003–2006; also shown Laboratory of the Children’s Hospital of Wisconsin pre-
are values obtained by the reference method. For each year pared 11 bilirubin standard solutions by adding National
from 2003 through 2006, one commutable sample was pre- Institute of Standards and Technology Bilirubin, Standard
pared and sent out as a specimen in the NB-Survey and Reference Material 916, to pooled human sera to a con-
C-Survey; thus, in 2003, NB-02 was identical to C-96; in centration of 20 mg/dL (342 ␮mol/L). Following prepa-
2004, NB-06 was identical to C-97; in 2005, NB-05 was ration, these solutions were analyzed by the reference
identical to C-96; and in 2006, NB-01, C-97, and NB-11 method and the molar absorptivity (␧) of the alkaline azo-
were identical. Shown in Figure 1 are differences between pigment was compared with the values established in 2
mean values of CAP All Data and those of the reference round robin studies with participation from national and
method for NB-Survey and C-Survey specimens 2003– international laboratories.5,6 Collaborating laboratories in
2006. The best overall laboratory performance was the 1988 round robin study were from the United States
achieved in 2003. The CAP All Data mean value for NB- (3) and The Netherlands (2); in the 1998 round robin
02 was identical to that for C-96 and the difference, 0.46 study, laboratories were from the United States (3), Ger-
mg/dL (7.9 ␮mol/L) or 2.4%, between the CAP All Data many (2), and The Netherlands (1). Data from the 2 round
mean and that of the reference method was the smallest robins and from the Children’s Hospital of Wisconsin Ref-
in 4 years.3 erence Laboratory, shown in Table 2, indicate that the es-
In 2006, the commutable specimen was mailed to par- tablished molar absorptivity of the alkaline azopigment,
ticipants on 3 occasions, in April and November with the because of its reproducibility and narrow range, can and
NB-Survey (specimens NB-01 and NB-11, respectively) should be used to assess the accuracy of bilirubin calibra-
and in June with the C-Survey (specimen C-97). The un- tors consisting of unconjugated bilirubin in human sera.
usually large difference, 0.64 mg/dL (10.9 ␮mol/L), be- The 20-mg/dL (342-␮mol/L) stock solutions are dis-
1782 Arch Pathol Lab Med—Vol 132, November 2008 Laboratory Performance in Neonatal Bilirubin Testing—Lo et al
Figure 2. College of American Pathologists Neonatal Bilirubin Surveys 2006. Shown are differences between total bilirubin mean values (milli-
grams per deciliter) for specimens NB-01 and NB-11 (yellow bars) and between specimens NB-01 and C-97 (red bars) (1 mg/dL bilirubin ⫽ 17.1
␮mol/L bilirubin).
Figure 3. College of American Pathologists (CAP) Neonatal Bilirubin Surveys 2003–2006. Shown are differences between total bilirubin mean
values (milligrams per deciliter) for selected field and reference methods. The numbers on the horizontal axis identify different instruments: 1
indicates CAP All Data; 2, Advanced BR2; 3, Advia; 4, Aeroset (2003 and 2004)/Architect (2006); 5, Cobas Integra; 6, Cobas Integra, DZ Salt; 7,
Dimension; 8, Hitachi, Roche Modular; 9, Olympus; 10, Synchron Diazo; 11, Synchron BO; 12, Unistat; 13, Vitros 250/350; 14, Vitros 950; and
15, Vitros 5,1 FS (1 mg/dL bilirubin ⫽ 17.1 ␮mol/L bilirubin).

pensed in airtight plastic tubes and kept at ⫺70⬚C until dL [85.5 ␮mol/L], and 10 mg/dL [171 ␮mol/L]) are an-
use. Each time the reference method is used, the stock alyzed to construct a calibration curve, the regression
solution is analyzed to verify that its ␧ value is within the equation of which is used to calculate the results of the
accepted limits (mean value ⫾ 3 SD). Next, the stock so- specimens analyzed. Shown in Table 3 are molar absorp-
lution and 3 dilutions (2.5 mg/dL [42.7 ␮mol/L], 5.0 mg/ tivities, slopes, and intercepts of calibration curves ob-
Arch Pathol Lab Med—Vol 132, November 2008 Laboratory Performance in Neonatal Bilirubin Testing—Lo et al 1783
Table 2. Molar Absorptivity of Bilirubin Alkaline Table 3. Long-term Calibration Stability and
Azopigment Results From 2 Round Robins and the Precision of the Reference Method for Total Bilirubin
Reference Laboratory of Children’s Hospital of
Calibration Curve
Wisconsin (CHW)*
(September 15, 2003–April 23, 2007)
n ␧ (L· mol⫺1 · cm⫺1)† SD
␧ (L· mol⫺1 · cm⫺1)* Slope Intercept
Round Robin 1988 (5 labs) 10 76 490 610
Mean 76 554 0.0770 ⫺0.0003
Round Robin 1998 (6 labs) 12 76 640 601
SD† 377 0.0004 0.0018
Reference Lab CHW 11 76 869 589
N 45 46 46
* n indicates number of determinations; ␧, molar absorptivity at 598 Min 75 579 0.0759 ⫺0.0038
nm. Max 77 257 0.0782 0.0043
† Corrected for purity (98.3%) of bilirubin.
* ␧ indicates molar absorptivity.
† One standard deviation for a single measurement.

tained during a 3.5-year period, and precision data on


controls prepared in our laboratory. The data in Tables 2 Accuracy
and 3 demonstrate the impressive stability and robustness The goal of limiting the total error to ⫾10% of the ref-
of the reference method in US and European laboratories. erence method value3 can be realized only by improving
In 2003, the CAP All Data mean values for NB-02 and accuracy. Since 2003 the gap between the reference values
C-96 were identical and within 2.4% of the reference value and the CAP All Data has progressively increased (Table
(Figure 1). In our opinion this was due to the use of cal- 1); this applies to most instrument groups with notable
ibrators with accurately assigned bilirubin values. In sub- exceptions of Synchron, and lately Advanced BR2 and
sequent years accuracy deteriorated. Results for 2006 are Cobas Integra. To state simply: Calibrators used by most
strange; there is good agreement for CAP All Data means manufacturers are inaccurate, that is, the assigned values
between specimens C-97 and NB-11, but the mean value are much higher than the actual bilirubin concentrations.
for NB-01 is greater than NB-11 by 0.64 mg/dL (10.9 Because for most major instrument groups precision is
␮mol/L). The lower mean bilirubin value for NB-11 was quite good (CVs, 2%–3%), further improvement in accu-
due to the downward shift of 4 instrument group mean racy would depend on the use of accurate calibrators.
values (Cobas Integra, Dimension, Hitachi/Roche Modu- Instruments from the same manufacturer using meth-
lar, and Vitros, which represent 59% of the reported re- ods based on different principles exhibit the widest vari-
sults) (Figure 2). The close agreement between the mean ation in results. Examples are as follows: (1) By direct
values for C-97 and NB-11, 22.44 mg/dL (383.7 ␮mol/L) spectrophotometry (neonatal bilirubin method), differenc-
and 22.32 mg/dL (381.7 ␮mol/L), indicates a small, if any, es between the reference method results and those re-
loss of bilirubin, in the 6-month period (June–November). ported by the Vitros are 2.74 mg/dL (49.9 ␮mol/L) (NB-
The most logical explanation for the difference between 01) and 2.06 mg/dL (35.2 ␮mol/L) (NB-11), respectively,
NB-01 and NB-11 would be a change in calibrators or re- but only 0.36 mg/dL (6.1 ␮mol/L) by the TBIL diazo
assignment of bilirubin values in the calibrators used by method; the reference method and the mean TBIL Vitros
these instruments. Consistent with this explanation, Dade values for C-97 were 21.04 mg/dL (360 ␮mol/L) and 21.40
Behring and Roche Diagnostics were among the diagnos- mg/dL (366 ␮mol/L), respectively. (2) Synchron values
tic companies that lowered the assigned values to their obtained by the diazo method differ by less than 1.0 mg/
bilirubin calibrators (Diane Stille, Dade Behring Inc, New- dL from those of the reference method while those of the
ark, Del, written communication, May 7, 2006, and Mela- bilirubin oxidase method differ (except for 2003) from 1.97
nie Swartzentuber, Roche Diagnostics Corporation, Indi- mg/dL (33.7 ␮mol/L) to 2.96 mg/dL (50.6 ␮mol/L) (Fig-
anapolis, Ind, written communication, August 18, 2006). ure 3). (3) According to Abbott package inserts, 2 methods
In addition, the Roche line of instruments revealed dra- are available for TBIL for both the Aeroset and Architect
matic improvement from 2005 to 2006 (Figure 3) and is analyzers. One is a diazo method (diazotized 2,4-dichlo-
consistent with an earlier change to lower assigned bili- roaniline) and the other a direct spectrophotometry. How-
rubin calibrator values (Melanie Swartzentuber, Roche Di- ever, because these are ‘‘open platform’’ instruments, the
agnostics Corporation, written communication, October 5, users may have chosen a variety of other methods (listed
2005). The large difference between C-97 and NB-01 mean in the participant summary report as ‘‘JG w/blank,’’
bilirubin values for the Vitros indicates a lapse in the cal- ‘‘Spectrophot w/o blank’’ or ‘‘w/blank,’’ ‘‘DZ Salt,’’ ‘‘ox-
ibration of the TBIL and neonatal bilirubin (Bu⫹Bc) meth- idation,’’ ‘‘vanadate oxidation’’), which yield very variable
ods; because NB-01 contained only unconjugated biliru- results. It is important to emphasize that calibrators con-
bin, one would expect the TBIL and neonatal bilirubin val- taining ditaurobilirubin or made in a protein base other
ues to be very close. than human serum may not be suitable for calibrating all
bilirubin methods,1 that is, assigned values to calibrators
Precision may have to be different for methods based on different
Coefficients of variation for CAP All Data increased in principles. A calibrator suitable for all methods is one
2004 and 2005, but in 2006 returned to the level of 2003 made with human serum enriched with unconjugated bil-
(Table 1). Lack of improvement is attributed to the large irubin. This is because it has the same composition as
differences in the mean bilirubin values reported by the specimens obtained from healthy neonates.
various instrument groups (Figure 3). It should be noted
that the CVs for the major instrument groups (Dimension, COMMENT
Olympus, Synchron, and Vitros), which report about 65% Because of the importance of accuracy in the measure-
of all results, vary from 2% to 3%. ment of bilirubin in neonates, we proposed setting a goal
1784 Arch Pathol Lab Med—Vol 132, November 2008 Laboratory Performance in Neonatal Bilirubin Testing—Lo et al
for limiting the total error (inaccuracy plus imprecision) prove the quality of bilirubin calibrators. A bilirubin ref-
to ⫾10% of the reference method value at bilirubin con- erence method and reference material, as well as an es-
centrations greater than 10 mg/dL (171 ␮mol/L).3 tablished molar absorptivity for the alkaline azopigment,
In 2003, the first year of the introduction of the com- can be used to assess the accuracy of calibrators made
mutable survey specimen consisting of human serum en- with unconjugated bilirubin in human serum. These cal-
riched with unconjugated bilirubin, that goal appeared to ibrators may then be used to assign values to secondary
be within reach for at least 80% of the reported results. calibrators for field methods.
That was presumably due to better bilirubin calibrators References
used in that year as judged by the small difference be- 1. Lo SF, Doumas BT, Ashwood ER. Performance of bilirubin determinations
in US laboratories—revisited. Clin Chem. 2004;50:190–194.
tween the CAP All Data and reference method mean val- 2. American Academy of Pediatrics, Provisional Committee for Quality Im-
ue. In 2006, for specimen NB-01, only the Advanced BR2 provement and Subcommittee on Hyperbilirubinemia. Practice parameter: man-
agement of hyperbilirubinemia in the healthy term newborn. Pediatrics. 1994;94:
and the Synchron CX4/5CE met the upper limit of the 558–565.
⫾10% goal; all instruments met the lower limit except the 3. Lo SF, Doumas BT, Ashwood ER. Bilirubin proficiency testing using speci-
Architect. The ⫾20% error allowed by the Clinical Labo- mens containing unconjugated bilirubin and human serum: results of a College
of American Pathologists study. Arch Pathol Lab Med. 2004;128:1219–1223.
ratory Improvement Amendments of 1988 implies that ac- 4. Doumas BT, Poon PKC, Perry BW, et al. Candidate reference method for
ceptable values for a specimen having bilirubin concentra- determination of bilirubin in serum: development and validation. Clin Chem.
tion of 25 mg/dL (428 ␮mol/L) would be from 20 mg/ 1985;31:1779–1789.
5. Doumas BT, Perry BW, McComb RB, et al. Molar absorptivities of bilirubin
dL (342 ␮mol/L) to 30 mg/dL (512 ␮mol/L); this very (NIST SRM 916a) and its neutral and alkaline azopigments. Clin Chem. 1990;36:
large window should be unacceptable to pediatricians and 1698–1701.
6. National Institute of Standards and Technology. Certificate of Analysis. Stan-
neonatologists. To improve bilirubin assays, manufactur- dard Reference Material 916a, Bilirubin. Gaithersburg, Md: National Institute of
ers of clinical analyzers and reagent kits will need to im- Standards and Technology; 2001.

Arch Pathol Lab Med—Vol 132, November 2008 Laboratory Performance in Neonatal Bilirubin Testing—Lo et al 1785

You might also like