You are on page 1of 2
DRS ENGINEERING 8 CALIBRATION SERVICES AN ISO/IEC-17025:2017/ ACCREDITED BY NABL_ cous oe, CONTACT US: (Gj 0120-4236630 © 8822229228 InfoGarstab.in Ss RVICES & A 82 SECTOR 65 NoIDA 201 301 €o3193 lIssued For MIS UNIQUE CALIBRATION JULR No, (6C319321000000597F SOLUTIONS LLP \cai - | 611-P, SECTOR-38,GURGAON, joa cenmenteNe. _Florsovavssy _J HARYANA-122001 RENEE De. prsisrri21266 | Date of Calibration 11.09.2021 [bisiptine : Mechanical Force Proving Ring [Calibration Duo on 0.142028 Details of Unit Under Calibration Issue Date of Certificate 13.09.2021 Type Load Cet with incicator Make MAKIN Date of Rec 70092024 Model 83 [Customer Ref. No. & Date: | BY HAND Serial Numbor wwrz0151 [Conaition on receipt OK Capacity 204N Cable Length S5METER Environmental Condition Details of Indicator JRoom Temperature mire Indicator Type Dita Huriaty 400% RH ee ma Mode of application of Force: Serial Number 1226 Equipment kt ucsu.cios Compression Resolution COUNTS Accossories Set atgning Compression Pad es, T0XN DEAD Wight Fore Calraion Nace (ORS/OWFON-GD) 200kN Fyeraue Caltaon Machine (ORSIMFCMO1) Associated Uncertainty 0.044% at 22) 8 0.040% a 422) trace ity FRVOSI20}0WENV021 Dated [3708/2020 Vai upto 030972022 FRUDS2O;DWCWVO22A Dated 0310812020 Vali upto 03 |Principle/Methodology of Calibration and Calibration Procedure: 1. Work Instruction Number: DRS/SOPIOIF Based on IS 4169-2014 & ISO 375: 2011 2022 2. No Load Output: The digital Indicator was switched on for 30 minutes to warm up and stabilized for no lead output before the start of [caltraton The no oad output was noted (before tarng) andthe calbration signal was noted. 3. Protoading : Before the aplication ofthe calibration forces, the instrument was preloaded tvice tos maximum capacity and Kept at fl jad for about 90 seconds, 4. Calibration The sequence of applied calibration force in Compression ‘given below: at 0° Two series of calitration forces in increasing values. At 120° and 240° postions: One series of calibration forces each increasing values. Creep testis performed by calculating the ference in output 190 obtained at 308 and i300 obtained at 300s afer the removal ofthe maximum calibration force and express this difference as percentage ofthe maximum defection. The calibration was made by using self laigning Compression Pad provided along wih he instrument to ensure axial application ofthe force. Between each senes, the instrument was| rated along its axis so as to occupy during calibration twee positions (0° 120",240") and the instrument was subjected to the fll load once fo Jabout 90 seconds each tie before starting in a new pesiton. Between the loadings, readings corresponding to no load after waiting at least 20 lseconds forthe return to zero were noted Se AS ys > Qs Gurdeep Singh KaUshal Kishore 3 (Technical Manager) [FORWAT NO_-DRSICF/18.01 Page 1 of 2 ALIBRATION *TE DRS ENGINEERING & CALIBRATION SERVICES <~ AN ISO/IEC-17025:2017/ ACCREDITED BY NABL CONTACT US: @ 0120-4236630 ( 8822229228 & Into@drsiabin SERVICE 2 A82 sector 65 NoIDa 201 301 CALIBRATION CERTIFICATE 03193 [Catibration Method : This Insrunent is Calbraied n Compression Modo as per] ULR No (60319521000000646F ORS Engineering & Calibration Services Procedure No. DRSISOPIONF Based | orccesreae | on 1s 4169-2014nS0 376-2011 cai comiteate no. | DRSIOE2H640 Result : Compression Mode No Load Output 3 Counts [Galbration Signal Na : sun ond rrcen] = Dato Coun) iH ‘sition 0° | Position” | Position 120° | Postion 240° Average series 4 series? | series 3 series 4 (sories 4.34) 7 3 ° o ° ° = 2 70 303 303 302 30 502 3 200 "ood 7005 7004 1003 7004 @ 500 2508 2506; 2505 2504 2506; 5 1000 ‘5005, ‘5000 ‘5003 ‘5000 7 5003. 5 7000 0024 0023 10020) 0005 70025 7 [4000 200s8 | 20028 2000 70022 20072 a 000 30025 30018 29998 30007 3000 3 0000) 50008 49968 49960 #0906) 43988 | 10 | 15000 74895) 74842 74832 “74860 74882 1 20000 ‘99727 99738 ‘99723 ‘99742 (99731 [12 0 0 co 1 0 = [Greop ato Load Reading After 20 Sec Reading After 300 Sec 1 Counts ° Counts Interpolation Equation: = 0.000001 16482 + §.0091735719% Viherox= Foc in Ys Average Reading in Couns cinssteaion: The Fore proving inant a fon a cmp wih he eure of 1:16.20 ssid a lows ees Moae From To |___ Unser of ensure Class 1 ‘Compression 20000 N 200N | 20.120% ciass2 camresion | 200000 | 100N 20% int is at coverage factor a2 which comesponde probably of approximately 85% for @ normal dtrbuton. Conssering the relative deviation ofdference components such as creep, zero, rsliion, repeat rereducbily & combining wih the uncertainty of applied tree 2 Caloraton nas boon permed using set-ingng Compression Pad proved bythe customer 3 Instruments has been calbatod as per case C cra, “4 Rosult ite only the tom cairates 5 Creep measure at zo fad 16 This conteate shal not be reproduced expect in fi Calbrated oy 1% Gurceep nah SCA oe ges (Techical Manager & i Page Fo FORMAT NO. DRSCHTEOT

You might also like