You are on page 1of 6

LABORATORY INVESTIGATION

Assessment of Needle Guidance Devices for Their


Potential to Reduce Fluoroscopy Time and Operator
Hand Dose during C-Arm Cone-Beam Computed
Tomography–guided Needle Interventions
Maarten W. Kroes, MSc, Wendy M.H. Busser, MSc, Jurgen J. Fütterer, MD, PhD,
Mark J. Arntz, MD, Caroline M.M. Janssen, MD, Yvonne L. Hoogeveen, PhD,
Frank de Lange, PhD, and Leo J. Schultze Kool, MD, PhD

ABSTRACT

Purpose: To assess whether the use of needle guidance devices can reduce fluoroscopy time and operator hand dose during cone-
beam computed tomography–guided needle interventions.
Materials and Methods: The freehand technique was compared with techniques employing two distinct needle holders and a
ceiling-mounted laser guidance technique. Laser guidance was used either alone or in combination with needle holders. Four
interventional radiologists were instructed to reach predetermined targets in an abdominal phantom using these techniques. Each
operator used all six techniques three times. Fluoroscopy time, procedure time, operator hand dose, and needle tip deviation were
obtained for all simulated needle interventions. All data are presented as median (ranges).
Results: All procedures were successfully completed within 2–4 minutes, resulting in a deviation from target of 0.8 mm (0–4.7). In
freehand procedures, the fluoroscopy time to reach the target was 50 seconds (31–98 s). Laser guidance, used alone or in combination
with needle holders, reduced fluoroscopy time to 31 seconds (14–68 s) (P o .02). The operator hand dose in freehand procedures was
275 mSv (20–603 mSv). Laser guidance alone or in combination with needle holders resulted in a reduction of the hand dose to o 36
mSv (5–82 mSv) per procedure (P o .001). There were no statistically significant effects on hand dose levels or fluoroscopy time
when the needle holders were employed alone.
Conclusions: Compared with the freehand technique, all three tested needle guidance devices performed with equivalent efficiency
in terms of accuracy and procedure time. Only the addition of laser guidance was found to reduce both fluoroscopy time and operator
hand dose.

The development of cone-beam computed tomography procedures involving targeted percutaneous needle inter-
(CT) within a C-arm angiography system has led to the ventions (3–11). During these procedures, a cone-beam CT
development of an image guidance technique for needle volume is used for needle path planning, while fluoroscopy
interventions in the angiography suite (1,2). Several and the rotational capabilities of the C-arm provide real-
authors have described the use of cone-beam CT in time feedback during needle placement and advancement
(2,12). Compared with conventional CT-guided needle
interventions, the use of cone-beam CT guidance has been
From the Department of Radiology, Radboud University Nijmegen Medical
Centre, Geert Grooteplein zuid 10, Internal postal code 766, Nijmegen, 6525 reported to improve patient access (12) and to reduce
GA The Netherlands. Received October 23, 2012; final revision received radiation exposure to the patient (4).
February 16, 2013; accepted February 24, 2013. Address correspondence One important limitation of cone-beam CT image
to M.W.K.; E-mail: m.kroes@rad.umcn.nl
guidance is that using two-dimensional fluoroscopy feed-
From the SIR 2011 Annual Meeting. back to monitor needle advancement, the operator lacks
M.W.K. is a PhD student working on an industrial PhD grant received from either depth or angular information of the actual needle
the Norwegian Research Council by the Department of Radiology of the path, depending on the viewing angle. To monitor both
RUNMC and NeoRad AS. None of the other authors have identified a conflict
of interest. depth and angular information, the operator needs to switch
between entry point view—an overlay of entry and target
& SIR, 2013
point in a bull’s-eye fashion—and progress view—
J Vasc Interv Radiol 2013; 24:901–906 perpendicular to the entry point view. If necessary, correc-
http://dx.doi.org/10.1016/j.jvir.2013.02.037 tions to the actual needle path angle can be made in the

Downloaded for Anonymous User (n/a) at Liberty University from ClinicalKey.com by Elsevier on March 07, 2019.
For personal use only. No other uses without permission. Copyright ©2019. Elsevier Inc. All rights reserved.
902 ’ Needle Guidance Devices during Cone-Beam CT Kroes et al ’ JVIR

Phantom; CIRS, Norfolk, Virginia) using an 18-gauge


needle. Four radiopaque 2.3-mm diameter spheres (CT-
SPOTS; Beekley, Bristol, Connecticut) placed inside the
phantom were used as targets (Fig 1). These small high-
density targets were chosen to determine accurately the
deviation from needle tip to target.
Cone-beam CT–guided needle interventions were per-
formed using a C-arm angiography system (Allura Xper
FD20; Philips Healthcare, Best, The Netherlands). Cone-
beam CT volumes were acquired and used to plan a needle
path between a randomly selected skin entry point and one
of the four targets using XperGuide ablation software
(Philips Healthcare) (2). Axial needle paths with max-
imum angulations of 45 degrees were used because this is
currently the maximum angle for the laser guidance device.
Figure 1. Cone-beam CT slice of the used phantom. The four Care was taken not to plan paths along previously followed
radiopaque targets were positioned at a depth of 9–13 cm inside
the phantom.
trajectories to prevent bias caused by previously deformed
phantom material. Cone-beam CT–guided needle interven-
entry point view (2,3,13). In our experience, these corrective tions involved locating and marking the skin entry point on
needle manipulations frequently result in placement of the the phantom and placing the needle in the planned angle
operator’s hand inside the primary radiation beam. under fluoroscopy guidance with the C-arm in entry point
For CT-guided needle interventions, the use of needle view. In the entry point view, the overlay of entry and
holders has been suggested to increase accuracy and aid in target point and the cone-beam CT volume are super-
decreasing the number of needle manipulations and control imposed on real-time fluoroscopy. The operator visualized
scans, saving time and limiting radiation dose to the patient the advancing needle in the progress view, perpendicular to
(14). Although cone-beam CT–guided needle placements the entry point view, along the needle shaft and with the
using needle holders have been described (11,13), the effect planned path as well as cone-beam CT superimposed on
of needle holders on these parameters in cone-beam CT– real-time fluoroscopy (2,12). The needle intervention was
guided procedures have not been reported. The purpose of completed when the needle was placed onto the target
this study was to assess whether the use of needle guidance according to the operator. A second cone-beam CT scan
devices could also aid in reducing the number of manipu- was obtained for accuracy measurements.
lations in cone-beam CT–guided needle interventions, reduc-
ing fluoroscopy time and operator hand dose. We further
assessed whether these devices could aid in optimizing these
Needle Guidance Devices
In this study, cone-beam CT–guided freehand needle
procedures in terms of procedure time and accuracy.
placements were compared with interventions employing
commercially available needle guidance devices. A ceiling-
MATERIALS AND METHODS mounted laser guidance technique and two distinct needle
holders were used.
Procedure The needle holders used were the SeeStar (AprioMed,
Needle interventions were performed on an abdominal Uppsala, Sweden) and Simplify (NeoRad AS, Oslo, Nor-
phantom (Model 057 Triple Modality 3D Abdominal way) (Fig 2). Both needle holders allowed rotation and

Figure 2. The two different types of needle holders used for assisting the operator in needle placement. (a) SeeStar. (b) Simplify.
(Available in color online at www.jvir.org.)

Downloaded for Anonymous User (n/a) at Liberty University from ClinicalKey.com by Elsevier on March 07, 2019.
For personal use only. No other uses without permission. Copyright ©2019. Elsevier Inc. All rights reserved.
Volume 24 ’ Number 6 ’ June ’ 2013 903

laser aimed at the previously marked skin entry point and


the plane laser beam was in alignment with the operating
table (Fig 3). The freehand technique was compared with
procedures performed with the aid of SeeStar, Simplify,
and SimpliCT alone and with procedures employing
SimpliCT-SeeStar and SimpliCT-Simplify combinations.
The simulated needle interventions were performed by
four interventional radiologists. Three interventional radi-
ologists had 3–5 years of experience in cone-beam CT–
guided needle interventions, and one (a fellow) had 1 year
of experience. Because the operators had no experience in
using the needle guidance devices, all operators performed
one needle intervention with each guidance device as
training. For this phantom study, all four operators per-
formed three needle placements per technique to verify
reproducibility and facilitate interoperator analysis.

Outcome Measures
Fluoroscopy time, procedure time, radiation exposure to
the hand of the operator, and accuracy were recorded.
Fluoroscopy time was obtained from the angiography
system software and was defined as the time in seconds
of real-time fluoroscopy necessary for locating the skin
entry point, placing and aligning of the needle, and
advancing the needle onto the target.
Procedure time of the simulated intervention was defined
as the time necessary to place the needle onto the target,
including the setup time for the needle guidance devices
but excluding the time required for placing the phantom on
Figure 3. Schematic presentation of the laser guidance setup.
The guiding laser of SimpliCT is aimed along a planned needle the table, acquiring both cone-beam CT scans and needle
path of 30 degrees in the axial direction (straight line), while the path planning. Procedure time was measured by one author
plane laser is aligned to the operating table (dashed line). The (M.W.K.) using a stopwatch.
C-arm is positioned in progress view. (Available in color online Radiation exposure to the hand manipulating the needle
at www.jvir.org.)
was measured in microsieverts (mSv) using a dosimeter
with a small-tip sensor (EDD-30; Unfors Instruments,
angulation of the needle, while maintaining it in a fixed
Billdal, Sweden). The sensor of the dosimeter was placed
angle during advancement. The metallic needle guide of
on the back of the hand manipulating the needle between
the SeeStar could be used to position the needle holder
the thumb and index finger.
because it was visible on fluoroscopy (14,15). A practical
Accuracy, defined as the distance between target edge
advantage of the Simplify was that it could be detached
and needle tip in the control cone-beam CT scan, was
without removing the needle.
measured on the angiography system’s dedicated three-
The laser guidance unit (SimpliCT; NeoRad AS) acted
dimensional workstation (XtraVision; Philips Healthcare)
as a laser pointing device with the possibility to visualize
and assessed to monitor the performance of the operators.
the planned needle path for the operator (16–18). The
ceiling-mounted version of the device used in this study
could be used to maximum angulations of 45 degrees in the
axial and craniocaudal direction. The guiding laser sus- Statistical Analysis
pension inside the unit was self-leveling, and the unit All statistical analyses were conducted in SPSS (version
contained an additional plane laser to align it with the 16.0.01; SPSS Inc., Chicago, Illinois). All results are
operating table. Using information obtained from the represented as medians with corresponding ranges. Differ-
vendor of the angiography system, the angles of the C- ences between the freehand technique and techniques using
arm position in the entry point view were recalculated to needle guidance devices were analyzed using the Mann-
yield the planned needle path angles in three-dimensional Whitney U test. Differences in performances between
views. These angles were subsequently fed into the laser operators were analyzed for all techniques using the
guidance unit. With the C-arm positioned in the progress Kruskal-Wallis test. Differences were considered statisti-
view, the laser unit was positioned such that the guiding cally significant for P o .05.

Downloaded for Anonymous User (n/a) at Liberty University from ClinicalKey.com by Elsevier on March 07, 2019.
For personal use only. No other uses without permission. Copyright ©2019. Elsevier Inc. All rights reserved.
904 ’ Needle Guidance Devices during Cone-Beam CT Kroes et al ’ JVIR

Figure 4. Box plot representing the fluoroscopy times in Figure 5. Box plot showing the measured radiation hand dose
seconds required to guide the needle onto the target per needle per needle guidance technique. The asterisk represents statis-
guidance device. The asterisk represents a statistically signifi- tically significant lower (P ¼ .001) hand dose for laser-guided
cant difference between the freehand technique and laser- procedures compared with the freehand technique.
guided procedures.

RESULTS Both needle holders alone showed no statistically significant


difference compared with the freehand procedure: SeeStar,
In needle interventions using the freehand technique,
298 mSv (80–876 mSv; P ¼ .478); Simplify, 167 mSv
median fluoroscopy time needed to reach the target was
(42–464 mSv; P ¼ .699).
50 seconds (31–98 s). The median fluoroscopy time for
The measured deviation from the target for the freehand
laser-guided needle interventions was 31 seconds (14–55 s)
technique was 1.2 mm (0.1–3.5 mm). There were no
for SimpliCT, 30 seconds (15–55 s) for the SimpliCT-
significant differences in accuracy between procedures
SeeStar combination, and 29 seconds (16–68 s) for the
using any of the needle guidance devices compared with
SimpliCT-Simplify combination. Compared with the free-
the freehand technique. Seestar-guided needle interventions
hand technique, significantly shorter fluoroscopy times
resulted in a deviation of 0.4 mm (0–1.5 mm; P ¼ .090);
were achieved using SimpliCT (P ¼ .004) and the com-
Simplify, 0.9 mm (0.1–4.7 mm; P ¼ .961); SimpliCT, 0.2
binations SimpliCT-SeeStar (P ¼ .001) and SimpliCT-
mm (0–1.8 mm; P ¼ .075); SimpliCT-SeeStar, 0.2 mm
Simplify (P ¼ .018) (Fig 4). Both needle holders alone did
(0–4.1 mm; P ¼ .091); and SimpliCT-Simplify, 0.3 mm
not result in a statistically significant difference in out-
(0–1.7 mm; P ¼ .106). Finally, the data did not show
comes compared with the freehand technique: SeeStar,
any statistically significant difference in interoperator
67 seconds (40–95 s; P ¼ .193); Simplify, 59 seconds
performance.
(30–138 s; P ¼ .898).
Except for the procedures using the SeeStar (2.8 min
[1.6–5.1 min]; P ¼ .098), the procedure time for the
freehand technique (1.9 min [0.8–3.6 min]) was statisti-
DISCUSSION
cally significantly shorter compared with all other guidance The most essential findings of this phantom study are that
devices. Procedure times were 3.1 minutes (1.7–5.8 min; adding laser guidance to cone-beam CT–guided needle
P ¼ .028) using Simplify and 2.9 minutes (2.2–6.1; P ¼ interventions significantly reduces fluoroscopy time and
.007) SimpliCT; for the combinations SimpliCT-SeeStar operator hand dose. Compared with freehand techniques,
and SimpliCT-Simplify, procedure times were 3.6 minutes laser guidance prolongs procedure times by only a few
(1.3–5.5 min; P ¼ .005) and 3.5 minutes (2.0–5.5 min; P ¼ minutes, a difference that likely would go unnoticed in
.001), respectively. clinical practice.
The measured operator hand dose for the freehand Hand doses during interventional radiology procedures
technique was 275 mSv (20–603 mSv). SimpliCT-guided can be high; dose levels up to several millisieverts per
needle interventions resulted in a significantly reduced procedure, consistent with exposure of the hand to the
operator hand dose of 33 mSv (8–82 mSv; P ¼ .001). Similar primary x-ray beam, have been reported (19,20). Sterile
results were obtained for SimpliCT in combination with the radiation-attenuating surgical gloves are commercially
needle holders (SimpliCT-SeeStar, 36 mSv [8–82 mSv]; available. These gloves can provide significant protection
SimpliCT-Simplify, 32 mSv [5–79 mSv]; both combinations from scatter radiation but provide only minimal protection
P ¼ .001) (Fig 5). All procedures with SimpliCT showed when hands are placed in the primary x-ray beam (21–23).
only a very small variation in hand dose measurements. The reduction in operator hand dose by employing laser

Downloaded for Anonymous User (n/a) at Liberty University from ClinicalKey.com by Elsevier on March 07, 2019.
For personal use only. No other uses without permission. Copyright ©2019. Elsevier Inc. All rights reserved.
Volume 24 ’ Number 6 ’ June ’ 2013 905

guidance is attributable to the more efficient placement of interoperator performance, the dataset may be too small to
the needle and the reduced number of corrective needle address questions relating to operator experience or learn-
manipulations in the entry point view, effectively prevent- ing curves. Second, at the present time, cone-beam CT
ing direct exposure of the hands to the primary beam guidance cannot compensate for target motion after the
leaving scatter radiation as the predominant contribution to needle path planning is done. In case the patient moves
the operator’s hand dose. with the needle already inserted, laser guidance could help
For CT-guided needle interventions, the use of the in getting the physician back on track, but more often
SeeStar needle holder has been suggested to increase patient movement would require a new cone-beam CT scan
accuracy and aid in decreasing the number of needle and planning procedure to be performed (11,12). When
manipulations and control scans, saving time as well as frequent motion compensations are required in clinical
radiation exposure to the patient (14). The authors practice, the reduction of fluoroscopy time by using laser
attributed these effects mainly to the metallic needle guidance would not reach the same level as presented in
guide creating a streak artifact pointing at the target in this study. However, even in these cases, laser guidance
the CT images, indicating the needle path. In our simulated would prevent exposure of the operator’s hands to the
C-arm cone-beam CT–guided interventions, the needle primary x-ray beam.
holders were employed only after the cone-beam CT scan The freedom of movement of the current version of the
was made and the streak artifact did not appear in laser guidance setup limits its applicability to guiding
fluoroscopy, so the effect on accuracy was not apparent. needle paths within the range of 45 degrees, in all
In addition, the design of the study was such that all directions from the vertical axis. The design of the current
techniques, including freehand, involved simultaneous study further involved simulated, in-plane trajectories only.
visualization of the advancing needle, the planned needle However, the combination of laser guidance and cone-
path, and the cone-beam CT volume superimposed on real- beam CT–guided needle interventions can be applied just
time fluoroscopy—essentially rendering all techniques as easily to double oblique trajectories.
equally accurate. In our simulated cone-beam CT–guided In conclusion, adding laser guidance to cone-beam CT–
interventions, the use of needle holders did not show an guided needle interventions provides visual feedback that
effect on the number of corrective manipulations compared effectively prevents exposure of the hands of the operator
with freehand, as evidenced by similar fluoroscopy times to the primary radiation beam. The use of laser guidance
and operator hand dose levels. Similar to the freehand maintains tactile feedback and facilitates accurate needle
technique, when employing the needle holders, the place- placements within essentially the same time frame as
ment of the needle still required frequent adjustments of freehand procedures. The extent to which laser guidance
the needle angle in the entry point view. Based on in cone-beam CT–guided needle interventions can aid in
our experience, the added value of the needle holders reducing fluoroscopy time and patient exposure in clinical
in cone-beam CT–guided percutaneous procedures is practice is currently under investigation.
in the physical support they provide while advancing the
needle.
Schultz et al (24) recently described a phantom study
in which the use of a table-mounted, remote-controlled, REFERENCES
small robotic device capable of holding and guiding 1. Siewerdsen JH, Moseley DJ, Burch S, et al. Volume CT with a
needles in aiding needle interventions was investigated. flat-panel detector on a mobile, isocentric C-arm: pre-clinical investigation
These authors indicated that accuracies o 5 mm were in guidance of minimally invasive surgery. Med Phys 2005; 32:
241–254.
feasible when using this device and that no direct 2. Racadio JM, Babic D, Homan R, et al. Live 3D guidance in the
exposure of the physician’s hands to the primary interventional radiology suite. AJR Am J Roentgenol 2007; 189:
radiation beam occurred owing to remote control. W357–W364.
3. Tam A, Mohamed A, Pfister M, Rohm E, Wallace MJ. C-arm cone
Compared with laser guidance as described in this study, beam computed tomographic needle path overlay for fluoroscopic-guided
one important drawback of using such automated guidance placement of translumbar central venous catheters. Cardiovasc Intervent
devices is that the needle holder is positioned in the Radiol 2009; 32:820–824.
4. Braak SJ, Van Strijen MJL, Van Es HW, Nievelstein RAJ, Van Heesewijk
primary beam during initial alignment. This positioning JPM. Effective dose during needle interventions: cone-beam CT guid-
may increase radiation exposure to the patient. In addition, ance compared with conventional CT guidance. J Vasc Intervent Radiol
because no table mounting is involved, setting up laser 2011; 22:455–461.
5. Morimoto M, Numata K, Kondo M, et al. C-arm cone beam CT for
guidance in a sterile environment in our daily clinical hepatic tumor ablation under real-time 3D imaging. AJR Am J Roent-
routine is completed within minutes. This is significantly genol 2010; 194:W452–W454.
faster than the reported Z 10 minutes of setup time 6. Hwang HS, Chung MJ, Lee JW, Shin SW, Lee KS. C-arm cone-beam
CT-guided percutaneous transthoracic lung biopsy: usefulness in evalua-
required for the robot (24). tion of small pulmonary nodules. AJR Am J Roentgenol 2010; 195:
There are several limitations to this study. For practi- W400–W407.
cal reasons, the operators were asked to simulate three 7. Spelle L, Ruijters D, Babic D, et al. First clinical experience in applying
XperGuide in embolization of jugular paragangliomas by direct
needle interventions for all six techniques. Although the re- intratumoral puncture. Int J Comput Assisted Radiol Surg 2009; 4:
sulting data did not provide evidence for differences in 527–533.

Downloaded for Anonymous User (n/a) at Liberty University from ClinicalKey.com by Elsevier on March 07, 2019.
For personal use only. No other uses without permission. Copyright ©2019. Elsevier Inc. All rights reserved.
906 ’ Needle Guidance Devices during Cone-Beam CT Kroes et al ’ JVIR

8. Nesbit GM, Nesbit EG, Hamilton BE. Integrated cone-beam CT and 15. Roberts CC, Morrison WB, Deely DM, Zoga AC, Koulouris G, Winalski
fluoroscopic navigation in treatment of head and neck vascular malfor- CS. Use of a novel percutaneous biopsy localization device: initial
mations and tumors. J Neurointervent Surg 2011; 3:186–190. musculoskeletal experience. Skeletal Radiol 2007; 36:53–57.
9. Busser WMH, Hoogeveen YL, Veth RPH, et al. Percutaneous radio- 16. Varro Z, Locklin JK, Wood BJ. Laser navigation for radiofrequency
frequency ablation of osteoid osteomas with use of real-time needle ablation. Cardiovasc Intervent Radiol 2004; 27:512–515.
guidance for accurate needle placement: a pilot study. Cardiovasc 17. Brabrand K, Aaløkken TM, Krombach GA, et al. Multicenter evaluation
Intervent Radiol 2011; 34:180–183. of a new laser guidance system for computed tomography intervention.
10. Braak SJ, Herder GJM, Van Heesewijk JPM, Van Strijen MJL. Pulmo- Acta Radiol 2004; 45:308–312.
nary masses: initial results of cone-beam CT guidance with needle 18. Krombach GA, Schmitz-Rode T, Wein BB, et al. Potential of a new laser
planning software for percutaneous lung biopsy. Cardiovasc Intervent target system for percutaneous CT-guided nerve blocks: technical note.
Radiol 2012; 35:1414–1421. Neuroradiology 2000; 42:838–841.
11. Kroeze SGC, Huisman M, Verkooijen HM, van Diest PJ, Bosch JLHR, 19. Vano E, Gonzalez L, Guibelalde E, Fernandez JM, Ten JI. Radiation
van den Bosch MAAJ. Real-time 3D fluoroscopy-guided large core exposure to medical staff in interventional and cardiac radiology. Br J
needle biopsy of renal masses: a critical early evaluation according to Radiol 1998; 71:954–960.
the IDEAL recommendations. Cardiovasc Intervent Radiol 2012; 35: 20. Sanchez RM, Vano E, Fernandez JM, et al. Staff doses in interventional
680–685. radiology: a national survey. J Vasc Intervent Radiol 2012; 23:1496–1501.
12. Braak SJ, Van Strijen MJL, Van Leersum M, Van Es HW, Van Heesewijk 21. Schueler BA. Operator shielding: how and why. Tech Vasc Interv Radiol
JPM. Real-time 3D fluoroscopy guidance during needle interventions: 2010; 13:167–171.
technique, accuracy, and feasibility. AJR Am J Roentgenol 2010; 194: 22. Schueler BA, Baiter S, Miller DL. Radiation protection tools in interven-
W445–W451. tional radiology. J Am Coll Radiol 2012; 9:844–845.
13. Leschka SC, Babic D, El Shikh S, Wossmann C, Schumacher M, 23. Neeman Z, Dromi SA, Sarin S, Wood BJ. CT fluoroscopy shielding:
Taschner CA. C-arm cone beam computed tomography needle path decreases in scattered radiation for the patient and operator. J Vasc
overlay for image-guided procedures of the spine and pelvis. Neuro- Intervent Radiol 2006; 17:1999–2004.
radiology 2012; 54:215–223. 24. Schulz B, Eichler K, Siebenhandl P, et al. Accuracy and speed of robotic
14. Magnusson A, Radecka E, Lönnemark M, Raland H. Computed- assisted needle interventions using a modern cone beam computed
tomography-guided punctures using a new guidance device. Acta Radiol tomography intervention suite: a phantom study. Eur Radiol 2013; 23:
2005; 46:505–509. 198–204.

Downloaded for Anonymous User (n/a) at Liberty University from ClinicalKey.com by Elsevier on March 07, 2019.
For personal use only. No other uses without permission. Copyright ©2019. Elsevier Inc. All rights reserved.

You might also like