You are on page 1of 6

ORIGINAL ARTICLE

Laser Targeting With C-Arm Fluoroscopy: Effect on Image


Acquisition and Radiation Exposure
Franklin D. Shuler, MD, PhD,* Justin L. Daigre, MD,† Danh Pham, MSIV,‡
and Vincent L. Kish, ASEE†

INTRODUCTION
Objectives: Reducing unnecessary radiation exposure from med- In 2010, the US Food and Drug Administration issued
ical imaging is paramount. This study assessed whether a laser a white paper and initiative to reduce unnecessary radiation
aiming guide for C-arm fluoroscopy reduced the number of exposure from medical imaging.1 Of the imaging modalities
exposures needed to obtain an acceptable image, thereby reducing addressed in this initiative, only one is in direct control of the
total fluoroscopy time for hip, knee, and ankle fluoroscopy. orthopaedic surgeon—intraoperative fluoroscopy. Increased
Methods: An obese cadaver was placed supine on a radiolucent use of C-arm fluoroscopy has been realized because of its utility
surgical table. Images were obtained by licensed radiologic technol- in fracture reduction, facilitation of internal fixation, and in the
ogists using a calibrated OEC 9900 Elite C-arm with laser targeting development of minimally invasive surgical techniques.1–7
(LT) and without LT (NLT). Dosimeters were placed 1, 3, and 6 ft Radiation exposure from intraoperative fluoroscopy is
(30.5, 91.5, and 183 cm) away from the center of the C-arm at derived from either primary radiation exposure (direct beam
90-degree angles at 2 levels, simulating thyroid and gonadal contact) or from scatter or stray radiation. Direct beam contact
exposure. Posterior–anterior (PA) images of the bilateral lower using a regular C-arm can result in exposure rates ranging
extremities were obtained with each technician acquiring 24 centered from 1200 to 4000 millirem per minute depending on the
images (hip, knee, and ankle) using both LT and NLT C-arm body part imaged.8 Scatter radiation is produced when the
fluoroscopy. direct beam interacts with objects in its path (ie, operating
room table, instruments, and patient) (Fig. 1). As the volume
Results: Total fluoroscopy time was reduced by 19% when using of tissue increases during C-arm fluoroscopy, the scattered or
LT with a 39% reduction for the knee and a 29% reduction for the stray radiation increases and provides the main source of
ankle. The addition of LT improved the likelihood of obtaining radiation dose to personnel in the operating room. Increasing
a centered image for knees and ankles but not for hips. The gonadal the distance from the radiation source reduces radiation
dosimetry data were significantly higher than the thyroid dosimetry exposure via the inverse square rule: increasing the distance
badges at 1 ft. At the 3-ft zone, only trace amounts of radiation were from the radiation source decreases radiation exposure
detected; the 6-ft zone reported no radiation exposure in either group. proportional to 1 by distance.2 Decreases in radiation expo-
Conclusions: LT helped with imaging knees and ankles with sure to the patient, surgeon, and ancillary staff are desirable.
statistically significant reductions in fluoroscopy time and a statisti- A decrease in radiation exposure during intraoperative fluo-
cally significant improvement of image quality defined as obtaining roscopy can be realized by decreasing fluoroscopy time and
a centered PA image faster. The dosimetry badges detected minimal number of digital spot images acquired, increasing the dis-
exposure at 3 ft and no detectable exposure at 6 ft at both levels. tance from the radiation source, using collimation, avoiding
magnification, using a mini C-arm, increasing patient and
Key Words: fluoroscopy, C-arm, laser, radiation exposure, fluoros- surgeon shielding, and allowing surgeon control of the
copy time C-arm.9–11 An additional tool that could potentially decrease
(J Orthop Trauma 2013;27:e97–e102) exposure time with fluoroscopy is laser targeting (LT).12,13
LT used in conjunction with C-arm fluoroscopy could
prove a useful tool in complying with mandates to reduce
unnecessary radiation exposure. However, previous studies
have demonstrated equivocal findings. Conn and Hallett12
Accepted for publication June 22, 2012. showed that a simple laser guide used in hip-fracture fixation
From the *Department of Orthopaedic Surgery, Marshall University, reduced screening time up to 40% thus limiting exposure
Huntington, WV; †Department of Orthopaedics, West Virginia University, time. In a prospective, randomized, controlled trial, Harris
Morgantown, WV; and ‡West Virginia University School of Medicine,
West Virginia University, Morgantown, WV. et al13 showed that a laser aiming guide did not reduce the
The authors declare no external funding for this study. number of exposures or total exposure time in fluoroscopy of
F. D. Shuler is a consultant for Medtronic and is on their speaker bureau. The several body regions in a non–case-controlled study. Our
remaining authors have no conflicts of interest to declare. study was designed to assess whether a laser aiming guide
Reprints: Justin L. Daigre, MD, Department of Orthopaedics, West Virginia
University, PO Box 9196, Morgantown, WV 26506-9196 (e-mail:
for C-arm fluoroscopy would reduce the number of exposures
jdaigre@hsc.wvu.edu). needed to obtain an acceptable image, thereby reducing the
Copyright © 2013 by Lippincott Williams & Wilkins total exposure time for hip, knee, and ankle fluoroscopy.

J Orthop Trauma  Volume 27, Number 5, May 2013 www.jorthotrauma.com | e97


Shuler et al J Orthop Trauma  Volume 27, Number 5, May 2013

FIGURE 1. C-arm scatter radiation:


surgeon radiation exposure during
C-arm fluoroscopy comes from 2
sources: primary (direct beam expo-
sure) and scatter radiation. Surgeon
exposure to scatter radiation increa-
ses as patients’ BMI increases (com-
pare A to B). This project used an
obese cadaver (BMI 43.6) to increase
the production of scatter radiation.
Source: An Investigation of Operator
Exposure in Interventional Radiology
permission from RadioGraphics.14

MATERIALS AND METHODS were placed under the lateral aspect of the lower legs to
The study protocol was approved by our Institutional facilitate image acquisition at the ankles.
Biosafety Committee as a biosafety level BSL2. C-arm A calibrated OEC 9900 Elite C-arm (GE Healthcare,
technologists registered with the American Registry of Salt Lake City, UT) was used for image acquisition. A
Radiologic Technologists at various levels of experience removable OEC laser aimer for the 9900 Elite C-arm (GE
(n = 16 with C-arm experience ranging from 1 year 8 months Healthcare) was used as the laser aiming guide with sterile
to 28 years with average experience of 5 years 8 months) drapes not interfering with its intensity. The attachment site
were brought to the operating room for live C-arm fluoros- for the laser was on the image intensifier. Optically stimulated
copy image acquisition. A fresh obese cadaver (body mass luminescence (OSL) badge dosimeters (Landauer Inc,
index or BMI 43.6; 122.5 kg) obtained from our Human Gift Glenwood, IL) were used to collect ionizing radiation doses
Registry was used in all experiments. This project specifically from the fluoroscopy. The OSL dosimeters detect x-rays
used an obese cadaver to increase the production of scatter or and gamma rays from 1 mrem to 1000 rem). The deep dose
stray radiation with PA projections used with the image (Hp10) is +/2 15% at the 95% confidence interval for
intensifier above the cadaver. The cadaver was placed supine photons above 20 keV, and the shallow dose (Hp0.07) is
on a Jackson radiolucent operating room table (Mizuhosi, +/2 15% at the 95% confidence interval for photons above
Union City, CA) at a fixed table height of 36 inches 20 keV. An apparatus was designed to hold the dosimeters at
(91.5 cm) from the floor. The cadaver was covered in a cool- fixed distances from the center of the C-arm (Fig. 2). The
ing blanket from the head to several centimeters above the apparatus fit on the C-arm image intensifier [vertically fixed
umbilicus to protect cadaver identity with an impervious body at 56 inches (142.3 cm) from the floor] and projected 4 arms
bag used for biocontainment. The pelvis and lower extremi- at 90-degrees from each other (45, 135, 225, and 315
ties were exposed during the testing protocol. Towel bumps degrees). The arms were radiolucent and more than 6 ft in

FIGURE 2. Detection of scatter radi-


ation during fluoroscopy with and
without LT. A, Three-dimensional
radiation detection was performed
both above and below the radiolu-
cent operating room table. Radiation
detection above the table simulated
surgeon thyroid exposure and
detection below the table simulated
surgeon gonadal exposure. B, Dos-
imeters were placed above and
below the operating room table
on radiolucent guidelines rigidly
attached to the C-arm intensifier at 1,
3, and 6 ft (30.5, 91.5, and 183 cm)
at four 90-degrees intervals. Images
used with permission from Markus
Wagner, Peter L. Reichertz Institute
for Medical Informatics, Hanover
Medical School, Germany.

e98 | www.jorthotrauma.com Ó 2013 Lippincott Williams & Wilkins


J Orthop Trauma  Volume 27, Number 5, May 2013 Laser Targeting with C-Arm Fluoroscopy

length. The length of these arms prevented us from obtaining An image was considered centered and acceptable if the
lateral images in this protocol. By rigidly attaching guidelines joint was at the center of the monitor screen within a 4-cm
to the C-arm image intensifier, distance from the center of the diameter (Fig. 3). Two-dimensional (x and y axis) deviations
radiation beam was kept constant during the horizontal C-arm from the center in centimeters were recorded for all images
motion required to complete the testing protocol. There were using a template directly applied to the C-arm monitor.
2 planes to the apparatus, one at the level of the thyroid and Repeat image acquisition was required for noncentered
one at the level of the gonads. OSL dosimeters were placed at images and the number of attempts for appropriately centered
1, 3, and 6 ft (30.5, 91.5, and 183 cm) away from the center of image acquisition was recorded. C-arm technologists were
the C-arm on all 4 arms and on both planes totaling 24 dos- not able to see measurements after image acquisition to avoid
imeters on the apparatus at one time. The dosimeters were potential bias on centered images. For noncentered images,
staggered on the apparatus so as not to block radiation to the the technologist was asked to take another image centered
other badges. Separate sets of badges were used in the LT and at the joint of interest. Once an acceptable image was
non-LT groups. A total of 56 dosimeters were used during the obtained, the technologist moved to the next body region.
testing protocol: 24 for the LT group and 24 for the non-LT Each technologist ultimately acquired 24 centered images
group. Eight background radiation badges were kept outside both with and without LT. All data were collected in a spread-
of the operating room as controls. sheet for statistical analysis.
Before entering the operating room, each American There were 4 outcomes measures assessed, (1) Does LT
Registry of Radiologic Technologist–registered radiologic decrease C-arm fluoroscopy exposure time for acceptable
technologist was randomized to laser or non-laser use. The centered image acquisition for a PA image series (hip, knee,
technologist’s name and years of intraoperative C-arm expe- and ankle)? (2) Does LT result in a decrease in the number of
rience were recorded. The technologist was escorted into the attempts for acceptable image acquisition? (3) Does C-arm
operative theater with the C-arm positioned above the umbi- experience decrease radiation exposure and improve image
licus with fixed operating room table and C-arm height used positioning accuracy either with or without LT? (4) Is there a
throughout the image protocol. If randomized to laser use, the learning curve associated with this protocol that can
technologist would obtain 6 PA images of the lower extrem- produce bias?
ities using the laser attachment. The following 6 PA images All data collected and analyzed were expressed as
were obtained: (1) PA of the right hip centered at the femoral means with statistical analysis performed using a Student t
test. A P value of ,0.05 was considered significant.
head, (2) PA of the left knee with the knee in the center of the
film, (3) PA of the right ankle with the ankle centered, (4) PA
of the left hip centered at the femoral head, (5) PA of the right RESULTS
knee with the knee in the center of the film, and (6) PA of the
The number of image acquisition attempts, fluoroscopy
left ankle with the ankle centered. The technologist used fluo-
time, and average image distance from center for hip, knee, and
roscopy (short taps of continuous mode with last-image hold) ankle fluoroscopy with and without LT is shown (Tables 1–3).
to obtain the images with no manual techniques. The technol- Use of LT resulted in statistically significant decreases
ogist was required to move the C-arm to each location to in the number of image attempts to obtain centered images at
obtain centered images. After the sixth image was obtained, the knee and ankle but not at the hip. The average distance
the technologist would then repeat the 6 images in order again from center per attempt is shown (Table 3). Each image
with the use of the laser. After the acquisition of 12 centered acquired had the average distance (in centimeters) from center
images, the technologist would repeat acquisition of the 12 calculated using the Pythagorean theorem. Statistically signif-
centered PA images without use of the laser. The same icant decreases were noted for the knee (4.95 vs. 3.24 cm;
sequence was followed for those who started with non-laser NLT vs. LT) and ankle (5.75 vs. 3.57 cm; NLT vs. LT) but
use. The technologist would then leave the room and the laser not for the hip (6.56 vs. 6.03 cm; NLT vs. LT).
attachment would either be applied to the C-arm image With use of LT, statistically significant improvements
intensifier or removed depending on the technologist’s ran- were noted for obtaining centered images of the knee and
domization. The OSL badges were changed as well to reflect ankle but not of the hip (total reduction in distance from
the laser or non-laser group and the protocol repeated. Only center: hip, 8%; knee, 35%; and ankle, 38%). Although
PA images were obtained during this protocol because the randomization was used in trials with or without LT, we
size of the apparatus precluded the technologists from swing- assessed for a possible learning curve between acquisition of
ing the C-arm to obtain a lateral image. centered images 1–6 (trial 1) and images 7–12 (trial 2). No
Eight technologists started the protocol in the LT group statistically significant differences were noted between the
and 8 started the protocol in the NLT group. Because the first 6 and second 6 centered images obtained with and with-
protocol repeated image acquisition after completion of the out LT (trial 1 vs. trial 2: 423 vs. 399 images; P = 0.2651).
first set of 6 centered radiographs, there was the potential for LT reduced the total fluoroscopy time by 19% for all 3
bias with improved image acquisition for subsequent trials. body regions imaged with an 81-seconds decrease in total
Data collected during protocol initiation did not show in-beam exposure using LT during the protocol. With the use
a statistically significant difference in number of images of LT, statistically significant decreases in fluoroscopy time
obtained during repeat image acquisition with this assertion were noted for image acquisition at the knee (39% reduction)
confirmed following final data analysis (see discussion). and ankle (29% reduction) but not at the hip.

Ó 2013 Lippincott Williams & Wilkins www.jorthotrauma.com | e99


Shuler et al J Orthop Trauma  Volume 27, Number 5, May 2013

FIGURE 3. Determination of centered fluoroscopic images. Two-dimensional determination (x and y axis) of centered fluoro-
scopic images was performed for all images captured during the protocol. In (A) and (B), a PA radiograph of the femoral head was
obtained. In (A), the hip was considered off center (27 cm x axis, +8 cm y axis). In (B), the hip was considered centered (+1 cm x
axis, 22 cm y axis). The scatter plot in (C) demonstrates that the addition of LT provided little benefit in obtaining centered hip
radiographs. In (D), the ankle was off center and in (E), the ankle was considered centered. The scatter plot in (F) demonstrates
that the addition of LT provided significant benefit in obtaining a centered ankle image.

A total of 56 OSL dosimeters were analyzed (24 from exposure for both laser and non-laser groups. There were no
LT, 24 from non-LT, and 8 controls). The average radiation statistically significant differences in radiation doses either with
dose in millirem is reported for simulated thyroid and gonadal or without LT.
exposure at the 1, 3, and 6 ft zones (Fig. 4). At 1 ft from center Technician experience (range of operating room C-arm
of the radiation beam, scatter radiation was 12-fold greater in experience from 1 year 8 months to 28 years) did not produce
the simulated gonadal dose (mean mrem: 66.9 NLT/68.3 LT) statistically significant differences in fluoroscopy time, num-
than the simulated thyroid dose (mean mrem: 5.4 NLT/5.6 LT). ber of image attempts, or average distance from the center.
At 3 ft, there was a significant reduction in radiation dose and
no radiation dose was detected at 6 ft. No difference in milli-
rem dosimetry data was noted between laser and non-LT DISCUSSION
groups. The gonadal level dosimetry badges at the 1-ft zone Because the US Food and Drug Administration issued
had the highest radiation exposure with a total of 273 mrem an initiative for reducing radiation exposure, many actions
with LT and 267 mrem without LT. The thyroid level dosim- have been taken to reduce this risk to patients, surgeons, and
etry badges at the 1-ft zone reported significantly less radiation ancillary staff. Our study was designed to assess whether
with a total of 22 mrem both with and without LT. At the 3-ft laser-guided C-arm fluoroscopy would help reduce radiation
zone, only trace amounts of radiation were detected at the exposure in the operating room because previous studies have
gonadal level, whereas the 6-ft zone reported zero radiation yielded conflicting data on whether laser guidance is helpful

TABLE 1. Number of Image Attempts TABLE 2. Fluoroscopy Time in Seconds


Region No LT With LT P Region No LT With LT P
Hip 164 173 0.5129 Hip 151 161 0.4433
Knee 143 83 ,0.0001 Knee 126 77 ,0.0001
Ankle 158 100 ,0.0001 Ankle 143 101 ,0.0001
Total 465 357 ,0.0001 Total 420 339 ,0.0001

e100 | www.jorthotrauma.com Ó 2013 Lippincott Williams & Wilkins


J Orthop Trauma  Volume 27, Number 5, May 2013 Laser Targeting with C-Arm Fluoroscopy

scatter radiation than the knee or ankle, the hip is the primary
TABLE 3. Average Centimeters From Center Per Attempt
contributor to the radiation exposure in this protocol.
Region No LT With LT P Although total fluoroscopy time was decreased by 19% using
Hip 6.56 6.03 0.1199 LT (Table 2), no difference in dosimetry data was noted
Knee 4.95 3.24 ,0.0001 between laser and non-LT groups (Fig. 4). The increase in
Ankle 5.75 3.57 ,0.0001 fluoroscopy time at the hip using LT would be expected to
Total 17.27 12.83 ,0.0001 offset any decrease in radiation exposure obtained by decreas-
ing fluoroscopy time at the knee or ankle. If different OSL
with C-arm fluoroscopy.12,13 The primary method of decreas- dosimeters were used for each body region studied in this
ing intraoperative radiation exposure is to minimize fluoros- protocol (hip, knee, and ankle), a decrease in fluoroscopy time
copy time. These data show that for common lower extremity would be expected to produce a decrease in radiation expo-
fluoroscopy procedures, the addition of C-arm LT helps sure. To reemphasize, in clinical practice, the best way to
reduce fluoroscopy time and obtain a more centered image decrease radiation exposure with C-arm fluoroscopy is to
with fewer attempts for knees and ankles but not for hips. decrease fluoroscopy time.
Statistically significant reductions in fluoroscopy time Our protocol simulated gonadal and thyroid radiation
were obtained following the addition of LT for the knee and exposure. A statistically significant increase in exposure was
ankle (39% and 29% reduction, respectively). We observed noted for the gonadal region versus thyroid at the 1- and 3-ft
that the addition of LT in joints with bony landmarks (ie, zones. These results were expected because of the effect of
patella, medial and lateral malleoli) aided in obtaining scatter radiation and were shown in a previous report
a centered image with deviations from center reduced 34% (Fig. 1).8,14 As x-rays travel through matter, scatter radiation
for knees and 38% for ankles using LT (both statistically is produced. Increasing the amount of matter through which
significant). Without the presence of defined bony landmarks x-rays travel increases the scatter radiation. In addition,
in our protocol using an obese cadaver (BMI 43.6), it was not increased scatter radiation was generated in the protocol
surprising to note that LT did not help with image acquisition because of the length of fluoroscopy time.
at the hip with no statistically significant reduction in Crawford et al15 showed that the average fluoroscopy
fluoroscopy time, no reduction in number of attempts to time for various orthopaedic trauma cases (ie, intramedullary
obtain a centered image, and no reduction in distance from nailing of femurs and tibias) ranged from 1.5 to 2.23 minutes
center per attempt. Support for this assertion was noted from depending on the case. Our total fluoroscopy time
the statistically significant increase in the average distance (12.65 minutes; 7 minutes NLT, 5.65 minutes LT) is about
from center per attempt comparing the hip to the knee and 6 times higher than an average orthopaedic trauma case. In
ankle with LT (Table 3). a 1997 report by Mehlman and DiPasquale,8 a 10-minute
It is important to highlight that fluoroscopy time C-arm exposure was used to establish the “3-ft rule,” where
actually increased using LT for the PA of the hip (151 vs. there was no detection of radiation at neck and waist level. It
161 seconds). This increase in fluoroscopy time at the thickest should be emphasized that the previous 1997 report had a limit
part of the patient would be expected to yield the highest of detection of 10 mrem with this report using a limit of 1
contribution to OSL dosimetry data because of scatter mrem. Our data demonstrated that, at 3 ft, there was only 1
radiation (Fig. 1). Because the hip would generate greater mrem of radiation detected at the thyroid level and a total of
10 mrem detected at the gonadal level for 12.65 minutes of
total fluoroscopy time. The previous report reinforced the use
of personal radiation protection for “individuals working 24
inches (70 cm) or less from a fluoroscopic x-ray beam.”8 With
improved sensitivity in this study, personal radiation protec-
tion should be worn even if greater than 3-ft away from the
C-arm radiation source.
LT helped with imaging knees and ankles with
statistically significant reductions in fluoroscopy time and
a statistically significant improvement of image quality
defined as obtaining a centered PA image faster. The
dosimetry badges detected minimal exposure at 3 ft and no
detectable exposure at 6 ft at both levels.

ACKNOWLEDGMENTS
FIGURE 4. Radiation OSL dosimeter data (n = 4 per location).
Scatter radiation is 0.1% of beam energy at 3 ft and 0.025% of The authors would like to thank Nina Clovis and
beam energy at 6 ft. At 1 ft, simulated gonadal dose was Suzanne Smith for their help with cadaveric acquisition and
12-fold greater than simulated thyroid dose. At 3 ft, there was manuscript review. They would like to thank Jackie Sions,
a significant reduction in radiation dose and no radiation dose Myra Jo Beach, and Brian Grose for granting them access to
was detected at 6 ft. the operating room to complete this research project. The

Ó 2013 Lippincott Williams & Wilkins www.jorthotrauma.com | e101


Shuler et al J Orthop Trauma  Volume 27, Number 5, May 2013

Radiation Safety Office and Nasser Razmianfar provided the 4. Verlaan JJ, Dhert WJ, Verbout AJ, et al. Balloon vertebroplasty in com-
dosimeters and a project specific radiation dosimetry report. bination with pedicle screw instrumentation: a novel technique to treat
thoracic lumbar burst fractures. Spine (Phila Pa 1976). 2005;30:E73–E79.
Dr Gerald Hobbs performed statistical analysis. The authors 5. Shortt CP, Al-Hashimi H, Malone L, et al. Staff radiation doses to the
thank Dee Headley for allowing them access to a C-arm unit lower extremities in interventional radiology. Cardiovasc Intervent
and providing the licensed technicians to complete this Radiol. 2007;30:1206–1209.
project. Specifically, they would like to thank Lauren Beiter, 6. Vano E, Ubeda C, Leyton F, et al. Radiation dose and image quality
Donna Bindernage, Amanda Brenneman Shelby Chasko, for paediatric interventional cardiology. Phys Med Biol. 2008;53:
4049–4062.
Ashley Clutter, Kevin Doorley, David Fries, Melissa 7. Ho P, Cheng SW, Wu PM, et al. Ionizing radiation absorption of vascular
May, Valerie Moyer, Luvinia Rice, Kirk Rogozinksi, surgeons during endovascular procedures. J Vasc Surg. 2007;46:
Geoffrey Rozak, Gina Short, Stephanie Sinsel, Abbey Swiger, 455–459.
and Melinda Wright for participation in this experiment. They 8. Mehlman CT, DiPasquale TG. Radiation exposure to the orthopaedic sur-
thank Markus Wagner for permission to use Figure 2 and gical team during fluoroscopy: “how far away is far enough?”. J Orthop
Trauma. 1997;11:392–398.
RadioGraphics and Beth Schueler for permission to use the 9. Singer G. Occupational radiation exposure to the surgeon. J Am Acad
diagram in Figures 1 and 4. Orthop Surg. 2005;13:69–76.
10. Giblin JG, Rubenstein J, Taylor A, et al. Radiation risk to the urologist
during endourologic procedures, and a new shield that reduces exposure.
REFERENCES Urology. 1996;48:624–627.
11. Giachino AA, Cheng M. Irradiation of the surgeon during pinning of
1. US Food and Drug Administration. White paper: initiative to reduce
femoral fractures. J Bone Joint Surg Br. 1980;62:227–229.
unnecessary radiation exposure from medical imaging. Available at:
12. Conn KS, Hallett JP. A simple laser guide to reduce the screening time
http://www.fda.gov/Radiation-EmittingProducts/RadiationSafety/ during the insertion of dynamic hip screws. Injury. 1998;29:539–541.
RadiationDoseReduction/ucm199994.htm. Accessed December 7, 2010. 13. Harris I, Walker PM, Trieu L. Radiation exposure using laser aiming
2. Levin PE, Schoen RW Jr, Browner BD. Radiation exposure to the sur- guide in orthopaedic procedures. ANZ J Surg. 2002;72:349–351.
geon during closed interlocking intramedullary nailing. J Bone Joint Surg 14. Schueler BA, Vrieze TJ, Bjarnason H, et al. An investigation of oper-
Am. 1987;69:761–766. ator exposure in interventional radiology. Radiographics. 2006;26:
3. Giordano BD, Baumhauer JF, Morgan TL, et al. Patient and surgeon 1533–1541.
radiation exposure: comparison of standard and mini-C-arm fluoroscopy. 15. Crawford CH, Seligson D, Malkani AL, et al. Time of fluoroscopy use in
J Bone Joint Surg Am. 2009;91:297–304. orthopedic trauma cases. Osteo Trauma Care. 2003;11:84–87.

e102 | www.jorthotrauma.com Ó 2013 Lippincott Williams & Wilkins

You might also like