You are on page 1of 5

SHORT COMMUNICATION

Radiation protection guidelines for the


practicing orthodontist
Muralidhar Mupparapu
Newark, NJ

This article summarizes the most recent (December 2003) dental x-ray guidelines from the National Council
on Radiation Protection and Measurements report #145. The guidelines are intended for all dental
health-care providers. They address radiation dose limits for occupational and nonoccupational exposure
and radiation protection for operators, patients, and the public. Equipment design can play an important role
in radiation protection, and recommendations from the guidelines are discussed. (Am J Orthod Dentofacial
Orthop 2005;128:168-72)

T
he use of ionizing radiation in medical and radiation-induced cancer occurs or it does not. There is
dental health care is well regulated in the no threshold.
United States. The federal government has es- The NCRP has established recommended dose
tablished performance standards for the manufacture limits for occupational and public exposure (Table I).
and installation of x-ray equipment designed for clini- Annually, the average person in the United States
cal use.1,2 In addition, some states have implemented receives about 100 mrem (1mSv) of natural background
regulations that govern users, including dentists. These radiation excluding the radon measurement. The radi-
regulations deal with the design of facilities, especially ation protection recommendations pertain to man-made
radiation shielding, and the use and maintenance of sources other than medical and dental diagnostic x-
equipment. radiation.
The National Council on Radiation Protection and For continuous or repeated exposures, the annual
Measurements (NCRP) is a nonprofit organization effective dose limit is 100 mrem or 1 mSv. Published
chartered by Congress in 1964 to collect, analyze, data3 indicate that average dental occupational expo-
develop, and disseminate information and recommen-
sures are generally a small fraction of the limit and are
dations in the public interest about (1) protection
far less than that of most other health-care workers
against radiation and (2) radiation measurements, quan-
(Table II).
tities, and units related to radiation protection. The
NCRP report #145 supersedes the report for den-
NCRP is working to develop basic guidelines about
tistry, #35, published in 1970.4 The NCRP uses shall
radiation quantities, units, and measurements and their
application to the field of radiation protection. and shall not as well as should and should not in its
report. Shall and shall not mean that “adherence to the
BIOLOGICAL EFFECTS OF RADIATION recommendation is considered necessary to meet ac-
Biological effects of radiation fall into 2 categories: cepted standards of protection.” Should and should not
deterministic and stochastic. Deterministic effects oc- are used to indicate “a prudent practice to which
cur in anyone who receives a dose of radiation that exceptions may occasionally be made in appropriate
exceeds a certain threshold. The severity of the effect is circumstances.”
proportional to the dose. Examples include acute radi- Although the guidelines cover broad areas for the
ation sickness, cataracts, and epilation. Stochastic ef- general dental practitioner, orthodontists should be
fects, such as cancer, are all-or-none effects. Either a especially concerned with the recommendations for
extraoral radiography, including panoramic, lateral, and
Associate professor and director, Division of Oral and Maxillofacial Radiology,
Department of Diagnostic Sciences, University of Medicine and Dentistry of posteroanterior cephalometric views. The general
New Jersey Dental School. guidelines govern the use of leaded aprons and thyroid
Reprint requests to: Dr M. Mupparapu, Diagnostic Sciences, D-860, New
Jersey Dental School, 110 Bergen St, Newark, NJ 07101-1709; e-mail,
collars, rectangular collimation for intraoral radio-
m.mupparapu@umdnj.edu. graphs, selection criteria, radiation protection programs
Submitted, November 2004; revised and accepted, April 2005. in offices, radiation safety training for all staff, mea-
0889-5406/$30.00
Copyright © 2005 by the American Association of Orthodontists. sures to minimize sight development for evaluation of
doi:10.1016/j.ajodo.2005.04.014 films, and qualified experts to determine the shielding
168
American Journal of Orthodontics and Dentofacial Orthopedics Mupparapu 169
Volume 128, Number 2

Table I. NCRP recommended radiation dose limits for occupational and public exposure
Dose limits

Basis Occupational Public

Stochastic 5 rem (50 mSv)* annual effective dose 0.1 rem (1 mSv) frequent exposure
1 rem (10 mSv) ⫻ age (y) cumulative effective dose 0.5 rem (5 mSv) for infrequent exposure
Deterministic effects 15 rem (150 mSv) for lens of eye 1.5 rem (15 mSv) for lens of eye
15 rem (150 mSv) for skin, hands, and feet 5 rem (50 mSv) for skin, hands, and feet
Embryo and fetus 0.05 rem (0.5 mSv)/month once pregnancy is known

1 mSv (0.001Sv) ⫽ 0.1 rem.


*1 Sv ⫽ 100 rem.

Table II.Average dental exposure by mean whole-body dose (WBD) to health-care workers compared with other
occupations (data from Kumazawa et al3 and NCRP1)
Measurement Occupational subgroup

Type of exposure Hospital Medical offices Dental Podiatry Chiropractic Veterinary Total

WBD for all workers with potential 1.4 1.0 0.2 0.1 0.3 0.6 0.7
occupational exposure (mSv)
WBD in workers who are exposed (mSv) 2.0 1.8 0.7 0.3 0.8 1.1 1.5

1 mSv ⫽ 0.1 rem.

requirements for new equipment or when an older 2. For symptomatic patients, the radiographic ex-
facility is remodeled. amination shall be limited to images that are
In addition, report #145 covers “collimation guide- required for the treatment planning of the current
lines for cephalometric and panoramic radiography.” disease.
This will be discussed later in this article. 3. For asymptomatic patients, the radiographic exam-
The NCRP report includes some general guidelines inations should be conducted based on published
for all clinicians: selection criteria.5-7
1. No one shall be permitted to receive an occupa- 4. Administrative use of radiation shall not be permit-
tional effective dose in excess of 5 rem (50 mSv) in ted.
any year (Table I). 5. Radiation exposure that is desirable per image is
2. Facility design, x-ray equipment performance, and generally a function of the film/receptor speed and
operating procedures should be established to main- based on published diagnostic reference levels at
tain patient, occupational, and public exposures as skin entry; their applicability varies from state to
low as reasonably achievable, with economic and state.8
social factors taken into account. 6. The x-ray equipment shall meet or exceed all
3. All radiographic examinations shall be performed applicable government requirements and regula-
only on direct prescription of the dental practitioner tions. Portable x-ray machines shall not be used
or physician after a clinical history and physical when fixed installations are available.
examination of the patient, and determination of a 7. The operating potential of the x-ray machines
reasonable expectation of a health benefit to the should be between 60 and 80 KVp. The operating
patient. potential should not be lower than 50 KVp or
higher than 100 KVp.
The report also lists radiation protection guidelines 8. All position-indicating devices shall be open-ended
for patients: with provision for attenuation of scattered radiation
1. For each new or referred patient, the dentist shall arising from the collimator or filter.
make a good-faith attempt to obtain recent, perti- 9. Source-to-image receptor distance for intraoral ra-
nent radiographs from the patient’s previous den- diography should not be less than 40 cm.
tist. Radiographic examinations shall be performed 10. Rectangular collimation of the x-ray beam shall be
only when indicated by patient history, physical routinely used for periapical and bitewing radiog-
examination, or laboratory findings. raphy (Fig 1).
170 Mupparapu American Journal of Orthodontics and Dentofacial Orthopedics
August 2005

11. Image receptors of speeds slower than American


National Standards Institute (ANSI)11 speed E
films shall not be used for intraoral radiography.
Faster films should be evaluated and used if
acceptable.
12. For extraoral dental radiographic projections,
high-speed (400 or greater) rare-earth screen-film
systems or digital imaging systems of equivalent
or greater speed shall be used.
13. Dental radiographic films shall be developed ac-
cording to the time-temperature method and the
chemistry according to the manufacturer’s instruc-
tions. Sight development shall not be used.
14. Use of leaded aprons for full mouth or panoramic
examinations is a prudent but not essential prac-
tice. Gonadal doses from current full-month series Fig 1. Rectangular collimator (arrow) attached to open-
or panoramic examinations do not exceed 5 ␮Gy ended round x-ray position-indicating device (PID)
(0.0005 rad).9 A significant portion of this gonadal
dose results from scattered radiation in the pa-
tient’s body. Leaded aprons do not significantly
reduce these doses.1 The guidelines state that
leaded aprons on patients are not required if all
other recommendations in the report are rigor-
ously followed.
15. Thyroid collars shall be provided for children and
should be provided for adults (Fig 2), when they
will not interfere with the examination.
The report includes guidelines to protect the oper-
ator:

1. New or remodeled offices should have shielding


designed by a qualified expert to create adequate
protective barriers. The barriers shall be con- Fig 2. Thyroid collar and lead apron for intraoral
structed so that operators can maintain visual con- radiography.
tact and communication with patients throughout
the procedures.
ties shall be designed so that no member of the public
2. The operator should be standing at a minimum
will receive an effective dose in excess of 100 mrem or
distance of 2 meters from the tube head. If this
0.1 rem (1 mSv) annually.
minimum distance cannot be maintained, a barrier
shall be provided. EQUIPMENT DESIGN
3. Monitoring of individual occupational exposures
For optimal use of radiation in dentistry, the Food
for office staff and personnel is required for those
and Drug Administration has developed performance
reasonably expected to receive a significant dose.
standards for medical and dental x-ray machines. Com-
The NCRP recommends personal dosimeters for
pliance with these standards at installation is required
external exposure measurement for workers who
for all machines manufactured in the United States after
are likely to receive an annual effective dose in
1975.2 The NCRP recommends that the x-ray machines
excess of 1 mSv and also for pregnant occupation-
shall provide the range of exposures that are suitable for
ally exposed personnel.
use with the fastest image receptors available.
Finally, the report outlines steps that should be It is the responsibility of the clinician who owns the
taken to protect the public, including people in recep- office to make sure that the tube head is free of drifts
tion rooms, other treatment areas in the same office, and and oscillations. The operator is prohibited from hold-
adjoining corridors in the building. New dental facili- ing the tube head during exposure.10
American Journal of Orthodontics and Dentofacial Orthopedics Mupparapu 171
Volume 128, Number 2

Equipment design for intraoral radiography shall be Clinicians who routinely perform cephalometric
capable of providing rectangular collimation. Collimat- radiography should note the following. Generally, the
ing the x-ray beam to the precise size of the image area of clinical interest in cephalometric radiography is
receptor eliminates scatter radiation. Scatter radiation significantly smaller than the image receptor. The
in general decreases the quality of the image. “Rectan- central axis of the beam is usually aligned through the
gular collimation of the beam shall be used routinely external auditory canals, positioned by the ear rods of
for periapical radiography. Each dimension of the beam the cephalostat. Imaging of the structures superior to
measured in the plane of the image receptor shall not the superior orbital rim, posterior to the occipital
exceed the dimension of the image receptor by more condyles, and inferior to the hyoid bone is clinically
than 2% of the source-receptor distance. Similar colli- unnecessary. This prevents unnecessary exposure to the
mation should be used when feasible for interproximal patient’s hard and soft tissues. It is also a standard
(bitewing) radiography.” recommendation from the NCRP that the soft tissue
facial profile should be imaged along with osseous
SPECIFIC GUIDELINES FOR THE ORTHODONTIST structures of the face. This is accomplished by reducing
For all intraoral radiography, the NCRP recommends exposure to anterior soft tissues.
using ANSI11 film speeds E and above when available. Clinicians should switch from the slower, blue
For extraoral radiography, blue fluorescing calcium tung- fluorescent screens to the faster green fluorescent
state screens that are considered slow are no longer screens. Kodak T mat G or Kodak Ektavision15 (East-
recommended because they increase the patient’s skin and man Kodak Company, Rochester, NY) screens with
absorbed x-ray dose. The specific NCRP recommendation their speed-matched film combinations are examples of
is: “The fastest imaging system consistent with the imag- fast screens. Filters for imaging the soft tissues of the
ing task shall be used for all extraoral dental radiographic facial profile and the facial skeleton shall be placed at
projections. High speed (400 or greater) rare-earth screen- the x-ray source rather than at the image receptor. This
film systems or digital systems of equivalent or greater equipment modification might not be easy to achieve
speed shall be used.” without purchasing new cephalometric or panoramic-
For direct digital radiograpy, in which the film- cephalometric combination machines (Fig 3). The new
based image is replaced by a digital image consisting of digital cephalometric machines come with this kind of
a 2-dimensional array of pixels, the latent image is built-in soft tissue filtering. This might be an added
recorded directly on a suitable digital sensor. Receptors incentive for the practitioner to convert to digital
include photostimulable phosphors, charge-coupled de- receptors when replacing older equipment. The dose
vices, and complementary metal-oxide semiconductors, reduction might be substantial, because digital systems
which are also known as active pixel sensors. The latent are much thriftier than existing screen-film systems,
image is later digitally processed to produce an elec- and the built-in soft tissue filters will eliminate the
tronic image. The electronic image can be displayed on external cassette-side filtration. According to the NCRP
a computer monitor, converted to a hard copy, or report, “Practitioners need to remember that all struc-
transmitted electronically. Although the guidelines dis- tures recorded on the image need to be interpreted for
cuss the use of direct digital radiography at length, evidence of disease or injury as well as for cephalo-
there are no specific recommendations to replace the metric analysis.”
existing fast film for intraoral radiography. The report
agreed with the current published literature regarding CONCLUSIONS
the detection efficiency of digital receptors with that of NCRP report #145 is relatively new, and many in
conventional film when investigating occlusal and the dental profession have not yet taken a serious look
proximal caries, periodontal bone lesions, periapical at it. This important report defines standards of care in
bone lesions, and root canal systems.12-14 terms of radiation protection for the operator, the
Rotational panoramic machines use a narrow verti- patient, and the public. It is certain to make an impact
cal beam, exposing only a small portion of the image on the radiographic practices in most oral health-care
receptor at any time. The NCRP’s recommendation is: settings.16 It is the responsibility of all dental health-
“The x-ray beam for rotational panoramic tomography care providers to reexamine their practices, equipment,
shall be collimated such that its vertical dimension is no protective devices, and receptor selection.
greater than that required to expose the area of clinical Clinicians should consider converting to digital
interest. In no case shall it be larger than the slit in the radiographic modalities for improved compliance, ra-
image-receptor carrier plus a tolerance of two percent diation dose reduction, and ease of storage, retrieval,
of the source-to-image receptor distance.” and transfer of information.
172 Mupparapu American Journal of Orthodontics and Dentofacial Orthopedics
August 2005

12. Kullendorf FB, Nilsson M, Rohlin M. Diagnsotic accuracy of


direct digital dental radiography for the detection of periapical
bone lesions: overall comparison between conventional and
direct digital radiography. Oral Surg Oral Med Oral Pathol
Radiol Endod 1996;82:344-50.
13. Hintze H, Wenzel A, Jones C. In vitro comparison of D and E
speed film radiography, RVG and Visualix digital radiography
for the detection of enamel approximal and dentinal occlusal
caries lesions. Caries Res 1994;28:363-7.
14. Sanderink GC, Huiskens R, van der Stelt PF, Welander US,
Stheeman SE. Image quality of direct digital intraoral x-ray sensors
in assessing root canal length. The RadioVisioGraphy, Visualix/
VIXA, Sens-A-Ray and Flash Dent systems compared with Ekta-
speed films. Oral Surg Oral Med Oral Pathol 1994;78:125-32.
15. Kodak T-MAT G dental film and Kodak Ektavision G dental
film. Available at: http://www.kodak.com. Accessed November
20, 2004. Accessed April 27, 2005.
16. Miles DA, Langlais RP. NCRP report no. 145. New dental x-ray
Fig 3. Newer-generation digital radiography machine guidelines: their potential impact on your dental practice. Dent
has built-in soft tissue filter. (Planmeca Promax pan- Today 2004;9:128-34.
oramic-cephalometric combination machine).
COMMENTARY

REFERENCES The NCRP guidelines state that “all radiographic


1. National Council on Radiation Protection and Measurements.
examinations shall be performed only on direct prescrip-
Radiation protection in dentistry (report no. 145). Bethesda, Md: tion of the dental practitioner or physician after conduct of
NCRP; 2003. a clinical history and physical examination of the patient,
2. Food and Drug Administration. Performance standards for ion- and determination of a reasonable expectation of a health
izing radiation emitting products, 21 CFR Sect. 1020 (1995). benefit to the patient.” The guidelines also direct clinicians
3. Kumazawa S, Nelson DR, Richardson ACB. Occupational ex-
posure to ionizing radiation in the United States: a comprehen-
to “make a good-faith attempt to obtain recent, pertinent
sive review for the year 1980 and a summary of trends for the radiographs from the patient’s previous dentist. Radio-
years 1960-1985. EPA 520/1-84/005 (National Technical Infor- graphic examinations shall be performed only when indi-
mation Service, Springfield, Va). cated by patient history, physical examination or by
4. National Council on Radiation Protection and Measurements laboratory findings.”
(NCRP). Dental x-ray protection (report no. 35). Bethesda, Md:
NCRP; 1970.
It has become common practice in orthodontic clinics
5. Joseph LP. The selection of patients for x-ray examinations: to “quick start” potential patients on the first visit to the
dental radiographic examinations. HHS Publication No. FDA office. Some practice consultants advise giving the pro-
88-8273. Rockville, Md: The Dental Radiographic Patient Se- spective patient an “office tour” that includes a stop at the
lection Criteria Panel, DHHS, Center for Devices and Radiolog- x-ray machine to take panormic and cephalometric films,
ical Health; 1987.
6. Matteson SR, Joseph LP, Bottomley W, Finger HW, Frommer
before introducing the patient to the orthodontist. This is
HH, Koch RW, et al. The report of the panel to develop done under the assumption that many patients will actu-
radiographic selection criteria for dental patients. Gen Dent ally begin treatment that day, including spacers.
1991;39:264-70. Of course, not all prospective patients are ready to
7. ADA Council on Scientific Affairs. An update on radiographic start treatment immediately or to commit to being treated
practices: information and recommendations. J Am Dent Assoc
2001;132:234-8.
at that particular office, so this practice results in unnec-
8. Conference of Radiation Control Program Directors, Inc. Patient essary exposure for some and is a violation of the guidelines.
exposure and dose guide. CRCPD publication E-03-2. Frankfort, Also, more effort should be made to obtain previously taken
Ky: Conference of Radiation Control Program Directors Inc; films to avoid duplication of exposure, even though it seems
2003. faster and more cost-efficient to take new films. This is a
9. White SC. 1992 assessment of radiation risk from dental radi-
ography. Dentomaxillifac Radiol 1992;21:118-26.
good example of the push/pull of orthodontic marketing and
10. International Electrotechnical Commission. Medical electrical the resulting impact on patient care.
equipment. Part I: general requirements for safety. 3. Collateral Jan Bell
standard: general requirements for radiation protection in diag- Seattle, Wash
nostic x-ray equipment. SS-EN 60601-1-3. Geneva, Switzerland:
International Electrotechnical Commission; 1994. Am J Orthod Dentofacial Orthop 2005;128:172
11. American National Standards Institute. Photography-intraoral 0889-5406/$30.00
dental radiographic film specifications: ANSI/ISO 3665. New Copyright © 2005 by the American Association of Orthodontists.
York: Amercican National Standards Institute; 1996. doi:10.1016/j.ajodo.2005.04.015

You might also like