You are on page 1of 6

Radiation Protection Dosimetry (2020), pp. 1–6 doi:10.

1093/rpd/ncaa026

Downloaded from https://academic.oup.com/rpd/article-abstract/doi/10.1093/rpd/ncaa026/5823517 by James Cook University user on 23 April 2020


EVALUATION OF THE RADIATION DOSE TO THE HANDS OF
ORTHOPAEDIC SURGEONS DURING FLUOROSCOPY USING
STORED IMAGES
Carlo Giordano1, *, Ivo Monica2 , Fabrizio Quattrini2 , Elena Villaggi1 , Rossana Gobbi1
and Lorenzo Barbattini3
1
Medical Physics Department, AUSL Piacenza, via Taverna 49, 29121 Piacenza, Italy
2
Orthopaedics and Traumatology Department, AUSL Piacenza, via Taverna 49, 29121 Piacenza, Italy
3
Radiology Department, AUSL Piacenza, via Taverna 49, 29121 Piacenza, Italy

*Corresponding author: c.giordano@ausl.pc.it

Received 20 September 2019; revised 16 December 2019; editorial decision 12 February 2020; accepted 12
February 2020

Data were collected from 642 orthopaedic interventions during which the images produced by X-rays were recorded. By examining
these images, it is possible to determine the time that the orthopaedic surgeons’ hands were exposed to the direct radiation beam.
The procedures with greater exposure to the direct beam were those involving the hand (median 15 s) and the wrist (median
13 s). Two surgeons wore a ring to measure the absorbed dose at the fingers: one on the dominant hand and the other on the
non-dominant hand. The two surgeons performed 34 and 48 operations, respectively, in 14 months. The total doses measured
with the rings were 2.30 and 1.04 mSv, respectively. The images of the interventions were examined, determining how much each
individual hand was exposed. The interventional reference point (IRPeff (left or right) ) was calculated by comparing the doses at
the IRP with the exposure times of the right or the left hand. Summing the IRPeff of the two surgeons in 14 months, it is obtained
the maximum values of 2.87 mGy for the left hand of one and 6.74 mGy for the right hand of the other, which are of the order
of 1/100 of the annual dose limit for the extremities.

INTRODUCTION In most cases, radiation exposure to the hands of


orthopaedic surgeons derives from scattered radia-
X-ray units have been used extensively in orthopaedic
tion but there are situations in which there is also a
operating theaters since the 1980s(1) . The image
contribution from primary radiation. According to
intensifiers of that period were not able to memorize
Arnstein et al.(13) , the dose within the primary beam is
the last image acquired, which led to an increase in
100 times higher than the scattered radiation at 15 cm
the times of X-ray delivery and consequently of the
and according to Mahajan et al.(14) , the scattered
exposure of the operators to radiation. Technological
dose at 46 cm is 0.1% of the dose from the direct beam.
developments subsequently led to the introduction of
International Commission of Radiological Protection
various improvements that have reduced the exposure
117(1) states that ‘exposure of staff may be unavoid-
of both patients and theater staff to radiation, such
able when maintaining a difficult reduction, confirm-
as the storage of images and greater filtering of the
ing adequate bony alignment or securing implant
radiation beams as well as marketing of specifically
placement’, even if in these cases ‘there is evidence of
designed equipment (mini C-arms).
poor radiological protection practices.’
The exposure of orthopaedic surgeons to radiation
The purposes of this study were to assess the
is heterogeneous, because they must normally wear
level to which the hands of orthopaedic surgeons are
lead gowns and therefore the parts exposed, in
exposed in our center and in particular if and how
particular the hands, are substantially those not
important the contribution of direct beam to the total
covered by the protective clothing. Many studies
absorbed dose is.
that have evaluated the exposure of staff involved in
orthopaedic procedures have been published in the lit-
erature. In particular, the review by Kesavachandran
et al.(2) presents data from various authors(3–11) who
MATERIALS AND METHODS
reported annual doses at the extremities far below the
dose limit (500 mSv). However, Tsalafoutas et al.(12) Data from 642 operations carried out between
believe that there is a lack of reliable data as many December 2017 and April 2019 were collected and
theater staff do not wear a dosemeter regularly. analyzed. The mobile C-arm system used in these

© The Author(s) 2020. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com
C. GIORDANO ET AL.
orthopaedic procedures was a Philips BV Pulsera 9R2 on the monitor in every image. Thus, the exposure
equipped with an image intensifier with a diameter of times to the direct beam of the dominant and
23 cm, with the tube positioned below the operating non-dominant hand of each doctor have been

Downloaded from https://academic.oup.com/rpd/article-abstract/doi/10.1093/rpd/ncaa026/5823517 by James Cook University user on 23 April 2020


table. The interventional reference point (IRP) was derived (T hand (left or right) ). Finally, a parameter IRP
at a distance of 30 cm from the intensifier in the effective (left or right) = IRP dose × T hand (left or right) /T fluoro
direction of the tube: this position approximately was also determined, which indicates the dose at the
represents where the beam enters the patient and IRP given during the exposure of each extremity to
also, occasionally, the surgeon’s hands. During each the direct beam.
intervention, the fluoroscopy images were stored,
at an average rate of two images per second, and,
at the end of each procedure, were sent to the
RESULTS
picture archiving and communication system (PACS)
together with the radiation dose structured report The data relating to the range, the median and the
(RDSR): the latter reports the kerma-area product 75th percentile of the fluoroscopy times, the dose to
(KAP), the air kerma at the IRP and the time of the IRP and the KAP are shown in Table 1. The IRP
fluoroscopy. For each patient, all the recorded images dose and the KAP were corrected for the calibration
(N im ) were visually examined, counting the number factors, which, for the C-arm used, were 1 for the IRP
in which the surgeon’s hands appeared (N hand ), which dose and 1.07 for the KAP. In 75% of the cases the
in this case intercepted the direct beam of radiation. fluoroscopy time was <74 s, with the exception of the
When counting the N hand images no distinction was ‘miscellaneous’ procedures, which were generally in
made between hands intercepted by the beam totally, patients with multiple trauma.
partially or individually. Given that the number of The medians of the data related to the wrist, tibia,
images (N im ) is proportional to the total time of hand, ankle, humerus, shoulder and femur are in
fluoroscopy (T fluoro ), the time during which the line with those reported by Giordano et al.(15) with
hands were exposed to the direct beam (T hand ) was differences being generally <30%.The data relating
determined as follows: T hand = (N hand /N im ) × T fluoro . to the evaluation of the contribution of the direct
The operations were divided according to the beam to the radiation dose to the hands are shown
anatomical district involved, but the type of interven- in Table 2. It can be seen that the anatomical districts
tion performed was not taken into account. Images that required longer operating times were the extrem-
that included different districts and that could not be ities of the upper limbs (hand and wrist). Among
unequivocally categorized were grouped in a separate the 17 ‘miscellaneous’ procedures, those in which the
category called ‘miscellaneous.’ The anatomical exposure times to the direct beam were greater than
districts identified were: hand, wrist, forearm, elbow, the median (8 in total) included four involving the
humerus, shoulder, foot, ankle, tibia-fibula, knee upper limbs. The others were operations for multiple
and femur. The interventions were performed by trauma to the lower limbs. It can be seen that, despite
13 surgeons, who were assigned, as normal, cuff the relatively long fluoroscopy times (median 38 s)
dosemeters to measure the radiation received by the (Table 1), the operator’s hands were barely exposed
extremities, and a dosemeter to be attached to the to the direct beam during operations involving the
lead gown to evaluate the effective dose. Two of the 13 femur. An intermediate situation was noted for oper-
surgeons made themselves available for more accurate ations to the humerus during which the fluoroscopy
measurements using an additional ring dosemeter. times were longer overall (median 53 s), and the
The dosemeters, whose reading is supplied in Hp extremities were exposed to radiation for a median of
(0.07), were produced by the Tecnorad company of 7 s.
Verona (Italy), and were replaced every 2 months. In the period between March 2018 and April 2019,
The types of dosemeter used were TLD LiF: surgeons A and B performed 46 and 58 procedures,
MCP GR-200A or MCP-N. The observation period respectively. In some interventions, the two surgeons
was 14 months and therefore there were seven were listed as ‘assistant surgeon’ and not as ‘first sur-
dosimetric results available per surgeon, although geon’; since it is the hands of the first surgeon that are
the same surgeons did not always work during these typically exposed to the direct radiation beam, only
periods. Surgeon A wore the ring dosemeter on the cases in which the two surgeons were recorded
the little finger of the non-dominant hand, while as ‘first surgeon’ were considered. The total number
surgeon B wore it on the ring finger of the dominant of interventions therefore fell to 34 for surgeon A
hand. Both surgeons had >15 years of experience, and 48 for surgeon B. The districts on which surgeon
neither is left handed. At the end of each 2-month A operated were the ankle (n = 1), femur (n = 23),
period, the surgeons involved returned forms with hand (n = 1), humerus (n = 2), wrist (n = 3), tibia
a description of each intervention carried out and (n = 2), knee (n = 1) and miscellaneous (n = 1); those
subsequently they viewed the images recorded on of surgeon B were the ankle (n = 2), femur (n = 16),
the PACS distinguishing if and which hand appeared elbow (n = 2), hand (n = 5), humerus (n = 2), tibia

2
EVALUATION OF THE RADIATION DOSE TO THE HANDS OF ORTHOPAEDIC SURGEONS
Table 1. Summary of fluoroscopy time, dose (mGy) at the IRP and KAP data.

N Time (s) Dose at IRP (mGy) KAP (μGy m2 )

Downloaded from https://academic.oup.com/rpd/article-abstract/doi/10.1093/rpd/ncaa026/5823517 by James Cook University user on 23 April 2020


Range Median 75th Range Median 75th Range Median 75th

FEMUR 218 1–194 38 54 0.01–27.3 3.40 5.6 0.14–508.2 63.7 106.9


KNEE 33 2–90 10 27 0.04–4.2 0.30 1.1 0.71–76.1 5.5 14.0
TIBIA 77 0–195 42 74 0.01–7.7 0.68 1.4 0.11–134.8 14.7 30.3
ANKLE 50 3–142 25 33 0.02–2.2 0.37 0.57 0.33–40.6 6.3 10.7
FOOT 50 1–102 5 13 0.003–1.3 0.05 0.15 0.06–24.4 0.90 2.4
SHOULDER 24 1–73 11 31 0.06–3.2 0.54 1.2 1.66–60.5 10.1 21.8
HUMERUS 51 3–139 53 74 0.03–4.9 1.38 2.0 0.47–101.2 25.8 41.1
ELBOW 27 4–202 19 40 0.05–4.5 0.25 0.62 0.85–82.8 4.5 11.5
FOREARM 13 1–100 10 18 0.01–2.0 0.11 0.19 0.19–14.9 2.1 4.0
WRIST 44 1–106 28 41 0.01–1.4 0.26 0.47 0.06–32.0 4.7 7.7
HAND 38 1–223 21 44 0.01–1.6 0.16 0.35 0.10–30.0 2.4 5.4
MISCELLANEOUS 17 29–222 76 144 0.49–18.7 1.91 9.2 9.13–348.8 35.6 170.1

Table 2. Summary of the exposure times of the hands to the direct beam according to the type of procedure.

N Hands under the direct beam (s)


Range Median 75th

FEMUR 218 0–49 0 1


KNEE 33 0–15 1 4
TIBIA 77 0–132 5 14
ANKLE 50 0–43 4 11
FOOT 50 0–50 3 7
SHOULDER 24 0–25 0 4
HUMERUS 51 0–73 7 14
ELBOW 27 0–154 5 8
FOREARM 13 0–39 1 4
WRIST 44 0–76 13 20
HAND 38 0–207 15 25
MISCELLANEOUS 17 1–49 14 25

(n = 7), wrist (n = 5), radius-ulna (n = 3), knee (n = 1), The correlation between the measured dose and
shoulder (n = 1) and miscellaneous (n = 4). the time of exposure of the hands to the direct beam
The dosimetric results and some of the two sur- is strong (R2 = 0.40) for the doctor A and very strong
geons’ operating data are shown in Tables 3 and 4. (R2 = 0.81) for the doctor B, while it is very good
The dose read by the dosemeter placed on the chest that between the dose measured with the ring and
on the outside of the leaded gown was zero in all the IRPeffective left or right parameter (R2 equal to 0.97 for
cases (nominal value attributed to dosimetric read- doctor A and 0.81 for doctor B [see Figures 1 and 2]).
ings lower than 0.02 mSv). Examining Tables 3 and 4, if the sum of the col-
The nominal reading of the ring dosimeter of the umn of ‘IRPeff (left or right) ’ is added, the dose delivered
period January–February 2019 of physician B was to the IRP is obtained during exposure of the hands
zero. However, it was considered more reasonable to the direct beam throughout the period considered.
to assign the minimum declared sensitivity value The maximum total values are 2.87 mGy for doctor
(0.02 mSv) since the dosimeter was nevertheless A’s left hand and 6.74 mGy for doctor B’s right hand.
exposed, albeit for a short time. In the same Tables 3
and 4, the value of extended uncertainty percentage
is provided by the Tecnorad company. DISCUSSION
Subsequently, the relationship between the dose
measured by the ring dosemeter and the exposure The RDSR data of the operations recovered from the
parameters reported in the RDSR was studied. PACS are in line with those collected in a study of

3
C. GIORDANO ET AL.

Table 3. Exposure data of the surgeon A.

Downloaded from https://academic.oup.com/rpd/article-abstract/doi/10.1093/rpd/ncaa026/5823517 by James Cook University user on 23 April 2020


Surgeon A

N Dosimetric Extended Screening time Time with left Time with right IRPeff left IRPeff right
reading (mSv) uncertainty (s) hand under hand under (mGy) (mGy
percentage beam (s) beam (s)

2018
March–April 4 0.09 60 159 8 10 0.05 0.08
May–June 10 1.57 45 470 41 6 2.02 0.29
July–August 0 0 - 0 0 0 0 0
September–October 8 0.28 50 238 17 3 0.27 0.04
November–December 0 0 - 0 0 0 0 0
2019
January–February 7 0.20 55 278 1 8 0.11 0.56
March-April 5 0.16 55 237 11 14 0.42 0.53
Total 34 2.30 1382 78 41 2.87 1.50

Table 4. Exposure data of the surgeon B.

Surgeon B

N Dosimetric Extended Screening time Time with left Time with right IRPeff left IRPeff right
reading (mSv) uncertainty (s) hand under hand under (mGy) (mGy
percentage beam (s) beam (s)

2018
March–April 10 0.15 55 399 58 48 1.15 0.82
May–June 7 0.16 55 427 32 54 0.83 1.60
July–August 8 0.20 55 474 71 67 1.14 1.24
September–October 8 0.33 50 569 250 174 2.13 2.03
November–December 11 0.18 55 393 26 20 0.93 0.96
2019
January–February 4 <0.02 70 211 6 2 0.34 0.09
March–April 0 0 0 0 0 0 0 0
Total 48 1.04 2473 443 365 6.52 6.74

Figure 1. Dosimetric reading surgeon A vs IRPeff left .

4
EVALUATION OF THE RADIATION DOSE TO THE HANDS OF ORTHOPAEDIC SURGEONS

Downloaded from https://academic.oup.com/rpd/article-abstract/doi/10.1093/rpd/ncaa026/5823517 by James Cook University user on 23 April 2020


Figure 2. Dosimetric reading surgeon B vs IRPeff right .

patients’ exposure to radiation performed in 2014– A second reason is that the exposure of the hand
2015 in the same center(15) . The operations that to the direct beam in the orthopaedic field occurs
involve exposure to higher radiation doses to the in an irregular and usually partial way. Therefore,
hands of the orthopaedic surgeon are essentially a precise measurement could be performed only by
those to the hand and the wrist, while exposure to the assigning more dosemeters to each surgeon’s hand, a
direct beam is almost negligible in the case of femoral procedure that is impossible to achieve systematically
surgery, regardless of whether the operation involved for obvious practical reasons.
intramedullary nailing, a cannulated hip screw or In certain types of interventions (hand, wrist), the
less invasive stabilization system plates. Tables 3 and surgeon’s hand is exposed entirely for relatively long
4 show, first of all, that the exposure times for the times to the direct beam without the patient’s attenu-
hands are greater for the non-dominant hand, even ation, for others (femur, tibia) the exposure is limited
for the doctor A almost double the other. The doses to the phalanges and often the beam is attenuated by
to the IRP are higher during the exposure of the the patient, so that despite the presence of relatively
non-dominant hand for the doctor A, while for the high IRPeff values, it is possible that the extremities of
doctor B there is a slight prevalence for the dose the surgeon have instead absorbed little dose as they
dispensed to the right hand. The dosages detected by are placed on the side of the patient opposite the tube.
the dosemeter are always low except for a reading, Therefore, it can be deduced that, at the end of the
they are always <0.5 mSv. However, the measured 14 months of observation, the doses absorbed by the
data are quite different between the two surgeons and two surgeons, were surely not higher than previous
this may be due to several factors. recorded data of ‘IRP eff (left or right) ’ converted into
First of all the ring dosemeters were worn at mSv. The conversion between mGy and mSv can be
different places: on the little finger of the non- performed since in the range of energies used (average
dominant hand of surgeon A and on the ring 30–50 keV) the conversion coefficient between air
finger of the dominant hand of surgeon B. The kerma and Hp (0.07) is only slightly higher than
choice of using different positions for the ring one(18) . The doses absorbed by the surgeons in our
dosemeter derives from the fact that the literature center are comparable to those found in other studies
is not clear on which is the most exposed hand in published in the literature. Kesavachandran et al.(2)
the case of orthopaedic procedures. According to reviewed a collection of publications with annual
Vosbikan et al.(16) , the non-dominant hand appeared dosimetric data. The measured doses ranged from
to be more exposed during operations to the hands 0.03 to 1.27 mSv(3–11) . In an analysis of 80 lower
and wrists, whereas Mahajan et al.(14) found no limb operations, Singh et al.(19) evaluated that the
differences. Hafez et al.(4) reported that the radiation surgeon’s right hand was exposed to a radiation dose
dose absorbed was higher for the dominant hand, of 2.4 mSv, while Khan et al.(20) considered that the
but only for trainee surgeons: there were no practical absorbed dose to the fingers during femoral surgery
differences for experienced surgeons. According was 0.13 mSv per procedure.
to Martin and Whitby(17) , ring dosemeters should
be worn on the little fingers of both hands in CONCLUSIONS
order to monitor the radiation dose correctly at
the upper limb extremities during interventional The hands of an orthopaedic surgeon are exposed
radiology. in different ways to the direct beam of radiation

5
C. GIORDANO ET AL.
during the various interventions performed. This 8. Rampersaud, Y. R., Foley, K. T., Shen, A. C.,
direct beam is the major source of irradiation, since it Williams, S. and Solomito, M. Radiation exposure to
is some orders of magnitude greater than the contri- the spine surgeon during fluoroscopically assisted pedicle

Downloaded from https://academic.oup.com/rpd/article-abstract/doi/10.1093/rpd/ncaa026/5823517 by James Cook University user on 23 April 2020


bution from scattered radiation. The time of exposure screw insertion. Spine (Phila Pa 1976) 25, 2637–2645
to direct beam irradiation can be estimated in our (2000).
9. Faulkner, K. and Harrison, R. M. Estimation of effec-
center from stored images. In the two cases analyzed, tive dose equivalent to staff in diagnostic radiology. Phys.
a correlation was found between dose absorbed by Med. Biol. 3, 383–391 (1988).
the hands and exposure time to the direct beam. In 10. Muzaffar, T. S., Imran, Y., Iskandar, M. A. and
the light of the results from our center, as well as data Zakaria, A. Radiation exposure to the surgeon during
published in the literature, it can be concluded that the femoral interlocking nailing under fluoroscopic imaging.
dose absorbed by the hands of orthopaedic surgeons Med. J. Malaysia 60(Suppl C), 26–29 (2005).
rarely exceeds one-tenth of the maximum annual dose 11. Biswas, D., Bible, J. E., Bohan, M., Simpson, A. K.,
limit (500 mSv); in our case, in which the exposure of Whang, P. G. and Grauer, J. N. Radiation exposure
two doctors is examined, it can be concluded that in from musculoskeletal computerized tomographic scans. J.
Bone Joint Surg. Am. 91, 1882–1889 (2009).
14 months a doctor certainly did not exceed 3 mSv 12. Tsalafoutas, I. A., Tsapaki, V., Kaliakmanis, A.,
and the other 7 mSv. Furthermore, it can be assumed Pneumaticos, S., Tsoronis, F., Koulentianos, E. D.
that it may be precautionary to wear the dosimeter at and Papacristou, G. Estimation of radiation doses
the ends of the non-dominant hand, as it is exposed to patients and surgeons from various fluoroscopically
to the direct beam for longer times. guided orthopaedic surgeries. Radiat. Prot. Dosimetry
128, 112–119 (2008).
13. Arnstein, P. M., Richards, A. M. and Putney, R. The
REFERENCES risk from radiation exposure during operative X-ray
screening in hand surgery. J. Hand Surg. Br. 19, 393–396
1. International Commission of Radiological Protection. (1994).
Radiological protection in fluoroscopically guided proce- 14. Mahajan, A., Samuel, S., Saran, A. K., Mahajan, R. K.
dures performed outside the imaging department. Ann. and Mam, M. K. Occupational radiation exposure from
ICRP 40(6), 50–60 (2010) ICRP Publication 117. C arm fluoroscopy during common orthopaedic surgical
2. Kesavachandran, C. N., Haamann, F. and Nienhaus, A. procedures and its prevention. J. Clin. Diagn. Res. 9(3),
Radiation exposure of eyes, thyroid gland and hands in RC01–RC04 (2015).
orthopaedic staff: a systematic review. Eur. J. Med. Res. 15. Giordano, C., Monica, I., Gobbi, R. and Barbattini,
17, 28 (2012). L. Determination of local levels of radiation exposure
3. Müller, L. P., Suffner, J., Wenda, K., Mohr, W. and in orthopaedic interventions using structured dose report
Rommens, P. M. Radiation exposure to the hands and data drawn from the hospital picture archiving and com-
the thyroid of the surgeon during intramedullary nailing. munication system. Radiat. Prot. Dosimetry 178(19),
Injury 29(6), 461–468 (1998). 82–86 (2018).
4. Hafez, M. A., Smith, R. M., Matthews, S. J., Kalap, G. 16. Vosbikan, M. M., Ilyas, A. M., Watson, D. D. and
and Sherman, K. P. Radiation exposure to the hands of Leinberry, C. F. Radiation exposure to hand surgeons’
orthopaedic surgeons: are we underestimating the risk? hands: a practical comparison of large and mini C-
Arch. Orthop. Trauma Surg. 125, 330–335 (2005). arm fluoroscopy. J. Hand Surg. Am. 39, 1805–1809
5. Miller, M. E., Davis, M. L., MacClean, C. R., Davis, (2014).
J. G., Smith, B. L. and Humphries, J. R. Radiation 17. Martin, C. J. and Whitby, M. Application of ALARP to
exposure and associated risks to operating-room per- extremity doses for hospital workers. J. Radiol. Prot. 23,
sonnel during use of fluoroscopic guidance for selected 405–421 (2003).
orthopaedic surgical procedures. J. Bone Joint Surg. Am. 18. ENEA. Dosimetria personale esterna in termini di
65, 1–4 (1983). equivalente di dose personale, Hp (d). (Atti giornata di
6. Sanders, R., Koval, K. J., Di Pasquale, T., Schmelling, studio) 28 February 2001.
G., Stenzler, S. and Ross, E. Exposure of the orthopaedic 19. Singh, P. J., Perera, N. S. and Dega, R. Measurement of
surgeon to radiation. J. Bone Joint Surg. Am. 75(3), the dose of radiation to the surgeon during surgery to the
326–330 (1993). foot and ankle. J. Bone Joint Surg. Br. 89-B, 1060–1063
7. Levin, P. E., Schoen, R. W. Jr. and Browner, B. D. (2007).
Radiation exposure to the surgeon during closed inter- 20. Khan, I. A., Kamalasekaran, S. and Fazal, M. A. Risk
locking intramedullary nailing. J. Bone Joint Surg. Am. of ionising radiation to trainee orthopaedic surgeons.
69, 761–766 (1987). Acta Orthop. Belg. 78, 106–110 (2012).

You might also like