INSTRUCTION MANUAL
Ver.3.1
IEI
’04/03/22
2
CONTENTS
1 General.................................................................................................................3
2 Standard Specifications ......................................................................................4
3 Features of ACCURA9/9L ................................................................................5
3.1 Features ..................................................................................................5
4 Component Names..............................................................................................6
4.1 Front Panel of Controller.....................................................................6
4.2 Rear Panel of Controller......................................................................7
5 Set up ................................................................................................................8
6 Before using ACCURA 9/9L ..................................................................... 9
7 Standard Operation .........................................................................................10
7.1 Key operation and selecting screen. ...............................................10
7.2 Mode selecting ..................................................................................11
7.3 Auto Mode.........................................................................................12
7.3.1 Shot screen......................................................................................12
7.3.2 Data editing screen ........................................................................15
7.3.3 D.SAMP setting screen..................................................................18
7.4 STEADY MODE...............................................................................20
7.4.1 Shot screen..........................................................................20
7.4.2 Data editing screen………………………………………..22
7.4.3 D.SAMP setting screen......................................................24
7.5 Viscosity Mode..................................................................................25
7.6 RS232C communication setting .......................................................26
7.7 Error screen .......................................................................................26
8 Data Sampling ................................................................................................29
9 Change barrels ................................................................................................33
10 Remote Mode .................................................................................................34
10.1 RS232C Communication...................................................................35
10.2 Parallel Communication ....................................................................50
10.2.1 Input signal………………………………………………. 50
10.2.2 Output signal ………………………………………… 53
11 Interface of Input/Output signals..................................................................55
11.1 Connector and Pin layout ................................................................55
11.2 Connect Input and Output Signal ...................................................58
11.3 Input/Output Timing Chart...............................................................62
12 Memory Backup Battery ...............................................................................64
13 Maintenance / Consumable Parts .................................................................65
14 Appearance of the Controller .......................................................................66
1 General
ACCURA 9/9L maintains shot volumes constantly at any variation of the material balance in the barrel by
means of IEI’s original shot volume control system. The newly incorporated shot-pressure compensation
function can acknowledge the material level in the barrel constantly and compensate the shot pressure
automatically, which solves the former problem of shot-volume variance.
Be sure to read through this instruction manual for correct handling and operation of Accura 9/9L dispenser.
Warning
z Never take the cover off or take the unit apart, or you may get an electric shoc
k.
Caution
z Be sure to read through the catalogue and the instruction manual for safety use.
z Do not exceed the applicable temperature range or you may get an electoric shock.
z Do not exceed the applicable air pressure source range or it may get damaged.
z Do not use the product in the environment under high-temperature, high-humidity, dust
and gas corrosion (chloride, acid and alkali) or you may get an electoric shock.
z Do not convert or remodel the products. We are not responsible for any accidents
and damages caused by it.
z Use the instructed voltage to input or it may get damage.
z When the fuse blows out, never replace it by yourself. Consult with the agent or us
to return for repair.
z Take preventive measure against the serge voltage such as thunderbolt, or it may get
damage due to the abnormal voltage.
z Keep the unit away from vibration or shock from outside.
z Never let the noise in Accura9/9L controller and the wiring cable.
z Ground lead is required, or you may get an electoric shock.
z Never operate the unit according with any other instructions not stated in the
instruction manual. Consult with us if you have any questions.
2 Standard Specifications
Function Explanation
Compensation for shot volume Compensate shot pressure automatically in order to
make constant shot volume by not only AUTO mode but
also STEADY mode
.
Compensation for vacuum Compensate vacuum-level automatically and prevent
bubbling in a barrel and dripping of materials.
Pressure compensation for Stabilize shot-volume for materials, which may change
viscosity its viscosity.
Hardening, softening or mixed with hardening softening
materials are applied.
Empty alarm system Buzzer rings when the material reaches the set level.
Pressure compensation for Line coating can be dispensed evenly and constantly by
cartridge means of pressure compensation when using170cc
cartridge
Communications By RS232 Communication, ACCURA 9 can be
controlled from external host computer.
4 Component Names
②
⑫ ⑬⑭
IEI
ACCURA9 ⑮
CLR RST
⑯
CTRL
POWER INDICATOR
ON H AIR OUT
OFF MODE /OFF REMOTE MENU SHIFT RETURN
① ④ ⑤ ⑥ ⑦⑧⑨⑩ ⑪ ③
② ⑥ ⑦
FUSE GND
SHOT CON. I/O PORT1
AIR IN AC IN
EJECTOR RS232C I/O PORT2
③ ④ ⑤
① ⑧
Components Names
① ... Air Import Port
② ... Exhaust silencer
③ ... Control connector
④ ... Outer input/output connector
⑤ ... Power source connector
⑥ ... Fuse
⑦ ... Ground
⑧ ... RS232C connector
Please take the following procedures for connections of power source and air piping.
S e t-u p P r o c e d u r e
Procedure
1 Connect AC plug with the power source connector on rear
panel.
6 Assemble the needle on the barrel and fill the material into the
barrel up to 70-80%. No bubble should be inside the barrel.
8 Wipe off the material stuck around 1-2cm from the upper side
of the barrel, and then connect the adapter assembly with the
barrel.
9 Now start.
Please take the following procedures when turning on the power source at the initial time to avoid incorrect
operation.
O p e r a tio n
4
Mode Select Screen The title screen is shown in a few seconds and then
** *
** *
**s*
MUe*
OTa*
DOd*
E*y*
S
Ev *
Lir*
Ess*
Cc2*
To3*
*ic*
*t *
*y *
* *
* *
s2*
7 Standard Operation
Key operation
Entered key is written in bold and italic in the sentence. SHIFT+MENU Key describes pressing
SHIFT KEY and MENU KEY simultaneously.
Press MODE KEY to change each screen in the AUTO-STEADY Mode. Each screen changes as
following steps.
MODE Key
Data Edit
MODE Key D.SAMP Function Setting
To select input-menu
If there are multiple input items on the screen, the following procedures should be taken.
1) Go forward the item with MENU Key.
2) Return to the previous item with SHIFT+MENU Key.
When power-switch is on, the mode screen appears after the title screen. In case of power off on the last
time, go to the mode directly as it was.
Functions
Messege Confirmation
A confirmation message appears on the screen when
DATA CLEAR?
enter the key. Press Return key for ‘carry-on’, and press
CLR Reset key for ‘cancel’. YES .. RETURN
NO .. CLR RST
Dispense automatically according to the input shot-data. This Auto mode can be used for shot compensation,
vacuum compensation and empty alarm function. Auto mode consists of three modes such as shot-screen,
data editing screen and D.SAMP function setting screen. There are 3 screens such as D.SAMP, Bottom &
Top level sampling screen and Empty Alarm Setting Screen under the D.SAMP function setting screen.
In the Auto Mode, shot screen is shown at first. In the screen, dispensing carries out according to the input
data on the data-editing screen.
L C D Display
Shot Screen(NoCompensation Shot Screen(Level Compensation)
Functions
Exchange barrels
To exchange barrels, press CLR RESET Key.
Dispense-pressure and vacuum-pressure returns to the initial setting figures. Dispensing counter
will be cleared ‘0’. The lapsed time is reset when viscosity-compensation is working.
Dispensing
One shot dispensing is operated with /OFF key. Level compensation and vacuum compensation works
while Bot-Level and Top-Level are set ‘ON’ on the D.SAMP function setting screen.
Continuous dispensing
Continuous dispensing is operated with SHIFT+ /OFF key. Continuous dispensing keeps on working
until the Sc (dispensing counter) reaches the initial setting at setting Sc (shot counter). Press /OFF key
in order to stop the continuous dispensing. While not setting the dispensing counter, the continuous
dispensing stops by pressing /OFF.
When changing barrels, fine-adjusted dispensing period is reset. If the adjusted dispense period i
sover ±20% or the dispense time is required to adjust soon after changing the barrels, a data
sampling is required by changing the dispense period and pressure.
※When only foot switch is used, it can be shifted to input through LCD panel by
pressing SHIFT+REMOTE key.
※Refer to 10 “Remote Mode” in the instruction manual for details.
Input and edit of channel data are operated on this screen. Input each necessary data while watching the
quantity of dispensing material with /OFF Key.
LCD Display
Data Edit Screen Data Edit Screen
(No-Compensation) (Level-Compensation)
*** Auto Edit *** *** Auto Edit ***
Ch 00 St 00.100 Ch 00 St 00.100
Sp 100.0 Sc 000300 Sp 100.0 Sc 000300
Cm OFF Bs 30 Cm LE +0 Bs 30
Data Edit Screen Data Edit Screen
(Viscosity compensation ) (Intermittent Compensation)
*** Auto Edit *** *** Auto Edit ***
Ch 00 St 00.100 Ch 00 St 00.100
Sp 100.0 Sc 000300 Sp 100.0 Sc 000300
Cm VI 01 Bs 30 Cm IN 500 100 BS 30
Ch ・・・ Input the channel number for data-editing.
Sp ・・・ Input the shot-pressure. Unit: kPa
St ・・・ Input the shot-time. Unit: second
If the level compensation is needed, input 30msec or more.
Sc ・・・ Input the shot-count. The count-up screen will appear when the shot-counter
reaches the setting shot-count. Input 0 (zero) when not use the count-up.
Cm ・・・ Select the shot-compensation.
OFF ・・ No shot-compensation
LE ・・ Level compensation
VI ・・ Viscosity compensation
IN ・・ Intermittent compensation
Bs ・ Input the size of barrel (5, 10, 30, 50 and 170 cc) by cursor keys(↑、↓)
Remark
If the barrel size is input incorrectly, each compensation function and empty alarm may not work
properly.
Channel selection
Enter the channel number into Ch for data editing. The data for a new channel is shown on the screen
when changing data.
To select input-menu
Select the input menu by MENU Key. Press the MENU key to go forward, press SHIFT+MENU key
to go back, and press H+MENU key to return to Ch.
After saving the data, the following screen is shown for confirmation. Press RETURN Key for saving
or press CLR RESET Key to stop Confirmation Screen Data Save Screen
save the data .
SAVE DATA? *** DATA EDIT ***
*
YES ..RETURN * SAVING DATA! *
NO .. CLR RST ********************
Remark
When the shot-pressure, shot-period and/or barrel sizes are changed and saved, the setting sampling data for
Bot-Level, Top-Level and Emp-Level become invalidly as NOD.
Level compensation
Compensate a shot volume corresponding to decreasing material volume in a barrel.
Set Cm on the edit screen LE for data sampling on the D.SAMP setting screen.
Level compensating coefficient is shown beside LE for adjusting the increasing pressure.
The cursor key of (↑、↓)can be used.
At normal time, +0 is shown.
To increase the shot pressure, direct toward +.
To decrease the shot pressure, direct toward -.
In the case of compensating level at AUTO mode, do not compensate the level at
STEADY mode in the same channel. Sampling data changed
Viscosity compensation
In accordance with the predetermined data for viscosity compensation at VISCOSITY mode,
the shot pressure will be changed automatically corresponding to the lapsed time. This will be used
for the material when viscosity varies. Set Cm on the edit screen ‘VI’, and input the number of
viscosity compensation table at VISCOSITY mode.
The lapsed time is shown at the upper-left of dispense screen while compensating viscosity. Press CLR
Key to clear the lapsed time and return shot pressure to initial setting values.
Intermittent compensation
Increase the shot pressure per the predetermined shot times. This can be effective when the level
compensation does not work. For example, when non-standard barrel is used or pipe is too long.
Set Cm on the edit screen ‘In’ and input the value as stated below.
3 digit from the left:Shot interval to increase the pressure.(Shot number)
3 digit from the right: Pressure increased value. ( kPa: 000.0Kpa )
This screen operates to memorize the sampling data for each channel into ROM and to set the empty alarm
function. In case of selecting ‘level compensation’, all functions can be set. In case of selecting other than
‘level compensation’, only the condition for vacuum data at Top-Level can be memorized into ROM.
LCD SCREEN
(※1)
ON : Function set
OFF : Function set off
NOD : No sampling data to be memorized.
(※2)
Except the level compensation, NOD is always shown no matter what the memorized data is.
(※3)
ON : Function set
OFF : Function set off
Remark
When the selected function is NOD, the cursor key (↑↓) will not work to set the function. The sampling
data should be memorized into ROM.
Memorize the sampling data at the level compensation into ROM in order of Bot-Level and
Top-level. The data sampling at Bot_Level and Top_Level enables shot compensation and vacuum
compensation, downloading the sampling data for empty level into ROM and setting empty alarm function.
The sampling data for vacuum condition besides the level compensation should be downloaded
into ROM at Top_Level. The vacuum condition is changeable.
STEADY Mode can be operated in manual to dispense material. Dispense out while the /OFF key is being
pressed.
The mode consists of three screens such as shot-screen, data editing screen, and D.SAMP setting screen.
The screens for sampling at the Bottom & Top levels and for setting of empty-level are located under the
D.SAMP setting screen.
In the STEADY Mode, the shot screen will appear at first. Dispense according to the set data in the data
editing acreen.
L C D Screen
Shot Screen Shot Screen
(No Compensation) (LevelCompensation)
*** Steady Shot *** ***
***Steady
SteadyShot***
Shot
Ch 00 St 00.000 Ch 00 St 00.000
Sp 100.0 Sc000300 Sp 100.0 Sc000300
Cm OFF Cm
Cm LE
LE VC VC EMP DE
Shot Screen
(Cartridge Compensation)
*** dy ShSteaot ***
Ch 00 St 00.000
Sp 100.0 Sc000300
Cm CA
Level compensation
VC ・・・Vacuum compensation is working.
EMP ・・・Empty alarm is working.
DE ・・・Empty alarm function is detected while selecting “continue to shot.”
(if ‘continue to dispense’ is selected.)
FUNCTIONS
Exchanging barrels
Press CLR RESET key when changed barrels.
The shot pressure and vacuum pressure return to the initial setting values. The shot counter is reset to ‘0’.
Dispensing
The shot keeps going only while pressing the /OFF key. When release /OFF key, the actual shot
period is shown at St.
The vacuum compensation and the level compensation work when Bot_Level and Bot_Level are set
‘ON’ in the D.SAMP screen.
Remote Mode
Enter into Remote Mode with Remote key. Remote Mode Screen
Dispensing is operated by signal from external input in the
remote mode. *** Steady Shot EXT
Ch 00 St 00.000
In order to meet the individual user’s requirements, the Sp 100.0 Sc 000300
foot switch (FSW), EXIT I/O(EXT)and R Cm LE VC
S232C(RS2)can be used for dispensing.
Each time when press the REMOTE key, FSW,EXT and RS2 will appear.
※When using only foot switch, it is switchable from LCD panel by pressing SHIFT
+REMOTE key
※Refer to Art. 10. Remote Mode for the detail of remote mode.
This screen is for editing of the channel data. Set each input data while watching the shot quantity with
SHOT key.
L C D SCREEN
Mt ・・・ When compensate the level, enter the minimum shot period for sampling.
Shown on the shot compensation in LE.
Enter more than 30msec as minimum.
Bs ・・・ Set the size of barrel by cursor key(↑、↓).
Range of barrels : 5cc, 10cc, 30cc, 50cc and 170cc.
Remark
: Each compensate function and empty alarm function may not be carried
out unless the size of barrel is set correctly.
Channel selection
Enter the channel number to shot on Ch.
Remark
If the saving data is carried out after having changed the shot pressure, the minimum shot period and the
size of barrel, the sampling data of Bot_Level & Top_Level and the setting of empty detection on Emp_Level
become invalid (NOD).
Functions
Level compensation
Compensate decline of shot volume caused by decreasing the material volume in barrel.
Cm at edit screen should be set LE, which enables data sampling at D.SAMP setting screen. The level
compensation coefficient beside LE will change the pressure inclease.
Operate with cursor key (↓↑).
At normal, set at +0.
To increase the shot pressure, operate toward +.
To decrease the shot pressure, operate toward –-.
When Cm is changed to LE, Mt is shown. Enter the minimum shot period at STEADY mode.
The compensation sampling is carried out in accordance with this set time. Therefore, set as long period as p
ossible for more accurate compensation.
Remark: While level compensating at STEADY mode, do not operate level compensation at AUTO mode in
the same channel.
Enter D.SAMP at the STEADY mode as the same key operation as at AUTO mode.
In this screen, set data sampling, vacuum compensation and empty detection of each channel
Refer to the D.SAMP screen for further detail in AUTO mode.
LCD SCREEN
FUNCTIONS
Enter the information of viscosity compensation into Tbl_No. New table number appears by changing
the Tbl_No.
Select the entering item with MENU key. Press the MENU key to go forward, and press
SHIFT+MENU Key to go back. Press H+MENU key to return to Tbl_No.
Enter the lapsed time (min.), the pressure rate of change (%), the hardening (UP), and softening
(DOWN) viscosity per data number. Enter the data number from the number 01 in order.
Enter the lapsed time by minute since changing barrels. Enter the lapsed time to get longer. Enter the
rate of pressure changes (%) against the set pressure.
Select changing viscosity, Up & Down, with cursor key.(↑、↓)
Saving data
Save the data by RETURN key. Confirmation screen is shown when completed saving the data.
RETURN key to save and CLR RESET key to stop.
LCD Screen
RS232C Setting Screen
BAUDRATE 2400
PARITY NONE
BIT LENGTH 8
STOP BIT 2
BAUDRATE・・・ Communication speed. (1200, 2400, 4800, 9600) The initial setting: 2400.
PARITY ・・ Parity check (NONE, ODD, EVEN). The initial setting: NONE.
BIT LENGTH・・ Data length (8-bits fixed.)
FUNCTIONS
Operation
Select the entering item with Menu Keyt. Press the Menu Key to advance one step, and press the
SHIFT+Menue Key to return to one step. Press H+Menue Key to return to Baudrate.
1.Shot count-up
This screen appears when shot-count reaches the initial setting shot-count.
2.Empty Alarm
This screen appears when the material quantity in the barrel or cartridge reaches the initial setting
minimum empty level.
This screen appears when there is difference between values of checksum transmitted in
RS232C and of checksum caliculated inside of dispenser
Check Sum Error Screen
*** RS232C
*
* CHECK SUM ERROR *
********************
This screen appears when Parity, Over-run or Framing error occurs during communication..
Status Error Screen
*** RS232C
*
* STATUS ERROR *
********************
This screen appears when not receive an acknowledge signal or a denial signal within 10sec.
after transmitting requirement signal.
Time Over Error Screen
*** RS232C
*
* TIME OVER ERROR *
********************
How to release:
The error with buzzer sound is cleared after releasing the error.
8 Data Sampling
**Sampling Mode**
Press CLR RESET key to stop.
*DATA SAMPLING!! *
********************
11
D.Samp Function Setting Screen
In order to use the ‘Empty Alarm’ function, set
*** D.Samp Menu *** Emp_Level ON on the ‘D.SAMP’ screen.
Ch 00 Bot_Level ON If setting OFF, the Empty Alarm function doesn‘t
Top_Level ON work.
Emp_Level ON
L C D screen PROCEDURES
*** Steady Shot *** When the remote mode is EXIT I/O (EXT), it is
Ch 00 St 00.000 possible to change the barrels by the signal to clear the
Sp 100.0 Sc 000300 setting signal from outside.
Cm LE VC
When the remote mode is RS232C (RS2), it is
possible to change barrels by the command to clear
the setting data.
When the above procedure shall not be taken despite changing barrels, each compensation function and
empty alarm function might not work correctly.
10 Remote Mode
※Refer to 11. Input & Output Signal Interface about Input & Output signals.
Operation procedure
2
Auto Shot Remote Screen After entering the REMOTE key, the condition of
remote mode will appear on upper right corner. The
*** Auto Shot EXT controller can be controlled by means of the Foot
Ch 00 St 00.100 switch (FSW), EXIT I/O(EXT) and RS232C
Sp 100.0 Sc 000300 (RS2).
Cm LE VC
***** on the upper right screen means a normal
condition.
Pressing REMOTE key every one time, "***"、
"FSW"、"EXT" and "RS2" is shown in order one
by one.(The left screen is on the EXIT I/O.)
The shot control procedure and communication example at RS232C are as follows:
The sender stops the transmission because of timeout when the sender does not receive the
acknowledgment signal or Non-acknowledgment signal within 10 seconds after transmitting the
transmission request signal. The receiver stops the transmission by sending Non-acknowledgment
signal because of timeout when the receiver does not receive the all data within 10 seconds after
receiving the signal to receive.
SIGNALS
Data Format
・・・
Every data is stored the first byte as the high end. Omit a decimal point.
List of Received-command description
① Clear count: .. Clear shot count (current count).
Kind Item Byte No. Figure
1 Data length 1 0Ch
2 Command 2 C
③ Request the shot data ..The shot data on the received channel data transmit.
④ Shot command .. Shot in the received channel mode and transmit the shot
end signal after shot is finished.
⑤ Shot end .. Complete to shot at the STEADY mode. Transmit the shot end
after shot is finished.
➉ Setting clear...Clear the shot count (present figure). Return the shot pressure and vacuum
pressure to setting figures. Return the compensating condition to setting figures.
*When data is stored, sampling is ON automatically at upper limit or lower limit of barrel.
Therefore set shot data and master data as sampling is carried out before storage of sampling
data.
*Take a note of shot data and master data, which are data while sampling is carrying out.
They are necessary when transmitting.
TRANSMISSION DATA
① Shot data ... The below data are transmitted responding to the request for the
shot data.
④ Master data...Responding to the request of master data, the following data are
transmitted.
Kind Item Byte Figure
1 Data length 1 28
h
2 Command 2 MD
3 Channel 2
Barrel capacity 1
Mode 1
(Auto mode) lower limit shot 4
pressure
(Steady mode) lower limit shot 4
pressure
Vacuum pressure at upper limit 4
Vacuum pressure at lower limit 4
Vacuum operating time 4
Vacuum operating off time 4
2 Command 2 “SD”
3 Channel 2
Mode 1
Data1 6
Data2 6
Data3 6
Data4 6
Data5 6
Data6 6
Data7 6
Data8 6
ACCURA9 HOST
Error message is shown on the screen.
However, above-mentioned error is rarely occurred if the communication setting and operation between
the Host computer and ACCURA 9 are properly done. When the error is occurred, it may be
impossible to transmit the error message command. In this case, please perform the following
procedure to release the failure condition of RS232C communication.
n Press the CLR RESET key and release the failure condition. (ACCURA 9 does not transmit the
error content (released ) command.)
o Check and fix the defects in the communication setting of the Host computer and ACCURA 9.
p Resume the communication from the Host computer.
<Pattern 2>
ACCURA 9/9L proceeds the job corresponding to the command from Host computer after
receiving it. No actual data should be transmitted to Host computer.
Host ACCURA9
Intsructed data
ENQ Data stored
EOT
Shot command(steady)
Instructed
data ACK Clear numbers
Executed Reset
Clear settomg
Stored master data
Change channels
Change remotes
Stored sampling data
Host ACCURA9
Error occurs
ENQ
EOT
Error detail
ACK (n byte)
ENQ
EOT
Alarm clear ACK
(n byte)
Release error
ENQ
EOT
Errror detail
Shot count up
Empty ditector
<Pattern 4>
ACCURA 9/9L will transmit a message of ‘NAK’ to Host computer when there is something
abnormal in the data transmitted from Host computer, since then will transmit the error status
to Host computer.
H ost ACCURA9
ENQ E rror occu rs
EOT Error detail
Command data
(n byte)
“ Check sum error
ACK
NAK
ENQ
Host make retry
HOST ACCURA9
ENQ
EOT
Command data
(n b yte ) All bytes of data can not be transmitted
Error within 10
occurs seconds after receiving ENQ and seneling EOT.
NAK
<Pattern 6>
Countermeasure for transmission time over of ACCURA9/9L
HOST ACCURA9
ENQ
EOT EOT can not be receired within 10 seconds
Error occurs after sending ENO ,or no-acknowledgment
for data transmission when receivingEOT.
NAK
Re-execution by
Host computer
ENQ
<Pattern 7>
Host computer should shift to Receiving if Host computer and ACCURA 9/9L start
transmitting (ENQ communication) simultaneously.
Data only for the error status should be transmitted by ACCURA 9/9L in one sided.
HOST ACCURA9
ENQ ENQ
EOT
Shift toReceiving
by Host computer Data
(n byte)
ACK
PC9801 ACCURA9
1. FG 1. FG
2.TXD 2.RXD
3.RXD 3.CTS
4.RTS 4.GND
5.CTS 5.N.C.
6.DSR 6.TXD
7.GND 7.RTS
8.DCD 8.DSR
20.DTR 9.DTR
The following example is for connection between IBM-PC AT Converter and ACCURA 9/9L for the
reference.
It is available to control Accura 9/9L dispensing by input/output signal from I/O Port.
S8 S4 S2 S1 Channel number
0 0 0 0 0
0 0 0 1 1
0 0 1 0 2
0 0 1 1 3
… … … … …
1 0 0 0 8
1 0 0 1 9
Remark
If the input route isn‘t connected to Channel Selection Signal, the channel number for dispense will be
shown ’00’.
Remark :
If the input route isn‘t connected to Channel Selection Signal, Auto Mode is shown.
Steady-Mode:
SHOT CONTROLL ACTION
START
8PIN
0 Stop dispensing.
This signal can be input in case of the busy signal 1.
1 Dispensing.
Input this signal to dispense material in linear.
Keep dispensing until inputting ‘0’.
During dispensing (*1), the busy signal shows ‘1’.
(※1) Dispensing time + Vacuum time
1 0 0 Reset alarm.
Input this signal to reset when alarm/error signal shows
a number ‘1’. Empty alarm is also reset.
Error screen and/or buzzer sound is reset.
0 0 1 Reset settings.
Reset compensating condition to an initial setting.
Reset dispensing pressure and vacuum pressure to the
initial settings.
Reset a number of dispensing
Reset the lapsed time during viscosity compensation in
use..
During alarm/error, this signal is ignored.
0 1 0 Reset.
After input this signal, the software of controller is reset
and restarts from the initial title screen.
During ‘Busy’, this signal is ignored.
Remark :
If the input cable is not connected in the control signal, the controller is regarded as no signal.
(※1) During dispensing (incl. vacuum time and shot time) and changing channel.
Remark :
There might is time lag in a few metric-seconds to turn ‘1’ on the busy signal after
being accepted the input signal by the dispenser controller.
Pay attention when checking BUSY signal.
34 GND GND
35 GND GND
36 GND GND
15 GND GND
16 GND GND
17 GND GND
34 GND GND
35 GND GND
36 GND GND
Remark
:
1. Do not connect all other ones except the above pin numbers.
2. Use cable with shielding to connect with external system and gound shielding beside the external
system.
3. Do not tie signal cable with power line such as AC100V and keep a distance between them.
4. Never connect with 24V outer power input when except at SW3: IN . It may cause malfunction.
+5V(INSULATION)
USER SOURCE
+8VDC∼+40VDC
ISO +5V +COM
REGULATOR
+5V(ACCURA 9) SW1 1,2,20,21
EXT
GND
15,16,17,
31,32,33
GND
(INSULATION)
IN PUT
GND (ACCURA 9)
+5V(ACCURA 9)
IN PUT
SHIELD
GND (ACCURA 9)
F.G
ACCURA 9
SW3
OUT
+5V(ACCURA 9)
IN
LOAD
LOAD
GND (ACCURA
GND (INSULATION)
OUT PUT
MAX 50mA
+5V(ACCURA 9)
OUT PUT
MAX 50mA
GND
15,16,17
SHIELD
F.G
ISO
EXT
IN
OUT
SW 1
SW 3
There are internal power source/external power source setting switches SW1, SW3, which are placed on
the second PCB of control PCB unit as above. The cobination of setting is as follows.
Setting of SW1,SW3
SW SW Explanations
1 3
Supplying power to port 1 of ACCURA9 is not required. External input
ISO IN port is required power because power is not done output from port 2 to
outside.
Supplying power to port 1 of ACCURA9 is not required. External input
OU port is not required because power is done output from port 2 to outside.
ISO
T In this case, insulate it with power of external auto system. Otherwise, it
may cause malfunction.(factory default setting)
EX IN Supplying power to port 1 of ACCURA9 is required. Additionally, it is
T available to turn on power from port 2.
EX OU Unusable
T T
+5V(ACCURA 9)
LOAD
1KΩ OUT PUT
GND
3
MAX 50mA
+5V(INSULATION)
+5V(ACCURA 9)
GND
(INSULATION)
INPUT
2
3mA
GND (ACCURA 9)
-COM
1
SHIELD
GND
(INSULATION)
F.G
SHOT START
More than 10ms
BUSY
SHOT END
SHOT OFF
Counter
Error output
Alarm clear
Remote change
Reset input
POWER ON
S1
Channel input
S2 DON'T CARE DON'T CARE
1 AUTO MODE
10msec or more
ON
Shot start
OFF
Shot
Shot motion
Stop
ON
Shot complete
OFF
2 Steady Mode
ON
Shot start
OFF
Shot
Shot motion
Stop
ON
Shot complete
OFF
RAM for ACCURA9 is backed up by lithium battery so that the data may not be erased if the
Power supply is turned off.
The life of battery is about 1.5 years (conditions on 16 hours of power off a day)
Memorized setting data and important data for the system may disappear completely if detaching the
backup battery while turning ACCURA 9 off for a long time (5∼6 hours). Replace batteries as the
following procedures.
In the event of data loss or request for us for barreries replacement,
please contact us .
Battery for use:CR2032 (Panasonic)
(1) Power off Accura9
(2) Take off the chassis cover of Accura 9
Battery holder
Front panel
Top face
Battery holder
Batterryy
Remark
Complete the procedures (1)∼(3) within 2-3 hours.
Maintenance
No special maintenance is required as far as the controller is operated under normal condition.
Consumable
・Do not use a cracked or flawed barrel. The barrel is disposable.
・When the O-ring of the adapter assembly is corroded by the solvent or when it hardens, replace
it with a new one.
・The adapter assembly, barrel, needle and O-ring are consumable. Specify the size and model
number when ordering.
Miscellaneous
When turning off controller, take off asapter assembly, which is connected to air outport or decrease air
source pressure .
Front view
Rear view