EPS 45-80 High-Voltage Generator

Technical Manual Supplement

2-TUBE OPTION

MANUAL P/N 11050.001.G3


2-Tube Option EPS 45-80 High-Voltage Generator

Copyright application © 2009, 2010 by emd technologies Company. All rights reserved. emd
technologies and the EMD Technologies logo are trademarks of emd technologies Company,
registered in the U.S.A. and other countries.

Contents of this publication may not be reproduced in any form or by any means, electronic or
mechanical, including photocopying and recording, or by any information storage or retrieval system
without the written permission of emd technologies Company., 400 du Parc, Saint-Eustache
(Québec), Canada, J7R 0A1.

IMPROVEMENTS
emd technologies reserves the right to make changes and improvements to its products and assumes
no obligation to incorporate any changes into units previously sold or shipped.

2-TUBE OPTION / Technical Manual Supplement P/N 11050.001.G2


Manual Revision History
N° Revision Number Date Note
1 P/N 11050.001.G1 23 May 06 Initial release
2 P/N 11050.001.G2 04 Feb. 09 H.V. cable length correction
3 P/N 11050.001.G3 07 June 10 Page 7: Warning
Page 10: Figure 7 for PC Tools 1.34

Head Office European Authorized Representative

emd technologies Biocordis France S.A.A


400 du Parc, 3, Rue du Bois Chaland
Saint-Eustache (Québec) F-91090 LISSES
Canada, J7R 0A1 France
Phone : 450-491-2100
Fax : 450-491-4138
www.emd-technologies.com

Document 11050.001.G3 2/10


2-Tube Option EPS 45-80 High-Voltage Generator

2-TUBE OPTION P/N 10020.000

TABLE OF CONTENTS

1 INTRODUCTION................................................................................................................................................................ 3
1.1 Description .......................................................................................................................................................................... 3
1.2 List of Material .................................................................................................................................................................... 4

2 INSTALLATION .................................................................................................................................................................. 5
2.1 2-Tube Option Interface Board Installation ................................................................................................................... 5
2.2 High-Voltage Cables Connection ..................................................................................................................................... 6
2.3 Low-Voltage Interconnection........................................................................................................................................... 7
2.4 Tube Stator Connection..................................................................................................................................................... 8
2.5 Tube Housing Connection ................................................................................................................................................ 9

3 SYSTEM CONFIGURATION ......................................................................................................................................... 10


3.1 2-Tube Option Interface Board...................................................................................................................................... 10
3.2 Tube Setup......................................................................................................................................................................... 10
3.3 Tube Assignment.............................................................................................................................................................. 10

4 TUBE CALIBRATION ...................................................................................................................................................... 10

5 DISPOSAL OF THE PRODUCT .................................................................................................................................... 10

1 INTRODUCTION

1.1 Description
The 2-Tube option allows interfacing a second X-ray tube to the EPS 45-80 High-Voltage
Generator. It contains a standalone tube switch assembly and the necessary hardware to operate the
two X-ray tubes in any RAD workstation.

If the EPS 45-80 High-Voltage Generator is fitted with the Epsilon Operator Console, this latter
will control the selection and the display of both X-ray tubes parameters. Fluoro is not allowed on
the second X-ray tube. RAD exposures cannot be done on both X-ray tubes simultaneously.

Document 11050.001.G3 3/10


2-Tube Option EPS 45-80 High-Voltage Generator

1.2 List of Material


The 2-Tube option is made up of a standalone 2-Tube option Assembly (Figure 1), a tube switch
interface board and the interconnection cables (Table 1).

Table 1-
Parts Shipped with 2-Tube Option for EPS 45-80 Generator
Qty Name Part Number
1 2-Tube Option Assembly 11024.000
1 2-Tube Option Interface Board 9996.00
1 2-Tube Option Manual 11050.001
1 Patch Cable RJ45 (8 feet) ----
2 High-Voltage Cables (6 feet) ----

Dimensions in inches [mm] Weight: 58 lb (26.3 kg)

Figure 1 – 2-Tube Option Assembly (dimensions).

Document 11050.001.G3 4/10


2-Tube Option EPS 45-80 High-Voltage Generator

2 INSTALLATION

All parts of the option are shipped in a single heavy-duty cardboard box. Open the box and inspect
the parts (Figure 2) for any mechanical damage that could impair the operation of the system.

Do not lift the Two-Tube Option Assembly by the motor box

½ 2-Tube Option Interface


Board P/N 9996.00

2-Tube Option Assembly ¾


P/N 11024.000,
With ground cable (18 in) ¾

½ High-Voltage Cables (2 x 6 feet)

Not shown ½ Patch Cable (8 feet)

Figure 2 – Parts shipped with the 2-Tube Option P/N 10020.000.

2.1 2-Tube Option Interface Board Installation


The 2-Tube Option Interface Board P/N 9996.00 must be installed in one of the four positions
available on the Expansion Motherboard (J4 through J1 connectors, Figure 3). Use the first free
position available beginning from J4 (right-most) connector.
In Figure 3, for example, the board should be installed on J2-position of Expansion Motherboard.

Document 11050.001.G3 5/10


2-Tube Option EPS 45-80 High-Voltage Generator

Figure 3 – Connect the 2-Tube Option Interface Board P/N 9996.00 on the first available
free connector on the Expansion Motherboard, beginning from the right-most position (J4).

Note that the position of the boards connected to the Expansion Motherboard must be specified in
the generator configuration. Refer to Section 3.1.

If the 2-Tube Option Interface Board P/N 9996.00 has to be installed in


J1 position (left-most connector), the side panel of the board must be
taken off before connecting it to the Expansion Motherboard. To do so,
loosen the two nuts of the 2-Tube Option Interface Board (Figure 4) and
pull out the side panel.

Figure 4 – Side panel removal of the 2-Tube Option Interface board.


Do this step only when this board is the fourth one to be installed on
the Expansion Motherboard. ¾

2.2 High-Voltage Cables Connection


1. Place the Two-Tube Assembly at less than six feet of the EPS 45-80 H.-V. Generator.

CAUTION– RISK OF ELECTRIC SHOCK

2. Make sure the mains line is turned off at the room circuit breaker.

3. Label one high-voltage cable supplied with the option as "Anode" and the other one as
"Cathode".

4. Take off the protective covers of the high-voltage output on both receptacles of the generator.
Do not discard them as they can be used later for shipping.

Document 11050.001.G3 6/10


2-Tube Option EPS 45-80 High-Voltage Generator

5. Ensure proper isolation of the receptacles by coating them with insulating grease. Isolation is
necessary to prevent electric arcs between the high-voltage receptacles and the tube switch
assembly.

6. Connect the anode high-voltage cable (6-foot) between the generator’s anode receptacle
and the anode high-voltage receptacle on the tube switch assembly (identified “A”). Make
sure the contact pins are engaged in the holes of the receptacles. Tighten the cable nut
while holding the cable firmly in place. Use an appropriate Allen ranch to tighten the
cable ring nut. If necessary, consult the installation instructions provided by the H.V.
cable manufacturer
Repeat last step to install the cathode high-voltage cable (6-foot) on the generator’s
cathode H.V. receptacle and the cathode H.V. receptacle (identified “K”) on the 2-tube
option assembly.

7. Repeat steps 5 and 6 to connect the high-voltage cables (not supplied with the
option) between the high-voltage receptacles “A1” and “K1” on the tube switch
assembly and the X-ray tube number 1 (figure 5).

• Use the same type and the same length of High-Voltage cables for both
anode and cathode sides.

• To make this connection, follow the installation procedures provided by the X-


ray tube manufacturer. Make sure to respect the maximum cable length between
the tube switch assembly and each X-ray tube.

8. Repeat step 7 to connect the high-voltage cables (not supplied with the option)
between the high-voltage receptacles “A2” and “K2” on the tube switch assembly
and the X-ray tube number 2 (figure 5).

Make sure that both cathode and anode high-voltage cables are installed between
the generator and the X-ray tube before making any exposures (4 cables).

2.3 Low-Voltage Interconnection


1. Connect the 2-Tube Option assembly ground cable (attached to the
ground of the 2-Tube Option Assembly) to the screw on the generator
chassis shown in Figure 5.

2. Install the network cable between the Tube Switch Interface Board
and RJ45 connector on the tube switch assembly (Figure 5).

Document 11050.001.G3 7/10


2-Tube Option EPS 45-80 High-Voltage Generator

Maximum length of Federal Cables: Cable lengths shown not to scale


“K” = “A” = 6 foot (cables supplied with 2-Tube Option)
“a1” = “k1” < Max. 94 foot and “a2” = “k2” < Max. 94 foot (cables supplied by customer)

Figure 5 – Two-Tube Option Connection.

2.4 Tube Stator Connection


Refer to your Technical Manual (Chapter 2) to connect the Tube stator to the Line Filter board. The
connection points are shown below in Table 2 and Figure 6.

Table 2- Tube stator connection to Line Filter board


Cable Signal Connector
J6 pin 1 – Main Tube 1
Black Stator Main
J6 pin 4 – Main Tube 2
J6 pin 2 – Common Tube 1
White Stator Common
J6 pin 5 – Common Tube 2
J6 pin 3 – Phase Tube 1
Green (Red) Stator Phase
J6 pin 6 – Phase Tube 2
Bare lug Stator Cable Shield Shield (on Line Filter Board Holder, Figure 6)

Document 11050.001.G3 8/10


2-Tube Option EPS 45-80 High-Voltage Generator

Figure 6 – Tube Stator and Tube Housing Connection


on the generator Line Filter Board.

2.5 Tube Housing Connection


Refer to your Technical Manual (Chapter 2) to connect the Tube housing (Thermal switch) to the
Line Filter board. The connection points are shown below in Table 3 and on Figure 6.

Table 3- Tube housing connection to Line Filter board


Signal Connector
Tube Housing J8 pin 1 – Tube 1 ; J8 pin 4 – Tube 2
Tube Housing Return J8 pin 2 – Return 1 ; J8 pin 3 – Return 2
Tube Housing Shield Shield (on Line Filter Board Holder, Figure 6)

If there is no tube housing safety on tube 1 or on tube 2, install a jumper in place of


the corresponding interlock switch on J8 connector.

Document 11050.001.G3 9/10


2-Tube Option EPS 45-80 High-Voltage Generator

3 SYSTEM CONFIGURATION

Refer to Chapter 4 of your Technical Manual to setup the following parameters.

3.1 2-Tube Option Interface Board


Make sure to select the position (Slot) in which the 2-Tube Option Interface Board is installed on
the Expansion Motherboard (Epsilon PC Tools, Options configuration, Figure 7).

Figure 7 – Options configuration screen (Epsilon PC Tools).

3.2 Tube Setup


Refer to Section 4.4 of your Technical Manual and to Epsilon PC Tools to setup both X-ray
tubes using the proper tube chart file of the Tube Package library. Usually, the tube library is
provided with your generator on a CD.

3.3 Tube Assignment


Refer to Section 4.5 of your Technical Manual and to Epsilon PC Tools to assign the proper X-
ray tube to every workstation.

4 TUBE CALIBRATION
Refer to Chapter 5 of your Technical Manual to perform the tube calibration on both X-ray tubes.

5 DISPOSAL OF THE PRODUCT


At the end of the 2-Tube Option Box Assembly usable life, proper environmental considerations
shall be taken to dispose of its dielectric oil (15 liters) according to local laws.
END OF DOCUMENT