Professional Documents
Culture Documents
2-TUBE OPTION
Copyright application © 2009, 2010 by emd technologies Company. All rights reserved. emd
technologies and the EMD Technologies logo are trademarks of emd technologies Company,
registered in the U.S.A. and other countries.
Contents of this publication may not be reproduced in any form or by any means, electronic or
mechanical, including photocopying and recording, or by any information storage or retrieval system
without the written permission of emd technologies Company., 400 du Parc, Saint-Eustache
(Québec), Canada, J7R 0A1.
IMPROVEMENTS
emd technologies reserves the right to make changes and improvements to its products and assumes
no obligation to incorporate any changes into units previously sold or shipped.
TABLE OF CONTENTS
1 INTRODUCTION................................................................................................................................................................ 3
1.1 Description .......................................................................................................................................................................... 3
1.2 List of Material .................................................................................................................................................................... 4
2 INSTALLATION .................................................................................................................................................................. 5
2.1 2-Tube Option Interface Board Installation ................................................................................................................... 5
2.2 High-Voltage Cables Connection ..................................................................................................................................... 6
2.3 Low-Voltage Interconnection........................................................................................................................................... 7
2.4 Tube Stator Connection..................................................................................................................................................... 8
2.5 Tube Housing Connection ................................................................................................................................................ 9
1 INTRODUCTION
1.1 Description
The 2-Tube option allows interfacing a second X-ray tube to the EPS 45-80 High-Voltage
Generator. It contains a standalone tube switch assembly and the necessary hardware to operate the
two X-ray tubes in any RAD workstation.
If the EPS 45-80 High-Voltage Generator is fitted with the Epsilon Operator Console, this latter
will control the selection and the display of both X-ray tubes parameters. Fluoro is not allowed on
the second X-ray tube. RAD exposures cannot be done on both X-ray tubes simultaneously.
Table 1-
Parts Shipped with 2-Tube Option for EPS 45-80 Generator
Qty Name Part Number
1 2-Tube Option Assembly 11024.000
1 2-Tube Option Interface Board 9996.00
1 2-Tube Option Manual 11050.001
1 Patch Cable RJ45 (8 feet) ----
2 High-Voltage Cables (6 feet) ----
2 INSTALLATION
All parts of the option are shipped in a single heavy-duty cardboard box. Open the box and inspect
the parts (Figure 2) for any mechanical damage that could impair the operation of the system.
Figure 3 – Connect the 2-Tube Option Interface Board P/N 9996.00 on the first available
free connector on the Expansion Motherboard, beginning from the right-most position (J4).
Note that the position of the boards connected to the Expansion Motherboard must be specified in
the generator configuration. Refer to Section 3.1.
2. Make sure the mains line is turned off at the room circuit breaker.
3. Label one high-voltage cable supplied with the option as "Anode" and the other one as
"Cathode".
4. Take off the protective covers of the high-voltage output on both receptacles of the generator.
Do not discard them as they can be used later for shipping.
5. Ensure proper isolation of the receptacles by coating them with insulating grease. Isolation is
necessary to prevent electric arcs between the high-voltage receptacles and the tube switch
assembly.
6. Connect the anode high-voltage cable (6-foot) between the generator’s anode receptacle
and the anode high-voltage receptacle on the tube switch assembly (identified “A”). Make
sure the contact pins are engaged in the holes of the receptacles. Tighten the cable nut
while holding the cable firmly in place. Use an appropriate Allen ranch to tighten the
cable ring nut. If necessary, consult the installation instructions provided by the H.V.
cable manufacturer
Repeat last step to install the cathode high-voltage cable (6-foot) on the generator’s
cathode H.V. receptacle and the cathode H.V. receptacle (identified “K”) on the 2-tube
option assembly.
7. Repeat steps 5 and 6 to connect the high-voltage cables (not supplied with the
option) between the high-voltage receptacles “A1” and “K1” on the tube switch
assembly and the X-ray tube number 1 (figure 5).
• Use the same type and the same length of High-Voltage cables for both
anode and cathode sides.
8. Repeat step 7 to connect the high-voltage cables (not supplied with the option)
between the high-voltage receptacles “A2” and “K2” on the tube switch assembly
and the X-ray tube number 2 (figure 5).
Make sure that both cathode and anode high-voltage cables are installed between
the generator and the X-ray tube before making any exposures (4 cables).
2. Install the network cable between the Tube Switch Interface Board
and RJ45 connector on the tube switch assembly (Figure 5).
3 SYSTEM CONFIGURATION
4 TUBE CALIBRATION