IMPROVEMENTS
emd technologies reserves the right to make changes and improvements to its products and assumes
no obligation to incorporate any changes into units previously sold or shipped.
TABLE OF CONTENTS
1 INTRODUCTION .................................................................................................................................. 3
1.1 Description................................................................................................................................................ 3
1.2 Generator Configuration .......................................................................................................................... 4
2 INSTALLATION..................................................................................................................................... 5
2.1 Power Line Connection ............................................................................................................................ 5
2.2 K2 Relay and Auxiliary Output Supplies Connection .............................................................................. 5
3 ELECTRICAL SPECIFICATIONS ........................................................................................................ 6
3.1 Input Specifications .................................................................................................................................. 6
3.2 Output Specifications................................................................................................................................ 6
3.3 Electric Schematics ................................................................................................................................... 6
1 INTRODUCTION
1.1 Description
The Auxiliary Power Supplies provides a medical grade power supply rated 24 VAC, 150 VA
(2.7A) and an optional 24 VDC, 65 W.
This option is used with an EPS generator and requires that the generator be supplied by a
400Y/230VAC electrical line (230 VAC Phase-to-Neutral). The electrical line must therefore
provide a neutral wire to be connected to J2 (Neutral) lug on the Line Filter Board P/N 9916.00.
Prerequisite for this option is the use of a 400Y/230 VAC three-phase line for
electrical supply of the EPS generator. The neutral must be connected on the
neutral lug of the Line Filter board; see Technical Manual for connection.
Connector
Terminal Strip
Medical Grade
Power Supply Power Transformer
Line-Access Panel
(Inner side)
To J7-4
To J7-3
Figure 1 – Auxiliary Power Supplies Option P/N 10 023.000 (24 VDC/24 VAC), installed on the inner
side of the Line-Access Panel. Note that the option version P/N 10 023.001 does not include
the Power Transformer and the 24 VAC output.
2 INSTALLATION
1. Refer to your Technical Manual for the complete power line connection instructions.
2. Make sure the mains line is turned off at the main circuit breaker.
3. When removing the Line-Access Panel at the back of the EPS High-Voltage Generator, it is
recommended to unplug the Auxiliary Output wires from J7 of Line Filter Board P/N 9916.00
(Figure 2). If you do not disconnect
them, be aware that all of the weight of
the panel does not apply directly on the
connected wires.
Figure 2 –
EPS 45-80 High-Voltage generator
with Auxiliary Power Supplies Option.
2. If used in your installation, connect K2 programmable relay,. K2’s contacts are programmed
using Epsilon PC Tools; see Technical Manual, Chapter 4.
4. Reinstall the Auxiliary Output wires on J7 pins 3 (H) & 4 (N) of Line Filter Board.
6. Firmly attach the 24 VDC (/VAC) output wires using the cable holder.
3 Electrical Specifications