EPS 45-80 High-Voltage Generator

Technical Manual Supplement

AUXILIARY POWER SUPPLIES OPTIONS


P/N 10023.000 & 10023.001

MANUAL P/N 11050.000.G2


Auxiliary Power Supplies Option EPS 45-80 High-Voltage Generator

Copyright application © 2006-2010 by emd technologies. All rights reserved. emd technologies
and the EMD Technologies logo are trademarks of emd technologies, registered in the U.S.A.
and other countries.

Contents of this publication may not be reproduced in any form or by any means, electronic or
mechanical, including photocopying and recording, or by any information storage or retrieval system
without the written permission of emd technologies, 400 rue du Parc., Saint-Eustache (Québec),
Canada, J7R 0A1.

IMPROVEMENTS
emd technologies reserves the right to make changes and improvements to its products and assumes
no obligation to incorporate any changes into units previously sold or shipped.

AUXILIARY POWER SUPPLY OPTIONS / Technical Manual Supplement P/N 11050.000.G2


Manual Revision History
N° Revision Number Date Note
1 P/N 11050.000.G0 03 April. 2006 Initial Release
2 P/N 11050.000.G1 29 Aug. 2008 F3 & F4 fuses new specifications
2 P/N 11050.000.G2 07 June 2010 Update

Head Office European Sales Representative


emd technologies Biocordis France S.A.A
400 du Parc, 3, Rue du Bois Chaland
Saint-Eustache (Quebec) F-91090 LISSES
Canada, J7R 0A1 France
Phone : +1-450.491.2100
Fax : +1-450.491.4138
www.emd-technologies.com

Document P/N 11050.000.G2 2/6


Auxiliary Power Supplies Option EPS 45-80 High-Voltage Generator

AUXILIARY POWER SUPPLIES OPTION

TABLE OF CONTENTS
1 INTRODUCTION .................................................................................................................................. 3
1.1 Description................................................................................................................................................ 3
1.2 Generator Configuration .......................................................................................................................... 4
2 INSTALLATION..................................................................................................................................... 5
2.1 Power Line Connection ............................................................................................................................ 5
2.2 K2 Relay and Auxiliary Output Supplies Connection .............................................................................. 5
3 ELECTRICAL SPECIFICATIONS ........................................................................................................ 6
3.1 Input Specifications .................................................................................................................................. 6
3.2 Output Specifications................................................................................................................................ 6
3.3 Electric Schematics ................................................................................................................................... 6

1 INTRODUCTION

1.1 Description
The Auxiliary Power Supplies provides a medical grade power supply rated 24 VAC, 150 VA
(2.7A) and an optional 24 VDC, 65 W.

This option is used with an EPS generator and requires that the generator be supplied by a
400Y/230VAC electrical line (230 VAC Phase-to-Neutral). The electrical line must therefore
provide a neutral wire to be connected to J2 (Neutral) lug on the Line Filter Board P/N 9916.00.

Prerequisite for this option is the use of a 400Y/230 VAC three-phase line for
electrical supply of the EPS generator. The neutral must be connected on the
neutral lug of the Line Filter board; see Technical Manual for connection.

Document P/N 11050.000.G2 3/6


Auxiliary Power Supplies Option EPS 45-80 High-Voltage Generator

Connector
Terminal Strip

Medical Grade
Power Supply Power Transformer

Line-Access Panel
(Inner side)
To J7-4

To J7-3

Figure 1 – Auxiliary Power Supplies Option P/N 10 023.000 (24 VDC/24 VAC), installed on the inner
side of the Line-Access Panel. Note that the option version P/N 10 023.001 does not include
the Power Transformer and the 24 VAC output.

1.2 Generator Configuration


When the Auxiliary Power Supplies option is fitted on the generator, the regular
generator hardware setup is modified for the following configuration.
1. The contacts for K1 power-on relay normally provided on pins 3 & 4 of connector J7
(Line Filter Board P/N 9916.xx) are not available with this option. Instead, the installer
can use the 24 VDC (+/-) output supply, located on the Auxiliary Power Supplies’ upper
terminal strip, inner side of the Line-Access Panel (see Figures 1 & 3).
2. Fuses F3 & F4 (4A-500VAC/DC) are installed on the board.
3. The quick connect jumper installed between TB3-TB4 on the regular generator is replaced
by the following two (2) quick connect jumpers: 1. TB3-TB1 and 2. TB4-TB5.

Document P/N 11050.000.G2 4/6


Auxiliary Power Supplies Option EPS 45-80 High-Voltage Generator

2 INSTALLATION

2.1 Power Line Connection

1. Refer to your Technical Manual for the complete power line connection instructions.

2. Make sure the mains line is turned off at the main circuit breaker.

3. When removing the Line-Access Panel at the back of the EPS High-Voltage Generator, it is
recommended to unplug the Auxiliary Output wires from J7 of Line Filter Board P/N 9916.00
(Figure 2). If you do not disconnect
them, be aware that all of the weight of
the panel does not apply directly on the
connected wires.

- The Line-Access Panel will be reinstalled at


the end of the generator installation only -

4. When connecting Unplug Auxiliary Output Wires


the mains line of the From J7- pins 3 & 4
disconnect switch to
the Line Filter Board, make sure that a
Neutral is connected to J5 lug. Do not
use DELTA line configuration.

Figure 2 –
EPS 45-80 High-Voltage generator
with Auxiliary Power Supplies Option.

2.2 K2 Relay and Auxiliary Output Supplies Connection

1. It is recommended to perform the X-ray tube connection before the following


section is done. Refer to your Technical Manual for instructions.

2. If used in your installation, connect K2 programmable relay,. K2’s contacts are programmed
using Epsilon PC Tools; see Technical Manual, Chapter 4.

Document P/N 11050.000.G2 5/6


Auxiliary Power Supplies Option EPS 45-80 High-Voltage Generator

3. On the inner side of the Line-Access Panel, connect the 24


VDC output on the upper terminal strip. If fit with the option,
connect the 24 VAC output on the lower terminal strip.

Figure 3 – Terminal Strips on the


Auxiliary Power Supplies.

4. Reinstall the Auxiliary Output wires on J7 pins 3 (H) & 4 (N) of Line Filter Board.

5. Reinstall the Line-Access Panel.

6. Firmly attach the 24 VDC (/VAC) output wires using the cable holder.

3 Electrical Specifications

3.1 Input Specifications


Shunted from the Epsilon High-Voltage Generator power line supply.
 400Y/230VAC electrical line, 50/60 Hz (Neutral wire mandatory).

3.2 Output Specifications


 24 VDC, 2.7 A
 24 VAC, 6.25 A, 65 W (fitted on Auxiliary Power Supplies P/N 11028.000 only).

3.3 Electric Schematics

Figure 4 –Auxiliary Power Supplies Electronic Schematics.

Document P/N 11050.000.G2 6/6