Contents
1. Components
2. Specification
3. Installation
4. S/W Update
6. Maintenance
7. Principle
양식-GE02-15 (Rev.03)
Basic Components
Optional Components
양식-GE02-15 (Rev.03)
Specification
Item Specification
Weight 1.3kg
DC 12V/5A
AC/DC Adaptor
Output: DC 12V/5A
Temperature 15-35℃
Operation temperature / humidity
Humidity 10~70% (Non-condensing)
양식-GE02-15 (Rev.03)
Installation of ichroma II
ichroma™ Ⅱ displays the product logo and loading screen serially. When loading, it takes few seconds. Please wait.
If loading screen lasts more than 5 minutes, please turn off and on ichroma™ Ⅱ once again.
If the problem continues, please contact Boditech Med Inc.’s technical service or authorized regional distributor.
양식-GE02-15 (Rev.03)
Installation of ichroma II
**The initial User ID and P/W are factory set-up. The P/W can be
changed from Setup→ User ID→ admin→ Change Password
양식-GE02-15 (Rev.03)
Use of Batteries
The use of batteries is highly recommended just in case of a blackout for instant
operation. It is not meant for normal use with batteries for a long time.
It is possible to run 30 multi tests or 2 single tests with batteries.
We suggest to use AC adaptor and power cable for stable and safe operation.
It is possible to use AC adaptor and batteries simultaneously.
However, we suggest to remove batteries when AC adaptor is used.
양식-GE02-15 (Rev.03)
Application Update
양식-GE02-15 (Rev.03)
Application Update
양식-GE02-15 (Rev.03)
Firmware Update
양식-GE02-15 (Rev.03)
Spot Calibration
양식-GE02-15 (Rev.03)
Calibration /Adjusting the
test values
(Target − Measured) 15 − 12
100 = 25 (%)
CAL = 100 12
Measured
14
14
Back up (To save raw data)
양식-GE02-15 (Rev.03)
Maintenance_Stepper moter replacement
Motor block
Stepper motor
Optic
Moving base
양식-GE02-15 (Rev.03)
Maintenance_Stepper moter replacement
1. Unscrew the bolts from 2. Remove the upper case 3. Turn on the reader
the bottom case and front case
양식-GE02-15 (Rev.03)
Maintenance_Stepper moter replacement
7. Unscrew 2 bolts of 8. Take out the barcode 9. Turn the axis of motor
barcode sensor sensor counterclockwise
10. Take out the holder 11. Unscrew 2 bolts from 12. Unscrew 2 bolts from
the moving base the motor bracket and
replace the stepper motor
with the new one
양식-GE02-15 (Rev.03)
Overview of i-Chroma diagnostic system
• Protein-based
• Antigen-antibody reaction (immuno..)
• Lateral Flow Immuno Assay
• Multi-parameter
• POC diagnostic system
• Fluorescence signal
• Remote operation supported
• Printer supported
양식-GE02-15 (Rev.03)
Detecting (Invisible) Antibodies
양식-GE02-15 (Rev.03)
Visualizing the Invisible
양식-GE02-15 (Rev.03)
Detecting the Invisible
Target
양식-GE02-15 (Rev.03)
Understanding Fluorescence
양식-GE02-15 (Rev.03)
Fluorescence
양식-GE02-15 (Rev.03)
Fluorescence
• Luminescence from certain materials under external illumination
• Emitted light is less energetic than the incoming
Excitation Emission
26
Paper Chromatography
http://www.youtube.com/watch?v=Oy_yTOwyRLM
27
Lateral Flow
Immuno-chromatography
ELISA on paper chromatography base
http://www.youtube.com/watch?v=FvIIozN58gw
28
ichroma Technology
29
ichroma Technology
Control Ab/fluo
Test Ab/fluo
Antigen
Detection buffer
Test Control
30
4500
4000
3500
3000
F-units
2500
2000
1500
1000
500
0
0 20 40 60 80 100 120 140 160
1st Test zone 2nd Test zone
31
Optics System
Detector 1
Detector 2
Laser
Barcode
Sensor
LED
Photo sensor
B.
B - 1.시료
B - 2.LE D 광원
B - 3.A p erture
B - 4.P ho to d etecto r
B -4
LED
B -3
B -2
B -1
Sample Well
33
Technology Development
WHAT NEXT?
Development direction
Price
Automation User-friendliness High sensitivity
competitiveness
34
35