Professional Documents
Culture Documents
i
Contents
Contents
1.1 General safety statements these Instructions for Use should only be made following
consultation with the respective manufacturers or an
z Before using the product, carefully read the Instructions expert.
for Use.
z Strictly follow the Instructions for Use. The user must fully Use in areas subject to explosion hazards
understand and strictly observe the instructions. Use the z Devices or components for use in explosion-hazard areas
product only for the purposes specified in the Intended use which have been tested and approved according to
section of this document. national, European or international Explosion Protection
z Do not dispose of the Instructions for Use. Ensure that they Regulations may only be used under the conditions
are retained and appropriately used by the product user. specified in the approval and with consideration of the
relevant legal regulations. The devices or components may
z Only fully trained and competent users are permitted to use not be modified in any manner. The use of faulty or
this product. incomplete parts is forbidden. The appropriate regulations
z Comply with all local and national rules and regulations must be observed at all times when carrying out repairs on
associated with this product. these devices or components.
z Only trained and competent personnel are permitted to
inspect, repair and service the product as detailed in these 1.2 Definitions of alert icons
Instructions for Use. Further maintenance work that is not
detailed in these Instructions for Use must only be carried The following alert icons are used in this document to provide
out by Dräger or personnel qualified by Dräger. Dräger and highlight areas of the associated text that require a greater
recommend a Dräger service contract for all maintenance awareness by the user. A definition of the meaning of each
activities. icon is as follows:
z Properly trained service personnel must inspect and
service this product as detailed in the Maintenance section WARNING
of this document. ! IIndicates an imminently hazardous situation which, if
z Use only genuine Dräger spare parts and accessories, or not avoided, will result in death or serious injury.
the proper functioning of the product may be impaired.
Do not use faulty or incomplete products, and do not modify CAUTION
z
the product. ! Indicates a potentially hazardous situation which, if not
z The threads for the explosion proof enclosure do not avoided, could result in physical injury, or damage to
conform to the minimum/maximum values in EN/IEC the product or environment. It may also be used to
60079-1. The threads must not be reworked by the user. alert against unsafe practices.
z The measuring function of the transmitter for explosion NOTICE
protection according to Annex II, Clauses 1.5.5, 1.5.6 and ii Indicates additional information on how to use the
1.5.7 of Directive 94/9/EG is not covered at present.
product.
z Only operate the product within the framework of a risk-
based alarm signalling concept.
2 Description
2.1 Product overview
2.1.1 Explosion proof device 2.1.2 Explosion proof device with increased safe wiring
compartment (e-Box)
1 3
2
4
3
5
4
6
5
7
8
6
001233300.eps 002233300.eps
2.2 Functional description 2.3.2 Dräger Polytron 8720 with PIR 7200 infrared gas
sensor Type CO2
The device is operated from a 10 to 30 volt DC supply.
The current gas concentration, status messages and menu For monitoring carbon dioxide.
items are shown on a backlit LCD display and by three colored z Measuring range (IDS 01x5): 0 to 0.2 up to 30 vol% CO2
LEDs. The measured gas concentration is output as a 4 to
20 mA analog signal or a digital HART® signal1 After the power
is switched on, the device detects whether the 4-20 mA WARNING
interface is to be operated as a current source or a current ! Explosive. Not to be used in oxygen enriched
sink. The flexible design permits fitting even in places that are atmospheres. None of the Polytron 8000 transmitters
difficult to access. The Dräger PIR 7000/7200 can be fitted on is certified and approved to be operated in oxygen
either side of the of the equipment and installation with up to enriched atmospheres.
30 m between the sensing head and the PIR 7000/7200 is
possible. In the variant with the integrated e-Box with the
"increased safety" protection class, the device can be easily
connected to the electrical field wiring. The device can be
2.4 Intended area and conditions of use
configured, calibrated and serviced without declassification of
the area. Hazardous areas classified by zones:
The device is only designed for use in Zone 1 or Zone 2
potentially explosive atmospheres for which a temperature
2.3 Intended use range as specified on the device must be observed and in which
The Dräger Polytron 87x0 is an explosion proof transmitter for gases of Explosion Groups IIA, IIB or IIC and Temperature
continuous monitoring of the concentration of gases and vapors Classes T4 or T6 (depending on maximum ambient
containing hydrocarbons or carbon dioxide. The housing for the temperature) or dust of Groups IIIA, IIIB or IIIC may occur.
device consists of robust, non-corroding stainless steel or
aluminum and is suitable for indoor or outdoor applications. Hazardous areas classified by divisions:
The unit can be connected to a Dräger monitoring system or The device is only designed for use in potentially explosive
a programmable logic controller (PLC) by means of a sealed atmospheres of Classes I&II, Div. 1 or 2, for which
conduit or an appropriately approved cable gland. With the a temperature range as specified on the device must be
optionally integrated relay module, the device can be operated observed and in which gases or dust of Groups A, B, C, D or
without a central controller (with additional local alarm E, F, G and Temperature Classes T4 or T6 (depending on
signalling). The device is designed for fixed installation and is maximum ambient temperature) may occur.
approved for use in potentially explosive atmospheres,
see chapter 2.5 on page 6. 2.5 Approvals
!
WARNING
To prevent ignition of explosive atmospheres, strictly
follow the warnings below:
D LA
Polytron 8700
ITR 0421
Dräger Safety 23560 Lübeck, Germany
0158
II 2G
II 2D
Gas Detector for Use in Hazardous Locations as to
SSI FI E
Part No: 4544638
Serial No. XXXX-9999
Sample
Fire, Electrical Shock and Explosion Hazards only
D
C
NOTICE
3.5 Electrical installation without e-Box
ii A dust cap can be fitted to the cable entry of the device.
This cap is only used for transport purposes and must CAUTION
be removed before the device is connected to a sealed ! First connect the cables for the relays and make
conduit. the connections to the sensor before connecting the
device to the power supply.
NOTICE
ii There must be a free space of 5 cm next to the
PIR 7000/7200 to permit fitting/removal of the splash
guard and the later fitting of accessories.
Graphic symbols facilitate navigation in the various menus. In addition, the following symbols may be displayed:
z " ", if the measurement range is exceeded
Together with the text » Back «, » Menu «, etc., exit the z " - - - - " and " X " in the event of a malfunction. The yellow
menu or go back 1 step. LED lights, (see chapter 4.2.2 on page 12).
Closed folder z "SIL", when SIL status is activated.
Further functions or sub-menus are located under
this item. When the optional relay PCB is used:
Opened folder z If the first alarm has been triggered, the red LED gives
The functions and sub-menus available here are listed single flashes.
under this item. z If the second alarm has been triggered, the red LED gives
Function double flashes.
When activated, functions can be executed in one or
more steps.
Selection activated
For functions that can be selected and activated,
the activation is initiated by pressing » OK «.
List closed above / complete
There are no further functions, menus or sub-menus
listed above.
List can be scrolled upwards
There are further functions, menus or sub-menus
listed above.
Information can be called up in Info mode, see chapter 4.3 z The Info mode can be ended at any time by selecting OK .
on page 12.
z If no button is selected within 30 seconds, the device
Error message present – to display the error, reverts automatically to the measurement mode.
see chapter 7.1.2 on page 17
Maintenance signal will be output on the interface, Example of the Info mode
see chapter 8.3.2 on page 21. (Displays vary depending on sensor):
Measurement range of analog interface exceeded 1. Display - Device information:
Measurement less than range of analog interface 07.11.2003 12:34
Analog interface is set to a fixed value (e.g. Multidrop HART SW Version : 1
communication) and is not transmitting any measurements. Part No. : 8317778
Serial No. : ARUA0001
"Preventive" maintenance: The sensor is ready for operation. DeviceCode: 00006317
1/3
"Preventive" maintenance: The sensor is ready for operation Instrument Info
but is close to the end of its life cycle. A001.eps
"Preventive" maintenance: The sensor is still ready for z Date and Time
operation but should be changed as soon as possible. z Software version
The datalogger is active in rolling mode. z Part number of the device
To activate/deactivate, see chapter 8.5.2 on page 25. z Serial number of the device
The datalogger is active in stacking mode. z Device code
To activate/deactivate, see chapter 8.5.2 on page 25.
SIL SIL is activated (observe Safety Manual - Order Number 2. Display - Sensor information:
9033307). Sensorname: O2
Part No. : 6809630
Serial No. : XXXXXXX
4.2.2 LED symbols EEPROM Typ : 1
EEPROM Vers.: 1
2/3
Sensor Info
Symbol LED Description A002.eps
z If the first alarm has been triggered, the red LED gives z Gas description
single flashes. A1 relay is activated.
z Maximum measurement limit and measurement units for
z If the second alarm has been triggered, the red LED gives the sensor
double flashes. A2 relay is activated.
z Measurement range and measurement units for the analog
z If an alarm can be acknowledged and is acknowledged, interface. Only displayed when the 4-20 mA interface is
the LED stops flashing and lights continuously instead until enabled.
the alarm condition is no longer present. The corresponding
relay is deactivated.
z Tap M
and hold for 1 to 2 seconds. The device uses various current values to indicate different
The Quick Menu will open. This menu is displayed without operating modes. This conforms to the NAMUR NE43
any password request. No changes can be made. recommendation. The 4-20 mA interface on the device can be
The selected information about status and device settings customized and configured to individual requirements.
can be retrieved (e.g. warning messages, error messages, The device is provided with suitable standard settings at the
installed modules and calibration data). factory (see chapter 13.3 on page 33).
Password presets:
Password for the Calibration menu: _ _ _ 1
Password for the Settings menu: _ _ _ 2
NOTICE
ii The correct date and time settings are important for
many functions (see chapter 8.2.3 on page 20).
5 Calibration
Repeat the calibration of the device at regular intervals as
specified in the particular sensor data sheet.
CAUTION
! Danger to health! Do not inhale the test gas. Observe
2
the hazard information in the corresponding safety
data sheets. Provide venting into a fume cupboard or
outside the building.
z Actual value within the permissible range: 5.2 Performing Auto Calibration
{ Select Next and confirm.
The device offers the option of an automatically sequenced
{ Shut off the test gas and remove the calibration adapter
calibration. This reduces the number of otherwise manual
from the PIR 7000/7200.
interactions with the device. The auto calibration is only
z Actual value outside the permissible range: suitable for experienced users, as calibration errors can occur
{ Select Next and confirm. here as a result of tightened procedures.
{ Wait until the measurement is stable.
NOTICE
5.1.2 Span calibration ii Not all of the available sensors and gases support auto
1. Select Calibration > Span calibration and confirm. calibration. If the function is not available, a manual
The Maintenance signal is transmitted. calibration must be performed.
The values for the calibration gas are displayed, e.g.:
Gas Methane 1. Select Settings > Sensor > Autocal. set. to enable or
Unit %LEL disable the function.
Conc. 50.0 2. Supply fresh air or feed nitrogen or compressed air at
a flow rate of 0.5 - 2.0 L/min to the sensor.
All three paramrters can be changed:
z Select Gas and confirm. WARNING
{ Select the desired calibration gas from the list and ! It must be insured that during the zero point calibration,
confirm. no measured gas is present in the supplied calibration
z Select Unit and confirm. gas. Particularly with a fresh air calibration, Dräger
{ Select the desired unit from the list and confirm. recommends performing a second, independent
z Select Conc. and confirm.
measurement. Some sensors may only be calibrated
with synthetic air or nitrogen (observe the relevant
{ Set the desired concentration of the calibration gas.
sensor data sheets).
When the settings are correct:
3. Select Calibration > Auto cal. and confirm.
2. Select Next and confirm.
{ Please wait... is displayed and a calibration of the zero
A message such as, e.g. Gas flow to METHANE
point is performed automatically.
is displayed.
{ After a successful zero point calibration, the span
3. Connect the calibration gas to the sensor.
calibration is started.
4. Select Next and confirm.
{ The calibration gas concentration, the units and, where
The current value is displayed. appropriate, the calibration gas are displayed and can
be changed.
After the displayed value is stable:
4. Start the calibration with Next; Back to menu aborts the
5. Select Next and confirm. calibration.
The message Please wait… is displayed.
The actual value is displayed.
NOTICE
z Actual value within the permissible range:
{ Select Next and confirm.
ii After 10 minutes without confirmation, the device
{ Shut off the test gas and remove the calibration adapter reverts to measurement mode without performing
from the PIR 7000/7200. a calibration.
z Actual value outside the permissible range:
5. Feed the relevant calibration gas to the sensor at a flow
{ Select Next and confirm.
rate of 0.5 - 2.0 L/min.
{ Wait until the measurement is stable.
{ When the measured value is stable, the device performs
the calibration automatically.
z A calibration can be aborted at any time. { After a successful calibration, the measurement and
the message Value OK? are displayed.
6. When the measurement is stable, shut off the calibration
gas and wait until the measurement is below any possible
alarm thresholds again.
z The calibration is performed again with Redo.
z Complete the calibration with Accept value.
The device returns to the measurement mode.
6 Menu overview
◄ Back to
measure
NOTICE
ii The fault code is very helpful for technical support and
should always be stated in full in the event of a fault.
8 Settings menu 1. Select Settings > Instrument > Alarm > Relay A1 or
Relay A2 and confirm.
This menu contains all the functions that are required for the 2. Select Normal energy supply oder Energy supply
individual configuration of the device. during alarm and confirm.
Overview of standard alarm settings (for Polytron 8700 1. Select Settings > Instrument > Alarm > Set Alarm A2
with PIR 7000, configured for methane, full scale reading and confirm.
of 100 %LEL) 2. Select On or Off and confirm.
When the function is activated, the relay and the interface are
A1/Relay 1 A2/Relay 2
set to the A2 alarm status and the red LED gives double flashes.
Alarm threshold 20 %LEL 40 %LEL
Testing Fault status
Hysteresis 0 %LEL 0 %LEL
NOTICE
Relay Normally
energized
Normally
energized
ii When the function Set fault is exited in the menu,
the Polytron 87X0 reverts automatically to measurement
Relay alarm: Rising Rising mode.
Direction
Latching mode Non-latching Latching This function simulates the Fault alarm.
Acknowledgement Acknowledgeable Not 1. Select Settings > Instrument > Alarm > Set fault
mode acknowledgeable and confirm.
2. Select On or Off and confirm.
The function for setting the A1 hysteresis is opened. When the function is activated, the Fault relay and the 4-20 mA
This function permits the setting of a zone within which interface are set to the fault current and the yellow LED lights.
a triggered relay maintains its status until the gas concentration
is outside the zone. This function prevents relays from
chattering at an alarm threshold.
Example: A2 at 40 %LEL methane
Maximum possible hysteresis: 3 %LEL
Alarm activation at measured values ≥40 %LEL
Alarm deactivation at measured values ≤37 %LEL (40 %LEL -
3 %LEL)
11. Select the line for entering the hysteresis and confirm.
12. Set the hysteresis for A1 and confirm.
The hysteresis setting for A1 is displayed.
13. Select Next and confirm.
The settings for A1 are displayed.
14. Select and confirm with Confirm.
The settings for the A1 alarm are now complete.
Testing Alarm A1
NOTICE
ii When the function Set Alarm A1 is exited in the menu,
the Polytron 87X0 reverts automatically to measurement
mode.
When the function is activated, the relay and the interface are
set to the A1 alarm status and the red LED gives single flashes.
Testing Alarm A2
NOTICE
ii When the function Set Alarm A2 is exited in the menu,
the Polytron 87X0 reverts automatically to measurement
mode.
1. Select Settings > Instrument > Function key and confirm. 8.2.8 Display settings
2. Select the desired function and confirm.
Changing the contrast
z Repeating the confirmation deactivates the selected
function. 1. Select Settings > Instrument > Display > Display contrast
and confirm.
z If Bumptest is selected:
The maintenance symbol is shown on the right hand 2. Change the contrast and confirm.
side of the display and the maintenance signal is output.
Repeating the confirmation ends the bump test and the Changing the display mode
device reverts to the normal measurement mode. The display of measured values can be turned on or off with
this function.
z In measurement mode, the selected function can be
activated by briefly pressing . 1. Select Settings > Instrument > Display > Display mode
and confirm.
Selection options: 2. Select Standard or Non display and confirm.
If "Non display" has been selected, the display shows the start
Graph (only with The measurements from the sensor screen and the symbols.
data logging) are displayed graphically on a time axis
(see chapter 7.3.2 on page 17). If the relay option is set and an alarm is triggered, the display
Faults Faults are displayed in plain text shows the current gas concentration and the red LED flashes,
(see chapter 7.1.2 on page 17). regardless of the selected display mode.
Switching the display test on or off 1. Select Settings > Communication > HART interface >
The function of the display and the LEDs is tested with Tag and confirm.
this function. 2. Select the tag line for the entry and confirm.
1. Select Settings > Instrument > Display > Displaytest 3. Set the tag and confirm.
and confirm. 4. Select Confirm and confirm.
2. Select On or Off and confirm.
Selecting Dräger REGARD
When this function is activated, the display is shown inverted This function enables switching between different HART
and all the LEDs are lit. protocols.
Test functions for the analog interface Testing the maintenance signal
This function is used to set the analog interface to the
maintenance signal.
NOTICE
ii Alarms in the central controller may be triggered by
1. Select Settings > Communication > Analog interface >
Set mainten. and confirm.
these functions! If necessary, the alarms in the central 2. Select On or Off and confirm.
controller must be disabled beforehand. { The current for the maintenance signal will be
transmitted on the analog interface.
Setting current signals Testing the beam block signal (for PIR 7000/7200 only)
Using this function, various currents in the range from 0 to This function is used to set the analog interface to the beam
22 mA can be set on the analog interface. block signal.
1. Select Settings > Communication > Analog interface >
1. Select Settings > Communication > Analog interface > Set beamblock and confirm.
Set current and confirm.
2. Select On or Off and confirm.
2. If the alarm in the control unit is disabled, confirm the
{ The current for the beam block signal will be
message Inhibit all alarms.
transmitted on the analog interface.
3. Select the current line for the entry and confirm.
4. Set the currrent and confirm. 8.3.3 Profibus address
The setting for the current is displayed.
5. Select Next and confirm. Not yet active - for future functions.
6. Select Set current out or Current off and confirm.
7. Select Next and confirm. The function is terminated.
8. After the alarm settings in the control unit have been re-
enabled, confirm the Enable all alarms message.
Example:
NOTICE
NIOSH Category, measured gas: Methane 100 %LEL
corresponds to 5.0 vol.% ii On means that the PIR 7000/7200 is in the "Fast
Customer specific: Measured gas: methane 100 %LEL response" mode. Off means that the PIR 7000/7200 is
corresponds to 4.7 vol.% methane in the "Normal response" mode.
NOTICE The results in the datalogger can only be evaluated with the
ii The calibration interval can be set between 0 and PolySoft 8000 PC software. An evaluation of the datalogger
720 days. contents over the last 15 minutes is available in the menu
under Information > Datalogger > Graph (see chapter 7.3.2
on page 17).
1. Select Settings > Sensor > Calibration interval
and confirm.
8.5.1 Switching the datalogger on or off
2. Set the calibration interval and confirm.
1. Select Settings > Datalogger > Datalogr. on/off and
8.4.9 Sensor test confirm.
2. Select On or Off and confirm.
NOTICE
ii This function is not available when the PIR 7000/7200
8.5.2 Setting up the datalogger
9 Troubleshooting
9.1 Fault reference:
Fault number Cause Remedy
#001 Serious device fault, various causes. Have the device checked by DrägerService®1.
#003
#004
#005
#011 - #014
#020 - #024
#043
#060
#067
#002 Serious data error in the device, Reset the device to factory settings (see chapter 8.2.6 on
#025 - #027 various causes. page 20).
#030 - #034 If this fault occurs again:
#041 Have the device checked by DrägerService.
#042
#044
#050
#052
#081
#085
#10 4-20 mA interface cable not connected. Disconnect the powwer supply, connect the 4-20 mA interface
and reconnect the power supply.
#051 Faulty zero-point. Perform zero-point calibration (see chapter 5.1.1 on page 14).
#054
#055 Faulty span calibration was performed. Perform new span calibration (see chapter 5.1.2 on page 15).
#064 Communications fault. Check contact with PIR 7000/7200.
#071
#083 PIR 7000/7200 optics dirty. Clean PIR 7000/7200 optics.
#086 Supply voltage outside range. Check supply voltage.
#087 Device fault, 4 - 20 mA PIR 7000/7200. Check 4-20 mA connection on PIR 7000/7200.
#090 Sensor lock function is activated. A sensor Disable the sensor lock function (see chapter 8.4.4 on
with a different part number than was page 24) or use a new sensor with the same part number.
previously configured has been used.
#094 Data error in the device. Reset the sensor to factory settings (see chapter 8.4.7 on
#095 page 25).
If this fault occurs again:
Have the device checked by DrägerService.
#096 Differing SIL password on PIR 7000/7200. Enter the password again.
®
1 DrägerService is a registered trademark of Dräger.
Warning
Cause Remedy
number
#101 Datalogger in Stack mode is 100 % full and is Read out the data, clear the datalogger and start again.
not logging any more data.
#102 Datalogger in Stack mode is 90 % full. Read out the data as soon as possible, clear the datalogger and
start again.
#103 Data error in the device. Functions such as Reset the device to factory settings (see chapter 8.2.6 on
#106 the datalogger or dongle functions may no page 20).
longer be available. If this fault occurs again:
Have the device checked by DrägerService.
#104 No valid date or time set. Set time and date (see chapter 8.2.3 on page 20).
#107 Battery almost used up. Have the device checked by DrägerService.
#110 SW dongle has been removed without Disable SW dongle (see chapter 8.2.7 on page 20).
#111 logging out.
#112
#113 Alarms disabled. Enable alarms (see chapter 8.2.1 on page 18).
#164 Sensor warm-up phase has not ended yet. Wait until the sensor has warmed up.
#165 Increased measurement error must be
expected.
#167 Calibration interval expired. Recalibrate the device (see chapter 5 on page 14).
#170
#171 Increased zero-point offset. Perform zero-point calibration (see chapter 5.1.1 on page 14).
#172 PIR 7000/7200 optics dirty. Clean PIR 7000/7200 optics.
#182 Autocalibration not possible with PIR 7200. Perform new zero-point and span calibration (see chapter 5.1.2
on page 15).
10 Maintenance WARNING
! Five threads must be engaged to insure explosion
10.1 Maintenance intervals protection.
10.1.1 During commissioning 8. Plug the sensor cable into the socket.
z Check calibration. 9. Re-fit the sensor electronics and enclosure.
z Check signal transmission to controller and triggering of 10. Screw the housing cover on again and tighten the retaining
alarms. screw.
11. Switch on the power to the device.
10.1.2 Every 6 months 12. If necessary, switch off SIL activation (see chapter 8.1 on
page 18).
z Inspection by trained service personnel.
13. Calibrate the device if necessary (see chapter 5 on
page 14).
The inspection intervals must be established in each individual
case and shortened if necessary, depending on technical 14. If required, switch on SIL activation (see chapter 8.1 on
safety considerations, engineering conditions, and the page 18).
technical requirements of the equipment.
Contact DrägerService if you wish to take out a service 10.3 Changing the main electronics
contract and have any repairs carried out.
1. Switch off the power to the unit or declassify the area
according to the local regulations.
10.1.3 As required
2. Release the retaining screw and unscrew the housing
z Change the sensor, see chapter 10.2 on page 29. cover from the device.
3. Lift the handle and remove the enclosure with the main
10.2 Changing the sensor electronics.
4. Remove the protective caps for the connectors.
5. Unscrew the connector locking ring for the power supply
WARNING and relay and withdraw the connector.
! As a matter of principle, always check a newly installed 6. Change the main electronics.
sensor with measured gas to insure correct operation. 7. Plug the sensor cable and the connector for the power
supply and relay into the corresponding sockets and screw
CAUTION them tight.
! When the sensor is changed, all settings and 8. Re-fit the dust caps for the connectors.
parameters must be checked for correctness. 9. Re-fit the sensor electronics and enclosure.
NOTICE 10. Screw the housing cover on again and tighten the retaining
ii The analog interface of the PIR 7X00 must be
screw.
11. Switch on the power to the device.
configured to the following values (PIR 7x00 factory 12. Check and adjust the configuration as required.
settings): 13. Calibrate the device if necessary (see chapter 5 on page 14).
Maintenance signal: 3.0 mA
Alt. maintenance signal: Off
Beam block warning: 2 mA (inactive) 11 Disposal
Alternating warning signal: Off
Fault signal: 1 mA
Disposal of electrical and electronic equipment:
The sensor is changed as follows: EU-wide regulations governing the disposal of
electric and electronic appliances which have been
1. Switch off the power to the unit or declassify the area defined in the EU Directive 2002/96/EC and in
according to the local regulations. national laws have been effective from August 2005
2. Release the retaining screw and unscrew the housing and apply to this device. Common household
cover from the device. appliances can be disposed of using special
3. Lift the handle and remove the enclosure with the main collecting and recycling facilities. However, as this
electronics. device has not been registered for household usage,
4. Disconnect the sensor cable. it must not be disposed of through these means.
5. Unscrew the PIR 7000/7200. The device can be returned to your national Dräger
6. Pass the sensor cable through the threaded connection in Safety Sales Organization for disposal. Please do not
the housing. hesitate to contact the above if you have any further
7. Screw in the PIR 7000/7200. questions on this issue.
Substance CAS No. Gas Code Substance CAS No. Gas Code
Acetaldehyde 75-7-0 402 Methyl acrylate 96-33-3 581
Acetone 67-64-1 453 Methyl tert-butyl ether 1634-04-4 355
Allyl chloride 107-05-1 723 Methyl isobutyl ketone 108-10-1 476
Benzene 71-43-2 260 Methyl chloride 74-87-3 701
1,3-Butadiene 106-99-0 172 Ethyl methyl ketone 78-93-3 454
n-Butane 106-97-8 104 n-Nonane 111-84-2 109
i-Butane 75-28-5 114 n-Octane 111-65-9 108
1-Butene 106-98-9 154 n-Pentane 109-66-0 105
i-Butene 115-11-7 164 1-Pentene 109-67-1 155
2-Butoxyethanol (n-Butyl glycol) 111-76-2 384 Propane 74-98-6 002
n-Butyl acetate 123-86-4 554 Isopropanol 67-63-0 313
n-Butyl acrylate 141-32-2 584 Propylene 115-07-1 153
Chlorobenzene 108-90-7 280 Propylene oxide 75-56-9 823
1-Chloro-1,1-difluoroethane Styrene 100-42-5 270
75-68-3 783
(R 142b)
Toluene 108-88-3 261
Cyclohexanone 108-94-1 496
o-Xylol (1,2-Dimethyl benzene) 95-47-6 265
1,2-Dichloroethane 107-06-2 732
The current gas library can be viewed at www.draeger.com.
Dichloromethane 75-09-2 730
1,2-Dichloropropane 78-87-5 734
1,3-Dichloro-1-propene 542-75-6 744
Diethyl ether 60-29-7 354
Dimethyl ether 115-10-6 352
Epichlorohydrin 106-89-8 820
Ethane 74-84-0 102
Ethanol 64-17-5 302
1-Ethoxypropan-2-ol 1569-02-4 392
Ethyl acetate 141-78-6 552
Ethylene 74-85-1 003
Ethylene oxide 75-21-8 822
n-Hexane 110-54-3 106
1-Hexene 592-41-61 156
Methane 74-82-8 001
Methanol 67-56-1 301
1-Methoxypropan-2-ol 107-98-2 391
2-Methoxy-1-methyl ethyl acetate 108-65-6 570
Substance CAS No. Gas Code Substance CAS No. Gas Code
Acetaldehyde 75-7-0 402 Isopropanol 67-63-0 313
Acetone 67-64-1 453 Propylene 115-07-1 153
Allyl chloride 107-05-1 723 Propylene oxide 75-56-9 823
n-Butane 106-97-8 104 Toluene 108-88-3 261
i-Butane 75-28-5 114 o-Xylol (1,2-Dimethyl benzene) 95-47-6 265
1-Butene 106-98-9 154 The current gas library can be viewed at www.draeger.com.
i-Butene 115-11-7 164
2-Butoxyethanol (n-Butyl glycol) 111-76-2 384
n-Butyl acetate 123-86-4 554
n-Butyl acrylate 141-32-2 584
Cyclohexane 110-82-7 186
Cyclohexanone 108-94-1 496
1,2-Dichloroethane 107-06-2 732
Dichloromethane 75-09-2 730
1,2-Dichloropropane 78-87-5 734
1,3-Dichloro-1-propene 542-75-6 744
Diethyl ether 60-29-7 354
Dimethyl ether 115-10-6 352
Epichlorohydrin 106-89-8 820
Ethane 74-84-0 102
Ethanol 64-17-5 302
1-Ethoxypropan-2-ol 1569-02-4 392
Ethyl acetate 141-78-6 552
Ethylene oxide 75-21-8 822
n-Hexane 110-54-3 106
1-Hexene 592-41-61 156
Methane 74-82-8 001
Methanol 67-56-1 301
1-Methoxypropan-2-ol 107-98-2 391
2-Methoxy-1-methyl ethyl acetate 108-65-6 570
Methyl acrylate 96-33-3 581
Methyl tert-butyl ether 1634-04-4 355
Methyl isobutyl ketone 108-10-1 476
Methyl chloride 74-87-3 701
Ethyl methyl ketone 78-93-3 454
n-Nonane 111-84-2 109
n-Octane 111-65-9 108
n-Pentane 109-66-0 105
1-Pentene 109-67-1 155
Propane 74-98-6 002
13 Technical data
CAUTION
! Specifications and restrictions in the Instructions for Use and/or data sheets for the sensors used must be observed.
For SIL applications the Dräger Polytron 8X00 Safety Manual must be observed.
The measuring range and performance characteristics are dependent on the sensor installed, see the Instructions for
Use and/or data sheets for the sensors used.
Supply voltage
Operating voltage 3-wire shielded cable, 24 V DC (10 to 30) V DC
Inrush current 2.3 A for 2 ms at 24 V DC, 10 Ω conductor resistance
Housing NEMA 4X
Dimensions (LxWxD)
Without e-Box approx. 150 x 290 x 130 mm (5.9" x 11.4" x 5.1")
With e-Box approx. 180 x 300 x 190 mm (7.1" x 11.8" x 7.4")
Weight
Without e-Box, aluminum approx. 3.9 kg (8.5 lbs.)
Without e-Box, stainless steel 316 approx. 5.7 kg (12.6 lbs.)
With e-Box, aluminum approx. 5.2 kg (11.5 lbs.)
With e-Box, stainless steel 316 approx. 7.1 kg (15.6 lbs.)
Environmental conditions Specifications for the sensor: see sensor data sheet
Pressure 700 to 1300 hPa (20.7 to 38.4 in. Hg)
Humidity 0 to 100 % R. H., non-condensing
Temperature
without relay -40 to +77 °C (-40 to +170 °F)
with relay -40 to +70 °C (-40 to +158 °F)
Yellow Fault LED Lights when a fault is present FSD Set 100 100 5
(see chapter 9 on page 27). Password 0000 0000 0000
Red Alarm LED Gives single flashes when an A1 alarm PIR Lock Off Off Off
is present. SIL password _ _ _2 _ _ _2 _ _ _2
Gives double flashes when an A2 alarm
is present. SIL status Off Off Off
If an alarm is configured as Language English English English
acknowledgeable and the alarm is
Fault current 1.2 mA 1.2 mA 1.2 mA
acknowledged, the LED lights.
Alarm hierarchy: A2 has a higher priority Warning current 3.0 mA 3.0 mA 3.0 mA
for the LED than A1. However the A1 and Maintenance
A2 relays operate independently of one 3.4 mA 3.4 mA 3.4 mA
current
another, i.e. if A1 is acknowledgeable and
A2 is not and the gas concentration is so Beam block Off Off Off
high that A1 and A2 are triggered, an HART address 0 0 0
acknowledgement results in the A1 relay
Auto-cal. Off Off Off
dropping out. However the red LED
continues to give double flashes as long Calibration
360 360 360
as the A2 condition exists. interval
Response Normal Normal Normal
Function key Faults Faults Faults
14 Order list
14.1 Dräger Polytron 8700 Type 334 14.3 Dräger Polytron 8720
Order Order
Name and description Name and description
Number Number
Dräger Polytron 8700 334 d A 4-20/HART 45 44 601 Dräger Polytron 8720 d A 4-20/HART 45 44 709
Dräger Polytron 8700 334 d A 4-20/HART Relay 45 44 602 Dräger Polytron 8720 d A 4-20/HART Relay 45 44 710
Dräger Polytron 8700 334 d S 4-20/HART 45 44 610 Dräger Polytron 8720 d S 4-20/HART 45 44 718
Dräger Polytron 8700 334 d S 4-20/HART Relay 45 44 611 Dräger Polytron 8720 d S 4-20/HART Relay 45 44 719
Dräger Polytron 8700 334 e A 4-20/HART 45 44 619 Dräger Polytron 8720 e A 4-20/HART 45 44 727
Dräger Polytron 8700 334 e A 4-20/HART Relay 45 44 620 Dräger Polytron 8720 e A 4-20/HART Relay 45 44 728
Dräger Polytron 8700 334 e S 4-20/HART 45 44 628 Dräger Polytron 8720 e S 4-20/HART 45 44 736
Dräger Polytron 8700 334 e S 4-20/HART Relay 45 44 629 Dräger Polytron 8720 e S 4-20/HART Relay 45 44 737
Dräger Polytron 8700 Remote 334 e A 4-20/HART 45 44 673 Dräger Polytron 8720 Re e A 4-20/HART 45 44 745
Dräger Polytron 8700 Remote 334 e A 4-20/ Dräger Polytron 8720 Re e A 4-20/HART Relay 45 44 746
45 44 674
HART Relay
Dräger Polytron 8720 Re e S 4-20/HART 45 44 754
Dräger Polytron 8700 Remote 334 e S 4-20/HART 45 44 682
Dräger Polytron 8720 Re e S 4-20/
45 44 755
Dräger Polytron 8700 Remote 334 e S 4-20/ HART Relay
45 44 683
HART Relay
Order
Name and description
Number
D-Junction Box, aluminum 45 44 099
D-Junction Box, stainless steel 316 45 44 098
Duct mounting kit (PIR 7000) Polytron 5000/
68 12 300
8000
Pipe mounting kit, Polytron 5000/8000 45 44 198
IR connection kit, 5000/8000 45 44 197
Magnetic wand with key chain 45 44 101
Order
Name and description
Number
Status indicator PIR 7000 68 11 625
Splash guard PIR 7000 68 11 911
Flow cell PIR 7000 68 11 490
Remote test adapter PIR 7000 68 11 630
Status indicator PIR 7200 68 11 920
Mounting kit PIR 7000 68 11 648
Splash guard PIR 7200 68 11 912
Flow cell PIR 7200 68 11 910
Remote test adapter PIR 7200 68 11 930
Insect guard PIR 7000 68 11 609
Hydrophobic filter PIR 7000 68 11 890
Calibration adapter PIR 7000 68 11 610
Flow cell PIR 7000 Duct 68 11 945
Remote test adapter PIR 7000 Duct 68 11 990
Process adapter PIR 7000 68 11 915
Process cuvette PIR 7000 68 11 415
Junction box Ex e PIR 7000 (EAC 0000) 68 11 898
15 Declaration of Conformity
Manufacturing Location:
Draeger Safety, Inc.
101 Technology Drive
Pittsburgh, PA 15275-1057, USA
Phone +1 412 7 87 - 83 83
Fax +1 412 7 87 - 22 07
9033303 - GA 4683.620
© Dräger Safety AG & Co. KGaA
Edition 01 - September 2011
Subject to alteration