Professional Documents
Culture Documents
In India the first Hemodialysis facility was established in 1961 at the CMC Vellore; soon it was
started at 3-4 major centers during that decade viz. CMC Vellore, KEM Mumbai, PGI Chandigarh &
All India Institute of Medical Sciences, New Delhi.
Unfortunately in our country the quality of dialysis delivered to patients can vary from center to
center. The quality could range from very poor to as good as any center in world. This is because
there are no minimum defined standards of care of maintenance dialysis.
The Indian Society of Nephrology therefore set up core group to write these minimum standards
that should be followed in our country. Our concern is to study the use of quality water for dialysis
delivered to our patients.
The average hemodialysis patient is exposed to approximately 25 times the amount of water
normally ingested by an individual. In addition he or she is deprived of the protective barrier of the
gastrointestinal tract and the detoxification function of the kidneys, increasing the risk several fold
of toxicity caused by the numerous chemical and microbiological contaminants in the water.
The final quality of the water is dependent on the configuration of the treatment system and the
quality of the feed water which itself may be highly variable. The processes of hemodialysis evolve
with use of high flux dialyzers and hemodiafiltration.
This requires a unit to design a system capable of delivering this very high quality from the worst
feed water and a monitoring system and quality assurance to prevent breakdown of the system.
The following guidelines describe the setting up, maintenance and monitoring of a system designed
to reliably provide ultrapure water.
1. Water treatment to achieve the following water quality (AAMI standards) is mandatory for all
hemodialysis units. (Table 1)
2. It is recommended that the hemodialysis units should try to achieve European Standards of
purity of water (Table 2)
3. It is recommended that there are three aspects of water quality that should be achieved.
a. Chemical purity b. Microbiological purity c. Endotoxin purity
8. Ideally storage tank should be avoided but this is practical impossibility hence, it is
recommended that a stainless steel (grade 316) or medical grade PVC water tank could be used for
water storage. The tank must have de-aeration valve & drain facility at the bottom, so that complete
water could be drained out. It should have an air tight lid.
9. It is strongly recommended that all pipelines after reverse osmosis system should be stainless
steel (grade 316) or medical grade PVC. All valves joints & connectors also should be of the same
material. It is recommended that bends & blind loops must be avoided.
10. It is preferable to have 0.22μ membrane filter & ultraviolet light after reverse osmosis online.
Ultraviolet light is also recommended after activated carbon filter & before reverse osmosis.
Monitoring
11. Chemical purity: Online conductivity meters are mandatory after deionizers & reverse osmosis.
There should be visible & audible alarm for improper conductivity in the dialysis technician’s
station. The alarm should lead to stoppage of water beyond reverse osmosis. The water should re-
start only after adequate conductivity is achieved.
Once in 3 months treated water sample must be sent for detailed chemical analysis to an
independent laboratory having adequate instrumentation for testing (see table 1). The results
should be mandatory part of the record system.
12. Microbiological purity: This should be checked once every 15 days to achieve the standards as
per table 2. It is strongly recommended that pour plate method on nutrient poor medium should be
used for cultures of treated water. Incubation should be at room temperature (200 C to 240 C) for 7
days. (See annexure)
13. Endotoxin levels: should be checked once in every 15 days to achieve the standard as per table
2. (Please see annexure)
Sterilization
14. It is strongly recommended that each component of the water treatment system must be
thoroughly cleaned & sterilized as per the manufacturer’s recommendation. The process of
sterilization should be carried out once every 15 days. After sterilization it is essential that the
sterilent is completely removed before the treated water is used for dialysis.
15. Cleaning & sterilization of the water storage tank & plumbing is necessary once every 15 days.
Depth Filter Pressure drop across Backwashing & Twice a week and
filter Rinsing daily during the
monsoon or when
water is contains
extra suspended
particles/impurities.
Storage tank and Bacterial counts >50% increase over Cleaning &
Pipeline baseline. Disinfection
a) Sodium tripolyphosphate and Sodium edetate at a pH adjusted to > 10 to remove calcium scales
and low level organic foulants.
b) 2% Citric acid (no pH adjustment required) which removes calcium carbonate, metal oxides and
inorganic colloidal compounds and also provides disinfection.
A water wash is recommended between the cleaning solutions. A final disinfection of the
membranes using 1% peracetic acid or 2% formalin (to be treated for at least 6 hours) is
recommended.
After a cleaning and disinfection cycle; the first 200 to 250 liters of water or permeate over 1 hour
should be discarded. The pH of permeate should be confirmed before using.