Quick-operating for Insufflator(F102)


1、Front of the Device

① On/Standby key
② LED for On/Standby key (green/
yellow)
③ Nominal pressure setting
④ Nominal pressure display
⑤ Menu key
⑥ Reset key
⑦ Nominal Flow Setting
⑧ LED VERESS mode
⑨LED INSUFFLATION mode
⑩ Connection for insufflation tube
⑪ Nominal gas flow display
⑫ Actual gas flow display
⑬Volume display
⑭ Actual pressure display
⑮ Start/stop LED
⑯ Start/stop key
⑰ LED > 25 bar (green)

2、Operating the Device


2.1 Switching the Device On
1)Connect the gas supply to the gas connection port,Open the gas supply line.
2)Connect the mains connection cable. The device is automatically switched on and performs a self-test for approx. 15
seconds.
3)The mains LED is illuminated green
4)The display shows Device check and all LEDs remain illuminated until the end of the self-test. All segments of the actual
value displays are illuminated.
Three warning signals can be heard.
The displays and indicators depend on the preset/set values in the user menu.
The following information is based on factory settings:
Actual pressure (PRESSURE) ⑭ 0
Actual gas flow (FLOW) ⑫ ---
Nominal pressure (SET mm Hg) ④ 10
Nominal flow (SET l/min) ⑪ 3
Volume ⑬ 0.0
Start/stop LED ⑮ Off
LED VERESS ⑧ ON
LED INSUFFLATION⑨ Off
LED On/Standby ② On (green)
The device test concluded successfully. The device is operational.
2.2 Preselecting the Nominal Pressure and Flow
This setting is possible in standby mode and during operation of the device
1)Before the start of surgery Preselecting Pressure ④(default value) is 10 mmHg, Flow⑪(default value) is 3L/mim,. this
is Veress needle mode(safe mode).
At the beginning of the surgery to press start / stop⑯to start Inflating;
2)when abdominal pressure value on display reaches the setting value, Change Veress needle to Trocar,According to the
requirements of the surgeon to adjust the setting pressure and flow(Setting pressure recommended value 10~15 mmHg;
flow recommended value6~20 L/mim;
3)After the operation have finished, press the start / stop⑯can stop inflating
4) shut down gas press ① for 3 second, can close device.
5)The nominal pressure value "15 mm Hg" is a threshold value.

DANGER!
Exceeding this safety limit is to be decided by and the responsibility of the user/operator.