You are on page 1of 2

FDR D-EVO Mobile Compo

How to Use the Communication Cable kit (Option) Relay


Connector for the Flat Panel Sensor
An optional cable kit is required to use a flat panel sensor (DR-ID 601SE/DR-ID 602SE/DR-ID 611SE/
DR-ID 612SE/DR-ID 613SE) and another flat panel sensor (DR-ID 1201SE/DR-ID 1202SE/DR-ID1211SE/
DR-ID 1212SE/DR-ID 1213SE) together. The connector of the flat panel sensor to be used can be replaced
by connecting/disconnecting the cable kit’s relay connector.
Make certain to read the Operation Manual prior to use of this product.

• Do not connect the flat panel sensor except to the interface unit of the FDR D-EVO Mobile Compo.
• Do not connect the flat panel sensor unregistered to the system. For details on the registration, contact a
FUJIFILM dealer.

Connecting/Disconnecting the Wired Communication Cable of


the Flat Panel Sensor
This section explains how to connect to or disconnect from DR-ID 601SE/DR-ID 602SE/DR-ID 611SE/
DR-ID 612SE/DR-ID 613SE as an example. The same procedure can be applied when connecting to or
disconnecting from DR-ID 1201SE/DR-ID 1202SE/DR-ID 1211SE/DR-ID 1212SE/DR-ID 1213SE.

1 Disconnect the connector. 2 Connect the connector.

Press the latches on both sides of the connector. Press the connector into the insertion section.

Make sure that the latches on both sides are


properly engaged when connecting the connector.
If the connector is inserted incompletely, the power
may turn off.

• Do not connect the wired communication cable to the flat panel sensor, except when switching the flat panel
sensor or making an exposure.
• When multiple flat panel sensors are connected, make sure that the READY lamp among the status lamps of
the flat panel sensor to be used for an exposure is not blinking.
• Connect/Disconnect the connector straight to the flat panel sensor. If connected/disconnected at an angle, the
connector may be damaged.

897N120755
Connecting/Disconnecting the Relay Connector of the
Communication Cable Kit (Optional)
Follow the procedure below to connect/disconnect the communication cable relay connector.

• When connecting/disconnecting the relay connector, always hold the grip of the connector.
The wire inside may be broken, if you connect/disconnect by holding the cable.
If you turn the outer bushing of the grip, the cable lock becomes loose, causing a short-circuit of the cable.
• Do not drop the relay connector when connecting/disconnecting it. Otherwise, personal injury may result, or
properties or the connector may be damaged.
• If the protective cover of the relay connector is damaged, contact a FUJIFILM dealer.

1 Before disconnecting the relay connector 3 To connect, align the positioning marks,
of the flat panel sensor, make sure that the and then push the connectors in.
READY lamp of the flat panel sensor is not
Align the positioning mark on the connector on the
blinking.
interface unit side with that on the flat panel sensor
side, and then insert the connectors by slightly
2 To disconnect, hold the grips A and B of turning them.
both the connectors, and then pull the grip Positioning mark
A of the flat panel sensor to unlock.

Flat panel sensor Interface unit

Green Grip Locking latch Grip Green

Green Grip A Grip B Green Push in until you feel a click.


2mm(0.08 in.) to 3mm(0.12 in.)
CAUTIONS
If you skip Step and perform Step , a
communication error occurs. In such a case, Green Green
turn the power back on to the FDR D-EVO
Mobile Compo. Note, however, that repeating Push further in to the position shown in the figure in
this action may result in damage to the Step until you feel a click again to lock them into
place.
equipment.

897N120755