Welcome to a new era of petroleum testing

Short Operation Instructions


ERASPEC – FTIR Analyzer

Setting up the analyzer Selecting a library


Position the analyzer properly on the bench, connect For calculated properties, like RON, MON, Cetane
the outlet tube and the waste container to the left-hand numbers and distillation temperatures, a data base or
side of the analyzer and connect the power cable to library is necessary.
the power inlet. Different libraries can be created and stored on the
instrument. To select a data base, press “SELECT
Turning on the analyzer LIBRARIES”. The display shows the currently used
Press the main power switch. The analyzer will data base in the right field “used”and other available,
automatically display the measuring menu: but currently unused libraries in the left field
“available”:

Reference spectrum Move the data bases between the fields using the “>>”
A reference spectrum can be measured any time by and “<<”buttons.
pressing the “REFERENCE”button (it turns green) and Press OK to return to the measuring menu.
then RUN.
It is recommended that 10 minutes after switching on, Filling the sample
a reference is measured this way. After that, Connect Luer tip of the yellow filling tube to the inlet on
references are measured automatically when needed. the right side of the analyzer and insert the tube into
the sample.
Selecting a method
Method “Gasoline”is the standard method for gasoline Starting the measurement
testing, and method “Diesel”the standard Diesel Start the measurement by pressing either the RUN
testing method. button on the display or the “RUN”hard key on the
Methods “EN238”for benzene content and “EN14078” front panel (directly below the displayed RUN button).
fro FAME are optional and not active unless As soon as the measurement is started, the screen
specifically ordered and installed. displays the current status of the measurement
sequence (Selecting cell, Filling, Stabilize IR-Source,
Sample ID and Operator name Stabilize Detector, Data Scan). The sample is
Enter the SAMPLE ID and the OERATOR name by automatically introduced into the ERASPEC.
touching the respective white field and entering the When the system is completely filled, please remove
names via the keyboard, or by using the drop down the inlet tube from the sample and confirm on the
lists (touch the red bar left of the respective field). display with “Ok”:
When the measurement is finished, the sample is
removed from the analyzer and disposed of into the
waste container. At the same time, the test results are
displayed (the example below is for gasoline analysis,
the result display for Diesel is similar):

Touch the red number field to the left of the list in the
line of the measurement for which you wan to see the
results. The result list of for this measurement will be
displayed in the same format as directly after a
measurement.
Use the buttons , , , and to move the list
down or up line by line or page by page, and to
return to the first page.
Press OK to return to the result menu.

Adding a sample to the data base


If properties like RON, MON, Cetane number, and
distillation temperatures are known for a sample, it can
be added to the data base. After measuring this
Use the red scroll bar on the right side to scroll through
sample, go to the result menu and mark this sample by
the entire list.
touching the name of the sample. Then press “Add to
Press “OK”to return to the measurement menu or
Lib”. Enter the known properties in the respective line.
“RUN”to immediately start a new measurement.
If a property or several properties are not known, leave
The entire measurement takes approximately one
the filed blank. Press OK to stored this sample in the
minute.
data base.
Press “Add”to add this sample to a data base (see
below in “Adding a sample to the data base”), or press
Import and export libraries using a memory stick
“Print”to print this result on a connected printer.
A memory stick can be connected to one of the USB
connectors of the instrument at any time
Results menu
If a memory stick is connected, the fields “Export to”
The results of previous measurements can be viewed
and “Import from”become active, and the directory on
in the results menu. The results are shown and are
the memory stick that is used for the data transfer is
numbered and ordered by date and time:
shown.
To export the libraries that are stored on the
ERASPEC to the memory stick, press “Export to”.
To import libraries stored on the memory stick to the
www.eralytics.com ERASPEC, press “Import from”.
After the data transfer, the memory stick is unmounted
Partners you can rely on. and can then be safely removed.
The data base is stored with the extension .lib. It is a
tab-separated data file which can be read using Excel.