piezoelecfr~ctrat~sducers& instrumentation for pressure. force.

multi-component force, & acceleratiorr

type 5004 dual mode amplifier


DATA BULLETIN K11.5004 4181

An exceptionally versatile and easyto-operate instrument for


use with either high or Piczotron'" low impedance piezoelectric
transducers, Type 5004 amplifier offers distinctive operational
features, such as:
Twelve ranges, calibrated i n mechanical uniis per volt and
adjustment of the transducer sensitivity
Wide frequency response which can be tailored by a
broad range of low pass bandwidth and notch plug-in
filters
* Unique scale indicator simplifies operation and prevents
improper setting
* Front panel zero adjustment with LEDs indicates any
zero imbalance
LEDs display overload conditiori of amplifier and proper
operation in Piezotron mode
Remote control of Reset-Operate function

OPERATION
Extreme ease of operation characterizes the 5004 which re-
quires no mathematical manipulations.
Simply set the sequence 01 numbers corresponding to the
transducer sensitivity an the 10-turn potentiometer ( 5 ) In DC/MU
or mVlMU (mechanical units e.g.: psi, g, Ib, etc.).
Following this, set the Transducer Sensitivity selector (2)to
the appropriate range.
Then, select the appropriate scale (3)from one of the 12
calibrated positions.
The selected charge amplifier scale in MUIV is displayed in
the window @).
Recording and display equipment connected to the voltage
output should be set for 1Vldiv. The selected amplifier scale
then becomes the display or recorder scale in MUldiv.

DESCRIPTION
Type 5004 Dual Amplifier in the charge mode contains an input
amplifier stage (A) having high input Impedance, high open FOR QUASISTATIC MEASURING, TOO
l o o p gain, and capacitive negative feedback. This dual
MOSFET stage converts the electrostatic charge generated by Contrary t o general impression, a quartz piezoelectric
the transducer into a proportional voltage. transducer, In addition to nieasuring dynamic events, can be
Because of the high open loop gain and the capacitive feed- used also for short duration static (quasistatic) measurements.
back, the voltage at the input of the amplifier remains practi- This is accomplished by operating the amplifier in Long time
cally at zero. The charge yielded by the transducer appears to constant. Measuring time i s dependent largely upon the
flow directly into the feedback (range) capacitor (C) and the transduccr signal strength and the amount of error that can be
resulting output voltage V = Q I C i s proportional to the tolerated. Presupposed is the condition that the input circuit
rneasurand. The amplifier can be reset by short.circuiting the (transducer and cable) has adequate insulation, 10 TO or more,
range capacitor locally (panel switch) or remotely, thereby and that the transducer is not experiencing any temperature
establishing the measuring zero for quasistatic measLrements changes.
(Long time constant). By shunting the feedback capacitor with A rough esiitnale of the possible measuring time can be
a resistance (R), the time constant (Medium or Short) is re- obtained by comparing the charge signal (PC) yielded by the
duced, allotwing dynamic meascrements without drift. trarisducer with t h e maximum specified amplifier d r i f t
Following the input amplifier is a plug-in filter stage (F) and (0.03 pCls). In certain measuring conditions, it is possible to
an operational amplifier output stage (H) with adjustable reset t h e amglifier periodically, thereby eliminating
negative feedback. In effect, settina the transducer sensitivity accumulated effects of drift.
controls the amount of negative feedback and the amplifica-
tion of the output stage.
Input signals that result in an overdrive of arnplifier (H) (out-
put voltage in excess of I 10.5:lare displayed and indicated by FREQUENCY RESPONSE ADAPTABLE
a Flashing Ovef!oad LED. Concurrently a logic signal appears TO SYSTEM REQUIREMENTS
at the rerncte control connector. The Overtoed display i s The frequency responsecf the 5004amplifier can be tailored for
cancelled with the Reset switch or done remotely. system requirements with a wide range of filters which are
Two LEOS indicate the ampllfier Zero level and a i out-of- plugged into the front panei and are either pnssivc (Series
range condition for proper amplifier operation. Any zero drift i n 5311A), or active (Series 5313A) low-pass design with l2dBioct
the amplifier can be corrected with the front panel Zero roll-off. The range of available filters is described in Data
adjustment. Bulletin 12.011. Filters with a Notch characteristic (Series
Integrity of the Piezotron Transducer circuit is indicated by 5312A) are available also for resonance attenuation.
the color sf an LED: red for short circuit, green for normal opera- Model 531 1 filter with a -3dB point at 180 kHr rs supplied as a
tion, and flashing redlgreen for opcn circuit. standard amplifier accessory.