Under normal conditions the diaphragm pump runs for about three minutes and then is stopped for about
seven minutes. The LED2 diode remains lighting while the pump turns. The three ways valve opens (open
way between the two pumps) about five seconds after the diaphragm pump is switched on.
This feature is inactive and the pump keeps turning in one of the two following cases:
Note: The LED2 (on the middle of the board) lights more than the LED1, that’s normal.
As long the LED1 lights, the diaphragm pumps works permanently. Also the LED1 remains light-
ing if BR1 is inserted.
In case of retrofit by an ARL MP1-V3 pump the bridge BR1 must remain inserted.
P1 potentiometer allows to adjust the function threshold. P1 must be adjusted so as the system switches when
the potential of the signal returned from the converter board crosses 40 mV (measured on TP2).
If the LED2 is lighting and the diaphragm pump does not work, check first the fuse. If the board is defective,
the complete board must be replaced (no troubleshooting at component level).
If the LED1 remains permanently lighting, check first if the BR1 bridge is removed. If it is the case, then
pumping line problems (leaks) or molecular drag pump problems must be suspected.
Version 2 Version 3
Number: A037396 A037396
Diagram: S316695 S429079
The board is placed horizontally on the bracket (version 2) or vertically against the panel (version 3). This is
an Alcatel product and is used as a frequency converter to power the molecular drag pump at 27’000 turns
per minutes. The principle diagram as well as the layout are also given in the Molecular Drag Pump manual
that is supplied with the instrument. There are some differences between the Alcatel manual and our appli-
cation, because we use only the converter board and the molecular pump. We do not use the Alcatel box
neither the CVF 100 interface. The elapsed time counter function is also not used.