Professional Documents
Culture Documents
net/publication/232072692
CITATIONS READS
2 4,440
2 authors, including:
John Hiltz
Defence Research and Development Canada, Atlantic Research center
77 PUBLICATIONS 339 CITATIONS
SEE PROFILE
All content following this page was uploaded by John Hiltz on 02 June 2014.
&
DEFENCE DÉFENSE
John A. Hiltz
Sophia Smith
John A. Hiltz
Sophia Smith
Approved by
© Her Majesty the Queen in Right of Canada, as represented by the Minister of National Defence, 2009
© Sa Majesté la Reine (en droit du Canada), telle que représentée par le ministre de la Défense nationale,
2009
Abstract ……..
Fluoroelastomers are synthetic elastomers containing fluorine. They have excellent chemical and
heat resistance and are specified for many applications where chemical and heat resistance is
required. They are among the best materials for use in the marine, automotive, aerospace, oil, and
chemical industries where they are used as seals, O-rings, gaskets, and fuel hoses. In this
memorandum four fluoroelastomers, A Type, B Type, GFLT Type and GLT Type, were
characterized using Fourier transform infrared (IR) spectroscopy, differential scanning
calorimetry (DSC), thermogravimetric analysis (TGA), dynamic mechanical analysis (DMA), and
pyrolysis gas chromatography/mass spectrometry/mass spectrometry (Py-GC/MS/MS). The
objective of this characterization was two-fold; firstly to amass physical and chemical
information on the elastomers and secondly to determine which technique or combination of
techniques could be used to positively identify the four fluoroelastomers.
IR spectroscopy indicated that there are peaks that could be used to differentiate between the four
types of fluoroelastomers. TGA weight loss versus temperature and temperature of maximum
rate of weight loss data was different for the four fluoroelastomers studied and in conjunction
with other experimental data could be used to identify the four elastomers. The glass transition
temperatures (Tg) of the four elastomers measured, using both DSC and DMA, were different.
For new unadulterated elastomer samples, the Tg could be used in conjunction with IR data to
differentiate between the four fluoroelastomers studied in this work. Py-GC/MS/MS could be
used to differentiate between the four fluoroelastomers. Each of the elastomers produced a
distinctive chromatogram (pyrogram). The mass spectra of the compounds giving rise to the
peaks in the pyrograms contained ions that could only be produced by elastomers containing, for
instance, perfluoro(methylvinylether) and therefore have the best potential to positively identify a
fluoroelastomer.
Résumé ….....
Les élastomères fluorés (ou fluoroélastomères) sont des élastomères synthétiques qui contiennent
du fluor. Ils possèdent une excellente résistance à la chaleur et aux produits chimiques et, par
conséquent, leur utilisation est recommandée lorsque ces propriétés sont requises. Ils font partie
des meilleurs matériaux ayant des applications dans les industries navale, automobile,
aérospatiale et pétrolière, ainsi que dans celle des produits chimiques, où ils servent à fabriquer
des joints d’étanchéité, des joints toriques, des joints statiques et des tuyaux de combustible.
Dans le cadre des travaux faisant l’objet du présent document, on a caractérisé quatre
fluoroélastomères, soit ceux des types A, B, GFLT et GLT, par spectroscopie infrarouge à
transformée de Fourier (IRTF), par analyse calorimétrique différentielle (ACD), par analyse
thermogravimétrique (ATG), par analyse mécanique dynamique (AMD) et par pyrolyse
et chromatographie en phase gazeuse/spectrométrie de masse/spectrométrie de
masse (Py-CG/SM/SM). L’objectif des travaux de caractérisation était bipartite, soit une première
étape visant à recueillir des renseignements sur les propriétés physiques et chimiques des
élastomères et une seconde ayant pour but de déterminer quelle technique, ou quelle combinaison
de techniques, permettrait d’identifier avec certitude les quatre fluoroélastomères.
Introduction: Fluoroelastomers are synthetic elastomers containing fluorine. They have excellent
chemical and heat resistance and are specified for many applications where chemical and heat
resistance is required. They are among the best materials for use as seals, O-rings, gaskets, and
fuel hoses in the marine, automotive, aerospace, oil, and chemical industries. Because of the
superior sealing capabilities and heat and chemical resistance of fluoroelastomers, their use in
these industries also helps address environmental concerns arising from chemical leaks and spills
related to elastomer failure.
Fluoroelastomers are differentiated based on the weight percent of fluorine in their make-up and
the method used to cure them (bisphenol or peroxide cure). The performance of a fluoroelastomer
is directly related to its chemical make-up and cure. The higher the fluorine content, the better the
fluid resistance. Peroxide cured fluoroelastomers have superior water and acid resistance
compared to bisphenol cured fluoroelastomers. To ensure that the fluoroelastomer with the
required properties for a particular application is used, techniques to positively identify the
elastomer are required. Several analytical techniques for the identification of fluoroelastomers
are investigated and discussed in this report.
Results: Infrared (IR) spectroscopy indicated that there were peaks that could be used to
differentiate between the four types of fluoroelastomers. Thermogravimetric (TGA) weight loss
versus temperature and temperature of maximum rate of weight loss data was different for the
four fluoroelastomers studied and in conjunction with other experimental data could be used to
identify the four elastomers. The glass transition temperatures (Tg) of the four elastomers
measured, using both differential scanning calorimetry (DSC) and dynamic mechanical analysis
(DMA), were different. For new, unadulterated elastomer samples, the Tg could be used in
conjunction with IR data to differentiate between the four fluoroelastomers studied in this work.
Pyrolysis gas chromatography/mass spectrometry/mass spectrometry (Py-GC/MS/MS) could be
used to differentiate between the four fluoroelastomers. Each of the elastomers produced a
distinctive chromatogram (pyrogram). The mass spectra of the compounds giving rise to the
peaks in the pyrograms contained ions that could only be produced by elastomers containing, for
instance, perfluoro(methylvinylether) and therefore have the best potential to positively identify a
fluoroelastomer.
La différenciation des fluoroélastomères est basée sur leur teneur en fluor, exprimée en
pourcentage massique, et la méthode employée pour les durcir (durcissement au moyen de
bisphénol ou de peroxyde). L’efficacité d’un fluoroélastomère est directement liée à sa
composition chimique et à la méthode de durcissement employée. Plus sa teneur en fluor est
élevée, plus sa résistance aux fluides est bonne. Les fluoroélastomères durcis au moyen de
peroxyde présentent une meilleure résistance à l’eau et aux acides, par rapport à ceux durcis au
moyen de bisphénol. Afin de garantir qu’un fluoroélastomère choisi pour une application
particulière possède les propriétés requises, il faut utiliser des techniques permettant de
l’identifier avec certitude. Le présent rapport contient des détails sur l’étude de plusieurs
techniques d’analyse pouvant servir à identifier des fluoroélastomères, ainsi qu’une discussion
connexe.
Portée : On a déterminé quelles sont les techniques qui permettent de caractériser et d’identifier
avec certitude des fluoroélastomères. Leur utilisation assurera que le fluoroélastomère adéquat est
utilisé pour une application donnée et permettra de réduire, voire d’éliminer, les risques de
défaillance que présente l’utilisation d’un élastomère n’offrant pas des propriétés optimales à ces
fins. Ceci entraînera une réduction des temps d’arrêt causés par les pannes d’instruments et des
coûts d’entretien, ainsi qu’un accroissement de l’état de préparation opérationnelle des dispositifs
et des composants dans lesquels ces élastomères sont utilisés.
Table 1: First weight loss, second weight loss, and temperatures of maximum rate of first and
second weight losses of the four fluoroelastomers...................................................... 10
Table 2: Average glass transition temperatures (Tg) from DSC data with standard deviation...... 13
Table 3: The glass transition temperatures of the four fluoroelastomers measured using DSC
and DMA.................................................................................................................... 16
Table 4: The elemental composition of the inorganic residues from the four fluoroelastomers
following heating in air to 800oC. They were determined by SEM-EDX. ................. 28
Fluoroelastomers are synthetic elastomers containing fluorine. They have excellent chemical and
heat resistance and are specified for many applications where chemical and heat resistance is
required. They are among the best materials for use in the marine, automotive, aerospace, oil, and
chemical industries where they are used as seals, O-rings, gaskets, and fuel hoses [1]. Because of
the superior sealing capabilities and heat and chemical resistance of fluoroelastomers, their use in
these industries also helps address environmental concerns arising from chemical leaks and spills
related to elastomer failure [2].
Fluoroelastomers are differentiated based on the weight percent of fluorine in their make-up and
the method used to cure them (bisphenol or peroxide cure). The performance of a fluoroelastomer
is directly related to its chemical make-up and cure. The higher the fluorine content the better the
fluid resistance2. Peroxide cured fluoroelastomers have superior water and acid resistance
compared to bisphenol cured fluoroelastomers [2].
There are several monomers used to prepare fluoroelastomers: vinylidene fluoride (VF2),
tetrafluoroethylene (TFE), hexafluoropropylene (HFP), and perfluoromethylvinylether (PMVE).
Because of the relationship between the structure and performance of these elastomers, it is
important to ensure that a fluoroelastomer with the required properties is used in a particular
application. Therefore analytical techniques that can positively identify the various
fluoroelastomer types are very important.
The objective of this work is to assess the ability of analytical techniques to distinguish between
different types of fluoroelastomers, specifically FR10/80 A TYPE, FR17/75 B TYPE, FR25/80
GLT TYPE, and LR6316/75 GFLT TYPE. Several techniques were investigated to determine
their usefulness in the characterization and identification of fluoroelastomers. These techniques
were Fourier transform infrared (FT-IR) spectroscopy, thermogravimetric analysis (TGA),
differential scanning calorimetry (DSC), dynamic mechanical analysis (DMA), pyrolysis-gas
chromatography/ mass spectrometry/ mass spectrometry (Py-GC/MS/MS) and scanning electron
microscopy coupled with energy dispersive x-ray analysis (SEM-EDX). In this Technical
Memorandum, the results of the study are reported and the applicability of the analytical
techniques to the positive identification of the four types of fluoroelastomers is discussed.
2.1 Materials
The four fluoroelastomers, FR10/80 A TYPE, FR17/75 B TYPE, FR25/80 GLT TYPE, and
LR6316/75 GFLT TYPE, used in this study were purchased from James Walker MFG. Co.
The B Type fluoroelastomer was a terpolymer of HFP, VF2, and tetrafluoroethylene (TFE). It is a
general purpose fluoroelastomer with improved fuel resistance compared to A Type
fluoroelastomers and can be used in a temperature range between −12°C and 210°C [1].
The GLT Type fluoroelastomer was a tetrapolymer composed of HFP, VF2, TFE, and
perfluoromethylvinylether (PMVE). It is a low temperature (LT) grade fluoroelastomer with
superior rapid gas decompression properties. It can be used in a temperature range between
−41°C and 200°C [1].
The GFLT Type fluoroelastomer was a tetrapolymer composed of HFP, VF2, TFE, and PMVE
and was formulated to have improved low temperature properties compared to A Type and B
Type fluoroelastomers and superior chemical resistance compared to GLT. It is a low temperature
(LT) grade fluoroelastomer with high fluorine content and can be used in a temperature range
between −29°C and 205°C [1].
H F F F
C C C C
H F F F
(VF2) (TFE)
F
F F
F F O C F
C C
F C C C F
F
F F F
(HFP) (PMVE)
All analyses were carried out in the high resolution mode. The temperature was equilibrated at
35°C, and then ramped at 50°C/min until the rate of weight loss exceeded an operator selected
rate. The temperature of the sample was then kept isothermal until the rate of weight dropped
below an operator selected rate. The temperature ramp of 50°C/min was then resumed until the
temperature reached 650°C. At 650 °C, the gas was changed from N2 to air and the temperature
ramp of 50°C/min resumed until the rate of weight loss exceeded the operator determined rate.
The temperature of the sample was then kept isothermal until the rate of weight dropped below an
operator selected rate. The temperature ramp was then resumed. The experiment was terminated
at 800 °C. Three samples of each elastomer were analyzed. The reported percentage weight
losses and temperatures of maximum rate of weight loss were the average values of three
analyses.
Four fluoroelastomers were analysed, A Type, B Type, GFLT Type, and GLT Type. The results
of FT-IR, TGA, DSC, DMA, and Py-GC/MS/MS analysis of the fluoroelastomers and energy
dispersive x-ray (EDX) analysis of the ash following TGA analysis are presented and discussed.
An expanded view of the IR spectra between 2900 cm-1 and 2800 cm-1 is shown in Figure 7. IR
absorbance peaks between 2950 cm-1 and 2800 cm-1 are characteristic of symmetric and
asymmetric C-H stretching. As vinylidene fluoride is the only monomer used in these
fluoroelastomers containing C-H bonds, the strength of the absorbance for a particular
fluoroelastomer should correlate with the vinylidene fluoride content of that elastomer. Typical
values [2] for the percentage of vinylidene fluoride in the four types of elastomers are 60%
vinylidene fluoride for A Type fluoroelastomer, 54% vinylidene fluoride for the GLT Type
fluoroelastomer, 45% vinylidene fluoride for the B Type fluoroelastomer, and 36% vinylidene
fluoride GFLT Type fluoroelastomer. Referring to Figure 7, the A Type fluoroelastomer has the
most intense absorbance and GFLT Type has the least intense absorbance. This corresponds well
with the percentage of vinylidene fluoride in the elastomers.
Figure 6: Overlay of the FT-IR spectra of Type A (blue), Type B (red), Type GFLT (green), and
Type GLT (black) fluoroelastomers.
Figure 7: Overlay of FT-IR spectra of Type A (blue), Type B (red), Type GFLT (green), and Type
GLT (black) fluoroelastomers betweem 2900 cm-1 and 2800 cm-1.
An expanded view of the IR spectra between 1500 cm-1 and 1300 cm-1 is shown in Figure 8.
There is a shoulder on the absorbance peak at 1350 cm-1 in the spectra of the A Type and B Type
fluoroelastomers that is not present in the spectra of the GFLT Type and GLT Type
fluoroelastomers. An expanded view of the spectra between 950 cm-1 and 700 cm-1 is shown in
Figure 9. Comparison of the spectra of the A Type and B Type fluoroelastomers indicates that
the B Type fluoroelastomer has a weak absorbance peak at ~930 cm-1 that is not present in the A
Type fluoroelastomer. In addition, the position of the absorbance peak between 850 cm-1 and 800
cm-1 is different for the A Type and B Type fluoroelastomers. The absorbance peak for the A
Type fluoroelastomer is at ~830 cm-1 and the B Type fluoroelastomer is at ~820 cm-1. The
presence of a shoulder on the absorbance peak at 1350 cm-1 peak and absorbance peaks at ~930
cm-1 and 820 cm-1 are characteristic of a B Type fluoroelastomer, whereas a peak at ~1350 cm-1,
no peak at ~930 cm-1 and an absorbance peak at ~830 cm-1 are characteristic of an A Type
fluoroelastomer.
Although the GFLT Type and GLT Type fluoroelastomers have no absorbance peak at ~ 1350
cm-1, there are subtle differences in their IR spectra between 950 cm-1 and 700 cm-1. For instance,
the GLT fluoroelastomer has a weak absorbance peak at ~830 cm-1 that is not present in the
spectrum of the GFLT Type fluoroelastomer whereas the GFLT Type fluoroelastomer has
absorbance peaks at ~820 cm-1 and ~ 790 cm-1 that are not present in the GLT Type
fluoroelastomer spectrum.
There is some variation in the position of the major absorption peak (between 950 cm-1 and 700
cm-1) in the IR spectra of the four fluoroelastomers. The major peak is at ~888 cm-1 for the B
100
90
80
Transmittance [%]
60 50
40
30 70
Figure 8: Overlay of FT-IR spectra of Type A (blue), Type B (red), Type GFLT (green), and Type
GLT (black) fluoroelastomers between 1500 cm-1 and 1300 cm-1.
100
80
Transmittance [%]
40 20
0 60
Figure 9: Overlay of FT-IR spectra of Type A (blue), Type B (red), Type GFLT (green), and Type
GLT (black) fluoroelastomers between 950 cm-1 and 700 cm-1.
A Type
61.52 ± 0.40 487.91 ± 3.01 29.79 ± 0.49 661.75 ± 4.54 8.32 ± 0.17
B Type
70.02 ± 0.68 501.06 ± 0.99 18.85 ± 0.57 661.43 ± 4.45 10.48 ± 0.12
GFLT
62.09 ± 0.18 504.67 ± 0.57 36.77 ± 0.57 672.94 ± 5.78 0.81 ± 0.23
Type
GLT
59.97 ± 0.08 497.11 ± 0.01 39.39 ± 0.10 677.53 ± 3.42 0.38 ± 0.15
Type
The percent weight loss in the first step was similar for A Type (61.52 ± 0.40%), GFLT Type
(62.09 ± 0.18%), and GLT Type (59.97 ± 0.08%) fluoroelastomers. However, the B Type
fluoroelastomer lost ~70% (70.02 ± 0.68) of its weight in the first step. The temperature of
maximum rate of weight loss, Tmax1, for the first weight loss step was similar for the B Type
(501.06 ± 0.99°C), GFLT Type (504.67 ± 0.57°C), and GLT Type (497.11 ± 0.01°C)
fluoroelastomers while the A Type fluoroelastomer had a Tmax1 (487.91 ± 3.01oC) that was
approximately 10°C lower than the others.
The percent weight loss for the second weight loss step was different for each of the
fluoroelastomers. The A Type fluoroelastomer lost ~30% (29.79 ± 0.49%), the B Type
fluoroelastomer lost ~19% (18.85 ± 0.57%), the GFLT Type fluoroelastomer lost ~ 37% (36.77 ±
0.57%), and the GLT Type fluoroelastomer lost ~40% (39.39 ± 0.10). The temperature of
maximum rate of weight loss for the second weight loss step, Tmax2, was similar for the A Type
(661.75 ± 4.54°C) and B Type (661.43 ± 4.45°C) fluoroelastomers. The Tmax2 for the GFLT Type
(672.94 ± 5.78°C) and GLT Type (677.53 ± 3.42°C) fluoroelastomers were higher than the Tmax2
for the A Type and B Type fluoroelastomers.
100
6
60
665.71°C
2
40
30.16%
0
20 (3.601mg)
Residue:
8.117%
(0.9693mg)
0 -2
0 200 400 600 800
Temperature (°C) Universal V4.5A TA Instruments
Figure 10: Plots of % weight loss vs. temperature and the derivative of % weight change vs.
temperature for Type A fluoroelastomer.
120 10
502.04°C
100 8
70.70%
Weight (%)
(6.443mg)
60 4
40 656.83°C 2
18.29%
20 (1.667mg) 0
Residue:
10.34%
(0.9422mg)
0 -2
0 200 400 600 800
Temperature (°C) Universal V4.5A TA Instruments
Figure 11: Plots of % weight loss vs. temperature and the derivative of % weight change vs.
temperature of Type B fluoroelastomer.
100 8
60 4
679.51°C
40 2
37.03%
20 (5.182mg)0
Residue:
0.8557%
(0.1197mg)
0 -2
0 200 400 600 800
Temperature (°C) Universal V4.5A TA Instruments
Figure 12: Plost of % weight loss vs. temperature and the derivative of % weight change vs.
temperature of Type GFLT fluoroelastomer.
The A Type fluoroelastomer (8.32 ± 0.17 %) and the B Type fluoroelastomer (10.48 ± 0.12 %)
had significantly larger amounts of residue compared to the GFLT Type (0.81 ± 0.23 %) and the
GLT Type (0.38 ± 0.15 %) fluoroelastomers. The larger percentage residue in the A Type and B
Type fluoroelastomers are most likely due to the presence of additives, such as fillers, in these
elastomers.
p Q
120 8
496.87°C
100
6
60
2
681.66°C
40
39.44%
(6.439mg)
0
20
Residue:
0.3076%
(0.05023mg)
0 -2
0 200 400 600 800
Temperature (°C) Universal V4.3A TA Instruments
Figure 13: Plots of % weight loss vs. temperature and the derivative of % weight change vs.
temperature of Type GFLT flurorelastomer.
The Tg values for the four fluoroelastomers are listed in Table 2. The B Type fluoroelastomer has
the highest Tg (−12.74 ± 0.23°C), followed by A Type (−16.79 ± 0.91°C), GFLT Type (−25.03 ±
0.17°C), and GLT Type (−31.29 ± 0.70°C). The Tg correspond well with published values for
these four types of fluoroelastomers [2].
Table 2: Average glass transition temperatures (Tg) from DSC data with standard deviation.
The results of DSC analysis, like those from TGA, provide good supporting evidence for the
identification of polymeric materials. However, the Tg of a polymer is related to its structure and
the presence of additives, and these can be affected by a number of factors while in service.
These factors include the absorption of fluids and aging (thermal, oxidative, and ultra violet
degradation). The absorption of fluids by the polymer most often leads to a reduction in the Tg of
the sample. This is referred to as plasticization. If a material has been in service and in contact
with a fluid or fluids that act to plasticize the sample the measured Tg will be lower than a sample
of the same material that has not been exposed to the fluid. Thermal, oxidative or ultraviolet
degradation can lead to a number of changes in the structure of a polymer. In some polymers
degradation leads to an increase in cross linking in the polymer while it others in make result in a
decrease in cross linking. Additives, such as plasticizers, can also be lost from a polymer over
time. These changes will result in changes to the Tg of the material.
-0.12
-0.14
-21.21°C
-17.81°C(I)
-0.16
-15.07°C
-0.18
-0.20
-60 -40 -20 0 20 40 60
Exo Up Temperature (°C) Universal V4.5A TA Instruments
-0.10
-0.12
-0.14
-15.94°C
Heat Flow (W/g)
-0.16 -13.00°C(I)
-9.69°C
-0.18
-0.20
-0.22
-60 -40 -20 0 20 40 60
Exo Up Temperature (°C) Universal V4.5A TA Instruments
-0.08
-0.10
-27.94°C
Heat Flow (W/g)
-25.02°C(I)
-0.12
-22.17°C
-0.14
-0.16
-0.18
-60 -40 -20 0 20 40 60
Exo Up Temperature (°C) Universal V4.5A TA Instruments
-0.12
-0.14
-33.98°C
Heat Flow (W/g)
-0.16 -31.47°C(I)
-28.86°C
-0.18
-0.20
-0.22
-0.24
-60 -40 -20 0 20 40 60
Exo Up Temperature (°C) Universal V4.3A TA Instruments
The B Type fluoroelastomer had the highest Tg (−10°C) followed by A Type (−15°C), GFLT
Type (−25°C), and GLT Type (−30°C) fluoroelastomers. The glass transition temperature of the
elastomers measured using DSC and DMA are shown in Table 3. Comparison indicates that the
measured values of the Tg for a particular elastomer are not the same using DSC and DMA.
However, the order of the Tg, from highest to lowest, is the same.
Table 3: The glass transition temperatures of the four fluoroelastomers measured using DSC and
DMA.
The results of DMA, like those from TGA and DSC analysis, provide good supporting evidence
for the identification of polymeric materials. However, as was noted for DSC, the Tg of a
polymeric is related to its structure and is affected by a number of factors. These factors will
affect the measured Tg.
800
6000
600
4000
400
2000
200
0 0
-60 -40 -20 0 20
Temperature (°C) Universal V4.3A TA Instruments
Figure 18: Plot of storage modulus and loss modulus vs. temperature for A Type fluoroelastomer
at 1 Hz.
10000 1000
-9.99°C
8000 800
6000 600
4000 400
2000 200
0 0
-60 -40 -20 0 20
Temperature (°C) Universal V4.3A TA Instruments
Figure 19: Plot of storage modulus and loss modulus vs. temperature for B Type fluoroelastomer
at 1 Hz.
1000
6000
4000 600
400
2000
200
0 0
-60 -40 -20 0 20
Temperature (°C) Universal V4.3A TA Instruments
Figure 20: Plot of storage modulus and loss modulus vs. temperature for GFLT Type
fluoroelastomer at 1 Hz.
8000 1400
-29.97°C
1200
6000
1000
800
4000
600
400
2000
200
0 0
-60 -40 -20 0 20
Temperature (°C) Universal V4.3A TA Instruments
Figure 21: Plot of storage modulus and loss modulus vs. temperature for GLT Type
fluoroelastomer at 1 Hz.
The pyrograms of A Type fluoroelastomer following pyrolysis at 900°C, 800°C, and 700°C are
shown in Figures 22 and 23. Figure 22 shows the chromatograms from 0.0 to 40.0 min while
Figure 23 shows the chromatograms between 3 and 15 min. The pyrograms for the B Type,
GFLT Type, and GLT Type fluoroelastomers following pyrolysis at 900°C, 800°C, and 700°C
are shown in Figures 24 and 25, Figures 26 and 27 and Figures 28 and 29 respectively. Pyrolysis
was carried out at three temperatures to determine if a change in the pyrolysis temperature would
affect either the number or intensity of the degradation products. Comparison of the pyrograms
for a particular elastomer indicated that the thermal degradation products were little affected by
the pyrolysis temperature. On the basis of this, it was decided to use a pyrolysis temperature of
700°C.
Figure 22: Pyrograms of A Type fluoroelastomer pyrolyzed at 900oC (top), 800oC (middle), and
700oC (bottom).
Figure 24: Pyrograms of B Type fluoroelastomer from 0-40 minutes pyrolyzed at 900oC (top),
800oC, and 700oC (bottom).
Figure 26: Pyrograms of GFLT Type fluoroelastomer from 0-40 minutes pyrolyzed at 900oC
(top), 800oC, and 700oC (bottom).
Figure 28: Pyrograms of GLT Type fluoroelastomer from 0-40 minutes pyrolyzed at 900oC (top),
800oC, and 700oC (bottom).
100 5.51E7
TIC MS
6.90 viton_b_pyro7
00-run8
50
3.53 3.98 5.73 7.39 7.95 9.53
10.38 11.61 12.78 14.10
0
6.92 7.36 NL:
100 3.23E6
TIC MS
viton_gflt_pyr
4.03 7.97 8.74
5.49 o700-run10
50
5.42 6.02 9.10 9.60 12.00
10.44 13.04 14.66
0
9.09 NL:
100 3.75E7
8.73
14.27 TIC MS
11.02
vtion_GLT_py
12.21 13.66 ro700_run5
50
9.68 9.99
3.98 4.63 7.36 7.69
6.43
0
4 6 8 10 12 14
Time (min)
Figure 30: Pyrograms of A Type (blue), B Type (red), GFLT Type (green), and GLT Type (black)
fluoroelastomers between 3.5 and 15 min. All samples were pyrolyzed at 700oC.
The mass spectra of the compounds giving rise to the most intense peaks in the pyrograms of A
Type, B Type, GFLT Type and GLT Type fluoroelastomers are shown in Figures 31 through
Figure 34 respectively.
100
viton_a_pyro700_run6
#1267 RT: 10.97 AV:
1 T: + c EI Q1MS
50 [50.000-500.000]
113
109 207 227 271
335 361 381 431 476
0
163 NL: 2.15E7
100
viton_a_pyro700_run6
#1346 RT: 11.66 AV:
1 T: + c EI Q1MS
50 [50.000-500.000]
113 133 165
77 207 229 288
335 379 399 431 451
0
163 NL: 4.70E7
viton_a_pyro700_run6
40000000
#1628 RT: 14.11 AV:
30000000 1 T: + c EI Q1MS
[50.000-500.000]
20000000
113 133 165
10000000 77 197 229
293 337 352 399 443 475
0
100 200 300 400 500
m/z
Figure 31: Mass spectra of peaks (from top to bottom) at 8.73 min, 10.97 min, 11.66 min and
14.11 min in the pyrogram of A Type fluoroelastomer.
20 165
77 197 229 263
327 371 391 459 485
0
113 NL: 1.54E7
100
viton_b_pyro700-
Relative Abundance
run9#798 RT:
80
6.90 AV: 1 T: + c
EI Q1MS
60
[50.000-500.000]
40
133
20 69 77 165 213 257 277 321
353 392 419 482
0
163 NL: 1.70E7
viton_b_pyro700-
15000000 run9#1009 RT:
113
8.73 AV: 1 T: + c
EI Q1MS
10000000 [50.000-500.000]
5000000 133
77 165 224
209 259 317 337 387 407 471 497
0
100 200 300 400 500
m/z
Figure 32: Mass spectra of peaks (from top to bottom) at 6.76 min, 6.90 min, and 8.73 min in the
pyrogram of B Type fluoroelastomer.
The major ions in the mass spectra of the three degradation products of the B Type
fluoroelastomer (Figure 32) vary. Ions at m/z 163, 113, and 133 are prominent for the
degradation products at 6.76 min and 8,43 min. However, the mass spectrum of the degradation
product at 6.90 min has no ion at m/z 163 but has an ion ion at m/z 165. Differences in the
structure and intensity of the degradations products produced by the B Type fluoroelastomer
compared to the A Type fluoroelastomer result from the difference in composition of the two
elastomers. The A Type fluoroelastomer is a copolymer of VF2 and HFP while the B Type
fluoroelastomer is a terpolymer of VF2, HFP and TFE. As was observed for the mass spectra of
the degradation products of the A Type fluoroelastomer, the mass spectra of the B Type
100
viton_GFLT_pyro700-
91 run8#849 RT: 7.35
AV: 1 T: + c EI Q1MS
50 113 [50.000-500.000]
Figure 33: Mass spectra of peaks (from top to bottom) at 6.91 min, 7.35 min, 7.96 min and 8.73
min in the pyrogram of GFLT Type fluoroelastomer.
The mass spectra (and most intense ions) of the major degradation products of the GFLT Type
fluoroelastomer (Figure 33) indicate that some of degradation products of this elastomer are the
same as found for the A Type and B Type fluoroelastomers. For instance, the mass spectrum of
the compound giving rise to the peak at 6.91 min is the same as that for the compound giving rise
to the peak at 6.90 min in the B Type fluoroelastomer pyrogram. The mass spectrum he
compound giving rise to the peak at 8.73 min is the same as that for the compound giving rise to
the peak 8.73 min in the A Type and B Type fluoroelastomer pyrograms.
The mass spectra of three of the four degradation products of GLT Type fluoroelastomer (shown
in Figure 33) also have an ion at m/z 85. The GFLT Type fluoroelastomer is made using PMVE
100
vtion_GLT_pyro700_ru
n5#1051 RT: 9.09 AV:
69 91 1 T: + c EI Q1MS
50 [50.000-500.000]
113 133
212 307
276
336 371 403 422 466
0
113 NL: 6.71E6
100
vtion_GLT_pyro700_ru
n5#1274 RT: 11.02
AV: 1 T: + c EI Q1MS
50 [50.000-500.000]
133
69 85 179 223 285 311 350 387 432 449 498
0
113 NL: 5.29E6
vtion_GLT_pyro700_ru
4000000 n5#1648 RT: 14.27
AV: 1 T: + c EI Q1MS
[50.000-500.000]
2000000 133
165
77 197 229 285 337 357 434 464 491
0
100 200 300 400 500
m/z
Figure 34: Mass spectra of peaks (from top to bottom) at 8.73 min, 9.09 min, 11.02 min and
14.27 min in the pyrogram of GLT Type fluoroelastomer.
The mass spectra of the four major degradation products of the GLT Type fluoroelastomer are
shown in Figure 34. Ions at m/z 85, m/z 91 and m/z 179 are present in the mass spectra and can
be attributed to degradation products from the portion of this elastomer containing PVME.
Although there are some similarities in the ions found in the mass spectra of the major
degradation products of the GFLT Type and GLT Type fluoroelastomers, there are a number of
differences that result from the difference in the composition of the two elastomers. These are
sufficient to differentiate the two elastomers.
Table 4: The elemental composition of the inorganic residues from the four fluoroelastomers
following heating in air to 800oC. They were determined by SEM-EDX.
The residue of the GFLT Type fluoroelastomer contained Ba, Ca, Al, zinc (Zn), iron (Fe), S, and
Mg. Ba was the major constituent and there were trace amounts of the other elements present
relative to Ba. The residue of the GLT Type fluoroelastomer had a slightly different composition
than that of the GFLT Type fluoroelastomer. Al was the major constituent with smaller amounts
of Zn, Ca, Fe, copper (Cu), Mg, Si, chromium (Cr), potassium (K), titanium (Ti), nickel (Ni),
chlorine (Cl), and S present.
Residue elemental analysis data is of little use in the identification of a particular type of
fluoroelastomer. However, if one wishes to confirm that a fluoroelastomer is the same as another
and a sample of the other is available, elemental residue analysis data is very useful. The nature
inorganic residue can provide information on additives, such as fillers, curing agents, uv
stabilizers and antioxidants, used in the preparation of the elastomer. If the constituents of the
inorganic residue are the same and their relative concentrations are similar then the two
elastomers are most likely the same.
Four fluoroelastomers have been characterized using FT-IR, TGA, DSC, DMA and Py-
GC/MS/MS. FT-IR provided information that could be used to differentiate the four
fluoroelastomers. Absorbance due to symmetric and asymmetric C-H bonding (2800cm-1 - 2950
cm-1) in the fluoroelastomers is indicative of the VF2 in the elastomer formulation and the strength
of the absorbance was found to correlate with the published percentages of VF2 in each of the
elastomers. Absorbance peaks at ~1350 cm-1, ~930 cm-1, and ~825 cm-1 can also be used to
differentiate between the elastomers.
Weight loss versus temperature plots and temperatures of maximum rate of weight loss,
determined using TGA, were different for the four fluoroelastomers. This information, in
conjunction with FT-IR or py-GC/MS/MS data, and a Tg from DSC or dynamic mechanical
analysis, would allow positive identification of the fluoroelastomers. The Tg of the four
fluoroelastomers, determined using either DSC or DMA, were different. As was stated for the
TGA results, this information in conjunction with FT-IR or Py-GC/MS/MS data would allow
positive identification of these fluoroelastomers.
Each of the fluoroelastomers had a unique pyrogram. This indicates that Py-GC/MS/MS can be
used to identify the fluoroelastomers studied in this report. The gas chromatographic separation
of pyrolytic degradation products and subsequent identification of the degradation products using
mass spectrometry are advantages of this technique compared to the others investigated in this
report. Although FT-IR can be used to identify the fluoroelastomer studied in this Technical
Memorandum, these fluoroelastomers had not been in service and therefore were not adulterated.
The presence of a fluid in a sample could mask some of the IR peaks used to identify an
elastomer. Adulteration could also affect the Tg of a sample. Although the presence of a
contaminant, such as oil, would affect the pyrogram of a sample, the pyrolytic degradation
products of the contaminant would be separated from the degradation products of an elastomer
prior to mass spectral detection. In addition, the mass spectra of the individual components
would enable identification of products arising from the elastomer or the adulterant.
[1] “James Walker: Sealing Products & Materials for the Oil & Gas Industry”. Retrieved May14
2009 from:
http://www.jameswalker.biz/docs/jwco/Oil_and_gas_guide.pdf
[2] Albert L. Moore, Fluoroelastomers Handbook: The definitive user’s guide and databook,
William Andrew Inc., New York (2006).
[3] CWC- “Fillers for rubber compounds”. Retrieved July 31 2009 from:
http://www.p2pays.org/ref/13/12534.pdf
[4] R. M. Paroli, and J. Penn, “Measuring the Glass Transition Temperature of EPDM
Roofing Materials: Comparison of DMA, TMA, and DSC Techniques”, Assignment of
the Glass Transition, ASTM STP 1249, R. J. Seyler, Ed., American Society for Testing and
Materials, Philadelphia, 1994, pp. 269-276.
1 Library and Archives Canada, Attn: Military Archivist, Government Records Branch
1 NDHQ/DRDC/DRDKIM3
Technical Memorandum
8. SPONSORING ACTIVITY (The name of the department project office or laboratory sponsoring the research and development – include address.)
10a. ORIGINATOR'S DOCUMENT NUMBER (The official document 10b. OTHER DOCUMENT NO(s). (Any other numbers which may be
number by which the document is identified by the originating assigned this document either by the originator or by the sponsor.)
activity. This number must be unique to this document.)
11. DOCUMENT AVAILABILITY (Any limitations on further dissemination of the document, other than those imposed by security classification.)
Unlimited
12. DOCUMENT ANNOUNCEMENT (Any limitation to the bibliographic announcement of this document. This will normally correspond to the
Document Availability (11). However, where further distribution (beyond the audience specified in (11) is possible, a wider announcement
audience may be selected.))
Unlimited
13. ABSTRACT (A brief and factual summary of the document. It may also appear elsewhere in the body of the document itself. It is highly desirable
that the abstract of classified documents be unclassified. Each paragraph of the abstract shall begin with an indication of the security classification of the
information in the paragraph (unless the document itself is unclassified) represented as (S), (C), (R), or (U). It is not necessary to include here abstracts in
both official languages unless the text is bilingual.)
Fluoroelastomers are synthetic elastomers containing fluorine. They have excellent chemical and
heat resistance and are specified for many applications where chemical and heat resistance is
required. They are among the best materials for use in the marine, automotive, aerospace, oil and
chemical industries where they are used as seals, O-rings, gaskets, and fuel hoses. In this
memorandum four fluoroelastomers, A Type, B Type, GFLT Type and GLT Type, were
characterized using Fourier transform infrared (IR) spectroscopy, differential scanning
calorimetry (DSC), thermogravimetric analysis (TGA), dynamic mechanical analysis (DMA),
and pyrolysis gas chromatography/mass spectrometry/mass spectrometry (Py-GC/MS/MS). The
objective of this characterization was two-fold; firstly to amass physical and chemical
information on the elastomers and secondly to determine which technique or combination of
techniques could be used to positively identify the four fluoroelastomers.
IR spectroscopy indicated that there are peaks that could be used to differentiate between the four
types of fluoroelastomers. TGA weight loss versus temperature and temperature of maximum
rate of weight loss data was different for the four fluoroelastomers studied and in conjunction
with other experimental data could be used to identify the four elastomers. The glass transition
temperatures (Tg) of the four elastomers measured, using both DSC and DMA, were different.
For new unadulterated elastomer samples, the Tg could be used in conjunction with IR data to
differentiate between the four fluoroelastomers studied in this work. Py-GC/MS/MS could be
used to differentiate between the four fluoroelastomers. Each of the elastomers produced a
distinctive chromatogram (pyrogram). The mass spectra of the compounds giving rise to the
peaks in the pyrograms contained ions that could only be produced by elastomers containing, for
instance, perfluoro(methylvinylether) and therefore have the best potential to positively identify a
fluoroelastomer.
14. KEYWORDS, DESCRIPTORS or IDENTIFIERS (Technically meaningful terms or short phrases that characterize a document and could be
helpful in cataloguing the document. They should be selected so that no security classification is required. Identifiers, such as equipment model
designation, trade name, military project code name, geographic location may also be included. If possible keywords should be selected from a
published thesaurus, e.g. Thesaurus of Engineering and Scientific Terms (TEST) and that thesaurus identified. If it is not possible to select
indexing terms which are Unclassified, the classification of each should be indicated as with the title.)