— @PICKER ‘A $C Company PICKER INTERNATIONAL LIMITED P.O BoxNe.2, East Lane, Ieove 2 Midéleser, England. HAS 7PR shtegan, won. SADEY A988: Installation, Commissioning and Service R500 Series 2 INSTRUCTION 2450 While every care is taken to ensure the accuracy of the description and illustrations in this publication we rescrve the right to ake any modifications in tho design of the apparatus concidered to be in the Interest of our customers. Reproduction of this docuneat or any part of it without prior written permission from PICK#K INTERNATIONAL LED ia prohibited. (92924) Instruction No 2450 Wendie, R900 SERIES 2 CONTROL CONSOLE (2450/1/ 385) RADIATION HAZARDS 6 SAFETY PRECAUTIONS It is dangerous for any person co operate this equipment without having appropriate training which will have included instruction in the means For using X-radfacfon without hazard to patient, user and surroundings. The purpose of thase instructions is to inform the user about the technical functioning of the equipment for the intended purpose(s); they do not cover any aspects of radiation protection or other aspects of safety relating ta the application of the equipment. The user must be avare of all regulations and requirenents that may be applicable governing the installation and use of equiprent producing fonieing radiation for medical purposes. BLECTRIC SHOCK HAZARD Do not remove any high voltage cable connection to an X-ray tube. Note also that such a cable aay retain an electric charge or be connected to other components retaining a charge after the equipment has been switched off. Do not renove any covers or panels giving access to live parts. Any cover requiring the use of a tool for its reroval can be assuned to be in this category. EXPLOSION HAZARD ‘This equipment is not classified as anaesthetic-proof and may ignite flamable anaesthetics. Flanmasle agents used for skin cleaning or disinfection may 21so produce an explosion hazard. HEALTH & SAFETY AT WORK ACT 1974 (UK installations only) ALL equipment manufactured and supplied by this Company has been tested and examined to ensure as far as is reasonably practicable, that it is safe and without undue risk to health when properly used. The conditions under which our equipment will operate safely and without undue risk to health are specified in our Operating Instructions and users should ensure that they fully understand the technical conditions rogerding cafe operating of the equipment and are conversant with and observe Regulations and Codes of Practice which relate to X-ray Equipment. It is also the duty of the enployer to ensure that his employees fully understand the Regulations and Operating Instructions. (2450/ 1/385) ALi/iv conrENts. Paragraph Radiation hazards and safety precautions General requirenents 1 Specification Ll output 1.2 Power Supply 13 Maing voltage compensacton 14 kW range 8 mA range 16 Timer cange M7 Repetition rate 8 Xcray cubes i) Gecomat 2 . 1:L0 Displaye [11 Interfacing . 2 Preliminary inetallation work 2 Checking pre-installat ton work 3 Inventory 3.1 Kit supplied with DaSS orders 4 Installation Preliminaries HT Transformer of1 Level Removing and refitting covers 1 Control eonsole +2 WT Transformer Connecting the control console Connecting the HT transformer +1 UT cables Preparatory work 1 Mains supply taps 2 Voltage and frequency taps 3 X-ray over-pressure thermal cut-out terminals 44 Xoray tube stator connections 23. Bucky connections +6 Tube exposure counters R500 Series 2 torminal etrip connections Common interface connections Commissioning the R500 Series 2 Preliminaries jectrical safety checks +1 Earth continuity resistance tests 22 Earth insulation resistance cescs Setting up the tube overload circuits +1 Preliminary work £or tube type X -2 Setting up tube type X overload parameters L 2 2 2. 3 220Y Line voltage adjustment 4 e~focus, high-speed ©.4.3 Setting up tube type X broad-focus, high-speed overload paranaters S.te4 — Setcing up tube type X low speed overload parameters (2450/1/385) Page aah 10 2 14 14 14 La 15 1s 5 7 19 21 23 23 23 23 23 23 27 27 33 3 vi CONTENTS (Continued) Page Tube type X without associated high-speed facility 35 Setting up tubes type ¥ and Z 35 Programming the overload board 36 WW meter calibration to HT primary voltage 28 Fluorosco py 38 Radiography 38 Adjuatnent of tube ad 39 Broad focus pre-heat adjustments for remaining tubes 40 Fluoro m4 adjustment 4L Fine focus pre-heat adjustuents for remaining tubes 41 Fluoro m4 capacity conpenss tion 41 Radiography exposure timer check 42 Radiography mA adjustments, non mammo 42 Space charge compensation 43 Mamnography WA adjustments (parallel operation) 46 Pinel cal thration of kV meter to line resistance 45 Method 1, with kV measuring device . 45 Nethod 2, without kV measuring device 45 Flucro k¥, with kV measuring device 46 Coumisstoning the Gecomat 2 Pe General 49 ‘Automate kV compensation controls ia Film density cailbration, Tube 1 50 Film density calibration, Tube 2 54 7 Final tests and handing over 57 ra Final tests 57 72 Hending over 37 8 Relay, contactor and wiereswiteh functions 59 8.1 Relays 59 8.2 Contactors . 67 8.3 Microswitches, rotary switches and pushbutton switches 72 9 Technical description on 10 Servicing Appendix 1 R500 Series 2 Interfacing connections Appendix 2 R500 Series 2 Circuit diagrams Appendix 3 R500 Series 2 Recommended spares ‘Supplement Section Always refer to the Supplement Section before conmencing installation and/or service work. (2450/2/ 0888) Figure Frontispiece (2450/1/385) TLLUSTRATIONS R500 Series 2 control console Gontrel coascle ~ control panel NY teanstormer top and connections Control console = covers Control console ~ terminal strips Control console ~ front internal view R500 series 2 wall box Control Console - rear internal view ka/time overload pcb ki/time overload test board kV adjust pob Xoray tube ratings ~ nigh speed (150 fz) R5U0 Sories 2 maximum rating chart Xcray tube ratings = low spead (50 Ha) Gecomat control unit ~ installation information Gecomat control unit ~ interconnections Gecomat electronics module-board U6 Gecomat interface circuit diagram Page a 4 13 16 18 25 26 28 3 32 37 a7 51 55 vii Will GENERAL REQUIREMENTS PRE-INSTALLATION INSTRUCTIONS The following iastruetions sre to be read in conjunction with the Layout drawing prepared by PICKER INTERNATIONAL LIMITED. GENERAL ELECTRICAL REQUIREMENTS ALL elacteical wiring and interconnestion must be strictly in accordance Wich che recommendacions contained in 'The regulations for the Electrical Equipment of Boildings' published by the Institute of Electrical Snginears or other appropriate national wiring regulations. Special care must be taken to ensure that all conduits are properly bonded and connected to earth. The following table shows the sizes of cables to be uaed for interconnecting equipment. Cross Metric Sizes Inporial Sizes Carrent Sectional Rating, area am Stranded | Flexible | Stranded | Flexible A 1.0 32/0420 401.0076 10 3/76 3/,02910 nominal | 30/0.25 15 3/.036 | 70/.0076 15 77,028, | 110/.0076 20 50/0430 20 7/0685 | 56/0430 25 1/036 | 162/.0076 25 105/,010 30 7jL.0% 30 7/044 36 2f1,35 7/4052 42 7170 7/064 33 19/ .044 62 7/2614 19/.052 B 19/1.53 90 19/064 91 19/1,78 145 19/083 160 19/244 LBs. ‘he metric sizes quoted are nominal, some variations are wade by suppliers. Note: Solid conductors are NOT suitable. Stranded or flexible cables MUST be used. (2450/1/385) ix ALL metal parts of items such as tables, tubestands and tracks must be securely earthad at the point indicated in the Pre-installation Schedule and layout drawiags, then returned to the main earth point at the main isolator in the X-ray room, A stranded conductor of suftable erossvsectional area may be used or a copper strip of cross-section not less than lomn x 2.5nm. A tail of 0.5 metre should be lett for connection by the Installation Engineer. The resistance between any earthed wetal part and the main earch point mist not excced 0. lohn as specified in paragraph 7 of DHSS General Technical Clauses 1973. All leads must be enclosed in steel conduit or trunking. Normally all connections to terminals will be made by PICKER Equi pment Engineers. GENERAL STRUCTURAL REQUIREMENTS Fach installation must be investigated by an Architett or Surveyor to determine the size of gitders and other bearers required to support the Weight of equipments installed. PREPARATION OF CEILING Gelling tracks can be fixed directly to the ceiling but, normally, « board is required which must be of hardwood nor less chan 40mm chick. For widen, lengch and positions, refer to layout drawing. Ceiling boards ust be level and capable of supporting the moving load stated in the relevant pre-installation instruction and mst be firmly bolted to the cefling or other fixings. Aecess to tho upper side of the board should be possible for the Installation Enginear co bolt the tracks in position. CEILING MOUNTED TUBE SUPPORTS The lateral movement of ceiling tracks mist not be greater then 3nmj if necessary f1t cross~aembers between girders or use some other nethod to prevent excessive lateral novenent. FLOOR TO CEILING MOUNTED TUBE SUPPORTS - MOBILE The ceiling board and floor track must be plumb level to within a tolerance of 3mm in che position #howa on the layout drawing. FLOOR TO CEILING/WALL MOUNTED TUBE SUPPORTS & BUCKY STANDS Provide and fit a hardwood fixing board to the ceiling or wall. The board should be a minimum size of 230m x 230um x 40mm. Fixings for the board mist not coincide with the f1xings on the extension tube plate or ceiling tracks. for the maxinum loading on the extension tube plate Fixings, refer to the pre=installacion instructions. 5 (2450/1/385) CBILING SUPPORT FOR HT CABLES A eetLing board SOmm wide {s normally required to support the Capitol Track and Fittings, together with the KT cables. for length and position refer to the layout drawing. PREPARATION OF FLOOR A level floor surface must be provided for all equipments, No part of the floor area concerned may be more than Imm from the nominal floor level. ‘The areas aver which the equipments are to he fixed mst be checked with che aid of a straight edge, and any high spots renoved uneil che areas are within tolerance. If che floor eannot be improved or is coo weak, bearers of the type as shown in the pre-installation instruction should be Let into tho floor. The upper surface of the bearers must be level in all directions and the bearink surface aust not be lower than the highest point of the adjacent floor. In rooms where a new concrete or composition floor is being laid, built in bearers may be considered. FLOOR TRACKS: When a flush floor track is to be used a suitable channel must be made in the floor, or a channelled bearer provided. (See pre-installation schedules). (2450/1/385) xi/aii qa NOTE 13 1 15 1 SPECLFICATION Output 37.5 kW (TEC Rating) 2 pulse Max mA ~ 500mA at 100kV on a line resistance of less than 0.3 ohm referred to 415v ac Max KV = 1258V at 300m Power Supply Nomina 415¥, 380V and 220V, single phase, 50/60H2 1. There ds reduced output performance with a 2207 power eupply- 2 The R500 Series 2 can be used on a 60He oupply after fectory sodifications have been cecried out. Tt is important, therefore, tha the correct supply frequency ie apecified at the time of ordering. Line-to-line resistance is not to exceed 0.33ohm at 415V, 0.3ohm at 380V and 0.150hm at 220V, Mains Voltage Compensation Manual adjustment range aveilabl Nominal 415 Adjustment range is 3674950 Nominal 380¥ Adjustment range 1s 332-4207 Nominal 2207 Adjustment range Lo 172-2607 KY Range Radiography: Fine and coarse ‘flovoltage controls with adjustment between 45k¥p to 125kVp in steps of approximately 1kV. Fluoroscopy: 60~110kV, contiauously adjustable. mA Range Broad Foes: 300aA at 125i 400mA 2 125%V 500mA at 100K Fine Focus: 50, 100, 200mA at 125kv Fluoroscopy: 0.5 ~ Sma, continuously adjustable ° (2450/2/0485) 1 qa 1.2 NOTE 13 14 15 1 SPECIFICATION Output 37.5 kW (IEC Rating) 2 pulse Max mA - 500mA at 100kV on a lino reaietance of less than 0.3 ohm referred to 415V ac Max kV - 125kV ar 300mA Power Supply Nominal 415V, 380V and 220V, single phase, 50/60Kz 1, There is reduced output performance with a 220¥ power supply. 2 The R500 Series 2 can be used on a 60Hz supply after factory modifications have been carried out. It is important, therefore, that the correct supply frequency is specified at the time of ordering. Lame-to-line resistance is not to exceed 0.33chm at 415V, 0.Johm at 380V and 0.15ohm at 220V. Mains Voltage Compensation Manual adjustment range available: Nominal 415v Adjustment range 1s 367-655 Nomina 380V Adjustoent range is 332-4200 Nominal 220¥ Adjustment range 1s 172-260V AV Range Radiography: Fine and coarse kiloveltage controle with adjustment between 45kVp to 125kVp in steps of approximately 1kV. Fluoroscopy: 60-110, continuously adjustable. mA Range Broad Foous: 300mA at 125V 400mA at 125kY 500mA at 1OCKY Fine Focus: 50, 100, 200A at 125KV Fluoroscopy 0.5 ~ Sma, continuously adjustable (2450/2/0485) 1 47 1.8 1.8 1.10 Timer Range Rediographyt 0.01 - 5 seconds 24 position rotary switch selector Fluoroscopy: 0-10 minutes with buzzer alara during last Se prior to automatic termination at zero. Repetition Rate Up to 6 exposures per second X-ray Tubes Sockets are fitted for the connection of 4 dovble-focus X-ray tubes, the fourth tube being for mammography only.. Automatic overload protection for a total of 6 different foci ts provided. Gecomat 2 If @ Gecomat is incorporated, automatic kV compensation is applied at GOKV, BOKV and 100KV. Displays. Figure 1 is a view of the control panel. Two meters are provided: L Voltmeter. This indicates the selected kV. During standby it indicates the kV preselected for radiography. 2 Combined voltmeter and mA meter. This hae two scales: 0-10mA for fluoroscopy; 0-1000mA for radiography. During standby, a red mark at the centre of the scale indicates the correct line voltage. This ie set by rotating the control at the left-hand side of the control console casing. The {1luminated display panel indicates the following: Pre-selected radiographic mA Technique selected . Tube focus selected Ready to expose Tube overload High speed stator selection (1f applicable) X-rays on Exposure time (2450/2/0486) (2430/1/385) FIG.1 CONTROL CONSOLE-CONTROL PANEL 1.11 Interfacing Facilities are provided for direct connection of the following equipment: PICKER X-ray Tables and Chest Bucky Stands (modification is required for standard sectograph) Apollo Ceiling Tubemount K-S80 ceries of Image IntensLfiere and TV Systena HSIS0 Frequency Converter Geconat 2 . CAUTION tnscallacion of equipment and accessories nét listed above must not be attenpted vithout prior consultation with Picker Internationsl Utd. Kembley. 112 Associated Documents Always refer to the relevent installation/service sanual of accessories to assist. 4 (2450/1/385) 2 PRELIMINARY INSTALLATION WORK 2.1 Checking Pre-Installation Work Ensure that che cequired pre-iastallatioa work has been completed In aceordance with the pre=tnstallations schedules and layout drawings provided. (2450/1/385) a FIG.2 HT TRANSFORNER TOP AND CONNECTIONS (2450/2/0486) 3 INVENTORY Vguipnent ts supplied as follows BOK 1 Coneatatng R500 Series 2 Control Console BOX 2 Containing 2500 Series 2 HT Transformer BOX 3 Containing the following itens: 1 off Wall Junction Box prewtred to 2 flexible wieing conduics for connection to the Control Console 1 off Wall Junction Box identical to the above for connection to the KT Transformer® 1 off Flexible wiring conduit to connect the Wall Junction Box to the ET Transformer 1 Kit Supplied with DHSS orders DHSS installations will be supplied vith an additional kit Stock No.M1286-517 containing the following 1cens: 4 off Exposure Counters X610-905 25 off Terminals Crimp Spade 4 BA 6233-201 1 off Wall Box containing: Mo22049 ‘Tube selected Lights and Auxiltartes Panel Room lights transformer ‘If any item is missing or damaged, inform Picker International and the carriers inmediately. (2450/1/385) 7 FLG.3 CONTROL CONSOLE - COVERS (2450/1/385) 4 INSTALLATION Preliminaries Ensure that the required pre-installation work has beea completed in accordance with tho pre~inctallation schedules and layout drawings provided. Circuit diagrams for the R500 Series 2 are given in Appendix 2. HT Transformer O11 Level Remove the HT transformer cover (see para. 4.3.2). Romove the oil-filler cover (see Fig. 2) from the iT transformer. CAUTION: Conpiete the following steps as soon as possible to prevent NOTE e contamination of the ofl. Cheek that the cil level is between 25am (lin) and 38am (1in) from the oll filler cop. Use a clean implement such as a screndriver as a dipstick. 1, The levels specified are at normal anbleat tenperature of 20°C. 2. Export units may heve 23 litre (5 gallons) of oil despatched separately. Transfer the oil to the HT transformer using a funnel with a fine gauze to filter out any impurities. Add the oil slowly, directing it so that the aininum nusber of air bubbles are formed. 3. Only use DIALA grade B Shell or equivalent ofl. If the ofl level is low, add oil as detailed in Note 2 above. Refit the oil-filler cover having first removed the vent screw from it. (The vent screw is fitted to provent oil spillege during transit.) CAUTION The vent screw mast be removed when the KT transformer is in P 43 43 Refit the UT transformer cover. Removing and Reficting Covers Control Console To gain access to the interior of the Control Console for installation (2450/1 /385) 9 connections and adjustuents, it 1s necessary to remove the covers (sce Figure 3)+ When work has been completed, the covers must be refitted. 4.3.2 HT Transformer To gain access to the HT transformer terminals, it is necessary to remove the cover. To do this, the four cover securing screws must be unscrewed (see Figure 2)- When work has been eoapleted, the cover mist be refitted. 4.4 Connecting the Control Console a The two flexible conduits from the previred wall box (see Figure 6) enter the Console through 2 holes in the base panel at the appropriate side. B Cable clampe are used -to gccure the conduits to the left-hand side panel of the Console. © The two conduits contain 120 light-duty conductors (5 of viich are spare) and 8 heavy-duty ones. D ‘The light-duty conductors are formed into 8 branches. Each branch is terminated by a multi-way terminal strip, with the first and lest vire on the etrip being ideated with numbers corresponding to those on the fixed terminal strip to which it is to be connected. E There are 8 fixed terminal strips on the left-hand side panel of the Console (eee Figure 4), the top strip being allocated for Gecomat connaction- P Connect the flexible conduit wiring stripe to the corresponding fixed terminal strips and check that the flexible condult terminal strip dents correspond to the numbers on the fixed terminal strips. 6 ‘The heavy-duty conductors are divided into three groups which are to be secured to the main terminal board on the Console as follows: 10 (2450/2/0486) (2450/2/0486) LINE COMPENSATOR SWITCH TERMINAL STRIPS CONSOLE LEFT HAND SIDE FIG. CONTROL CONSOLE-TERMINAL STRIPS 45 Conductor Ferainal connection tent Colour u blue ul w Black Lose a rom a EB Creen/ Yellow t = 1 hue u r lack ras 3 Brom 5 The earth stud is located on the left-hand front stanchion of the Console, ‘These conductors ere not used and should be insulated and bound back to the cable. Connecting the HT Transformer One flexible conduit is provided for connection between the Prewired wall box (see Figs 6) and the HI transformer (see Fig.2). A cable clamp is used to secure the conduit at the HT transformer. The conduit contains 15 light-duty conductors, 3 heavy-duty ones, and 1 earth wire. ‘The 1ight-ducy conductors are terminated at each end by a nulti-wsy terainal strip one of which is to be connected to the 15 way terminal etrip on the transforuer top panel, and the other to the corresponding terninal strip in the wall box. Check chat the flexible conduit terminal strip idents correspond to the nunbers on the fixed terminal strip. ‘The heavy-duty conductors are divided into two groupe which are to be secured to the appropriate terminals at the wall box and at the end of the HT tranaformer neareat the Tube 4 anode socket (see Fig. 2) as follows: (2450/1/385) MAINS VOLTAGE EXP 292288 50 Fer TAPPING PANEL OUNTERS B88; ooo eins raps RELAY PANEL, ) =f) “a \ ) RIT 0 Oy [ft ) FIG.5 CONTROL CONSOLE-FKONT INTERNAL VIEW (2450/1/385) 13 4yoel Ab 4.6.1 4.6.2 1s tack me brown " 2 oy | tem Green/Yellow Earth % “This conductor ig sot used and should be Insulated and bound back to the cables UT Cables : ‘The HT transformer ts equipped to supply up to four double-focus X-ray Lubes. The positions of the ET cable sockets are shown in Figure 2. Preparatory Work Mains Supply Taps On the mains voltage tapping panel (wee Figure 5), set the CT link to the terminal (415, 380 or 2207) nearest to that of the mains supply volrage tn che building. Voltage and Frequency Tepe Line Contactor. Op transformer TRI? (cee Figure 7), connect the yellow wander lead at the left-nerd terminals (primary winding) to the terminal dented with the voltage (220, 380, 415Y) nearest to that of the maine supply voltage. On the right-hand terminal strips connect the green lead to the terminal marked 20 (secondary winding) for a 50Hz supply or to the one marked 24 (secondary winding) for a 60H supply. Magnetic Stabiliser. On transformer TRIO (see Figure 5), connect the red eleoved lead and the greon lead to their respective cappinge for 50R2 or 60Hz according to rhe maine aupply frequency Contactor and Anode Starrer 1 Connect the violet lead to the 220V terminal (at fuse F5) fer 2 50ltz mains supply or co the 260 terminal (at fuse Fé} for a 60H usin supply. al Connect the single thick white lead to the 220V cerninal (et fuso ¥5) for 2409 operation of Bucky and aucillaries. (2550/2/0486) 4.6.3 X-Ray Tube Over-Pressure Therual Cut-Out Terminals A Qn delivery these wf1L de Linked on the Consle terminal stripe {sce Figure 4) as Eollow TUBE 1 120 to 117 TUBE 2 ILL to LB TUBE 3 112 vo 119 TOBE 4 313 to 120 If over-pressure thera cut-outs or X-rey table interlocks have to be ficted, the secoctated Links must be renoved, Terninals are avallable cn the wall box to accommodate che wires fron the tube cnereal cut-outs. (Soe Appendix 1 for details of the toterconestions.) NOTE If tube or tubes are not Insralled, che corresponding link must be open (relay Pl is opea and Peep and. Flucra are disabled). - 4.6.4 Xray Tube Scator Connections : Stator vires should be coanected to the wall box. The ternlcal numbers are a5 follows: muse wise [ emamaL maze STATOR WIRE COLOUR 1 qa Brown (common) v2 Blue (eapacitive) 123 Black (nein) z . Creen/Yellow i ! 2 124 Brown (comon) 135 j Blue [eapacitive) 126 Black (wain) z Green/ Yellow 3 2 | Brown (common) 128 Biue (capacitive) 129 Black (nein) r Greon/ Yellow 4 : 330 Beown (conon) at Biua (capacitive) i 132 Black (oa1n) z Grean/Yel low 4.6.5 Rocky Connections The standard Bucky connections are as follows: 133 = 220V Backy 1 start 138 = 720v sucky 2 start 135 ~ 2208 Bucky 3 start 136 = 2200 Bucky 4 start 137 = 220v Bucky 5 stare 138 - 2209 Bucky 6 stact 141,142 - Common Bucky returne (220¥) 146-148 - Common OV ciner stare from Buckys (2450/2/0486) 8 FIG,6 R500 SERIES 2 WALL Box (2450/1/385) For twehniques hat ay nut coymire buckfes, link the extoctated bucky Stare earninals to Leeminal 187. This can be effected on che Gonsule terninal strips. Tue bueky wiring should be connoceed Eo the wall box teenioals, Jypleal interennnections are derailed below Veebalque | Tube Bucky. R500 terminals | Bucky terminals 1 L Not used Link 133 te 148 | tot uaed a a Table 134 12s . 146) ies met Ls 3 2 versatilt 135. 126 Lae 128 rae Re 4 2 otner 136 Burky stact Las Txp start ba Bucky return 5 3 other 1397 Bocky stare Las. exp start Lag Bucky revurn 6 a Manne 138 Bucky start 1 Exp start Lan Bucky reeura Note? There is no provision to switeh a bucky in or ont once @ partlealar tarknique haa baen aclactady 4.6.6 Tube Exposure Counters Sere the counter assemblies to the pane] (see Tige 5) ® Connect each pate of wires ta the appropriate patr of terminals allocated to cach tube Cac, Ly 2, 3 and 4). (2450/1/385) " mre — rs QOg0 000 EASE ELG.7 CONTROL CONSOLE-REAR INTERNAL VLEW (2450/2/ 0486) 4.7 R300 Series 2 Torminel Strip Connectiona Figure 4 shows the locations of the terninal strips on the Console. TERMINAL NUMBER. Bey Shaw lat 2200 TERMINAL DESIONATLON HT tank ~ve supply to «etering HY cank te supply to metering Xoray tube filament transformer connon X-ray tube filament transformer gupply - sroad focus Xeray cube filawent transforner supply ~ fine focus Supply for cube 1 UT changeover cizeult Supply for cube 2 HT changgover atrcutc Supply for tube J HT changeover cizcuit Supply for cube 4 WT changeover circuit Tron tube 1 thermal interlock ewiteh ) Veriffes positfon Froa tube 2 therual iaterloek switch | of changaover From tube 3 thermal interlock switeh / suite afver From tube 4 thermal Laterlock switch J remate operation Wain 24V de supply for NT changeover cixesit Spare tecninel . Be tabe 1 thermal tnteriock owicoh ] Operates SPL relay Zo tube 2 therm Interlock swicch | LF tabe teaperature fe sabe 3 thermil interlock seiteh fis nowmal and cnrFoct Ze tobe 4 therm tnceriock swicen J tate is selected Tube 1 common sector winding s8oply Tube 3 main stator winding supply Tobe 4 conmon stator winding supply Tube 4 capacitive ctator winding supply Tube 4 sain stator winding supply W Ducky 1 start Ov Bucky 2 start Ov Bucky 3 start OW Bucky 4 start DY Bueky 5 start OV Bucky 6 start Bxpose ready - epare N/0 contact Tube 1 capacitive stator winding supply Tube 2 mala scator winding supply ‘Tube 2 conmon stator winding supply Tube Z capacitive stator winding supply Tube 2 main stator winding supply Tube 3 common stator windiag supply Tube 3 capacitive stator winding supply 3 4 4 220Y ae eupply for ancillaries Spare cerninal (2450/2/0686) 1 ‘TERMINAL, SUMBER 146 1a? 148 128 150 ast 152 153 134 455 156 i57 asa 159 160 151 162 Les 164 165 166 167 168 169 azo V7 2 173 174 1s 176 ur 178 179 120 121 182 183 ise 185 186 188 189 190 191 192 193 1s4 20 } ‘TERMINAL DESTONSTION Bucky feedback concn #24 cachatque 420Y technique #24 ceehndque 428 technique 420V technique 4200 technique 428V enable to external exposure interlocks Fluoroscopy extornal footawitch feaoto rediography preparation comand Totoro rasiogeseny Eevee eseeac Preperation enable edveute Common % Anode speed Rovmally closed (150¥) contact | geleceton = Torsslly opan' (3UH2) cooeact | incase satomelc eaposure device ctveett couman for HY canpensaeton Sire hiniuua supply (fron 79) extersat Flecroceopte Herioum caply (fron Foy A muriac connect tees Slider inteonal floocsscony Wi varie eonsect on Slider ancernai fiustoscory WY variae conection Interac! ek detec cones put Tneerned eh drive cceeest oocpet Outqut vom exterael fluocesceny aA delve US safety interlock + 24 enable from externel exposure interlocks Photo cima tn etreatt Photo tine stop signal (stop exposure) Autonatis exposure device circuit contacte (kV compensation) Spare teeadnal Spare teclutgve ewlech common tn 5 Spare technique eviteh our for Bucky 1 Spare tectmique eviteh out for Bucky 2 Spare teckuigue eviteh cut for Bucky 3 | signal Spare tectnique euiteh out for Bucky & (contacts Space technique switeh cut for Bucky $ Spare toctinique awiteh out for Bucky 6 | Spare X-ray radLography contact (RTE~3) Spare terminal Common spare fluoroscopy contact Normal closed esare fluoroscopy contact Wormally open spare fluoroscopy contact Spare tevuinal (2450/270486) TERNAL, NOKBER 195 196 4s? rs } 199 ano | 201 202 203, 204 ty 208 4.8 TERMINAL DESIGNATION Spere N/O radiography preparation contact Noraally epen spare fluerossopy contact Wormily open spare fluoraceopy contact Gomon stator vinding supoly for 150z starter Gapscitive stator winding supply for 150Ke starter Main stator winding supply for 150Hz starter mAneter (must be nornilly short circuited) Comme Interface Connections Interfacing comections are detatled in Appendix 1. (2450/3/0888) 22 54 NOTE Tuva the line compensator switch on the Loft-hand side of che Console until 220V 15 indicated on che test meters Cheek thar the pointer of the @A mater is coincident wen (te red dati sak. Adjust 217 (see Tig.5), if necessary, to achteve chise Swicch [OFF] the R500, Masaonager the test veilcweter. Setting up the Tube Overload Cireuits Three types o£ Keay tube can be deployed to make up che ental comlamenc of 4 cubes. on delivery, che kW/Time Overload ped (see Pigs 6) is net up as detailed below: Tans Nunber tube | Tube Focal | Rotor | Mexinun [Auxiliary Device type | Identification | Stee, | Drive | Voltoge | (working Station a Be, ww X {Dynamax 50-69 J 0.6, 4.3 | 50 1s, 1 x |Dynanax 50-69 | 0.6, 1.3 | 50 125, uy ety ay y |oynanas wo 49 [1 ~ 2 50 125 v 2 | tynamax WD D8 50 ao vr 4 20 NOTE on the K/Tise Overload pch, the three types of tube are referred boas X, Vand Ze he tube selector switch mst be programed for ihe cube combination deployed. ‘This 13 achieved by re-arranging the Links on che KW/Tine Overload pcb. Fig. 8 shows Links A arcenged for on application where tubes 1 and 2 are type X, tube 3 is type ¥ and tube 4 is type Z, three types of tube being used. link Bis shown areanged such thet oply tuhes 1 and 2 are operated at high speed. Technique 6 (tube 4) {2 axelusively reserved Lor mammography. {f chis factlicy is insta¥led, the kW/Time Overload pcb «iLL have beea premired to accowmedate the ratings of a Dynamax Hi) 40 MO 0.8 mm Cube vith paralleled-filanent. ‘on deldvery the tude focus combination is as follows FINE FOCUS 50, 100 & 20004 BROAD Foous 300, 400 & 500A These can be rearranged to suit individaal tube requirements by following the example show an Tig 10 page 28. (24 50/2/785) SLIT EL pd td ts) Le SH TTT il mM) = ML OW WU (2450/2/ 0086) HIGURE § MAX LOAD FOL Sehed NOTE 524.2 NOTE Ensure that the main oupply to the R500 islo#e,| Remove the Eroat panel of the Console. Remove connectors El, f2 and £3 frow the kK/Time Overload pcb and plug them date the corresponding comectors on the tost hoard (sus Fig. 9%). Preliminary Work For Tube Type X Tarn potentfometers P19 te P24 on the kV Adjuat peb (seo Fig, 10) ful ly eaunter-elockwise. On the teut bourd Link ceeminal X in the GROUP CHULCE teeation to the tube number(s} to be displayed with Group X. swice [in] che R500. : Rotate che contrel panel, switch €o kelect an auxiliary device assoctated with tube type Ks TE avy of che auxiliary devices associated with ehe tube cype % LS Fiteed with « high-speed Facility, Link tube type X to lIS at the TO NS CHOLCE Location on tho test board. ‘Then curcy out the HS proceduces detalled In paras 5.4.2, 5.4.3 and 5.4e4e TE chece are ao high-speed facilities associated with cube type X proceed to paras 5efe5e A tolerance of + 2 positions of the clme-secting control ts sLlowsble on all the time settings given in the following procedures. The Cima settiags available by adjustnont of the time Fontgol switch are shows oa Fig. ly Setting up Tube Type X Fine-Focus, High-Speed Overload Paraueters Fine focus 1s autonatically selected whea one of the 30, 10) or 2U0 mA valucs is selocted, The following details the setting-up procedures for tha tine focus sottings, 50 aA Setting £ Sadech [ON] che R500 and aelest tube X and SU my SL Gmmect a 1OY dice voltmeter across capacitor €25 on the kV Acjust pod 4037, Nocate the radiography kV eonceols until che voltnecer indicates 745 Vs Check thar the indication on the kV readout is 15042 RY, Tf necosnary, néjuse 25 00 pcb 4037 to give that Tndication. ‘Macennect the voltecter. On carlior models, check the munber on KY Adjust PUR, If it is type 3504 the abave 025 should read Clq. For location refer to drawing Wo 3594 in Appendix 2 (2430/2f 785) 27 8 TUBE FOCUS CommINaTTON — EXAMPLE (See raving No 3760) TURE 1 FINE $0 and 100mA ONLY TER 2 FINE §0, 100, 200 & 300mA CRLY TUBE 3 FINE 50, 300 & 200mA ONLY = FIG 10. KV ADJUST PCR (2450/2/785) Lit ascertain from the rating table supplied with cube type € the tube restriction at S0m\, (Fg. LA in for a Bynanax Super S~b8 X-ray tube at iilgh speed, 150 He, and fine Tocau, 0.6 mm, and it is used as on example to illustrate the setting up procedures.) Thera ix no tube restedetian in this case because 50 mA exposures are permieced wichin the full cenge of the So tiser ewitch. The Rbill) generatoe Tatings mst Limte the voltage to 125 KP (see ig, 12) mA Solectad 4¥500 Serioe 2 aaxinum allowable L voleage, kV 50 125 how. 225 200 125, 300 325 ano 125, 5m 100 x vk vit (2450/1 /385) Select 50 mAs On the Lest hoard, Link the rollewing teentaais: TKA and 0.01s (the Lodest tine sezting) KXA ond 125 kv VXA and 125 KY Turn the radiogranhy kV controts until the OVERLOAD Lamp fluvinatess Check thet the indication ou Lie kV readout ts 1254249. Th 4e ts not, move tho onde of the KAA and VA Liaks attached to the 125 kv terminal co a tigher voltage vermaal (LE the OVERLOAD Lamp L1lanianted at less than 125 k¥) or co a lower voltage tarninal (if the OVERLOAD tonp {iluninated at more than 125 x0}, Advance the time control co 1¢s maximan setting and check that the operation of the OVERIOAD lama at 12) kV is uot arfected. Mark the actual Link positions on a copy of the matrix diagran for the overload board, e.ge, 1f TXA Lak is ac D.Ols, mark the point whore the Linea IMA and .U1 intersects TE che KxA and VR4 iinke are at 125 KV, mark the points where the KHA and 125, and the VHA and 125 lines mect. These are the points marked Aon Pig. Se Kenave tho Linke fitted at otep 5.462 B ve 29 30 100 ' a aut a tw mA Setting. Ascertatn €rom the rating table, Getgella), the tune restriction at 100 m4: Tt is possible to use the Lube al 125 kY up to Se, 60 the time and yolrana sectings shoold be the sane as for the W) mA setting. Selece 10) ma. eu the test board, liak the following tereinals: TXB ond 0.01e RXR ond 125 bE VEB and 125 kV Repeat steps B vi aad vit for the 50 mA setting procedures Meck cle actual Link position on the matrix diagram, The Link postticna are markad with a Bon Hg. @ aucuniug that the TXB 1fpk was ar O.00s and che KKB and VXA Links vere ben at 125 ky, Kemove tha links firted at step 5.4.2 ¢ iti. mA Setting. Ageertaia from the cating table (Wig. 1LA) Lhe tube restriction at 200 mA. This time, 21 is only possihle to bake 125 KW expoowren up to 0.60 but ner Uwe which meena that 1c WL11 also he necessary to ostablish the maxiuan reraitted kV at 32. trom Figs LIA 70 KY is perubtted but aot 80 KV, Tho maxiava porsitced time for an 80 kV expusure Le therefore 4s and not Ss. Select 200 mA. Gn the test board, Link the following terminals: THE and 086 KEG and 125 RV VXC and 80 kV Set the radiography kV eonrrale ta give a resdnut of 80 ky ¢2450/1/383) Exposure Time, # kp d.0L | 0.02} 0.05f da} o.2fas}1 | 2 [5 | 10] a | so} ive Maximum Current, aA 50] 392 | a92 | 392 | a92 | 392 | 292 p392 | 392 | 305} 23 | 167 J 00 | ae ev faai | arr far | arr | 471 | ase |4o2 | 34 | 254} 193 [139 | ga | 30 70] 480 | 477 | ase | 456 | 435 | 390 | 345 | 292 | 218] 165 | 119 | 72 f 43 ao] 420 | 417 | aio | 399] 380 | 342 Ja02 [ass [ig1] 14a [ine | os foe 30] 374 | 321 | 364 | 355 | 338 | 304 [26% | 227 F169] 12a] 93 | 36 | 93 100] 336 | 334 | 328 | 319] 304 | 273 f2aa [204 [153] 16 | 83] 50} 30 110] 306 | aoe | 298 | 9p | 277 | 243 220 | 188 F1a9] 105 | 76 | 46 | 27 225] 269 | 267 | 262 | 255 | 243 | 219 f193 fae3 fiz2} 92] 67 | 40} 24 150} 224 | 223 | 219 | 213) 203 J 192 Ji61 | 1a8 fies} 77] Se} 33] 20 A RINE POCUS (0,6 mn) Fxposure Tine, & kvp g.a1} v.02} 0.03] orf oe2fas| 1) 2 | s | toy 20] 30 five Haximm Current, mA sa] 250 | 750 | 750 | 750] 750] 750} 750 | 750] 593 | ai2] 273 j 420 f 60 en] 92a | yaa | 924 | 924] 924] 924] 924 | 739 | 494 | 343 | 228 | 200 F 50. 70 i098 |rosa |ro¥s |ro9e jing} 979] sas | 634] 423 | 294] 195 | ao | ay 80 fiz0u fitas f1isa [1x01 f20%7} #30] 708 | 354 | s70 | asa) 71 | a5 | 38 30 f1o7e |iosa Jin26 | 979] gua) 761] 629 | 493] 329 | 229} 192 | 47 | 33 uw f yes | vss | 923 | 281] sia] 685] 566 | 444 | 296 | 205) 137 | 50] 30. Lil e76 } 866 | 839 | gov] 740] 623] sis | 403 | 269 | 197) 124] 5a] 27 tas} 7z. | 762 | 73% | 7oa| s51} ss} 453 i355 | 237 | tes) tog | aw | 28 150] a2 | 695 | 61s | 987] 543] 457] 378 {296 fase | 147) 91] 49 | 20 BROAD HOCUS (1.3 mm) MOTE these ratings are for a Dynavax Super 50-69 tube with « single-phase supply. FLG.i1 X-RAY TUBE RATINGS - EIGH SPEED (15U Bz} (2450/1/385) a od r1G.12 500 SFREES 2 AAXIMIM RATING CHART (2450 /1/385) NOTE Behe v Advance the time concrol te 4s (the next lowest cine serting to Ss) and check that the OVEKLUAD lanp is extinguished. Select 55 and check that the OVEKLOSD lamp Itlum{nates. LE it goon not, move the VEC Link co progressively tower voltake terminals (75kY, 7OKY, ete.) uncil the OVERLOAD Lamp illuniuates when the time setting is 3s. vi Set che radfogcaphy kV controls to give a readout of 125K¥. vit Advance the time control and check chat the OVERLUAD Lamp (luntnaces 2t 0.83. Tf At does nor, move the KXC Link to progressively lower voltage terminals until it does, If the OVERLOAD Lamp (1iuminates a 2 Line serting less than 6.68, move the KXC link to progressively higher voleay= Lerninals until the Tamp illugingtar at tho 0.35 setting. vidi Repeat uteps D iv and v to check chat the OVERLOAD Lanp Iluminares at the saccing combinacion cf SOkV and Se. TE re-adjustent of the XC link is cecessary, chen steps D vi aad vil will have to be ropeated uatil the settinga are correct at both 125kY and BORY. ik Now check the setcings a all other intermediate voltages. For example, at 110k¥ check thet the OVERLOAD lamp iThuminates 2¢ a Line setting of i.6e, IF the teseking 48 poor around the high KW/shore tise zone on the tuhe rating shart, readjust matrix KXC. Lt the teucking 1a poor around the low k¥/long tise zone en the tube rating chart, readjust matrix: VEC. If there is st{11 roow for improvement in overload crack ing around the high kV/shoré time zone, readjust matedx TKC followed by readjustment of matctx VXC to achieve an overall compromises Adjustoent of the time Link on che test board has a greacer etlect at high voltages than it docs at Low ones. The Eime link is moved ia the sane direction as the required tine change, i.e., tf che OVERLOAD lamp com=s on at Loc high a cine setting, the Cine Link nust be uoved to a Lower time value, xi Mark the actual link positions on the matrix diagran es before. Rif Hemove the Links fitted at stop 5.442 D iids Setting up Tube Type X Broad-Focus, High-Speed Overload Parameters Broad focus 1s automatically selected whea one vf the 300, 400 or 500mA pusbbuttons 1s pressed. (2450/1/385) 3 3a 34 Figure 118 is the rating chart for the Dynamax Super 50-69 at high speed, 150K2, and bread focus, l.dan. From this it can pe saea Chat the cane exposure blockages for 125 kV ace 3s and Lefis For the JOD and 40UwA settings, respectively« uhe peicedures for sercing up the broad focus, high speed overload eiecuits are oe detailed in para. 3.4,2 tor the fine focns, high spaed ongs. However, 42 msc be noted that at 50) mA the tube rating can exceed the 37.3 EW vating of the 2500 Series 2, Under thase circunseances, the overload circuit mist be suapted to restrbet expsuces to WO RY mex at 500 aA, MULs means chat at 900 aA the overload circult must be anda to Iollaw the tube vating curve from [00 &V ang not 125 KY. The &xF aud TXt Caps mist cherefore be see by trial aud error to back LDL RV exposures on tine ovicch nettings below "I.6s. Examples of Final tapping polnts for a DX50-69 1. aa HS, Tahe nee showa below: 300 mA XRD 135 kV m™ 28 veo 80 RY 400 n& RXE 140 KY TE 1,26 VRE 60 RY 30D WA KX 110 KW TRF 1.25 VER 45 kv Setting up Tube Type X Low Speed Overload Parameters Having set up tha high speed overinad paramecers as detatled in betas 504.2 and 5.443, 1b is now necersary Eo set up che low spaed Overload paraneture fer asch mi valne for any Cube ubleb Nas un associated High-speed facility. This ensures that tubes wilt enly be switched to the high-wpesd mode of control vhon the uernsl epord overload ratings are exceeded. Hofer to rhe Luba rating charts for low speed use, snd aljust the pocentioneter ou the KY Adjust rch 4037 (see Tig, 10) avsuciated with each aA Listed below clockwige uatil the HIGH SPEED Lamp illuminates at ar before the overload point calculated om every &Y rating curves (2450/1 /385) 78, Potent tomezer Selected a poh 4037 50 219 209 220 200 Pai 300 P22 400 223 500, 26 c Procead to para. 5.4.6 for setting up the overload parauaters for any other type of cube used. Tf only one cype of tube 1s used, proceed to para. 544,75 5,4,5 Tube Type X Without Associated High-Speod Facility a Carry out the procedures detatled Im paras Se4et, 54442 and Se4ed for tube cype X. Ta this cage, the tube restrictions will de ascertained from the low-speed tating tables given tm Fig. (3 instead of the high-speed rating tables given in Fig, 1ly x Proceed to paras 54446 for secting up the overluad parameters tor any other type of tuhe used, Lt only one type of tube is weed, procesd to parade 5+4475 5.4.6 Setting up Tubes Type ¥ and Z 4 ‘The procedures for seccing up Fhe overload paranccers for tuse type Yare identical to those already detatted for tube type X Lo pare, Sele] bo 544.5. 8 The procedure for setting up the overload circuits for tube trve % if this is a Dynsmax HD 40 4o G,#nm cube is ce follows: Exposure Tine, § kup ot foalas| 1 | 2]a ja fs Maximum Current, aA 2 250 | 250 1250 | 250 | 250 |248 /232 j 216 30 300 | 30y 293 | 260 | 226 [205 [193 | iso 40. 200 | 255 f220 | i95 | azo fa5s [145 | 135. SINGLE EXPOSURE RATINGS FOR FD 40 TUBE (1 ph, 30 Hr) 4 On the test board 1ink terminal Z to tersiaal 4 at the CROUP CHOICE Jocation, and connect che other Links as follows: (2450/1/385) 35 SDA Tea 0.918 } wa PRANTL RLCEKED wa 1 looms RO. me ow wm 200 mA Gas we aw vo 204 300 mk m1 mo Ke ub tome 400 mh me O01 Ke lou PERAANBNELY BLOGKED ve 10 wy 500 mE OO fe TM) ramasrer scoccn ver owe i swigen [DnJehe R500 and check at1 the overload points as given in the Table above "Single Exposure Ratings for HD4l) Tube ph, 50 He)". If the tracklag is poor around the high kV/short tie zone on the tube rating chert, readjust the matrix kZ. Tf the tracking £9 poor around tle ow kV/long time zone, readjust che macrix V2. If there is still room for improvenent in overload tracking around the high kV/enort time zone, resdjust nacrix 72 Folloved ty readjuscwent vf natrix YZ te obtein an overall comronise. ¢ When the overload perameters for all che cube types have beon set up, proceed to para, 5+4y7e 5.4.7 Programming the Overload Board a With the Overload Hoard renoved from the R500, disconnect any Liake from the time and voltage matrixes that have not been marked on Fig, & during the overload cirentt setting procedures. Solder in the remaining Links as marked on Fig. 6+ CAUTION ensure that wateix Lines are cot bridged daring che | soldering pracesr. 8 Solder in the links A and 8 as appropriate (see Fig. 8 and pars. 3.4 C, for tube progeamxing and high speed progranming). 36 (2450/1/385) Exposure Time, + kVp o.a1| o.o2} a.03fo.1]o.2/a5} 1] 2 | 5 | i] 20 | 30 | 109 Maxiaum Current, ma so {sau | a9. | sea | sas | 377] 357 | 33z| 298 [2ar | 193 | 1as | 93 | 60 eo }426 | 326 | 32a | 320 | 314 | 297 | 277 | 24a foo {161 [122 | 77 | 50 70 | 279 | 279 | 278 1275 | 269 | 255 | 237 | 213 fare | 13e | toa | a6 | 43 ao asa | 249 | 243 | 240 | 225 | 223 | 208 | 1a6 [150 [120 | a1 | 58 | 38 eo f2i7 [217 | 2 | 214 | 209 | ge | 189] ves F134 [107 | si] s2 | 33 aon five } 19s } 19s 192 | 19a ]178 :166 | 149 }120 | 96 | 73 | 46 | 30 tio fuze [avs | avy [ays jira | rez faor| vas figo | a8] 66 | 42 | a7 125 |156 | 156 | 156 414 [isi] 143 [193 fits | 964) 77) 48 | a7 | 26 150 |130 [130 [x30 4128 126 frig au] 99 | a0] 04 | as | aa | 20 A FIRE Focus (0,6 mm) txposure Time, = lev or} o.02}o.05| ot /e.2}a.s]1 | 2} 5 | 10] 20] 50 100 Maximun Current, mA sa [se | 730 | 750 | 750 | 730 | 750 | roo jut | 513 | 373 | 256 [220 | 60 40 |924 | 924 | 924 | 907 | 874] 796 j706 |392 | 42a | 311 | 213 |100 | 50 7a }gai | sor | 793 | 777 | 749 | 83 |o05 | S08 ] 306 | 267 | 18a | 6 | 43 a0 {701 | 701 | 693 | 680 } 636 | 597 520 | 44s [321 | 233 | 160 | 75 | 38 90 |623 | 923 | 616 | 604 | 583 | 531 Jari | 395 | 285 | 207 | 142 | 67 ] 33 109 [561 | 561 | 555 | 544 | 524 | ars |aze | 455 | 257 [167 | 128 | 60 | 30 140 |s10 | 520 | 504 | 495] 477 | ase faas a2] 233 fa70 | ase | 55 | 27 125 449) 449 | 444 | 435 | 620 | 382 [339 |2—% | 205 J 149 | 302 | 4a | 24 aso |3z4 | 372 | 370 | 363 | 350} 319 |2ez [237 [171 | 124] a5 | 40 | 20 NOTE these rarings are for 2 Dyranex Super 50-69 tube with a B BROAD FOCUS (1,3 mm) single-phase supply. FLG.L3. K-RAY TUBE RATINGS — Low SPE (2450/1/385) a (50 Ha) KY Meter Calfbracion to WY Primacy Voltage 5.5.1 Flucroseopy ‘ Connect a 500¥ a.n., f.8.d., cost meter across fuse lu (soe Fig. 7) and terminal 168 (on the terminal strips at the left-hand side of the Console). 8 Vemporarity Link terminals 155 and 156, © switeh Pa} rhe R500, d Press cha fluoroscopy pushbutton and set the Cluocoscepy KY control (nce Figs 1) to give the Eollowing ceadings oa the cest sete Tont Motor Beading, ¥ KV Meter Reading ut 30 167 15. 224) 100 f Li the test meter reading ut SU kV difiere by more than 1% from TV4¥, or che one at 100 KY by more than 2% from 223V, adjust che wore kV contcol te give the correct reading on the test meter and then adjust potentionater P26 oa the kV Adjust peb (see Fig. 10) to obtain the associated k¥ wexer reading on the console display 7 Ronove the Link fitted ac Be 5.5.2. Radiography a Connect @ 10000 acce, fisad. meter across fuses 11 and 12 on the manok e. 3B Snsure that potentionstare 27, PY, FLL, P13, P1S and P17 on the kV. Adjust pcb 4037 are all curned to their extrene counter=clockwise positions. c Switch [ON] the #500 and check the UT primary voltages for each ms and KY Indicated in che Table baloy. (It does wot matter which tube is selected). 8 (2450/1385) Kv A 50 75, 100 125 Wt Frimary Voltage, 50 a4 160 20 260 aon Le 186 2 2er 200 136 185 232 280 300 158 203 249 294 490 190 233 276 220 300 218 ree} ond 360 b Ur che voltages ar 50 kV are not within 2% edjust the appropriate nA comensatloa potentioneter on the kV Adjust pcb 4057 (see Fig. 10). + S0mA / SURY PI | Le THERE 15 2ODAA {SKY PZ | TNSUFFIGLENT Abdus 200A / S0kY —-E3 | MENT IN ANY OF TuESe, 300mA / SOKY PA f POTENTIOMETERS, MAKE 400n8 / 50k¥ PS | FUTAEN ADJUSTMENT WITH 500aA / SULT BG | P28, e Lf the voltages at 125 kY axe aot within 1% adjust ths appropriate TA compensation potentionecer oa the KY Adjust peb 6037: 50n8 / 1odma / 200mA # 300mA f 400mA f 500KA / 125K¥ 125K 125K¥ 125k¥ 125k 125kY va | ue tuxne os FID] INSUFHICTENT ADJUST vi2\ amNT Tn ANY OF THESE ¥14/ POTANTIDMETSRS, MAKE P16] FURTHER ADJUSTMENT Pls) ETH ¥27. NOTE Tf a particular aA potentioneter, evg. the 50kV one, has to be adjusced the associated i25kY potentiomater must alva be readjusted because they are Lareractives F Repeat steps 0 and Euncil che 1254V and SOkV vaTues are correct. 5.6 Adjustment of Tube mA a Connect the primary of the AT Transformer. (2450/2/0486) 39 fo To sereen from the Console on all tubes, temporarily conneet a cuitable eviteh (to be used se the “Eluoro test euiten") across terminals 156 and 157 on the Gonsole and link terminals 156 and 155. To aet up the imdertable tube mAs output frou the Consele, connect = temporary link acroca terviaals 155 and 160 on JB4 and use the Prop/dapose switch on the control facia. A Bucky 1ink ust also be connected across terminals 133 and 147. Broad Focus Pre-Hear Adjustuent for All Tubes Disconnect the broad and fine tube filament supply wires ac cerminals 10% and 103 on the AT transformer. sutten [i the e500 series 2. Select tube 1 (undertable). Press the Fluoroscopy kY puehbutton and set the flurorseapy KV control to indicate 7OkY on the merer. : Press tho fluoro test switch and record the reading on the mA weter. Ths will be very sual beceuse Lt will only be recording the HT secondary capacitive current. Release the fluoro test switch. suited Biche 9500 series 2. Reconneet the broad focus wire to terminal 104 on the ET transforner. switen Pix] ehe R500 Serses 2. Press the fluore tect awiteh. Kote the tube emission on the mA moter. Thig must be O.1mA in excess OF the reading recorded ta paragraph 5.6.1 B. Lf the pro~host omission {2 ineorreer, adjuer the tapping band on restster RIS (see Figure 5) until O.1mA is added to the capecitive current recorded in paragraph 5.6.1 Es WARNING surrcu OFF THe R500 Soriee 2 WHEN ADJUSTING B13 (2450 /2/0486) 54602 WARNING switce ore THe 1500 WHEN ADSUSTING R13. 5.803 A 5.644 Pluoro mA Adjustment Check that the R500 ts/orF, Reconnect the fine focue wire to terminal 105 on the lit Transformer. switch ox ene R500. Select tube L. Bross tho fluoro test switch and select 70k, Increase the Fluore wA, With the Fluero aA coatrol cured fully clocksise, the mA meter should indicate set more that Sn, Lf necessary, adjust the tapping band corresponding to tube 1 oa resfstance R13 (see Fig 5) until this value is dbrsined. Fina Focus Pre-heat Adjustuenta for Renaining Tubes settch[or]ene R500, Select the tube Lo be adjusced. Frese the Cuore test switch and select 70kv, Increase the Fluoro mA, With the Fluere mA contro} rotated fully clockwise, the mA meter should Indicate ner more then Sma. If necessary, adjust the 213 capping hand corresponding co che Cubs undergoing adjustment for SaA maximum enisston. Fluoro mA Capacity Compensation sutech[ov ene R500. Select Tube 1 Switch [Oi] the fluoro test awitch and select 10UkY and Sad oa the EVuore Goarrots. Adjust R26 (soe Fig.) until the A meter recding begins to decrease. Continued adjustment in the eaze direction will cauxe the mA to dectease Further until a point is reached when tt will begin co vise ayain, ‘The correct adjustment is the wininum nA which caa be cbtaiaed on the mA nerer, Reuove the fluoro tact owitch and the link aeross termfaals 155 and 1564 (2450/2/0486) 41 566.5 Radiogeaphy Exposure Timer Check Road paras 546 (exposure Links fur teebaique 1) Um the Hiner poh (see Fig. 7), conneet the signal Lead of 8 DM6% storage CRO cr equivalent to the 6 pin Ansley edge connector A4 at pin 6 (anoccupiod outlet) and chip tha garth lead €9 chassis ground. Check that che coloured guysors op the mA adjustment TheostaLs BLO and RIL (see ¥ig.3) ave at Lhe butcom of thele teavel. > satued {BR} che R500 and setect 0kY, SOn4 and 0.01 seconds. Take @ radfography exposure and chock ehe trace as thdteaced on the OHO. This stould bave en anclitude of I5V anda width of Ul ws. F Advance the Tiner switch to Del seconds and toke another exposure the CW tease Luis thee should neagure 10) no on the tise buses @ Advance the Hyer switch to 1 secon and take another exposurs The CRO trace trie ting should meawure J ees on the time hase, AE the Tier Lo inuccurate, ceter Uo the aetting-up procedure 3.6.6 Sadiography ah Adjustnente, now Mammo A tho tubo Filogent drive aujusting resiotore axe show in Figs M10 As used foe tubes 4 and 2 Wii dg eed for tuber 3 aed 4 fice has G2 auto of ellding curuors shich are colour-coded Lor Cie identitieation and alee ietter-coded [oe mA identi Peat Lan ae In the table below! MITE een ‘eu xia . Tune 1 Tue 2 Tune 3 Tone & Sond ona Sonn TA 5 ree 1 rene 2 mwe 3 roe 4 Goes 008A 100s Tha D re 1 ross 2 swe 3 rue 4 200n8 200na 2o0ne 200mA a ruse 1 rose 2 ane 3 aver 4 Homa Daa 200m Soak ress 1 ruse 2 use 3 swe 6 100m avoma 400n8 ma “ rime 1 ane 2 tune 3 supe 4 50m soon 100n8 um 42 (2450/1 4385) © Connect an mAs meter to the oxteTnal mAs meter Tink and adjust all the tube mA values at 75kV by the sliding cursors tabulaced above. LL there i incuffieient adjustment to set up the high mA values, recurn #11 sliding cursors to the borton of chete travel, tava the blue wander lead tap on TRO (the stabilised filament deive Cransfnrner, see Fig-6) to the aext highest voltnge for fine Focus. Move the orange wander lead tap to the aext tughest voltage for broad focus. AdJuce all the tube mA velues at 75kV by the slidiag cursors tabulated above. Space Charge Coupensation Toke a series of radiographic cxposurce at different EV settings on every position af the mA switch as tabulated below: A SHLgCTED ‘ev SETTINGS 30 50 5 10a 120 109 30 6 100 120 200 so 5 too 20 00 so 1 ry Lo 00 80 5 io us 900 50 15 ” 100 Compare the mae soter readings obtained within the kV cepge specified for cach setting of the mA svitch, ‘The overall yartatioa should not exceod 5% of the nominal value geasured ab 75k¥, e.g, for a coninal value of 20m\s a change of amas is perniasible withia the kV range specifind. Adjustment facilities are aveileble on che sceondary of the space charge conpensstion trancforner TR? (ace Figs 7). (2450/1 /383) 3 There eve eix wander lead tapa identad ee followst CORRESPONDING mA 50 Af the indicated uA goes higher ven the WY Is Incressed, move the wander lead correspocding to the mA oolected to a higher tap aumber and vice versa. D —-soAE ald tuben are of the same type, the space charge compensation for one should be edequate for the other Jn an installation with very different foct, such as 0.3 and lum and I and 2am, not all the foct can be compenssted within 5%. A compromise will have ta be mde. 5.6.8 Mammography mA Adjustments - Simultaneous Ealesion of both Tube Filenenta (Parallel Operation) A The tube fflanent drive adjusting resintora are sham tn Pigure 5. R27 Ja used for adjusting the emall focus. RU is used for adjusting the large focus. Fach hes a set of three aliding cursora whlch are letter-coded for mA Adeati fication as in the table below: R27 SMALL =©—-RLL. LARGE B 100 aa 100 mA G 200 mA 200 aA D300 wa 300 24 NOTE After calibration, ach focus will deliver half the indicated uA. B Connects 100 volt fed geter across terminals 104 and 105 m the HT task. The object ia to obtain meter readingo a9 close to zero aa possible during setting up. In this wey both filaments will ¢ balanced and thus capable af equal emtsaton. e Switch ON and select 35 kV and Technique 6. D Take a series of radiographic expoaures at 20 mas starting at 100 ma, Agjuet each rheostat R27 sad 211 in turn to obtain the selected mAs on the wae uater. 4 (2450 /2/ 0888) 37 3.7.2 During float adjuarment of each mA value closely observe the test volt seter during preparation, and fine tune R27 and 211 for minimum readings coincident with 20 mis output during exposures. PINAL CALIBRATION OF LY METER 10 LINE RESISTANCE At is strongly recommended thet en approved kV acasuring device is used for these tests, such as a Voltim Shock Proof High Valtage Measuring Frobe with suitable C80, oc a Winsconsin Penatrometer Cassette. If these are available, proceed as detailed tm METHOD 1 (see paragraph B.TADe If an approved kV measuring device i¢ not available, praceed aa éetatled in METHOD 2 (see paragraph 5.7.2), but note that this ts not considered accurate, especislly on the 400 and SCO mA bangs of the R500 Sertes 2. Method 1, with kV Measuring Device With the @A switch at 50 MA, set the KY meter th indicate 80 kT, Take a suitable exposure end note the reading recorded on the ®¥ measuring device, If this differs fron the kY display indtcatton, take nore exposures with the « coarse and fine ecatrols reset accordingly util the kV teasuring device records 60 k¥ peak. Adjust potentiometer PT oa the tv Adjuer peb 4037 (sea Figure 1D) until the kV display indicates 80 ky. The kV display is now compensated to prevselect the actual kV wiiich with be generated ON LOAD, Ropeat this procedure for 100, 200, 300, 400 and 50D mA using potenrionerers P9, P11, P13, PIS and P17 respectively on pcb 4037. Method 2, with ¢ kV Meaguring Device aaa a 1 CAUTION: avo OVERHEATING X-RAY TUBZS BY KEEPING EXPOSURE TIMES AS | L SHORT AS POSSIBLE WHEN OBSERVING ON-LOAD kV MRIKR READINGS. With the oA switch at 509A, ser the KV display to read 73 kv. Press the radlograrhy Prepare/Fxpose switch and note the kV display indication OW LOAD. hen the expoaure is terminated, recallbrate the kY display to indicate the kV noted ON LOAD by adjusting potentiometer P7 on the kV Ad Just pcb 4037 (see Figure 10). ‘The ‘V¥ display is acw compensated to pre-select the actual kV which will be geaerated ON TOAD. (2490/2/0888) as. 5.7.3 46 Repeat the procedure for 100, 200 sad 300 mA using potentionetera P9, Pil and P13 respectively. Ac 400 asd $00 ma it iil be difficult to observe the ON DAD kW display readings because of the exposure Lime liaftations. Kv compenzstion will therefore have to be eatimated by waltiplying the kV crop noted at 100 mA by @ factor of 4 and 5 respectively. Thase corrections can be applied by adjustment of poteationerer P15 (400 wa) and P17 (500 mA), Fluoro k¥ with kV Measuring Device Take a fluoro exposure at 1 at and adjust WY controls to give a readiag of 100 ky ca the kV weasuriag device. Tf the fluoro kv display does not indicate 100 kV, sdjust potearioneter £26 oo the KV Adjust pob 4037 until the Fluor kV display Indicates 100 Kv. (2450/2/0868} a COMMTSSTONING THE GEGONAT 2 Ihe following pages detail comissioning the R500 Series 2 Generater with the Geeomat 2 Refer sls to Gecomat 2 Installation ond Service yomuil Ne 248 (2450/1/385) INSCRIBING AND DRILLING DETAIL FOR THE GECOMAT CONTACL UNIT GECOMAT IDENTIFICATION PANEL FIG.14 GECIMAT CONTROL UNIT - INSTALLATION INFORMATION 4 (ares 14s bal 6.2 General, NOTE Vurticr inforaatina is given 4 Inst, 2243 (Geconat 2). Tnseribe and drill oles for the Gecomar contre? Unit (see Fig.l). If 4 bucky ds to be used, the Link between terminal 148 and the relevanz terminals 133 t> 137 on the #500 coatrol console (see Fig’) wit he removed. Interface che Geeomat 2 co che 4500 (nee #igsL3 and Fig 17) NOTE the terminals 40L to 420 on che Eront left-hand side af the R500 control cansole are wor used for thie Interfaces Fir the new ddeatification panel (supplied with the Geeamac) (wee Fig.14) into junction box 4. NOTE Terminals 72 and P24 fn the junction box are noe used. Antomatic kV Compensation Controls Switeh [or] che R500 ana the Gacomat. Select 50k, Odseeve chat both celays KYO and K125 ere closed on the RSA) relay panel. Confirm chac relay K9U epans when the kV enntrals are advanced to select 80KV, Adjustment en be effected by patentiomster P31 on the kV Adjust peb [occ Fige LO). Confiem that reley K40 remains open and that relay KI23 opens when che kT controls are advanced to eeleet 100k% Adjestuent can be effected by pocentioneter P30 on tha kY Ad jioe Pah. On hoard U6 (see Fig. 16) in the Gecomat Electronics Mudie eheck that Talays d0U9 cud 2608 open et BURY and Ta0ky, respectively, and chat they both stay open when the kV is advanced beyond MWOkV to maximms Af there is any discrepancy check the interface wiring, (2450)1/385) 49 6.3 a 50 Fila Density Calibration, Tube 1 Prelfninaries. vil This procedure is a guide for setting up the Gecowat in conjunction with casaettex fitted with par seceens (mediua speed). Screen 2 {5 thecefore chosen for chis calibration. When adjustments are completed the hase donsity junpers on board Ua ehould be adjusted Lo give approx, —25% mas on seceen 1 and +25% aAx on sereen 3. LE che X-ray department uses only Cast cangecate er eare earth screens, sceeen 1 ust be used for calibration and +25% Se must be added between sercon 1 and cercea 2, aad botweon screen 2 and screar 3 after calibration. chance that euffictent phantom thickneas 1x adopted ta arndace exposures in excess of the mininun switch tire of the ayatem, ‘wis checks will conficn thia and the rule in to od@ pore phantom If mas drops below & més. Lf a dencitoneter 1s not available, regard raferancas to base density 1 as a level of density vhere bold black print ou a white backgreand is just ebout legible when the exposed tila is used to mask the print in good daylight. Prepare board lb for the chanber/sctacn/donsity combination. For example, for a screwing room with chest bucky, start with a jumper between sercea 2, chamber 1 to density tap No 9, a jumper beevoen screan 2, chamber 2 to dansity tap No.9 and a jumper between screen 2, chamber 3 to density tap No a on board U? jumper LINK 1 for autowetic selection of centre cell when under rable cechaique is selected. Suiceh[GN]che R500, aelect tube 1, 70kV, 200m ane 1 second. Seicch[Oh|che Geconat with cencre celt and screen 2. on che X-ray Cable, load a full size casaetce vith no (Lim inside, Enouce that grid is IN, cone OUT, shutters fully open and no phantom in the oxpocure area. Take an expoaure and check that Cermination is Lamediare with @ very suall value of mAs indicated on the actual mas display. If che exposure cuns on to 1 second, check the Interface wiring, chawher wiring and Geconat for faults, Hepest the sbove procedure for the evertable and chest ghaubers oa each cell and eacure that in euch cage the overtable rube is carefully coned aver the cell which ts selected. Other cells must be blanked off vith lead. (2450/1385) AINHOZSNVHL LH 00S 9 X08 NOWLINAL oa t—on TNdNT SNIV oie O19 | 310SNOI_T0¥LNOI 90S 038N LON (Fo) AMONY 1 S3NIT a Fava UNNUIMOd Z LYWOIED yOIdNY v feva 2 27a) AINA 10YLNOT @ avWo3a9 smMReTTONS GECONAT ENNTRUL UNET = THTFROY lS 2450/1985) & face Deasicy Undertable Tube Ie t u ar switch [x] the R500, selece tube 1, 60kY, 200uA and 1 second Also switet [ON] #he “Gecons ¢. tm tha table, inseet_a full oize cassette Ioaded with film and switen che geid TW snd cone O07, In che exposure area, place a phantom of approximately len of hardboard. {£ tardboard is not available, a suitaale Phantom caq be made cut of @ pile of books, a 3 gallon rectangular plastic container filled with water or galt metal blocks such as atuminiumy Cone the tnbe shutcers down tn the size ef che centre cell and ensure thet acreen 2 is selected on the Gecomat and that the density conerel is turned to zero. Jake an exposure and note the mau readout after the exposure, Tf {e {9 leas than 8 mAs mare phantom thickness 1s required. Develop che f1Im and ascertain whether i¢ matehes up te a density of 1, TZ Lhe filw is under exposed, move the Jumper from chamber 1, gereen 2 to a higher density tan auabar. [f the film is over exposed, move this tap to a Lover density Lap numbers Take a series of expocurcs adopting the ahove procedure wntil a film density of 1 ts obtained. © BOY Density Compensation Tube 1. 1 ie 52 Change only the kV to 8OkY and check that relay K9D in the 12500 opens. Take an exposure and chack the developed fila for density L Tf it las becone Lighter, curn 7623 in bosrd U6 counter clockwise. Lf the filu has yone darker cura R623 clockwise. Using this procedure take 2 series of axpooures uuc&? Eiln density Lis obtatued. Do nor forge to add more phanton if actual mas readout drops below 8 miss (2450/7 1265) u InURY Density Compenaation lube 1. range only the k¥ to LOOKY and chesk that relay 126 in the R500 opens. ii Take an exposure aad check the developed film for dencity \. Check also més for useful phantan tifekness. If che fiin hag bescma Lighter, turn RA In board Us couter clockwise. Tf the f1lm has goue dacker tucn RAL4 clockwises Using thie proeadure take 2 cerise of expouures until film deasity Lis ebtained. Att Repeac para. 6.38, Cand Duneil the game Flim density is obtained at OUkV, SOkY, and 1OOkW. NOTE 11 is quite aormal for films co becom blacker berween Lhese stepr expecially when fact screens are used+ % Gone Uensity Gounensation for Serial Radiography TUBE i. Ou the board Ub, connect 3 Jumper between sereen 2, chauher 1 ¢T) and the next lowest density tap cunber to the tap previously aycd for the base densicy. ‘Thaw setting ic generally acceptable in combating the effects of dye or arian, bur may have to be changed by one wore step pending user comments NOTE The taps are not used for overtable techniquens PF Sexeen | and Screen 3 Density Adjusements TUBE | 1 Seesen 1 should genotally give suorcer exposures (approx 25 less mAs). Sereen $ Should generally jive longer exposures (Approx 25% nore mAs). 41 The remaining juepers on Boeed Us for chanhar 1 should therefore be positioned am Follows uatil user preference ts ascertained. Seren 1 Wo donaity teps nuvevically BOWS fcom sereen 2 Sereen 1 (1) Wo density taps nunerically BORY fron seceen ray Screen 3. Two density raps mmerteatiy UP from sereen 2 Seroen 3 (T) Two density taps nnnerically UP from sereen 2 (2450 /1/385) 8a 6.4 Film Density Calibration, Tube 2 a Overtable Bucky. 1 On Che board Us, reposteton the jumper cummecting chauber on screen 2 watrix te the same densicy as tube |. Ji Using this tap as a staeting point, procced ta set up the overtable bucky chamber usiag the came procedura already applied co the undectable tube (see para, 6.3). NOTE The table conc suiteh tape T on eaeh sercen marete are unt used and Lt is also not necessary to check H0kU end 1DUKY compensation again. 8 Chest Bucky 4 Op the board U6, reposition the remaining jumper to connect ehanber 3 on ceteen 2 matrix to the sono dankity tap as the owertable bucky. tL Using ChLs cap a6 a Starting poine proceed to set op che chest stand bucky chamber using the sane procedures already upplied to undertable and overtable buckles. NOTE The table cone switch tap: T on cach screen matrix are not used and it is also not naceswary to check BOKV and :0OkT compensation again. « Finally, foserc che jumpers for chambers 2 and 3 between screen L and matrices to che density caps suggested in para. 6.3 F, e-g+, two down for screen I and tye up for sercen J, 34 (2450/1385) NOTE 1 NOTE 2 NOTE 3 NOTE 4 an DENOTES TEREACE mIRING DENOTES EXISTING WiRING In the Gecamat Control cun in a new were trom NO contect 7 ot RLY/3 and Use 4 spare terminat to connect this wire to termingt 176 in the RECO via Ampick cable C (brown wire), The blue and brown wires from termina 10 & 1) in Geco Control which run through exble A to the: clectronics unit must be disconnected and insulated wih tape. In the Gecn Control terminals 10 8 11 ‘(Dust be jumpered as shown in the wtertace diagram. Disconnect kV Comp pushbuttons to Geco Etectronic Unit. aasefoes) a R500 (Series 2) CONTROL CONSOLE a | R500 ‘Series 2) WALL BOX a sree aes tee woasroa sme oy | eseas ad cus yo I sicen wl vee Heats | ke L._._.,. Oe. GECOMAT 2 ELECTRONIC UNIT GECOMAT 2 CONTROL ST z ee Fe z n s ay | | Se é ipl est) athe Big £ eres | t lh i i. eS aL fe eH 2 Lpts eis nd Soca ont comme ant cane @iucer [ \ ose west rage wv FrOMO SUCE v4 cone 7 FINAL TESTS AND HANDING OVER Final Tests Remove all tomporary eoanecting Linka. Check all modes of control for correct functioning. Take a stepwedge film for reeord purposes and recheck that che correct mAs is indicated Zor cach exposure, Bemove all test equipaent, and ensure that any Links removed tor test purposes are refitted. At exposure counters arc fitted, reeord their readings. Replace ald Control Console, Wall Box end Uf Transforner covers. Handing Over Demoustrate the equipaent to the ueor aad ensure that they know how to use ite Bnsure that she usec has a copy of the Operator's Manual INST. 2449, (2450/1. /385) 57/58 8 RELAY, CONTAGTOR AND MICROSWITCH FUNCTIONS &1 Relays DESTONATTON DRAWLNG: REFERENCE PUNCILON. FAL 247 coil (UE CHARCEOVER) (Nounced on IT tank} RL way? 37608 37608 37603 F6 ¥5/6 F6 Closed ta rotate KT avirch cam plate c/clockwise during tube selection, Open to rotate HT switch cam plate clockwise during ube selection. Changes drive sense of switch motor #1, Fold oa for RM during efclockwise rotation after Tete has opened. SPL 12¥ oft CHUBE THERMAL O72) sPLL SPL/Z SPL/3 37608 37608 37638 Fy Flo Ha 7 Closed when tuhe 48 Selected and Cube thermal switch is closed at safe. Blocks Fluore, peep and expose eouhling ctrewits if tuba Js overbested oc NT switch ts net homed to the eoreect tube. Disconnects the lawps ia the tube/tachnigne push buttuns if the tube Ls overheated or ET owitch is not honed to the correct tubes Not used OAD 24V coll (OVERLOAD) oap/1. oaD/2 oaD/3 37603 37608 3763/8 37608 F2d Fo ass FI Closed Lf not tube overload Blocks the line to the fluoro, prep aad expose control eireuite tf che pre-set kW/tine overload factors ere exceaded. Switches OFF the ready lamp gnd suitehes ON tha overload lamp if the pre-set kW/ting overload Taccors are exceeded. Nnt used. (2650/2/0285) 38 8 RELAY, CONTACTOR AND MICROSWITCH FUNCTIONS &L Relays DESTONATION DAWLNG: REFERENCE PUNE LON Fal 247 coil (EE CHARCEDVEa) (Mounted on UT tank} wo wos? 37608 37608 37608 F6 w5/6 6 Closed to ratate UT swirch cam plate c/clockwi se duzing tuna selection, Open to rotate HT switch cam plate clockwise during eube selentinn. Changes drive sense of switch nator #1, Bold oa for RM during efclockwise rotation after I-16 has opened. SPL 12¥ eoLt CHUBE THERMAL O/2) sPL/L spurs SPL/3 37608 37608 37638 Fy Flo He 7 Closed when tuhe 48 Selected and cube thermal switch is closed at safe. Blocks fluore, prep and expose enahling eirewits if cube is overhested oc UT switch ts not homed 2 the coreeet tube. Disconnects the Leaps ja the tupe/taenniqne push- buttons if the Lube 1s overheated or ET evitch is not honed te che correct, tubes Not used OAD 24v coll (OVERLOAD) oap/1 oaD/2 oaD/3 37603 37608 3763/8 37608 Fad F1o ass FI3 Closed Lf nct tube overload Blocks the Vine to the fluoro, prep aad expose concral eireuite if the pre-set kW/tine overload factore ere exceaded. Switches OFF the ready lamp snd switches ON the overload lamp if the provset kW/tine overload Tactors sre exceeded. Not used. (2450/2/0885) 38 nS LGN TON DRAWING REFEMENGE UNCTLON cf 24 coil (HOCLS SELECT) or cee cess 708 vex 37588 37588, 24 1 Ds. Closed when small focus 1s selected. This normally occurs ag SOmA, 100m and 200nA. Connects smald foens: iadicator lamp shea energised. Cennects Large focus indicator Janp wien de~energised. Supplies ceeontury ol space charge compensation (eansiormer TR? with 110t on anal] foens ané 1409 on large focus. Selects mull Locus filament transformer when Ok and large Fooue filament transforner when orF, EPH 24 eof] (UGH SPER set.80-7) sey tas eea/3 37608, 37638, 37020 37608 22 HUL " Lass Closed when high speed ant requested. Connects high speed Indicator lanp when de-energiseil. Connects 161 to 163 for nomel speed start and 161 to 162 for high speoe start, These ternioals for interface to the MCU, Not useds x9) 46V eat (GECN 90ky) xaogt Kona RvD/s 37608 37620 37608 37008 P19/20 19 rt aL Closed below B0kV. Disconnects 164 From 178 for CHCOMAT interface to secivate bhe low k¥ coapensation. Not used. Noe useds 60 (2650/1/383) DRS CUNATTON DRARING REFERENCH FUNCTION KL23 24v eat (GEO 125K) w125/1 1235/2 KL2s/3 37608 37626 37603 37603 Fao fe Closed below 10UKV. Discennects 164 From 179 for GECONAT intertaca tp activate the igh kV comensation, Not used. Not used. sca 24¥ coil CFLUORO ORDER seas sca/3 37608 37618 37620 37558 6 15 ay Closes whea fluora is ordered feon footswitch provided techaigue | ie selaeted and the fluoro timer is reset with ne radingeapay prep comand present. Energises fluoro evatactor 808 when ON, Spare Interface contacts with 19! normully elesed te 192 and 197 makiag co 193 during fluoro. Change 10mt mA meter scale from Line indication to fluora mA indication. EGE 24V coil PREP ORDER) POEs ee/2 voz/3 37608 37618 37608 27608 ¥AS ar BS #5. Closes when radiography prep Is ordered, provided that the Eubo overload, ‘ube thermal, tube selected Lights and tube selection circulte are enabled safe. Blocks fluoro sec contactor for 0.80 after prep and expose Not used. Wot used. «2450/2/0486) 61 DESIGNATION DRAWING REFERENCE FUNCTION oR 2uv eos1 {FALSE mAs DETECTOR) PoB/L Pon /2 PoE sa 37608 a6 37608 37565 Fb cio Pia Closes when prep is commanded only if BEE is not indfeared on the As displey. Blocks PGA, PGC and MA during prep if « feloe ate te selected. Tf mas 1s normal 1t also freezes the Lina Autowatic Detver priox to expose and also prepares che axposure counter to operate view exposure stertay Blocks flucte SCA relay and flora display during rad prep and expose. Change aA acter scsles frou 20 to 1000mA (Fron fluoroscopy to radiography). EDD 24v coil (PREP DELAY) EDI/t por/2 eor/3 37608 27618 37620 37578 F23 giz 110 Closes 0.88 after prep is ‘commanded. Enables the radiography timer st the end of the prep boost delay tive, provided that the erator detector circuits are cloned, the tube filenent ip intact, no fluoro ts commaiced and the operators expose button is pressed. Spare NO contact connected to 139 and 140 for interface purposes. Change uA meter Input from Line indicatien te mA radiography. 62 (2450/2/0486) bzSLonarTroN DRAWING REFERHUCE wUNCTTON SE 6V coil 375au Lo Closes during radiography CFLLAMENT SRUIRTTY) prep and expose 4f current in pravewt La the tobe Titanenc. Bed 37elw G18 Blocks the radiography exposure timer if the Lube Filsmont te oped cirenited. sera a758u La Noe nsed, SEs3 S75BB Das Hot used. AHP L.5 A Cold aol G18 @lcses to indicate that (STATOR MATH) current is present In che stator main winding ducing preparation run ups ae? a7elB 612 Blocks SIA contactor if ntarne main wading ¢3 open ducing prep. ange BToLe Blocks exposure riaor stazt LLoe if AMP docs ant open efter the end of th: prep boost pertad. anys 37o18 g17/18 Delays uhangeover from 220 status start to 40v stator run to minimise areing on STAW3 contact. wee 2200 coik 37618 cls. (STATOR CAPACLTOR) present across stator phase shLicing C4 and ch Auring stator rin up. woes 37618 G1? Blovks the exposnee Liner stact Line 3f the phase sotre capacitor is Shart cfrenfied or 1 the supacicive winding goee open circuit. ator wecys ele cla Nor used. yee! a761B Gls Nor vod. (2450/1385) 63 bESIGIATION DRAAING REFERENCE FUY ION ER LWW entt (STATOR 90? WINDING) ueeyy zeup2 gocs3 37618 37618 aoe 37618 039 ez 14/15 @1asis Closes te decece voltage seeoes the ecator capacitive vioding during prep boosts Blocks the exposure tiger etart lisa if the stator capacitive winding 18 ahort circuited, Hot used. Hor used. RAY 24V aod (ERPOSE URUER) kant Ray RAN 37608 37618 37568 37608 Uy 13 Plo Closes ven radiography exposures are counanded by the operators control button provided that the correct technique is selected and that che tube overload, tube Lhecmal, tube reldctnd Lights ané tube selection elreuits are enabled safc Closes one vf the interlosks iq che exposure conmand 1192 ta the radiography exposure timers Open cirouits che external was mover Link when a radicgtaphy exsosure is commanded. Mot vaed. 64 2450/1/385) DESTONATION DRAWING REFERENCE FUKSTION RIE 24V coil {EXPOSE ON) RTEst puK se? Rre/3 37608 37618 37638 3762e FIB en a/6 13 Dloses afeer radlopraphy Hiner comands an exposure via MU/L. Coaplecor acnmon retura for exposure councers via POB/L prep und SCC/6 fluoro. Geangcts 43¥ te radiography X-rays ON tanp fer a period long eaough co give adequate visual Indication ox short exposures. Spare NO contact counceted Fo 188 and 188 tor interface purposes. NOTE: Thig renaing closed for a short peciod after each expoaure due te the slugging action of XL. MA 220N coil (QMO PREP) mas maya Mal? 7612 37602 37612 37618 ou F26 an2 12 Closes on prop when Tube 4 fe aeleated. Completes circuit for relays MB, MC and HD during prep if TQN 6 and TUBE 4 are aclacted. Not used. Not used. HB 26 coil (OAIHO 100A) was) MB/2 Ms /9 37605 37588 37588 37608 25/26 pi2/13 v3 me Closes on prep when TOK 5, TUBE G and 100ma are selected. Consects supply to 1000 tapping end on R11 for adjusting amsll faeus on Maun. Connects supply ta 100mA tapping bend of R2/ foe adjusting large focus oo Manno. Wor used (2450/2/ 0486) 6 bEstUNATION DEARING KEFERENGE FUNCTION Ma 24¥ coil (iuaaso 200ma) nosh mep2 vers 37603 37588 37583 37608 F26 els bla 20 Closes om prep when TON 6, TWBE 4 and 20008 is selected. Connects supply tapping band of adjusting smal Mannie. Coanecta supply tapping band of adjusting Large Manno. Nor deed. co 200mA, RIL €or forus ua be 200mA 27 Eve Focus on ab 24 coil OLN 3oOHA) MD/1 wo/2 vo/3 37608 3758/8 37588 ‘37608 28 pias bia p22 Cinses on prep when 1QN 6, TURE 4 snd 00nA io pelected. Connects supply tapping band of adjusting small Mame. Connects supply Tapping pend of adjusting large Not used. bo 300 mA KIL for focus on to 300m k2? for Focue on DK 24 oid (TIMER) (PCB 3474 93554) BEAL 2554 37598 37678 3554 HL 2 Closes if there prinary current is excess ia the HT tank during exposure or if there fe X-rays exposure. ourside Reuovas supply trom 1Lae contactor Cl to isulare tho RBUD from the room supply Xl, DAV eot} (CLIMFR) (pce 3371 03555) xL/L 3554 3534 37608 8 Clowes during radiography exposures and is held OW azter termination for short period to peratc adequate ewitebing tine during the shortest exposure durations. tnergises relay RTE during radLlography exposures. 66 (2450/1 /385) 8.2 Gentactors DESTGUATION DRANING REFERENCE, BUNCIION OL 24 60Rz eott 367A Pz Cloaes whan ON switch I-15 (LINE oWTACTOR) is pressed via OFF suttch I-16 and breaker relay contaet BR/1. L/L ae 367A. 2 Not used. couse BETA BD Hold on for CL after TH18 ig released. ost 767A BB Comneets LI to xed phase 3755B az Line Compensator Switeh : Pole. cLsa 37678 2 Not used. Lys. 3767A 3 Connects L3 to blue phase 37558 AZ Lime change voltage Links cus nfo 3767A 2 Not used. ly? ae 37672 Not used. Pca 2z0y 37e1B co Closes end cf pres boost {SCR ISOLATOR) except {f a HS NCU ts feted. In this case it closer during prop. PoA/L 3761B cig Net useds PGA/2 3761B Gc Wot used. Pasa 3755B a? Connects TRIA Avto to SoRs. PoAl4 a761B G10 Not uaed. POA/S 3795B A? Completes RT primacy eixeuit to TRIA Auto. Poale 3761B Go Not used. Poa? nfe 2761 10 Not used. (2450/2/0486) 7

You might also like