Installation Instructions
Orion Rotary
Chair
Micromedical by Interacoustics
D-0120794-C – 2019/11
Copyright © November 1, 2019 by Interacoustics. All rights reserved. Information in this document is subject
to change without notice. Companies, names, and data used in example herein are fictitious unless otherwise
noted. No part of this document may be reproduced or transmitted in any form or by any means, electronic or
mechanical, for any purpose, without express written permission of Interacoustics A/S, Micromedical
Technologies or its licensees.
Windows® is a registered trademark of the Microsoft Corporation, registered in the United States and other
countries. FireWire® is a registered trademark of the Apple Inc., registered in the United States and other
countries. Other trademarks are the property of their respective owners.
Table of Contents
1 INTRODUCTION ........................................................................................................................... 1
1.1 About this manual ................................................................................................................ 1
1.2 Intended use ........................................................................................................................ 1
1.3 Contraindications ................................................................................................................. 1
1.4 Environmental Conditions .................................................................................................... 1
1.5 Required tools...................................................................................................................... 1
1.6 Marking ................................................................................................................................ 2
1.3 Contraindications
More intensive rotational chair tests may be contraindicated in patients less than 5 years of age or greater
than 110 years of age.
D-0120794-C – 2019/11
Orion Rotary Chair Installation Instructions Page 1
1.6 Marking
The following marking can be found on the instrument:
Symbol Explanation
An applied part that includes a patient connection that is intended to deliver
electrical energy or an electrophysiological signal to or from the patient shall
be a Type BF part. An EOG amplifier is considered a Type BF part.
WEEE (EU-directive).
This symbol indicates that when the end-user wishes to discard this product,
it must be sent to separate collection facilities for recovery and recycling.
Failing to do so may endanger the environment.
0123
Manufacturer.
Year of Manufacture.
Do not re-use.
Parts like foam cushions and irrigator tips are for single use only.
Do not push symbol used with components that could easily tip over, such
as the Digital Light Bar
Chinese RoHS compliance standard where the product contains less than
the maximum concentration value of lead, mercury, cadmium, hexavalent
chromium, polybrominated biphenyls, and polybrominated diphenyl ethers.
D-0120794-C – 2019/11
Orion Rotary Chair Installation Instructions Page 2
2 Unpacking and inspection
2.1 Inspection of equipment
Check shipping containers and contents for damage
When the system is received, please check the shipping crate and pallet for rough handling and damage. If
damage has occurred, shipping containers should be kept until the contents of the shipment have been
checked mechanically and electrically.
If the instrument is found to be faulty, please contact your local distributor. Keep the shipping material for the
carrier’s inspection and insurance claim.
Chair in crate Chair wrapped on pallet chair with head support & cables in box
D-0120794-C – 2019/11
Orion Rotary Chair Installation Instructions Page 3
Chair pallet with shipping cover removed Ramp secured to base
Chair is strapped to a 2-wheel dolly inside crate. A wooden ramp, that is part of the equipment on the pallet,
should be fastened with screws to the chair pallet to form a ramp. Tilt the chair and roll it off the pallet and
down the ramp. Position the chair in the exam room such that the cable opening in the chair base is directed
toward the equipment cart.
D-0120794-C – 2019/11
Orion Rotary Chair Installation Instructions Page 4
3 Computer configuration
3.1 Tower configuration
The small form factor tower computer will come with an all-in-one
monitor and computer stand. The computer slides into the back of the
stand.
• The goggle cameras plug into the front USB ports of the
computer.
• The speaker bar sits under the touch screen monitor and the
room camera sits above the monitor. Both the speaker bar and
room camera connect to the two USB ports on the back of the
monitor.
• The monitor’s USB hub cable and DisplayPort cable connect to
the back of the computer.
• The TV connects to the computer either by HDMI (if available)
or by a DisplayPort to HDMI adapter.
• The RF Remote plugs into the USB port on the keyboard.
• The Orion rotary chair’s USB cable will connect to the back of the computer.
• Printer USB cable connects to the back of the computer.
• Power cables for the monitor and tower are provided by the stand, and the stand is powered by an
external power cable connected to the isolation transformer.
Left to Right: Power to Isolation Transformer, Power Switch, USB to PC, E-stop
D-0120794-C – 2019/11
Orion Rotary Chair Installation Instructions Page 5
3.4 Isolation Transformer configuration
• Orion Reclining Chair uses 110V for operation. The Isolation
Transformer is set from the factory for the intended source voltage.
• Specialized power cables connect the isolation transformer to the
computer and peripherals.
• Laptop computer and printer may connect to the external power
strip or to power mains.
Example of a chair equipment cart with Aqua Stim and water tank on the same shelf
D-0120794-C – 2019/11
Orion Rotary Chair Installation Instructions Page 6
4 Anchoring chair to the floor
4.1 Computer setup
The chair base must be anchored to maintain proper position of the chair at the correct distance from the
wall mounted TV. The TV should be mounted on the wall before anchoring the chair to the floor.
For this procedure the installer will need to use the following additional items:
1. Hammer drill
2. Hammer
3. Adjustable wrench or socket wrench
4. Ear plugs
5. Disposable dust mask
6. Safety glasses
D-0120794-C – 2019/11
Orion Rotary Chair Installation Instructions Page 7
3. The Orion Reclining Chair should be situated away from the TV so the patient seated in the chair
(center of chair base) is between 30” and 39” (0.75 and 1.00 m) from the front panel of the TV (not
the wall). Adjacent walls should have the same clearance to the base of the Orion Reclining Chair.
4. After Chair has been carefully placed in the proper position, mark the area where the three holes are
to be drilled.
5. Move the chair to the side, or keep the chair in place and drill the holes using the base as the
template.
Secure the drill bit in the chuck of the hammer drill. Place the drill in the middle of the target hole. Keep the
drill bit vertical. Be careful not to drill at an angle, because this may cause difficulty in keeping the anchor
screw head flush on the washer.
When drilling in concrete cement dust will be produced. The cement dust must be carefully removed from the
area after drilling. Sweep the dust and pick up remaining particles with a water dampened cloth. Do not
breath the dust, use a mouth cover as appropriate. Wear ear plugs and safety glasses while drilling into the
concrete floor and cleaning up the dust.
Ceramic Tile
If the floor is ceramic tile, then obtain a different drill bit appropriate for drilling through ceramic tile.
Vinyl Flooring
If the floor is vinyl tile over concrete, the anchor holes can be drilled with the supplied bit.
Wood Floor
D-0120794-C – 2019/11
Orion Rotary Chair Installation Instructions Page 8
If the floor is hard wood, then consult with the building owner about how they want to mount the chair.
Consider lag screws, toggle bolts or molly anchors.
Insert the split anchor end into the drilled hole. Make sure the anchor is below the concrete surface.
Complete this process for all three anchors. Move the chair back over the anchors.
Stack the lag screw with the washers, first the small steel washer then the large black
washer. Place the screw with washers into the anchor and tighten firmly with a wrench.
The lag screw will expand the anchor in the concrete hole.
If the anchor spins in the hole, remove the anchor, clean out the concrete dust and try
again. If the anchor does not hold, contact Interacoustics or Micromedical Technologies for
further instructions.
D-0120794-C – 2019/11
Orion Rotary Chair Installation Instructions Page 9