Professional Documents
Culture Documents
OPTICAL PROPERTIES
Optical properties of a substance are depends on the directional distribution of light after
impacting on the sample. Molecular structure, morphology, type and amount of
additives and processing conditions decide the optical behaviour of the plastic
materials. Additives like optical brighteners and nucleating agents can be used to
increase the surface optica and transmission properties respectively. The sensation of a
typical colour can be explained by the selective absorption of radiations of certain wave
lengths in the visible region of the spectrum.
(a) Technical use is made of polymers as optical materials in lenses or light fixtures.
(b) The relatively large change of refractive index with deformation (Photoelasticity)
make polymers ideal materials for qualitative determination of stress deformation
fields.
General criteria for the degree of transparency of commercial polymeric materials may
be classified according to the following:-
Significance:
Refractive index can be used to calculate the bend angle of a beam of light as it
passes from one transparent medium to another.
The index of refraction is a measure is how much light bends when it hits an object.
If the object has in index of refraction near that of air, the object will be transparent.
Glass, polycarbonate, polystyrene, and acrylic all have indices of refraction close to
air.
Thus, the index of refraction is used in designing optical lenses and as a quality
control check of transparent plastics.
Calculation:
The value of Refractive index is calculated by dividing the actual thickness of the
specimen to apparent thickness of the specimen.
Definition: Light Transmittance is defined as the ratio of transmitted light to the incident
light.
Plastics differ in their ability to transmit light.
Transparent plastics allows light to pass throw them
Opaque plastics do not allow any light to pass through.
Translucent plastics allow some light to pass through them, but the images are
cloudy and unfocused.
Semi-opaque plastics allow enough light to pass through them and able to make out
shadows and vague outlines.
Significance:
Light that is scattered upon passing through a film or sheet of a material can
produce a hazy or smoky field, when objects are viewed through the material.
Haze luminous:– transmittance value are especially useful for quality control &
specification purpose.
Window glazing requires low haze & light transmittance light fixture requires both
high total transmittance & haze.
Test Methods: ASTM D 1003, ASTM D 1746, ASTM D 1494
Specimen Dimension: Approximately 50mm in diameter.
Conditioning: Test specimens conditioned at 23 2C and 50 5% relative humidity
for 40 hours prior to the test.
Test Procedure:
The test specimen must be large enough to cover the aperture, but small enough to
be tangent to the sphere wall.
A disc of 50 mm in diameter is most commonly used.
The test is conducted by taking four different consecutive readings and measuring
the photocell output as follows:
Td = [ T4 – T3 (T2 / T1)] / T1
Haze % = Td / Tt x 100
Factor Influencing:
Absorption of water
Temperature: Transmission is inversely proportional to the temperature of the
medium.
Moulding conditions
Additives used
Irregularities within or at the surface of the material due to molecular defects.
Gloss Meter
It consists of a source optical assembly which houses an incandescent light source,
a condenser lens and a projection or source lens. The light beam is directed towards
the specimen at a specified angle.
Three basic angle of incidence 60, 20 & 85 are used for specular gloss
measurement.
A sensitive photo detector collects the reflected light and generates an electrical
signal that is amplified to energize an analog or digital panel meter to display the
value in gloss unit.
Test Procedure:
The instrument is turned on and placed on a black glass primary standard. The control
knob is adjusted so that the meter indicates the value assigned to the primary standard.
Next, the sensor is placed on the specimen surface and the gloss value is read directly
from the analog or digital display.
Calculation:
The instrument is calibrated during the operation at sufficiently frequent intervals to
assure that the instrument value is practically constant. If at any time an instrument fails
to repeat readings of the standard to within 2 percent of the prior setting the intervening
result should be rejected. The gloss value is directly read on the instrument.
Factor Influencing:-
Surface property
Temperature
Humidity
Polymer molecular weight, Melt pressure, injection pressure & mould
temperature also influences the gloss of a product.
3.4 COLOUR
Definition: The colour we see in an object is not in the objects but in our brains. This
brings a primary definition of colour which is defined as brain’s perception of light that
has been modified by the absorption of parts of its rays.
Significance:
The colourability of plastics is much higher than that of other materials.
Colours can be incorporated throughout the mass of the material.
An aesthetic component can be incorporated with the use of a wide range of dyes
and pigments.
Colours are also importantly used as a very simple means of part identifications.
o Colour ranges from dark to light – black is darkest, white is lightest, and grey is in
between. These are called neutral colours. This aspect of colour is termed
‘value’ or ‘lightness’.
o There is another basis of colour difference – red differs from blue, green or
yellow. These distinctions are called ‘hue’. Hue is defined as the attribute of
colour perception by means of which an object is judged to be red, yellow, green,
blue, violet or their intermediates.
o Another dimension of colour is ‘saturation’ or ‘chroma’. This is defined as the
attribute of colour perception that expresses the degree of departure from grey of
the same lightness.
Test Method: ASTM D 2244, ASTM D 1925, ASTM D 172, ASTM E 308, ISO 3557
ISO 4582
o
Conditioning: The test specimen is conditioned at 23 + 2 C and 50 + 5% relative
humidity for 40 hours prior to testing.
Equipment:
Equipment for applications dealing with colour and colour related properties falls into
two groups:
(i) Visual Devices: - When the human eye serves in a primary role in the
measurement pro-processes.
(ii) Instrumental: - When the human eye serves in a supporting role only. Instruments
for colour are further divided into two groups.
a) Tristimulus colorimeters for measurement of materials on either a reflectance or
transmittance basis and with measurements weighted with the standard observer
functions.
b) Reflectometers: - Reflectometers of a wide variety for the measurement of light
energy on a reflectance basis. (In selected cases the instrument is also available for
transmittance measurement). Usually reflectometer provides a single purpose
measurement such as total reflectance, lightness or yellowness, etc.
Test Procedure:
a) Visual Assessment of Colour and Appearance: -
The visual examination and evaluation of colour is the most commonly used
technique in the plastics industry:-
The two essential elements of visual colour evaluation are: -
(1) The viewing of both the sample and a standard by an experienced observer.
(2) The use of a predetermined light source at a fixed distance.
Instrumental:
Where: X / X0, Y / Y0, Z / Z0 > 0.01 and X0, Y0, Z0 define the colour of the
nominally white object-colour stimulus.
Calculation
Colorimeters & spectrophotometers calculate yellowness index as well as whiteness
index and is directly read by instrument.
Factor Influencing:-
Type of illuminant
Wave length of light
Angle of illumination
3.5 CLARITY
0
Definition: Clarity is defined as ‘Light flux deviating less than 2.5 from the normal, on
the average’.
Significance:
The Clarity is the distinctness with object appears when viewed through a film. Clarity,
which may also be prescribed as the quality of image formation through a sheet, ends
upon the linearity of the passage of light ways through material. These deflections are
much small than those registered in haze measurements. While haze measurement
depends an wide-angle scattering, clarity is determined by small angle scattering Wide-
and small-angle scattering are not directly related to each other, i.e. haze
measurements cannot provide information about the clarity of a specimen and vice
versa.
Specimen: Specimen may be in form of sheet, or film.
o
Conditioning: The specimen is conditioned at 23 + 2 C and 50 + 5% relative humidity.
Equipment: It consists of clarity sensing unit and photometric unit. Clarity sensing unit
consist of a light source, focusing lens, specimen holder & test cell.
Test Procedure:
The equipment is turned on. With no sample in position the unit is standardized until a
reading of 100.0 relative to air is obtained. If a calibrated standard is to be used, it is
inserted in the specimen holder, and the instrument brought to balance upon the value
assigned to the standard. The beam is blocked completely with a piece of block flet of
paper. ‘Zero adj’ knob is then turned until reading of zero is obtained. The specimen is
placed in the specimen holder. The reading is recorded as percentage clarity directly
from the measurement unit. Reading at several different position or orientation may be
taken and the results averaged for a more representative determination of clarity.
Factor Influencing:
Humidity
Wavelength
Temperature