D. CHECK THE SKIN TEMPERATURE ALARM


Inset two skin temperature sensors into the socket of skin sensor 1 and skin sensor 2
separately as the following figure shows (Connecting socket of sensor is located on the left side of
bed), the arrow sigh of the plug should aim to the opening on the sensor socket so that the sensor is
inserted correctly.

Socket of Skin Sensor 1 Socket of Skin Sensor 2

Sensor socket panel Alarm light

Arrow Opening Opening

Socket of Skin Sensor 1 Socket of Skin Sensor 2


Skin temperature sensor 1: central skin temperature sensor connector
Skin temperature sensor 2: external skin temperature sensor connector
Alarm light: when there is one of the following conditions, the alarm indicator will flash with continues
audible sound.
� The internal independent probe of mattress-temp sensor is failure;
� The mattress over-temp alarm;
� The internal independent probe of skin-temp sensor is failure;
� In baby mode, the skin over-temp alarm;
� If relevant system failure alarm is activated on the device, please stop using it and
hand it over to the qualified person to repair it.

1. Insert or pull out the skin sensor, you must hold the plug of skin sensor, pulling
the leads is forbidden.
2. Please do not bend the connection of sensor.

In baby mode, pull off the skin temperature sensor 1, the sensor failure alarm should active.
Both controller and panel of sensor socket will flash with continues sound, and controller's message
Skin temperature Sensor Fault
status bar of HKN-2001 will show "S ault", and the controller's message
Skin temp. sensor
indicator of HKN-2001L will show "S ensor". Pull off the skin temperature sensor 2, the
sensor 2 will display "--.-", the device is without alarm indication.