You are on page 1of 8

CORTECS 2.

7 µm Columns

CONT ENT S I. INT RODUC T ION


I. INT RODUC T ION Thank you for choosing a CORTECS Column. CORTECS 2.7 µm
Columns are based on the same solid-core particle technology
II. G E T T ING STA RT E D as CORTECS UPLC® 1.6 µm Columns, enabling seamless method
a. Column Connection transfer between UPLC, UHPLC, and HPLC platforms. All CORTECS
b. Column Installation Columns are manufactured in cGMP, ISO 9001:2000 certified
c. Minimizing Band Spread Volume facilities. Strict quality procedures for the manufacture of the
d. Column Equilibration solid-core particle, the bonded phase synthesis, and the column
e. eCord™ Installation packing result in a high quality and reproducible final product. Every
f. Functional Tests for Benchmarking a New Column column is individually tested and a Performance Chromatogram
g. VanGuard™ Cartridge Columns and Certificate of Batch Analysis are provided on the eCord
Intelligent Chip. The eCord is available on all 2.1 mm and
III. COLUMN US E 3.0 mm column dimensions.
a. Sample Preparation
b. pH Range CORTECS 2.7 µm Columns will exhibit maximum chromatographic
c. Solvents performance when used on a member of the ACQUITY UPLC® System
d. Pressure family since they were created and designed to operate with it.
e. Temperature

IV. COLUMN C L EANING, R EG EN E RAT ION,


AND STO RAG E
a. Cleaning and Regeneration
b. Storage for Reversed-Phase and HILIC Columns

V. eCO RD INT E L LIG ENT C HI P T EC HNOLOGY


a. Introduction
b. Installation
c. Column Use Information

V I. A DDIT IONA L INFO RMAT ION


a. Tips for Maximizing Column Lifetime
b. Getting Started with CORTECS® HILIC Columns
c. Troubleshooting Questions

V II. C AUT IONA RY NOT E


II. G E T T ING STA RT E D c. Minimizing Band Spread Volume
Each CORTECS 2.7 µm Column comes with a Certificate of Band spreading is the measurement of the system dispersion,
Analysis, which includes the bonded phase batch number and which impacts the chromatographic performance. Internal tubing
the analytical test results for the unbonded and bonded particle. diameter and fluidic connections can significantly impact system
T he included Performance Test Chromatogram summarizes band spreading and chromatographic performance. Larger
the performance of each individual column and provides batch tubing diameters cause excessive peak broadening and reduced
number, column serial number, USP plate count, USP tailing sensitivity (Figure 1).
factor, retention factor, and chromatographic test conditions.
These data should be recorded and stored for future reference. 0.005 inches
When available, the information can be accessed via the
ACQUITY UPLC Console using the attached eCord. 0.020 inches
0.040 inches

a. Column Connection
CORTECS 2.7 µm Columns are designed to operate on any HPLC,
UHPLC, or UPLC system. Due to the absence of an industry
standard, be aware that the type of fittings and connection on Diluted/distorted sample band

each system will vary by manufacturer. All tubing connections


Figure 1. Impact of tubing diameter on band spread.
must be reseated to match the new column when it is installed.
If the style of the column end fitting does not properly match the d. Column Equilibration
system connections, the chromatographic performance of your CORTECS 2.7 µm Columns are shipped in 100% acetonitrile. It is
new column will be negatively impacted, or leaking can occur. important to ensure mobile phase compatibility before changing
to a different mobile phase system. To avoid precipitating mobile
b. Column Installation phase buffers within the column or system, flush the column with
Note. The flow rates given in the procedure below are described for five column volumes of a water/organic solvent mixture, using
a 2.1 mm I.D. column. Adjust the flow rate up or down based on the the same, or lower, solvent content as in the desired buffered
column pressure and system limits. mobile phase (for example, flush the column and system with 60%
1. Purge the pumping system of any buffer-containing mobile methanol in water prior to introducing 60% methanol/40% buffer
phases using 100% HPLC-grade water. mobile phase).

2. Purge the pumping system to 100% organic mobile phase Note. If mobile phase additives (i.e., ion-pairing reagents) are
(methanol or acetonitrile). present in low concentrations (<0.2% v/v), 100 to 200 column
3. Connect the inlet of the column to the chromatographic system. volumes may be required for complete equilibration. In addition,
4. Flush the column with 100% organic mobile phase (methanol mobile phases that contain formate (i.e., ammonium formate,
or acetonitrile) by setting the pump flow rate to 0.1 mL/min. formic acid) may require extended equilibration times.
Increase the pump flow rate to 0.5 mL/min over 5 minutes.
For reversed-phase separations, equilibrate the column with a
5. Once the mobile phase is flowing from the column outlet, stop
minimum of 10 column volumes of the mobile phase to be used
the flow.
(refer to Table 1 for a list of column volumes). The column may be
6. Attach the column outlet to the detector. This prevents air
considered fully equilibrated once a constant backpressure and a
from entering the detector flow cell.
stable detector baseline is achieved.
7. Increase the flow rate as described in step 4.

8. Monitor until a steady backpressure and baseline have


been achieved.

CORT ECS 2.7 µm Columns 2


1. Run an initial benchmark test to monitor key performance
Table 1. Empty Column Volumes (mL)
criteria such as retention time, peak area, peak tailing,
I.D.
resolution, response, system pressure, etc.
Column length (mm) 1.0 mm 2.1 mm 3.0 mm 4.6 mm
2. Repeat test daily to track column and system performance
30 0.024 0.10 0.21 0.50
over time.
50 0.039 0.17 0.35 0.83
For more information on benchmarking performance, visit
75 0.059 0.26 0.53 1.25 www.waters.com/QCRM
100 0.079 0.35 0.71 1.66
Note. If an analyte mixture found in the Performance Test
150 0.12 0.53 1.07 2.49 Chromatogram is used to benchmark, the isocratic efficiencies
measured in your laboratory may be less than those given on
For CORTECS HILIC Columns, flush with 50 column volumes of the Waters Performance Test Chromatogram. This is normal
50:50 acetonitrile/water with 10 mM final buffer concentration. and expected. The Waters isocratic column testing systems
have been modified in order to achieve extremely low system
Prior to the first injection, equilibrate with 20 column volumes
dispersion. This presents a more challenging test of how well
of initial mobile phase conditions. If gradient conditions are the column was packed. This also guarantees the highest
used, equilibrate with 8–10 column volumes between injections. quality packed column. These special testing systems have
Failure to appropriately equilibrate the column could results in been modified to such an extent that they are not commercially
viable and have limited method flexibility other than isocratic
drifting retention times. See “Getting Started with CORTECS HILIC
column testing.
Columns” under “Additional Information”.
3. Determine the number of theoretical plates (N) and use this
value for periodic comparisons.
e. eCord Installation
4. Repeat the test periodically to track column performance
eCord technology represents a significant advancement in column
over time. Slight variations may be obtained on different
usage tracking management when connected to an ACQUITY UPLC LC systems due to the quality of the connections, operating
System. The eCord Intelligent Chip can be read by connecting the environment, system electronics, reagent quality, column
yellow fob to the reader/writer located on the right hand side of condition, and operator technique.
the ACQUITY UPLC column heater module. Embedded information
such as the column manufacturing QC data and Certificates of g. VanGuard Cartridge Columns
Analysis may be accessed via the ACQUITY UPLC console. The VanGuard Cartridge Columns are a universal solution to extend all
eCord is available on all 2.1 mm and 3.0 mm column dimensions. HPLC, UPLC and UHPLC CORTECS column lifetimes. The chemistries
offered for VanGuard Cartridge Columns are identical to the available
f. Functional Tests for Benchmarking a New Column CORTECS chemistries. The reusable holder is designed with an
Waters recommends performing a benchmarking test upon receipt adjustable ferrule so that it can be attached to any column inlet.
of your column and throughout lifetime usage. By using a standard VanGuard Cartridge Columns come in two sizes, 2.1 mm I.D. for
mix, such as the appropriate QC Reference Material, you can: columns I.D.’s 3.0. mm or less, and 3.9 mm I.D. for column I.D.’s 3.9 mm
or greater.
■■ Verify the performance of your column upon receipt.
■■ Monitor the health of your column and system over time.
■■ Troubleshoot separation difficulties that arise.

The Neutrals QC Reference Material (p/n 186006360) or HILIC QC


Reference Material (p/n 186007226) are appropriate mixtures with
which to benchmark your CORTECS Column and system. Other QC
reference materials are available depending on the application and
detection mode.

CORT ECS 2.7 µm Columns 3


VanGuard Cartridge Column Installation Instructions
VanGuard Cartridge Columns are recommended to provided superior column protection for all HPLC and UHPLC columns up to
4.6 mm inner diameter.

VanGuard Cartridge Columns require the VanGuard Cartridge Holder (p/n 186007949) for proper assembly.

Cartridge Holder Parts list


Two Piece Ferrule Holder Fitting Protective Plug Tools required
/ " wrench
5 16

/ " wrench
7 16

/ " wrench
3 8
O-Ring Compression Screw Transition Tube Assembly

ATTENTION: READ AND 1. Remove protective o-ring. 2. T hread the holder assembly 3. With holder on top, tighten holder
finger tight into the to the column inlet using wrenches
UNDERSTAND ALL ASSEMBLY Keep assembly upright to
keep ferrule from falling off. column inlet. (1/4 turn after finger tight).
INSTRUCTIONS PRIOR TO USE.
Hold assembly securely
DO NOT remove the protective Two Piece Ferrule to prevent the holder Ensure that the
plug from the holder assembly from slipping from holder is firmly
the column inlet. seated into the
until Step 4. T he plug is used to column inlet before
seat the tubing securely prior to setting the ferrule.
final assembly. HPLC column

In order to ensure void-free


and leak-free connections, the ⁄ " wrench
5 16

VanGuard Cartridge Holder is


shipped with the ferrule NOT
permanently attached. Care must
be taken when assembling the 4. Remove plastic plug. 5. Insert guard cartridge 6. Tighten guard cartridge.
holder to the column. into holder.
Holder housing should
7 16
⁄ " wrench
rotate freely until
cartridge is tightened.

⁄ " wrench
38

Guard cartridge

Guard Cartridge Replacement Instructions

1. Loosen guard cartridge 2. Insert new guard cartridge 3. Tighten guard cartridge.
and remove from holder. into holder. 7 16
⁄ " wrench
The guard cartridge can be replaced
without removing the holder from
the column inlet.

⁄ " wrench
7 16

⁄ " wrench
38

38
⁄ " wrench

CORT ECS 2.7 µm Columns 4


III. COLUMN US E 3. If the sample is not prepared in the mobile phase, ensure that
To ensure the continued high performance of CORTECS 2.7 µm the sample, solvent, and mobile phases are miscible in order to
Columns, follow these guidelines: avoid sample precipitation or buffer precipitation.
4. Filter sample with a 0.2 µm membrane to remove particulates.
a. Sample Preparation If the sample is dissolved in a solvent that contains an organic
1. Sample impurities and/or particulates often contribute modifier (e.g., acetonitrile, methanol, etc.) ensure that the
to column contamination. One option to avoid column membrane material is compatible with the solvents in use.
contamination is to use Oasis ® or Sep-Pak ® solid-phase Alternatively, centrifuge the sample for 20 minutes at
extraction (SPE) devices. To select the appropriate sorbent for 8,000 RPM, followed by the transfer of the supernatant
a specific sample type, visit www.waters.com/sampleprep to an appropriate vial.
2. It is preferable to prepare the sample in the initial mobile 5. For hydrophilic-interaction chromatography (HILIC) separations,
phase conditions or a weaker solvent for the best peak shape the samples must be prepared in a high percentage of organic
and sensitivity. solvent (e.g., 95% acetonitrile). See “Getting Started with
CORTECS HILIC Columns” under “Additional Information”.

Table 2: CORTECS Chemistry Characteristics


Name of Surface charge End cap Carbon Ligand pH Temp.
Ligand type
column modification style load density limits limits
C18 Trifunctional C18 None Proprietary 6.6% 2.7 µmol/m2 2–8 45 °C
C18 + Trifunctional C18 + Proprietary 5.7% 2.4 µmol/m 2
2–8 45 °C
C8 Trifunctional C 8 None Proprietary 4.5% 3.4 µmol/m 2
2–8 45 °C
T3 Trifunctional C18 None Proprietary 4.7% 1.6 µmol/m 2
2–8 45 °C
Monofunctional
Shield RP18 None Proprietary 6.4% 3.2 µmol/m2 2–8 45 °C
Embedded Polar C18
Phenyl Trifunctional phenyl None Proprietary 5.9% 3.2 µmol/m2 2–8 45 °C
HILIC None None None Unbonded n/a 1–5 45 °C

Table 3. Mobile Phase Buffer Recommendations for CORTECS 2.7 µm Columns


Volatility Used for
Additive/Buffer pK a Buffer range Comments
(±1 pH unit) mass spec
Ion pair additive, can suppress MS signal, used in
TFA 0.30 Volatile Yes
the 0.02–0.10% range.
Maximum buffering obtained when used with
Acetic Acid 4.76 Volatile Yes
ammonium acetate salt. Used in 0.1–1.0% range.
Maximum buffering obtained when used with
Formic Acid 3.75 Volatile Yes
ammonium formate salt. Used in 0.1–1.0% range.
Acetate Used in the 1–10 mM range. Note that sodium or
4.76 3.76–5.76 Volatile Yes
(NH4 CH2COOH) potassium salts are not volatile.
Formate Used in the 1–10 mM range. Note that sodium or
3.75 2.75–4.75 Volatile Yes
(NH4 COOH) potassium salts are not volatile.
Phosphate 1 2.15 1.15–3.15 Non-volatile No Traditional low pH buffer, good UV transparency.
Above pH 7, reduce temperature/concentration
Phosphate 2 7.20 6.20–8.20 Non-volatile No
and use a guard column to maximize lifetime.

CORT ECS 2.7 µm Columns 5


b. pH Range solvent to remove the non-polar contaminant(s), taking care not to
precipitate any buffered mobile phase components. If this flushing
Column lifetime will vary depending on the combination of
procedure does not solve the problem, purge the column with the
temperature, mobile phase pH, and type of buffer or additive
following cleaning and regeneration procedures.
used. Table 3 lists the recommended buffers and additives for
CORTECS Columns. 1. Use a cleaning routine that matches the properties of the
samples and stationary phase type (reversed-phase, normal-
Note. Working in combinations of extreme pH, temperature, and phase, or HILIC) to help solubilize the suspected contaminate.
pressure may result in reduced column lifetime.
2. Flush with 20 column volumes of solvent at 45 °C.
c. Solvents 3. Return to the initial mobile phase conditions by reversing
To maintain maximum column performance, use high quality HPLC the sequence.
or MS-grade solvents. Filter all aqueous buffers prior to use through
a 0.2 µm filter. Solvents containing suspended particulate materials 4. If using a reversed-phase column, purge the column with a
will generally clog the outside surface of the inlet of the column. sequence of progressively more non-polar solvents (i.e., water-
This may result in higher backpressure or distorted peak shape. to-methanol-to-tetrahydrofuran-to-methylene chloride).

5. If using a HILIC column, purge the column with a sequence of


d. Pressure progressively more polar-organic solvents (i.e., acetonitrile-
CORTECS 2.7 µm Columns are compatible with HPLC, UHPLC, to-acetonitrile/methanol-to-acetonitrile/water-to-water).
and UPLC pressures. Table 4 provides the maximum operation
6. If column performance has not improved after regeneration/
pressure for each column dimension.
cleaning procedures, contact your local Waters representative
Table 4. Maximum Operation Pressure for additional support.
Column I.D. Maximum operating pressure
2.1 mm 15,000 PSI [1034 bar] b. Storage after Reversed-Phase and HILIC Use
3.0 mm 15,000 PSI [1034 bar] For periods longer than four days, store the column in 100%
4.6 mm 9,000 PSI [620 bar] acetonitrile. For separations utilizing elevated temperature, store
immediately after use in 100% acetonitrile. Do not store columns
in buffered eluents. If the mobile phase contained a buffer salt, flush
e. Temperature the column with 10 column volumes of HPLC-grade water (see Table
CORTECS 2.7 µm Columns can be used at temperatures up to 45 °C 1 for column volume information) followed by 10 column volumes
to enhance selectivity, reduce solvent viscosity, and increase of acetonitrile. Failure to perform this intermediate step could
mass transfer rates. result in precipitation of the buffer salt in the column when 100%
acetonitrile is introduced. Completely seal the column to avoid
solvent evaporation and drying out of the chromatographic bed.
IV. COLUMN C L EANING, R EG EN E RAT ION,
Note. If a column has been run with a formate-containing mobile
AND STO RAG E
phase (e.g., ammonium formate, formic acid, etc.) and is purged
a. Cleaning and Regeneration with 100% acetonitrile, slightly longer equilibration times may be
Changes in peak shape, peak splitting, shouldering peaks, shifts in necessary when the column is re-installed and re-wetted with that
retention, change in resolution, or increasing backpressure may same formate-containing mobile phase.
indicate contamination of the column. Flush with a neat organic

CORT ECS 2.7 µm Columns 6


V. eCO RD INT E L LIG ENT C HI P T EC HNOLOGY c. Column Use Information
The eCord intelligent chip provides the user with specific column
a. Introduction
information as well as column use data including; chemistry
The eCord intelligent chip provides a paperless tracking history
type, column dimension, serial number, and part number. The
of the column’s performance and usage throughout its lifetime.
overall column use information includes; total number of samples
The eCord is permanently attached to the column body via a
injected, total number of injections as well as the maximum
tether that cannot be removed. This ensures that the history of the
pressure, and temperature that the column has been exposed to.
column is always accessible to the user of that column. The eCord
Additionally, detailed column history includes the sample set
intelligent chip is available on CORTECS Columns with 2.1 mm
start date, user name, and system name.
and 3.0 mm inner dimensions.

eCord–
intelligent chip V I. A DDIT IONA L INFO RMAT ION

a. Tips for Maximizing Column Lifetime


1. To maximize column lifetime, pay close attention to:
■■ Water quality (including water purification systems)

■■ Solvent quality
Figure 5. eCord intelligent chip identification.
■■ Mobile phase preparation, storage, and age

■■ Sample, buffer, and mobile phase solubility


At the time of manufacture, the Performance Test Chromatogram, ■■ Sample quality and preparation
analytical data for the particles, and Certificate of Batch Analysis is
downloaded onto the eCord. This information may then be accessed 2. When problems arise, systematically troubleshoot potential
cause one variable at a time in a systematic fashion.
via the ACQUITY UPLC console once the column is installed.
3. Always remember to:
■■ Use an ACQUITY UPLC Column In-Line Filter Kit
b. Installation of the eCord
(p/n 205000343) or a suitable Vanguard
The eCord device can be read by connecting the yellow fob to the
Cartridge Column.
heater compartment receptacle located on the right hand side
■■ Discourage bacterial growth by minimizing the use of
of the ACQUITY UPLC Column heater module. Once the eCord 100% aqueous mobile phases where possible.
is connected, column identification, and overall column usage ■■ Reduce the chances of mobile phase contamination or
information can be accessed. degradation, prepare enough mobile phase to last for
3–4 days. Alternatively, store excess bulk quantities in a
refrigerated environment.
■■ Discard and re-prepare aqueous mobile phase every
24–48 hours (if 100% aqueous mobile phase is required).
■■ Add 5–10% organic modifier to aqueous buffer to minimize
bacterial growth (adjust gradient profile as necessary).
eCord fob ■■ Filter aqueous portions of mobile phase through a
0.2 µm filter.
■■ Routinely maintain your water purification system to
ensure it is functioning properly.
Figure 6. Installing the eCord intelligent chip. ■■ Only use ultra-pure water (18 Ω-cm) and highest quality
solvent possible.
■■ Consider sample preparation (e.g., SPE, filtration,
centrifugation, etc.) when possible.

CORT ECS 2.7 µm Columns 7


4. Avoid when possible: 2. Avoid water and dimethyl sulfoxide (DMSO) in the injection
■■ 100% aqueous mobile phases. solvents. These solvents will produce very poor peak shapes.
■■ HPLC-grade bottled water. In HILIC, it is important to remember that water is the
strongest eluting solvent. Therefore, it must be eliminated or
■■ “Topping off” your mobile phases.
minimized in the injection solvent. Exchange water or DMSO
■■ Using phosphate salt buffer in combination with
with acetonitrile by using reversed-phase SPE. If this is not
high acetonitrile concentrations (e.g., >70%)
possible, dilute the water or DMSO with organic solvent.
due to precipitation.

c. Troubleshooting Questions
b. Getting Started with CORTECS HILIC Columns
1. Are you using 100% aqueous mobile phases?
Note. CORTECS HILIC Columns are designed to retain very polar
bases. Acidic, neutral, and non-polar compounds will have 2. W hat is the age of the mobile phase?
limited retention. 3. Is the mobile phase filtered through a 0.2 µm membrane?
4. Was the mobile phase prepared fresh or topped off?
Mobile Phase Considerations:
1. Always maintain at least 3% polar solvent, such as water, 5. Is the water source of adequate quality?
in the mobile phase, or gradient. T his ensures that the
6. W hen was the last time the water system was serviced or
CORTECS HILIC particle is always hydrated.
was the bottle of water unopened?
2. Maintain at least 40% organic solvent (e.g., acetonitrile)
in your mobile phase or gradient. 7. Is bacterial growth a possibility (pH 7 phosphate buffer is
susceptible to bacterial growth within 24 hours)?
3. Avoid phosphate salt buffers to avoid precipitation in high
organic content mobile phases used for HILIC separations. 8. If a neat standard is prepared in the initial mobile phase
Phosphoric acid is acceptable. conditions and injected, are the problems still observed?
4. Buffered mobile phases such as ammonium formate or 9. If the sample is filtered/purified (i.e., SPE, filtration… etc.) is
ammonium acetate will produce more reproducible results
the problem still observed?
compared to unbuffered additives such as formic acid or acetic
acid. For best peak shape, maintain a buffer concentration 10. Has the quality of the samples changed over time?
of 10 mM.
5. If using an ACQUITY UPLC System, the weak needle wash
solvent should closely match the percent organic solvent V II. C AUT IONA RY NOT E
present in the initial mobile phase conditions, otherwise, Some products may be hazardous during and after use and are
analyte peak shape distortion can occur. to be used by professional laboratory personnel trained in the
competent handling of such materials. The responsibility for the
Injection Solvent Considerations: safe use of products rests entirely with the purchaser and user.
1. Whenever possible, injection solvents should contain 95% The safety data sheets (SDS) for these products are available at
acetonitrile with the polar solvent (i.e., water, methanol, www.waters.com /SDS.
isopropanol) component to be no more than 25% of the total
volume. A generic injection solvent is 75:25 acetonitrile/
methanol. T his is a good compromise between analyte
solubility and peak shape.

Waters Corporation
34 Maple Street
Milford, MA 01757 U.S.A.
Waters, T he Science of W hat’s Possible, CORTECS, UPLC, ACQUITY UPLC, Alliance, Oasis, and Sep-Pak are registered trademarks of Waters
Corporation. eCord and VanGuard are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
T: 1 508 478 2000
F: 1 508 872 1990
©2016 Waters Corporation. Produced in the U.S.A. September 2016 720005024EN IH-PDF www.waters.com

You might also like