Professional Documents
Culture Documents
Flash 4000 OPM
Flash 4000 OPM
FLASH 4000
Elemental Analyzer
Operating Manual
Trademarks
Flash 4000 and Eager Xperience are trademarks of Thermo Fisher Scientific Inc.,and its subsidiaries. Other
brand and product names may be trademarks or registered trademarks of their respective companies.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-
free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Release history:
For Research Use Only. Not regulated for medical or veterinary diagnostic use by U.S. Federal Drug
Administration or other competent authorities.
C
Contents
Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ix
Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xi
Safety Alerts and Special Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xi
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xi
Types of Alerts and Signal Words . . . . . . . . . . . . . . . . . . . . . . . . . . . . xii
Special Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xii
Instrument Markings and Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . xii
Using the Flash 4000 Document Set . . . . . . . . . . . . . . . . . . . . . . . . . . . xiv
Part 1 Description
Part 2 Preparation
Part 3 Analysis
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
Thermo Fisher Scientific is the manufacturer of the instrument described in this manual
and, as such, is responsible for the instrument safety, reliability and performance only if:
• installation
• re-calibration
• changes and repairs
Thermo Fisher Scientific is not liable for any damages derived from the non-compliance
with the aforementioned recommendations.
Elemental Analysis
Voltaweg 22
2627 BC Delft
The Netherlands
Preface
This Operating Manual contains descriptions of the features and components of the Flash
4000 elemental analyzer, inside, you will find all of the information necessary for routine
operation of your elemental analyzer, including operating sequences, sample injection
techniques, diagrams and descriptions of the major components.
• Part 2, “Preparation,” provides description of the reactors and adsorption filters. Moreover,
it contains instructions to install and remove the system reactors, to perform sample
preparation and weighing, and also how to start up the instrument before running
analyses.
− Chapter 5, “Analyzer Installation,” provides instructions to install your Flash 4000
elemental analyzer.
− Chapter 6, “Reactor Preparation and Connections,” provides instructions for the
preparation and the installation of the reactors into the elemental analyzer and how to
remove them.
− Chapter 7, “Sample Preparation,” describes the technique for the sample preparation,
also it provides basic instructions to weigh the sample.
− Chapter 8, “Analytical Methods,” describes the analytical methods used for all
configurations of the Flash 4000 elemental analyzer.
• Part 3, “Analysis,” contains information and operating sequences to perform the analysis
preparation, execution and interpretation.
− Chapter 9, “Instrument Set-up,” contains information and operating procedures to set
the instrument for running analyses.
− Chapter 10, “Instrument Start-up,” sequences to prepare the instrument for running
analyses.
− Chapter 11, “Guide to Run Analyses,” contains information and operating sequences
to run sample analyses, and it also describes the comparison methods for a correct
evaluation of results. Practical advise for daily operation is also provided.
• Part 4, “Maintenance and Troubleshooting,” contains information and operating sequences
to perform the necessary maintenance of the instrument and also information concerning
troubleshooting in case of malfunctioning.
− Chapter 12, “Maintenance,” provides information on the current and periodic
maintenance of the instrument, and it also contains the operating sequences for the
maintenance of the MAS 4000 autosampler.
− Chapter 13, “Troubleshooting,” provides the information necessary to give you some
hints on instrument problems and how to solve them.
The Abbreviations contains definitions of terms used in this guide. It also includes
abbreviations, acronyms, metric prefixes, and symbols.
The Index contains an alphabetical list of key terms and topics in this guide, including cross
references and the corresponding page numbers.
Compliance
We thoroughly test and evaluate our products to ensure full regulatory compliance with
applicable domestic and international regulations. Your system (hardware and software) is CE
Compliant and meets Electromagnetic Compatibility (EMC) and safety standards.
CAUTION Instrument Damage. Flash 4000 systems operate safely and reliably under
carefully controlled environmental conditions. If the equipment is used in a manner not
specified by the manufacturer, the protections provided by the equipment may be impaired.
If you maintain a system outside the specifications listed in this guide, failures of many
types may occur. The repair of such failures may be excluded from the documents regarding
your standard warranty and service contract coverage.
Safety Symbols
Safety and special notices that may be found on your instrument and in this manual include
the following:
This symbol indicates that an electrical shock hazard will, could, or may occur.
This symbol indicates a hot surface. Make sure the instrument is at room
temperature before touching, or else you will, could, or may incur burn injuries.
This symbol indicates that chemical damage or physical injury will, could, or
may occur.
This symbol indicates that eye damage will, could, or may occur. Eye protection
must be worn.
This symbol indicates the user must wear gloves when performing the
sequence.
DANGER safety alerts an imminent hazard exists that WILL result in death or serious
personal injury.
WARNING safety alerts you to an action or sequence that, if improperly performed, could
result in damage to the instrument or possible physical harm to the user. This symbol may
be followed by icons indicating special precautions that should be taken to avoid injury.
CAUTION safety alerts you to an action or sequence that, if performed improperly, could
damage the instrument.
Special Notices
Tip Helpful information that can make a task easier.
Symbol Description
Direct Current
*
Alternating Current
Symbol Description
Equipotentiality
*
On (Supply)
*
Off (Supply)
* Instruction manual symbol affixed to product. Indicates that the user must
refer to the manual for specific Warning or Caution information to avoid
personal injury or damage to the product.
*
Caution, hot surface
*
Caution, biohazard
* + -
Jack socket
The Flash 4000 Document Set (PN 317 110 50) as paper copy is also available. Furthermore,
Thermo Fisher Scientific part numbers (PN) for the paper copy manuals are provided for each
book title.
• Site Preparation and Installation Requirements (PN 317 110 51)
This manual describes how to set up a workspace for the instrument and accessories and
how to connect the main unit to the gas supplies and peripheral devices.
• Operating Manual (PN 317 110 52)
This manual provides the descriptions of the hardware and software and detailed
instructions for their use.
• Consumables and Spare Parts Catalog (PN 317 082 50)
This catalog contains a list of consumables and spare parts for the Flash 4000 elemental
analyzer.
Preliminary Information
This chapter provides information on the classification of the equipment, its safety and
configurations. Basic technical features are also described.
Contents
• Classification of the Instrument
• Technical Features
• Safety Information
• Safety Cut Off Device
• Instrument Cleaning
• Instrument Configuration
• Standard Outfit
Technical Features
The following table summarizes the major technical features of the Flash 4000 elemental
analyzer.
Table 1. Technical features of the instrument
Features Description
Instrument configurations One
Detector Thermal conductivity detector (TCD)
External interface RS 232 serial line, Local Area Network (LAN)
Instrument control Eager Xperience for Windows™
Power supply 230 Vac; 50/60 Hz; 2300 VA
Dimensions (cm) Height 47 (60 with fittings); Width 67, Depth 62
Mass (kg) 80
Safety Information
WARNING The instrument must be used according to the specifications of this guide.
Improper use can adversely affect the instrument protection. If the equipment is connected
to optional instruments, such as computer, balance, etc., the degree of insulation of
peripheral devices should be equivalent or higher (double or reinforced) than that of the
Flash 4000. The analyzer operation requires the use of chemical substances having different
hazard specifications. Before using chemicals, please read the hazard indications and
information reported in the Safety Sheet supplied by the manufacturer referring to the
relevant CAS (Chemical Abstract Service) number.
Use of Gases
The following gases are used with the instrument:
• Helium (He) as carrier, reference and purge gas.
• Oxygen (O2) as gas for sample oxidation.
WARNING Before using gases, carefully read the hazard indications and information
reported in the Safety Sheet supplied by the manufacturer referring to the CAS (Chemical
Abstract Service) number. It is the user’s responsibility to see that all local safety regulations
for the use of gases are obeyed.
IMPORTANT In absence of Helium the Flash 4000 goes in Safety Cut Off. See “Safety Cut
Off Device” on page 17.
Purity of Gases
The Flash 4000 uses Helium and Oxygen with 99.995% minimum purity.
Instrument Cleaning
Y Instruments Cleaning
WARNING Cleaning must be performed with the instrument off, the furnaces at room
temperature and the power cord disconnected.
1. Externally clean the instrument with a soap and water solution, or with a household
non-abrasive product, carefully avoid seeping of the products used inside the instrument.
2. If you just suspect that a substance used for cleaning or a product submitted to analysis has
infiltrated inside the instrument, immediately shut down the instrument and call an
authorized customer support engineer for proper actions.The service engineer must be
fully informed on the nature of the concerned substance.
Instrument Configuration
The Flash 4000 elemental analyzer is available in the following version:
Table 2. Instrument configuration
Configuration Description
N/Protein Analyzer For the determination of the Nitrogen and protein amount
in food and animal feed products. It can also be used in
determining Nitrogen content in samples of different nature,
such as agricultural products, marine science, geological,
human/animal samples, material characterization.
Standard Outfit
Flash 4000 elemental analyzer is provided with its own standard outfit.
Use the standard outfit checklist accompanying the instrument to verify that all items have
been received.
Description
2
Contents
• The Units Constituting the Instrument
• Analytical Section
• Control Section
• Automation
Analytical Section
It comprises the following major components:
Autosampler
It performs the automatic injection of the samples into the combustion reactor.
Reactors
These are tubes made of special alloy filled with different materials. Refer to Chapter 6,
“Reactor Preparation and Connections,” on page 73.
Furnaces
The Flash 4000 analyzer is equipped with two furnaces: LEFT and RIGHT. Each furnace
consists of a quartz candle surrounded by an electrical resistor. The candle is plunged in a
refractory material housed in a metal compartment.
• Furnace Temperature
The temperature is monitored by a thermocouple appropriately located in the furnace.
• Furnace Cooling
The cooling time varies according to the operating temperature setting.
Chromatographic Column
It performs the chromatographic separation of the reaction products generated during the
combustion process.
Control Section
The control section consists of two major compartments:
Pneumatic Compartment
It consists of two pressure regulators, three pressure gauges and several lines fitted with an
thermoregulator electronic flow controller (EFCt), which ensures the switching between
helium and oxygen, and controls the flow values.
For more details, please refer to Chapter 4, “Pneumatic Circuit,” on page 45.
Electronic Compartments
They comprise the electronic boards for the instrument power supply and control.
Automation
The instrument is fully controlled by the computer through a dedicated software which is also
used for data acquisition, data handling and interpretation of the acquired results. Therefore
the instrument is not provided with independent keyboard and display. On the instrument
front there three status LEDs where you can monitor the instrument statuses.
Instrument Description
This chapter gives you a detailed description of the instrument components.
Contents
• Front View
• Rear Panel
• Top View
• Furnaces Compartment
• Fittings for Gas Connections
• Right Front Compartment
• Regenerative Filters Compartment
• Right Side Compartments
• Description of the Detection System
• Water Condensate Drainage
• Electrical Compartment
• Connections Panel
• Status Panel
• Autosamplers
Front View
The front part of the instrument, shown in Figure 1, comprises:
• Furnaces compartment
Also refer to “Furnaces Compartment” on page 28.
• Status panel
Also refer to “Status Panel” on page 41.
• Right Front compartment
It includes:
− Pressure regulators and pressure gauges
− Water condense collector
Rear Panel
The rear part of the instrument, shown in Figure 2, comprises:
• Panel for connections including:
− interface section
− gas inlets section
Also refer to “Connections Panel” on page 39.
• Cooling fan
• Main voltage electronic section
Also refer to “Electrical Compartment” on page 35.
• Cooling fan of regenerative filters See “Regenerative Filters Compartment” on page 31.
• Transformer compartment
Also refer to “Electrical Compartment” on page 35.
• Breaker and Mains socket
Gas inlet
section
Cooling fan
Interface
section
Breaker
Mains
socket
Top View
The top part of the instrument, shown in Figure 3, comprises:
• Top cover
• MAS 4000 autosampler
• Fittings for mounting and securing the reactors into the furnaces
• Fittings for gas connections
Furnaces Compartment
The furnaces compartment, shown in Figure 4, is accessible from the instrument front
removing the front cover.
To remove the front cover it is necessary to open the lower door unscrewing the proper fixing
screw.
CAUTION Do not open the furnaces compartment during operation because very high
temperatures are reached.
The front cover can be removed only when the furnaces temperature shown is near the
room temperature.
Gas pressures must be set and controlled through the pressure regulators and the gauges of the
instrument.
Pressure Gauges
Pressure Regulators
Water collecting
bottle
• Pressure gauge (0-300 kPa) marked Combustion Reactor visualizes the pressure of Helium
in the pneumatic circuit. This pressure is not adjustable.
• Pressure gauge (0-300 kPa) marked He visualized the pressure of Helium.
Set this pressure to 120-150 kPa by turning the adjusting screw of the relevant pressure
regulator.
This allows to obtain a gas flow of 300 mL/min ± 50 mL/min (250÷350 mL/min) for the
filters regeneration.
• Pressure gauge (0-500 kPa) marked O2 visualized the pressure of Oxygen.
Set this pressure to 450-460 kPa by turning the adjusting screw of the relevant pressure
regulator. This allows to obtain a gas flow suitable for the combustion.
Filters
WARNING Do not remove the filter from its housing. The filter must be removed, only
when required, by authorized technical personnel.
WARNING Do not expose gas tanks or bottles, chemicals, regulators, electrical cords, or
other temperature-sensitive items to exhaust vents.
Filters valve
manifold
Check valve
Chromatographic
column
TCD Detector box
Water condensate
drainage box
The detector essentially consists of a stainless steel block provided with two pairs of filaments
(generally of tungsten/rhenium) having the same electrical resistance. The detector is housed in
a thermally insulated metal block (detector oven) and maintained at constant temperature.
The two pairs of filaments are electrically connected according to a Wheatstone bridge circuit
powered at constant voltage.
The first pair of filaments is fed with pure carrier gas (reference channel), whereas the second
pair is fed with the gas flowing from the reactor (analytical channel).
When the bridge is powered, the filaments heat at a temperature (resistance) that is a function
of the thermal conductivity of the gas feeding the filaments. The reference channel is exposed
only to pure carrier gas, whereas the analytical channel is exposed to the reactor effluents
(carrier gas + sample).
When pure carrier gas flows through both the reference and the analytical channels, a constant
temperature gradient is established between the elements and the detector walls, and the
Wheatstone bridge is balanced, namely there is no output signal.
As a component is eluted, a change in heat transfer occurs, with consequent variation of the
filaments temperature. Since electrical resistance is a function of temperature, the bridge
unbalances and the detector generates a signal proportional to the difference in thermal
conductivity between the eluted component and the carrier gas. The output signal is then sent
to the data acquisition board.
Note The filaments are powered at 5V constant voltage and are electrically protected if
their temperature exceeds 220 °C (Safety Cut Off ).
CLR-EAX board
LED
Water condensate
drainage box
Every analysis, the condense is drained in the collecting bottle located into the right front
compartment of the instrument.
Electrical Compartment
CAUTION Before opening the electrical compartments, cut off power supply to the
instrument and disconnect the power cord.
EFC-t Box
Table 6. Description of the function of the low voltage electronic boards, continued
Board Description
CPU-EAX CAUTION The rechargeable battery replacement must be performed
(continued) by authorized Thermo Fisher Scientific field service engineers.
HWD-EAX It allows the detector oven thermoregulation and also amplifies and
converts the detector signal to send it to the PC.
TST-EAX This board provides test points for technical diagnostic.
On the motherboard is present the slot for this board which is not
included with the instrument.
CAUTION This board must be used only by authorized Thermo
Fisher Scientific field service engineers.
EXT-EAX It provides the circuits for the communication between the elemental
analyzer and the external world through RS 232 serial line or Local
Area Network (LAN)
FRT-EAX Status panel.
This board is located on the instrument front.
CLR-EAX Peltier element control.
This board is located in the low temperature analytical section.
This compartment contains the mains power circuits with the Safety Cut Off device and the
electrical devices to power the furnaces, the filters and control their temperature.
Solid state relays (SSR) are coupled with a proper safety sensor, which
detects any malfunction. The SSRs control the power supply to the
relevant furnace and cut off power to the heating resistor when the
thermocouple detects temperature values exceeding the setpoint.
Connections Panel
The panel, shown in Figure 14, is subdivided into four sections:
• Gas inlets
• Interface
• Communication
4 Selection Switch
Interface Section
It comprises:
• 15-pin connector marked MAS for the MAS 4000 autosampler for solid samples.
Communication Section
It comprises:
• 9-pin connector marked RS 232 to dialog with the computer via serial line.
• LAN communication port. See LAN Section.
• Selection switch
Note By using the selection switch located between the RS 232 and LAN connectors it is
possible to choose the desired type of communication.
LAN Section
The LAN (Local Area Network) section is easily recognizable by the presence on the analyzer
rear panel of a RJ45 connector, 2 LEDs for LAN activity and a reset button.
The Flash 4000 LAN is shipped with a default IP address, which may not match the LAN
needs of the sites where the analyzer must be installed.
To change the default values, contact your LAN administrator and ask for the IP address to be
assigned, the netmask, and eventually the port. The IP address and communication port may
be reset to default value by pushing the reset button, accessible from the rear panel of the
analyzer, for at least 5 seconds.
CAUTION The reset button is accessible introducing a little screwdriver into the relevant
passing hole.
To set the IP address and the LAN communication port, follow the instructions reported in
paragraph “LAN Set-up” on page 63.
Status Panel
Status panel, shown in Figure 15, is the user interface of the GC. It consists of a series of three
Status LEDs showing the instrument’s status.
The LEDs (Light Emitting Diodes) indicate the operating status of Flash 4000 elemental
analyzer. Each status LED indicates the current operating mode and special settings activated
by the operator. Table 10 lists and explains each status LED.
Autosamplers
The Flash 4000 elemental analyzer can be configured with the following autosamplers.
Tray
Motors MAS-EAX
board
Viewer
Sampler body
Pneumatic Circuit
This chapter describes the pneumatic circuit of the configuration in the pre-run condition.
Contents
• Introduction
• Pressure Regulators and Gauges
• EFC-t Module
• Water Condensate Drainage
• Solenoid Valves Manifold
Introduction
The pneumatic circuit, schematically shown in Figure 18 and Figure 19 has the following
common components:
• Inlet gases pressure regulators and relevant gauges
• EFC-t Electronic Flow Controller for gases
• Two reactors
• Valves manifold
• Water condensate drainage (Water drain)
• Regenerative carbon dioxide filters
• One gas chromatographic column
• TCD Thermal Conductivity Detector
He Pressure F2
Valves
Regulators EFCt
Manifold
O2 and Gauges
F1
Vent AS
EV5
Water
Carrier Drain CC
Filters Purge R1 R2
Reference
Water Drain
CC = Chromatographic Column
F1 = Adsorption filter
F2 = Adsorption filter TCD
R1 = Oxidation Reactor
R2 = Reduction Reactor
F
He PRV1
SVP2 T He
1
(5) 2
PI2
F PRV2
F O2
IN O
OUT SV1 2
(6) 2
2 2 2
Green Reduction furnace
CV S1 1
Vent
4
(4) 1
2
SVP1
(1) (2)
R1 R2
Reference
Autosampler AS
F F
600 sccm @ 30 psi - AIR Water Drain WD
1
SV8 3 1 3 SV9 SV5
1
2
2 Vent
2
Bottom - BF White (6) F F (4) Top - TF
White
B
Blue (5) F F (3) Blue
2 2
3 1 1 3
SV7 SV6
(8) (1)
Front Panel - FP
PI1
PRV1
He
PRV2
O
2
PI3
Carrier
To Autosampler
Carrier
Reference
EFC-t Module
The electronic flow control thermostatic (EFC-t) module, schematically shown in Figure 21,
contains some solenoid valves that are placed in the manifold maintained at the constant
temperature of 42 °C. Every variation of temperature does not influence the flows of the gas.
See Table 13.
EFC-t Module
S2
F
SVP2 1 He
(5) From
2
To TCD Pressure Regulators
F
F O
(6) 2
SV1
2
2 2 2
S1
To TCD
(4) 1
2
Reference SVP1 To Autosampler
(1)
(2)
To Autosampler
The gas is used to eliminate air from the MAS 4000 autosampler.
The valve is closed during the leak test.
S1 Electronic flow sensor for Helium as carrier gas and Oxygen during the
sampling stage. It cooperates with the SVP1 electronic controller
(proportional valve).
S2 Electronic flow sensor for Helium as reference gas.
Carrier
From Reactor R2
Water Drain WD
1
SV5
2
Vent
Carrier Helium
To Solenoid from regenerative
Valves Block CO2filter
He To Chromatography
Column
(7) (2)
F F
600 ml/min @ 30 psi - AIR
SV8 3 1 1 3 SV9
2 2
(5) F F (3)
2 2
3 1 1 3
SV7 SV6
(8) (1)
Preparation
5
Analyzer Installation
This chapter provides instruction to install your Flash 4000 elemental analyzer.
Contents
• Preliminary Information
• MAS 4000 Autosampler
• Sample Tray
• LAN Set-up
Preliminary Information
This chapter contains the preliminary information for the installation and connection of the
Flash 4000 elemental analyzer and for the electrical requirements.
Standard Outfit
Use the standard outfit checklist accompanying the instrument to verify that all items have
been received.
Electrical Requirements
The instrument must be electrically supplied as indicated in “Technical Features” on
page 16.
CAUTION The power line and the connections among the instruments must maintain
good electrical grounding. Poor grounding represents a danger for the operator and may
seriously affect the instrument performance.
Do not connect the Flash 4000 elemental analyzer to lines feeding devices of a heavy duty
nature, such as motors, UV lamps, refrigerators and other devices that can generate
disturbances.
If other instruments, such as computer, balance, printer, etc., have to be connected to the
same electrical line as the Flash 4000, ensure that such electrical line is capable of
withstanding such electrical consumptions by calculating the total absorption.
Laboratory Temperature
The instrument is generally best located in an environment where atmospheric conditions can
be controlled and extreme temperatures avoided.
The precision of the instrument can be regarded as a function of the precision of the
laboratory temperature control. Therefore do not place the instrument near heating or air
conditioning vents which may cause differential heating and drafts.
In laboratories with windows, venetian blinds should be installed to protect the instrument
from direct sunlight.
Pneumatic Connections
Verify that the pneumatic connections have been properly carried out as described in Chapter
3 of the Site Preparation and Installation Requirements manual.
Electrical Connections
Verify that the electrical connections have been properly carried out as described in Chapter 3
of the Site Preparation and Installation Requirements manual.
If LAN connection has been carried out, perform the set-up following the instruction
reported in paragraph “LAN Set-up” on page 63.
The base unit is provided with one 31-position sample tray. Up to three optional 31-position
sample trays can be added to reach a capacity of 124 samples.
Each sample tray is installed in a specific position defined by the numbering, and therefore
they are not interchangeable.
4 Sample tray
5 lens
Optical
Sampler3 body 1 Purge/reference gas inlet
2 Carrier gas inlet
6 Sampler fixing nut
Reactor fittings
WARNING Before starting, make sure that the analyzer is powered OFF.
1. The autosampler must be directly installed on the connecting nut of the combustion
reactor.
2. Place the autosampler on the top cover of the analyzer in correspondence of the
combustion reactor.
3. Manually screw the autosampler nut on the combustion channel.
4. Fix the autosampler.
5. Connect the tubes coming from the gases connections 1 and 2 located on the analyzer, to
the relevant connections 1 and 2 of the autosampler.
6. Connect the signal cable of the MAS 4000 autosampler to the 15-pin connector, marked
MAS, on the real panel of the analyzer.
WARNING To avoid damaging the MAS 4000 autosampler, NEVER connect/disconnect
the signal cable when the analyzer is powered ON.
Sample Tray
The sample tray, shown in Figure 25, can contain up to 31 samples. Up to four sample trays,
numbered 1to 4, can be added. The sample numbering is detailed in Table 16. Each sample
tray is provided with a locating mark to be aligned to the seat marked 0 (zero).
Any sample has to place in the locating mark position. The correct alignment of the
locating mark is important for the installation of the sample tray on the MAS 4000
autosampler.
Under the tray, in correspondence of the locating mark, there are present:
• the sampling hole through which the sample will pass during the sampling sequence.
• the guide pin
A B
2
3
1
2
1
1. Before starting, verify that the guide slot, located on the tray holder hub, is perfectly
aligned with the sampler body. If not, click on the red switch, located on the left hand side
of the sampler body. The tray holder hub turns until the guide slot reaches the correct
position.
Figure 26 shows the wrong and the correct position of the guide slot on the tray holder
hub.
WRONG CORRECT 1
A
1
3
2 2
1
2
4
5
1. Tray holder hub 2. Guide slot
3. Sampler body 4. Red switch
5. Collar
2. Take the sample tray and manually move the locating mark until it is aligned to the seat
marked 0 (zero).
3. Place the sample tray on the tray holder hub paying attention to properly insert the guide
pin, located under the tray in correspondence of the locating mark, into the guide slot.
Verify that also the sampling hole is perfectly positioned on the white collar on the
sampler body. If not, adjust manually.
4. Place the samples to be analyzed, previously prepared, in the sample tray seats starting
from the seat #1. Up to 31 samples can be allocate.
5. If additional sample trays are required, install them in the correct order one over the other,
paying attention that the relevant locating marks are in correspondence with the relevant
seats marked 0 (zero), and accommodating the samples properly. See Table 16.
IMPORTANT Before installing an additional sample tray, make sure that the samples to be
analyze have been placed in all the seats of the previous tray installed.
Figure 27 shows the result of the trays installation with the relevant locating mark aligned to
the respective seat marked 0 (zero).
Sample tray #4
Sample tray #3
Locating mark Sample tray #2
Sample tray #1
Sampling Phases
Sampling cycle happens in three phases as schematically shown in Figure 28 on page 61.
• Loading
The sample that is going to be analyzed drops in the sampler body.
• Purging
The sample is purged by a flow of purge Helium (reference gas).
• Sampling
The sample drops into the combustion reactor.
4. When a new start signal is send, the sampling cycle is repeated for the sample #2 and so
on for all the samples put in the sample tray.
5. If additional trays are present, after the sampling of the sample #31, at the next start
command, the locating mark (seat 0) of the first tray hooks the locating mark (seat 0) of
the second tray.
This allows the correct sampling of the samples put into the seats provided in the second
tray.
The same mechanism is repeated also for the third and the fourth tray if installed. See
Figure 29 on page 62.
2
LOADING
3
4
8
1 5
6
7
PURGING
OUT
Purge
He IN
SAMPLING
LAN Set-up
This paragraph contains a few notes on how to set-up and start using the LAN port of the
Flash 4000 elemental analyzer.
Introduction
The Flash 4000 featuring LAN (Local Area Network) capabilities is easily recognizable by the
presence on the rear panel of a RJ45 connector, 2 LEDs for LAN activity, a reset button and a
switch to select the communication via RS232 or LAN.
5
3
6
4 1
LAN Area
To change the default values, contact your LAN administrator and ask for the IP address to be
assigned, the netmask, and eventually the port.
• The IP address is a 3 digits x 4 fields number given by the network administrator e.g.
192.168.127.10
• The netmask is a 3 digits x 4 fields number given by the network administrator e.g.
255.255.255.0
• The port is a 4 digits number given by the network administrator e.g. 4001 (4001 is
usually the CPU/LAN default)
Reset Button
To reset the IP address and communication port to default value, push the reset button,
located on the rear panel of the Flash 4000 for at least 5 seconds.
Network Cables
Two network cables are included in the standard outfit of the Flash 4000:
• a reversed RED patch for computer to Flash 4000 direct connection. This is used for the
initial set-up operation and in the case you do not want connecting the instrument in
remote mode.
• a standard GREY patch for Flash 4000 to local area network connection.
The principle of connection between Flash 4000 and PC is schematically shown in the
following figure.
Set-up
To set the desired IP and set up the LAN communication port of the Flash 4000 then to
configure the data system, please follow the instruction reported in To Set- up the Flash 4000
operating sequence.
Material Required
PC (desktop or portable)
Network connecting RED cable
Network connecting GREY cable
To properly set the desired IP address and set up the LAN communication port of the Flash
4000, perform the following steps:
1. Verify that the instrument and the PC are switched off.
2. By using the LAN reversed RED patch included in the standard outfit, connect a PC
(desktop or portable) directly to the RJ45 connector marked LAN located on the rear
panel of the Flash 4000.
1. Select the Local Area Network connection properties of your computer (please refer to
specific operating system instructions to access this configuration).
4. Select an IP in the range 192.168.127.1 to 192.168.127. 253 and set the subnet as shown
in the next figure.
5. At this point confirm by clicking OK, and restart from point 3 to make sure the Flash
4000 default IP is now reachable.
Note Should you still have problems, please check the cable connection, and go through
the reset procedure of the Flash 4000. The CPU/LAN is hence reset to the default IP
address. Refer to “Reset Button” on page 63.
6. Proceed following the instructions reported in the operating sequence To operate when
the IP address is reachable.
When the IP is reachable it means that the PC is communicating with the Flash 4000 LAN. It
is now possible to set the LAN communication port.
1. Start a command prompt and type telnet 192.168.127.254
2. The telnet program connects to the IP 192.168.127.254 and shows the following page:
3. Confirm the default selection (1) by pressing Enter. The following page will be visualized.
With arrows, Enter and Esc keys you’ll navigate the program.
4. Select server Config to enter the new IP address and netmask.
With arrows and Enter keys you’ll point to the different parameters to be set.
In this page you’ll need to enter the new assigned IP needed for the LAN environment
where the Flash 4000 is installed. Also the netmask is entered in the same menu section.
In this example the IP set is 192.168.0.101.
Press ESC when done with the settings of this menu, to return to upper level menu.
Press the right arrow key to select OP_mode, and then press ENTER.
5. By moving the selection with the up/down arrow keys, make sure that Raw connection
(TCP Server) is selected and then confirm with Enter key.
Press ESC when done, to return to upper level menu.
6. Use the arrow key to select Serial Port.
CAUTION Please confirm to make the changes effective, and remember that from
now on, the Flash 4000 LAN will respond to the newly assigned IP Address.
8. Disconnect the Flash 4000 from direct RED patch, and connect it to its final destination
LAN environment by using the grey patch.
At this point check the communication opening again the Internet Protocol (TCP/IP)
Properties page and select Obtain an IP address automatically.
9. As a final check, start from the computer that should be used for controlling the newly
installed Flash 4000 and run Command Prompt.
Then type ping xxx.xxx.xxx.xxx, where the xxx.xxx.xxx.xxx is the new IP you just
configured.
The Flash 4000 should answer as follows:
The Flash 4000 is now ready for LAN control through the Thermo Fisher Scientific data
systems. Now it is necessary to configure the data system to access the Flash 4000 through
the configured IP address.
The data systems that support the Flash 4000 control is Eager Xperience.
Data System Configuration
It is advisable to use the self adhesive labels you find in the standard outfit of the Flash
4000 to annotate the IP address and the TCP Port that have been set in the elemental
analyzer. Stick the label in a place that can be easily referenced when it will be necessary to
configure the data system.
10. Install, and start the Data System as described in the specific manual, and go to
Instrument configuration. The Flash 4000 configuration page features the possibility to
control the instrument through the COM ports or to specify a direct TCP/IP address.
11. Just enter the IP address of the Flash 4000 and complete the configuration in the usual
way.
When the Advanced button is pressed, it is possible to set the communication port used
by the TCP/IP protocol and the timeout.
The parameters set by default are those necessary for standard operations; however your
LAN may be provided with Firewall services that may prevent the Port 4001 to be used.
For this reason you have here the possibility to set an alternative port number. However,
the number of the port entered in this box must correspond to the port assigned to Flash
4000 setup described on step 5.
After selecting Raw connection (TCP server), it will be possible to enter the Select for
more setting menu and to enter a different TCP port rather than the default 4001.
Moreover the Flash 4000 Instrument configuration advanced settings feature a box for
timeout. This timeout is set by default to 500 ms, and it’s appropriate for most of the
LAN environments. However should the LAN be extremely slow, this time can be
increased to allow slower access to the autosampler.
Note If the connection is performed through hubs over a 10 Mbit/s network, it is
suggested that no more than five Flash 4000 are connected on the same network
trunk. In the case of switched network, this warning can be ignored.
Alternative Configuration
An alternative to the direct setup of the Flash 4000, it is also possible to use a management
program designed for sites where multiple Flash 4000 elemental analyzers with LAN option
are available.
In this case a specific program must be installed and used to monitor, setup, or update each of
the instruments connected.
The program to be installed is named DSSETUP.EXE, and it is in the CD of the data system,
as well as included in the CD of this manual.
Deselect the check box COM Port Mapping Tools, since it is not required in the management
of Flash 4000 elemental analyzers.
Please just follow the detailed route proposed step-by-step by the installation program, and
when finished, you can run the program. The first time after the installation the program
starts automatically.
12. From main menu select the Locate Server and the program will automatically search for
all Flash 4000 elemental analyzers connected to the network.
By double-clicking on any of the found Flash 4000, you can setup the program directly
Any change can be entered individually for a certain parameter, and when OK is pressed,
the parameter is updated to the instrument on edit.
Contents
• Introduction to the Preparation of Reactors
• Preparing Reactors and Crucible
• Reactors Installation and Replacement/Remotion
The following Table 17 and Table 18 report the characteristics of the components required
for N/Protein determination and the type and size of the filling materials to be used for a
proper preparation of reactors.
Quartz Wool
R2 Reactor Material: Special Steel Quartz Wool
Reduced Copper
CR Crucible Material: Special Steel Ultra temperature ceramic
disk
CC Gas chromatographic Material: SS
column Length: 80 cm
Diameter: 10 x 9 mm
R1 R2
2 40 mm
CR
260 mm
CR R1 R2
4 360 mm
10 mm 1
15 mm 2
145 mm 3
40 mm
20 mm 2 2
CC
CAUTION Before using the filling materials required for this operation, please read the
hazard warnings and information reported in the Safety Data Sheets provided, referring to
the relevant CAS (Chemical Abstract Service) number.
Always use original Thermo Fisher Scientific materials and products. The use of materials
not meeting the technical specifications of our products does not ensure a good operation
of the instrument and may even damage it.
The filling of reactors and crucible requires the use of quartz wool and ceramic
disks respectively. Before handling these materials, we recommend to wear
gloves and face protection.
The filling procedures should be carried out on a wide and clean workbench.
The following operating sequence provides instructions for filling the combustion and
reduction reactors.
Material Required
Special steel reactors
Compression rod
Filling material
Lubricant (silicon grease)
Figure 31. Introduction of quartz wool into the bottom end of the combustion reactor
Reactor
Quartz wool
2. Mix a proper amount of platinum supported on alumina (PtAl2O3) and copper oxide
(CuO) as shown in Figure 32.
Figure 33. Filling of the combustion reactor with the mixed filling material (1)
4. Introduce a sufficient quantity of quartz wool to form the last required layer.
5. Gently push down the quartz wool by using the rod provided without overpressuring.
6. Place and smear a slight layer of silicon grease on the conical bottom end of the reactor.
Figure 34. Introduction of quartz wool into the bottom end of the reduction reactor
Reactor
Quartz wool
2. Introduce a proper amount of reduced copper into the upper end of the reactor as shown
in Figure 35.
Figure 35. Filling of the reduction reactor with the filling material
Reduced copper
3. Introduce a sufficient amount of quartz wool to form the last required layer.
4. Gently push down the quartz wool by using the rod provided without overpressuring.
5. Place and smear a slight layer of silicon grease on the conical bottom end of the reactor.
The following operating sequence provides instructions for the preparation of the crucible you
will use with the special steel reactors required for combustion.
2
1
3
Material Required
Ultra temperature ceramic disks
2. Hold the crucible bottom end as shown in Figure 38. Introduce two ultra temperature
ceramics disks into the bottom end of the crucible.
Figure 38. Introduction of ultra temperature ceramics disks into the crucible
3. Hook the crucible bottom end to the crucible body as shown in Figure 39.
CAUTION The reactors must be installed with the furnaces at room temperature.
Material Required
Tool for steel tube
O-ring
Lubricant (silicon grease)
Proceed as follows:
1. Remove the autosampler, if installed, from the connecting fitting of the reactor tube by
manually undoing the fixing nut, as shown in Figure 40.
Note It is not necessary removing completely the autosampler from the analyzer and
disconnecting the signal cable.
2. On the conic part of the reactor smear a small quantity of o-ring lubricant.
3. Introduce the reactor into the furnace as shown in Figure 41.
Combustion reactor
4. Guide the reactor into the furnace. The conical part should slide into the coupling union
located on the base of the furnaces compartment, as shown in Figure 42.
Furnace
Reactor
Coupling union Base
5. Push the reactor up to reach the correct placement, as shown in Figure 43.
O-ring
7. Using the tool for special steel reactors, introduce the crucible into the combustion
reactor, which is in the left furnace. See Figure 45.
Crucible
removing
tool
Crucible
Material Required
Tool for steel tube
O-ring
Lubricant (silicon grease)
Reduction
reactor
3. Guide the reactor into the furnace. The conical part should slide into the coupling union
located on the base of the furnaces compartment, as shown in Figure 48.
Furnace
Reactor
Coupling Base
Union
4. Push the reactor up to reach the correct placement, as shown in Figure 49.
Reactor
removing
tool
O-ring
CAUTION The reactors must be removed with the furnaces at room temperature.
Make sure that the gas flows are OFF.
The following operating sequence contains the instructions to remove the combustion reactor
from the furnace.
Material Required
Tool for special steel reactor
Tool for crucible
Proceed as follows:
1. Remove the autosampler, if installed, by manually undoing the fixing nut as shown in
Figure 40 on page 81.
2. Remove the O-ring from the top of the reactor as shown in Figure 52.
O-ring
3. Remove the crucible from the combustion reactor (left furnace) using the appropriate tool
as shown in Table 53.
Crucible
4. Using the proper tool, remove the reactor turning it counterclockwise and simultaneously
pulling it upwards, as shown in Figure 54.
Reactor
removing
tool
The following operating sequence contains the instructions to remove the reduction reactor
from the furnace.
Material Required
Tool for special steel reactor
Proceed as follows:
1. Remove the autosampler front cover.
2. Remove the fixing nut as shown in Figure 55.
3. Remove the O-ring from the top of the reactor as shown in Figure 56.
Figure 56. O-ring
O-ring
4. Using the proper tool, remove the reactor turning it counterclockwise and simultaneously
pulling it upwards, as shown in Figure 57.
Sample Preparation
This chapter describes the technique for the sample preparation and the basic instructions to
weigh the sample.
Contents
• Sample Weighing Technique
• Weighing Technique for Solid Samples
To define the weighing range, you should know the nature (organic, inorganic,
metal-organic) and origin (pure chemical, natural product) of the substance to be analyzed.
Solid Samples
Solid samples are introduced directly into the tin container by using a spatula. According to
the sample quantity to be analyzed, refer to “Weighing Technique for Solid Samples” on
page 90.
Liquid Samples
The sampling will be according to the sample nature.
To introduce the sample, take some using a spatula and let it without touching the container
walls. Only for N Determination, according to the sample viscosity, it may be necessary to
adsorb the sample on a Chromosorb WAW® layer previously introduced into the tin container.
Material required
Tin disks
Capsulator
Cylindrical tool
Spatula
Tweezers
Balance
Tweezers Capsulator
Tin disk
Spatula
Cylindrical
tool
1. By using tweezers, take a tin disk and rest it on the cavity of the capsulator.
2. By using the cylindrical tool, press the tin disk and make it enter the cavity of the
capsulator, as shown in Figure 59.
3. By using a spatula, introduce the sample into the tin container until it is sufficiently filled
as shown in Figure 60.
4. Delicately close the tin container with your hand as shown in Figure 61.
5. By using the cylindrical tool, press the tin container filled with the sample into the cavity
of the capsulator, as shown in Figure 62.
6. Remove the container filled with the sample from the capsulator as shown in Figure 63.
Note The tin container filled with the sample should have a spherical shape. See
Figure 64.
Analytical Methods
This chapter describes the analytical methods used for the Flash 4000 elemental analyzer.
Contents
• Introduction
• Analytical Method for N/Protein Configuration
Introduction
The analytical method includes different subsequent steps to determine the percent
composition of Nitrogen through the transformation of the solid or liquid sample into gas.
CAUTION For correct sample analyses the pneumatic lines must be leak-free. Therefore, a
preliminary leak check is recommended before starting analytical cycles.
AS
R1 R2 WD F1 F2 CC
TCD
Diagram
An autosampler AS is connected to a special steel reactor R1 placed in an furnace at the
temperature of 950 °C.
This reactor is connected to a second reactor R2 placed in an furnace at the temperature of
840 °C. The R2 is connected to the condensate water drainage WD that is connected to two
adsorption filters F1 and F2. These are connected in parallel and working alternatively. The
outlet of the filter F2 is connected to the analytical column CC, on its turn connected to the
thermal conductivity detector TCD.
PI1
S2
F
He PRV1 He
SVP2 T
1
2
(5)
PI2
F
O 2 PRV2
F O 2
SV1
IN OUT (6) 2
Carrier Red
2
SV3 1 SV4 SV2 3 1
1
AS
2 2 2 Green
CV 1
S1
Combustion furnace
Reduction furnace
Vent
4
(4) 1
SVP1
(1) (2)
R1 R2
Reference
(7) (2)
Water
F F drain
Vent
2 2
Blue Blue
B
(5) F F (3)
2 2
3 1 1 3
SV7 SV6
(8) (1)
Reference helium He flows to the flow sensor S1, through the solenoid valve SV2. Both flow
sensors are connected to the relevant proportional valves SVP1 and SVP2.
The reference helium flows to the TCD through the flow sensor S2. Then, the reference
helium is used as purge gas by the sample into the autosampler.
The carrier helium is controlled by the flow sensor SV2. At the moment of the combustion,
the carrier helium is exchanged with the oxygen flow through the solenoid valve SV2.
When Start Analysis is pressed, and the combustion is started, the solenoid valve SV1 opens,
whereas the solenoid valve SV2 switches to allow Oxygen to flow in as far as the combustion
reactor R1 for a preset time.
The sample, weighed in a tin container and stored in the autosampler, is dropped into the
combustion reactor after a few seconds. Tin, coming in contact with an extremely oxidizing
environment, triggers a strong exothermic reaction.
Temperature rises to approximately 1800 °C instantly causing the sample combustion.
At the end of the time set for Oxygen introduction, the solenoid valves SV1 and SV2 return to
their original position restoring the Helium flow. The gas mixture (N2, CO2, H2O and SO2)
generated by combustion is conveyed across the reactor R1 where oxidation is completed.
Then the mixture crosses the reactor R2 where Nitrogen oxides possibly formed are converted
into elemental Nitrogen, and the Oxygen excess is retained.
The gases pass in the water drain WD where water is condensed and separated from the
stream. Then the CO2 filters F1 and F2 adsorbs the Carbon Dioxide and the water fraction
which has passed the WD. Nitrogen is then eluted in the chromatographic column CC and
conveyed to the thermal conductivity detector TCD that generates an electrical signal, which,
properly processed by Eager Xperience, provides the Nitrogen-Protein percentage.
Analysis
9
Instrument Set-up
This chapter contains information and operating sequences to set the instrument for
running analyses.
Contents
• Introduction
• Powering on the System
• Installation of Eager Xperience
• Eager Xperience Main Menu
• Analytical Configuration
• Edit Method
• The Elemental Analyzer Parameters
• Analyzer Status
• Detector Signal Level
• Leak Test
Introduction
To analyze any type of sample the instrument must be in the correct operating conditions.
Proceed according to the following operating sequences:
• “To power the system on” on page 102
• “To Install Eager Xperience” on page 102
• “To Configure the Analyzer” on page 106
• “To Check Leaks” on page 117
• “To Adjust the Detector Signal Level” on page 116
CAUTION Before starting the operating sequences, make sure that instrument, reactors,
regenerative CO2 filters, autosampler (or manual injection device for liquids) and any
complementary units are properly installed.
1. Switch on the instrument lifting the breaker, located at the back of the instrument, on the
position I. At the powering on, the indicating LED Power on the status panel lights up
green.
2. Switch on the computer and any complementary units by means of relevant switches.
Material required
Eager Xperience package
The software Eager Xperience can be installed in a system provided with Windows™
2000/XP/Vista operating systems. The free space on the PC hard disk must be at least 30 MB.
Eager Xperience is installed by using the CD provided in the standard outfit and operating as
follows:
1. When the CD is introduced into the CD driver of the PC, the installation menu shown
in Figure 67 is displayed.
Note If the installation menu does not automatically appear, start the CD Autorun
program through the Windows™ Start-Run command.
2. Start installation by clicking the push-button Install Eager Xperience for Flash.
3. Follow the instructions prompted step by step.
4. At the end of installation, in the page Start-Program Eager Xperience, double-click the
Eager Xperience for Flash icon. The window of Figure 68 is displayed.
Analytical Configuration
The analytical conditions are set in our laboratories during the final test of the analyzer. To put
the analyzer in operating conditions, you only have to follow the instructions reported in the
next operating sequence.
1. In Main Menu of Figure 69 on page 103, choose File > Instrument Configuration. The
following window is displayed.
a. In the section Instrument name, type the instrument serial number (6 digits; for
example 991234). See the label located on the instrument rear panel.
b. In the section Analytical instrument configuration select N/Protein option.
Note The option Undefined can be used in case the desired operating conditions
are different from those defined for the instrument configuration.
c. Click Elemental analyzer setup to enter the dialog window of Figure 71 where the
configuration parameters have to be set.
d. In the section Elemental Analyzer Connection select the computer serial port
(COM1, COM2 etc.) to which the instrument is connected.
If the LAN control of the Flash 4000 is required, select network address and enter the
I.P. address. Use Advanced button to set the LAN communication port used by the
TCP-IP protocol and timeout.
e. In the section Type verify that the instrument in use is Flash 4000.
f. In the section Instrument Settings choose the following settings:
i. Line Frequency = 50 Hz
ii. TCD Settings Source = Internal
iii. TCD Settings Polarity = Positive
g. In the section Sampler Setting select the type of autosampler installed on the
instrument.
i. In the case of autosampler for liquid sample, also specify the computer serial port,
to which the autosampler is connected, and the number of vials.
ii. Click Ok to go back to the window of Figure 71, then click Ok to return to Main
Menu.
Edit Method
In Main Menu choose Edit > Edit Method or just click the icon . The Edit Method dialog
box is visualized.
Method editor tabs allow to modify the method parameters and to access the following
sections:
Method Title
This tab allows to add a comment to the method. This comment will be displayed in Main
Menu and saved with the parameters of the analysis.
Detection Parameters
Detection parameters is the section of the analytical method that contains all the parameters
for the data acquisition.
Function Description
Time Base This box specifies if you want to work with a time base in seconds
or minutes.
Actual Sample This box indicates the sample being acquired.
Baseline Background It permits to compensate the acquired chromatogram with the
Compensation baseline background.
Integration Parameters
Integration parameters is the section of the analytical method that contains all the parameters
for the integration.
This dialog box contains all the parameters used for the integration algorithm. The
integration algorithm is based on the following parameters.
Parameter Description
Peak Width It specifies the width of the peaks to be integrated.
Peak Threshold It is used by the integration algorithm to discriminate the peaks
from the baseline or the signal noise.
Parameter Description
Minimum Area The peak having an area lower than the entered value are rejected.
Skim Ratio It determines whether overlapping peaks are integrated by using
tangential skim or perpendicular drop methods.
Calculation Parameters
Calculation parameters is the section of the analytical method that contains all the parameters
for the calculation of the element percentage.
Function Description
Calibration Method It allows to specify the calibration method by which the analyzed
sample will be processed. See Instrument Calibration.
Heat Value It allows to perform heat value calculations or to estabilish the
Calculation calorific content of solid or liquid samples in an automatic manner.
Protein It allows to perform the calculation of the protein content a sample
in a fully automatic manner.
CO2 Emission Factor This option allows to perform the CO2 emission factor either for
liquid or for solid samples in an automatic manner. The calculation
of the Carbon Dioxide emission factor is based on the percentage
composition of the elements and on the Gross Heating Value.
Report Parameters
Report parameters is the section of the analytical method that contains all the parameters for
the report.
Function Description
Report Type Choose among the following report types:
• Standard report
• Extended report
• Summary report
• Custom report
Include Signal to It permits to choose the peak of interest and the baseline portion
Noise Report on which the noise evaluation should be done.
Include Calibration Check this option to have the printout of the samples calibration
Report report for each of the calibration levels.
Append for
Summarize Click the icon and select the result to append for summarize,
normally Element %.
Standard Report On Choose among the following reports:
• None: It will not generate any report at all.
• Printer: It will send the report directly to the printer.
• ASCII: It will send the report to an ASCII disk file.
• DIF: It will send the report to a DIF file.
• Excel: It will send the report to a XLS file.
• HTML: It will send the report to HTML format.
Concentration Unit Normally %
Report Title It permits to enter a text to be printed as a title of the report
generated by the system.
Report Publisher It is the tool to create or modify the file report layout.
Report Stripchart
Report stripchart permits to select and to set the stripchart options to be included in the
report printout.
Function Description
Stripchart Options This box contains the parameters used for the generation of the
stripchart report.
Stripchart Title This box permits to enter a text to be printed as a title of all
stripchart generated by the system.
Strichart Annotation Check the options of interest.
Operator ID/Info
In this tab the operator can store the informations for future reference.
1. In Main Menu, choose File > Instrument name and configuration and enable the
elemental parameter setting by ticking the relevant check box as shown in Figure 73.
2. In Main Menu, choose Edit > Edit Elemental Analyzer Parameters or just click the icon
. The following tabs are visualized where the analyzer operating parameters are
displayed. Refer to Table 21.
Tab Description
Oxygen/CO2 Traps In this tab it is possible to set:
• the trap heating time in minutes
• the trap cooling time in minutes
Analyzer Status
This fuction allows to visualized the status of the elemental analyzer through five tabs. This
function is accessible from Main Menu choosing View> View Elemental Analyzer Status or
just clicking the icon .
1. In Main Menu choose View > View Elemental Analyzer Status or just click the icon .
The dialog window of Figure 75 is displayed. Select the tab Detector.
2. Click Auto Adjust Level at 1000 μV. At the end of the operation, the value 1000 is set
representing the analysis starting point.
Leak Test
The leak test must be performed any time a component of the pneumatic circuit is replaced.
Operate according to the instructions reported in the following operating sequence.
Y To Check Leaks
1. In Main Menu choose View > View Elemental Analyzer Status or just click the icon .
The dialog window of Figure 76 is displayed. Select the tab Special Function.
2. In the section Command click Leak Test. The dialog window with the leak test
parameters is displayed as shown in Figure 77.
a. Click Start to begin the operation. A window will appear where you will be requested
to perform the leak test with the CO2 trap regeneartion on the last use.
b. A second window will appear where you will be requested to perform the Autozero.
c. After some seconds (Leak test time), Carrier flow and Reference flow must be
within 0 and 5mL/min. Higher values indicate that the system is not leak-free.
Note Leaks in the system are generally due to incorrect closure of the reactors
locking nuts.
3. To terminate the leak test and restore the flow operating values, click Stop and Done.
Instrument Start-up
This chapter contains information and operating sequences to prepare the instrument for
running analyses.
Contents
• Start-up
CAUTION Before starting the following operating sequences, make sure that instrument,
reactors, autosampler (or manual injection device for liquids) and any complementary
units are properly installed.
Start-up
Y To Start-up the instrument
1. In Main Menu choose Edit > Edit Elemental Analyzer Parameters or just click the
icon .
2. Select the Flow/Timing tab and set 300 mL/min for both carrier and reference flows as
shown in the following Figure 78.
Set 300
Set 300
3. Click Send to trasfer the operating parameters to the instrument. Click Ok to exit.
4. In Main Menu choose View> View Elemental Analyzer Status or just clicking the icon
.
5. Select the Special function tab and start the leak test by clicking Leak test as shown in
Figure 79. For detail refer to “Leak Test” on page 116.
Set temperatures
8. Click Send to transfer the operating parameters to the instrument. Click Ok to exit.
9. In Main Menu choose View> View Elemental Analyzer Status or just clicking the icon
.
10. Select the Detector tab and turn on the filaments of the TCD as shown in Figure 81.
Turn filaments On
11. Wait the Ready of the instrument and perform again the leak test. For details refer to
“Leak Test” on page 116.
12. Wait the stabilization of the TCD signal, then adjust the detector signal. For details refer
to “Detector Signal Level” on page 116.
13. It is possible to start the analysis.
Contents
• Introduction
• Directory for Analyses
• Current Maintenance Program
• Instrument Calibration
• Sample Table
• Sequence of Analyses
• Quality Control and Check of Analytical Results
• Post-Analysis Operations
• Analytical Troubleshooting
Introduction
To program and analyze any type of sample, do the following:
• Create a directory for analyses
• Program the current maintenance (recommended operation)
• Choose the calibration
• Set a sample table
• Run the sequence of analyses
• Determine the blank value
CAUTION Before starting operating sequences, make sure that the instrument start-up
operations have been performed as described in Chapter 9.
1. In Main Menu select View > View Maintenance; a window like the following one is
displayed.
Component Description
Left 1 It represents the crucible.
Left 2 It represents the oxidation reactor.
Right It represents the reduction reactor.
The example above shows the analytical circuit components for which the maintenance
routine is required.
a. In the diagram the active components are indicated with colored areas and show on
the left the numerical scale of their lifetime. The meaning of each colored area is
indicated in the upper section of the diagram.
b. The components not present in the concerned instrument configuration are indicated
by a dashed line.
2. To view in detail the default conditions of the components of the concerned instrument
configuration, choose Edit > Set Maintenance >Default. A window like the one shown
in Figure 83 is displayed. The values shown cannot be changed.
3. To use a different maintenance program from the default one, choose Edit > Set
Maintenance > Manual. A window like the following will be displayed.
a. In the window of Figure 84, you can change any value by clicking on the different
boxes and entering the desired new value.
Instrument Calibration
Eager Xperience offers three calibration methods:
• K-Factor
• Linear
• Non Linear
All tests are performed with the K-Factor method that is generally used by most users. This
method consists in obtaining a constant of calculation by means of the following formula:
K =%Th*(I-b)/w
where:
For the calculation of an Unknown sample, Eager Xperience will use the reverse formula:
%Unknown = K*w/(I-b)
where:
K = K-factor
w = Weight in milligrams of the sample
I = Area integral of the sample
b = Blank area integral
The Linear method is generally used when samples very different from each other are
analyzed in the same analytical sequence. In this way the errors due to the detector response
linearity are minimized.
The Non-Linear method is used when the analyzer is connected to another detector having a
response of exponential type. To select the calibration method or view the calibration curves of
a memorized method, do as described in the following operating sequence.
b. Select Calib. Method. According to the calculation method, the calibration points
with peak area and concentration will be displayed.
Sample Table
The sample table of the analytical method contains all information concerning the series of
samples to be acquired and processed. To enter the sample table, do as described in the
following operating sequences.
In Main Menu select Edit > Sample table or just click the icon . Sample Table will be
displayed.
1. Select Edit sample >Fill Sample table or just click the icon . The following window
will be displayed, which offers all the functions necessary to fill, change or cancel Sample
Table.
iii. In the box Number samples, enter the number of samples, up to 200, to be
analyzed.
iv. Check that Unknown is ticked
2. Leave Sample name idx and Filename idx set as 1.
a. In the section Sampler, enter the parameters if the instrument is equipped with the
autosampler for liquids.
b. In the left bottom section, enter the following.
i. Leave Weight set as 1.
The weight is entered time by time. In case of direct injections of constant
volumes, the sample volume can be entered directly.
ii. Set Protein factor.
3. At the end of editing, click Replace. The sample table appears again. In the sample table
grid you will find all information entered.
Note In the case of liquid samples analyses with the autosampler for liquid, the
column Density will be displayed in the sample table grid in which insert the density
of the liquid sample that will automatically be turned into weight.
Figure 89 shows an example of edited sample table referred to the N/Protein instrument
configuration.
The sample weights may be manually entry or automatically transferred from the balance to
the PC. These two modes are explained below.
• “Mode 1: Manual Entry of the Weights” on page 130
• “Mode 2: Automatic Entry of the Weights from the Balance to the PC” on page 130
3. Select the computer serial port to which the balance is connected. Pay attention that the
serial port must be different respect the serial port selected for the analyzer.
4. Select Balance menu.
a. Choose the type of balance in use.
b. Press OK. The dialog window of Figure 90 is displayed again.
c. Press OK to return to Sample Table.
5. In Sample Table select Balance > Receive weight. After you have selected the sample
number desired (e.g. sample number 3), click on Weight. Pressing the down arrow the
following window is visualized:
6. Put the container on the balance plate. Perform the tare pressing Tare. Introduce the
sample then wait for stabilization.
7. Clicking Weight, the value of the weight is automatically transferred. The sample table is
ready to acquire the value of the next weight.
Note If the balance Mettler Toledo AB54S is used, perform the tare by using the
command located on the balance control panel.
1. In the sample table grid, click on the column Type. Click the arrow shown to display the
window for the sample type selection.
2. According to the sample type, select Bypass, Standard, Unknown or Blank as shown in
Figure 92.
Note Selecting Standard, the following window appears. Set here the desired standard.
Note New standards may be memorized by clicking Edit Standard Table. After the
number 5, add the standards of interest and the percentages of the relevant elements.
Refer to the following Figure 94.
To obtain precise and reproducible results the sample must be completely oxidized.
If the major objective of the analysis is to obtain the best precision of results, it is particularly
important that catalysts, and specially Copper, last as long as possible.
Therefore you have to establish how much Oxygen is required to burn a sample of that
particular nature as a function of its weight.
For example, chocolate requires more Oxygen than a soil or an organic substance, though at
equal weight.
We recommend to always use the OxyTune® function (Automatic Oxygen Dosing System).
Selecting, after the weight entry, the category to which the sample belongs, the system will
deliver the right Oxygen quantity required for a complete combustion. This condition is
obtained by a feature provided thanks to the method below. It is important to select the
correct sample category to avoid an incomplete combustion (not enough Oxygen) and the
premature exhaustion of the Copper.
Eight fixed categories can be choose in order to have a correct Oxygen dosage for different
sample matrices. There are:
• Glucose
• EDTA
• Pasta, flours, beans, gluten, cereals
• Starch
• Cocoa
• Ham
• Salame
• Cheese
The selection is performed by clicking the category of interest. For special samples do not
included in the previous categories it is possible to develop a specific category.
Note This operating sequence gives you a general indication to set the correct category
according to the sample nature.
Flows
Flow 1 and Flow 2 are the flows of Oxygen. Both can assume the following values: 125, 250,
500, 750, 1000, 1250, 1500 mL/min.
• If the sample is rich in fat
In this case the combustion could happen with difficulty with the risk of cracked. To
avoid the cracked it is suggested to set a low flow of Oxygen (125 - 1000 mL/min).
• If the sample is rich in water
In this case the combustion happens more easily. It is suggested to set a low flow of
Oxygen and an high Oxygen Delay time.
Time Ratio
Oxygen Caps
It is the quantity of Oxygen required to oxidize the tin caps. For the 63-mm tin caps it is 0.1
liters.
Oxygen Sample
Oxygen Delay
When this time is elapsed, the Oxygen is injected into the circuit.
Purge
It is the duration of the sample purge. It depends on the concentration of nitrogen in the
sample. If the percentage of nitrogen in the sample is lower than 0.5%, it is suggested to set
Trace Purge to 50 s.
1. In the Humidity column enter the percentage of humidity present in each sample when
known.
Sequence of Analyses
CAUTION Before shipment, every instrument is submitted to an analytical final test
procedure according to the concerned instrument configuration.
The results of this test are included in the documentation set accompanying the
equipment. When the instrument is used for the first time, before analyzing any unknown
sample, it is advisable to repeat the test maintaining the selection of standards and their
weighing range.
To correctly run the analysis, proceed as described in the following operating sequence.
4. Select Bypass, then click OK to confirm and go back to the Sample Table.
5. Repeat the step 3 for all the other samples selecting the proper type. Then click OK to
confirm.
6. At the end of Sample Table editing, to automatically stop the sample sequence, select the
number following the last edited sample.
Select Edit sample > Insert line or just click the icon .
7. In Main Menu, select Run > Start sequence of sample.
8. Click Start Now.
At the end of the analytical sequence, the results obtained must be compared with those of the
final test, doing what described in the operating sequence “Comparison Between Analytical Re-
sults and Final Test Results” on page 135.
At the end of the analytical sequence, the results obtained must be compared with those of the
final test, doing what described in the operating sequence “Comparison Between Analytical Re-
sults and Final Test Results” on page 135.
Interpretation of Results
If the results obtained are satisfactory, go on with your samples sequence. If on the contrary
the results are not correct, try to identify the cause and find the remedy.
Tip It is suggested to see Chapter 7, “Sample Preparation,”
The cause of the error is generally due to incorrect sample weighing. Always observe the
indicated weighing ranges using, if possible, the direct connection between balance and
computer, selecting in the Sample Table the option Receive Weight from Balance in Balance
menu.
3. If the weight of an unknown sample is wrong, the error immediately becomes clear from
its percentage result. On the contrary, if the error is due to the weight of one or more
standards, the three results of the unknown sample will all be wrong.
4. Another cause of error is the incorrect integration of the peak. If it happens, the correction
of the baseline is required proceeding as follows:
a. Open the chromatogram selecting the sample to adjust. In Main Menu choose View
> View Chromatogram > Peak.
b. To modify the baseline, choose the option Move peak start or Move peak end
accordingly.
c. Save the new chromatogram.
d. Recalculate the chromatogram following the instruction in sub-paragraph Sample
Recalculation.
Sample Recalculation
1. In Main Menu select Recalculation > Recalculation. The relevant dialog window is
displayed.
The blank value is a function of the type of containers used and of Oxygen purity. Check that
the values found are within acceptable limits versus those reported in the final test certificate.
If the values found are higher, refer to the following Table 24.
IMPORTANT All blank values are memorized and subtracted to the sample values. As a
consequence, definitely high values may affect analytical precision.
After having analyzed the standard samples and checked the instrument precision, do the
following:
1. In Main Menu, select Recalculation > Summarize results or just click the icon . The
Summarize result table is displayed. See Figure 96 on page 136.
2. Select Edit > Select reference compounds. A table like the following will appear.
b. In the text box Nitrogen enter the supposed theoretical percentage of the element to
be analyzed.
c. Click OK. For each value a minimum-maximum percent deviation (box +%) versus
the entered value is automatically calculated. The calculation of this deviation is made
by an algorithm, which takes into account the entered percentage and the
instrumental error.
Note If the acceptable error range is wider than that automatically calculated, it
can be manually changed.
At the end of the analytical cycle, to compare the values found with the theoretical ones, do
the following:
1. Go to the table of Figure 99 doing as described in points 1 and 2 of the previous operating
sequence. Select the sample to be compared then click OK. The Summarize Results table
is displayed.
2. Select the sample to be compared as shown in the example of Figure 99.
a. Select View > Compare to reference compound. On the left side over the
Summarize Results table, as shown in Figure 100, you will read the name of the
analyzed sample followed by one or more LEDs according to the number of analyzed
elements.
Green light = Indicates that the values found are close to the theoretical values and
within the preset error limits.
Red light = Indicates that the values found are far from the theoretical values.
If a single element is analyzed, the result can be graphically displayed. This is particularly
useful when the sample is routinely analyzed and when it is more important to have the result
within the preset error limits rather than the absolute value found.
c. If the analyzed sample, represented in the graph by a red square, is within the two
green lines, it means that, independently of its absolute value, it has a value close to
the theoretical one and is within acceptable error limits.
The red square, when clicked, becomes white and the percent value is displayed, as
shown in Figure 103.
Post-Analysis Operations
To obtain precise accurate and constant in time results, and at the same time to reduce
operating costs, we recommend you follow some practical suggestions.
When the instrument goes in Standby Mode, the deactivated adsorption filter is regenerated.
After the regeneration the carrier flows through both adsorption filters and the filter purge is
stopped.
The Standby function can be activated manually or automatically at the end of the analytical
sequence.
Proceed as follows:
1. In Main Menu, select View > View Elemental Analyzer Status or just click the icon .
The page of Figure 104 is displayed.
Proceed as follows:
1. In Main Menu, click the icon . The window of Figure 105 is displayed.
2. In the section Elemental analyzer conditions while start sequence is finished, enable
the Force to Stand-by function by ticking the appropriate check box.
The analyzer will automatically go to the Standby condition, when the last sample has
been analyzed.
1. In Main Menu, just click the icon . The window of Figure 106 is displayed.
2. In the section Elemental analyzer conditions while start sequence is finished, enable
the function Shut-Off temperature, detector and gas by ticking the appropriate check
box.
3. When the oxidation reactor requires cleaning, the furnace should be switched off and
then the Helium flow reduced using the Stand-by function.
a. In Main Menu, select Edit > Edit Elemental Analyzer parameters or just click the
icon . The window of Figure 107 on is displayed.
b. In the section Furnaces, tick the box Left Furnace to enable the Off condition.
c. In the section Other, enable the function Set instrument to Stand-by by ticking the
appropriate check box.
d. Click Send to send the command to the instrument, then click OK.
a. In the section Set Wake-up, set the desired date and time of activation of the
Wake-up function.
b. Click Activate and then OK to confirm.
Auto-Start Function
If you desire Auto-Ready to be followed by Auto-Start, programming an analytical sequence,
do the following:
2. Enable the function Enable time programmed sequence start by ticking the appropriate
check box.
a. In the section Starting time click Now.
b. In the Start Date/Time text box enter the date and time of the function activation.
IMPORTANT The function activation should be programmed with a delay of at least
60 minutes versus the time programmed for the Auto-Ready Function to allow the
analyzer to reach a good thermal equilibrium.
Analytical Troubleshooting
If the instrument has been correctly installed, the gas characteristics are as required and
maintenance has been regularly carried out, Flash 4000 will provide correct results.
The lack of the above conditions will be indicated by anomalies in the chromatograms and the
relevant analytical reports. The following table reports the most common anomalies with the
relevant diagnosis and remedy.
Maintenance and
Troubleshooting
12
Maintenance
This chapter provides information on the current and periodic maintenance of the
instrument, and it also contains the operating sequences for installation and maintenance of
the MAS 4000 autosampler.
Contents
• Instrument Maintenance
• Reactors Replacement
• Replacement of the Gas Chromatographic Column
• Servicing the MAS 4000 Autosampler
WARNING If, for technical reasons, it is necessary to work on parts of the machine that
may involve hazardous operations (moving parts, components under voltage, etc.).
Thermo Fisher Scientific authorized Technical Service has to be called.
This situation can be identified because the access to these moving parts is possible only
using a tool and because the concerned removable protective covers bear a warning symbol
that draws the operator’s attention to the specific warnings included in the documentation
accompanying the instrument. In case the work has to be carried out by the operator, the
latter must prove to be adequately trained to perform the specific maintenance operation.
WARNING When the instrument is switched off, consider that its does not cool down
immediately, but heat tends to concentration in the upper part of the furnaces area. The
openings provided for the chamber aeration will cause a slow cooling of same, which
however, in the vicinity of the areas marked with the symbol “hot surfaces”, might even
reach temperatures higher than ambient temperature. Therefore in the minutes
immediately following the instrument switching off, the operator must consider this risk
and pay adequate attention during any maintenance operations following the use of the
instrument.
Instrument Maintenance
The instrument will be generally serviced by Thermo Fisher Scientific authorized technical
personnel for all the warranty period or, after warranty, possibly according to a Programmed
Service Contract. For more information contact your local Thermo Fisher Scientific office.
Current Maintenance
Replacement of crucible, reactors and their filling material.
The instrument configuration uses special steel reactors for combustion it may be necessary
also to clean the crucible done with same material.
• Periodic Maintenance
Replace the o-ring of the reactors coupling unions placed on the furnace compartment
base.
For some maintenance operations, furnaces and oven need to be at room temperature. Follow
the instructions given in paragraph “Shutting Off Furnaces, Detector and Cutting Off Gas
Flows” on page 143.
Reactors Replacement
The replacement of reactors is performed when required by the maintenance program
schedule according to the setting entered in paragraph “Current Maintenance” on page 155.
Replace and install reactors according to the operating sequences described in Chapter 6
“Reactor Preparation and Connections” on page 73.
Perform these operations according to the instructions given in the following operating
sequences.
Material Required
Tool for cleaning special steel reactors
Filling materials
WARNING Before starting the operation, check that the furnaces are at room
temperature and the gases are turned off.
1. Remove the special steel reactor from the furnace following the instructions given in the
operating sequence “To Remove the Combustion Reactor” on page 85, then do the
following operation.
WARNING Before starting check that the gases are turned off.
Material Required
Tool for cleaning quartz reactors
Ultra temperature ceramic disks
1. Remove the crucible following the instructions given in the operating sequence “To
Prepare the Crucible” on page 78, then do the following operations:
WARNING The crucible can be removed paying proper attention even when the reactor
is hot. Rest the crucible on a surface resistant to heat and remove ashes only when the
crucible is cold.
Material Required
XX-inch wrench
Screwdriver or spatula
Spare seals
WARNING Before performing this operation, check that the furnaces are at room
temperature and the gases are turned off.
4. Disconnect the coupling unions as shown in Figure 111, and rest them on a clean surface.
5. By using a small screwdriver or a spatula, remove the o-ring from each union, as shown in
Figure 112.
6. Put a new o-ring into each union making sure, by using an appropriate tool, it correctly
fits its seat, as shown in Figure 12-10.
7. Reconnect the coupling union and remount it in its seat as shown in Figure 114.
WARNING Before starting check that the gases are turned off.
Material Required
Open end wrenches for the column fittings
1. Access to the side compartment removing the right side panel. The chromatographic
column is located in the low temperature analytical section as shown in Figure 115.
Chromatographic
column
Low temperature
analytical section
2. By using the wrenches for column fittings, disconnect the inlet and outlet tubings
unscrewing the relative retaining nut, then remove the column from the fixing clips as
shown in Figure 116.
3. Fix the new column with the fixing clips, then reconnect the inlet and outlet tubings.
4. Finger-tighten the retaining nuts until them start to grip the column.
5. Use the wrenches to tighten the retaining nuts. Use no more pressure than is necessary to
obtain a good seal (1/4 to 1/2 turn).
This operating sequence provides instructions for the MAS autosampler maintenance.
Material Required
Dry clean cloth
Sample room
O-Rings
Rack
6. Re-introduce the piston into the autosampler until in place keeping its rack turned
downward, as shown in Figure 119.
7. Slightly pushing the piston with one finger, as shown in Figure 120, click Piston
movement again. The autosampler mechanism first tries to eject the piston, then, on the
motor reversal, the mechanism will hook the piston and draw it inside the autosampler.
8. Click Ok.
9. Turn down the autosampler front cover.
Troubleshooting
This chapter provides information necessary to find out instrument troubles and to solve
them.
Contents
• Safety Cut Off
• EFC-t Module
When lit, this LED indicates that the furnaces and detector oven power has been cut off for
safety reasons.
The Safety Cut Off status is followed by an error message about the possible cause of error.
Proceed as follows:
1. In the Main menu select View > View Elemental Analyzer Status or just click the icon
.
2. In the displayed dialog window, select the option Special Functions. The dialog
window of Figure 121 is displayed.
3. Read the error message in the reading box located on the lower right side of the window,
below the buttons Help and OK.
4. Refer to Table 26 to find out the error status and have more information.
1
1. Error Message Reading Box
The following Table 26 reports the error messages and the explanation of the relevant
correlated problem.
safety cut off Right furnace has exceeded the set temperature limit.
out of limit right Temperature has been set with furnace missing
safety cut off Left furnace has exceeded the set temperature limit.
out of limit left
safety cut off The detector oven temperature exceeds the 220 °C.
Oven limit Error may be due to a probe malfunction, etc.
CAUTION The error message is generally due to the specific cause indicated. Sometimes,
it may generated by different electric factors or caused by failures not depending on the
system. In this case contact the Technical Service.
EFC-t Module
The failures that may be generated on the EFCt Module are connected to the breakage or to
the malfunctioning of solenoid valves and flow sensors.
Refer to the Table 27 to find the component responsible of the EFCt module malfunctioning
and to solve the relevant problem.
Customer Communication
Thermo Fisher Scientific provides comprehensive technical assistance worldwide and is
dedicated to the quality of our customer relationships and services.
How to Contact Us
This appendix contains contact information for Thermo Fisher Scientific office. To contact
your local Thermo Fisher Scientific office or affiliate, please refer to:
Voltaweg 22
2627 BC Delft, the Netherlands
Tel: +31 15 257 13 14
Fax: +31 15 257 22 97
This appendix also contains a one-page Reader Survey. Use this survey to give us feedback on
this manual and help us improve the quality of our documentation.
Reader Survey
Product: Flash 4000 Elemental Analyzer
Manual: Operating Manual
Part No.: 317 110 55
Please help us improve the quality of our documentation by completing and returning this
survey. Circle one number for each of the statements below:
If you would like to make additional comments, please do. (Attach additional sheets if
necessary).
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
Abbreviations ft foot
g gram
B byte (8 b) He Helium
baud rate data transmission speed in events per second HPAR High Performance Alloy Reactor
d depth K Kelvin
ºF Fahrenheit lb pound
mL o ml milliliter w width
mm millimeter W Watt
m/z mass-to-charge ratio When a unit of measure has a quotient (e.g. Celsius degrees
per minute or grams per liter) this can be written as
N nitrogen
negative exponent instead of the denominator:
n nano (10-9) For example:
°C min-1 instead of °C/min
O oxygen g L-1 instead of g/L
Ω ohm
p pico (10-12)
Pa pascal
PN part number
RF radio frequency
s second
S sulphur
G O
Oxygen Caps 133
Gas Chromatographic Column
Oxygen Samp 133
Replacing 153
Gas Connection
Fittings 28 P
Gas inlets 39 Pneumatic Circuit 45
Gas Supply 40 Components 47
Pneumatic Compartment 22
I Pneumatic Connections 56
Pollution degree 15
Installation 55, 55
Post-Analysis 142
Instrument
Classification 15 Power On 102
Cleaning 17 Power Supply 38, 39
Configuration 17 Pressure Gauges 47, 47
Safety 17 pressure gauges 25, 45
Sleeping Mode 142 Pressure Regulators 47, 47
Instrument Configuration 106 Pressure regulators 25
Instrument Data System 23
Integration parameters 124 Q
Interface 39, 40
Quality Control 138
IP Address 40, 63
Default 63
R
L Reactors 21
Filling Materials
Laboratory Temperature 56
Replacement 150
LAN Communication Port 63
reactors 45
LAN Option 39, 40 Reactors Coupling Unions 151
LAN Set-up 63
Recalculation 135
Leak Test 116
Reference Compound Selection Table 138
Liquid Samples 89 Results
Low Temperature Analytical Section 32 Graphic Display 140
Low Voltage Electrical Section 33 Interpretation 136
lower door 28 Right Front Compartment 25, 30
Right Side Compartments 32
M Run Analyses 123
Mains Voltage Components 38
Maintenance 124, 149 S
Current 150 Safety Cut Off 157, 157
Default Conditions 125
Safety Cut Off Device 17
Periodic 150
Safety Information 16
set 125
Nominal Pressure of Gases 17
Start-up 124
T
TCD chamber 32
Technical Features 16
Temperature 15
Theoretical Values
Comparison 139
Thermal Conductivity Detector (TCD) 22, 45
Time Ratio 133
Transients 15
Troubleshooting 157
V
Valves manifold 45
Voltage variations 15
W
Wake-up Function 144, 144
Water Condensate Drainage 50