You are on page 1of 1

SERVICE MANUAL FOR INFANT RADIANT WARMER

TABLE 3.1 SPECIFICATIONS (CONTINUED)

ALARM(See table 3.2)


The internal probe of skin temp sensor fault………………………………skin temp sensor alarm
The internal isolated probe of skin temp sensor fault…………isolated sensor of skin temp alarm
The deviation of the main probe and isolated probe of skin temp sensor is more than
0.8°……………………………………………………………the skin temp sensor differences alarm
The internal probe of mattress temp sensor fault……………………mattress temp sensor alarm
The internal isolated probe of mattress temp sensor
fault………………………………………………………isolated sensor of mattress temp alarm
The deviation of the main probe and isolated probe of mattress temp sensor is more than
0.8°…………………………………………………………mattress temp sensor differences alarm
Skin sensor over-temp alarm…………………………………………………skin over-temp alarm
Mattress sensor over-temp alarm………………………………………mattress over-temp alarm
The upper deviation fault tested by skin sensor………………………skin upper deviation alarm
The lower deviation fault tested by skin sensor…………………………skin lower deviation alarm
The upper deviation fault tested by mattress sensor……………mattress upper deviation alarm
The lower deviation fault tested by mattress sensor………………mattress lower deviation alarm
In baby mode, the temperature measured by skin sensor is lower 0.5 than the setting
temperature, and the deviation alarm does not active……………check skin temp setting alarm
In manual mode, prompting every 15min………………………………………check manual mode
Main CPU internal system failure……………………………………………………system alarm 01
Main CPU internal ROM failure……………………………………………………system alarm 02
Main CPU internal EEPROM failure………………………………………………system alarm 03
Main CPU and auxiliary CPU communication failure……………………………system alarm 04
Auxiliary CPU internal system failure………………………………………………system alarm 05
Auxiliary CPU internal ROM failure…………………………………………………system alarm 06
Auxiliary CPU internal EEPROM failure……………………………………………system alarm 07
Main CPU and independent protection CPU communication failure……………system alarm 08
Independent protection CPU internal failure………………………………………system alarm 09
Independent protection CPU internal ROM failure………………………………system alarm 10
Independent protection CPU internal EEPROM failure……………………………system alarm 11
Radiation heating system failure……………………………………………………system alarm 12
Mattress heating system failure……………………………………………………system alarm 13
Main CPU external ADC failure……………………………………………………system alarm 14
Auxiliary CPU external ADC fault……………………………………………………system alarm 15
independent protection CPU external ADC failure………………………………system alarm 16
SRAM failure…………………………………………………………………………system alarm 17
The clock chip RTC failure…………………………………………………………system alarm 18
Battery failure…………………………………………………………………………system alarm 19
NOTE: For HKN-2001, when system failure, the indicator shows the system alarm content. For
HKN-2001L, when system failure, the indicator shows the system alarm code.