Professional Documents
Culture Documents
Manual Centrifuga Unico-Powerspin-C8624-24
Manual Centrifuga Unico-Powerspin-C8624-24
PowerSpin™ C8624
24 Place Centrifuge
OPERATORS MANUAL
1
Introduction
Features
Safety Precautions
Cleaning Centrifuge
Problem Solving
2
Introduction
This is a bench top centrifuge with a space saving design and intuitive controls.
The unit has provision for 24 tubes in the rotor, cycle timer, automatic lid lock and
release. A safety device is installed to prevent the lid from opening while the centrifuge is
running. Each PowerSpin™ C8624 centrifuge comes equipped with an angled rotor with
24-place tube shields already installed.
Items included
Operators Manual
Centrifuge
Power Cord
Other
Digital Timer 1 – 60 Minutes
Rotor 24 place angled
Capacity 24 x 10mL (nominal)
3 – 15mL tube range
Dimension (L x W x H) 13 3/4“ x 15” x 11”
Net Weight 31 lbs
Features
3
Memorized time and speed settings: The centrifuge will memorize the last time
and speed settings when the unit is turned off. It will resume the previous time
and speed settings when the unit is turned on.
Maintenance free brushless AC motor.
Safety: The centrifuge lid cannot be opened until the unit stops spinning. It will
not run unless the lid is properly closed.
Suction cup rubber feet for stable bench top operation.
Power switch with indicator light tells when unit is powered on.
Power on/off switch located on back of unit.
Inspect upon receipt. If you note any damage, please contact Ascend Clinical.
Save the original packaging. If it becomes necessary to return the centrifuge to
us, we ask that it be returned in the original packaging.
Place the centrifuge on a level, horizontal surface.
A safety distance of 12 inches is needed around the centrifuge during operation.
No objects which could cause additional damage, or could themselves be
damaged in the event of a centrifuge crash should be positioned in this space.
Before plugging in the centrifuge compare your main supply with the electrical
requirements of the centrifuge.
The rotor is protected by cushion foam in the rotor chamber. Do not operate the
centrifuge before the foam is removed. This special packing inside the chamber of
the centrifuge prevents the rotor and shields from moving during shipping.
1. Use a small screw driver (or a rod with .45mm or less in diameter) to open the lid.
Insert the tool into the emergency lid-opening hole on the up-right corner on the
left side of the front panel of the centrifuge. Push the tool to release the safety
lock and open the lid.
4. The power inlet is located in the back of the centrifuge. Plug in one end of the
power cord to the inlet and the other end to an 115V power outlet.
4
Safety Precautions
Always
The rotor and lid must always be securely fastened.
If the centrifuge makes unusual noises when started, or exhibits excessive
vibration, switch the unit off immediately by turning the power switch OFF.
The centrifuge must be loaded symmetrically.
Do not use centrifuge with wet hands.
Take all precautions when using the centrifuge for toxic, radioactive, or material
with pathogenic micro-organisms, with respect to your safety and the safety of
others.
Do not place any sample containers or test tubes directly into the rotor.
Reminders
If any function of the centrifuge does not operate correctly do not use the
centrifuge.
Do not use a damaged centrifuge.
Do not move the centrifuge during the run.
After prolonged use, or if the motor was stalled, the bottom of the centrifuge may
become very hot. Handle the centrifuge by the sides only.
5
Operation & Control Panel
CAUTION!!
Make sure that the specimen tubes are balanced.
Always balance rotor with tubes of equal size and equal fill.
Note: when the centrifuge run is complete, the timer will be automatically reset to
the last timer setting and the centrifuge is ready for the next run.
Timer setting
Use the up and down arrow to preset the running time. The timer range is from
1 to 60 minutes. The display will show counting down time.
Speed setting
Use the up or down arrow to preset the running speed.
The speed range is from 1,000 RPM to 3,400 RPM in 100 RPM increments.
The display will show actual running speed and RCF (g) level.
6
Start button
Press Start button to spin sample.
The centrifuge will not spin is the lid is not closed or not properly closed.
Stop button
Stop the centrifuge at anytime.
Open button
Open the centrifuge lid only when the centrifuge is not spinning.
Emergency lid-opening
The lid is designed to prevent it from being opened accidently while the
centrifuge is spinning. There may be a situation when the lid needs to be
opened manually (For example if you lose electrical power and samples are
still in the centrifuge).
To open the lid manually, use the lid-opening tool (or a small screw driver).
Insert the tool into the emergency lid-opening hole on the upper right corner
on the left side of the front of the centrifuge.
Cleaning
7
Disinfect using a 10% bleach solution (one part bleach to nine parts water).
Exterior of the centrifuge should be disinfected weekly.
Interior should be disinfected monthly.
Prepare a fresh 10% bleach solution each time. Once mixed with water, bleach
loses effectiveness after 24 hours. The solution should not be stored for future
use.
Cleaning Rotor
There is no need to clean the rotor.
If cleaning the rotor is required please call Ascend Clinical Client Services at
800.800.5655, Option 1 and we will provide a new centrifuge.
Problem Solving
8
Fuse Replacement:
o CAUTION: DISCONNECT THE POWERCORD BEFORE REPLACING
FUSE.
o Follow the procedures below to replace the fuse:
o Unplug the power cord. Turn the on/off switch to “off”.
o Remove the power cord from the power inlet on the back of the
centrifuge.
o Locate the fuse holder. The fuse holder is a part of the power inlet.
o Use a Philip type screwdriver to take the snap-in type holder cover off.
Remove the blown fuse and replace it with the same type and rating fuse
which is 250V 4A
o Put the snap-in fuse holder back.
Ascend Clinical Client Services is available to assist you will all issues regarding
centrifuges.
If you feel that your centrifuge needs work, please call Ascend Clinical and
speak to a customer service representative to determine the nature of the
problem. Simple problems may be resolvable over the telephone. If the
centrifuge needs to e returned to Ascend Clinical a customer service
representative will assist you.
Please do not hesitate to call us at 800.800.5655, Option 1 with any questions