Professional Documents
Culture Documents
net/publication/342901878
CITATION READS
1 1,515
3 authors, including:
Michael W Dong
99 PUBLICATIONS 1,451 CITATIONS
SEE PROFILE
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Michael W Dong on 13 July 2020.
C. Paul, F. Steiner, and M. W. Dong, HPLC Autosamplers: Perspectives, Principles, and Practices, LCGC North Am.
37(8), 514-529, 2019.
PER SPEC TI V E S
IN MO DERN HPLC
Key Components.
An autosampler comprises two major
subunits: a sample injector and a sam-
ple compartment. The sample injec-
Pump
tor’s role is to deliver a precise sample
Pump
aliquot to the column, and typically
Wash Wash consists of a sample metering device
Port Port
Column Column or sampling syringe, a rotary injec-
tor valve (or valve system), a sample
FIGURE 3: Schematic diagrams showing (a) the sample aspiration, and (b) the sample loop, a sampling needle or capillary,
injection steps of the pulled-loop autosampler design. and a needle port or seat (depend-
ing on design). The role of the sample
ducted in (gradient) reversed-phase dient elution (1,3). The exceptions compartment is to provide storage of
chromatography (stability-indicating are for early-eluting peaks in gradient a large number of sample solutions
analysis) where pre-column sample chromatography. Another important in sample containers (vials or micro-
dispersion has little impact due to exception is in size-exclusion chroma- plates) under ambient or thermo-
“peak compression effect” during gra- tography (SEC), using UHPLC columns stated conditions to allow sequential
520 LCGC NORTH AMERICA VOLUME 37 NUMBER 8 AUGUST 2019WWW.CHROMATOGRAPHYONLINE.COM
tion principles, which are pushed-loop, aspiration capillary. During that phase, the injection valve is in the load position.
pulled-loop, and split-loop injection the sample loop is bypassed and After the loop is filled (either partially or
modes. This section also describes the vented. In a second step, the needle overfilled for full-loop operation), the
consequences of each injection princi- moves to a low-pressure needle port. injection valve can shift to its inject posi-
ple in regard to engineering needs and When inside the port, the sampling tion, and the sample aliquot is delivered
benefits or limitations for the user. Addi- syringe pushes the sample aliquot into to the HPLC column.
tional discussions of the different types the sample loop. Finally, the injection Without the needle port, this setup is
of injection approaches can be found valve is switched to the inject position. more robust from an engineering per-
elsewhere (9,15). A flush solvent is used to fill the spective, and has less contribution to
metering syringe as a “carrier” solvent pre-column dispersion and dwell vol-
The Pushed-Loop Design and provides a continuous fluidic path ume. This sampler type has two main
The principle of the pushed-loop for sample withdrawals and delivery. Air disadvantages. First, such design has a
design is perhaps closest to a manual gaps are often used to segment the higher sample consumption as not all
injection process, utilizing a low-pres- sample plug from the carrier solvent. aspirated sample is transferred to the
sure sample needle port built into or One example of the push-loop autos- column. In the full-loop injection mode,
external to the injection valve. This ampler is the PerkinElmer Series 200, the consumption is often a factor of 3
design contains two fluidic sections, as described elsewhere (3). higher than the injected sample. The
illustrated in Figure 2. The first section, . second disadvantage is an increased
called the aspiration capillary, is used The Pulled-Loop Design carryover, which may be reduced with
for the initial sample aspiration and The pulled-loop design shown in Fig- multiple rinsing of the sample loop
storage. While the injection valve is in ure 3 is perhaps a simpler approach, with flush solvents (which negatively
its load position, the sampling needle because it does not require a needle impacts the cycle time).
moves to the sample and aspirates the port. Here, during sample aspiration, A modern example of the pulled-
sample aliquot by moving the plunger the needle dips into the sample liquid, loop design is included in the Waters
of the sampling syringe backward. The and the sampling syringe directly draws Acquity UPLC, which can operate
sample is now temporarily stored in the the sample into the sample loop while in the “partial loop needle overfill”
WWW.CHROMATOGRAPHYONLINE.COM AUGUST 2019 LCGC NORTH AMERICA VOLUME 37 NUMBER 8 525
(PLNO) mode typically using two wash solvents (a weak and pulled-loop injector in this discussion though the description
a strong solvent) and air-gap segmentations of the sample is relevant for all injector types.
plug to maintain precision and carryover performance (16). .
Vial-to-Needle Autosamplers
The Split-Loop Design Vial-to-needle autosamplers (Figure 5a) require a robotic grip-
The most popular design in modern auto-samplers is per that captures the vial from the tray, and moves it below the
the split-loop design (often called integrated-loop, flow- injection needle, as exemplified in the Agilent 1100 Standard
through needle, or needle-in-loop design). The split-loop Autosampler or Agilent 1260 Infinity Standard Autosampler
autosamplers operate with the sampling needle as part of (but not the Agilent High Performance or Micro Well-Plate
the sample loop. It has two major instrumental setups— Autosamplers). For this (mostly split-loop) type of autosampler,
with the metering device as part of the high-pressure flow the vial is placed on the needle seat, and is removed prior to
path (Figure 4a) or placed outside of the high-pressure flow the sampling needle moving down into the high-pressure nee-
path (Figure 4b). dle seal (Figure 5a). With this design, the needle only has to
In the split-loop design, the injection needle is directly con- perform short distance movement in z-direction. This design
nected with the sample loop with a high-pressure needle seal. eases the mechanical stress on the high-pressure needle seal
During sample aspiration, the needle moves to the vial, and and the whole sample loop, and gives more flexibility for loop
splits the loop into two portions. The sample aspiration can be volumes and loop materials. At the same time, the gripper can
done in two ways. The first option is to use a metering device shake the vial to ensure sample solution homogeneity. Never-
in the mobile phase flow-path with the benefit of more pre- theless, the big disadvantages are the longer injection cycle
cise injection and the constant flushing of the device, so that times and incompatibility with microplates.
no air bubbles can accumulate (Figure 4a, such as in Thermo
Scientific Vanquish and Agilent Infinity II 1290). Alternately, the Needle-to-Vial Autosamplers with Static Sample Tray
sampling syringe can also be placed at a dead end with- This approach is a variant for microplate-compatible autos-
out being flushed throughout the run (Figure 4b, such as amplers, where the sample containers in trays are stationary
in Thermo Scientific UltiMate 3000 and Waters I-Class FTN).
However, special care must be taken that no air accumulates
in the syringe, for instance, by using degassed liquids. The
design has the advantage of reduced dwell volume for high-
throughput screening (HTS) applications.
Comparative Performance
of Different Injector Designs
Table II summarizes of the comparative performance of autos-
amplers with different injector designs, indicating a strong
superiority for the split-loop design, with better precision and
carryover performance.
6
WVL:254 nm The injection cycle time is shorter than
with vial-to-needle types, but still not
150
120 min
1
400
2
3
the shortest possible. Due to the rapid
5 7
200
4
8 9
10
movement of the sampling needle,
there can be more mechanical stress on
-100
min the drive mechanism. Since the vials or
0.00 1.00 2.00 3.00 4.00 5.00 6.00
microplates are static, the only way to
FIGURE 6: Effect of SmartInject technology on the elimination of pressure dips during homogenize the sample solution is to
the injection cycle and the resultant improvements of retention time precision show- program the needle movement in the
ing (a) without, and (b) with, for an in average of six-fold improvement of precision for vial by sample aliquot withdrawal and
retention time. release sequences.
1,500
tion, the elimination of pressure shock
2,000
improves column lifetime and lessens
2,500 potential interference with the analyte
Infliximab min
mAU detection and quantitation of peaks
0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0 22.5 25.0 27.5 30.0
near the solvent front.
Figure 7 shows another benefit of
FIGURE 8: Simultaneous parallel characterization of a monoclonal antibody sample SmartInject technology during a size
(Infliximab) by (a) strong cation exchange (SC-UV (with pH gradient) for charge variants, exclusion chromatography analysis
and (b) reversed-phase LC–MS (with acetonitrile gradient) for intact protein analysis,
using a multi-angle light scattering
using a Thermo Scientific dual split sampler with two separate sample flow paths.
(MALS) detector for the aggregate
most modern autosamplers are Pel- illustrating the modern trends of high- analysis of a sample of bovine serum
tier devices, but the way to perform end HPLC autosamplers. albumin (BSA) (11). The use of SmartIn-
sample cooling may differ. The typical ject eliminates a substantial interfering
lower temperature limit is around 4 oC, Case Study of Design and Perfo- artifact peak caused by flow disruption
as lower temperatures may illicit issues mance of an UHPLC Autosampler associated with the injector valve rota-
associated with condensed moisture (Thermo Scientific Vanquish) tion of a standard injection process.
and ice formation. One operational There is a diversity of design types and A note worth mentioning is that the
principle is to cool the sample con- technical implementations by different metering device used for the split-
tainer in an air stream, which requires manufacturers, making it difficult to loop injection mode contributes to
a well-enclosed compartment for best comment on their impacts on perfor- the system dwell volume. In general,
temperature control. The other is a mance characteristics. To illustrate the that device has a flexible volume, and
contact-based cooling, using metal design rationales and innovative impli- can thus be adjusted to change the
blocks (aluminum) as a temperature cation, we will describe the technical overall dwell volume of a system. In
sink near the samples and allowing for details of one modern UHPLC autosam- this autosampler, this adjustment of
a more open design. The closed com- pler as a case study (Thermo Scientific dwell volume can be implemented
partment concept with ventilation has Vanquish) (17–18). in the chromatography data system
advantages regarding the reduced for- Introduced in 2014, the Vanquish (CDS), which can be helpful in method
mation of condensed water or expo- autosampler has a default sample transfer scenarios where differences
sure to the laboratory atmosphere. capacity of four trays hosted in a car- of dwell volume across different
The alternate mode of contact cooling ousel (Figure 5b3), resulting in a total instrument platforms are sources of
has the advantages for faster cooling capacity of 216 for 2-mL vials. In addi- potential issues (10).
and easier insertion of last-minute tion, this autosampler can work with Finally, a unique dual split sampler
vials or microplates. multiple vial sizes or standard microti- is available to enhance laboratory pro-
ter plates, with barcode identification ductivity by providing a second injec-
Market Landscape of different plate formats for added tion flow path (each with its separate
Table III is a summary of the design and security. The trays are open to facili- pump and detector) for performing
specifications of modern UHPLC autos- tate air circulation for efficient sample two assays in parallel to double sam-
amplers from four major manufacturers cooling. The injection mode is a split- ple throughput as illustrated in Figure
(standard configuration of the high- loop design with a metering device in 8. Here, the same monoclonal anti-
end instruments). The information is the high-pressure flow path and a fast body sample is characterized simul-
extracted from a recent article (9), sci- X-Z needle drive, while the carousel taneously by strong cation exchange
entific literature, and updated manufac- performs the movement in Y-direc- chromatography for charge variant
turers’ specifications. Note that all four tion simultaneous to the X-drive. This analysis and reversed-phase chroma-
autosamplers employ the split-loop design enables fast injection cycles tography for intact protein analysis
injection and needle-to-vial approaches, down to 8 seconds with an injection (11). This dual split sampler is capable
528 LCGC NORTH AMERICA VOLUME 37 NUMBER 8 AUGUST 2019WWW.CHROMATOGRAPHYONLINE.COM
of performing two entirely different using carbon-reinforced polymer or eral maintenance comments of HPLC
assays at the same time. ceramic, with lower friction and better autosamplers are included here as a
wear properties (9). reference.
Modern Trends in Autosamplers The reliability of HPLC injectors
In recent years, HPLC autosamplers Autosampler Selection, Opera- has increased significantly in recent
have become more sophisticated, tion, Maintenance, and Trouble- years. The typical wearable item is the
with excellent reliability and perfor- shooting injector rotor seal for a rotary injec-
mance. This is particularly obvious Best practices in HPLC operation, main- tion valve, which should be replaced
for newer UHPLC autosamplers with tenance, and troubleshooting have periodically (annually) or when leaks
higher precision even for small-vol- been described in books (3,22–23), occur. Under ideal conditions, a rotor
ume injections (<1 µL). Two applica- journal articles (24–25), and manual or seal can last 30,000 to 100,000 injec-
tion areas have seen the most recent publications from manufacturers (26– tions, though 10,000 to 20,000 injec-
progress in autosampler technolo- 27). A summary of practical guidelines tions may be more typical for many
gies: high-throughput screening (HTS) for autosampler selection (for specific applications (13,15). This replacement
and automated sample preparation applications and options), daily opera- for HPLC injector rotor seal can be
for complex sample matrices. Recent tion, maintenance, and troubleshoot- accomplished by users with some
examples of new products designed ing is included here as a general refer- practice following the manufacturers’
for HTS include the Agilent Infinity ence. Since these procedures are highly instructions.
II Multisampler, the Thermo Scien- dependent on autosampler type and The sampling needle and the sam-
tific Dual Split Sampler, and the Shi- design, the reader is referred to manu- pling syringe (low-pressure type) are
madzu Nexera UHPLC Series Autos- facturers’ manuals or resources on the also wearable items that require peri-
ampler, which can be equipped with specific model for details. odical replacement to restore sam-
up to three plate chargers capable pling precision. An external sampling
of a total sample capacity of ~17,000 Selection Guidelines for Specific syringe with a Teflon (PTFE) tip is
(19–20). Examples for automated sam- Applications and Optional Equipment easily accessible and replaced by the
ple preparation systems include LEAP Table IV lists some of our general rec- user. The replacement procedure of
PAL3 autosamplers (CTC Analytical) ommendations on autosampler types the sampling needle or capillary can
with customizable sample prepara- for specific applications and optional be more elaborate and might require
tion modules and Shimadzu Clini- equipment. Note that the availability a service specialist on some models.
cal Laboratory Automation Module of software control drivers on the exist- For a split-loop autosampler, the
(CLAM) 2030 for blood analysis using ing CDS for the specific HPLC system high-pressure needle seat seal also
triple-quadruple MS detection (20). being considered for purchase may be requires annual replacement.
Other autosamplers designed for an overriding consideration in most pur- In regulated laboratories, all HPLC
specific applications include Waters chasing decisions. modules, including autosamplers, must
Alliance Dissolution System with a be “calibrated” after annual preventive
modified 2695 separation module Recommended General Practices maintenance to maintain a “qualified”
capable of automated collection for Autosampler Operation instrument status. The performance
and injection dissolution samples in Table V lists general recommendations verification procedures and acceptance
the sample carousels (21) and HTA’s for autosampler operation, including criteria are often similar to those found
HT4000E SPE-LC sample preparation the selection of various autosampler- in operational qualification (4).
workstation for direct injections from related solvents and sample diluents.
solid-phase extraction cartridges. The reader is referred to other pub- Autosampler
Many autosamplers can perform lications for a more detailed recom- Troubleshooting Strategies
additional simple liquid handling steps mendation of general best practices The reader is referred to manufactur-
such as sample dilution, mixing, inter- (3), including operating manuals or ers’ manuals and resources for details
nal standard additions, and pre-col- resources from manufacturers on spe- on troubleshooting specific autosam-
umn derivatization, though these pro- cific autosampler models. pler models. In one case, an existing
cedures are infrequently performed, mobile app is available for a compre-
except for dedicated analyzers such as Autosampler Maintenance hensive troubleshooting guide for a
those for precolumn amino acid analy- The actual maintenance required is particular UHPLC system (28). Note
sis (20). Another area of rapid develop- dependent on the type and design that some seemingly autosampler-
ment is the replacement of traditional on the particular autosampler, and related problems (imprecision of peak
materials in the most problem-prone recommendations from the manufac- areas, no peaks, bad peak shapes) are
key components, such as the rotor seal turers should be followed. Some gen- often caused by other malfunction-
WWW.CHROMATOGRAPHYONLINE.COM AUGUST 2019 LCGC NORTH AMERICA VOLUME 37 NUMBER 8 529
ing modules, columns, or method- dad, C. Poole, P. Schoenmakers and D. (22) P.C. Sadek, Troubleshooting HPLC Sys-
Lloyd, Eds. (Elsevier, Amsterdam, the tems: A Bench Manual (John Wiley &
related parameters (mobile phases,
Netherlands, 2013), Chapter 1. Sons, New York, New York, 2000).
samples, injection volume, diluent,
(3) M. W. Dong, HPLC and UHPLC for Prac- (23) J. Dolan and L.R. Snyder, Troubleshoot-
and so forth). Diagnosing problems
ticing Scientists, (John Wiley, & Sons ing LC Systems (Humana Press, Totowa,
may require a more holistic system Hoboken, New Jersey, 2nd ed., 2019), New Jersey, 1989).
approach (3). Common autosampler- Chapters 2,4,5,7–9, and 11. (24) J. Dolan, “HPLC Troubleshooting,” col-
related problems and solution scenar- (4) S. Ahuja and M.W. Dong, Eds., Handbook umns in LCGC North Am. 1983–2016.
ios, focusing on the performance or of Pharmaceutical Analysis by HPLC (Else- (25) E. Grushka and I. Zamir, Chemical Anal-
other issues such as peak area impre- vier/Academic Press, Amsterdam, the ysis, Volume 98, P. Brown and R. Hart-
Netherlands, 2005), Chapters 11, 12, 15. wick, Eds. ( Wiley Interscience, New
cision, carryover, cycle time, ghost
peaks, leaks, or blockages, are listed (5) C. H. Arnaud, Chem. Eng. News, 94(24), York, 1989), p. 529.
29–35 (2016). (26) U. Neue, HPLC Trou ble shooting
in Table VI as a reference.
(6) D. Guillarme and M. W. Dong (Eds), Guide, Waters Applications Notes,
Trends in Anal. Chem. (Special Issue) 720000181EN, Waters Corp., Milford,
Summary and Conclusions 63, 1–188 (2014). Massachusetts, 2002.
This installment provides an updated
(7) T. Fehrenbach and S. Wiese, The HPLC (27) Thermo Scientific Vanquish UHPLC Sys-
overview of the design of the injector Expert II: Find and Optimize the Ben- tem, Operating Manual, 4820–3601-En,
and sampling approaches of modern efits of Your HPLC/UHPLC, S. Kromidas, Revision 3.0a, 2017, Thermo Fisher Sci-
autosamplers. The split-loop injector Ed. (Wiley-VCH, Weinheim, Germany, entific, Germering, Germany.
design with the “needle-to-vial” sam- 2017), pp. 223–267. (28) ht tps://play.google.com/store/apps/
pling configuration appears to dominate (8) M.W. Dong, LCGC North Am. 35(6), d e t a i l s? i d = c o m . t h e r m o s c i e n t i f i c .
374-381 (2017). troubleshootingguide&hl=de
most high-end autosamplers because of
its superior precision and carryover per- (9) J. De Vos, K. Broeckhoven and S. Eel-
tink, Anal. Chem. 88, 262 (2016).
formance. Technical details and speci-
(10) M.W. Dong, LCGC North Am. 35(11),
fications of several autosamplers have ABOUT THE AUTHORS
818-823 (2017).
been summarized, followed by a brief
(11) T. Zhang, C. Quan, and M.W. Dong, Carsten Paul
discussion of duo-path or dual-injector is a Product Marketing Man-
LCGC North Am. 32(10), 796–808 (2014).
multisamplers for high-throughput ager for UHPLC products at
(12) Model U6k Universal Liquid Chroma-
screening, and others for automated Thermo Fisher Scientific, in
tography Injector Instruction Manual, Germering, Germany.
sample preparation. General recom- Waters, Milford Massachusetts, 1973.
mendations for autosampler selection (13) Operating Instructions Model 7125 Frank Steiner
and operation were described, together Manual Sample Injector, Rheodyne is a Scientific Advisor at
with guidelines on common mainte- LLC (a subsidiary of IDEX Corporation), Thermo Fisher Scientific, in
nance and troubleshooting procedures. Cotati, California, 2320157D, June 2001. Germering, Germany.
(14) J.L. DiCesare, M.W. Dong, and J.R.
Gant. Chromatographia 15, 595 –598
Acknowledgments_
(1982).
The authors thank the following col-
(15) J. W. Dolan, LCGC North Am. 34(5),
leagues for their review of the manu- ABOUT THE AUTHOR
342-329 (2016).
script: Ken Broeckhoven and Jelle De
(16) ACQUITY UPLC I-Class, System Speci- Michael W. Dong
Vos of Vrije University of Brussel, Chris fications, Revision A, Waters Corpora- is a principal of MWD
Pohl of Thermo Fisher Scientific, Davy tion, Milford, Massachusetts, 2011. Consulting, which provides
Guillarme of University of Geneva, (17) Vanq uish Split Samplers VH - A10, training and consulting
Mark Hayward of Stemline Therapeu- VF-A10, VH-A40, VF-A40 Operating services in HPLC and UHPLC,
tics, Wulff Niedner of Bruker Dalton- Manual, 4828.50 01-EN Revision 3.0, method improvements, pharmaceutical
Germering, Germany, February 2018. analysis, and drug quality. He was formerly
ics, He Meng of Pfizer, Peter Ringeling a Senior Scientist at Genentech, Research
of Spark Holland B.V., and Konstantin (18) Thermo Scientific Vanquish–Split Sam- Fellow at Purdue Pharma, and Senior
pler HT/FT, Product Specifications, Staff Scientist at Applied Biosystems/
Shoykhet of Agilent Technologies.
PS71198-EN 09/16M, Germering, Ger- PerkinElmer. He holds a PhD in Analytical
many, 2016. Chemistry from City University of New York.
References He has more than 100 publications and
(19) M. W. Dong, LCGC North Am. 36(4),
(1) L.R. Snyder and J.J. Kirkland, Introduc- 256–265 (2018). a best-selling book in chromatography.
tion to Modern Liquid Chromatography He is an editorial advisory board
(20) M. W. Dong, LCGC North Am. 37(4),
(John Wiley & Sons, Hoboken, New member of LCGC North America and
252–259 (2019).
Jersey, 3rd ed., 2010), Chapter 3. the Chinese American Chromatography
(21) Alliance HPLC Dissolution System, Association. Direct correspondence to:
(2) L.R. Snyder and J. W. Dolan, Liquid
720000251EN, Waters Corporation, Mil- LCGCedit@ubm.com
Chromatography, S. Finali, P. Had-
ford, Massachusetts, 2013.