You are on page 1of 14

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/342901878

HPLC Autosamplers: Perspectives, Principles, and Practices

Article  in  Lc Gc North America · August 2019

CITATION READS

1 1,515

3 authors, including:

Michael W Dong

99 PUBLICATIONS   1,451 CITATIONS   

SEE PROFILE

Some of the authors of this publication are also working on these related projects:

A series of 3 white papers on uhplc View project

All content following this page was uploaded by Michael W Dong on 13 July 2020.

The user has requested enhancement of the downloaded file.


514  LCGC NORTH AMERICA  VOLUME 37 NUMBER 8  AUGUST 2019WWW.CHROMATOGRAPHYONLINE.COM

C. Paul, F. Steiner, and M. W. Dong, HPLC Autosamplers: Perspectives, Principles, and Practices, LCGC North Am.
37(8), 514-529, 2019.

PER SPEC TI V E S
IN MO DERN HPLC

HPLC Autosamplers: Perspectives,


Principles, and Practices
This installment is the second of a series of four white papers on high-performance liquid chromatography (HPLC)
modules focusing on pumps, autosamplers, ultraviolet (UV) detectors, and chromatography data systems, respectively.
This installment provides a technical overview of HPLC autosampler design, historical perspectives, and recent trends.
Recommendations for selection of autosamplers and guidelines for operating and troubleshooting are provided.

Carsten Paul, Frank Steiner, and Michael W. Dong

T he autosampler, an automated sam-


ple injector, introduces a precise ali-
quot of a sample solution from a sample
prises a sample storage compartment
with an injector, consisting of a valve,
a sample dosing or metering device,
larly for small-volume injections, but
requires more sample solution to over-
fill the loop (three times the sample
container to the high performance liquid and a moving sampling needle. loop volume to ensure that the entire
chromatography (HPLC) column under • Injection Valve: A device such as a loop is filled with the sample solution).
high-pressure flow conditions. It is a rotary valve that allows the introduc- The partial-loop mode offers more
sophisticated automation device, capa- tion of a sample solution in the HPLC flexibility than the full-loop mode by
ble of great precision and long-term reli- column under high-pressure and flow its ability to inject different sample vol-
ability. The historical evolution of manual conditions. An injection valve can be umes without a physical change of the
injectors and auto-samplers have been a standalone manual valve, or part of sample loop.
reviewed in many textbooks (1–4) and an autosampler. It typically consists • Sampling syringe or metering
journal articles (5). Their performance, of a needle port, a rotor and stator device: A precise volume dosing unit
pressure ratings, and reliability have combination, and an exchangeable required for sample aspiration. Two
increased dramatically in the last two sample loop. types are commonly used. One is a
decades to accommodate demands from • Pulled-loop, pushed-loop, or split- flow-through metering device, which
the use of narrower columns in ultrahigh- loop design: These are the three com- is a part of the high-pressure flow
pressure liquid chromatography (UHPLC) mon modes of sample introduction path of a split-loop autosampler;
(6–9). In this article, an overview of HPLC designs. A split-loop design includes the other is a low-pressure dead-end
autosamplers is provided by describing the sampling needle as part of the sampling syringe.
their requirements, historical perspec- sample loop, and aspires an exact • Sample container: Typically a 2-mL
tives, modern design and trends, oper- sample aliquot to be introduced. A glass vial containing the sample dis-
ating principles, sampling configurations, pulled-loop design aspires the sample solved in a diluent (final sample solu-
and common operating, maintenance, by vacuum from an external sampling tion). Vials are placed in an autosam-
and troubleshooting procedures. The syringe into the loop. A pushed-loop pler-compatible vial holder, typically
focus here is on analytical-scale applica- design fills a loop with pressure. rectangular vial trays (vial racks), or
tions, which require high-precision analy- • Full-loop or partial-loop mode: Two vial carousels. Other configurations
sis in regulated laboratories. modes of injection are used with are microplates, commonly used in
pulled- or pushed-loop autosamplers. high-throughput screening (HTS)
Glossary of Key Terms In a full-loop (or fixed-loop) mode, the applications.
and Definitions entire sample loop is filled with the • Sampling needle: Typically a 22-gauge
• Autosampler: A device that auto- sample solution. In a partial-loop or blunt-tip needle of a manual sample
mates the introduction of an aliquot of variable-loop mode, only a fraction of syringe connected to the end of a
sample solution to the HPLC column. the loop (10%–50%) is filled. The full- movable stainless steel or polyether
An HPLC autosampler typically com- loop mode is more precise, particu- ether ketone (PEEK) sampling capillary,
516  LCGC NORTH AMERICA  VOLUME 37 NUMBER 8  AUGUST 2019WWW.CHROMATOGRAPHYONLINE.COM

containers, which is commonly ther-


TABLE I: Requirements and desirable characteristics of HPLC autosamplers
mostated. An optional device (plate
Requirements changer) can be used to transport
• P rovides automated and precise injections of sample solutions from sample additional sample trays or microplates,
containers under high-pressure and flow conditions. Typical injection volume
and increase sample capacity.
ranges in quality control applications with a precision of 0.2– 0.5% RSD and
upper pressure limits are shown below. • Needle wash port: A needle port con-
- Conventional HPLC (2–1000 μL, 6000 psi or 400 bar) nected to waste for parking the sam-
- UHPLC (0.5 –100 μL, >15,000 psi or 1000 bar)
pling needle and rinsing the outside
- Micro and nano LC (1–1000 nL, 6000 –15000 psi or 400 –1000 bar)
of the sampling needle with a solvent
• H as a sample flow path consisting of metals and other materials compatible with
common (U)HPLC solvents, and resistant to the maximum system pressure that is supplied by a peristaltic pump
• H as low carryover characteristics of <0.005%, fast injection cycle <1 min and (or by the sampling syringe flushing
long-term reliability through the inside and outside of the
• Controlled by a controller or chromatography data system (CDS) needle). The needle wash is used to
Desirable Characteristics reduce carryover from the previous
•G
 eneral: Dimensions compatible with other HPLC modules (stackable to minimize sample, which adheres to the outside
footprint); easy to operate and maintain (front-panel access to sample trays and
of the needle.
key components, like sample loop, sampling needle, and injection valve)
• Needle seat/port: For pushed-loop
• L abile samples and sample-limited microanalysis compatibility: sample cooling
capability for labile samples and reduced evaporation; ability to precisely inject small autosamplers, an orifice through
sample volumes (<1 μL) from a sample with limited availability in a vial (<200 μL) which the sampling needle can fill
•U
 HPLC compatibility: withstands pressure >15000 psi; precise injections for ~1 μL; the sample loop; for split-loop autos-
fast injection cycle (<20 s); low contributions to extracolumn band broadening
amplers, a counterpart port for the
• H igh-throughput screening (HTS): Fast injection cycle (<10 s); injects from
microplates; available plate changer to increase sample capacity; low
sample needle that must seal to the
contributions to system dispersion and dwell volume; low carryover; maximum system pressure level. This
multisamplers with dual-injectors or flow paths for higher throughput. port is not required for pulled-loop
• Biocompatibility: Flow path from titanium, tantalum, PEEK, Tefzel, and other salt-and autosamplers.
corrosion-resistant materials for IEX, HIC, IC, SEC and applications under high-pH or
corrosive salt content in mobile phase or column wash solution. • Carryover: The amount of the main
• L iquid handling capabilities: The ability to perform sample dilution and component from the prior injected
pre-column derivatization. sample, which is observed in a subse-
• I njection linearity and accuracy: linearity is important when calibration is quent injection of the sample diluent
performed with different volumes of the same standard, or when injection (blank) without analyte. Carryover per-
volume is varied in sequences. Accuracy may be relevant when methods are
transferred to another instrument platform. formance is highly analyte-dependent,
• Has these features: and is particularly problematic for very
- Wash port: Needle wash capabilities are important to reduce carryover in trace analysis. basic compounds and proteins.
- L eak detector: Improves instrument robustness and prevents faulty analysis by • Injection volume precision: Preci-
detecting leaks in fluidics.
sion reflects the variation (scatter) of
the delivered sample volumes during
TABLE II: Summary of comparative performance of different injector designs repetitive injections as a relative stan-
Feature Pushed-Loop Pulled-Loop Split-Loop (Flow-Through Needle) dard deviation (RSD) of peak area or
Precision High to very high Very High peak height.
Linearity Medium (partial-loop mode) Very High • Injection volume accuracy: Accuracy
Injection Vol- Medium Very High describes the deviation or systematic
ume Flexibility different sample loops required only one sample loop required error of the absolute injection volume
Very Low from the intended volume. The vol-
Carryover Low Sample loop, needle and needle seat con-
ume accuracy requirement is less rel-
stantly flushed with mobile phase from inside
evant in practice since method calibra-
Very High (full-loop mode)
Sample Sample Vol. > 3x Inj. Vol. Very Low tion (standardization with a reference
Consumption Medium (partial-loop mode) Sample Volume = Inj. Volume compound) is performed in quantita-
Sample Vol. > Inj. Vol.
tive HPLC analysis.
Shortest Cycle
Short Very Short • Injection cycle time: The time required
Times
to accomplish a sample injection that
typically used with pushed-loop type. the needle-port, and must be capable includes device movements to bring
Split-loop autosamplers use highly of operating at high pressure. the sampling needle to the sample,
specific sampling needles that match • Sample compartment: An enclosed sample volume aspiration, and needle
compartment for storing sample wash or loop cleaning. The injection
WWW.CHROMATOGRAPHYONLINE.COM AUGUST 2019  LCGC NORTH AMERICA  VOLUME 37 NUMBER 8  517

cycle can be reduced by overlapping


(a) Manual injection valve Sample loop with the ongoing separation cycle time
(if allowed by the instrument and con-
trol software).
Handle
Assembly • Pressure transducer: The pressure
sensor in the autosampler flow path
used in some autosamplers to moni-
Valve Body
tor the injection sequence. The pres-
Needle Port
sure sensor can be used to optimize
the injection routine timing, or warn
from from
(b) Load position Pump (c) Inject position Pump the customer in case of any irregular
behavior.
1 1
2 2 Sample Plug
6 to 6 to Solvents used in the autosampler
3 Column Column
3 include:
5 Needle Port 5
4 4
Sample Plug • Wash solvent: the rinsing liquid(s)
Vent Needle Port
used to clean the sampling needle and
loop to reduce carryover.
Sample Sample
Loop Loop • Carrier solvent: the liquid used to
aspire the sample into the sampling
FIGURE 1: Schematic diagram of (a) the manual injection valve (Rheodyne model 7125);
needle, needle capillary, and sample
(b) schematic of the 7125 injector in Load position; (c) and in Inject position. Figures loop, typically in direct contact with
adapted from reference (3). the sample solvent. The mobile phase
can also be used as a flush or carrier
518  LCGC NORTH AMERICA  VOLUME 37 NUMBER 8  AUGUST 2019WWW.CHROMATOGRAPHYONLINE.COM

scientist, thus allowing the implemen-


TABLE III: Design and specifications of UHPLC autosamplers from four manufacturers
(standard configurations of the high-end instruments) tation of better operating practices or
purchasing decisions.
Thermo
Vendor Agilent Shimadzu Waters Besides obvious requirements such
Scientific
as the ability to withstand high pres-
Vanquish
Instrument 1290 Infinity II Nexera X2 Acquity I-Class sure (6000 psi to 22,000 psi or 400 to
UHPLC
1500 bar) and compatibility to all com-
Pressure Limit 19,000 19,000 22,000 18,000
(psi, bar) 1300 1300 1500 1240 mon HPLC solvents and mobile phase
Injection volume additives, the most important require-
0.1–20 0.1–50 0.01–25 0.1–10
range (μL)1 ment is sampling volume precision
Minimum 14 (7 for Nexera (<0.5% RSD for regulatory analysis) and
<10 8 15
cycle time (s) UHPLC Series) flexibility of sampling volume (2–1000
Accuracy 0.7% at 10 μL ±1.0% at 10 μL ±0.5% at 10 μL ±2.0% at 10 μL μL) (3). Low carryover is important for
132 for G7129B 105 for SIL-30AC critical assays such as bioanalytical or
Sample Capacity
(2-mL vials)
and 432 for or 324 for 216 96 multi-residue assays, and purity analy-
G7167B SIL-30ACMP sis, while a fast injection cycle time is
Precision at important for high-throughput appli-
<1.0 <0.5 <0.25 <1.0
1 μL (%RSD)2
cations. For many biopharmaceutical
Carry over (%) <0.004 <0.0015 <0.004 <0.004 applications using ion-exchange chro-
Majority of data extracted from reference 9 or published manufacturers’ specification sheets. 1Standard matography, hydrophobic interaction
injection volume range, which can be extended by using an optional loop and sampling syringes.
2Different test probes may be used to generate precision data. chromatography (HIC), or size exclu-
sion chromatography (SEC) separation
modes that use mobile phases with
high-salt content and column rinsing
(a) Sample Aspiration (b) Sample Injection with basic eluents (>pH 10), the use
of “biocompatible” materials such as
Aspiration Aspiration titanium, Ni-Cr-Co alloy (MP35N), and
Capillary Capillary
base-resistant polymers in the fluidic
Sampling
Sampling
pathway is highly recommended (3,7,9).
Syringe
Syringe Note that sample volume accu-
Waste Waste racy and linearity are less important
than precision performance in prac-
tice, because most quantitative HPLC
analyses use a calibration process with
Pump Pump a reference standard solution. Never-
theless, these attributes are often veri-
Low pressure Low pressure
Needle Needle fied during HPLC system operational
Port Column Port Column qualification in some pharmaceutical
laboratories (4).
Sample Loop
Other autosampler attributes, such
Sample Loop
as contributions to system dwell vol-
FIGURE 2: Schematic diagrams showing (a) the sample aspiration, and (b) the sample umes or extracolumn dispersion, are
injection steps of the pushed-loop autosampler design. less important in practice and become
more relevant in highly optimized
solvent in some designs (the flow- Requirements and Characteristics UHPLC methods. The contributions
through metering device in split- Table I summarizes the requirements to dwell volumes from auto- samplers
loop samplers). and desirable characteristics of a are usually insignificant in comparison
• Diluent: the solvent for diluting the modern HPLC injector or autosam- to those from HPLC pumps (except for
sample in the final sample solution, pler. What follows below are discus- UHPLC binary pumps with small mix-
which should be weaker than the ini- sions of historical perspectives, injec- ers) (8–10). Pre-column system disper-
tial mobile phase, to avoid potential tor designs, operating principles, and sion is mostly important for isocratic
peak distortions. Note that the sam- sampling configurations. Our goal is analysis, due to the absence of the
ple injection volume is proportional to increase the understanding of the sample zone focusing at the column
to the column void volume (3). HPLC autosampler to the laboratory inlet (3,10). Most critical assays are con-
WWW.CHROMATOGRAPHYONLINE.COM AUGUST 2019  LCGC NORTH AMERICA  VOLUME 37 NUMBER 8  519

where extracolumn band broadening


(a) Sample Aspiration (b) Sample Injection can cause significant miscalculation of
molecular weight averages for organic
polymers (3), or in the determina-
Sample Loop Sample Loop
tion of aggregates levels of monoclo-
Sampling Sampling
nal antibody (mAb) therapeutics (11).
Syringe Syringe

Key Components.
An autosampler comprises two major
subunits: a sample injector and a sam-
ple compartment. The sample injec-
Pump
tor’s role is to deliver a precise sample
Pump
aliquot to the column, and typically
Wash Wash consists of a sample metering device
Port Port
Column Column or sampling syringe, a rotary injec-
tor valve (or valve system), a sample
FIGURE 3: Schematic diagrams showing (a) the sample aspiration, and (b) the sample loop, a sampling needle or capillary,
injection steps of the pulled-loop autosampler design. and a needle port or seat (depend-
ing on design). The role of the sample
ducted in (gradient) reversed-phase dient elution (1,3). The exceptions compartment is to provide storage of
chromatography (stability-indicating are for early-eluting peaks in gradient a large number of sample solutions
analysis) where pre-column sample chromatography. Another important in sample containers (vials or micro-
dispersion has little impact due to exception is in size-exclusion chroma- plates) under ambient or thermo-
“peak compression effect” during gra- tography (SEC), using UHPLC columns stated conditions to allow sequential
520  LCGC NORTH AMERICA  VOLUME 37 NUMBER 8  AUGUST 2019WWW.CHROMATOGRAPHYONLINE.COM

of different types of manual injec-


TABLE IV: General recommendations for selecting autosampler type and options
tors and autosamplers leading to
Specific Applications General Recommendations the modern renditions in use today.
Consider a UHPLC autosampler with sufficient pressure rating
UHPLC applications
and precision performance for small-volume injections.
Early Manual Injectors
Consider autosamplers compatible with microplates with a Before the 1970s, injections were
High-throughput fast sample cycle time (<10 s) or samplers with dual injectors
screening applications or pathways. Consider the purchase of an optional sample tray performed manually, using a micro-
chargers to increase sample capacity. syringe with a sharp needle through
Consider temperature control autosampler with cooling a septum, similar to the injection
Labile samples (pharma-
during the purchase (post-purchase installation may not be process in gas chromatography. The
ceuticals and biofluids)
available for integrated HPLC systems).
injection process was unreliable and
Bioanalytical Choose split-loop autosamplers or pulled/pushed loop types the weakest link in the HPLC process,
applications and those with multiple washing options. Different rotor seals materials
requiring low sample should be explored for better carryover performance (Tefzel, due to frequent leaks from the punc-
carryovers PEEK). tured septa under high pressures.
Biocompatibility and Consider a biocompatible autosampler with salt-resistant
compatibility to high-pH fluidic flow path and rotor seal materials compatible with high Waters U6K Septum-less Injector
mobile phases pH (Tefzel or PEEK).
The first reliable manual injector was
Consider a split-loop sampler requiring lower sample amounts, a septum-less loop-based Model U6K
Samples of limited
or a UHPLC sampler with high precision for low-volume
availability or volumes
injections. Consider using sample vials with micro-inserts or injector introduced by Waters Corpora-
(bioscience research)
vials with conical bottoms. tion in 1973 (12). The U6K uses a set of
Most standard configurations of UHPLC autosamplers have manual switching valves, a sample loop,
Injection volume range injection volume range <20 μL. Consider larger optional sample and a needle port, where samples are
loops and/or sampling syringes. introduced into the loop under no-flow
conditions. The needle port is then
closed with a plug, and the sample
Sample Aspiration Sample Injection
is delivered into the flow stream by
(a) switching another valve, which also
Sample Loop Sample Loop
triggered a recorder chart spike for
signifying the start of the chromato-
Metering
Device
Metering Device gram. While the U6K is cumbersome
in high pressure
flow path
by today’s standard, it was perhaps the
only injector with adequate reliability
until the advent of the rotary injector
Pump Pump
valve a few years later.
Needle Needle
Port Port
Column at system Column
pressure Rheodyne 7125 Manual Sample Injector
(b) Sample Loop Sample Loop
The Model 7125 manual sample injector
Sampling
introduced by the Rheodyne Corpora-
Sampling
Syringe
Syringe
tion (acquired by IDEX in 2002) quickly
became the industry-standard injec-
tor in the late 1970s (13). The compact
injector consists of a two-position, six-
Pump Pump port valve with a rotor, an interchange-
Needle
able sample loop (6 μL to 2 mL), and
Port at system
pressure Column Column
a front-loading needle port (Figure 1a).
The rotor is turned by a handle, and
FIGURE 4: Illustration showing the sample aspiration and injection steps of the split-
loop autosampler design: (a) Using a flow-through high-pressure metering device; and has a high-pressure rotor seal pressed
(b) using a dead-end sampling syringe. between the valve body and a sta-
tor with six connection ports. A man-
injections and analysis. The sample Brief Historical Perspectives ual syringe with a 22-gauge blunt-tip
containers are organized in sample There are numerous historical needle is used to introduce a sample
trays with additional sample capacity accounts of HPLC and its instrumen- aliquot through the needle port con-
provided by optional plate and vial tal development (1–2,5). Here are nected to the sample loop during the
tray changers (or chargers). brief highlights on the development loadposition (Figure 1b). This clever
WWW.CHROMATOGRAPHYONLINE.COM AUGUST 2019  LCGC NORTH AMERICA  VOLUME 37 NUMBER 8  521

ered by switching the valve to the inject


(b) Needle-to-vial configuration position (Figure 1c).
(a) Vial-to-needle configuration
(b1) Static tray with xyz
For best precision, the sample loop
is overfilled with a minimum of three
times the sample loop volume in a
“full-loop” injection mode. A sample
loop of a different volume is required
to change sample volumes in this full-
or fixed-loop mode. For better sam-
pling volume flexibility, a partial-loop
fill mode is used where different sam-
ple aliquots can be injected, as long as
(b2) Static tray with xy theta (b3) Sample carousel with xz drive they do not exceed 70% of the sample
loop volume for this manual injector.
Note that the sample is back-flushed
into the column to minimize band
broadening in the sample loop (1,3).
The 7125 injector was further refined
into the model 7725, which incorpo-
rates a “make-before-break” bypass
feature to prevent the momentary pres-
sure cutoff during the injection process,
which reduces column lifetime (14).
FIGURE 5: Diagrammatic illustrations of technical concepts of sample configuration: (a)
vial-to-needle; (b) needle-to-vial; (b1) static tray with x-y-z drive; (b2) static tray with x-y Early Autosamplers
theta drive; (b3) sample carousel with x-z drive.
The development of a reliable rotary
injector led to the implementation of
TABLE V: General recommendations on autosampler operation many early autosamplers. The first ones
Procedure General Recommendation were simple devices using a rotary turn-
Carrier solvent: mixtures of solvents without buffers and an elution table containing filled sample vials and
strength weaker than starting conditions. nitrogen pressure to overfill the sample
Sample loop purge solvent: A strong and weak wash solvent may be loop of a remotely actuated injector
Selection of solvent
recommended for pulled-loop samplers.
Needle wash solvent: A mixture of water and organic solvent without valve. Other early autosamplers such
buffers (either in wash vial or reservoir for the flush port). as the PerkinElmer Series 200 autos-
Diluent of appropriate solvent strength without buffer with good ampler, introduced in the 1990s and
Selection of solubility of analytes. Mobile phase A or diluent of weaker solvent described elsewhere (3), mimics the
sample diluent strength than the initial mobile phase is preferred for large-volume
manual injection process with a moving
injections to avoid potential peak distortions.
sampling needle, which can access up
Selection 2-mL vials (clear or amber) as recommended by autosampler
manufacturers. Use screw caps with thin or pre-slit septa. Consider higher to 100 sample vials in a tray, and deliver
Selection of
quality vials for trace analysis (extractable and leachable, multiresidue the sample aliquot into a fixed-position
sample vial
assays) for higher recovery and fewer ghost peaks issues (for example, 7125 injector valve that is remotely
Thermo Scientific Chromacol vial with gold-grade glass).
actuated by a contact closure.
Typical injection volume range for quantitative analysis is 5–50 μL for
Selection of
HPLC or 1–20 μL for UHPLC autosamplers. Injection sizes of <3 μL
sample injection Types of Autosamplers
might lead to poorer precision for many HPLC autosamplers. Most
volume
UHPLC autosamplers can inject precisely down to 1 μL. and Operating Principles
Autosampler For autosamplers using low-pressure sampling syringes, purge the While all autosamplers consist of key
purging fluidics daily to remove air bubbles. components serving similar needs
Inject a blank sample at the end of a sample sequence before (injection valve, metering device, sam-
Instrument
system shutdown to reduce any significant adhesion of sample
shutdown pling needle, sample loop, and sample
components on injector surfaces.
compartment), there are significant
differences in the operating principles
design with a built-in needle seat in the be transferred to the sample loop under between various types of autosamplers.
rotor seal coupled directly to the sample atmospheric pressure. While the needle This section explains the functional dif-
loop allows the entire sample aliquot to tip is still in the seat, the sample is deliv- ferences between the three main injec-
524  LCGC NORTH AMERICA  VOLUME 37 NUMBER 8  AUGUST 2019WWW.CHROMATOGRAPHYONLINE.COM

TABLE VI: Autosampler troubleshooting strategy: problems and solution scenarios

Problem Type Category Likely Cause Suggested Action

Air bubble trapped in Purge injector, replace sampling


Gross imprecision
Peak area external sampling syringe syringe with PTFE tips
imprecision
With small-volume
Sampling syringe mismatch Use smaller sampling syringes
injection only
Typically observed Previous injected sample
Post-injection loop and needle rinsing
at low levels adhering to needle or loop
High carryover
For basic analytes Previous injected sample Change to Tefzel or PEEK rotor seal;
or proteins adsorbed on rotor seal change pH of wash solvents
Malfunction of autosampler, Diagnostics of the root cause and
No peak
detectors or wrong methods remedy the module or method conditions
Use binary pumps; Eliminate loop/needle rinsing
Injection Usually caused by dwell volume
Too long for HTS steps or choice of more effective wash solvents, Use
cycle time issues or long rinsing steps
“work-ahead or prepare next sample” injection approach
Sample diluent too strong or
Poor peak shape Split peaks Use weaker diluent or smaller injection volume
mismatch of pH or ionic strength
Trace contaminants of mobile phase
Ghost peaks Used high-purity mobile phase A or sample vials certi-
A (or sample vials/microplates;
from blank fied for trace analysis; rinse mobile phase reservoirs
contaminated mobile phase bottles
Plugged needle or Coring of sample vial or microplate
Blockage Change to thin septa or split septa; filter samples
sampling capillary septa; particles from the sample
Caused by worn injector Replace rotor seal or injection capillary: Replace high-
Leaks rotor seal or sampling needle pressure needle seat seal in the split-loop sampler
Leaking injector fittings Tighten or replace fittings

tion principles, which are pushed-loop, aspiration capillary. During that phase, the injection valve is in the load position.
pulled-loop, and split-loop injection the sample loop is bypassed and After the loop is filled (either partially or
modes. This section also describes the vented. In a second step, the needle overfilled for full-loop operation), the
consequences of each injection princi- moves to a low-pressure needle port. injection valve can shift to its inject posi-
ple in regard to engineering needs and When inside the port, the sampling tion, and the sample aliquot is delivered
benefits or limitations for the user. Addi- syringe pushes the sample aliquot into to the HPLC column.
tional discussions of the different types the sample loop. Finally, the injection Without the needle port, this setup is
of injection approaches can be found valve is switched to the inject position. more robust from an engineering per-
elsewhere (9,15). A flush solvent is used to fill the spective, and has less contribution to
metering syringe as a “carrier” solvent pre-column dispersion and dwell vol-
The Pushed-Loop Design and provides a continuous fluidic path ume. This sampler type has two main
The principle of the pushed-loop for sample withdrawals and delivery. Air disadvantages. First, such design has a
design is perhaps closest to a manual gaps are often used to segment the higher sample consumption as not all
injection process, utilizing a low-pres- sample plug from the carrier solvent. aspirated sample is transferred to the
sure sample needle port built into or One example of the push-loop autos- column. In the full-loop injection mode,
external to the injection valve. This ampler is the PerkinElmer Series 200, the consumption is often a factor of 3
design contains two fluidic sections, as described elsewhere (3). higher than the injected sample. The
illustrated in Figure 2. The first section, . second disadvantage is an increased
called the aspiration capillary, is used The Pulled-Loop Design carryover, which may be reduced with
for the initial sample aspiration and The pulled-loop design shown in Fig- multiple rinsing of the sample loop
storage. While the injection valve is in ure 3 is perhaps a simpler approach, with flush solvents (which negatively
its load position, the sampling needle because it does not require a needle impacts the cycle time).
moves to the sample and aspirates the port. Here, during sample aspiration, A modern example of the pulled-
sample aliquot by moving the plunger the needle dips into the sample liquid, loop design is included in the Waters
of the sampling syringe backward. The and the sampling syringe directly draws Acquity UPLC, which can operate
sample is now temporarily stored in the the sample into the sample loop while in the “partial loop needle overfill”
WWW.CHROMATOGRAPHYONLINE.COM AUGUST 2019  LCGC NORTH AMERICA  VOLUME 37 NUMBER 8  525

(PLNO) mode typically using two wash solvents (a weak and pulled-loop injector in this discussion though the description
a strong solvent) and air-gap segmentations of the sample is relevant for all injector types.
plug to maintain precision and carryover performance (16). .
Vial-to-Needle Autosamplers
The Split-Loop Design Vial-to-needle autosamplers (Figure 5a) require a robotic grip-
The most popular design in modern auto-samplers is per that captures the vial from the tray, and moves it below the
the split-loop design (often called integrated-loop, flow- injection needle, as exemplified in the Agilent 1100 Standard
through needle, or needle-in-loop design). The split-loop Autosampler or Agilent 1260 Infinity Standard Autosampler
autosamplers operate with the sampling needle as part of (but not the Agilent High Performance or Micro Well-Plate
the sample loop. It has two major instrumental setups— Autosamplers). For this (mostly split-loop) type of autosampler,
with the metering device as part of the high-pressure flow the vial is placed on the needle seat, and is removed prior to
path (Figure 4a) or placed outside of the high-pressure flow the sampling needle moving down into the high-pressure nee-
path (Figure 4b). dle seal (Figure 5a). With this design, the needle only has to
In the split-loop design, the injection needle is directly con- perform short distance movement in z-direction. This design
nected with the sample loop with a high-pressure needle seal. eases the mechanical stress on the high-pressure needle seal
During sample aspiration, the needle moves to the vial, and and the whole sample loop, and gives more flexibility for loop
splits the loop into two portions. The sample aspiration can be volumes and loop materials. At the same time, the gripper can
done in two ways. The first option is to use a metering device shake the vial to ensure sample solution homogeneity. Never-
in the mobile phase flow-path with the benefit of more pre- theless, the big disadvantages are the longer injection cycle
cise injection and the constant flushing of the device, so that times and incompatibility with microplates.
no air bubbles can accumulate (Figure 4a, such as in Thermo
Scientific Vanquish and Agilent Infinity II 1290). Alternately, the Needle-to-Vial Autosamplers with Static Sample Tray
sampling syringe can also be placed at a dead end with- This approach is a variant for microplate-compatible autos-
out being flushed throughout the run (Figure 4b, such as amplers, where the sample containers in trays are stationary
in Thermo Scientific UltiMate 3000 and Waters I-Class FTN).
However, special care must be taken that no air accumulates
in the syringe, for instance, by using degassed liquids. The
design has the advantage of reduced dwell volume for high-
throughput screening (HTS) applications.

Comparative Performance
of Different Injector Designs
Table II summarizes of the comparative performance of autos-
amplers with different injector designs, indicating a strong
superiority for the split-loop design, with better precision and
carryover performance.

Autosampler Types Based


on Sampling Configurations
Autosamplers can be categorized by sampling configurations
such as the “vial-to-needle” or “needle-to-vial” approach,
with additional considerations for the ability to handle micro-
plates. These configurations are illustrated in Figure 5, and
further discussed here.
The two major subunits of autosamplers are the sample
compartment to host the sample containers or vessels, and
the injection device, which also includes the flush and needle
wash solvent flow paths. The actual design of each subunit
depends on the technical implementation to bring the sam-
ple aliquot from the designated sample vial or microplate
well to the injector. In this regard, there are two different
implementations. One is the “vial-to-needle,” and the other
is the “needle-to-vial” approach, which can have a static tray
or carousel format. For simplicity, we assume a split-loop or a
526  LCGC NORTH AMERICA  VOLUME 37 NUMBER 8  AUGUST 2019WWW.CHROMATOGRAPHYONLINE.COM

ment in the y-direction), and a screw


drive mechanism to move all these
250
(a) Without SmartInject parts together with the needle in the
200
WVL:254 nm
150 6 x-direction (17–18). The static tray and
120

1 injection valve configuration makes the


400 3
2
4
5 7
best use of space in the sample com-
200 8 9
10 partment, so that the compartment can
be more compact. Unlike the carrousel-
-100
0.00 1.00 2.00 3.00 4.00 5.00
min
6.00 based autosamplers, there is no moving
250 (b) With SmartInject sample tray in the sample compartment.
200

6
WVL:254 nm The injection cycle time is shorter than
with vial-to-needle types, but still not
150

120 min
1
400
2
3
the shortest possible. Due to the rapid
5 7
200
4
8 9
10
movement of the sampling needle,
there can be more mechanical stress on
-100
min the drive mechanism. Since the vials or
0.00 1.00 2.00 3.00 4.00 5.00 6.00
microplates are static, the only way to
FIGURE 6: Effect of SmartInject technology on the elimination of pressure dips during homogenize the sample solution is to
the injection cycle and the resultant improvements of retention time precision show- program the needle movement in the
ing (a) without, and (b) with, for an in average of six-fold improvement of precision for vial by sample aliquot withdrawal and
retention time. release sequences.

0.25 Needle-to-Vial Autosamplers


LS
with Sample Carousels
0.20
This alternative concept employs a carou-
0.15 sel that can host either vial trays or micro-
Relative Scale

plates (Figure 5b3, for example, Thermo


0.10
Scientific UltiMate 3000 WPS, Thermo
0.05 Scientific Vanquish Autosampler, and
Waters Alliance). Since the carousel per-
0.00
forms the Y-movement with the sample
1.0 2.0 3.0 4.0 5.0
containers, the needle drive movement
Volume (mL)
is reduced to X-Z-direction with the max-
06 Blank 10 µL with SmartInject 07 Blank 10 µL without SmartInject 05 BSA 10 µL with SmartInject
08 BSA 10 µL without SmartInject
imum distance from the needle seat to
the closest to carousel axis position. The
design restriction on the sample loop
and the mechanical stress are less than
FIGURE 7: Size exclusion chromatography (SEC) using an autosampler with and with-
out SmartInject Technology in combination with a Wyatt multi-angle light scattering with static tray. Due to the independent
(MALS) detector. The chromatograms illustrate the elimination of a significant artifact Y-movement of the carousel, this type
peak (first peak) of a bovine serum albumin (BSA) sample in the two lower chromato- allows for the shortest cycle times while
graphic traces. The color key to the traces are as: (orange) 06 blank 10 µL with SmartIn- it still moves and shakes the sample con-
ject, (blue) 07 blank 10 µL without SmartInject; (green) 05 BSA 10 µL with SmartInject,
tainers for homogenization to a certain
(black) 08 BSA 10 µL without SmartInject.
extent. The disadvantage to the carou-
after they are placed in the compart- drive mechanisms controlled by stepper sel design is the inefficient usage of the
ment. Hence, the needle performs an motors, while the up and down motion compartment space in the carousel cen-
x-y-z movement (Figure 5b1, for exam- (z-direction) is driven by another motor, ter, thus reducing the sample capacity
ple, PerkinElmer Series 200 and Waters that typically travels with the moving and the efficiency in thermostatting.
Acquity Classic Autosampler) (3); or an sampling needle.
x-z-theta movement (Figure 5b2), for The typical set-up of x-y-theta (figure General Comments on the
example., Agilent 1260 Infinity High 5b2) movement is a screw drive mecha- Sample Trays and Injector Designs
Performance Autosampler) to approach nism to move the rotating needle arm Regarding the general design of the
the containers. In case of the x-y-z in the x-direction, a motor for rotating sample compartment, reliable cool-
movement, the x-y travels of the sam- (theta) the arm that holds the sam- ing is an important functional aspect
pling needle are provided by two screw pling needle to substitute the move- of this subunit. Cooling devices in
WWW.CHROMATOGRAPHYONLINE.COM AUGUST 2019  LCGC NORTH AMERICA  VOLUME 37 NUMBER 8  527

peak area precision of <0.25% RSD for


mAU
Infliximab
injection volume as low as 1 µL.
(a)
80
70 2 K+ This autosampler uses SmartInject
60
50
K+ technology which eliminates the pres-
40 sure drop associated with the injector
30
20
valve flow cut-off during rotation and
n=10
10 improves the retention time precision
n=10 significantly (by a factor of 6 in the
(b)
500
example shown in Figure 6). In addi-
1,000

1,500
tion, the elimination of pressure shock
2,000
improves column lifetime and lessens
2,500 potential interference with the analyte
Infliximab min
mAU detection and quantitation of peaks
0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0 22.5 25.0 27.5 30.0
near the solvent front.
Figure 7 shows another benefit of
FIGURE 8: Simultaneous parallel characterization of a monoclonal antibody sample SmartInject technology during a size
(Infliximab) by (a) strong cation exchange (SC-UV (with pH gradient) for charge variants, exclusion chromatography analysis
and (b) reversed-phase LC–MS (with acetonitrile gradient) for intact protein analysis,
using a multi-angle light scattering
using a Thermo Scientific dual split sampler with two separate sample flow paths.
(MALS) detector for the aggregate
most modern autosamplers are Pel- illustrating the modern trends of high- analysis of a sample of bovine serum
tier devices, but the way to perform end HPLC autosamplers. albumin (BSA) (11). The use of SmartIn-
sample cooling may differ. The typical ject eliminates a substantial interfering
lower temperature limit is around 4 oC, Case Study of Design and Perfo- artifact peak caused by flow disruption
as lower temperatures may illicit issues mance of an UHPLC Autosampler associated with the injector valve rota-
associated with condensed moisture (Thermo Scientific Vanquish) tion of a standard injection process.
and ice formation. One operational There is a diversity of design types and A note worth mentioning is that the
principle is to cool the sample con- technical implementations by different metering device used for the split-
tainer in an air stream, which requires manufacturers, making it difficult to loop injection mode contributes to
a well-enclosed compartment for best comment on their impacts on perfor- the system dwell volume. In general,
temperature control. The other is a mance characteristics. To illustrate the that device has a flexible volume, and
contact-based cooling, using metal design rationales and innovative impli- can thus be adjusted to change the
blocks (aluminum) as a temperature cation, we will describe the technical overall dwell volume of a system. In
sink near the samples and allowing for details of one modern UHPLC autosam- this autosampler, this adjustment of
a more open design. The closed com- pler as a case study (Thermo Scientific dwell volume can be implemented
partment concept with ventilation has Vanquish) (17–18). in the chromatography data system
advantages regarding the reduced for- Introduced in 2014, the Vanquish (CDS), which can be helpful in method
mation of condensed water or expo- autosampler has a default sample transfer scenarios where differences
sure to the laboratory atmosphere. capacity of four trays hosted in a car- of dwell volume across different
The alternate mode of contact cooling ousel (Figure 5b3), resulting in a total instrument platforms are sources of
has the advantages for faster cooling capacity of 216 for 2-mL vials. In addi- potential issues (10).
and easier insertion of last-minute tion, this autosampler can work with Finally, a unique dual split sampler
vials or microplates. multiple vial sizes or standard microti- is available to enhance laboratory pro-
ter plates, with barcode identification ductivity by providing a second injec-
Market Landscape of different plate formats for added tion flow path (each with its separate
Table III is a summary of the design and security. The trays are open to facili- pump and detector) for performing
specifications of modern UHPLC autos- tate air circulation for efficient sample two assays in parallel to double sam-
amplers from four major manufacturers cooling. The injection mode is a split- ple throughput as illustrated in Figure
(standard configuration of the high- loop design with a metering device in 8. Here, the same monoclonal anti-
end instruments). The information is the high-pressure flow path and a fast body sample is characterized simul-
extracted from a recent article (9), sci- X-Z needle drive, while the carousel taneously by strong cation exchange
entific literature, and updated manufac- performs the movement in Y-direc- chromatography for charge variant
turers’ specifications. Note that all four tion simultaneous to the X-drive. This analysis and reversed-phase chroma-
autosamplers employ the split-loop design enables fast injection cycles tography for intact protein analysis
injection and needle-to-vial approaches, down to 8 seconds with an injection (11). This dual split sampler is capable
528  LCGC NORTH AMERICA  VOLUME 37 NUMBER 8  AUGUST 2019WWW.CHROMATOGRAPHYONLINE.COM

of performing two entirely different using carbon-reinforced polymer or eral maintenance comments of HPLC
assays at the same time. ceramic, with lower friction and better autosamplers are included here as a
wear properties (9). reference.
Modern Trends in Autosamplers The reliability of HPLC injectors
In recent years, HPLC autosamplers Autosampler Selection, Opera- has increased significantly in recent
have become more sophisticated, tion, Maintenance, and Trouble- years. The typical wearable item is the
with excellent reliability and perfor- shooting injector rotor seal for a rotary injec-
mance. This is particularly obvious Best practices in HPLC operation, main- tion valve, which should be replaced
for newer UHPLC autosamplers with tenance, and troubleshooting have periodically (annually) or when leaks
higher precision even for small-vol- been described in books (3,22–23), occur. Under ideal conditions, a rotor
ume injections (<1 µL). Two applica- journal articles (24–25), and manual or seal can last 30,000 to 100,000 injec-
tion areas have seen the most recent publications from manufacturers (26– tions, though 10,000 to 20,000 injec-
progress in autosampler technolo- 27). A summary of practical guidelines tions may be more typical for many
gies: high-throughput screening (HTS) for autosampler selection (for specific applications (13,15). This replacement
and automated sample preparation applications and options), daily opera- for HPLC injector rotor seal can be
for complex sample matrices. Recent tion, maintenance, and troubleshoot- accomplished by users with some
examples of new products designed ing is included here as a general refer- practice following the manufacturers’
for HTS include the Agilent Infinity ence. Since these procedures are highly instructions.
II Multisampler, the Thermo Scien- dependent on autosampler type and The sampling needle and the sam-
tific Dual Split Sampler, and the Shi- design, the reader is referred to manu- pling syringe (low-pressure type) are
madzu Nexera UHPLC Series Autos- facturers’ manuals or resources on the also wearable items that require peri-
ampler, which can be equipped with specific model for details. odical replacement to restore sam-
up to three plate chargers capable pling precision. An external sampling
of a total sample capacity of ~17,000 Selection Guidelines for Specific syringe with a Teflon (PTFE) tip is
(19–20). Examples for automated sam- Applications and Optional Equipment easily accessible and replaced by the
ple preparation systems include LEAP Table IV lists some of our general rec- user. The replacement procedure of
PAL3 autosamplers (CTC Analytical) ommendations on autosampler types the sampling needle or capillary can
with customizable sample prepara- for specific applications and optional be more elaborate and might require
tion modules and Shimadzu Clini- equipment. Note that the availability a service specialist on some models.
cal Laboratory Automation Module of software control drivers on the exist- For a split-loop autosampler, the
(CLAM) 2030 for blood analysis using ing CDS for the specific HPLC system high-pressure needle seat seal also
triple-quadruple MS detection (20). being considered for purchase may be requires annual replacement.
Other autosamplers designed for an overriding consideration in most pur- In regulated laboratories, all HPLC
specific applications include Waters chasing decisions. modules, including autosamplers, must
Alliance Dissolution System with a be “calibrated” after annual preventive
modified 2695 separation module Recommended General Practices maintenance to maintain a “qualified”
capable of automated collection for Autosampler Operation instrument status. The performance
and injection dissolution samples in Table V lists general recommendations verification procedures and acceptance
the sample carousels (21) and HTA’s for autosampler operation, including criteria are often similar to those found
HT4000E SPE-LC sample preparation the selection of various autosampler- in operational qualification (4).
workstation for direct injections from related solvents and sample diluents.
solid-phase extraction cartridges. The reader is referred to other pub- Autosampler
Many autosamplers can perform lications for a more detailed recom- Troubleshooting Strategies
additional simple liquid handling steps mendation of general best practices The reader is referred to manufactur-
such as sample dilution, mixing, inter- (3), including operating manuals or ers’ manuals and resources for details
nal standard additions, and pre-col- resources from manufacturers on spe- on troubleshooting specific autosam-
umn derivatization, though these pro- cific autosampler models. pler models. In one case, an existing
cedures are infrequently performed, mobile app is available for a compre-
except for dedicated analyzers such as Autosampler Maintenance hensive troubleshooting guide for a
those for precolumn amino acid analy- The actual maintenance required is particular UHPLC system (28). Note
sis (20). Another area of rapid develop- dependent on the type and design that some seemingly autosampler-
ment is the replacement of traditional on the particular autosampler, and related problems (imprecision of peak
materials in the most problem-prone recommendations from the manufac- areas, no peaks, bad peak shapes) are
key components, such as the rotor seal turers should be followed. Some gen- often caused by other malfunction-
WWW.CHROMATOGRAPHYONLINE.COM AUGUST 2019  LCGC NORTH AMERICA  VOLUME 37 NUMBER 8  529

ing modules, columns, or method- dad, C. Poole, P. Schoenmakers and D. (22) P.C. Sadek, Troubleshooting HPLC Sys-
Lloyd, Eds. (Elsevier, Amsterdam, the tems: A Bench Manual (John Wiley &
related parameters (mobile phases,
Netherlands, 2013), Chapter 1. Sons, New York, New York, 2000).
samples, injection volume, diluent,
(3) M. W. Dong, HPLC and UHPLC for Prac- (23) J. Dolan and L.R. Snyder, Troubleshoot-
and so forth). Diagnosing problems
ticing Scientists, (John Wiley, & Sons ing LC Systems (Humana Press, Totowa,
may require a more holistic system Hoboken, New Jersey, 2nd ed., 2019), New Jersey, 1989).
approach (3). Common autosampler- Chapters 2,4,5,7–9, and 11. (24) J. Dolan, “HPLC Troubleshooting,” col-
related problems and solution scenar- (4) S. Ahuja and M.W. Dong, Eds., Handbook umns in LCGC North Am. 1983–2016.
ios, focusing on the performance or of Pharmaceutical Analysis by HPLC (Else- (25) E. Grushka and I. Zamir, Chemical Anal-
other issues such as peak area impre- vier/Academic Press, Amsterdam, the ysis, Volume 98, P. Brown and R. Hart-
Netherlands, 2005), Chapters 11, 12, 15. wick, Eds. ( Wiley Interscience, New
cision, carryover, cycle time, ghost
peaks, leaks, or blockages, are listed (5) C. H. Arnaud, Chem. Eng. News, 94(24), York, 1989), p. 529.
29–35 (2016). (26) U. Neue, HPLC Trou ble shooting
in Table VI as a reference.
(6) D. Guillarme and M. W. Dong (Eds), Guide, Waters Applications Notes,
Trends in Anal. Chem. (Special Issue) 720000181EN, Waters Corp., Milford,
Summary and Conclusions 63, 1–188 (2014). Massachusetts, 2002.
This installment provides an updated
(7) T. Fehrenbach and S. Wiese, The HPLC (27) Thermo Scientific Vanquish UHPLC Sys-
overview of the design of the injector Expert II: Find and Optimize the Ben- tem, Operating Manual, 4820–3601-En,
and sampling approaches of modern efits of Your HPLC/UHPLC, S. Kromidas, Revision 3.0a, 2017, Thermo Fisher Sci-
autosamplers. The split-loop injector Ed. (Wiley-VCH, Weinheim, Germany, entific, Germering, Germany.
design with the “needle-to-vial” sam- 2017), pp. 223–267. (28) ht tps://play.google.com/store/apps/
pling configuration appears to dominate (8) M.W. Dong, LCGC North Am. 35(6), d e t a i l s? i d = c o m . t h e r m o s c i e n t i f i c .
374-381 (2017). troubleshootingguide&hl=de
most high-end autosamplers because of
its superior precision and carryover per- (9) J. De Vos, K. Broeckhoven and S. Eel-
tink, Anal. Chem. 88, 262 (2016).
formance. Technical details and speci-
(10) M.W. Dong, LCGC North Am. 35(11),
fications of several autosamplers have ABOUT THE AUTHORS
818-823 (2017).
been summarized, followed by a brief
(11) T. Zhang, C. Quan, and M.W. Dong, Carsten Paul
discussion of duo-path or dual-injector is a Product Marketing Man-
LCGC North Am. 32(10), 796–808 (2014).
multisamplers for high-throughput ager for UHPLC products at
(12) Model U6k Universal Liquid Chroma-
screening, and others for automated Thermo Fisher Scientific, in
tography Injector Instruction Manual, Germering, Germany.
sample preparation. General recom- Waters, Milford Massachusetts, 1973.
mendations for autosampler selection (13) Operating Instructions Model 7125 Frank Steiner
and operation were described, together Manual Sample Injector, Rheodyne is a Scientific Advisor at
with guidelines on common mainte- LLC (a subsidiary of IDEX Corporation), Thermo Fisher Scientific, in
nance and troubleshooting procedures. Cotati, California, 2320157D, June 2001. Germering, Germany.
(14) J.L. DiCesare, M.W. Dong, and J.R.
Gant. Chromatographia 15, 595 –598
Acknowledgments_
(1982).
The authors thank the following col-
(15) J. W. Dolan, LCGC North Am. 34(5),
leagues for their review of the manu- ABOUT THE AUTHOR
342-329 (2016).
script: Ken Broeckhoven and Jelle De
(16) ACQUITY UPLC I-Class, System Speci- Michael W. Dong
Vos of Vrije University of Brussel, Chris fications, Revision A, Waters Corpora- is a principal of MWD
Pohl of Thermo Fisher Scientific, Davy tion, Milford, Massachusetts, 2011. Consulting, which provides
Guillarme of University of Geneva, (17) Vanq uish Split Samplers VH - A10, training and consulting
Mark Hayward of Stemline Therapeu- VF-A10, VH-A40, VF-A40 Operating services in HPLC and UHPLC,
tics, Wulff Niedner of Bruker Dalton- Manual, 4828.50 01-EN Revision 3.0, method improvements, pharmaceutical
Germering, Germany, February 2018. analysis, and drug quality. He was formerly
ics, He Meng of Pfizer, Peter Ringeling a Senior Scientist at Genentech, Research
of Spark Holland B.V., and Konstantin (18) Thermo Scientific Vanquish–Split Sam- Fellow at Purdue Pharma, and Senior
pler HT/FT, Product Specifications, Staff Scientist at Applied Biosystems/
Shoykhet of Agilent Technologies.
PS71198-EN 09/16M, Germering, Ger- PerkinElmer. He holds a PhD in Analytical
many, 2016. Chemistry from City University of New York.
References He has more than 100 publications and
(19) M. W. Dong, LCGC North Am. 36(4),
(1) L.R. Snyder and J.J. Kirkland, Introduc- 256–265 (2018). a best-selling book in chromatography.
tion to Modern Liquid Chromatography He is an editorial advisory board
(20) M. W. Dong, LCGC North Am. 37(4),
(John Wiley & Sons, Hoboken, New member of LCGC North America and
252–259 (2019).
Jersey, 3rd ed., 2010), Chapter 3. the Chinese American Chromatography
(21) Alliance HPLC Dissolution System, Association. Direct correspondence to:
(2) L.R. Snyder and J. W. Dolan, Liquid
720000251EN, Waters Corporation, Mil- LCGCedit@ubm.com
Chromatography, S. Finali, P. Had-
ford, Massachusetts, 2013.

View publication stats

You might also like