You are on page 1of 11

1.

Scope of application
Widely used in microbial cell culture, seed germination, seedling test, plant cultivation,
and insects, small animal husbandry etc.. Accurately simulate different environmental
and climatic conditions.

2.The main technical parameters


Model
BJPX-A450
Parameter

Humidity Range 50-90%RH


Humidity stability ±5%-7%RH
Without Lighting: 4~60℃
Temperature range
With Lighting:10~60℃
Temperature
±2℃(Balanced work)
fluctuations
Cycle range of 0 to 100, of 30 segments, each time setting range: 0 to
Working Hours 99 hours
0 to 59 minutes, if the period is set to 0, for the continuous work
Supply voltage 220V、50Hz
Input power 1780W
Shelf number 4Block

3.The product structure characteristics


● Box with high-quality cold-rolled steel, the surface electrostatic powder coating, the
coating is hard and solid, with strong antirust ability.

● The studio material quality brushed stainless steel, rounded shape, smooth,
smooth, easy to clean.

●Microcomputer temperature controller with RS485 communication using Adaptive


temperature control, automatic judgment compressor on off temperature points,
high precision and high temperature without overshoot

●Forced-circulation air duct to ensure temperature uniformity studio.

● Duct heater, refrigeration equipment, a temperature sensor is mounted.。

4.Working conditions
1. Ambient temperature:±5~35℃

2. Relative humidity:≤80%RH

3. The surrounding without strong vibration and corrosive gases

4. Should avoid direct sunlight or other hot and cold sources

5. High concentrations of dust around, in addition to maintain the level of installation, should
reserve some space between the device and the wall

6. Should be installed in a well-ventilated place

5、Safety Precautions
1。. Transport box against collision, inverted or ≥ 450 diagonal.

2. In order to ensure the safety of the device test, install an external protective grounding,
and press the device requires a power supply.

3.For inflammable, explosive, toxic, the strong corrosive goods test is strictly prohibited.

4. The equipment should ensure that the level of installation, to prevent the transport of the
compressor refrigeration oil outflow caused by parking 24 hours to start running.

5. Non-professionals do not disassemble, repair.

6. The alarm must be excluded, not mandatory boot.

6.Handling Precautions
1. The equipment first start, the controller internal parameter (in addition to the set
temperature, time) Do not modify.

2. Lighting Do not open unless necessary.


3. Work indoor vertical wind circulation and each bracket items should not be placed too
full.
4. Ambient temperature is above 35 ° C or box temperature higher than 50 ° C,
non-running compressor.
5. Should cut off the power supply when the device is not in use, keep dry, clean inside and
outside.

◆ Special Note
1. All products manufactured before the rigorous testing, generally do not correct, such as

1
the use of the environment a little evil, ambient temperature outside the appropriate range,
errors cause the value of the temperature display inside the actual temperature, as beyond
the scope of technical indicators, you can refer totemperature controller operating
instructions to be corrected.

2. Instrument under normal operating conditions, such as opening the door long time, the
door is closed temporarily some of the inside temperature fluctuations, which is a normal
phenomenon.
3. Test Note: The test temperature in the box, the application the 0.1 precision mercury
thermometers and mercury-side on the inside the geometric center position.

7.Temperature Controller Description

Display instruction
(1) Heating light: lighten when it is heating. (2) Cooling light: lighten when it is cooling,
cooling delayed is twinkle. (3) Humidification light: lighten when humidification (4)
Auxiliary output light: turn on when it has auxiliary output (5) Normal light: lighten
when the controller set the constant value or day/night (6) Program: lighten when the
controller set the 24 segment program (7) Day light: lighten when setting the day
parameter (8) Program light: twinkle when setting the segment No. of program or
lighten when display current working segment No. (9) Night light: twinkle when
setting the night parameter or lighten when operating night (10) Loop light: lighten
when switching to loop checking (11) segment No. or display loop or display
parameter code (12) Display operating time or setting parameter or booking time, and
Display End when it stop (13) Display temperature measurement value. (14) Display
humidity measurement value (Note: no humidity show ----) (15) Display temperature
setting value (16) Display humidity setting value (Note: no humidity show ----) (17)

temperature unit: ℃ or ℉. (18) Humidity unit: %RH (19) Alarm light: twinkle when
it stop or unusual situation (20) Water level alarm light: Lighten when keyboard lock
(21) Power light: lighten when it turns on (22) Keyboard lock light: lighten when
keyboard lock (23) Booking time light (24) Illumination or ultraviolet light (25)
2
Defrosting light (26) Fan light (27) Display the level of illumination (light lighten
when illumination) or heating output percent (output lighten when no illumination)
Keyboard instruction
A Power:turn on or turn off。
B Lock/Shift:press this button 2 second, the other button doesn’t work. Then, press
this button 2 second, the other button work. Or this button combines with other
button to operate different function
C Run/Stop:press this button 2 second, combine with LOCK/SHIFT button could enter
into next stage
D Lamp:Turn on or off the Illumination or UV light
E SET:Enter into Parameter setting or press this button 2 second enter into LK or quit
from parameter setting, or combine with LOCK/SHIFT button could Shift with program
and constant value
F Shift:Parameter shift changing or combine with LOCK/SHIFT button could check the
rest of operation loop
G ∨:Revise Parameter value or revise operation time or combine with LOCK/SHIFT
button could turn on or off the PID about the Humidity
H ∧:Revise Parameter value or revise operation time or combine with LOCK/SHIFT
button could turn on or off the PID about the Temperature

3. Wiring Diagram

Note: illumination 1: illumination 1 working; illumination 2: illumination 2 working;

3
illumination 3: illumination 1 and illumination 2 working (illumination1,2,3 all need to
looping in with COM); Auxiliary output: If one of relays broke, you could setting Fu
parameter or F1 parameter, put the broken relay shift to Auxiliary relay.

4. Operation
1. User’s parameter setting: on standard status, press SET button for a long time,
the LCD display will show LK, then change LK to 8, Then, press SET button enter:
Revise
-100-10 HC=Humidity actual value –
SC humidity 0
0% Humidity measurement value
sensor
Revise near zero, in order to revise
error of the measurement of
Null (-19.9 ~
Pb temperature sensor. Pb=Mercury
adjustment 99.9) 0.0℃
thermometer – current
measurement value
Near to Setting temperature,
temperature appear deviation,
Full degree
(-199 ~ Adjust this value PK=1000 *
Pk adjustment 0
999) (mercury thermometer – current
(slope)
measurement value) / current
measurement value
Alarm 1, upper and lower deviation
alarm, When the temperature
Deviation over (temperature setting value
alarm 0.0 ~
AL +AL) ℃, turn off heating. When
setting 6.0℃
80.0℃
value the temperature lower
(temperature setting value

-AL) ℃, turn off cooling. Alarm


Number of 0:operation; Other value;repeat
Cy 0~99 0
replication light twinkling,
operating buzzer
Cy, then turn offopen.
KS 0:Power on, state of rest 1:Power
Power on on, operation from first
operation 0-2 stage(day) ; 2 : Power on, 1
selection operation from interruption of
power supply
KA Selection of 0 : no booking powering-up 1 :
booking 0-1 booking starting-up ; Display this
starting-up value when the stop status

4
2. Humidity and temperature control parameter setting: on standard status, press
SET button over 2 second, the LCD display will show LK, then change LK to 18, Enter
Humidity and temperature control parameter setting. If it doesn’t have humidity
control, from AS to SL parameter don’t display.
Co Name Setting Instruction Original
de range value
Humidifica When setting value lower than AS,
AS 0-99.9 10.0
tion it doesn’t work
Humidity Lower percent, PS lower, got quick
PS 2~99% 10
percent response on humidity
Humidity The value of integral time, IS lower,
IS integral 1~3600S the deviation of humidity 100
time eliminates quicker
Humidity The value of rate time, ds bigger,
dS 0~3600S 25
rate time humidity moderate oversh
Overshoot In order to overshoot suppression,
LS suppressio 0~100% don’t be very small value, it may 100%
n cause humidity can’t increase
Humidity
Control humidification output
tS control 1~99 秒 3 second
cycle
cycle
When the temperature setting
Open
within±1℃, humidity
Dehumidif (0 ~
Sr 4%RH
ication 60) %RH measurement value ≥humidity
deviation setting value:+Sr, and twice
compression engine starting-up
Cancel time ≥Ct, begin dehumidification
Humidity measurement value ≦
SL dehumidifi (0 ~
humidity setting value: +Sr-SL, 1%RH
cation 60) %RH
cancel dehumidification
deviation
Turn off
Temperature measurement
temperatu
value≥temperature setting value:
Cr re (0.0 ~
Refrigerati +Cr℃, and twice compression 0.4℃
60.0)℃
on
engine starting-up time ≥Ct, begin
deviation
Refrigeration
compressi
(60 ~ compression engine protection
on engine 120
Ct 3600)seco delay time, twice starting-up time
protection second
nd ≥Ct
delay
Pu Heating 1.0 ~ Adjust Heating proportional zone, 2.0

5
proportion P lower, response time quicker
60.0℃
al zone
Heating 1 ~ Adjust integral time vslue: Iu lower,
Iu integral 3600secon corrected static error ability 240
time d stronger
Heating Heating differential time: du
du differential 0~3600S bigger,Al has stronger ability to 60
time prevent overheating
Heat Used to suppress temperature
overshoot overshoot, if the value is too small,
Lu 0~100% 100%
suppressio which may cause the temperature
n can’t rise。
Heating
tu control 1~99 秒 Control heating output period 3 second
cycle
3. Factory parameter setting:In standard state,Long press SET key,The left
upper row display of the LCD screen shows LK,change LK=168,press SET to enter。
Code Setting Original
Name Instruction
range value
Program 0:Fixed value control or day and
md control 0-1 night ; 1: 24 segment program 1
selection control
External 0 : None ; 1 : External input
In input 0-2 disconnected valid 2:External input 1
selection closure efficiency
uA External
function 0-1 0:water line; 1:gating 1
selection
Illuminatio
0:level 3 illumination 1:level 5
bG n series 0-1 0
illumination
selection
degree
CF Celsius or 0-1 0:Celsius 1:Fahrenheit 0
Fahrenheit
Temperatu
0:NO decimal 1:With a decimal
dP re decimal 0-1 1
point
selection
Temperatu
re lower -60.0-9 Setting the lower limit of the
rL 0.0
limit 9.9 measuring temperature
setting
Upper -60.0-9 Setting the upper limit of
rH 99.9
limit of 9.9 measurement temperature

6
temperatu
re
Alarm when the measured
Allowable
temperature is higher than this
maximum 37.0-99
AP value, When it's above this 65.0
temperatu .9
temperature for more than 2
re
minutes, auto-stop.
Temperatu When temperature setting 《 =Cb,
re Compressor is often open ,
0 ~
Cb refrigerati Equilibrium mode ; If compressor
0.0℃
on starting 60.0℃ start and stop mode is needed,set
point Cb to smallest;
When temperature setting>Cb,But
Dehumidif
(0 ~ humidity setting <=JS,Compressor is
ication
JS 100) % often open,Equilibrium mode;If 0%RH
starting
RH compressor start and stop mode is
point
needed,set JS to smallest.
0 : When the set temperature is
greater than or equal to bt,Auxiliary
output;
1:When the set temperature is less
than or equal to bt , Auxiliary
output;
2:When the set humidity is greater
than or equal to HS , Auxiliary
output;
3:When the set humidity is less
Auxiliary than or equal to HS , Auxiliary
relay output;
Fu output 0-4 4 : Auxiliary output as automatic 0
function water supply output ; Water
selection shortage output
5 : Measurement temperature
greater than or equal to bt,Auxiliary
output;
6:When the measured humidity is
greater than or equal to HS ,
Auxiliary output;
7:Defrosting output 8:Lighting or
UV output 9:Fan output
10 : Illumination 1 output 11 :
Illumination 2 output 12 :

7
Illumination 3 output
13:Start the second manual output
by shifting the key
Auxiliary
output
HS 0-100 Set Fu parameter as 2,3,6 is working 85RH%
humidity
point
Auxiliary
output -60.0-9 Set Fu parameter as Fu 0,1,5 is
bt 30.0
temperatu 9.9 working
re point
After compressor continuous
operates dt hours , Defrosting
output. Set Ht as 0,And dt is Double
compressor switching time, dt=1,
Defrosting
1-9999 Fixed for compressor 1 working,
dt output 1
小时 dt=2, Fixed for compressor 2
interval
working; For other values, A
compressor works after DT hours,
automatically switch to another
compressor.
Defrosting Ht seconds later,Stop
Defrosting 0-9999 defrosting;Set to zero,The working
Ht 0
time 秒 mode of double compressor can be
done.
0:Defrosting output 1:Lighting or
UV output 2 : Fan output 3 :
Illumination 1 output 4:Illumination
Backup
2 output 5:Illumination 3 output
F1 relay 0-8 8
6:Automatic water supply output
selection
7:Auxiliary output 8:Start the
second manual output by shifting
the key
High and low wind speed:Setting 0,
Output 3 has output;When 100.0,
output 2 (fan) has output;For other
values, when the measured
Fan output temperature is higher than the
Fn 0-100.0 100.0
selection temperature setting value FN or
when the measured temperature is
lower than the temperature set
value, Output 3 has output,
otherwise output 2 (fan) has output

8
-Fn, Output 3 can be used as high
speed fan output, output 2 can be
used as low speed fan output; When
there is only one fan, the fan
outputs 2 and sets the value to
100.0.
4. Further explanation of the operation:
4.0:Set value modification:Press SET key, corresponding time, temperature,
humidity and light setting area to flash, modify the setting time, set temperature, set
humidity and illuminance in turn. The first period of time is set to 0, only the first
section is run, and there is no timing. Run the first paragraph without timing. For
multiple programs, if you run only three segments, set the time of the fourth segment
to zero, And so on; if the current run time to 0, the program has been running in this
segment, and stop periodically.
4.1:Ultraviolet disinfection function:If you want to use ultraviolet disinfection
lamp, set LK to 3, press set key to enter, modify Ut disinfection time, and then set the
external input as gated input; At this time, the lamp will do UV ultraviolet disinfection
output; according to the LAMP key to start disinfection, disinfection Ut seconds,
automatically turn off the disinfection lamp; in the disinfection process, automatically
shut off the disinfection lamp, after the door closed automatically start disinfection.
4.2:Book boot:If you need to power on after a period of time after automatic
boot, please start the reservation boot function. First stop the control machine, then
set the parameter Ka to 1 and exit, then press the SET key to flicker the time lamp,
then set the appointment time and press the SET key. The countdown and end of the
reservation boot are alternately displayed in the time display area; when the
countdown is zero, KS automatically turns on the machine
4.3:Adjust the number of running segments and run time:When the controller
works, the current run time can be modified by pressing the increase key or reducing
the key by more than 2 seconds in the standard state; Hold down the LOCK/SHIFT and
RUN/STOP keys at the same time to enter the next paragraph. This method allows you
to adjust the run time to synchronize with Beijing time.
4.4:Fixed value and 24 segment program keystroke switch:When the LOCK/SHIFT
and SET keys are held down for 2 seconds in the downtime state, it can be switched
between 24 stages of program control and day and night (or fixed value control). Hold
down the LOCK/SHIFT and shift keys and display Loop. indicating that you are viewing
the rest of the work cycle.
4.5:Self-tuning function:If the temperature and humidity control is not ideal, start
temperature and humidity self-tuning when the compressor is open and in balance
mode.
① In the standard state, by pressing LOCK/SHIFT bond and increasing bond for
2 seconds, starting temperature self-tuning, flickering at this time, and then turning on
at normal temperature after self-tuning, a new set of PID temperature control
parameters are obtained, and the temperature control effect is greatly improved. If
you want to cancel the temperature self-tuning, press the LOCK/SHIFT key and add the

9
key for 2 seconds.
② In the standard state, the RH flashes and the RH is illuminated long by
pressing the LOCK/SHIFT key and reducing the key for 2 seconds, and the humidity
control effect is greatly improved by using the new PID humidity control parameters
after the end of the self-tuning.If you want to cancel the humidity self-tuning, long
press the LOCK/SHIFT key and reduce the key for 2 seconds.,
③ Self-tuning timing can only choose temperature self-tuning or humidity
self-tuning, they can not be done at the same time, and temperature self-tuning
timing is not humidifying.
4.6:Double compressor working mode:In the factory parameters, set the H t to
0 t as the time required for the compressor to switch each other. Compressor 1
control output and compressor 2 control output because the current is small, so need
to connect AC contactor to control the two compressors. The two compressors are
connected to the cooling output after a series of AC contactors.
4.7:Rich peripheral functions:Through the RS485 interface, the data can be
connected to U disk controller, micro printer control board, mobile phone APP or PC
software.
4.8:Standby and auxiliary output:If a relay is broken at the scene, the broken
relay can be switched to the backup relay or auxiliary relay by setting the Fu
parameter or F1 parameter.