2010
Description for mixing
HAENSCH standard sugar solution
Optisch- elektronische Seite 1 von 2
Messinstrumente seit 1864 CRM 3
General:
The delivered Sugar Solution are manufactured by company Schmidt + Haensch GmbH &
Co. in two to each other-due plastic sample container.
In the first container is purest Saccharose (Part A), in the second container purest distilled
water (Part B). For this, please see also illustration 01.
Both quantities were manufactured accurately by Schmidt and Haensch in a complex drying
and a cradle procedure. The reason for the separation of these two materials, is to extend the
storage time. We assume the unblended components can be kept with dry storage under
ambient temperature (25 ºC) at least 6 months. After merging of sugar and distilled water the
use time amounts to maximally 14 days (under normally ambient conditions).
To get an indicated appropriate material mixture, please proceed extremely carefully. Very
important is also to close as firmly as possible the bottles after each opening to avoid
evaporation or the admission of water into the crystal sugar. To exchange the cover is
absolutely to be excluded, since this could lead to an irreparable deviation.
Mixture:
SCHMIDT+HAENSCH
Optisch- elektronische Messinstrumente seit 1864
7. Vibrate now the solution container approx. 3 min. well through. Make sure that all
existing water drops are mixed with. Afterwards leave the sample for approx. 5 min.
rest. You repeat this procedure please several times (at least 3 times), until all sugar
crystals dissolved completely. Unfortunately for this procedure could be not given an
exact description, since the dissolving process is strongly temperature and on the
mixing proportion dependent. After last vibrating please leave the mixture rest for
minimum 3 minute, before you measure the sample.
8. Example Values:
Brix Ri Anmerkungen:
5.00 BX ± 0,02 BX 1.34027 ± 0,00004 [T= 20 ºC / ••••589.3 nm]
15.00 BX ± 0,02 BX 1.35568 ± 0,00004 [T= 20 ºC / ••••589.3 nm]
20.00 BX ± 0,02 BX 1.36384 ± 0,00004 [T= 20 ºC / ••••589.3 nm]
25.00 BX ± 0,02 BX 1.37233 ± 0,00004 [T= 20 ºC / ••••589.3 nm]
50.00 BX ± 0,02 BX 1.42009 ± 0,00004 [T= 20 ºC / ••••589.3 nm]
55.00 BX ± 0,02 BX 1.43080 ± 0,00004 [T= 20 ºC / ••••589.3 nm]
9. Picture:
SCHMIDT+HAENSCH