You are on page 1of 1

DOCUMENT NO.

: R/INST/001
PAGE : 01 OF 01
CALIBRATION FORM FOR ISSUE NO. : 01
O2 ANALYSER ISSUE DATE : 01/07/02
REVISION NO. : 02
(PARAMAGNETIC TYPE) REVISION DATE : 13.05.2011

DP : XIT731A1 / 732 / 738


TAG NO. UNIT DP-17 HP-10
HP : 2AT1311A/2AT1351/2AT1321
LOCATION DP:D-1731 / D-1739 / D-1738 OUTLET HP:2R-1311 / 2T-1351 / 2R-1321 OUTLET
PURPOSE 1. OSM 2. SDM 3. TROUBLE
NAME OF INSPECTOR DATE OF INSPECTION
MODEL 1100/2200 MAKE SERVOMAX
TRANSDUCER UNIT TYPE 1131/2210A1 CONTROL UNIT TYPE 1100A/2223A1
MEASUREMENT INDICATION / ALARM /
0-10% VOL LOOP DETAILS
RANGE INTERLOCK
AMBIENT TEMP. o
C ACCURACY (±) 0.05Vol %
ANALYSER READING BEFORE MAINTENANCE 1. % (ANALYSER) 2. % (DCS)
1. INFORMATION TO DCS OPERATOR 2. INTERLOCK BYPASS
SAFETY MEASURES
3. ALARM OFF
REF. WI NO. WI/INST/001
CHECK POINTS
SEQUENCE PARAMETERS BEFORE AFTER REMARKS
ANY CORROSION, BREAKAGE
MOUNTING STATUS
PRESENCE OF ANY NOISES
DIRTYNESS OF FLOW
METER TUBE
DIRTYNESS & WETNESS
SAMPLING SYSTEM
OF FILTER ELEMENT
LEAKAGE FROM NEEDLE
VALVE
PREVIOUS LOOP GAIN
(DISPLAY : 6-9)
PRESENT LOOP GAIN (DISPLAY :
TRANSDUCER CHECK 6-9 AFTER SETTING 6-8 AT 0.0)
CELL TEMP. (DISPLAY : 1-7)
RAW DATA (DISPLAY : 1-4)
REF. GAS ANALYSER ERROR,
SEQUENCE PARAMETER REMARKS
COMPOSITION READING % OF O2
ZERO CHECK % VOL % VOL
BEFORE
SPAN CHECK % VOL % VOL
MAINTENANCE
LIN. CHECK % VOL % VOL
ZERO CHECK % VOL % VOL
CALIBRATION AFTER SPAN CHECK % VOL % VOL
MAINTENANCE LIN. CHECK % VOL % VOL
RESPONSE TIME FOR 90% OF SPAN SEC.
RESTORATION FOR 1. INFORMATION TO DCS OPERATOR 2. INTERLOCK ON
NORMAL WORKING 3. ALARM ON
ANALYSER READING AFTER RESTORATION 1. % (ANALYSER) 2. % (DCS)

RESULT PASSED NOT PASSED


SIGNATURES
LEADER AREA IN-CHARGE DEPTT. IN-CHARGE

DATE DATE DATE