You are on page 1of 18

MR and MR Arthrography 31

3 MR and MR Arthrography
Josef Kramer, Gerhard Laub, Christian Czerny, and Michael P. Recht

CONTENTS 3.1
Introduction
3.1 Introduction 31
3.2 General Considerations 31
3.2.1 Contrast-to-Noise Ratio 32
During the past decade MR imaging of the hip joint
3.2.2 Signal-to-Noise Ratio 32 has gained high value. Nowadays, this method plays
3.2.3 Spatial Resolution 33 an important role in the evaluation of hip disor-
3.2.4 Scan Time 34 ders. Scanning techniques have improved dramati-
3.2.5 Image Contrast 34 cally and provide MR images with excellent qual-
3.3 Pulse Sequences 35
3.3.1 T1-Weighted Spin Echo Sequence 35
ity. However, an optimal MR examination obtains
3.3.2 T2-Weighted Spin Echo Sequence 35 a maximum amount of information in a given time
3.3.3 Proton Density-Weighted Spin Echo Images 36 period. Local practices and other factors dictate cer-
3.3.4 Fast Spin Echo Sequences 37 tain aspects of the examination. In every clinical set-
3.3.5 Gradient Echo Sequences 38 ting there must always be a compromise between a
3.3.6 Echo Planar Imaging 38
3.3.7 Steady State Free Precession Sequences 39
long examination, which may be unnecessarily thor-
3.3.8 Dual Echo in Steady State Sequences 39 ough, and a cursory study, which leaves major ques-
3.3.9 Fat Suppression Techniques 40 tions unanswered. Normally, most institutions use a
3.3.10 Magnetization Transfer Technique 41 routine protocol which provides adequate visualiza-
3.4 MR Arthrography 41 tion of the hip joint and demonstrates the majority
3.4.1 Introduction 41
3.4.2 Techniques for MR Arthrography of the Hip 41
of pathologic changes. Additional protocols may be
3.4.3 Anatomy of the Hip Joint 43 tailored to specific clinical problems. The demon-
3.4.4 Labral Lesions 43 stration of most pathology within the hip requires
3.4.5 Loose Bodies 45 pulse sequences that provide optimal visualization
3.4.6 Cystic Lesions 46 of mobile water-bound protons. Increased water
3.4.7 Cartilage Lesions 46
3.4.8 Conclusions 46
content is seen in joint effusions, and in ligaments,
References 47 tendons and muscles following injury. It is also pres-
ent in inflammatory changes, in soft tissues, and in
bone marrow edema following contusions. However,
some questions can only be answered with sufficient
accuracy by using intraarticular administration of
contrast agents. A high spatial resolution is a further
requirement for imaging of the hip joint, especially
if subtle pathologic changes are suspected.
J. Kramer, MD, PhD
Röntgeninstitut am Schillerpark, Reanerstrasse 6–8, 4020 Linz,
Austria
G. Laub, PhD
Siemens Cardiovascular Center, Peter V. Ueberroth Bldg. Suite
3.2
3371, Los Angeles, CA 90095-7206, USA General Considerations
C. Czerny, MD
General Hospital/University Medical School Vienna,Währinger Many factors contribute to image quality. In addi-
Gürtel 18–20, 1090 Vienna, Austria tion to the hardware employed, these include the
M. P. Recht, MD
The Cleveland Clinic Foundation, Diagnostic Radiology/Mus-
signal to noise ratio (SNR), contrast to noise ratio
culoskeletal Section, 95 Euclid Avenue, Cleveland, OH 44195- (CNR), spatial resolution, pulse sequences, and scan
5145, USA time. The field strengths of the main magnet field
32 J. Kramer et al.

vary from low to high (0.2–1.5 T). More recently,


whole body 3 T magnets have become commercially
available, and are used for routine clinical imag-
ing. The field strength influences directly the SNR.
In a first approximation, the MR signal increases
linearly with the main magnetic field, and, there-
fore, a higher field strength magnet offers a better a
SNR and, consequently, a shorter examination time
may be expected or the spatial resolution can be
increased without the drawback of loss of signal
to noise difference. Because of claustrophobia in a
some patients, there is increasing interest in open
MR units operating at magnetic field strengths of
0.2–0.7 T for use in body imaging. The advantages
of these systems are that they are easier to install
and cheaper to maintain than whole body imaging
systems operating at higher field strengths. Recent
studies employing ultra-high field strengths of 3.0 T
are very promising in producing higher resolution
images (Niitsu et al. 2000; Peterson et al. 1999). In
fact, musculoskeletal imaging is considered to be one
of the major reasons for continued developments of b
3.0 T systems. Besides the main magnetic field, there
Fig. 3.1a,b. T1-weighted spin echo images of both hips (a) and
are many other technical factors which influence the
the right hip (b). Normal bone marrow shows hyperintense
SNR. Most importantly, dedicated coils are essential signal intensity, musculature is low to intermediate signal
for optimal signal. However, simultaneously phased
array coils have demonstrated certain advantages in
hip examinations, as well as offering the possibil-
ity to image both hip joints with reasonable spatial field strength, molecular structure of the tissue,
resolution (Fig. 3.1). proton density, and T1 and T2 relaxation times;
operator-dependent factors: coil design, field of
view, number of excitations, sampling bandwidth,
3.2.1 matrix size, slice thickness, TR, TE, and flip angle).
Contrast-to-Noise Ratio Changes in TR influence the degree of T1 weighting
and hence the signal attenuation. In T2-weighted
The contrast-to-noise ratio describes the system’s images, variations in the TE influence the T2 weight-
ability to differentiate two tissues under consider- ing and consequently the signal attenuation. Proton
ation. The contrast-to-noise ratio (CNR) depends density-weighted images are only slightly attenu-
on the difference in signal strength between both ated by T1 or T2 relaxation and thus have a higher
tissues, and the SNR. Pulse sequence type and SNR than either T1-weighted or T2-weighted images
sequence parameters need to be selected carefully (Heron and Hine 2002).
to obtain optimal contrast, and SNR. In some situ- Voxel size is one of the most important determi-
ations where the intrinsic tissue contrast is inad- nants of the SNR. Larger voxels have a higher SNR
equate, the administration of MR contrast agents than small voxels because they contain more pro-
can be very helpful to provide excellent tissue con- tons to produce signals. The SNR is influenced by
trast. any change in parameter which alters the voxel size.
It can be appreciated, for instance, that doubling the
field of view results in a doubling in the length of
3.2.2 two sides of the voxel and hence a fourfold increase
Signal-to-Noise Ratio in the SNR, whereas doubling the slice thickness
doubles the length of only one side of the voxel and
The signal-to-noise ratio is influenced by many hence only doubles the SNR. Both matrix size and
factors (non-operator-dependent factors: magnetic number of acquisitions affect the SNR. Doubling the
MR and MR Arthrography 33

matrix size in both the phase encoding and frequency signal intensities of the individual tissues and this
encoding directions for a fixed field of view results in results in partial volume averaging. There is a direct
the SNR being reduced by a factor of the square root relationship between the size of the voxel and the
of 2, but the scan time is doubled because there are resolution, with a small voxel size resulting in better
twice as many phase encoding steps to sample. Dou- resolution. Therefore, the voxel size in an imaging
bling the number of acquisitions doubles the scan sequence needs to be adjusted to the anatomic struc-
time but only increases the SNR by a maximum of tures which need to be resolved.
the square root of 2 (about 1.4). In cases where SNR The voxel size is a volume of sampled tissue within
is not sufficient, repeating excitations at any phase the patient, and the relationship between voxel size
encode setting is recommended. The number of rep- and field of view is indicated in the following equa-
etitions is referred to as Nex, and SNR increases with tion:
the square root of Nex.
In the frequency encoding axis the sampling Voxel size = (1)
bandwidth determines the range of frequencies (FOVread × FOVphase × slice thickness) / matrix size / Nphase
sampled. Reducing the sampling bandwidth results
in less high frequency noise being sampled and con- where FOVread and FOVphase refers to the field
sequently, as the signal remains unaltered, there is of view in the read, and phase encode directions,
an increase in the SNR. In general, the SNR scales respectively. The matrix size is usually referred to
inversely proportional to the sampling bandwidth. as the number of frequency encodings, and Nphase
Therefore, from a SNR perspective, a lower band- refers to the number of phase encodings. In most
width is preferable. Other factors, in particular the imaging systems the matrix size is limited to a power
chemical shift effects, will usually set a limit for the of 2, or multiples of 64. The number of phase encod-
bandwidth which is applied. Furthermore, a lower ings is flexible, and can be given any number equal
sampling bandwidth usually increases the mini- or less to the matrix size. Based on Eq. 1 any desired
mum TE that can be employed in the sequence, as resolution can be achieved (Fig. 3.2). For practical
the readout period becomes longer with a smaller reasons, however, limitations exist with respect to
sampling bandwidth. the selection of the parameters which are described
The choice of pulse sequence also influences the in the following:
SNR. In spin echo sequences all of the longitudinal
magnetization is converted into transverse magne- x FOVread: the field of view in the read direction
tization, but in gradient echo pulse sequences only a must be selected to cover the region of interest.
proportion of the longitudinal magnetization is con- Due to the low-pass fi ltering properties in the
verted into transverse magnetization as in this case frequency encoding direction there is no wrap-
the flip angle is less than 90°. As a result of this, the around artifact even if the object is larger than
signal is generally greater in spin echo sequences. the FOV. This is the reason why the read direc-
In summary, SNR can be increased by using a large tion is always selected to view along the longest
field of view, a coarse matrix and a large slice thick- dimension along the body.
ness. Spin echo sequences, and in particular those x FOVphase: the field of view in the phase direction
with proton density weighting, produce the high- usually corresponds to the FOV in the read direc-
est SNR. Additionally, the number of acquisitions tion. A partial FOV can be selected on state-of-
should be as high as possible. the art imaging systems, i.e. the FOV in the phase
direction is a fraction of the FOV in the read
direction. Partial FOV is useful for saving scan
3.2.3 time as fewer phase encode steps are necessary to
Spatial Resolution achieve the same spatial resolution. However, one
must be careful in selecting a sufficiently large
The spatial resolution is the ability to distinguish FOVphase to avoid wrap around artifacts. Unlike
between two points and it improves as voxel size in the frequency encoding direction, there is no
decreases. Separate tissues within the same voxel cutoff of signal intensities outside the dimen-
are not separately visualized on the MR image but sion of the FOV in the phase encode direction.
separate tissues in adjacent voxels are differentiated. If the object is larger than the FOVphase, it will
When a voxel contains more than one type of tissue, be wrapped to the other side, and may obscure
the signal intensity of that voxel is the average of the important information.
34 J. Kramer et al.

Nphase: The number of phase encodes determines


the spatial resolution in the phase encode direction.
On most commercial imaging systems, the number
of phase encodes can be selected independently of
the matrix size. A larger Nphase results in a better
spatial resolution due to a smaller voxel size. How-
ever, the scan time increases as only one phase
encode value is acquired per TR interval. Also, SNR
decreases with the smaller voxel size.

3.2.4
Scan Time

The scan time Ts of a sequence is given by:


a
Ts = TR × Nphase × Nex (2)

where TR denotes the pulse repetition time of the


sequence and matrix size and Nphase denote the
number of frequency and phase encodes, respec-
tively. Nex refers to the number of excitations per
each phase encode setting (Nex is also referred to as
the number of acquisitions). Ideally, the data acqui-
sition should be completed in the shortest possible
time without compromising image quality. Nphase
needs to be selected with respect to the necessary
spatial resolution in the phase encode direction. As
shown before, there are situations where one exci-
tation does not provide sufficient SNR, and mul-
b
tiple repetitions are necessary (Nex > 1). The SNR
Fig. 3.2a,b. The effect of FOV on image quality can be observed. increases with the square root of Nex, but scan time
A FOV of 180 cm in (a) and 500 cm in (b) was used. Large increases accordingly.
FOV (b) causes a decrease of spatial resolution and blurring
of image details

3.2.5
Image Contrast

The matrix size should be selected large enough to All imaging sequences are based on the idea of gen-
obtain sufficient spatial resolution. A larger matrix erating a spin echo which is frequency and phase
size will show more details in the object because encoded to create the spatial information. Two major
of the improved spatial resolution (Fig. 3.3). How- methods are commonly employed to generate the
ever, the SNR in the image will be correspondingly echo following the initial excitation. Each method
smaller as shown before. A compromise needs to has specific characteristics which are important for
be made between spatial resolution and SNR based musculoskeletal imaging. In the so-called spin echo
on sequence characteristics, system performance, technique, a second radio-frequency (RF) pulse is
and clinical needs. The scan time is not primarily used to reverse the magnetization and form an echo
affected by the matrix size, as each frequency read- at a time equal to the interval between the two RF
out is acquired during one TR interval. However, on pulses. In addition to refocusing of the magnetiza-
most imaging systems, the number of phase encodes tion, this second RF pulse also reverses the dephas-
is correlated to the matrix size, i.e. doubling the ing of magnetization which has occurred due to field
matrix size will also double the phase encodes which inhomogeneities (T2* relaxation). The signal inten-
will double the scan time as will be shown later. sity in this pulse sequence decays with a time equal
MR and MR Arthrography 35

gradient. Due to the absence of an 180° refocusing


pulse the dephasing of the transverse magnetiza-
tion due to magnetic field inhomogeneities cannot
be compensated for, and the signal decays with T2*
(T2* relaxation). T2* relaxation is generally shorter
than T2 relaxation time, and can be much shorter
depending on the actual field inhomogeneities. The
longer the echo time (TE), the stronger the signal
attenuation due to T2* will be. A good homogeneity
is required to obtain optimal results. In a gradient
echo sequence the flip angle can be smaller than 90°,
as the echo is created by gradients, not a refocusing
RF pulse. Therefore, much shorter TR times can be
applied without saturation of the spins. This allows
new applications such as 3D imaging which is not
a possible with the spin echo sequences due to exces-
sive scan time because of the long TR times.

3.3
Pulse Sequences

3.3.1
T1-Weighted Spin Echo Sequence

TI-weighted spin echo images (T1-SE) are obtained


with a relatively short TR. The echo time TE is chosen
to be as small as possible to minimize T2 signal
decay, and keep a pure T1 contrast. As a result of the
short TR, structures with a long T1 relaxation time
b become progressively saturated and demonstrate
low signal intensity (e.g. muscles and fluid-contain-
Fig. 3.3a,b. A body array coil was used. Images (a) and (b) ing structures). Fat, however, which has a very short
show the effect of the matrix on image quality. In (a) the
T1, demonstrates high signal intensity. T1-SE in the
matrix was 320u320 and in (b) 103u128. There was no change
of further image parameters. Image quality in (b) is definitely hip is used primarily for evaluation of bone marrow
worse compared to (a) disorders such as avascular necrosis and detecting
fractures or overuse syndromes (Fig. 3.4).
to the T2 relaxation time, and the T2* decay has been
eliminated by the refocussing process. As a result of
the 180° refocusing pulse, the contrast in spin echo 3.3.2
images is related to the T1 and T2 relaxation times of T2-Weighted Spin Echo Sequence
the tissues and to the chosen TR and TE. As a result
of the 90° excitation pulse in the spin echo sequence, T2-weighted spin echo images (T2-SE) are obtained
the TR must be relatively long in the order of the with a relatively long TR to minimize T1 satura-
T1 times of the tissue to make sure that the spins tion effects. The echo time TE is chosen to show
have sufficiently recovered between TR intervals to differences in the T2 relaxation times. This occurs
produce enough magnetization. at echo times of around 60–100 ms. At this TE all
The other method, gradient echo imaging, simply body fluids produce a high signal intensity due to
uses gradient pulses to generate a gradient echo. A its T2 relaxation characteristics. Muscle, however, is
negative dephasing pulse is followed by the positive of relatively low signal intensity and fat is of inter-
readout gradient, and an echo is formed right when mediate signal intensity. The SNR in these images
the dephasing is compensated for by the readout is inferior to the SNR in proton density-weighted
36 J. Kramer et al.

c
Fig. 3.4a–d. T1-weighted SE images. a Bilateral avascular necrosis of the femoral head. The necrotic area is clearly delineated
from adjacent bone. On the right side concomitant bone marrow edema is visible. b Perthes disease of the right hip. In contrast
to the normal fatty marrow of the epiphysis of the left hip on the right side signal alterations due to edema, granulation tissue,
and sclerosis can be observed. c In the weight-bearing zone of the femoral head a fine fissure almost imitating an early change
of necrosis can be seen. d A typical stress fracture at the medial aspect of the left femoral neck is evident

images, but the pathology, which is characterized general, the choice of the TR also depends on the
by an increased fluid content, is accentuated. T2-SE field strength as the T1 relaxation times tend to
are rarely used in evaluation of the hip because of increase with higher field strengths. Similar to the
the advantages of fast (turbo) spin echo T2 weighted T1-weighted sequences, the echo time TE is chosen
images. to be as small as possible to minimize T2 signal
decay. Because the T1 and T2 effects are minimized
the PD-SE images have a high SNR and are useful
3.3.3 for providing anatomical detail. In an effort to keep
Proton Density-Weighted Spin Echo Images the scan time relatively short, the TR is not always
used as long as it should be to reduce T1 saturation
Proton density-weighted spin echo images (PD- effects. The choice of TE is also subject to some con-
SE) are obtained with a long TR and short TE straints. For bandwidth and gradient performance
and provide signal which reflects the density of considerations the minimum TE is usually around
protons within the imaging field. The TR must 10–20 ms. Therefore, some T1 and T2 effects are
be long enough to make sure that all tissues have still noticeable in the image contrast of a PD-SE
sufficiently recovered during the TR interval. In imaging sequence.
MR and MR Arthrography 37

3.3.4 the individual echoes due to T2 decay. As a result fast


Fast Spin Echo Sequences spin echo images are susceptible to blurring and edge
artifacts which is particularly noticeable for very long
One of the time-limiting factors with the spin echo echo trains. In practice, these artifacts can be mini-
sequences is the fact that only one echo is generated mized by using a short echo train length and a long
per TR interval. In a modification called fast SE (FSE), effective TE. Another difference to the conventional
or turbo SE (TSE) several echoes are generated using spin echo sequence is the appearance of fat. Due to the
multiple 180° refocusing pulses. Each echo is indi- many refocusing pulses applied in the fast spin echo
vidually phase encoded, and scan time is reduced technique, fat appears of higher signal intensity, while
proportionally. The number of echoes produced after solid structures are of lower signal intensity than in
one excitation pulse is called echo train length (ETL). conventional spin echo images. FSE sequences are
Using the fast spin echo technique, proton density, T2- routinely used in the evaluation of the hip, often with
weighted, and T1-weighted SE images can be obtained the addition of fat suppression. These sequences are
in a fraction of the time required for conventional excellent at detecting bone marrow edema as well as
spin echo sequences. The downside of the fast spin soft tissue pathology such as masses, infection, or
echo technique is related to the signal variation of muscle injuries (Fig. 3.5).

Fig. 3.5a–c. a T1-weighted SE (spin echo) image; b fast T2-


weighted SE image; c T1-weighted SE after intravenous contrast
administration. Adjacent to the major trochanter a fluid collec-
tion (bright on T2-weighting, rim enhancement after contrast can
be seen due to bursitis) c
38 J. Kramer et al.

3.3.5 their susceptibility to magnetic field inhomogene-


Gradient Echo Sequences ities.
Gradient echo sequences are frequently used to
In a gradient echo sequence the transverse magne- obtain volume, or 3D acquisitions. This technique is
tization is generated by a flip angle which is smaller particularly useful for the detection of subtle altera-
than 90°. Magnetic field radiant pulses are used to tions of joint structures. With volume acquisitions a
dephase and rephase the transverse component of sub-millimeter, isotropic resolution can be obtained.
the magnetization, and generate an echo. Any mag- The small voxels also help to minimize susceptibil-
netic field inhomogeneities will effect the amount of ity-related signal intensity losses which are associ-
signal at the echo. This is known as the T2* effect. ated with gradient echo sequences. An entire volume
Some institutions use gradient echo images to of tissue is imaged and the volume is subdivided into
assess hyaline cartilage, particularly in conjunc- thin sections using a second phase encoding gradi-
tion with fat suppression. Gradient echo images do ent along the slice selection gradient. The scan time
not provide adequate visualization of bone marrow in the volume acquisition is defined as:
edema and they are not recommended for the evalu-
ation of bony pathology (Fig. 3.6). However, they Ts = TR × Nphase1 × Nphase2 × Nex (3)
may be useful if assessment of lesions of the hyaline
cartilage has to be performed. where Nphase1 and Nphase2 refer to the number of
Similar to spin echo sequences the tissue contrast phase encodings in the conventional phase encode
can be adjusted by proper selection of the imaging direction, and in the slice select direction, respec-
parameters. T1-weighted gradient echo images are tively. Nex is usually set to be 1 to keep the scan
obtained by the combination of a large flip angle, time within an acceptable range. Unlike in multi-
short TR and short TE. T2*-weighted images can be slice 2D imaging, there is no inter-slice gap and no
generated by a small flip angle and a TR which is cross-talk between individual slices which enables a
relatively long in order to permit sufficient recov- perfect means of postprocessing. In order to be able
ery of the longitudinal magnetization of the spins. to reconstruct the object in multiple planes without
Gradient echo images have scan times which can be a reduction in image quality it is essential that the
significantly shorter than spin echo images. This is voxels are isotropic, or nearly isotropic.
due to the shorter TR which is associated with the Volume imaging generates large datasets with
smaller excitation angle. The main disadvantage is a requirement for considerable storage capac-
ity. Obtaining hardcopy of all available images is
impractical. Images are typically viewed on image
processing workstations which allow quick and easy
multi-planar reconstructions. The isotropic three-
dimensional resolution is particularly useful for
visualizing very small structures and for reformat-
ting images of structures which do not lie in a single
anatomical plane, such as ligaments or the labrum.

3.3.6
Echo Planar Imaging

Echo planar imaging (EPI) is a fast imaging tech-


nique in which multiple gradient echoes are
acquired for each excitation. In effect, EPI is the
gradient echo equivalent of fast spin echo imaging.
There is only one excitation pulse and all of the
echoes are obtained from this using gradient rever-
sal. Therefore, movement artifacts do not occur with
Fig. 3.6. T1-weighted two-dimensional gradient echo (2D GE) these very fast imaging times. However, due to the
image in axial orientation. Hyaline cartilage appears bright long echo readout most of the signal is lost by T2*
and can be differentiated from adjacent bony structures decay, and, therefore, single-shot EPI techniques are
MR and MR Arthrography 39

not used for imaging of the hip. The T2* sensitivity with SSFP sequences are often lower than those
is reduced considerably when using only a small obtained for SE-based techniques. In addition, true
number of echoes after the excitation pulse. In a T2 weighting cannot be obtained, because the signal
so-called multi-shot approach all of the raw data decay for long echo time GRE and SSFP sequences
are acquired in repeated excitations with different includes the effect of field inhomogeneities. In some
phase encoding values. In the future it may play a applications this sensitivity to field inhomogeneities
role in dynamic studies of the hip joint during move- is an important contrast mechanism, but in imaging
ment and in MR fluoroscopy for the assessment of of the hip joint it mostly degrades the image qual-
the containment of joint structures and for guiding ity.
interventional procedures.

3.3.8
3.3.7 Dual Echo in Steady State Sequences
Steady State Free Precession Sequences
A modification of the basic SSFP sequence pro-
For steady state free precession (SSFP) sequences, duces the so-called dual echo in steady state (DESS)
the pulse sequence is designed so that the phase sequence. In this sequence a dual echo readout is
coherence of the transverse magnetization is main- applied to collect both the FISP and the PSIF signal
tained from repetition to repetition. This is different simultaneously. Both images show the identical
from the gradient echo sequences where the trans- anatomical structures with different contrast. The
verse magnetization is spoiled after each repetition overlay of both images creates an unusual contrast
using spoiler gradient and RF spoiling techniques. in which anatomical structures are displayed with
Thus the transverse magnetization generated by good overall contrast, and fluids are shown with
a given excitation RF pulse may contribute to the very high signal intensity. With the inherently high
signal measured in many succeeding echo periods. SNR the DESS technique can provide for tissues
Because phase coherence is maintained, the com- with long T1 and T2 relaxation times, the DESS
plete magnetization vector is in a steady state from sequence is well suited for applications that benefit
repetition to repetition during the free precession from high contrast between fluid and surrounding
between RF pulses. tissue (Fig. 3.7).
The SSFP pulse-sequence configuration of the two
signal components generated by a train of equally
spaced RF pulses. Images can be formed from the
FID component (fast imaging with steady state
precession, FISP), from the echo component (time-
reversed FISP, i.e. PSIF), or from a combination of
the two components. Because both the FID and echo
signals represent a superposition of many individ-
ual coherent pathways, the equations describing the
signal behavior for these two components are much
more complicated than those for a spin echo, or gra-
dient echo sequence. In general, the signals depend
on both T1 and T2 in a complex and mathematically
inseparable way. Nonetheless, some statements can
be made concerning the general contrast properties
for FISP-type pulse sequences: first, contrast weight-
ing that depends on T1 and T2 occurs at short TR
and high flip angle. Second, proton-density weight-
ing occurs at low flip angle and short TE, and third,
T2* weighting occurs at low flip angle and long TE.
In summary, imaging with SSFP sequences is a
flexible technique for the rapid acquisition of MR
images that can provide a wide variety of image Fig. 3.7. DESS image. Fluid shows high signal intensity and
contrast behaviors. However, the SNRs obtained thus enables differentiation from hyaline cartilage
40 J. Kramer et al.

3.3.9 art imaging systems provide adequate field homo-


Fat Suppression Techniques geneity to get good fat saturation, and, if necessary,
the homogeneity can be further improved by shim-
Three methods of fat suppression are in use (Del- ming techniques. The water component of fat and
faut et al. 1999). These are fat saturation, inversion some fatty acids are not suppressed and can result
recovery technique and opposed phase imaging. The in inadequately fat-suppressed images. Fat suppres-
last-mentioned technique is principally used for the sion may be poor with magnets of low field strength
detection of small amounts of fat in lesions, e.g. because the chemical shift between lipid and water
adrenal gland tumors. It may also be used for the increases with the magnetic field strength. Another
evaluation of bone marrow diseases (Disler et al. disadvantage of the CHESS technique is that the
1997). fatsat pulse requires extra time. This results in fewer
In the selective fat saturation technique an RF slices in a multi-slice imaging sequence (assuming
pulse with the same resonance frequency as fat the TR is kept constant), or requires an increase in
is preceding every slice-selection RF pulse. The the TR in an effort to keep the same number of slices
magnetization from fat is dephased using a spoil- with a corresponding increase in scan time.
ing gradient pulse to remove all signal from lipid The inversion recovery sequence (short T1 inver-
structures. This technique is referred to as CHESS sion recovery – STIR – sequence uses the differences
(chemical selective saturation) as it saturates tissues between the T1 of fat and water. Following a 180°
according to their chemical nature (which is related inversion pulse the magnetization of fat recovers
to the resonance frequency). Consequently, no much faster than that of water. By applying the 90°
signal from fat is visible on MR images. The advan- pulse at the null point of fat (the point at which the
tages of this technique are that only fat is suppressed longitudinal magnetization equals zero), the signal
and all other tissues are not affected, and it may be from fat is completely suppressed. One advantage of
combined with any imaging sequence. It is useful in this technique is that it can be employed on low field
post-contrast scanning, e.g. following intravenous strength magnets and it is not affected by main mag-
contrast enhancement and MR arthrography. The netic field inhomogeneities. The T2 and T1 differ-
fat saturation technique has a number of disadvan- ences are combined, resulting in high tissue contrast
tages, however. Fat saturation may be unreliable due (water appears very bright). Therefore, edema in soft
to inhomogeneities in the static magnetic field. These tissues as well as in bone marrow is detected with
inhomogeneities result in imperfect fat saturation, high sensitivity (Fig. 3.8). However, there are several
and partial saturation of other tissues. State-of-the disadvantages of this technique. SNR is relatively

b
Fig. 3.8.a,b. Turbo inversion recovery sequence. In a normal hip (a) the signal from fatty marrow is suppressed. Minimal joint
fluid (hyperintense signal) is visible. b Patient had suffered a hip trauma. Moderate hyperintense signal alteration in the femoral
head (typical for bone contusion) is visible
MR and MR Arthrography 41

low, which influences the spatial resolution directly. regarding the interaction between water and mac-
Tissues with a similar T1 to fat are also suppressed, romolecules.
for example mucoid tissue, hemorrhage, protein-
aceous fluid, melanin and gadolinium. For this
reason inversion recovery images are not suitable
for MR arthrography or examinations after intrave- 3.4
nous application of contrast material. Second, tis- MR Arthrography
sues with a short T1 and long T1 may have the same
signal on inversion recovery images, and cannot be 3.4.1
distinguished from each other. Spin echo and fast Introduction
spin echo sequences are the only sequences which
are suitable for use with inversion recovery tech- MRI has been shown to be very useful in the diagno-
niques. Fat suppression techniques are employed sis of several joint disorders. The use of MR arthrog-
in the hip for improvement of sensitivity of bone raphy is increasing due to its improved accuracy
marrow edema and the suppression of normal fat in compared with conventional MRI for the diagnosis
post-gadolinium imaging. This may either be in MR of several intra-articular disorders.
arthrography or following intravenous gadolinium.
The fat saturation technique (CHESS) is most effec-
tive for post-gadolinium scanning. Fat suppression 3.4.2
techniques are also used to assist tissue characteriza- Techniques for MR Arthrography of the Hip
tion. This is especially true for the detection of bone
marrow edema and demonstration of the extent of In our institution MR arthrography of the hip is
infiltration by tumor. Either fat saturation or inver- performed as a two-step procedure. Joint injection
sion recovery techniques may be employed for these is usually performed under fluoroscopy followed by
purposes and these sequences are a mainstay of hip MR imaging. The patient is in supine position with
MR imaging. the leg extended and slightly internally rotated. The
puncture site is marked on the skin between the
subcapital and transcervical portions of the femo-
3.3.10 ral neck. This point is lateral to the femoral artery
Magnetization Transfer Technique and below the inguinal ligament. Under sterile con-
ditions, a 20-gauge disposable needle is directed
In tissues the hydrogen being imaged consists of straight onto the femoral neck (Fig. 3.9). A few drops
two exchanging pools: freely mobile water protons of iodinated contrast agent are injected through an
and restricted motion macromolecular protons. extension tube (to avoid radiation exposure to the
Magnetization transfer contrast makes use of the hands) to confirm intra-articular location; follow-
fact that protons bound in macromolecules do not ing this 10–20 ml of the Gd-DTPA solution is injected
participate directly in the production of the MR into the joint.
signal whereas free protons in water do. Although An alternative method of streamlining intra-
the center frequencies of these two groups are the articular injections when performing conventional
same, the restricted protons have a much shorter MR imaging prior to the MR arthrographic portion
T2 and thus a much broader resonance. With mag- of the examination was reported recently (Miller
netization transfer contrast sequences, low-power 2000). By this technique the total MR examination
radiofrequency irradiation is applied off resonance time can be shortened by eliminating a visit to the
to selectively saturate protons with a short T2. Con- fluoroscopy suite in the middle of the MR study. The
sequently, the partially saturated protons emit less lower extremity is held in neutral or slight internal
signal and therefore tissues which contain bound rotation by tapering the feet together. A metal marker
water demonstrate lower signal intensity on magne- is placed over the middle of the femoral neck, and the
tization contrast images (Wolf et al. 1991; Balaban skin is marked. If contrast is needed, the injection
and Ceckler 1992). In this way the contrast gener- can be performed on the MR table. An MR-compat-
ally resembles that of T2-weighted images and con- ible plastic cart with drawers containing arthrogra-
siderable signal is lost in solid tissues but this does phy supplies is used. The feet are taped together as
not occur in fluid or adipose tissue. Magnetization they were in the fluoroscopy suite, and a 22-gauge
contrast, therefore, provides unique information needle is placed straight down until bone is reached.
42 J. Kramer et al.

Cartilage

Acetabular
Labrum

Zona
Orbicularis

Ring of contrast medium Fig. 3.9. A schematic drawing of the


around the femoral neck right hip joint with the arrow pointing
to the injection site (circle). The dark
Transverse Acetabular
area shows the distribution of contrast
Ligament material initially after injection

Low resistance to the injection confirms the needle formed immediately after the intra-articular injec-
position. Although this technique uses fluoroscopy tion of contrast medium, to prevent absorption of
to landmark the joint of interest, the fluoroscopy contrast solution and guarantee the desired cap-
portion of the procedure can be performed at any sular distention, although imaging delays of up to
time prior to the MR examination, thus uncoupling 1–2 h are tolerated (Wagner et al. 2001). Gd-DTPA
the fluoroscopy and MR schedules from each other and iodinated contrast material can be mixed before
and allowing the arthrographic portion of the study MRI without any release of free gadolinium and is
to be performed in the MR suite immediately after safe for confirming the intra-articular placement of
the conventional MR portion. contrast material (Brown et al. 2000).
Because of the spherical nature of the hip, imag- It has been shown that intravenous adminis-
ing in all three planes is necessary to evaluate joint tration of Gd-DTPA also leads to an enhancement
and in particular labrum. T1-weighted imaging is effect of the joint cavity (indirect MR arthrogra-
used to visualize the high signal of the intra-articu- phy). This technique has been proposed as an alter-
lar contrast solution. Fat saturation increases con- native to direct MR arthrography (Vahlensieck et
trast between the intra-articular gadopentetate al. 1995, 1996, 1997; Winalski et al. 1991; Kramer
dimeglumine and the adjacent soft tissues. Fat sup- et al. 1997; Peh et al. 1999). The enhancement effect,
pression is crucial in MR arthrography because fat however, is only mild and often heterogeneous,
and contrast medium have similar signal intensi- although exercise improves both the homogeneity
ties on T1-weighted images. The combination of and amount of enhancement in the joint. However,
MR arthrography and three-dimensional gradient there are several drawbacks of this technique. The
recalled echo-imaging shows a higher sensitivity for main limitation is the relative lack of joint disten-
subtle lesions than MR arthrography with spin echo tion, which is frequently necessary to accurately
sequences (Kramer et al. 1992). diagnose capsular trauma and soft tissue injury.
MR arthrography is performed by injecting a A further drawback of indirect MR arthrography
gadolinium-diethylenetriamine pentaacetic acid is that juxta-articular structures, such as vessels,
(Gd-DTPA) mixture into the joint. A Gd-DTPA mix- and the synovial membranes of bursae and tendon
ture consisting of 0.2 ml of a standard Gd-DTPA sheaths also demonstrate enhancement, which may
solution (469.01 mg/ml, Magnevist, Schering AG, lead to confusion with extravasation of contrast
Germany) mixed with 50 ml of saline (2 mmol/l medium or the presence of abnormal joint recesses.
Gd-DTPA solution) allows excellent delineation of Patients who have undergone MR arthrography
contrast medium and the intra-articular structures reported discomfort to be less than expected (Rob-
or abnormalities (Engel et al. 1990). No side effects bins et al. 2000). Arthrography-related discomfort
have been reported to date that are attributable to was well tolerated and rated less severe than MRI-
intra-articular Gd-DTPA. Imaging should be per- related discomfort.
MR and MR Arthrography 43

3.4.3 occur following minor or major trauma particularly


Anatomy of the Hip Joint hip dislocations. Degenerative tears may also occur.
Early recognition and debridement can result in
The hip is a ball-and-socket joint, which exhib- substantial pain relief and may prevent development
its a wide range of motion in all directions. The of degenerative disease (Altenberg 1977). Conven-
spherical acetabular socket covers the femoral head tional MR imaging is not accurate for the diagnosis
nearly completely except for its inferior medial of labral tears (Czerny et al. 1996, 1999; Leunig et al.
aspect, known as the acetabular notch, where the 1997; Petersilge et al. 1996; Petersilge 1997, 2000,
socket is deficient. The transverse acetabular liga- 2001; Kramer et al. 2002; Steinbach et al. 2002).
ment spans this deficient portion of the acetabulum. There are several causes for this, the most important
The fibrocartilaginous labrum rims the acetabulum being inadequate joint distention that prevents sepa-
and is triangular in cross section. The labrum is ration of the labrum and capsule.
thicker posterosuperiorly and thinner anteroinfe- A classification system for the evaluation of
riorly (Keene and Villar 1994; Tschauner et al. acetabular labral lesions by MR arthrography was
1997). The acetabular labrum consists of fibrocar- established by Czerny et al. (1996). Normal labra
tilaginous tissue with fibrovascular bundles. This (stage 0), are of homogeneous low signal intensity,
fibrocartilage lacks the highly organized structure triangular shaped, and a continuous attachment to
seen within the fibrocartilaginous meniscus of the the lateral margin of the acetabulum without a notch
knee. The labrum is attached directly to the osseous or a sulcus is visible. A recess between the joint cap-
rim of the acetabulum. It blends with the transverse sule and the labrum, a so-called labral recess, is
ligament at the margins of the acetabular notch. observed. Stage 1A lesions are characterized by an
Contrary to the shoulder, the acetabular labrum area of increased signal intensity within the center
increase the depth of the joint rather than increas- of the labrum that does not extend to the margin of
ing its diameter (Engel et al. 1990). Clinical and the labrum, a triangular shape, and a continuous
arthroscopic studies have documented the impor- attachment to the lateral margin of the acetabulum
tance of the acetabular capsular-labral complex as and a labral recessus (Fig. 3.10). Stage 1B is simi-
a biomechanical component of the hip joint (Suzuki lar to stage 1A, but the labrum is thickened and in
et al. 1986; Ueo et al. 1990). The joint capsule inserts most patients no labral recessus can be observed.
onto the acetabular rim. Along the anterior and pos- These findings are similar to the mucoid degenera-
terior joint margins, the capsule inserts directly at tion seen in low-grade meniscal lesions. In stage 2A
the base of the labrum; a small perilabral recess is lesions an extension of contrast material into the
created between the labrum and joint capsule. The labrum without detachment from the acetabulum
iliopsoas bursa, directly anterior to the hip joint, is visible (Fig. 3.11). In this stage the labrum is tri-
communicates with the joint in 10%–15% of normal angular, and has a labral recessus. Stage 2B are the
anatomic specimens and may be involved in patients same as stage 2A except the torn labrum is thick-
with synovitis (Williams and Warwick 1980). ened (Fig. 3.12). Stage 2 findings are consistent with
partial labral tears. Stage 3A labra, complete labral
tears, are detached from the acetabulum but are of
3.4.4 triangular shape, whereas stage 3B labra are thick-
Labral Lesions ened and detached from the acetabulum (Fig. 3.13).
Contrast material enables the visualization of the
Among several possible causes for chronic hip pain labral recessus and the clear delineation of the cap-
acetabular labral tears are a relatively rare entity. sular-labral complex (Hodler et al. 1995). However,
Labral lesions may be observed in patients with devel- labral lesions may be staged incorrectly even with MR
opmental dysplasia of the hips, and in patients with a arthrography if only a small volume of contrast mate-
history of hip trauma (Dameron 1959; Dorell and rial is injected or there is a scarred, shrunken joint
Catterall 1986; Fitzgerald 1995; Klaue et al. capsule caused by previous surgical interventions.
1991; McCarthy and Busconi 1995a,b). In patients A wide spectrum of appearances has been
suffering from hip dysplasia, the increased stress described for the labrum in asymptomatic individu-
on the acetabulum and on the superior labrum as als. However, the majority of these findings have
it assumes more of the weight-bearing function is been based on studies performed without joint dis-
believed to contribute to development of the tears tention. It is unclear whether the described appear-
(Klaue et al. 1991). Posttraumatic labral tears may ances represent variations in normal anatomy or
44 J. Kramer et al.

a b

c d
Fig. 3.10a–d. After administration of contrast material. T1-weighted SE (a,b) and T1-weighted SE sequence with fat suppression
(c,d). a,c Normal labrum (type 0). b,d Labrum is enlarged (type 1B). No tear is visible. The contrast material is delineating the
intraarticular joint structures very well

a b
Fig. 3.11a,b. T1 weighted (SE sequence) MR arthrogram with and without fat suppression. A small tear is visible in the cranio-
lateral portion of the labrum (type 2A)
MR and MR Arthrography 45

has been raised (Walker 1981). It is believed that


any intralabral collection of contrast material should
not be mistaken for a normal sublabral sulcus but,
rather, should be suspected to be a tear or detachment
(Czerny et al. 2002). In contrast Petersilge (2000)
observed such a sulcus commonly on MR arthro-
graphic images, located consistently at the anterosu-
perior aspect of the joint. Because of the consistent
location of this finding and the well-defined margins
of the sulcus as well as the presence of a well-defined
insertion onto the subchondral bone some believe
that this appearance is likely an anatomic variant.
Another potential pitfall with MR arthrography
is a groove separating the acetabular labrum from
acetabular articular cartilage in the region of the
acetabular fossa (Keene and Villar 1994). This
Fig. 3.12. T1-weighted SE image after intraarticular instilla-
groove should not be confused with a labral tear
tion of a Gd-DTPA solution. A patient with severe hip dyspla- because its location is different from that of labral
sia. The labrum shows signal alterations and it is enlarged. A tears, which occur most frequently anterosuperiorly.
horizontal tear is visible (type 2B) No instances of absent labra are reported (Lecouvet
et al. 1996). Sensitivity, specificity, and accuracy for
detection and correct staging of labral lesions with
MR arthrography has been reported to be 91%, 71%,
and 88% respectively, while the sensitivity and accu-
racy of conventional MR imaging was 30% and 36%
(Czerny et al. 1996, 1999). Because labral abnormali-
ties are believed to be a precursor of osteoarthritis,
early interventions is thought to be critical in the
treatment of labral abnormalities. The determination
of the type, extension, and location of labral abnor-
malities as shown with MR arthrography is helpful in
such treatment (Hodler et al. 1995).

3.4.5
Loose Bodies
Fig. 3.13. Fat suppressed T1-weighted MR arthrogram. Patient
suffering from hip dysplasia shows a completely torn labrum Loose bodies in the hip joint are rare but may cause
which is slightly detached from the acetabular rim (sur-
chronic joint pain, locking, and limited range of
rounded by contrast material) (type 3B)
motion and may occur with or without associated
osteochondral defects. In patients with suspected
asymptomatic abnormalities. Controversy remains intra-articular loose bodies, the value of an imag-
regarding the reality of a sublabral sulcus of the ing examination is determined by its capabilities for
anterosuperior aspect of the labrum. The presence of excluding the presence of loose bodies and decreas-
a sublabral sulcus has been raised as the possibility of ing the need for diagnostic arthroscopy (Haims et
a potential pitfall in the diagnosis of acetabular labral al. 1998; Palmer 1998). Frequently, intra-articular
abnormalities with MR imaging (Lecouvet et al. loose bodies cannot be differentiated from sur-
1996). However, the clinical implications of an absent rounding tissues on conventional MR imaging. If
labrum and of a sulcus between the labrum and the there is joint fluid present conventional T2-weighted
acetabulum anterosuperiorly are not known. Such a sequences may help to differentiate loose bodies from
sulcus has been described in histologic specimens of surrounding structures. In some patients specificity
fetal hips, however, although the possibility that its may be decreased, however, because of the difficulty
existence is artifactual because of fixation techniques in differentiating true loose bodies from osteophytes,
46 J. Kramer et al.

synovial folds, and hypertrophic synovitis. Charac- (Fig. 3.14). These defects are mostly found in the
teristically, on MR arthrography loose bodies are sur- anterosuperior part of the acetabulum. The diagnos-
rounded by contrast material, which improves their tic performance of MR arthrography in the detec-
detection and enables an accurate determination of tion of articular cartilage damage in the hip joint is
their number and topography. Arthroscopic tech- inferior to that in the knee joint with sensitivities
niques offering the least traumatic method of remov- and specificities of MR arthrographic detection of
ing loose bodies of foreign bodies from the hip joint cartilage damage slightly below 80%. This may be
are preferred to open arthrotomy, although arthros- due to the relatively thin cartilage in the hip lead-
copy of the hip is still not widely practiced (Die et al. ing to volume averaging of the cartilage layer with
1991). Only the accurate detection of these lesions may adjacent bone, and intraarticular contrast material
allow early therapeutic intervention, however, and due to the spherical shape of the femoral head.
relief of pain, and prevent or delay the development
of osteoarthritis (Czerny et al. 1996, 1999, 2001). In
addition, the diagnostic sensitivity and specificity of 3.4.8
MR arthrography in detecting intra-articular bodies Conclusions
is improved because contrast solution separates loose
bodies from the capsule and completely surrounds MR arthrography has become an important tool
them, whereas osteophytes and synovial projections for the evaluation of a variety of articular disorders
are only partially outlined. providing new insights into the lesions underly-
ing mechanical hip pain. Although not necessary
in all patients, MR arthrography may facilitate the
3.4.6 evaluation of patients with suspected intra-articular
Cystic Lesions pathology in whom conventional MR imaging is not
sufficient for an adequate therapy planning. This
Ganglion cyst have been reported in association technique combines arthrographic advantages, like
with labral tears (Haller et al. 1989; Czerny et joint distention and delineation of intra-articular
al. 1999; Petersilge et al. 1996; Klaue et al. 1991; structures, with the high quality of MR imaging.
Magee and Hinson 2000; Schnarkowski et al. Diagnostic confidence can frequently be improved
1996). These cysts typically start out as extra-articu- by MR arthrography, particularly in the assessment
lar soft tissue collections and may eventually erode of subtle lesions and complex anatomic structures.
into the adjacent acetabulum. The erosion created MR arthrography has been proven especially useful
in the acetabulum may be visible on plain films and in patients with suspected acetabular labral lesions,
may be a clue to labral pathology as the underlying osteochondral defects or loose bodies.
cause of hip pain. Additionally it has been suggested
that when perilabral cysts are identified a search
for an underlying labral tear should be undertaken.
These cysts appear to have an increased frequency
in those patients with associated developmental dys-
plasia (McCarthy and Busconi 1995b).

3.4.7
Cartilage Lesions

On MR images and at histologic analysis hyaline


cartilage of the hip joint has shown a thickness of
1–2 mm (Hodler et al. 1992). In contrast, articular
cartilage of the knee may be up to 7 mm in diameter,
with the highest values in the patella (Ahn et al.
1998). Cartilage lesions of the hip are not uncom-
Fig. 3.14. Fat suppressed T1-weighted image after intra-articu-
mon in young and middle-aged patients who are lar contrast application. At the lateral aspect of the femoral
suspected of having femoroacetabular impingement head a severe cartilage thinning and a small osteophyte is vis-
and/or labral abnormalities (Schmid et al. 2003) ible. The labrum appears torn craniolaterally
MR and MR Arthrography 47

References
Ahn JM, Kwak SM, Kang HS et al (1998) Evaluation of patellar Keene GS, Villar RN (1994) Arthroscopic anatomy of the hip:
cartilage in cadavers with a low-field-strength extrem- an in vivo study. Arthroscopy 10:392–399
ity-only magnet: comparison of MR imaging sequences, Klaue K, Durnin CW, Ganz R (1991) The acetabular rim syn-
with macroscopic findings as the standard. Radiology drome. J Bone Joint Surg Br 73:423–429
208:57–62 Kramer J, Recht MP (2002) MR arthrography of the lower
Altenberg AR (1977) Acetabular labrum tears: a cause of hip extremity. Radiol Clin North Am 40:1121–1132
pain and degenerative arthritis. South Med J 70:174–175 Kramer J, Stiglbauer R, Engel A et al (1992) MR contrast
Balaban RS, Ceckler TL (1992) Magnetization transfer contrast arthrography (MRA) in osteochondrosis dissecans. J
in magnetic resonance imaging. Magn Reson Q 8:116 Comput Assist Tomogr 18:254–260
Brown RR, Clarke DW, Daffner RH (2000) Is a mixture of gado- Kramer J, Scheurecker A, Engel A et al (1997) Magnetic reso-
linium and iodinated contrast material safe during MR nance arthrography: benefits and indications. Adv MRI
arthrography? AJR Am J Roentgenol 175:1087–90 Contrast 4:104–119
Czerny C, Hofmann S, Neuhold A et al (1996) Lesions of the Lecouvet FE, Vande Berg BC, Malghem J et al (1996) MR imag-
acetabular labrum: accuracy of MR imaging and arthrog- ing of the acetabular labrum: variations in 200 asymptom-
raphy in detection and staging. Radiology 200:225–230 atic hips. AJR Am J Roentgenol 167:1025–1028
Czerny C, Hofmann S, Urban M et al (1999) MR arthrography Leunig M, Werlen S, Ungerböck A et al (1997) Evaluation of
of the adult acetabular capsular-labral complex: corre- the acetabular labrum by MR arthrography. J Bone Joint
lation with surgery and anatomy. AJR Am J Roentgenol Surg Br 79:230–234
173:345–349 Magee T, Hinson G (2000) Association of paralabral cysts
Czerny C, Kramer J, Neuhold A, Urban M, Tschauner C, Hof- with acetabular disorders. Am J Roentgenol 174:1381–
mann S (2001) Magnetresonanztomographie und Mag- 1384
netresonanzarthrographie des Labrum acetabulare: Ver- McCarthy JC, Busconi B (1995a) The role of hip arthroscopy
gleich mit operativen Ergebnissen. Fortschr Röntgenstr in the diagnosis and treatment of hip disease. Can J Surg
173:702–707 38:13
Czerny C, Oschatz E, Tschauner C, Hofmann S, Kramer J (2002) McCarthy JC, Busconi B (1995b) The role of hip arthroscopy in
MR-Arthrographie des Hüftgelenks. Radiologie 42:451–456 the diagnosis and treatment of hip disease. Orthopedics
Dameron TB (1959) Bucket-handle tear of acetabular labrum 18:573–576
accompanying posterior dislocation of the hip. J Bone Miller TT (2000) MR arthrography of the shoulder and hip
Joint Surg Am 41:131–134 after fluoroscopic landmarking. Skeletal Radiol 29:81–84
Delfaut EM, Beltran J, Johnson G et al (1999) Fat suppression Niitsu M, Nakai T, Ikeda K et al (2000) High-resolution MR
in MR imaging: techniques and pitfalls. Radiographics imaging of the knee at 3 T. Acta Radiol 41:84–88
19:373–382 Palmer WE (1998) MR arthrography of the hip. Semin Muscu-
Die T, Akamatsu N, Nakajima I (1991) Arthroscopic surgery of losk Radiol 2:349–361
the hip joint. Arthroscopy 7:204–211 Peh WC, Cassar-Pullicino VN (1999) Magnetic resonance
Disler DG, McCauley TR, Ratner LM et al (1997) In-phase and arthrography: current status. Clin Radiol 54:575–587
out-of-phase MR imaging of bone marrow: prediction of Petersilge CA (1997) Current concepts of MR arthrography of
neoplasia based on the detection of coexistent fat and the hip. Semin Ultrasound CT MR 18:291–301
water. AJR 169:1439–1447 Petersilge CA (2000) Chronic adult hip pain: MR arthrography
Dorell JH, Catterall A (1986) The torn acetabular labrum. J of the hip. Radiographics 20:S43–S52
Bone Joint Surg Br 68:400–403 Petersilge CA (2001) MR arthrography for evaluation of the
Engel A, Hajek PD, Kramer J (1990) Magnetic resonance knee acetabular labrum. Skeletal Radiol 30:423–430
arthrography. Acta Orthop Scand 61 [Suppl 240]:1–47 Petersilge CA, Haque MA, Petersilge WJ et al (1996) Acetabular
Fitzgerald RH (1995) Acetabular labrum tears. Clin Orthop labral tears: evaluation with MR arthrography. Radiology
311:60–68 200:231–235
Haims A, Katz LD, Busconi B (1998) MR arthrography of the Peterson DM, Carruthers CE, Wolverton BL et al (1999) Appli-
hip. Magn Reson Imaging Clin North Am 6:871–883 cation of a birdcage coil at 3 Tesla to imaging of the human
Haller J, Resnick D, Greenway G et al (1989) Juxtaacetabular knee using MRI. Magn Reson Med 42:215–221
ganglionic (or synovial) cysts: CT and MR features. J Robbins MI, Anzilotti KF Jr, Lange RC (2000) Patient percep-
Comput Assist Tomogr 13:976–983 tion of magnetic resonance arthrography. Skeletal Radiol
Heron C, Hine A (2002) Magnetic resonance imaging. In: 29:265–269
Imaging of the knee – technique and applications. Davies Schmid MR, Nötzli HP, Zanetti M, Wyss TF, Hodler J (2003)
AM, Cassar-Pullicino VN (Editors) Springer, Berlin Hei- Cartilage lesions in the hip: diagnostic effectiveness of
delberg New York, pp 41–63 MR arthrography. Radiology 226:382–386
Hodler J, Trudell D, Pathria MN, Resnick D (1992) Width of Schnarkowski P, Steinbach LS, Tirman PF et al (1996) Mag-
the articular cartilage of the hip: quantification by using netic resonance imaging of labral cysts of the hip. Skeletal
fat-suppression spin-echo MR imaging in cadavers .Am J Radiol 25:733–737
Roentgenol 159:351–355 Steinbach LS, Palmer WE, Schweitzer ME (2002) MR arthrog-
Hodler J, Yu JS, Goodwin D et al (1995) MR arthrography raphy. Radiographics 22:1223–1246
of the hip: improved imaging of the acetabular labrum Suzuki S, Awaya G, Okada Y et al (1986) Arthroscopic diag-
with histologic correlation in cadavers. Am J Roentgenol nosis of ruptured acetabular labrum. Acta Orthop Scand
165:887–891 57:513–515
48 J. Kramer et al.

Tschauner C, Hofmann S, Czerny C (1997) Hüftdysplasie: Mor- arthrography: techniques and applications. Semin Ultra-
phologie, Biomechanik und therapeutische Prinzipien sound CT MR 18:302–306
unter Berücksichtigung des Labrums Acetabulare. Ortho- Wagner SC, Schweitzer ME, Weishaupt D (2001) Temporal
pade 26:89–108 behavior of intra-articular gadolinium. J Comput Assist
Ueo T, Suzuki S, Iwasaki R et al (1990) Rupture of the labra Tomogr 25:661–670
acetabularis as a cause of hip pain detected arthroscopi- Walker JM (1981) Histologic study of the fetal development of
cally, and partial limbectomy for successful pain relief. the human acetabulum and labrum: significance in con-
Arthroscopy 6:48–51 genital hip disease. Yale J Biol Med 54:255–263
Vahlensieck M, Wischer T, Schmidt A et al (1995) Indirekte Williams PL, Warwick R (eds) (1980) Arthrology: the joints of
MR-Arthrographie: Optimierung der Methode und erste the lower limb-the hip (coaxial) joint. In: Gray’s anatomy.
klinische Erfahrung bei frühen degenerativen Gelenkss- 36th edn. Saunders, Philadelphia, pp 477–482
chäden am oberen Sprunggelenk. Fortschr Röntgenstr Winalski CS, Weissmann BN, Aliabadi P et al (1991) Intrave-
162:338–341 nous Gd-DTPA enhancement joint fluid: a less invasive
Vahlensieck M, Peterfy GG, Wischer T et al (1996) Indirect alternative for MR arthrography. Radiology 181:304
MR arthrography: optimization and clinical applications. Wolf SD, Eng J, Balaban RS (1991) Magnetization transfer con-
Radiology 200:249–254 trast: method for improving contrast in gradient-recalled-
Vahlensieck M, Lang P, Sommer T et al (1997) Indirect MR echo images. Radiology 179:133

You might also like