Professional Documents
Culture Documents
INSTRUMENT CALIBRATION
Any questions or feedback about this Global Engineering Standard should be addressed to
the author:
Stephen Tate, Instrument Engineer, Pharma Supply Chain Engineering
Version
Date Revision Details
No
Section 7.8.2
• Addition of “It is common practice at many sites and across
the pharmaceutical industry to attach labels to instruments, it
must indicate the current instrument calibration status and the
next due calibration date (Notes 1 & 2)”.
• Changed Conform (or Pass) to “Pass (or Conform)” and added
“acceptable” to the requirement.
• Changed Non-Conform (or Fail) to “Fail (or Non-Conform)”
and added “acceptable” to the requirement.
• Addition of “the use of Failed or Overdue equipment” to the
Authorized requirement.
• Addition of “All the above have been the source of regulatory
observations in GSK” to note 1
• Addition of Note 2
Section 7.9 Addition of “appropriately as dictated by their
criticality”.
Section 7.10 Addition of “Incident and Investigation” and inclusion
4 Dec-20 of “appendix 4” reference
Section 7.10.1
• Addition of table number and description
• Addition of “It is important that all calibration deviations with
the potential to have a negative impact on Quality, Safety,
EHS and Compliance are investigated”.
• Addition of Guidance for reportable deviations
• Addition of figure 7.10.1.b
Section 8.3.1 Changed, day off sites to “days off site”. Inclusion of
“time since” explanation note
Section 8.5
• Change “Self-monitoring can be used to enhance confidence
between calibrations and can also be used to justify and
increase calibration interval. However self-monitoring should
not replace calibration activities”.
• Addition of “This can be achieved by Diagnostic and alarm
maintenance management software i.e. PDM (Siemens) or
AMS (Emerson)”
Table of Contents
DOCUMENT REVISION HISTORY...................................................................................... 2
1 RATIONALE ............................................................................................................... 6
2 POLICY ...................................................................................................................... 6
3 PURPOSE .................................................................................................................. 6
4 SCOPE ....................................................................................................................... 6
4.1 Within Scope....................................................................................................................... 6
4.2 Excluded from Scope ......................................................................................................... 6
4.3 Definitions ........................................................................................................................... 7
5 QUALITY COMPLIANCE REQUIREMENTS ............................................................ 10
5.1 Calibration ......................................................................................................................... 10
5.2 Contamination and Cross-Contamination ........................................................................ 10
6 EHS COMPLIANCE REQUIREMENTS .................................................................... 10
7 ENGINEERING MANDATORY REQUIREMENTS ................................................... 11
7.1 Overall Calibration Management Process ........................................................................ 11
7.2 Instrumentation List .......................................................................................................... 12
7.3 Instrument Criticality Assessment .................................................................................... 12
7.4 Set Up Critical Calibration Plans ...................................................................................... 14
7.5 Test Instruments and Working Standards ........................................................................ 16
7.6 Complete First Use Checks .............................................................................................. 17
7.7 Non-Routine Work ............................................................................................................ 17
7.8 Complete Calibration Activities ......................................................................................... 17
7.9 Create & Retain Calibration Records ............................................................................... 19
7.10 Deviation and Failures ...................................................................................................... 19
7.11 Manage Overdue Routines and Exceptions ..................................................................... 21
7.12 Defective Equipment ........................................................................................................ 21
7.13 Temporary Repairs ........................................................................................................... 22
7.14 Review of Calibration Plans .............................................................................................. 22
7.15 Skipping or Discontinuing Calibration Plans ..................................................................... 22
7.16 Out of Service Equipment ................................................................................................. 22
8 ENGINEERING BEST PRACTICE ........................................................................... 22
8.1 Overall Calibration Management Process ........................................................................ 22
8.2 Device assessment .......................................................................................................... 23
8.3 Calibration scheduling ...................................................................................................... 23
8.4 Instrument Verification ...................................................................................................... 24
8.5 Self-Monitoring.................................................................................................................. 24
9 REFERENCES ......................................................................................................... 25
Appendix 1 – Flowcharts for instrument criticality assessment .................................. 26
Appendix 2 - Examples of Critical Instruments by Category ........................................ 27
Appendix 3 – Flowchart for considerations to be taken on event of instrument failure
29
Appendix 4 - Decision Table for Deviation Impacts....................................................... 30
Appendix 5 - Audit Checklist ........................................................................................... 31
1 RATIONALE
Usage of inappropriately calibrated instruments during manufacturing operations can
adversely affect product quality, data integrity or increase the risk of EHS incidents.
A robust and effective instrument calibration management process is therefore
necessary to minimise the above risks.
2 POLICY
All sites must identify Quality critical, EHS critical and Business critical instruments
they must also define a calibration management process to ensure the calibration of
critical instruments are within appropriate ranges and tolerances.
Sites must also comply with any relevant local legislation or regulatory requirements in
addition to the requirements defined in this Global Engineering Standard (GES).
3 PURPOSE
This GES aligns and supports GES103 (Maintenance), it defines the minimum
mandatory requirements, engineering best practices and guidance for the
management of Quality critical, EHS critical and Business critical instrument
calibrations.
This GES aligns with and supports GSK’s QMS and EHS standards and guidelines.
4 SCOPE
4.3 Definitions
ASSET
Any physical item or equipment that is maintainable and brings value to the
enterprise. This includes but is not limited to:
• temporary and permanent facilities and equipment,
• Utilities,
• critical monitoring devices, and
• fixed, mobile and portable manufacturing and test equipment
MAINTENANCE
Activities necessary for retaining or restoring an asset to the condition that achieves
its maximum useful life.
It includes corrective, preventative, predictive maintenance, inspection, testing,
calibration, adjustment, refurbishment and/or replacement of asset or components.
It includes routine and non-routine maintenance.
INSTRUMENT
A device that produces an output signal (pneumatic, electronic / electrical, serial
data, digitally encoded etc.) in a defined relationship to a physical variable. For the
purposes of this GES, this definition does not apply to hand / limit switch devices.
CALIBRATION
The demonstration that an instrument or a measuring device produces
measurements within specified limits by comparison with those produced by a
reference or traceable standard over an appropriate range of measurements.
CALIBRATION TOLERANCE
MAINTENANCE PLAN
A programme of planned maintenance activities completed at regular intervals to
ensure that the asset continues to operate within its required parameters and that
any risks to product quality are eliminated or minimised.
ROUTINE MAINTENANCE
A set of pre-defined activities completed at regular intervals as part of a Maintenance
Plan.
NON-ROUTINE MAINTENANCE
Any maintenance that is not part of a Maintenance Plan.
SCHEDULING TOLERANCE
A defined interval before or after the scheduled due date for completion of
maintenance activities in a Maintenance Plan. This is used to define the earliest and
latest acceptable completion dates for routine maintenance activities.
The latest acceptable date to complete Routine Maintenance is the date defined by
the scheduled due date plus the scheduling tolerance. Any maintenance plan
activities completed after this date are considered overdue.
AS-FOUND RESULTS
The inspection results before any adjustment or other maintenance activities are
applied.
AS-LEFT RESULTS
The calibration or inspection results, after any adjustment or other maintenance
activities have been completed and before returning the asset to service.
If there has been no adjustment or other activities affecting the calibration or
inspection, then no ‘as left’ results are required.
CMMS
A Computerised Maintenance Management System used to record asset details
and/or spare part details and to plan and record all maintenance activities / spares
usage.
5.1 Calibration
Quality requirements for the calibration of instruments are defined in the following
QMS documents:
• GQMP 1008 – Validation Life Cycle
• GQP 1204 – Validation Life Cycle
• PSC TS 002 – Maintenance and Calibration
Range +
Device assessment * 7.3.4 Calibration
Tolerance
Calibration
Interval + Set Up Critical Calibration Plan 7.4 Calibration Schedule
method
The instrument list (DDS 09) is derived from the Asset register and must identify the
hierarchical relationship between parent systems/equipment and child
instruments/loops (as a single item on the register list or include in a parent system).
Record and approve the results of the instrument criticality assessment, including
rationale for the decision, and retain the record. Quality must approve the Quality
Critical Instrument assessment.
Note:
The term “robust” in this sense means that you must consider the environment for the label design:
temperature / humidity may impact the label, sterilization agent in classified room.
Critical Process Parameters (CPP) and Process Parameters (PP’s) with Quality Impact
have specified levels or, in some cases, absolute limits for acceptable operation.
Review and update these values when you change the process change or introduce a
new product.
Fig 7.3.4 - Instrument permitted limits
Adjustment Limit is the maximum calibration error, above which you must adjust the
instrument. The ‘As Left’ calibration error, after adjustment, must be less than the
adjustment limit.
Notes:
The adjustment limit needs to consider the uncertainty of measurement, else time will be lost adjusting
the instrument. Therefore, it should be specific for each measurement and part of the calibration system
Alternatively, you can use a generic site approach like “Adjustment limit = calibration tolerance x 50%”
The calibration points must encompass the Proven Acceptable Range in full, (Refer
to Fig 7.3.4 - Instrument permitted limits)
If it is not possible or if the proven acceptable is unknown, the calibration range must
encompass the normal operating range. In this case, you need to define the process
control (alert/alarms triggers) accordingly.
Express the calibration tolerance in engineering units (e.g. oC, kg, mbar, Ohms).
Do not use percentage of full scale as it creates confusion on the scale to consider
(process range, sensor technical capacity, display range).
If you are using percentage of reading, consider the risk of generating out of tolerance
around zero.
Notes:
Setting up the calibration tolerances in the unit used by the process enhance a better understanding for
users and facilitates the unplanned event management.
•Traceable in order to allow identification of all the devices that have been
calibrated with the test instrument, in case it is found to be out of tolerance,
damaged or lost
When an out of tolerance or an overdue calibration occurs, you must:
• investigate in the same way as for field instruments as described in GES 103
Section 7.12
• assess the impact on all the device that were calibrated with the test instrument
Notes:
If there is no recognised national, international or industry recognised physical standard, you must use an
independent reproducible standard. Where no applicable standard exists, you must establish an in-house
standard)
Wherever possible, the calibration tolerance of any test instrument or working standard should be less
than 25% of the calibration tolerance of the field instrument which it is being used to calibrate.
Similarly, the reference standard which is used to calibrate the test instrument or working standard should
be at least four times more accurate than the test instrument, and so on up through the traceability ladder.
It is recognised that this may not be possible in all cases (e.g. relative humidity measurement).
If the working standard is a buffer solution, the traceability is required, especially to ensure that it must be
purchased from accredited laboratories, storage and use before expiry date.
To allow calibration of the installed instrumentation, ensure that the required test
equipment is available on site for the first and future on-site calibrations.
Note:
It is recognised for specialist calibration laboratories to provide a test certificate showing the raw calibration
data, but not a PASS / FAIL result. In these cases, you must determine and record the PASS / FAIL result
on receipt of the test certificate in relation to the defined internal calibration tolerance.
Where new instruments are supplied with a calibration certificate, this may in certain
circumstances and with agreement be used as the first calibration without the need to
carry out a first calibration on site.
Note:
1. The recommendation (not mandatory) is to remove the labels for the follow reasons:
2. Several highly regulated sites in GSK do NOT use calibration date labels and have successfully
demonstrated to some regulators that a robust calibration management system is adequate to control
calibration completion dates, due dates and overdue calibrations.
However, the removal of calibration labels will require site representation approval i.e. Quality, EHS etc
and, must be accepted by all and agreed with regulatory agencies impacted by the site activities.
Table 7.10.1a - Describes the action which must be taken in event of a calibration FAIL
It is important that all calibration deviations with the potential to have a negative
impact on Quality, Safety, EHS and Compliance are investigated.
• Any device that was not found or was unavailable for calibration.
• Any of the above that have an impact to GxP, Statutory or EHS
In addition, in order to increase the calibration interval, you must consider the following:
Where loops have a number of successive successful calibrations (without adjustment)
then you can extend the calibration interval in line with the diagram below not by more
than two thirds of the time period over which the loop has proven itself.
Fig 7.14 - Increasing the calibration interval
Original Interval
6 Months
6 Months
1 Year
Always consider the point where deteriorative effects of age could affect the instrument
and the potential impacted period to investigate when the instrument will actually break
or drift.
Before an extension to a quality critical instrument interval, you must consider with
quality the potential impact to Production batches, if the instrument were to fail its
calibration.
It represents the percentage of calibration failures with potential product impact. Only focus
on As Found value. Consider Out Of Tolerance result or no result (broken, lost).
• Late execution (overdue) Usual target <3%
This ratio corresponds to the number of expired work order divided by the number of work
orders planned (equal to the invert of the schedule adherence).
Example: 100 work orders of calibration have been planned for January, 2 work orders have
not been completed. The late execution ratio is 2% (schedule adherence of 98%)
The Vaccines SPC 130 “Technical specification for Determination of adjustment limits
and calibration tolerances” suggest pre-established sensor criticality and tolerances
for standardized equipment and provide rationales. It provides a criticality analysis
method applicable to Calibration activities which can be used to prioritize and review
calibration plan / metrological specification
In this case, take into consideration the full scheduling (due date and calibration
interval of both work orders).
Using “days off site” can also be option, considering the maximum timeframe accepted.
Note:
Working with “Time Since” schedule for external calibration increases equipment’s availability. Schedule
dates will shift one transport period each year.
“Time since” = when the unit has been calibrated and returned to site, the date on the calibration certificate
is the schedule start date. So, if the off-site calibration takes 1 month then each year the removed
equipment / calibration date would move forward one month to allow the equipment’s full years calibration
period to be utilised
8.5 Self-Monitoring
Self-monitoring can be used to enhance confidence between calibrations and can also
be used to justify and increase calibration interval. However self-monitoring should not
replace calibration activities.
One practical way to realise it is to define one business critical instrument which
realizes the same measurements to a quality critical instrument and to monitor the
difference between the measurements. When this difference gets too large, start a
calibration to check which sensor has drift.
Smart instrumentation can also provide a similar benefit through detecting unusual
tendencies, when the instrument drifts. This can be achieved by Diagnostic and alarm
maintenance management software i.e. PDM (Siemens) or AMS (Emerson)
9 REFERENCES
GQP 3201B Data Management and Control
GQP 4301 Facilities Operation
GQP 4303 Maintenance
GQP 4304 Calibration (Retired)
GQP 4309 Cosmetic Products: Maintenance and Calibration
GQP 6302 Deviation Handling
GQMP 1005 Change Control
GQMP 1008 The Validation Life Cycle
GQMP 6007 Deviation Handling
GQP 2101 Organisation, Personnel Management and Training
GQMP3001 Documents and data – life cycle management
GQMP 4006 Outsourcing of engineering And facility services: assessment and
control of risks
PSC TS 002 Maintenance and Calibration
EHS SG 2.03 Investigation and Reporting
EHS SG 2.01 Risk Assessment and Control
EHS SG 5.16 Electrical and Machinery Safety
GES 101 Commissioning & Qualification
GES 102 Plant Modifications
GES 103 Maintenance
GES 110 Machinery Safety
GES 113 Permit To Work
GES 114 Contractor Management
GES 502 BPCS Alarm Design & Management
GES 503 Safety Instrumented Functions
GES 601 Design of Document Archives
POL-GSK-506 Global Records Retention Policy
POL-GSKF-507 Computerised System Validation
SPC130 Technical specification for Determination of adjustment limits and
calibration tolerances for manufacturing equipment
Calibration Optimisation Blueprint
CAP045 Compendium of Analytical Procedures for Analytical Equipment
Performance/Maintenance requirements and Standard Platforms
NO
NO
N
NO
N Is this “parallel” or
Could instrument O “downstream”
malfunction result in product YES instrument/measurement a NO
being misidentified? Quality Critical
instrument/measurement
NO
NO
NO
NO
NO
NO
Instrument is Instrument is
Not EHS Critical EHS Critical
NO
NO
Instrument is Instrument is
Not Critical Business Critical
in C&M system
CONTROLLED COPIES OF THIS DOCUMENT ARE ACCESSED FROM A CONTROLLED SERVER OR PRINTED ON
OFFICIAL (E.G. RED HEADED) PAPER.
USERS MUST CHECK THEY HAVE THE CURRENT VERSION