Professional Documents
Culture Documents
5、Component 04-chamber.......................................... 37
5.1 layout of M600-04-E00-Chamber PCB and function........... 37
5.1.1 M600-04-E00 PCB ................................................... 37
5.1.2 M600-04-E00 function describing ............................. 37
5.1.3 M600-04-E00 electronic components ....................... 37
5.2 Chamber function and trouble shooting .............................. 39
5.2.1 Chamber interfaces function description .................. 39
5.2.2 Possible problems and solution for Chamber ........... 40
5.2.3 Chamber pumps trouble shooting ............................ 42
5.2.4 PMT trouble shooting ............................................... 44
1、Whole analyzer
1
Figure 1.2 The DC voltage output of the switching power supply
3) If the switching power supply have +24V output, then pull out the
cables between M600-01-E01 and other main board one by one, and
check which PCB board is short circuit.
Figure1.3 M600-01-E01
4) If there is no output of the switching power supply, then check the input
terminal whether there is 220V or not.
5)If the switching power supply input voltage is 220V, but there is no 24V
output, that means the switching power supply is damage.
2
6)If there is no 220V input voltage, check the two fuses inside the main
switch and the cable from the main switch to the switching power
supply.
(2)Deputy switch unable to control the power supply of each main boards
Possible reasons:
1)Damage of the deputy switch
2)Damage of the cable which connecting to the deputy switch
Solutions:
1) Check the soldered dot of the red and blue wires
2) If the connection is OK, then the problem come from the deputy switch
3
1) Test the voltage of the +5V、-15V、-15V testing holes on PCB 04-E00
The heating module and cooling module are circuit loops which are
composed of power supply switch, sensor and temperature control device. The
sensor is used to check whether the temperature is normal. Usually one
centigrade is equal to 10 mV output voltage of the sensor.
4
D1 of the 03-E05 PCB is on, Test whether the voltage of the
temp-sensor signal wire (TP2 test hole) is normal; if it’s normal, then
check whether the cable of the temperature sensor is fixed; if it’s fixed,
check the resistance between the J6 heating metal, the resistance is
24±4Ω normally. If the resistance is damage, the multi-meter shows no
resistance; if all above are OK, then change the 03-E05 board.
2) Cooling function of the reagent area is abnormal. Check whether the
fans under the reagent area work normally, if not, check the soldering
cables or change the fans (The temperature will be too high to melt the
fuse without the fans). Test whether the temp-sensor signal wire (J4
yellow signal wire) is normal. If all above are OK, change the -02-E04
PCB.
The liquid levels of the starters are detected by capacitance method. And
the system liquid and waste levels are detected by reed switch.
5
5) Damage of the reed switch.
Solutions:
1) Check whether the detecting stick is clean; Check whether the
rubber o-ring is missing; Check whether the plate metal under the
starters is connected with the analyzer’s shell correctly; Check
whether there’re liquid drops on the M600-01-E02 PCB
O ring
信号接头
LLD connector
短路档
Short-circute function
Reed
磁簧开关Switch
6
1.4 Check the grounded status of the analyzer
(1)If the power supply is grounded, choose the short-circuit function of the
multi-meter. Use the two heads of the multi-meter to touch the analyzer
(metal shell or the metal screw) and the ground wire interface of the
multi-socket respectively. If the multi-meter alarm and beep continuously
(the resistance is 0), it means that the analyzer is connected good with
the multi-socket properly
Figure 1.12
(2)Soldering joint of Sensor
7
Figure 1.13
(3)Soldering joint of Encoder
Figure 1.14
(4)Soldering joint of heating metal
Figure 1.15
8
2、Component 01-COP
ID
Type Explanatory Notes Remarks
number
D4 Indication light of communication between Green
the 485 BUS line and COP sending end,
LED normally appears as twinkle
D5 Indication light of communication between Blue
the 485 BUS line and COP receiving end,
9
normally appears as twinkle
D3 Indication light of power supply(voltage
yellow
3.3V)
D2 Indication light of power supply(voltage
yellow
5V)
D1 Indication light of power supply(voltage
yellow
24V)
D6D7 Attended to be used by technician for Blue
system development
Fuse F1 24V Fuse 5A
Jumper JP2 Please insert JP2 and JP3 when proceed
JP3 program upgrading for processor and
remove them after update finished(insert
to JP1 and JP4)
3.1.3 M600-02-E00-Sample&Reagent
D13 Indicating light and signal testing hole for the sensor of green
TP13 shake speed monitoring.
When the sensor of shaker motor is not blocker, the D13
LED is on, the voltage at TP13 is below 300mV;
When the sensor is blocked by sheet metal, the D13 is
off, and the voltage at TP13 is above 4.5V;
D14 Sensor LED for inner pipetting arm initialization and green
LED TP9 hole for signal test.
When the sensor is not blocked by the sheet metal on
the inner arm, the D14 LED is on, and the voltage at
TP9 is below 300mV;
When the sensor is blocked, the D14 LED is off, and the
voltage at TP9 is above 4.5V
D15 Sensor LED for pipettor’s vertical initialization and hole green
TP11 for signal test.
When the sensor is not blocked by the vertical sheet
metal, the D15 LED is on, and the voltage at TP11 is
below 300mV;
When the sensor is blocked, the D15 LED is off, and the
voltage at TP11 is above 4.5V
D16 Sensor LED for outer pipetting arm initialization and green
TP10 hole for signal test.
When the sensor is not blocked by the sheet metal on
the outer arm, the D16 LED is on, and the voltage at
12
TP10 is below 300mV;
When the sensor is blocked, the D16 LED is off, and the
voltage at TP10 is above 4.5V
D1 LED for +24V power supply yellow
D2 LED for +15V power supply yellow
D3 LED for +5V power supply yellow
D4 LED for +3.3V power supply yellow
D5 Program debugging LED, should only be used by yellow
program developer
Fuse F1 +24Vpower fuse 5A
Upgrade J4 Connector for upgrade the left arm
connector
(1)outer and inner pipetting arm, vertical motor and encoder transfer
(2)liquid level detection signal transfer
13
3.3 Possible problems and solutions for pipettor
2)If the sensor is ok, but the motor unable to move and is not locked when doing
inner arm initialization, it suggests that the M600-02-E00 PCB is damaged and
need to replace a new one.
14
3)If the initialization and wire connection are ok, the external resistance does not
exist and the arm is not locked when power on, it suggests the inner arm encoder is
damaged and need to change it together with the motor.
(1)Unable to initialize the vertical direction, the probe cannot move up during
initialization
Possible reasons:
1)The probe hits something in the vertical direction
2)The sensor of the vertical direction is damaged
3)The M600-02-E00 PCB is damaged
4)The vertical motor is damaged
Solutions:
1)Adjust the probe to the correct position if it had hit something
2)Check whether the D15 LED changes its status when the sensor is blocked and
unblocked. If not, the sensor is damaged and need to be replaced.
16
3.3.4 Pipetting possible problems and solutions
17
Fig 3.6 “Detect” function in service software
4)If all above are ok, change the sensitivity parameter of LLD gradually(the value is
25 normally),the sensor becomes less sensitive as the value decrease.
5) If the needle is clashed, adjust it back to original position.
6) If the needle is rusty, please check whether there are some rust or crystal material
sticks to the welding point of needle PCB board.
(1) Unable to read the sample barcode information which is closest or farthest to
the barcode reader
Possible reasons:
1) Barcode reader installation problem
Solutions:
Check whether the barcode reader is installed properly, the barcode reader should be
perpendicular to the instrument bottom. The distance between laser release point and
reagent/sample area should be more than 60mm.The Laser bean should not be blocked
by anything.
18
(3)Sample/reagent area door switch doesn’t work
Possible reasons:
1) Connection cable not well connected
2) Magnetic switch malfunction
3) The magnet inside the Sample/reagent area door unable to be close to the
magnetic switch
4) M600-02-E04 PCB board malfunction
Solutions:
1) Check the cable connection between M600-02-E04 and magnetic switch. Check
the flat cable connection to the main PCB board of the Maglumi analyzer.
2) Use a magnet material to get close to the magnetic switch, and the buzzer on
M600-02-E00 PCB board should have a “beep” sound. At the same time, it
should be short circuit between both cables of the magnetic switch. If not, then
the magnetic switch should be replaced.
19
Fig 3.8 M600-02-E04 PCB board
(5)Abnormal speed of the reagent shaker
Possible reasons:
1)Affected by strong mechanical resistance
2)06-E02 PCB board malfunction
Solutions:
1)Check the cable connection of shaker motor and 06-E02 bonding wire
2)If you can move the shaker freely and feel no resistance, this indicates
M600-02-E00 malfunction and need to be replaced
3)If the shaker blocked when move to a certain position, this indicates this position
exist some mechanical resistance.
4)If there is a warning even the shaker moving normal, use the multi-meter to
measure the voltage at TP13 testing hole in M600-02-E00 PCB.
20
Possible reasons:
1)Temperature sensor malfunction
2)Fan malfunction
3)M600-02-E04 PCB board malfunction
4)Peltier malfunction
Solutions:
1)Use multi-meter to measure the yellow signal wire at J4,to check the output signal
of the temperature sensor.
2)Check whether the fan can work normally. If it doesn’t move, then check the
voltage of J8 socket. If the voltage is +24 Volts, then you need to replace the fan.
3)Check the voltage between red and black wire of J3 socket. If the voltage is +24V,
it indicates that the peltier is failure and need to be replaced. If there is no voltage or the
voltage is very low, then replace M600-02-E04 interface board.
21
1)Diluter pump is blocked by system liquid crystal
2)optical sensor in diluter pump is corroded by system liquid
3)M600-02-E00 PCB board malfunction
Solutions:
1) Check whether there are some crystals block the motor or the connector of the
diluter pump. Power off the analyzer, use screw driver to rotate the screw at the
bottom of the motor and check whether the motor can move smoothly. If there is
some friction or the motor can’t rotate totally, this indicates the connecting point
between the motor shaft and the pump is blocked and the pump needs to be
replaced.
Fig 7.7 Use screw driver to rotate the screw at the bottom of the motor
2) Disassemble the diluter pump optical sensor and check whether it is corroded by
system liquid. At the same time check the D12 LED indicator light and measure
the voltage of TP12 test hole to check whether the sensor is damaged, if it is,
then change the pump.
3) If the pump works well and no blocking by any crystal material, replace the
M600-02-E00 PCB.
(3)Abnormal washing liquid injection of the pipetting needle
Possible reasons:
1)Washing control valve malfunction
2)Vacuum pump malfunction or not connected well
3)Pipetting probe or tubing blocking
3)M600-02-E00 PCB board malfunction
Solutions:
1)Check the cleaning valve voltage, the voltage should be >23V if it works well
2)Check the voltage of vacuum pump. If it work well, the voltage should be >23V
3)If above voltages are normal but the liquid in the tube cannot move, this indicates
the valve or vacuum pump malfunction; If the voltages abnormal, this indicates
cable connecting not well or M600-02-E00 PCB board malfunction.
4)Check whether there are some crystal or other material block the tubing system.
22
4. Component 03-incubator &Wash station &Pusher
24
optical sensor is not blocked , LED on, TP9 <200mV
D10 Incubator cuvette detection LED and test hole green
TP10 When there is cuvette under the sensor, the LED will on,
TP10 voltage <300mV; No cuvette under the sensor,
LED off, Tp10 voltage >4.5V
D12 Incubator initial sensor LED and test hole green
TP11 When incubator at initial position, LED on, Tp11<300mV
When incubator at other positions, LED off, TP11 >4.5V
D17 Pipetting load incubator position LED and test hole green
TP16 When load at incubator position, incubator sensor is
blocked, LED off, Tp16 voltage >4.5V
Other positions, the sensor will not blocked, LED on,
and tp16 voltage <200mV
D18 Pipetting load back transport position LED and test green
TP17 hole.
The wash station sensor can be blocked when pipetting
load at back transport position, LED off, tp17 >4.5V
Other positions, LED on, tp17 <200mV
D9 Program debugging indicator LED, developers use only blue
D11 Program debugging indicator LED, developers use only blue
D16 +3.3V power indicator and test hole yellow
TP14
D15 +5V power indicator and test hole yellow
TP13
D13 +24V power indicator and test hole yellow
TP12
Fuse F1 +24V fuse 5A
upgrade J4 upgrade interface
interface
(1)03-E01
Right Pipetting area temperature control, wash station interface PCB and Circuit switching
device
25
Fig4.2 03-E01 PCB
(2)03-E02
Hall sensor to detect the initial position of wash lift
(3)03-E03
Optical sensor to detect initial sensor of the wash transport
(4)03-E04 interface PCB
Back Transport initialization and positions detecting by optical sensor
(1)M600-03-E05
Control for the incubator temperature, heating, motor, encoder, Hall sensor, temp
sensor and other parts. The Incubator movement and position record are controlled by the
encoder. Hall sensor can sense the magnet field, then output different voltage signal to
control the incubator to find the initial position.
(1)M600-03-E08
Interface for motors and sensors which controlled the washer loader and sample
loader. Each loader has two optical sensors located in the left and right, using for control
the movement of the loader.
(1)02-E02
Pusher includes Motor, encoder, optical sensor and other parts.
Encoder can control the Pusher movement and record the position. And the initial
position is detected by the optical sensor.
27
Figure4.5 Pusher
(4)Back Transport move abnormally, not initialize or move left/right not possible
Possible reasons:
1) Optical sensor mistake or broken
2) Blockage in the drill hole of the Back Transport fork
3) Optical sensor displacement due to large gap between geared fork and motor
wheel
4) Not proper matched between incubator, pusher and back transport
5) Motor is loose
Solutions:
1) Disassemble the fork to clean the blockage/dust in drill hole
2) Check the cooperate between geared fork and motor wheel
3) Push the fork manually, observe the indicator LED D3 on the M600-03-E00 PCB,
normally the LED should have status changes during the movement of the fork,
and only the sensor match the drill hole it will on.
4) If D3 doesn’t change status from on to off all along the movement, please check
the output voltage of testing hole TP18 on 03-E00(Normally the value is over 4V
when drill hole of fork faces against optical sensor, otherwise is less than 0.5V ).
If the voltage is out off the normal value, it needs to change the optical sensor. If
the voltage is ok, it means there is problem from M600-03-E00 PCB which
should be replaced.
5) Check each position fitted between the right Pipetting channel and the incubator
slot (16 slots) with a piece of cuvette. Adjust the position properly in the service
software if necessary. Otherwise, replace the fork if it’s bended.
6) Check and adjust the driving motor position to match with the fork gear
30
(6) Peristaltic Pump aspiration abnormal, overflow detection failure
Possible Reasons:
1)Too much contact pressure of the peristaltic pump wheels
2)one or more washer needles are clogged
3)Overflow sensors(a pair of hooks) broken or incorrect connection
4)liquid drops remains between injection needle and aspirate needle
5)M600-03-E00 PCB failure
Solutions:
1) If peristaltic pump could not rotate during aspiration, check the wire connection
between the pump and back penal PCB. Otherwise, replace M600-03-E00 PCB.
2) Check the tension of the aspirate tubing by adjusts the screw.
3)Clean the washer needle.
4)Check the connection of the flooding sensor and the soldering junction on 03-E01
Interface PCB, normally there are two groups of connectors, one group connect to the
injection needles and the other connect to three aspiration needles( Figure 4.9 ).
Check the sensor connection by multi-meter.
(7) Washer Reflect sensor can’t detect cuvette when initialize with clear cuvette.
Possible Reasons:
1) The sensitivity of the sensor is not adjusted well.
2) Broken of the sensor
3) PCB M600-03-E00 failure
Solutions:
1) Measure the voltage of TP7 located on PCB M600-03-E00, normally the voltage
should be > 4V when there is cuvette there, otherwise the voltage should be <
500mv. If the voltage is abnormal please check the connection of the sensor,
whether it’s good or not, or replace the sensor.
31
Figure 4.10 installation of the reflect sensor.
2) Measure the voltage of TP10 located on PCB M600-03-E00. Enter service
software and use the washer loader to push a cuvette to the wash station, then
adjust the W2 potentiometer, to make the D10 on (clockwise make it off, and
anticlockwise make it on). After adjustment the indicator light D10 should be on
when have cuvette block the sensor and off without cuvette.
33
Figure 4.13 W1Potentiometer on PCB M600-03-E05
4) If the temperature is far away from normal range, this could not be correct by
adjusting the potentiometer, it needs to check the temperature sensor or replace
the PCB M600-03-E05.
(1) Washer loader move abnormal, not initialize or move forward\backward failure
Possible reasons:
1) Cuvette crash between the washer and incubator
2) The sensor is not adjusted well or broken
3) PCB M600-03-E00 broken
Solutions:
1) Check whether there is cuvette in incubator or washer cause the cuvette crash.
2) Check the incubator positions, whether all the slots match the wash station well
3) Check the optical sensor of the loader; observe the indicator light D5 and D8 locate
on PCB M600-03-E00, whether the light has signal change according to the sensor
status, otherwise the sensor is broken.
4) When the loader move to one side (left or right), make sure the plate which on the
belt should block the senor. Otherwise it may have a dead zone and cause the
initialize fail. Need to adjust the sensor position.
35
Figure 4.15 the sensor of the loader
36
5、Component 04-chamber
Fig5.1 M600-04-E00
38
5.2 Chamber function and trouble shooting
39
Connect with pump motor of starter1and 2, optical sensor
Connect with chamber waste pump
Figure 5.3 Interface PCB of the optical sensor for chamber transport
40
Possible reasons:
1) Damage of the optical sensor for the chamber lift.
2) The position of the chamber lift is too high before initialization.
3) Damage of the M600- 04-E00 PCB or 04-E01 interface PCB.
Solutions:
1) Move the chamber lift manually to block or unblock the optical sensor, check
whether the status of D3 on the 04-E01 interface PCB will be changed. (If the
sensor works normally, when being blocked, the LED is on, otherwise the LED is
off.)
2) If the status of the D3 doesn’t change, test the voltage of the test hole on the
04-E01 interface PCB. Replace the optical sensor interface PCB when the
voltage is abnormal, while change the 04-E01 PCB when the voltage is normal.
3) If the status of the D3 changed normally, the M600-04-E00 PCB needs to be
changed.
4) The entry and exit of the chamber should not be higher than the chamber channel.
Otherwise it’s unable to initialize.
41
5.2.3 Chamber pumps trouble shooting
42
2) Observe whether the pump works when the analyzer do the soak process. If the
pump doesn’t work, test the voltage of the red power cable, the normal range
should be 14¬ 20V, if the voltage is out of range, change the M600-04-E00 PCB.If
the pump works, but the waste cannot be taken out, check whether the suction
needle or the soak pump is blocked by some solid particles.
43
5.2.4 PMT trouble shooting
Figure 5.9
(1)Measuring result is 0
Possible reasons:
1) Bad contact of the power wire or signal wire of the PMT.
2)Damage of the PMT PCB.
3)Damage of the M600-04-E00PCB.
Solutions:
1)Check the connection of the signal wire and power wire of the PMT.
2) Disconnect the power wire of the PMT, run the “background” in Service software.
At the same time, test the voltage of the test hole on 04-E01 PCB. The normal
range should be 1.5V ~ 2.2V. If the voltage is out of range, change the
M600-04-E00 PCB.
3)If all the above checking are Ok, change the whole PMT component.
(2)the value of the background is high and unstable
Possible reasons:
1)light leakage of the PMT or chamber.
2)PMT or the ground wire is not grounded well.
3)Damage of the PMT component.
Solutions:
1)Re-install the PMT component and lock it tightly; Use some black cloth cover the
chamber and then test the background again. If the value of background is
decrease, maybe there is light leakage from the PMT or the chamber. If the value
still be high and unstable, change the whole PMT component.
2)Check the analyzer grounding. Check the connection between the PCB of PMT
and the analyzer’s metal shell. There should not be insulative materials contact to
the metal.
3)If there is no light leakage and the grounding is good, but the value of the
background is still high and unstable. Change the PMT component.
(3)Blue LED defect: “No cuvette in chamber”, or the value of “Ref led
on” in the background is >30W or <1W
Possible reasons:
1)Damage or aging of the blue LED.
2)Damage of the 04-E07 PCB
44
3)Damage of the M600-04-E00 PCB
4)Parameters of the“D/A Ref led” and “Min.value.cuv” not accurate
Solutions:
1) Enter service software; test the value of “Ref led on” from the background.
Normally, the value should be >14W and <23W, if the value has a little difference,
adjust the value of the “D/A Ref led” to make it in the normal range.
2) If the value of the “Ref led on” in the background is near the value of
“background/closed”, and while increasing the “D/A Ref led” have no effect to
increase the value of “Ref led on”. That means the blue LED is not on. Check
whether the red LED on the 04-E07 PCB is on, if the red LED is on, change the
04-E07 PCB, if it’s off, then change the M600-04-E00 PCB.
45