You are on page 1of 19

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles

for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Designation: D8185 − 18

Standard Guide for


In-Service Lubricant Viscosity Measurement1
This standard is issued under the fixed designation D8185; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1. Scope D445 Test Method for Kinematic Viscosity of Transparent


1.1 Significance and Determination of Viscosity—The pur- and Opaque Liquids (and Calculation of Dynamic Viscos-
pose of this guide is to provide sufficient knowledge for a ity)
person with some technical background in lubrication or D446 Specifications and Operating Instructions for Glass
condition monitoring from which they can determine the best Capillary Kinematic Viscometers
choice for measuring viscosity of an in-service oil. Such D2983 Test Method for Low-Temperature Viscosity of Au-
information from this guide should enable the user to engage in tomatic Transmission Fluids, Hydraulic Fluids, and Lubri-
productive discussions with colleagues, service providers, cants using a Rotational Viscometer
managers, and service personnel about obtaining and using D4057 Practice for Manual Sampling of Petroleum and
information on and from viscosity. There are a number of Petroleum Products
different approaches to viscometric measurement, and this D4378 Practice for In-Service Monitoring of Mineral Tur-
guide is intended to be a helpful resource in selecting the most bine Oils for Steam, Gas, and Combined Cycle Turbines
appropriate viscometric approach to gain information for the D4683 Test Method for Measuring Viscosity of New and
in-service fluid. Used Engine Oils at High Shear Rate and High Tempera-
ture by Tapered Bearing Simulator Viscometer at 150 °C
1.2 The values stated in either SI units or inch-pound units D5133 Test Method for Low Temperature, Low Shear Rate,
are to be regarded separately as standard. The values stated in
Viscosity/Temperature Dependence of Lubricating Oils
each system may not be exact equivalents; therefore, each
Using a Temperature-Scanning Technique
system shall be used independently of the other. Combining
D5293 Test Method for Apparent Viscosity of Engine Oils
values from the two systems may result in non-conformance
and Base Stocks Between –10 °C and –35 °C Using
with the standard.
Cold-Cranking Simulator
1.3 This standard does not purport to address all of the D5478 Test Methods for Viscosity of Materials by a Falling
safety concerns, if any, associated with its use. It is the Needle Viscometer
responsibility of the user of this standard to establish appro- D6224 Practice for In-Service Monitoring of Lubricating Oil
priate safety, health, and environmental practices and deter- for Auxiliary Power Plant Equipment
mine the applicability of regulatory limitations prior to use. D6299 Practice for Applying Statistical Quality Assurance
1.4 This international standard was developed in accor- and Control Charting Techniques to Evaluate Analytical
dance with internationally recognized principles on standard- Measurement System Performance
ization established in the Decision on Principles for the D6304 Test Method for Determination of Water in Petro-
Development of International Standards, Guides and Recom- leum Products, Lubricating Oils, and Additives by Cou-
mendations issued by the World Trade Organization Technical lometric Karl Fischer Titration
Barriers to Trade (TBT) Committee. D6616 Test Method for Measuring Viscosity at High Shear
Rate by Tapered Bearing Simulator Viscometer at 100 °C
2. Referenced Documents
D6896 Test Method for Determination of Yield Stress and
2.1 ASTM Standards:2 Apparent Viscosity of Used Engine Oils at Low Tempera-
ture
D7042 Test Method for Dynamic Viscosity and Density of
1
This guide is under the jurisdiction of ASTM Committee D02 on Petroleum Liquids by Stabinger Viscometer (and the Calculation of
Products, Liquid Fuels, and Lubricants and is the direct responsibility of Subcom-
mittee D02.96 on In-Service Lubricant Testing and Condition Monitoring Services. Kinematic Viscosity)
Current edition approved April 1, 2018. Published June 2018. DOI: 10.1520/ D7110 Test Method for Determining the Viscosity-
D8185-18.
2
Temperature Relationship of Used and Soot-Containing
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Engine Oils at Low Temperatures
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on D7279 Test Method for Kinematic Viscosity of Transparent
the ASTM website. and Opaque Liquids by Automated Houillon Viscometer

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States

1
D8185 − 18
D7483 Test Method for Determination of Dynamic Viscosity flow in the process of lubrication and hydraulic service. All of
and Derived Kinematic Viscosity of Liquids by Oscillat- these patterns are expressed by Eq 1 if the fluid is Newtonian
ing Piston Viscometer in behavior. However, many fluids and lubricants have flow
D8092 Test Method for Field Determination of Kinematic patterns called “non-Newtonian” and these important lubri-
Viscosity Using a Microchannel Viscometer cants will be discussed further on in this guide.
2.2 SAE AIR Standard:3 3.2 Two Frequently Used Forms of Viscosity Measurement
SAE AIR 5704 Field Viscosity Test for Thickened Aircraft that Produce Different Values:
Anti-icing Fluids
3.2.1 In the early years of measuring the viscosity of
3. Definitions and Terms lubricants, most viscometric measurements of fluids were done
using glass capillary viscometers, and these are still used
3.1 What is Viscosity? widely today. However, in the mid-1930s, rotational viscom-
3.1.1 Viscosity is commonly recognized as the ease or etry was commercially introduced and has since become
difficulty with which a fluid flows—that is, its fluidity. Often it widely used, particularly in high shear rate viscometry, which
is very evident that temperature has a strong effect on was introduced particularly for viscometric information on
fluidity—viscosity always increases with decreasing tempera- non-Newtonian oils.
ture and vice versa. 3.2.2 However, regarding information on Newtonian oils
3.1.2 A fluid’s viscosity arises from the degree of its internal (which are non-shear rate susceptible), capillary and rotational
molecular resistance to motion and a fluid flows only under viscometers—at the same temperature, produce different vis-
sufficient force whether that force is gravity or some other cometric values for a very simple reason. The reason is that
source. Stirring, pumping, causing fluid to flow in a pipe or gravimetric viscometry is also a function of the density of the
lubricating a machine are all examples of shear—applying a fluid that causes the fluid to flow through the capillary. Thus,
force to cause a fluid to move. For a simple example, consider the rate at which the fluid flows through the capillary is not
filling a glass of water and a separate glass full of thick used only dependent on a fluid’s viscosity but also on its density.
oil, and stir each glassful at the same velocity (shear rate) with This form of viscometry has been termed “kinematic viscos-
an identical spoon held in the same manner. The thicker used ity.” It was formerly measured and reported in units of
oil will require more force to move the spoon than the water, centiStokes, cSt. This unit became obsolete in 1976 when
which is consistent with the used oil having a higher viscosity worldwide System International, SI, unit of mm2/s was
than the water. introduced, (for information = 1.0 cSt = 1.0 mm2/s).
3.1.3 Isaac Newton defined viscosity originally as the ratio
3.2.3 True viscosity is called “dynamic viscosity” and was
of the force moving the fluid over the rate at which the fluid
formerly measured in the units of centipoise, cP. This unit
moves in response to that force. Fig. 1 helps to visualize this
became obsolete in 1976 when the worldwide System
relationship. The edges of the two plates are shown with fluid
International, SI, the unit of mPa·s was introduced, (for
between. As predicted from Newton’s law regarding viscous
information 1.0 cP = 1.0 mPa·s).
flow, when the upper plate is moved under a steady force over
the stationary bottom plate, this produces a linear shear 3.2.4 With Newtonian fluids, the two viscosity values dif-
gradient through the fluid as shown. Depending on the viscos- fered from one another as shown by Eq 2 in which ν is the
ity of the fluid between the plates, the ratio of the force per area kinematic viscosity, η is the true, dynamic viscosity and ρ is the
(technically named ‘shear-stress’ and indicated in Eq 1 by τ) fluid’s density at the temperature of measurement.
causing motion of the upper plate at the shear gradient (termed ν 5 η⁄ρ (2)
shear rate and indicated by δu/δy) is given by Newton’s law as:
Both gravimetric capillary and rotational approaches to
η 5 τ⁄δu/δy (1) measuring viscosity have remained popular and, over time,
3.1.4 Fig. 1 is, of course, a very simple example of fluid other viscometric instruments have become available.
flow to clearly show the relationship in Eq 1. However, fluid 3.2.5 However, when selecting a viscosity-measuring tech-
flow can and does take many much more complex patterns of nique for an in-service fluid or lubricant, it is critically
important to clearly understand which unit of measurement has
previously been used to characterize its viscosity—that is,
3
Available from SAE International (SAE), 400 Commonwealth Dr., Warrendale, whether it is dynamic or kinematic viscosity. Conversion by
PA 15096, http://www.sae.org. density according to Eq 2 above may be required to achieve the
desired unit of measurement. To repeat, viscosity data must be
in identical units before comparison, otherwise incorrect con-
clusions and false expectations of responses to conditions may
result in poor choice of lubricants with a waste of time, effort,
and resources or may even result in choices harmful to the
device using the fluid or those depending on it.
3.2.6 However, there is another very important factor in the
measurement of viscosity, and that is whether the fluid is
Newtonian or non-Newtonian and that subject is opened in the
FIG. 1 Depiction of Velocity Gradient in a Flowing Fluid following section.

2
D8185 − 18
3.3 Newtonian and Non-Newtonian Viscosity and Its Mea- temperatures. However, as noted above, much of this viscosity
surement: increase is subject to TVL.
3.3.1 As noted earlier, Newton’s definition of viscosity was 3.3.5 Unfortunately, there is another factor to be considered.
given in Eq 1, in which the ratio of shear stress to shear rate is At the shear rates and other, non-viscous forces applied in
constant for a fluid. Many simple fluids such as mineral oils are lubrication, the larger of these polymeric additives are vulner-
Newtonian. However, with some more complex fluids or when able to mechanical rupture. This causes a significant and
certain additives are dissolved in some Newtonian fluids, the unrecoverable loss of lubricant viscosity, a so-called “perma-
ratio of shear stress to shear rate is not constant and these fluids nent viscosity loss” or PVL, depending on how vulnerable the
are called non-Newtonian. Today, many lubricating oils are dissolved macromolecules are to the forces and conditions to
non-Newtonian in behavior, and this fact also can affect their which they are exposed during use. PVL may reach 20 % or
ability to lubricate or the energy required to overcome viscous more during use of the lubricant.
resistance to flow.
3.3.2 Non-Newtonian flow may take many forms in com- 4. Measuring Viscosity
parison to Newtonian flow, two of which are shown in Fig. 2 4.1 Some viscosity measurement techniques are performed
compared to Newtonian flow. Shear-thickening by lubricants is at specific shear rates and temperature conditions and are
rarely shown, however, shear-thinning is very commonly designed to simulate the conditions of the device being
encountered especially in lubricants. This shear-thinning is lubricated, for example, the operating conditions of the lubri-
often called “temporary viscosity loss” or TVL, since on cant serving an automotive engine. For the automotive engine,
lowering shear rate, the viscosity “lost” returns. The shear rates Test Method D5293, Test Method D4683, and Test Method
producing TVL are in the millions of units (called “reciprocal D6616 are examples of viscosity measurement techniques that
seconds,” denoted frequently by 1/s or s–1). Such high values fix shear rates, shear-stresses and temperatures to obtain the
of shear rate are the normal operating level for most lubrication resultant viscosity.
of automotive engines and other machinery. Temporary viscos-
4.2 Measurement Consistency:
ity loss may reach levels of 30 % or more depending on the
4.2.1 The precision of the measurement method should be
lubricant formulation.
well within that required for the changes that can occur with
3.3.3 The base mineral oils used to manufacture lubricants
use of the lubricant, otherwise inadequate information may
are essentially Newtonian. For many years, one of the impor-
generate needless or erroneous time, effort, and expensive
tant additives used to formulate lubricants have been so-called
actions because of inadequate information from the viscomet-
“viscosity index (VI) modifiers.” These additives are com-
ric test method chosen. Thus, before choosing a viscometric
posed of polymeric molecules which are dissolved in relatively
test, it is wise to review the repeatability, r, and the
small amounts in the mineral oil base stock to impart desirable
reproducibility, R, of viscosity measurement the particular
viscometric properties at both low and operating temperatures
viscometric method can provide.
at low shear rates.
4.2.2 As well, programs such as the ASTM Proficiency
3.3.4 Conventional wisdom has held that on dissolution, the
Testing Programs (PTP) for Petroleum Products and Lubricants
volume of these VI modifiers increases greatly and thus impart
can provide valuable insights into a particular viscometric
an accompanying large viscosity increase to the base oil. The
method, such as how well the variability in reported values by
polymer volume increases when it is dissolved and is at higher
the number of laboratories participating in the program for a
temperature, thus giving a greater increase in viscosity than at
given method, compares to the reported repeatability (r) and
lower temperature, but more recent investigations suggest that
reproducibility (R) for the viscosity measurement method.
this phenomenon may occur only with certain polymeric
Analysis of PTP data can be especially insightful for viscosity
structures, and that for different structures, there is a different
measurement methods included in the In-Service Oil Monitor-
mechanism. Using much lower viscosity base oils, this results
ing of Hydraulic Fluids/Oils PTP or the In-service Diesel
in formulations that at low shear rates have viscosities similar
Lubricating Oil Monitoring PTP. The test oils are actually
to those of non-polymer containing lubricants at operating
in-service oils and can provide insights on measurement
variability applied to such in-service oils, rather than fresh oil
samples. If variability information is not available, periodic
redundant sampling may also provide insight into measurement
method repeatability.
4.3 Specifications and Warranty Testing—Lubricants are
often specified by their viscosity level at a given temperature.
For example, in the case for engine oils, the Society of
Engineers (SAE) viscosity grade of the fresh, unused oil
defines the criteria by which they are identified—as SAE 20,
SAE 10W-40, and others. The methods used to determine an
SAE grade or lubrication specification can be applied to a used
oil to determine if the oil is still within reasonable proximity to
the recommended viscosity specification specified by the
FIG. 2 Newtonian and Two Forms of Non-Newtonian Behavior manufacturer of the equipment. Warranties may require the

3
D8185 − 18
lubricant meet the requirements of a given specification or the 4.4.7 The level of analysis at which action should be taken
warranty is subject to voiding. should also be clearly identified as part of the measurement
4.4 Establishing a Monitoring Plan for Viscosity Measure- plan, along with what actions should be taken, who should
ment: perform them, and the timeliness expected in completing them.
4.4.1 Often, a series of in-service viscosity measurements In some cases, actions may require additional measurements of
are helpful and informative when taken in a systematic way. other properties. This should be taken into consideration when
Such data can be of interest, for example in comparison to the sizing the sample vessels. The most subjective factor of the
initial fresh oil value. Or, depending on the equipment, if there sampling plan is the frequency of sampling. This will depend
is a minimum value below which the viscosity of the fluid highly on the application, the particular oil type, and take into
should not fall or a maximum above which viscosity should not account the following:
rise. Documenting a plan for an in-service series of viscosity 4.4.7.1 Cost of analyses,
measurements will likely also improve one’s skill in good 4.4.7.2 Expected sample degradation rate under normal
decision making regarding the information generated by the conditions,
series. 4.4.7.3 The likelihood of equipment failure, and
4.4.2 Information on specific measurement plans advice can 4.4.7.4 Potential cost of a failure if serious degradation is
also be found in the following ASTM methods: not detected fast enough.
4.4.2.1 Practice D6224 for monitoring of turbine oils, 4.5 Obtaining a Representative Sample:
4.4.2.2 Practice D4378 for mineral turbine oils, and 4.5.1 There are inherent limitations when performing any
4.4.2.3 Guide to ASTM Test Methods for the Analysis of type of sampling, any one of which may affect the information
Petroleum Products and Lubricants.4 obtained on the sample. For example, a small sample for
4.4.3 Measurement plans should be designed to ensure determining the viscosity of the oil in a pipeline provides a
reasonably trouble-free operation of machines and equipment. sample from only one particular location. Therefore, when a
An effective plan will often include several different pieces of representative sample cannot be obtained from a single port, a
information, starting with the technique to ensure a represen- better approach is to collect samples from multiple locations.
tative sample. The amount of sample should be sufficient for 4.5.2 It is important to clearly understand that the objectives
the planned slate of tests. If the fluid to be analyzed is well of sampling an oil or lubricant are to:
circulated in the mechanism, the sample can be taken at any 4.5.2.1 Maintain data integrity,
site containing the circulating fluid—for example, the sump of 4.5.2.2 Maintain proper frequency of sampling,
an engine. 4.5.2.3 Set proper targets and alarms for each type of
4.4.4 In some cases, lubricant viscosity may be best ana- equipment, and
lyzed under temperatures and shear rates similar to in-service 4.5.2.4 Minimize data obstructions.
conditions. Frequently, measurements are also made under 4.5.3 Moreover, another area of planning involves strategic
conditions specified for new oils (for example, 40 °C or 100 °C sampling conditions that include such matters as:
at atmospheric pressure). 4.5.3.1 Safety considerations,
4.4.5 Regardless of where the samples are obtained and 4.5.3.2 Sampling location, and
under what conditions they are analyzed, consistent sampling 4.5.3.3 Sampling tools.
location(s) are important for meaningful interpretation of the (1) Hardware.
results. If multiple sample ports are desired, the location at (2) Bottles.
which the sample is taken should be recorded with analysis (3) Sampling and procedure practices (Practice D4057).
values, together with any machine identification, and date and 4.5.4 Under all circumstances, it is necessary to follow safe
time. In some cases, it is helpful to have the identification of and well-planned sampling procedures for each type of equip-
the person collecting the sample as well. ment whether they be:
4.4.6 For samples likely to contain volatile components/ 4.5.4.1 From pressurized or non-pressurized systems.
contaminants, sample-containing vessels should be chosen 4.5.4.2 From small reservoirs (such as bearing housings).
which are readily and completely closable and thus suitable to 4.5.4.3 From a drum or reservoir without a test port.
consistently preserve the sample from changes that may occur 4.5.4.4 From otherwise difficult or dangerous environments.
between the time the sample is taken and the measurement 4.5.5 All sophisticated oil analysis tools, techniques, and
performed. They should also be sized to hold adequate sample diagnostic processes are meaningless if the oil sample fails to
for the measurement method or methods being applied. The effectively represent the actual condition of the oil in service in
measurement plan should define the range of time allowed the machine. Proper sampling procedures build the foundation
between sampling and the time a measurement should be made of an effective oil analysis program. Without good sampling
to minimize sample degradation and to provide actionable data procedures that obtain a representative quantity of the fluid or
as quickly as possible. Ideally this time should be as short as lubricant in service, both time and money are wasted, and
possible. incorrect conclusions are likely if the data are faulty.
4.5.6 Another caution with sampling is the importance of
consistency of the method and the analytical technique. It is not
4
Nadkarni, K., Guide to ASTM Test Methods for the Analysis of Petroleum a good idea to trend sample results taken using different
Products and Lubricants, 2007. sampling methods or different test methods.

4
D8185 − 18
4.6 Temperature of the Viscometric Test and Measurement: will produce erroneous test results. Obviously, poor tempera-
4.6.1 Viscosity of an in-service oil will change significantly ture control while taking a viscosity measurement will very
with temperature. Proper viscosity measurement requires the likely lead to:
instrument or technique to control the temperature of the 4.6.2.1 Inaccurate results,
sample appropriately during measurement. Some ASTM meth- 4.6.2.2 Wrong conclusions, and
ods are written for elevated temperature measurement, and 4.6.2.3 Poor reproducibility and poor consistency,
others have been specifically developed to evaluate viscosity of 4.6.3 Checking the temperature measurements devices, such
the oil at low temperatures. ASTM methods typically indicate as the thermometer or PT-100, against a reference on a
the required temperature control limits during the measure- semi-annual basis (as a minimum) will help to ensure that the
ment. For some time, 40 °C, 100 °C, and 150 °C have been the viscosity measurements being taken are accurate. Each instru-
three most common above-ambient temperatures at which ment is likely to have specific procedures for calibrating the
viscosity data are collected and compared. temperature measurement and temperature control devices.
4.6.2 When performing different test methodologies, it is ASTM methods often detail the procedures for calibrating a
important to have an understanding that time to reach the device. A laboratory practicing a given form or forms of
temperature equilibrium at which measurements will be made viscosity measurement should be able to demonstrate profi-
differs with instrument and technique. For example, the kine- ciency in the performance of the calibrations to the frequency
matic capillary viscosity test, Test Method D445 requires the required by the ASTM method or provide a traceable certificate
use of clear liquid baths which may need 30 min or more to to the organization which provides the calibration services.
become constant depending on the temperature of measure-
ment prior to performing a test run. As another example, with 4.7 Turnaround Time Requirements for Results:
the low-temperature dynamic viscosity Test Method D5133, 4.7.1 Different viscosity measurement techniques can re-
the test requires cooling at a rate of 1 °C per hour after quire different amounts of time to produce the data. Table 1
stabilizing either the liquid bath or the refrigerated metal provides insights on the elements that affect turnaround time
dry-bath at –5 °C. Not waiting sufficient time for equilibrium to for each of the viscosity measurement methods. Included in
take place will produce test results outside the expected Table 1 is a brief description of the viscometric measurement
repeatability/reproducibility stated in the respective test technique and the ASTM method number. The information in
method and operators need to be aware that such poor practice the Typical Location column indicates where the viscosity

TABLE 1 Elements Affecting Turnaround for Each Viscosity Measurement Method


Description ASTM Method Temperature Typical Location Elements that Affect Turnaround
Capillary D445 Typical 40 °C and 100 °C, Oil Analysis • If the measurement is made by an automatic unit or
Viscosity Kinematic Viscosity wide range possible; –40 °C to Laboratory manual procedure.
150 °C • Temperature of measurement as a sample must equili-
brate at the test temperature. Standard temperatures such
as 40 °C and 100 °C are more likely to be ready on an au-
tomated unit or have a bath ready at temperature for a
manual measurement.
Low Temperature D2983 –55 °C to 20 °C Oil Analysis • This method requires the sample to equilibrate to tem-
Viscosity Dynamic Viscosity Laboratory perature for 16 h before measurement.
TBS HTHS Viscosity D4683 40 °C to 200 °C Oil Analysis • Non-standard temperatures will require a calibration pro-
150 °C Dynamic Viscosity Laboratory cedure
Scanning Brookfield D5133 –5 °C to –40 °C is typical, Oil Analysis • Test method requires cooling at 1 °C per hour from –5 °C
Technique Dynamic Viscosity can be run to –70 °C Laboratory to –40 °C. Test will take at least 48 h to complete.
Cold Cranking D5293 –5 °C to –30 °C, Oil Analysis • Test takes 1 h to 2 h to complete.
Simulator Dynamic Viscosity depending on viscosity grade Laboratory
Falling Needle D5478 –40 °C to 350 °C Field, • Field measurements take 2 min to 3 min.
Viscometer Dynamic Viscosity Oil Analysis • Tests completed in the lab take 2 min to 3 min after tem-
Laboratory perature equilibration of approximately 5 min to 10 min.
TBS HTHS Viscosity D6616 100 °C, see D4683 Oil Analysis • Method takes 5 min for results when unit is calibrated and
100 °C Dynamic Viscosity Laboratory ready to test.
Yield Stress and D6896 –15 °C to –40 °C, depending Oil Analysis • Depends on W grade, and the temperature of the test.
Apparent Viscosity Dynamic Viscosity on viscosity grade Laboratory Lower temperatures can require 48 h to 72 h to complete
the test.
Dynamic Viscosity D7042 –60 °C to 135 °C Oil Analysis • Test results take 1 min to 3 min, depending on precision
Stabinger Dynamic Viscosity Laboratory mode selected.
• For samples containing ferrous materials, a magnetic par-
ticle trap can be used with to improve performance by re-
moving larger fragments.
Houillon Kinematic D7279 20 °C to 120 °C Oil Analysis • A typical test takes 2 min to 3 min to complete. Sample
Viscosity Kinematic Viscosity Laboratory must equilibrate to test temperature.
Oscillating Piston D7483 –40 °C to 190 °C Oil Analysis • A typical test takes 3 min to 25 min, depending on sample
Dynamic Viscosity Dynamic Viscosity Laboratory viscosity and temperature differential at the time the new
sample is introduced to the viscometer.
Microchannel D8092 40 °C Field • Test takes approximately 1 s per centistoke.
Viscometer Kinematic Viscosity

5
D8185 − 18
measurement is most likely to be executed if the nature of 4.8.1.3 Calibration verification of a viscometer by use of a
measurement is crucial enough to support investment in standard is measured by the instrument and the resultant value
equipment and facilities to execute the method and personnel compared to the certified value of the standard. If the two
are trained to run the test method. values agree within an acceptable tolerance as specified by the
4.7.2 Elements That Affect Turnaround column summaries ASTM Test Method, then the viscometer can be put into use. If
assume that the instrument is properly set-up, calibrated, and the two values do not agree within the acceptable tolerance,
ready for sample testing. Such time to execute the test may be then the viscometer or its manual should be checked for help or
extended if calibration and reference procedures are not the manufacturer contacted as needed. (For manual glass
included unless required by the method to be completed for capillary viscometers, see Test Method D445, subsection 9.2
each test. Time estimates obviously do not include time to ship (Certified Viscosity Reference Standards) and Annex A4 (Cal-
the samples to a laboratory nor do they include response time culation of Acceptable Tolerance Zone (Band) to Determine
of the chosen laboratory if the measurement technique is not Conformance with a Certified Reference Material).) The cer-
immediately available at or associated with the location at tified reference materials are typically manufactured for an
which the sample is obtained. intended use and the data corresponds to the appropriate
4.7.3 The rate at which data is collected must be balanced temperature and viscosity range.
with the information that is required for the information 4.8.2 Non-certified Reference Fluids:
desired. An in-service oil analysis plan that utilizes the most 4.8.2.1 Suitable reference fluids can be chosen and used in
appropriate measurement technique and allows for the results a laboratory for a daily or periodic verification of an instrument
to be turned around in the appropriate time will achieve the and these measurements can become part of a statistical quality
best results with the minimum delay in appropriate response to control (SQC) system for a laboratory. This system would track
the data’s information. Make sure that the laboratory you changes or drifts in the measurement system. If the measured
choose is qualified, ISO 9001 or 17025 certified, and experi- result for a SQC sample does not fall within a defined
enced in running the test method. tolerance, then the viscometer can be calibrated again or the
4.8 Availability of Certified Reference Fluids—Reference cause of the error found and repaired. The tolerance should be
fluids are an important tool that a laboratory must use to ensure based on the ASTM test method’s precision. Non-certified
that the viscosity measurement equipment and procedures of viscosity reference fluids are also available from accredited
the technicians are both correct and consistent. When devel- laboratories. However, many companies use their own base
oping an in-service viscosity measurement program, it may be stock or formulated engine oils as the non-certified reference
advantageous to understand the reference fluids available and material. When choosing a fluid to use, some important
how they can be used by a laboratory to ensure consistent characteristics are:
results. Either the viscometer manufacturer or the ASTM (1) The material must be homogeneous,
committee responsible for viscosity reference material pro- (2) Stable over time,
ducer can give assistance with selection of appropriate refer- (3) Free of contaminants, and
ence materials for a particular application. Reference materials (4) Of sufficient volume such that the material will be
are generally classified as certified and non-certified. Here we available for an extended time.
will describe these and how they can be used. 4.8.2.2 The viscometer manufacturer should be able to
4.8.1 Standard Viscosity Reference Fluids: assist with selection of an appropriate material for the desired
4.8.1.1 Special standardized fluids are necessary for: application. Some viscometers are made to measure shear
(1) Viscometer calibration. sensitive fluids or low-temperature sensitivity to gelation, and
(2) Verification of a viscometer’s calibration or perfor- therefore a fluid with similar non-Newtonian characteristics is
mance. preferred. For example, some reference materials are fully
(3) Or in the case of non-Newtonian behavior, to appraise formulated motor oils and can be used to verify that the applied
an instrument’s response to certain test conditions. shear rate corresponds to the method. Practice D6299 is an
available resource for this further information on this topic.
4.8.1.2 Certified standard viscosity reference materials
Both certified and non-certified viscosity reference materials
should be sourced from the instrument manufacturer, the
should be considered to ensure the accuracy and consistency of
ASTM or other group responsible for the viscometric test
the viscosity measurements of the in-service fluids.
method, or an ISO 17025 or Guide 34 accredited laboratory
capable of manufacturing the certified standard viscosity ref-
5. Fluid Rheology; the Flow Behavior of Fluids
erence material. Some methods specify qualifications of the
source for these reference materials. Calibration of various 5.1 Previous sections of this guide have presented a some-
viscometers are defined in their ASTM test methods. Calibra- what simple view of flow behavior of fluids. Understanding of
tion of automated viscometers may be defined in the corre- the property of viscosity of a fresh lubricant is not simple, and
sponding ASTM method or the manufacturer’s operating this property becomes more complex with use of the lubricant.
manual. A calibration interval for the viscometer may be Moreover, the conditions under which the lubricant is used can
suggested in its ASTM test method and should be further add further complexity to its viscous behavior. This section
determined by each laboratory. A calibration verification inter- extends the degree of understanding of how lubricant viscosity
val should also be determined to ensure accuracy of the affects and is affected by its hours of use in an effort to remove
measurement results. some of the complexity.

6
D8185 − 18
5.2 Newtonian Fluid Behavior:
5.2.1 As previously noted, the viscosity of a simple fluid
was defined initially by Newton as the ratio of the force
moving the fluid to the rate at which the fluid moves. This ratio
is apparent in a restatement of Eq 1 where Newton’s Equation
is stated as:
η 5 τ⁄Ġ (3)

where:
η = is the viscosity (most often expressed in milliPascal
seconds, mPa·s, or centipoise, cP),
τ = is the force moving the fluid (called shearstress, ex-
pressed experimentally in whatever units of force are
applied) and
Ġ = is the rate at which the fluid moves in response to the FIG. 4 Higher Shear Rate Newtonian Response
force (called shear rate and expressed in terms of
reciprocal seconds, 1/s or s–1 ).
5.2.2 As stated earlier, an important point is that a fluid Newton’s concept of viscosity shown in Eq 3 when analyzing
having Newtonian properties will, at constant temperature, the viscosity of Newtonian fluids.
have the same viscosity no matter what the shear-stress or 5.3 Non-Newtonian Fluids and Their Complex Behavior:
resultant shear rate is. That is, since Eq 3 is a linear equation, 5.3.1 Many simple fluids, such as basic mineral oils, are
a plot of Eq 3 should be a straight line with the slope equal to Newtonian. However, with some more complex fluids or when
viscosity and a zero intercept. This is shown in Fig. 3 for a certain additives are dissolved in some Newtonian fluids the
fresh Newtonian engine oil measured at three temperatures as ratio of shear stress to shear rate is not constant, and these
an example. Of further interest in Fig. 3 is the evident effect of fluids are called non-Newtonian. Today, many lubricating oils
temperature on viscosity. Not only does the viscosity of a fluid are non-Newtonian in behavior, and this fact also can affect
increase with decreasing temperature but, as suggested by the their ability to lubricate and the energy required to overcome
slopes of the engine oil at the three temperatures, the effect of their viscous resistance to flow.
the temperature is exponential. 5.3.2 Non-Newtonian flow may take many forms in com-
5.2.3 That the principles of Newton’s equation remain parison to Newtonian flow. An important expression of such
constant to the physical limits of high shear rate viscometry has non-Newtonian flow in lubricants is shown in Fig. 5 compared
been shown by studies made, for example, over a very large to Newtonian flow and is termed “Temporary Viscosity Loss”
shear rate range by the tapered bearing simulator (TBS) (TVL). Its history is interesting.
viscometer. A plot of the shear rate versus the shear-stress on 5.3.3 Since the 1940s, some of the important additives used
several Newtonian mineral oils in the TBS viscometer over a to formulate lubricants have been so-called “Viscosity Index
wide range of shear rates is shown in Fig. 4. The plot again (VI) Modifiers.” These additives are composed of very large
emphasizes the linearity and zero intercept associated with polymeric macromolecules which, in relatively small
concentrations, are dissolved in the mineral oil base stock.
These macromolecules impart desirable viscometric properties
at both low and operating temperatures at low shear rates.

FIG. 5 Example of a Non-Newtonian Oil Showing Classic Tempo-


FIG. 3 Viscometric Response of a Newtonian Lubricant rary Viscosity Loss

7
D8185 − 18
5.3.4 On dissolution, the molecular size of these VI modi- 5.3.8 Specific to the Falling Needle Viscometer—For shear
fiers increases greatly, and by doing this impart a resistance to thinning fluids, drop a needle with hemispherical ends and
flow that results in an accompanying large viscosity increase to extension bar plus external weights (different densities with the
the base oil in which they are dissolved as shown in Fig. 6. same geometry) and measure their terminal velocities by the
Using much lower viscosity base oils results in formulations amount of time taken to travel between two of the measure-
that at low shear rates have viscosities similar to those of ment lines by using a stopwatch or an automatic sensing
non-polymer-containing lubricants at operating temperatures. device. The viscosities and shear rates can be determined based
This enables the formulator to use lower viscosity base oils and on the needle velocities, the fluid and needle densities, the
still achieve the same viscosity that higher viscosity oils show needle and system geometries, and gravitational acceleration.
at operating temperatures. However, as noted above, much of
this viscosity increase is subject to TVL. This approach has 6. Care to Avoid Adverse Effects of Contaminants5
since become a well-known practice in lubricant formulations 6.1 Sampling Procedures—Sampling procedures must be
and is very common, for example, in blending so-called developed and applied with care particularly when sampling
“multi-grade engine oils” to meet the Society of Automotive from large containers or sumps of machines. This is especially
Engineers (SAE) classifications of engine oils such as SAE true when sampling from the bottom of a tank, container, or
10W-30, SAE 0W-20, and so on. Such non-Newtonian behav- sump, where heavier particulates and water—both heavier than
ior and its variation with usage of the lubricant has been the the oil—will accumulate. Viscosity of the oil sample will
subject of much dialogue and technical publications over the obviously be influenced by the water content and particulate
last several decades. content and perhaps even prevent a viscosity measurement
5.3.5 Shear-thinning is very commonly encountered espe- from being made depending on the instrument. It is certainly
cially in lubricants in which VI improvers are used. The fact important to choose a viscometric instrument and method to
that this shear-thinning is often called Temporary Viscosity avoid significant error in the effort to obtain meaningful
Loss (TVL) is because on lowering shear rate any viscosity information.
“lost” returns. Shear rates producing TVL are in the millions of
reciprocal seconds. Such high values of shear rate are the 6.2 Effect of Solids or Semi-solid Particles in the Sample:
normal operating level for most lubrication of automotive 6.2.1 Some viscosity measurement techniques, such as cap-
engines and other machinery. Temporary viscosity loss may illary viscometry, can be sensitive to the presence of particu-
reach levels of 30 % or more depending on the lubricant lates in a sample and may require the sample to be filtered prior
formulation. to analysis. Depending on the purpose of the analysis, filtration
5.3.6 Another important reason for high technical interest in prior to measurement may be an acceptable practice.
the contribution of viscosity to mineral oils by the addition of Alternatively, obtaining the viscosity of both the unfiltered and
VI improvers is particularly in regard to engine oils. filtered oil sample may provide desirable information. On the
Unfortunately, at the shear rates and other, non-viscous forces other hand, it may be of primary interest to analyze the
applied in lubrication, the larger of these polymeric additives unfiltered sample to learn about the cause or character of the
are vulnerable to mechanical rupture. This causes a significant particles.
and unrecoverable loss of lubricant viscosity—a so-called 6.2.2 Once more, it is informative to consider why moni-
“Permanent Viscosity Loss” or PVL. The amount of loss is toring oil viscosity is very important. Most lubricated, load
dependent on how vulnerable the dissolved macromolecules bearing moving surfaces are separated by a lubrication film
are to the forces and conditions to which they are exposed thickness of 10 µm or less. Many of these loaded surfaces in
during use. It has been found that PVL may reach 20 % or near-contact are prevented from contacting only by the viscos-
more during use of the lubricant. ity of the lubricant at that temperature. This viscosity-imposed
5.3.7 Another factor that can influence the viscosity/shear separation is dependent on the viscosity of the continuously
rate response is the increase of viscosity due to oxidation of the supplied lubricating oil forced into that zone of separation. If
lubricant. Usually, the effect of lubricant oxidation is shown by the lubricant cannot escape the compressive pressure fast
viscosity increase which may or may not be affected by shear enough, the surfaces cannot come into contact because the
rate. lubricant cannot be compressed beyond a certain thickness,
again dependent upon the lubricant. Especially for systems
such as engines or high speed turbochargers, maintaining the
correct oil viscosity is quite critical. Any significant decrease of
oil viscosity may permit contact of the two surfaces in relative
motion and cause damage. In high-pressure hydraulic systems,
an increase in viscosity may adversely reduce pumping effec-
tiveness.
6.3 Water Saturation of the Sample—Once the saturation
point of water for the oil is exceeded, the viscosity of the oil

FIG. 6 Distention of the Polymeric Macromolecule in Mineral Oil 5


Additional resources available from Trico Corporation, 1235 Hickory St.,
Solution Pewaukee WI 53072, http://www.tricocorp.com.

8
D8185 − 18
will be less accurate. If it is desired to determine the amount of contamination is identified. It is important to note that some-
water in the lubricant, analysis by Karl Fischer, Test Method times two factors may be acting in opposite directions, result-
D6304 is the most accurate test for water up to a concentration ing in an acceptable viscosity range, for example soot (in-
of about 1 % in the oil. Over 1 %, Fourier-transform infrared creases viscosity) and fuel dilution (reduces viscosity) for
spectroscopy (FTIR) is a more informative but less precise test. engine oil. Both factors can be measured separately, and if
See Table 2. either of them exceed acceptable limits, this may be detrimen-
6.4 Viscometric Suspension of Solids or Semi-solid Particles tal to the machine. Hence, viscosity analysis alone is not
in the Sample: sufficient, but extremely important.
6.4.1 Particulate or soot levels depend on the viscosity of 6.4.6 In conclusion, as the most important characteristic for
the oil. That is, depending on the densities of both the oil and lubricating oil, viscosity of in-service oil must be monitored as
the particulates, the particulates will either sink or rise to the a routine parameter in oil analysis. However, in order to get the
surface over time. Also, the viscosity of the oil will determine most benefit from oil analysis, other important parameters must
the rate at which particles will settle out or rise. Of course, be analyzed as well to get the holistic view of oil and
depending on the viscometric information desired, the oil may machines’ condition.
be measured with its particles well-distributed by shaking or
agitation or by removal through filtration or centrifugation. 7. In-Service Oil Viscosity Measurement Techniques
6.4.2 A brief summary of potential effects of a viscosity
Test Method D445
change in an in-service oil:
Reduction in Viscosity:
7.1 Test Method D445 for Kinematic Viscosity of Transpar-
Loss of oil film resulting in excessive wear. ent and Opaque Liquids—As previously discussed, kinematic
Increased mechanical friction causing excessive energy consumption and heat viscosity measurement of in-service lubricants provide a value
generation.
Increased sensitivity to particle contamination due to reduced oil film thickness.
that is dependent on the ratio of the actual viscosity divided by
Oil film failure at high temperatures, high loads during start-ups or coast-downs. the fluid density. The use of glass capillary kinematic viscom-
Increase in Viscosity: eters presents several issues to consider. This section covers
Excessive heat generation resulting in oil oxidation, sludge and varnish build-up.
Gaseous cavitation due to inadequate oil flow to pumps and bearings. both the manual viscometers and automated kinematic viscom-
Lubrication starvation due to inadequate oil flow to pumps and bearings. eters.
Oil whip in journal bearings. 7.1.1 Factors Affecting Kinematic Viscosity
Excess energy consumption to overcome fluid friction. Poor air release or
demulsibility. Measurements—Items discussed here include:
Poor cold start pumpability. 7.1.1.1 Sample preparation.
6.4.3 The end results of above effects are shorter oil 7.1.1.2 Samples containing water.
lifespan, shorter components lifecycle, increased oil 7.1.1.3 Samples containing particulate.
consumption, higher power consumption, and reduced ma- 7.1.1.4 Opaque samples.
chine reliability. It is extremely important for the end user to 7.1.1.5 Solvent selection.
find the root cause of oil viscosity change and to act accord- 7.1.1.6 Samples containing volatiles.
ingly at the earliest possible time to minimize the damage. 7.1.1.7 Solvent selection.
6.4.4 A brief summary of potential root causes of a 7.1.1.8 Sample coating of the kinematic capillary viscom-
viscosity change in an in-service oil: eter walls.
Reduction in Viscosity: 7.1.2 Sample Preparation—Particularly if a sample is non-
Thermal cracking of oil molecules.
Shear thinning of VI improvers (for multigrade engine oil).
homogeneous, the sample should first be thoroughly mixed,
Fuel dilution. stirred, or shaken to improve the homogeneity but assurance of
Cross mixing with lower viscosity oil. homogeneity by such means for any sample is good practice.
Increase in Viscosity:
Oxidation.
After mixing, stirring, or shaking, permit the sample to set for
Water (emulsion). a few minutes to permit any bubbles of air to rise and leave, or
Formation of carbon and oxides that create insoluble soot. else the density of the fluid needed to drive the fluid through
Antifreeze (Glycol).
Cross mixing with higher viscosity oil.
the capillary will be slightly less and result in a slightly higher
kinematic viscosity. Do not use a vacuum on the sample or heat
6.4.5 Most of the factors, except water ingression, may be
it prior to determining its kinematic viscosity, as this may
rectified by changing the oil. However, further analysis is
remove more volatile components of the fluid also resulting in
required to confirm the root cause before any actions are taken.
a somewhat higher kinematic viscosity.
Once the root cause(s) is/are confirmed, remedial action should
7.1.3 Samples Containing Water:
be taken to prevent the incident from recurring, especially if
7.1.3.1 Samples that have a high water content will likely
not give repeatable or valid results at temperatures ≥100 °C
TABLE 2 Examples of Saturation Levels for Various Oils due to boiling of the water in the sample. However, it must be
Oil Type Dissolved (ppm) Emulsified (ppm) Free (ppm) considered that any attempt to remove water by heating the
New Hydraulic Fluid 0 – 200 200 – 1000 >1000 sample to or above 100 °C before measuring its kinematic
Aged Hydraulic Fluid 0 – 600 600 – 5000 >5000 viscosity will also have an effect on removing more volatile
New R&O Oil 0 – 150 150 – 500 >500
Aged R&O Oil 0 – 500 500 – 1000 >1000 components of the sample. Samples of engine oil often have
some water and volatiles present as a consequence of their use

9
D8185 − 18
and different viscometric instrument might be considered for 7.1.6.2 For automated kinematic viscometers, verify the
in-service viscosity studies. solvents are compatible with all wetted components in the
7.1.3.2 Emulsions are encountered in lubricants and other viscometer. The manufacturer can provide information for this
mineral-oil-based fluids. A number of these emulsions contain choice.
water. Again, kinematic viscometry at 100 °C and above may 7.1.6.3 Verification that the kinematic viscometer has been
be subject to error both because of the water content and the adequately cleaned after measuring a particular type of sample
size of the emulsion particles. In such cases, it would be wise can be accomplished by measuring a certified reference fluid
to contact the manufacturer of the capillary viscometers to using two or three measurements using the kinematic
obtain guidance for kinematic viscometry. viscometer, and then verifying that the repeatability meets the
7.1.4 Filtering Particulates: Test Method D445 precision statement for it. Also, verify the
7.1.4.1 If the presence of particles is undesirable for the kinematic viscosity result agrees with the certified value.
information desired, particulate filtering may be necessary 7.1.7 Measuring Samples Containing Volatile Components:
prior to measuring kinematic viscosity in accordance with Test
7.1.7.1 Volatile types of samples can offer several issues.
Method D445. This method specifies glass capillary kinematic
For manual measurements, boiling can occur while measuring
viscometers and the capillary size would determine the filtering
at elevated temperatures, such as 100 °C and above, kinematic
requirements if one knows the diameter of the particulates.
Specification D446 specifies the capillary diameter for many of viscosity, and more volatile components of the oil can be lost.
the manual glass kinematic viscometer tubes. For the auto- Light ends can also be lost when the sample is drawn into the
mated kinematic viscometers, the manufacturer should be viscometer tube under vacuum. Moreover, the more volatile
contacted for information regarding the capillary diameter. One components can separate and air pockets or foam can develop
of the problems is if the sample is not filtered, it is possible to within the sample column. Timing in determining kinematic
pull particulate up through the capillary and it can become viscosity can be more difficult if bubbles develop at the
lodged at the capillary opening, thus changing the flow rate meniscus.
through the capillary. This scenario will cause the reported 7.1.7.2 Automated instruments may exhibit the same issues
viscosity to be higher than the actual value. as in manual testing, and the column separation and bubbling
7.1.4.2 Filtering particulates will prevent partial ‘clogging’ can be exasperated by the sample being drawn up through the
or interference with the flow of the fluid through the capillary. capillary at the test temperature.
Any particulates, unless very small in comparison to the 7.1.7.3 Measuring volatile samples at 40 °C may be an
capillary diameter, will cause kinematic viscosity values to be alternative to 100 °C measurement.
generated that, on measuring density to convert the value to 7.1.8 Sample Coating on the Viscometer Walls:
true dynamic viscosity, will be in some degree of error.
7.1.8.1 Some samples may adhere and coat the wall of the
Filtering samples containing particles before measuring their
glass capillary of the kinematic viscometer over time. The
kinematic viscosity is desirable, although it will not provide the
length of time a sample requires to coat the wall of the
accurate kinematic viscosity of the particle-containing oil
lubricating the mechanism. viscometer tube will vary depending on what is coating it,
cleaning effectiveness, and the measurement temperature.
7.1.5 Measuring Opaque Samples—Reverse-flow manual
Among these coating-forming components adhering to the
kinematic viscometers are available for measuring opaque
glass may be additives used to improve the lubrication or
samples, such as highly-oxidized or particle-containing engine
oil. Some automated viscometers use thermal detection of the engine performance. Also, soot can adhere to the glass or to a
meniscus rather than optical detection, and the opacity of the coating on the surface. Kinematic viscometer cleaning with
sample does not offer any obstacle to the sample measurement. chromic acid can be performed to remove such coatings.
Consult the manufacturer to determine the ability to measure Alternative products to chromic acid are available and may be
opaque samples. used, or sulfuric acid may be used, to clean the viscometer
7.1.6 Solvent Selection for Cleaning Kinematic Viscom- without etching the glass. Glassware detergents are also
eters: available and will work to clean the kinematic viscometer.
However, some of these are highly alkaline and could etch the
7.1.6.1 To clean the kinematic viscometer according to Test
glass wall of the capillary and thus change the viscometer
Method D445, one must choose a cleaning solvent miscible
constant. In this case, the viscometer would need to be
with the sample as well as a drying solvent miscible with the
cleaning solvent and water. Prior to selecting the cleaning calibrated again, or at a minimum a certified reference material
solvent, determine the likely contaminants that may be present measured to verify the calibration constant for the particular
in the sample. After removing the bulk of the sample by kinematic viscometer has not changed.
pouring it out or vacuum-extracting it, a suitable solvent can 7.1.8.2 Some automated kinematic viscometer customers
then be selected to adequately remove the remnant sample use an alkaline cleaner, but the cleaning solution should not be
from the viscometer. Soot, glycol, fuels, water, and metals will held in the viscometer for more than a few minutes. A certified
potentially be in a sample and need to be removed. While the reference fluid should always be used to verify that the
metals are normally flushed out with the sample; soot, glycol, viscometer’s calibration constant has not been altered by the
and water may require different solvent selections than would cleaning solution. The instrument manufacturer should be able
be needed for new lube oils. to offer further assistance with these cleaning techniques.

10
D8185 − 18
Test Method D2983
7.2 Test Method D2983 for Low-Temperature Viscosity of
Automatic Transmission Fluids, Hydraulic Fluids, and Lubri-
cants Using a Rotational Viscometer—The low-temperature,
low-shear rate dynamic viscosity behaviors of automatic trans-
mission fluids, gear oils, torque and traction fluids, and
industrial and automotive hydraulic oils at low temperatures
are of considerable importance to the proper operation or even
continued operation of a number of mechanical devices.
Measurement of dynamic viscosity for these oils and fluids at
low temperatures is often used to specify their acceptance for
service. Test Method D2983 uses a rotational viscometer with
LV torque (max torque = 0.0670 milliNewton-meter) and 4B2
spindle (equivalent dimensionally to Brookfield LV-4 spindle).
The instrument measures low shear rate viscosity of lubricants
in the viscosity range from 300 mPa·s to 900 000 mPa·s at low
temperature down to –12 °C, –40 °C, or –55 °C. The back-
ground of the test is interesting. FIG. 7 Original D2983 Rotor/Stator Assembly
7.2.1 History of Development of the Low-Temperature
Brookfield Method: cooling period, it is necessary to quickly place the sample
7.2.1.1 In the early years of the automatic transmission container of Fig. 7 into the cooled carrier to move it to the
(early 1950s), it was suspected that the low-temperature viscometer head and measure its viscosity.
pumpability properties of some automatic transmission fluids 7.2.2 Development of Alternative Test Procedures Capable
(ATF) were the cause of frequently occurring failures during of Giving Very Similar Results:
cold weather operation of the vehicle caused by insufficient 7.2.2.1 In the development and later studies of the technique
rate of clutch plate closure as a consequence of the inadequate of Test Method D2983, it was found most informative to obtain
flow rate of the cold ATF activating the closure mechanism. At the highest rotor speed at which readable torque response of the
that time there were no developed techniques for measuring viscometer could be obtained. Overall, the repeatability of this
low temperature viscometric properties. Viscosity values of low-temperature viscometric technique was acceptable,
0 °F or lower were calculated by empirical extrapolation of but—as would be expected—less precise than other forms of
viscosities obtained at 100 °F. Initial studies of viscosity at low viscometry at higher temperatures. It has been found that while
temperatures immediately showed that the extrapolation tech- some of the variation in determining viscosity was caused by
niques were in serious error by indicating much lower viscosi- sample heating during analysis, what we know as “gelation” of
ties obtained at 100 °F and 210 °F. Initial studies of viscosity at the sample plays an important role. Repeatability is improved
low temperatures immediately showed that the extrapolation using two back-to-back samples of the same material and for
techniques were in serious error by indicating much lower the first, finding the highest rotor speed giving the highest
viscosities than were truly correct. possible torque in determining the viscosity reading. The same
7.2.1.2 As a consequence, by the 1950s, a low-temperature speed is then set for the second sample, and this viscosity
viscometric bench test was developed utilizing a Brookfield recorded.
viscometer, which showed good correlation with cold-room 7.2.2.2 In an effort to improve the low-temperature Brook-
ATF tests and automatic transmission failures. In 1971, after field method by maintaining the test sample temperature, the
considerable efforts in the 1960s to encourage the use of this use of liquid baths have been explored and used. As demon-
technique and a successful ASTM round robin involving strated in Fig. 8 when the test sample is put into a liquid bath
several laboratories, this bench test became Test Method that is at the desired temperature, the cooling rate of the sample
D2983. is much more rapid than in an air bath. More rapid cooling can
7.2.1.3 The original Test Method D2983 method used a significantly affect the viscosity values. As a consequence,
glass test tube stator into which a thin metal spindle was efforts were undertaken to mimic the air bath cooling rate. One
inserted into the fluid and spun by a Brookfield viscometer. approach has been to use a liquid bath held at the final
(See Fig. 7). A wooden or plastic stopper was placed in the temperature and transfer the samples from the air bath about an
mouth of the stator to hold the rotor in proper position. Cooling hour before analysis. Another approach has been to use a more
of the test sample to the desired temperature in a refrigerated energetic refrigeration system for the liquid bath and apply a
cabinet requires a number of hours depending on the final temperature control program emulating the cooling rate of the
temperature desired and 16 h of sample cooling in the refrig- air bath.
erated cabinet is necessary for any temperature desired. 7.2.2.3 A simpler approach is to modify the stator with a
7.2.1.4 To eliminate any warming of the sample when thin gas-filled Dewar section to modulate the heat transfer from
moving the rotor-stator cell from the air bath to the viscometer, the test fluid to the liquid bath. This form of stator was called
a specially made balsa carrier is used. This carrier is also the SimAir (trademarked) Cell. The cooling curve for the
cooled in the refrigerated cabinet. After the 16 h sample SimAir stator is very similar to that of the original air bath even

11
D8185 − 18
torque reading is between 20% and 80% on the torque scale.
After an acceptable spindle speed has been attained, record the
stabilized torque reading observed between 60 s and 180 s.
7.2.4 Recommended Low-Temperature Report of Rotational
Viscometer Results:
7.2.4.1 A routine report should include:
(1) The fluid identification.
(2) The measured viscosity.
(3) The test temperature (°C).
(4) The Procedure used for temperature control
(A, B, or C).
(5) The spindle speed.
7.2.4.2 Spindle speed data are needed to ensure that differ-
ent laboratories use the same shear rates. Data where the
reference fluid values are beyond the precision of the method
shall not be reported.
7.2.4.3 In cases where this test method is used for reference
FIG. 8 Comparison of Bath Design Cooling Rates
testing between laboratories, a full report including:
(1) The identity of the Newtonian reference fluid used in
bracketing the test fluid.
though the sample is in a liquid or metal bath held at the
(2) Its reference viscosity at the temperature of measure-
desired temperature. Accordingly, the SimAir stator eliminates
ment.
the need for sample bath transfer and the need for two sample
(3) Its apparent viscosity through measurement by the
analyses. Thus, there is a choice of three different procedures
laboratory.
for achieving temperature control of the test sample prior to
(4) The spindle speed at which the test was run should
viscosity measurement: Procedure A—air bath; Procedure
accompany the test fluid data.
B—mechanically refrigerated programmable liquid bath; Pro-
cedure C—mechanically refrigerated constant temperature liq- 7.2.4.4 Over time, application of the Test Method D2983
uid bath or refrigerated metal block, using the SimAir Cell. has been extended beyond ATFs and has been applied to gear
7.2.3 A Summary of the Test Procedure: oils and hydraulic fluids. Moreover, Test Method D2983 is part
7.2.3.1 In summary, the test fluid sample is first preheated to of various international lubricant specifications. The test
either 50 °C for 30 min for lower viscosity fluids at room method can be applied to in-service oils as a routine check that
temperature and 90 °C for 30 min for higher viscosity fluids at the oils have not exceeded the low-temperature pumpability
room temperature to remove any “memory” that might affect limits of the equipment. A viscosity versus temperature curve
its low temperature responses. Afterward, it is allowed to developed for an in-service oil compared to the same curve
stabilize at room temperature and is then poured into the glass developed on the fresh oil can be used to determine if
stator cell to the proper spindle depth, keeping in mind that the equipment is at risk when operating in low-temperature con-
oil will contract somewhat at measurement temperatures. ditions. Marketers of transmission fluids and lubricants can use
7.2.3.2 The spindle is then inserted into the glass cell Test Method D2983 on in-service fluids to determine how
through the special stopper and suspended at the proper depth formulations’ low temperature viscosity changes with
by the clip. An alternative approach is to first fill the glass cell oxidation, particulate contamination, and breakdown of addi-
with the test sample to the proper depth, insert the spindle tives. OEMs may set maximum low-temperature viscosity
through the special stopper, and after suspending the spindle by requirements and periodic measurement of in-service fluids
the clip, preheat the sample to 50 °C or 90 °C in the glass cell may be recommended.
in which its viscosity will be measured. In either case, the test Test Method D4683
sample is allowed to come to room temperature after preheat-
ing. A reference fluid with known viscosity value is also 7.3 Test Method D4683 for Measuring Viscosity of New and
prepared to verify the test temperature for the purposes of Used Engine Oils at High Shear Rate and 150 °C by Tapered
greater accuracy. Bearing Simulator:
7.2.3.3 Since lubricants such as ATFs are likely to be 7.3.1 History of the Development of the Tapered Bearing
non-Newtonian at low temperatures, the spindle speed selected Simulator (TBS) Viscometer:
for measuring the viscosity of a test fluid can strongly influence 7.3.1.1 The tapered bearing simulator (TBS) viscometer
the measured viscosity. Unless the approximate viscosity is was developed to provide a precise measurement of the
known, it is important to begin at one of the viscometer’s lower rheological properties of lubricating fluids over at least a
rotational speeds and work up. Initial speeds of 0.3 r ⁄min or temperature range of 100 °C to 150 °C, and at a shear rate
0.6 r ⁄min are appropriate speeds. When measuring viscosity, range of one hundred thousand (100 000; or 1.0 × 105) to seven
the instrument torque reading should be between 20 % and 80 and one-half million (7 500 000; or 7.5 × 106) reciprocal
% of full scale on the viscometer (50 % is optimal). If the seconds (s–1). The instrument was developed with a slightly
torque value is below 20 %, increase the spindle speed until the tapered rotor/stator cell, as shown in Fig. 9, to determine the

12
D8185 − 18
collected data are extrapolated to a value of zero reciprocal
torque to obtain the rotor height at theoretical zero gap. Using
the known taper of the rotor/stator set, any shear rate can be
calculated. This technique can be further simplified by calibra-
tion using a TBS-established non-Newtonian reference oil.
Once the gap is set that provides the desired shear-rate, other
oils of unknown viscosity may be examined at the specified
shear rate. For reference, the rotor/stator gap is quite
narrow—on the order of 3.5 µm at 1.0 × 106 s–1
7.3.5 Operating Technique for the TBS Viscometer:
7.3.5.1 Operation of the TBS viscometer is achieved with a
minimum of about 3.5 mL of test oil injected if the rotor/stator
chamber is clean and dry, unless special techniques are used, in
which case less than 1 mL can be sufficient. However, at least
30 mL of sample is needed for effective use of the ‘chase-flush’
technique of completely forcing out the previous fluid with the
FIG. 9 Cutaway Diagram of a TBS Viscometer introduction of the new oil for test (a 50 mL sample is
recommended). Under standardized conditions, the viscosity of
a fluid can be measured in approximately 5 min to 7 min. A
operating properties of automotive engine oils in their journal periodically replaced 10 µm filter is placed on the injection
bearings—hence, the name. The design of the TBS instrument tube just before the test oil is injected into the rotor/stator cell.
incorporates the ability to exactly change shear rate by precise This filter is important in removing particulates in the test oil
adjustment of the rotor’s position in the stator as well as to that might otherwise cause damage to the rotor/stator interface.
exactly determine the shear rate during operation of the TBS. Such filtration also permits analysis of in-service oil samples
This ability made the TBS an absolute viscometer as will be making the TBS a capable viscosity measurement technique
discussed in a following portion. for both new and in-service viscosity oil samples.
7.3.1.2 An interlaboratory study program, sponsored by
7.3.5.2 Considering the:
ASTM Subcommittee D02.07 on Flow Properties, culminated
(1) High shear rate viscometric information carried by
in Test Method D4683, outlining the application of the TBS
in-service oils.
specifically to automotive engine oils. Test Method D4683
(2) Critical nature of engine oils.
calls for viscosity determinations in the TBS viscometer at a
(3) Lubricants that are used in other demanding applica-
shear rate of 1.0 × 106 s–1 and temperatures of 150 °C.
tions such as the power industry’s various turbine engines and
7.3.2 Society of Automotive Engineers Request to the
the critical lubricants in wind turbines.
ASTM—This ASTM program was part of a direct response to
7.3.5.3 All of these need periodic viscometric analysis at the
a request from the Society of Automotive Engineers (SAE) to
shear rates at which they serve. Such analysis is critical in
“develop a test method that incorporates high-temperature,
regard to lubricant life duration and consequences if over-
high-shear viscometric and the rheological characteristics, and
looked. It is evident from the foregoing information that the
which predicts the performance of both single and multi-grade
viscosities of non-Newtonian oils containing VI improvers
[i.e. both Newtonian and Viscosity Index (V.I.) improved]
may change radically with use.
engine oils, in engine bearings and/or ring and cylinder areas.”
7.3.3 Other Applications of High Shear Rate Viscosity—In 7.3.5.4 Similar high-shear-rate viscosity measurement can
addition to fresh automotive engine oils, the TBS viscometer be used to monitor in-service machining lubricants where there
has been used in a number of other applications including is a need to maintain an appropriate viscosity level in the
in-service automotive engine oils, hydraulic fluids, automatic presence of other fluids used as coolants or for other lubrication
transmission fluids, gear oils, printing inks, water-based fluids, needs.
and shock absorber fluids. The TBS viscometer has obtained Test Method D5133
reliable viscosity data on oils from 40 °C to 200 °C over a
range of viscosity up to 30 mPa·s. 7.4 Test Method D5133 for Low Temperature, Low Shear
7.3.4 The TBS—An Absolute Viscometer—Of importance in Rate, Viscosity/Temperature Dependence of Lubricating Oils
the high shear rate viscometry of non-Newtonian fluids, the Using Temperature Scanning Technique:
TBS is an absolute viscometer on which both the shear rate and 7.4.1 A Low-Temperature Method to Measure Undesirable
shear-stress can be measured and set while the instrument is Gelation Tendencies:
operating. The principle involved uses the TBS’s variable-gap 7.4.1.1 Low-temperature viscometry data obtained from the
identically-matched, tapered rotor and stator geometry. The Scanning Brookfield Technique has provided important infor-
spinning rotor exhibits a reactive torque response to the viscous mation regarding the effects of base oils, wax content, VI
resistance of an oil filling the gap between the rotor and stator. modifiers, pour point depressants, and other additives on the
Using a variant of Newton’s equation, the gap between the pumpability of engine oils and automatic transmission fluids
rotor and stator is established by using the reciprocal of the since the early 1980s. It was developed to understand and to
torque at several precisely measured rotor positions. The predict the negative response of engine oils that may generate

13
D8185 − 18
undesirable gelation tendencies at low temperatures (Fig. 10). Test Method D5293
The method has become Test Method D5133 for fresh oils and
7.5 Test Method D5293 for Apparent Viscosity of Engine
Test Method D7110 for used and soot-laden oils.The standard
operating range is from –5 °C to –40 °C up to 100 000 mPa·s Oils and Base Stocks Between –5 °C and –35 °C Using
but can be easily modified to ranges well outside of those ColdCranking Simulator:
limits. 7.5.1 History of a Low-Temperature Viscometric Method to
7.4.1.2 Continuous measurement of low-temperature vis- Determine Engine Oil Startability Resistance:
cosities from –5 °C to –40 °C or below is used to characterize 7.5.1.1 In the early 1950s, the problem of starting automo-
the low-temperature rheological performance of lubricating tive engines due to elevated viscosity of the engine oil at lower
oils and other fluids including jet fuels. The Scanning Brook- temperatures was finally seriously attacked by use of a viscom-
field Technique provides the most complete iscometric analysis eter developed to measure engine oil viscosity at fairly high
available on engine oil pumpability in the critical sump-to- shear-stresses. Up to that time, engine oil viscosity at low
pump screen zone and is the only technique capable of temperatures was estimated by a technique of extrapolating
generating valuable information on adverse low temperature kinematic viscosities determined at 100 °F and 210 °F. Actual
gelation tendencies with the gelation index and gelation index measurement of engine oil viscosity at low temperatures
temperatures. Automated data acquisition and analysis can be showed that such extrapolation produced much error by
carried out with Test Method D5133 as well as the tabular and seriously underestimating the viscosity of these engine oils.
graphical printout of both parameters. The viscosity/ Over a several-year period of use in several laboratories, this
temperature curve is continuously updated and displayed on
experimental viscometer was considered not as precise as
the monitor while data are written to disk for later analysis.
desired although for several years it produced comparable and
7.4.1.3 Specifically, the Scanning Brookfield Technique
meaningful viscometric data that were used to establish better
measures the low-temperature pumpability and gelation index
blending techniques for engine oils.
of fresh, sooted, and highly-oxidized engine oils by slowly and
continuously lowering the test temperature while recording the 7.5.1.2 However, in the early 1960s, a viscometer that
two values. Values of gelation index above 15 have resulted in became known as the cold-cranking simulator (CCS) was
engine failures under certain field conditions of cooling. developed and, with its improved precision, became a desirable
7.4.1.4 The automation program calculates the first deriva- test device for producing high-shear-stress, low-temperature
tive of the ASTM-Walther curve in order to provide still more viscometric data and ultimately both SAE viscosity classifica-
information about the nature of the low-temperature gelation tion values and, in 1991, Test Method D5293.
phenomenon in engine oils and ATFs, which as noted above, 7.5.2 Operation of the Cold Cranking Simulator:
was later named the gelation index. 7.5.2.1 This test method provides a viscosity value related
to the cranking and startability resistance of an engine at low
temperatures. This method is part of the SAE J300 Specifica-
tions for SAE multi-grade engine oils and is thus an important
test for both passenger car and light and heavy-duty truck
engine oils.
7.5.2.2 The low-temperature cold-cranking simulator (CCS)
utilizes a rotor/stator test cell similar to that of the TBS
previously discussed in 7.3. In this application, sample filtering
is not recommended unless particles are larger than 5 µm in
size. In such cases, a centrifuge may be used to remove
particulates and the supernatant decanted off for viscometric
analysis in the CCS.
7.5.2.3 Aerated sample measurement results may be incor-
rect. As a consequence, it is best to avoid shaking the sample
thus minimizing air entrapment. However, entrapped air/gas
bubbles can be removed by inserting the sample container into
an ultrasonic bath (with no heat) for 5 min or 10 min.
7.5.2.4 When possible, it is recommended to certify new
lube oils and measure in-service lube oils in different instru-
ments to minimize potential damage to the instrument used for
certification.
7.5.2.5 A nitrogen purge option is also recommended to
avoid moisture in the cell area.
7.5.2.6 Calibration verification should be performed rou-
tinely to monitor performance and detect any damage to the
FIG. 10 Sketch of Rotor and Stator rotor/stator cell.

14
D8185 − 18
Test Methods D5478 upon the shear rate, a needle with extension bar plus external
7.6 Test Methods D5478 for Viscosity of Fluids by a Falling weights (different densities with the same geometry) is
Needle Viscometer: dropped. The falling needle is an absolute method of viscosity
7.6.1 Viscometric Techniques—The falling needle viscom- measurement that does not need any instrument calibration.
eter (FNV) was originally developed to measure the rheologi- 7.6.3.3 There are three main models (minimum sample size:
cal properties of Newtonian and non-Newtonian fluids. 0.6 mL, viscosity and shear rate and temperature ranges: 0.5 cP
However, the FNV is a versatile instrument that can also to 108 cP, 10–4 1/s to 104 1/s, and –40 °C to 350 °C, respec-
experimentally determine other physical properties of liquids. tively) of the falling needle viscometer that are currently
In addition to measuring viscosity, it can measure liquid available: general laboratory, disposable sample tube, and
density, very low shear rate viscosity and yield stress, mixture portable field. The absolute accuracies are better than 61 %
concentrations, relative particle settling rates, intrinsic viscos- and 62 % for electronic and visual falling time measurements,
ity and molecular weight, high-pressure (up to 344 738 kPa or respectively.
50 000 psi) and high-temperature (up to 350 °C) viscosity. 7.6.3.4 Because of the simplicity of the FNV principle,
7.6.2 A Viscometer That Can Be Used in the Field—Because portable and/or disposable viscometers are made available for
of the simplicity of the FNV principle, portable and/or dispos- lubricants and greases even under in-service or field condi-
able viscometers can be made available for lubricants and tions. These types of viscometers greatly reduce testing time
greases even under field conditions. These types of viscometers because sample tube cleaning is not required. The portable
greatly reduce testing time because sample tube cleaning is not field FNV consists of a viscometer main body with a dispos-
required. able (or reusable) syringe (tube) and a falling needle attached
7.6.3 Configuration of the Falling Needle Viscometer and to an extension bar with two markings shown in Fig. 11. The
How it Works: viscosity of the sample is determined by measuring the falling
7.6.3.1 Fig. 11 shows a schematic of a laboratory FNV, time of a needle between two markings on the extension bar
which consists of an inner cylinder containing the liquid whose attached to the needle.
properties are to be measured. A constant temperature liquid is 7.6.3.5 Using the portable field FNV, the entire process of
circulated between the measuring cylinder and the coaxial obtaining in-service lubricant (engine oil) viscosities can be
outer cylinder to control the temperature of the test liquid. completed within only a few minutes. This includes sampling,
7.6.3.2 A slender hollow cylinder (needle) with hemispheri- testing, sample unloadingand cleaning. Also, it conforms to
cal ends and a weighted tip or a needle with hemispherical ends SAE AIR 5704. It is simple to use, accurate, low-cost, and an
and extension bar plus external weights falls through the test alternative to larger full-featured laboratory viscometers. Com-
liquid with the longitudinal axes of the needle and the pact and lightweight, the portable field viscometer can be taken
measuring cylinder parallel to the gravity vector. The needle anywhere to check the viscosity of even non-Newtonian fluids.
quickly reaches its terminal velocity, which is then measured
either visually (by a stopwatch) or electronically (by Hall Test Method D6616
sensors). Measurements of the terminal velocity, the fluid and 7.7 Test Method D6616 for Measuring Viscosity at High
needle densities, and the needle and system geometries are Shear Rate by Tapered Bearing Simulator Viscometer at
sufficient to determine the viscous properties of Newtonian 100 °C:
fluids. For non-Newtonian fluids whose viscosity depends 7.7.1 Application of the Tapered Bearing Simulator (subsec-
tion 7.3) at 100 °C—Test Method D6616 is conducted using
the same tapered bearing simulator (TBS) viscometer as Test
Method D4683, which was covered in 7.3. The main difference
is essentially the temperature of the measurements and any
reference oils used to more quickly set the rotor position
relative to the stator for a desired shear rate. For a detailed
description of the tapered bearing simulator viscometer, see
7.3.
7.7.2 History of Lower-Temperature Operation of the TBS
Viscometer—When the TBS viscometer was developed, a
temperature of 150 °C was desired to meet motor torque
limitations. With equipment power advancements, measure-
ment of high shear rate viscosity at 100 °C became desired and
Test Method D6616 was developed, a round-robin was run and
published. Several sectors in the lubricants industry are now
using the HTHS viscosity at lower temperatures. Specifications
are also beginning to adopt the 100 °C test. The industry has
used the approximate value of kinematic viscosity information
at 100 °C for many years. Having a true dynamic viscosity at
100 °C provided by the TBS has been perceived as more
FIG. 11 Schematic of the Falling Needle Viscometer desirable and useful.

15
D8185 − 18
7.7.3 Importance of Lower-Temperature High Shear Rate Test Method D7042
Viscosity to Diesel Engines: 7.9 Test Method D7042 for Dynamic Viscosity and Density
7.7.3.1 For condition monitoring of diesel engines, TBS of Liquids by Stabinger Viscosity (and the Calculation of
high shear rate viscosity at 100 °C is especially useful. Diesel Kinematic Viscosity):
engines are known to run cooler than gasoline engines and 7.9.1 Description of Test Method D7042 and Instrument:
having a lower temperature measurement point is important. In 7.9.1.1 This test method measures both dynamic viscosity
Europe, diesel engines are much more popular in the passenger and the density of in-service oils which can be opaque. This
car segment, and TBS viscosity at 100 °C can be used to serve test method is applicable to petroleum liquids which primarily
multiple segments. exhibit Newtonian flow behavior. Many petroleum products
7.7.3.2 Any laboratory that runs the Test Method D4683 such as in-service oils are needed for the correct operation of
method at 150 °C should also be able to run the Test Method the equipment, which depends upon the appropriate viscosity
D6616 method at 100 °C with information and guidance from of the oil being used throughout the life cycle of the equipment.
the manufacturer of the TBS. 7.9.1.2 As shown in Fig. 12, the viscosity cell consists of a
pair of rotating concentric cylinders and an oscillating U-tube.
Test Method D6896
The dynamic viscosity is determined from the equilibrium
7.8 Test Method D6896 for Determination of Yield Stress rotational speed of the inner cylinder under the influence
and Apparent Viscosity of Used Engine Oils at Low Tempera- in-service oil’s resistance to flow and an eddy current brake for
ture: determining the dynamic viscosity measurement. The density
7.8.1 History of the Mini-Rotary Viscometer—In the 1970s, is determined by the oscillation frequency of the U-tube. The
considerable work on low-temperature engine startability also kinematic viscosity is calculated by dividing the dynamic
revealed that at low temperatures, depending on cooling rate, viscosity by the density.
some engine oils could exhibit a tendency to form a waxy 7.9.1.3 To perform a measurement, an in-service oil sample
structure, a phenomenon that later was called ‘gelation.’ As a is injected into the viscosity and density measuring cells by
consequence, the Mini-Rotary Viscometer was developed to be means of syringe or auto sampler. If necessary, the Stabinger
sensitive to such formation. However, shortly after its intro- viscometer can utilize a magnetic particle trap to screen iron
duction in 1979 and inclusion in the SAE Viscosity Classifi- filings from the sample. Test results can be achieved in 1 min
cation System, during the winter of 1979-80, a catastrophic set to 3 min per sample, depending on the level of precision with
of field failures occurred in engines that had been freshly filled which the viscometer is set to perform a measurement.
with a popular engine oil. This brought in another low 7.9.1.4 Verification of the viscosity and density measuring
temperature test (see 7.5) and results from both methods are cells should be performed monthly while the calibration the
now part of the SAE Viscosity Classification System. density and viscosity measuring cells should be checked
7.8.2 Mini-Rotary Viscometer Method: annually. This should be performed to monitor performance
7.8.2.1 As mentioned, this viscometric technique became and detect any damage to the measuring cell.
Test Method D6896 in the year 2003. The sample is heated to
Test Method D7279
remove any crystalline structure present, and then cooled at a
controlled rate. A chosen, fixed force is applied to the rotor and 7.10 Test Method D7279 for Kinematic Viscosity of Trans-
the motion of the rotor observed to determine if the rotor will parent and Opaque Liquids by Automated Houillon Viscom-
rotate freely under that force or if a further yield force is eter:
required to begin the rotation. Once rotating, the speed of 7.10.1 Description of Houillon Viscometer—This test
rotation is used to determine the low temperature viscosity. The method covers the measurement of the kinematic viscosity of
force required to begin rotation is also recorded. transparent and opaque liquids by automated viscometry, using
7.8.2.2 Filtering is not specified for this test method. The Houillon viscometers (see Fig. 13). These viscometers were
rotational speed at which the rotor moves is quite slow and thus designed as a high-speed, low-volume alternative to other
damage due to contaminants unlikely. However, this instru- commonly used viscometer types. The sample volume is very
ment utilizes a rotor/stator that could be damaged if larger low (typically less than 1 mL) and as a consequence, the entire
metal particulates are contained with the sample specimen.
Caution should be used to avoid testing samples with large
metallic particulates.
7.8.2.3 Sample preparation is defined within the method.
The sample must be preheated, shaken and then allowed to
settle for 10 min.
7.8.2.4 It is recommended to isolate the test cell by using the
supplied cell caps to avoid condensation and ice crystals at the
top of the sample. A nitrogen purge option is also recom-
mended to avoid moisture in the cell area.
7.8.2.5 Calibration verification should be performed rou-
tinely to monitor performance and detect any damage to the
rotor/stator cell. FIG. 12 Diagram of Stabinger Viscometer

16
D8185 − 18
7.10.2.2 The sample cycle starts with injecting an exact
volume of the sample by micro-pipette into the viscometer
tube. The sample travels down towards the detection sensors
and heats up to the test temperature during flow. Upon reaching
the first sensor, the time measurement by stopwatch is started.
When the meniscus reaches the second sensor, the measure-
ment of flow time is stopped. The time interval thus measured
allows the calculation of the kinematic viscosity using a
viscometer tube constant determined earlier by calibration with
certified kinematic viscosity reference standards. On some
instruments, the results are either displayed on-screen or
through the PC software.
7.10.2.3 After an analysis, a predefined cleaning/vacuum
cycle automatically commences, cleaning and drying the vis-
cometer completely. Some units use a second solvent for
drying the tube. This operation is repeated several times until
FIG. 13 Houillon Tube Schematic Diagram the viscometer tube is clean.
Test Method D7483

measurement cycle is very short, allowing high sample 7.11 Test Method D7483 for Determination of Dynamic
throughputs. These two significant features privilege this Viscosity and Derived Kinematic Viscosity of Liquids by
method in testing kinematic viscosity of in-service oil samples. Oscillating Piston Viscometer:
Additionally, this test method is used for analysis of base oils, 7.11.1 Description of Oscillating Piston Viscometer—This
formulated oils, diesel oil, biodiesel, biodiesel blends, residual test method allows viscosity measurement of a broad range of
fuel oils and marine fuels. materials including transparent, translucent and opaque liquids.
7.10.1.1 The small sample and cleaning volumes used are The measurement principle and stainless steel construction
helpful where quick results are necessary, for example, in a makes the oscillating piston viscometer resistant to damage and
continuous feedback process for a production line or where a suitable for portable and in-line operations. The measurement
large number of viscosity tests are performed on a daily basis itself is automatic. The electromagnetically driven piston
to in-service oils. Houillon viscometry uses much smaller mixes the ample while under test. The instrument requires a
quantities of both sample and solvents and thus, reduces minimum sample volume of a few milliliters and typical
measurement time for cleaning and sample warm-up. solvent volume of less than 10 mL which minimizes cleanup
7.10.2 Description of Houillon Viscometer Method: effort and waste.
7.10.2.1 The Houillon kinematic viscometer uses a thermo- 7.11.2 Description of Test Method—A specimen of sample
statically controlled liquid bath equipped with heater and is introduced into the thermally controlled measurement cham-
stirrer. The kinematic viscosity is determined by measuring the ber (See Fig. 14) where the piston resides. The piston is driven
time required for a sample to fill a calibrated volume at the bath into oscillatory motion within the measurement chamber by a
temperature. controlled magnetic field. Once the sample is at the test

FIG. 14 Diagram of the Oscillating Piston Viscometer

17
D8185 − 18
temperature, as determined by the temperature detector, the measurement cycle must be easy and quick. Further, issues
piston is propelled repeatedly through the liquid (by the with highly sooty samples are addressed as the plates may be
magnetic field). A shear-stress ranging from 5 Pa to 750 Pa is directly wiped, and the short “capillary” of only approximately
imposed on the liquid under test due to the piston travel. The 5 in. means that there is a minimum amount of surface area for
dynamic viscosity is determined by measuring the average the soot to attach to and potentially alter the measurement.
travel time of the piston. The kinematic viscosity is derived by 7.12.2 Description of Test Method—60 µL of the sample to
additionally measuring the ratio between the up and down test are gathered into a disposable pipette. With the hinged
travel times. This information is then applied to an internal plates closed, the sample is deposited into the area in Fig. 15
calibration curve output of the dynamic viscosity or kinematic marked “Liquid Loading.” As it travels down the channel, its
viscosity of the liquid under test. velocity will be marked by the LED/photodiode pairs # 1,2,3 as
the sample passes by each. When the sample passes LED/
Test Method D8092 photodiode pair #3 the measurement is signaled as complete.
7.12 Test Method D8092 for Field Determination of Kine- The viscosity is calculated by using the time it takes the liquid
matic Viscosity Using a Microchannel Viscometer: to travel between pair #1 and #3. Typical flow times are
7.12.1 Description of Microchannel Viscometer—The speci- approximately 1 s/cSt. Linear flow is verified by comparison to
men in this test method flows by means of gravity between two pair #2. The user is then prompted to open the viscometer and
parallel plates held at 40 °C in a channel area approximately wipe the channel clean. As the user prepares to run the next
2 mm wide. The parallel plates are spaced 100 µm apart. This sample, a drop of the new liquid is placed on each channel and
is illustrated in Fig. 15. Note that the flow is unbounded, in the wiped clean as a pre-wetting phase. The plates are then closed
sense that the sample is open to air on two sides as it is only and the next sample is run. A calibration check should be
constrained by the two plates. The fact that it is unbounded performed once per day on the device or when a new user is
helps when the sample contains a significant portion of gaseous operating the device. A full calibration is recommended once
material, as the gas has a path to escape from the channel. The per year, or when repeated calibration checks fail.
flow time down channel area is noted by three pairs of optical
sensors. This flow time is linearly related to the kinematic 8. Closing Summary
viscosity of the sample. The advantage of this method is that 8.1 This guide has been written and reviewed by a task force
these parallel plates are hinged, so when the measurement is assigned by the D.02.96.01 sub- group on Emerging Viscosity
complete the plates may simply be opened and wiped directly. Measurement Technologies.
In this way, the “capillary” may be opened, cleaned, and
reassembled for the next measurement. Thus, the viscometer 8.2 The purpose of this guide is to provide sufficient
requires no liquid cleaning flush cycle. Only a clean cloth is knowledge for a person with a technical background in
needed to wipe off the sample. The plates themselves are lubrication or condition monitoring by which they can deter-
constructed from aluminum so that the sample rapidly in- mine the options for testing viscosity of an in-service oil. Such
creases temperature to 40 °C (within milliseconds) without the information from this guide should enable the user to engage in
need for a temperature bath. The viscometer is designed for one productive discussions with colleagues, service providers,
sample at a time field use for in-service lubricants where there managers, and service personnel about obtaining and using
are challenges with using liquid cleaning materials and the viscosity data on in-service oils. There are many different
approaches to viscosity measurement and this guide is intended
to be a helpful resource in selecting the right viscosity
measurement approach for the in-service applications.
8.3 The introductory Section 1 discusses the meaning and
importance of lubricant and fluid viscosity in the modern
world. Essentially it is one of the most critical properties of
lubricants of all forms in ensuring to some important degree
that the mechanical devices of this civilization will give
adequate functionality.
8.4 In the third and fourth sections, the development of
viscometric understanding and definition of the terms related to
viscosity, such as Newtonian and non-Newtonian fluids, kine-
matic viscosity, dynamic viscosity, and shear rate and shear
stress are presented. In relatively few pages, the reader can
obtain a basic understanding of viscometric principles and their
application. Certainly, there are many technical papers and
publications covering the principle of viscosity in greater
depth, and the reader is encouraged to explore the Related
Material section of this guide for information and assistance.
8.5 Section 4 covers the topics associated and critical to the
FIG. 15 Diagram of the Microchannel Viscometer measurement of viscosity—the factors to consider for any of

18
D8185 − 18
the measurement methods. The experience of the authors principles of these instruments and techniques. The reader can
provides the reader guidance on best practices to obtain understand the basic principles of each measurement approach
reliable, repeatable viscometric data, and insights on how to and match the approaches to the need for information. No
use that data to gain understanding and maintenance of good specific approach is recommended. Cost information has been
lubrication practices. excluded due to the complexity in determining a cost versus
value created by having reliable and meaningful viscosity
8.6 Section 5 goes into greater depth on the presence and
information to solve lubrication problems.
causes of distinctions between Newtonian and non-Newtonian
fluids and lubricants. Some of these differences were the cause 8.9 There were a number of contributing authors and
of development of several viscometric methods that have reviewers of this In-Service Viscosity Measurement guide. It is
successfully been used to make better lubricants and fluids. worth noting the authors include instrument manufacturers
who build viscometers, laboratories who measure viscosity as
8.7 Section 6 presents the cautions to be exercised in
a service, consultants who advise clients on technical matters
dealing with various contaminants affecting the viscosity of
related to viscosity, equipment manufacturers who must under-
lubricants and other fluids. The contaminants—particularly
stand how viscosity of lubricants affects the service life of their
when in particulate form—can erode the precision of a
products, and lubricant formulators who use viscometric infor-
viscometer or actively damage it.
mation to optimize formulations to intended applications. The
8.8 Section 7 contains descriptions of a number of today’s task force team was carefully selected to provide the reader
instrumental methods to measure the viscosity of an In-service with a balanced perspective and useful information on In-
oil. This guide presents information on the basic operating service lubricant viscosity measurement.

RELATED MATERIAL

Toms, L. A, Toms, A. M., “Machinery Oil Analysis,” 2008, pp. 103–108. Oils at Engine Shear Rates—Application of the Automatic TBS
Selby, T., “Problems in Bench Test Prediction of Engine Oil Performance Viscometer,” SAE International Powertrain Fuels and Lubricants
at Low Temperature,” SAE International Fuels and Lubricant Meeting Congress, SAE 2002-01-2799, 2002.
and Exposition, SAE Paper #922287, 1992. Park, N. A. and Irvine, T. F. Jr., “Measurement of rheological fluid
Selby, T., “Instrumental Approach and Improvement of the Low- properties with the Falling Needle Viscometer”, Rev. Sci. Instrum., Vol
Temperature Determination of ATF Fluidity,” 63rd Annual STLE 59, No. 9, 1988, pp. 2051–2058.
Meeting & Exhibition, Cleveland, OH, 2008. Park, N. A. and Irvine, T. F. Jr., “The Falling Needle Viscometer—A new
Selby, T., “Instrumental Approach and Improvement of the Low- technique for viscosity measurements,” Warme und Stoffubertragung,
Temperature Determination of ATF Fluidity,” 63rd Annual STLE Vol 18, 1984, pp. 201–206.
Meeting & Exhibition, Cleveland, OH, 2008. Park, N. A. and Park, H. L., “Rheological Property Measurements of
Selby, T., Piasecki, D. A., “The Tapered Bearing Simulator—An Absolute Newtonian and Non-Newtonian Fluids with the Falling Needle
Viscometer,” SAE International Congress and Exposition, Detroit, Viscometer,” STP 1564: Viscosity & Rheology of In-Service Fluids as
Michigan, SAE Paper #830031, 1983. They Pertain to Condition Monitoring, edited by A. Rishell and T.
Selby, T., Hite, R. J., Miiller, G. C., “Viscometry of New and Used Engine Selby, ASTM International, Pennsylvania, 2013, pp. 88–100.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

19

You might also like