This document outlines the standard operating procedure for operating a high performance liquid chromatography (HPLC) instrument. It details the responsibilities of technical staff, instrument details, safety precautions, and step-by-step instructions for degassing solvents, starting up and shutting down the instrument, creating and running methods, and analyzing samples. The objective is to ensure proper operation of the HPLC to separate and quantify organic compounds in pharmaceuticals and biological fluids.
This document outlines the standard operating procedure for operating a high performance liquid chromatography (HPLC) instrument. It details the responsibilities of technical staff, instrument details, safety precautions, and step-by-step instructions for degassing solvents, starting up and shutting down the instrument, creating and running methods, and analyzing samples. The objective is to ensure proper operation of the HPLC to separate and quantify organic compounds in pharmaceuticals and biological fluids.
This document outlines the standard operating procedure for operating a high performance liquid chromatography (HPLC) instrument. It details the responsibilities of technical staff, instrument details, safety precautions, and step-by-step instructions for degassing solvents, starting up and shutting down the instrument, creating and running methods, and analyzing samples. The objective is to ensure proper operation of the HPLC to separate and quantify organic compounds in pharmaceuticals and biological fluids.
1. Objective: To lay down the procedure for the operation of HPLC.
2. Responsibility: It is the responsibility of the Technical Assistant, Technical Manager, Quality Manager, and all those individuals who are working on the instrument. 3. Instrument Details: Make- .................., Model- ....................................
4. Procedure: Enter in the log book.
4.1. Purpose: The following document describes the procedure for the Operation of the HPLC. 4.2. Scope: This instrument is used to separate and quantify organic compounds in pharmaceutical formulations and biological fluids. 5. Precautions: · Always use a guard column to avoid clogging of HPLC analytical column. · Filter the mobile phase with 0.45µm filter paper. 6. Procedure: 6.1. Degassing Solvents · Ensure there are enough degassed solvents before running a sample. · Get a 500 mL or 1000 mL Erlenmeyer filter flask. · Sonicate the solvent/Mobile phase for 5-10 mins · DO NOT HEAT. · The solvent is degassed when bubbles no longer appear 6.2. Starting the Instrument · Turning on the Instrument · Turn on the Column oven · Turn on the Liquid Chromatographs · Turn on the UV Detector · Starting the Computer · If the computer is off, log in to the system · Open the program by Clicking on the HPLC Icon. · Click OK (No password) · Click Empower LC Icon 6.3.Creating a Method · Click the Method drop-down menu · Click LC Set-Up · Set up the instrument mode · Set the run Time of your method · Define the flow rate · Percent Concentration Adjusts the graph limits. · Click parameters, Adjust flow rate, and Adjust the pressure. · Click OK. · Click the File drop-down menu Click Exit 6.4.Saving the Method · Click the file drop-down menu · Click save as · Enter desired name of the method · Click Save 6.5.Aquisition Set-Up · Click the set-up icon · Select Sample Frequency · Ensure your frequency is 1 unit greater than your sample requires · Enter Run Time · Default is 5 minutes · Save the method again 6.6.Running the sample · Select the run single icon (Which has a blue arrow pointing to the right) · Enter sample identification name · Select the desired method (Select from browse button) · Select the file destination (Select from browse button) · Click start 6.7.Injecting the sample · A blank should be run first · Make sure the knob is turned to load · Insert sample using a blunt tip syringe · Quickly turn the knob up to inject the sample 7. Select analyze icon · This will display retention times and peak areas 8. Shutting down the instrument · Wash the column with 50% water and 50% methanol or acetonitrile for 10 min. · To Stop the pump change the flow rate to 0 ml/min. · Click on the pump icon. · Turn off the Column oven & Turn off the Liquid Chromatography · Turn off the UV Detector · Exit from the HPLC software. · Shut down the computer. END OF THE DOCUMENT