3 Installation (3 2)
3 Installation (3 2)
Ver3.2
INSTALLATION
EVA-HF325/525
EVA-HF325/525
Installation
TABLE OF CONTENS
1. INTRODUCTION ----------------------------------- 3
2. COMPOSITION -------------------------------------- 4
3. ELECTRIC ENVIRONMENT------------------------------ 5
4. PACKING ------------------------------------------ 7
5. DIMENSION------------------------------------------ 8
6. INSTALLATION---------------------------------------- 9
1
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
1. INTRODUCTION
2
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
2. COMPOSITION
①TUBE ②COLLIMATOR STAND ③
④TOP TABLE
⑪STAND
GUIDE
⑩CABLE
DUCT
⑤ ⑥ ⑦ ⑬ ⑧ ⑨
X-RAY CONTROLLER HV TRANSFORMER RAIL RAIL SURPPORT
①TUBE
②COLLIMATOR
③STAND
④TOP TABLE
⑤X-RAY CONTROLLER
⑥TABLE MOVEMENT S/W
⑦HV TRANSFORMER
⑧RAIL
⑨RAIL SURPPORT
⑩CABLE DUCT
⑪STAND GUIDE
⑫HANDLE BAR
⑬TABLE
3
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
3. ELECTRIC ENVIRONMENT
Table. Line Power requirement
4
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
208VAC 273 A
230VAC 245 A
240VAC 232 A
Differential Sensitivity 30 mA
Below 210VAC requires an auxiliary boost transformer to adequate the line voltage to the
generator input
Wire size at ;
Line voltage
15m 30m
240 VAC 50 ㎟ 95 ㎟
5
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
4. PACKING
5. DIMENSION
6
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
207
74
20
200
110
120
245
7
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
6. INSTALLATION
Top table
1. Put in top table to table.(first, remove rubber stopper and insert rubber stopper after insert
top table)
2.connect braker into table.
Breaker
interval : 4 - 6mm
When heavy person be on the bed, consider interval of locker and bed (about 5 mm)
And fix with volts and wind tightly.
8
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
Stand support
9
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
Tube support
Fixing tape
Fixing bolt
Weight balance
wire
A angle
B angle
A angle
B angle
Six bolts
10
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
8.fixing the tube assembly into tube support and cable connection.
C
o n
n e
c t
11
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
12
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
HV cable Connection
When you are connecting high voltage cable, be careful of its polarity.
*CAUTION: Cable must be connected deep into the hole.
Cathod(-)
Anode(+)
Terminal connection
Connect 10p cable to the terminal according to the following labels.
13
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
GND
14
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
2)If x-ray film image moved, you should fix bucky to wall
1. DC+(locker power)
2. DC-(locker power)
3. START(BUCKY2—moving motor power110VAC)
4. START(BUCKY2— moving motor power 110VAC)
15
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
16
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
17
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
2. remove locker.
grid
spring
18
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
4. check hitter
hitter
ready s/w push -- hitter hitting spring
(check 24VDC in hitter when ready s/w is pushed)
5. check bucky signal
19
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
7. bucky squency
ready s/w
ready
exposure s/w
x-ray
hitter signal
20
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
Move
first, remove rubber stopper on top-table and remove breaker and move top-table to right-
side little.(consider room size).
①rubber stopper on top-table ②breaker
6800 bearing
M5 bolts
21
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Installation
M6 bolts
interval : 4-5mm
when pushed foot s/w, check interval of locker
push foot s/w release locker
not push foot s/w attached locker
22