You are on page 1of 10

Europace (2009) 11, 850–859 REVIEW

doi:10.1093/europace/eup102

Physiotherapy and cardiac rhythm devices: a


review of the current scope of practice
Geneviève C. Digby 1, Marguerite E. Daubney 2, Jim Baggs2, Debra Campbell1,
Christopher S. Simpson 1, Damian P. Redfearn 1, F. James Brennan 1,

Downloaded from https://academic.oup.com/europace/article/11/7/850/498953 by guest on 29 November 2022


Hoshiar Abdollah 1, and Adrian Baranchuk 1*
1
Cardiac Electrophysiology and Pacing, Arrhythmia Service, Kingston General Hospital K7L 2V7, Queen’s University, Kingston, Ontario, Canada; and 2Blaser’s Physiotherapy and
Acupuncture Clinic, Kingston, Ontario, Canada

Received 15 December 2008; accepted after revision 4 April 2009; online publish-ahead-of-print 2 May 2009

Aims Several case reports have demonstrated negative interactions between various physiotherapy modalities and cardiac
rhythm devices (CRD). Fear of these potential interactions may lead to suboptimal utilization of physiotherapy treat-
ments in CRD patients. No prior review of available guidelines, or management strategies, on the interaction between
physiotherapy modalities and CRD patients has been reported. To review existing guidelines regarding the use of
physiotherapy modalities in patients with pacemakers and/or implantable cardioverter-defibrillators (ICDs). To retro-
spectively analyse CRD patient encounters at a local physiotherapy facility during a period of 2 years.
.....................................................................................................................................................................................
Methods A review of the literature regarding the potential interactions between physiotherapy modalities and CRDs was
and results performed. Next, a 2 year retrospective analysis of patient encounters at a physiotherapy facility was conducted.
In addition, seven international physiotherapy societies and four CRD manufacturers were surveyed with respect
to recommendations regarding physiotherapy treatments in device patients. The local physiotherapy facility
treated 25 patients with CRD (22 pacemaker and 3 ICD patients) for a total of 230 visits (9.2 visits/patient). Five
patients received transcutaneous electrical nerve stimulation (TENS) and all 25 were administered additional
treatment in the form of ultrasound (15), acupuncture (19), Laser (7), traction/manual therapy (12), exercise (8),
education (18), taping (5), and/or moist heat (5). No complications occurred. Meanwhile, international societies
and device manufacturers offered few specific or consistent recommendations.
.....................................................................................................................................................................................
Conclusion There are no specific international policies regarding the administration of physiotherapy modalities in CRD patients
and, thus, there are no specific guidelines to be implemented at the local level. Review of the literature and of rec-
ommendations from CRD manufacturers suggests that TENS, Diathermy, and Interferential Electrical Current
Therapy are best avoided in patients with CRDs. However, there is no consensus and it may be possible to safely
deliver these modalities in a proper setting with device and patient monitoring. Although further research is required
in this regard, active collaboration between physiotherapists and CRD clinic physicians should allow for the safe
application of most physiotherapy modalities.

these devices have been shown to prolong survival in well-defined


Introduction patient populations4 – 7 and that implant rates are increasing, it is
The interactions between cardiac rhythm devices (CRD) and phys- anticipated that an increasing number of CRD patients will be pre-
ical therapy procedures using sources of possible electromagnetic senting to physiotherapy centres requiring treatment with phy-
interference (EMI) have been reported for decades.1 – 3 Even so, siotherapy modalities.
the extent to which physical therapy procedures are safe in There is a concern that patients with CRD may be less likely to
patients with CRD, namely permanent pacemakers and implanta- receive optimal care as a result of the preconception that some
ble cardioverter-defibrillators (ICDs), is still unclear. Given that physical therapy modalities are potentially harmful to this patient

* Corresponding author. Tel: þ1 613 549 6666, Fax: þ1 613 548 1387, Email: barancha@kgh.kari.net
Published on behalf of the European Society of Cardiology. All rights reserved. & The Author 2009. For permissions please email: journals.permissions@oxfordjournals.org.
Physiotherapy and cardiac rhythm devices 851

population. Given the paucity of good data on safety, it is under- obtained, a contact by telephone was attempted until all societies
standable that physiotherapists tend to adopt a conservative had been contacted.
approach to decision making in this patient population,8 and this
often leads to a decision to forego treatment in a patient who Recommendations from cardiac rhythm
would otherwise be expected to derive benefit. In fact, it is not device manufacturers
uncommon to see a patient with a pacemaker or ICD who may Cardiac rhythm device manufacturers were contacted to obtain their
have benefited from electrical pain treatment or the use of current international recommendations regarding the administration
another physiotherapy modality—but who has been denied this of various physical therapy modalities in patients implanted with
treatment on the basis of the fact that they have a CRD.9 their devices. The Canadian representative of each company was con-
tacted by email (Medtronic, Boston Scientific, St Jude Medical and
In this paper, we review the current standards that exist at the
Sorin Group– ELA). Electronic copies of the manufacturer-specific rec-
local and international levels regarding the use of various physical
ommendations were obtained from each company.
therapy modalities in patients with pacemakers or ICDs. In order

Downloaded from https://academic.oup.com/europace/article/11/7/850/498953 by guest on 29 November 2022


to assess safety issues and to develop recommendations, it is Commonly used physiotherapy modalities
important to understand the indications for and the technical con-
and documented interactions with cardiac
siderations of each modality, as well as to review the current evi-
rhythm devices
dence regarding potential interactions of the various treatment
The following is a summary of the more frequently used physical
modalities with pacemakers or ICDs. Thus, we begin with a brief
therapy modalities and documented interactions between physical
review of the literature regarding these potential interactions.
therapy modalities and CRD.
Then, at the local level, we review the policy of care of a phy-
siotherapy centre in a city of 150 000 inhabitants (Kingston, Acupuncture
Ontario, Canada). Finally, at the international level, we investigate Acupuncture involves the insertion of needles into the skin at specific
the standards that are used throughout the world, including the points in order to relieve pain. This therapeutic modality is generally
recommendations from cardiac device manufacturers and inter- considered to have a low risk of interfering with pacemakers or
national physiotherapy societies. ICDs because of the lack of electrical current being used.10
However, case reports suggest possible adverse reactions to acupunc-
ture in patients with CRD. Apaloo11 reported a case of a patient with a
pacemaker experiencing an episode of syncope during acupuncture
Methods treatment. Notably, this patient also experienced several episodes of
syncope during electro-acupuncture treatment, which involves electri-
Current standard of practice for patients with cal stimulation of the acupuncture needles by various frequencies and
cardiac rhythm devices at a local voltages to provide a longer-lasting pain blocking effect. The author did
physiotherapy facility not report abnormal findings on pacemaker interrogation.
In order to assess the current standard of practice in a local phy- Acupuncture-related syncope may have a different aetiological origin
siotherapy centre, we conducted a retrospective analysis of all than pacemaker interference, given several reports of syncope during
patient encounters that occurred within a 2 years period. The acupuncture in patients with no cardiac devices.12
number of individual patients seen by this centre as well as the total Overall, it seems as though acupuncture may be a relatively benign
number of patient encounters was recorded. A specific analysis was treatment in CRD patients, but the extent to which adverse effects are
conducted on patients who presented with a CRD. All of these likely to occur, especially in the case of electro-acupuncture, needs to
patients were analysed by retrospective chart review in order to deter- be further investigated.
mine the nature of the treatment recommended by the clinic staff and
the clinical outcomes. Diathermy
Diathermy refers to the deep heating of tissues by either ultrasound
(US) or electrical means in order to relieve pain. Short-wave diathermy
Recommendations from international (SWD) specifically refers to the therapeutic application of radiofre-
physiotherapy societies quency electromagnetic radiation in order to achieve deep tissue
To assess the existing international policies and recommendations heating.8 Treatment may be applied as either a continuous beam of
regarding the use of physiotherapy modalities in CRD patients, we energy (CSWD) or pulses of energy (PSWD). Because of the nature
contacted the national physiotherapy societies or associations govern- of this treatment, it is potentially dangerous to use it in patients with
ing physiotherapists in the English-speaking world, namely: the CRD. The generated heat can cause burning to surrounding tissues
American Physical Therapy Association (APTA), the Australian Phy- and damage the circuitry of the pacemaker, but also, the electromag-
siotherapy Association (APA), the Canadian Physiotherapy Association netic fields generated may directly interfere with the performance of
(CPA), the Chartered Society of Physiotherapists (CSP), and the Irish the device.8
Society of Chartered Physiotherapists (ISCP). We also contacted the There is very little evidence-based literature that addresses the issue
Ontario Physiotherapy Association (OPA) as well as the International of potential interference between SWD and CRD. In fact, as pointed
Society of Education in Physiotherapy (ISEP) to obtain more local and out by Shields et al.,8 electrotherapy textbooks provide only broad
international guidelines. Societies were initially contacted by email contraindications to SWD and the literature available is not sufficiently
using addresses acquired from their websites. An inquiry was posed comprehensive to substantiate all the suggested contraindications to
as to whether the society had implemented any policy or regulation electrotherapeutic agents. The few studies that exist are largely out-
regarding the administration of various physiotherapy modalities in dated, and may no longer be applicable to modern CRD. One
patients with pacemakers or ICDs. If a response had not been review13 noted interference in pacemakers due to CSWD application
852 G.C. Digby et al.

and found that patients wearing demand pacemakers were more vul- No specific recommendations exist surrounding the safety of various
nerable to interference than asynchronous ones. Valtonen et al.14 massage therapy and traction procedures in patients with CRD. This
found that CSWD caused pacemaker output to decrease and pulse is likely a result of the lack of electrical current being used and the
rate to increase. These induced changes were dependent on applied lack of deep tissue heating involved in this form of physiotherapy—a
power, electrode-pacemaker distance, and orientation of the pace- putative mechanism for device interference has not been postulated.
maker and leads relative to the electrical field. Although the effects
of SWD on pacemakers have been investigated in the laboratory, Transcutaneous electrical nerve stimulation
very little evidence exists regarding these interactions in the human Transcutaneous electrical nerve stimulation is a commonly used treat-
body. Thus, in order to err on the side of caution, it has generally ment for the relief of acute and chronic musculoskeletal pain. It
been accepted that SWD of any form should be contraindicated in involves placing electrodes on the skin, through which TENS is
patients with both dependent and non-dependent pacemakers.8 In applied at varying frequencies, intensities, and pulse durations of stimu-
our service, dependency is defined as an absence of intrinsic signals lation in order to maximize pain relief.20 Pinski and Trohman21
when the pacemaker is programmed at 40 b.p.m. Another study15 reviewed the evidence indicating adverse interactions between TENS

Downloaded from https://academic.oup.com/europace/article/11/7/850/498953 by guest on 29 November 2022


found no disruption in eight pacemaker patients being treated with and CRD. They concluded that TENS could be used safely in patients
SWD, a finding that was believed to reflect modern developments in with modern implanted bipolar pacemakers and in patients with uni-
pacemaker shielding. It has also been suggested that treatment of polar pacemakers if sensitivity were reduced. Eriksson et al.1 reported
extremities in patients with pacemakers is permissible given the remo- inhibition by TENS in four patients with non-programmable unipolar
teness of the treatment site from the device.16 The review from pacemakers, whereas other studies found that inhibition by a TENS
Shields et al.8 compiles the contraindications as advocated by text- machine could be eliminated by making the pacing device less sensi-
books, published articles, and device manufacturers’ literature into a tive.22 However, the risk associated with reducing the sensitivity of
list to be used by physiotherapists in clinical practice. On the basis the device is that if the native signals are smaller than the EMI
of the evidence listed above, they included pacemakers among their signals, this may predispose to undersensing and competition with
list of contraindicated devices. Implantable cardioverter-defibrillators the intrinsic rhythm. Meanwhile, other authors have been unable to
are not specifically mentioned, but as the review contraindicates confirm the observation that TENS may lead to interference of
implanted metal devices as well, it is likely that ICDs would fall into pacing devices; they found no interference, inhibition, or automatic
the contraindicated category along with their pacing counterparts. In reprogramming.2
fact, even the US Food and Drug Association (FDA) has broadcasted Despite several cases demonstrating EMI between TENS and pace-
a warning against the use of diathermy in a patient with any implanted makers, fewer cases have been reported on the interactions between
metallic lead in the body, including cardiac pacemakers and implantable TENS and ICDs. One report describes a patient in whom a TENS unit
defibrillators, after two patients with implanted devices (neurostimula- created an electrical artefact that was interpreted by a transvenous
tors) died as a result of diathermy treatment.17 Even so, Shields et al.8 ICD as ventricular fibrillation, leading to spurious shock treatment.23
point out that this is a complex and poorly evidenced area, and while A safety study24 showed that neuromuscular electrical stimulation
limited (largely anecdotal) evidence is available, some of the infor- could cause EMI leading to interference in atrial and ventricular
mation presented is deductive and therefore contraindications are sensing or intermittent ventricular undersensing due to post-sense
typically based on consensus rather than evidence. blanking. Although EMI did not trigger an inappropriate ICD therapy
in this study, interference was seen when stimulation was applied
Interferential electrical current therapy close to the ICD and during leg muscle therapy, regardless of interelec-
Interferential current therapy is a treatment to aid the relief of pain and trode spacing or sensing mode. The authors suggested that not only
promotion of soft-tissue healing. It is similar to a transcutaneous elec- could EMI affect ventricular pacemaker or ICD tachycardia detection
trical nerve stimulation (TENS) unit, but provides deeper penetration algorithms, but could also trigger atrial mode switch episodes or inap-
and less patient discomfort. Case reports specifically relating to this propriate atrial therapies. Siu et al.25 published another case in which
form of treatment are not available, but deductive reasoning has led TENS therapy applied to the mid-back resulted in two consecutive
one device manufacturer (Medtronic) to advise that the electrical shocks to a patient with an ICD. The shocks were attributed to EMI
current applied to the body in this form of treatment may affect simulating a series of low-voltage-ventricular signals that were inter-
implanted devices.10 preted as ventricular fibrillation, resulting in shock delivery. Several
other cases have confirmed interference of ICD function with TENS
Light amplification stimulated emission of radiation treatment, including interference with pacemaker function in patients
Light amplification stimulated emission of radiation (Laser) therapy is with biventricular ICDs,26 as well as several cases of inappropriate
the application of red and near infra-red light to improve tissue shocks with exposure to TENS.27
healing and relieve pain. It has become a popular modality to treat More recently, a study on EMI in 30 patients with ICDs was con-
musculoskeletal and skin disorders with an athermic effect and has ducted.9 In this study, it was found that TENS produced EMI triggering
been reported to have a substantial therapeutic role in rheumatoid inappropriate detection of VT/VF. Interference was found more fre-
arthritis, post-traumatic joint disorders and myofascial pain.18 quently in systems with integrated vs. dedicated bipolar leads. Interfer-
However, little has been reported with regards to the safety of ence was also more likely to occur when TENS treatment occurred at
Laser therapy in patients with CRD. A review by Volkman et al.19 the mammilla than at the hip, and at a frequency of 80 Hz as opposed
suggests that Laser application poses no risks to the pacemaker to 2 Hz. Because of the potentially serious consequences of interfer-
patient and that its relative safety is comparable to that of US, direct ence, they concluded that TENS should not be recommended for
currents, thermotherapy, and phototherapy. patients with ICDs.
Transcutaneous electrical nerve stimulation is perhaps the best-
Manual therapy/traction studied physiotherapy modality in terms of its interactions with
Manual therapy and traction involve the mobilization of the soft tissues CRD. Nonetheless, the clinical evidence is limited and, consequently,
and joints of the body in order to relieve pain and promote healing. the medical community acts conservatively by contraindicating the
Physiotherapy and cardiac rhythm devices 853

use of this device in CRD patients. Overall, the risk of TENS interfer- occurred to discuss the relative safety of administering TENS to
ence seems to be lower when electrode placement is further away the patient which was later applied with no adverse outcome.
from the device and when a lower frequency is used. Perhaps under
such situations of controlled distance and frequency, and by sub-
selecting populations with lower risk—such as bipolar systems in non-
Recommendations from international
dependent patients—TENS may be safe enough to be used. Finally, physiotherapy societies
one could perhaps envision the delivery of TENS in a monitored The responses acquired from surveying the designated inter-
setting, with ICD therapies temporarily turned off, to facilitate the national physiotherapy societies and associations are summarized
treatment required. in Table 1. The majority of these societies had no specific rec-
ommendations regarding the use of various physiotherapy modal-
Ultrasound ities in patients with CRD. The APA as well as the International
Ultrasound therapy uses high-energy sound waves to help relieve mus- Society of Educators in Physiotherapy, based out of Australia,
culoskeletal pain. Although no evidence-based literature specifically both referred us to the Guidelines for the Clinical Use of Electro-

Downloaded from https://academic.oup.com/europace/article/11/7/850/498953 by guest on 29 November 2022


refers to the safety of US therapy in patients with CRD, the manufac- physical Agents as developed by Robertson et al.28 in 2001 for the
turers of these devices report few EMI concerns in this regard. Med- APA. This document states that the provided guidelines err on the
tronic suggests maintaining a distance of 15 cm (5.9 in.) between the conservative side and that some historically accepted contraindica-
US transducer head and the implanted device to minimize mechanical
tions and precautions are repeated due to the absence of adequate
damage to internal circuitry of pacemaker or ICD.10
research findings to exclude them. These guidelines suggest that
deep heating in the form of SWD should be contraindicated in
patients with inbuilt stimulators (such as pacemakers) within 3 m
Results of the device, whereas microwave diathermy and US therapy
(with or without heat) should be contraindicated directly over
Current standard of practice regarding the pulse generator device. The use of electrical stimulators (e.g.
patients with cardiac rhythm devices TENS) with transthoracic application or long duration direct
at a local physiotherapy facility current is also contraindicated. No precautions are listed for
Laser therapy or for superficial heating in the form of infra-red
The retrospective data collected from the local physiotherapy facil-
or moist heat.
ity analysed all patient encounters that occurred within the 2 years
period from 1 June 2006 to 1 June 2008. During this time period,
2780 patients presented to this physiotherapy centre a total of Recommendations from cardiac rhythm
31 918 times, with an average number of patient visits of 11.5 device manufacturers
(range 1–27). Of these patients, 25 had an implanted CRD and Electronic copies of recommendations regarding the use of phys-
accounted for 230 of the visits to this clinic, (average of 9.2 visits ical therapy modalities in patients with CRD were obtained from
per patient). The patients with implanted devices included four CRD manufacturers. These data are summarized in Table 2.
22 patients with pacemakers and 3 patients with ICD (single The following physical therapy modalities are reported to be
chamber 13 patients, dual chamber 10 patients, biventricular associated with a risk of affecting implanted pacemaker or ICD
2 patients). function and are advised against being administered to patients
Of the 25 device patients, five presented on two separate with CRD by the device manufacturers listed in parentheses: Inter-
occasions for treatment of different clinical problems. Thus, ferential Electrical Current Therapy (Medtronic),10 Microcurrent
there were a total of 30 occasions in which a CRD patient pre- Electrical Therapy (Medtronic),10 TENS (Medtronic, Guidant–
sented to the physiotherapy clinic for treatment. None of the Boston Scientific, St Jude Medical)10 (Hospital EMI and the
patients that presented to the physiotherapy clinic with a CRD Pacemaker/Implantable Cardioverter Defibrillator (ICD) Patient,
were denied care. In five cases (occurring in five different patients), personal communication; Diathermy Warning from User’s Manual
TENS was administered despite the presence of a CRD. Alterna- & TENS Therapy and ICD, personal communication), Diathermy
tive forms of treatment commonly believed to be associated (Guidant –Boston Scientific, St Jude Medical, Sorin Group–ELA)
with a lower risk of interference with CRD were administered (Hospital EMI and the Pacemaker/Implantable Cardioverter Defi-
to the remaining 25 cases, as well as to the five patients receiving brillator (ICD) Patient, personal communication; Diathermy
TENS therapy. This care occurred in the form of US (15 cases), Warning from User’s Manual & TENS Therapy and ICD, personal
acupuncture (19 cases), Laser (7 cases), traction or manual communication; Sorin Group, personal communication), Ultra-
therapy (12 cases), exercise (8 cases), education (18 cases), sound (Sorin Group–ELA) (Sorin Group, personal communi-
taping (5 cases), or moist heat (5 cases). The body parts being cation). The following physical therapy modalities are reported
treated in all the CRD patients included lower back (18), neck to have a low risk of interfering with CRD by the manufacturers
(8), shoulder (4), hip (4), arm/hand (2), leg/foot (5), or other (2). listed in parentheses: Acupuncture (Medtronic),10 Magnetic
With regards to the five patients who received TENS therapy, Therapy (Medtronic),10 Pulse Radiation Therapy (Medtronic),10
the body parts being treatment included lower back (1), neck Ultrasound (Medtronic, Guidant –Boston Scientific)10 (Hospital
(1), arm/hand (1), and leg/foot (2). In one of the 30 CRD patient EMI and the Pacemaker/Implantable Cardioverter Defibrillator
cases, the Arrhythmia Service at the local tertiary care centre (ICD) Patient, personal communication), as well as TENS (Sorin
was consulted regarding the care of the patient. The consult Group–ELA) (Sorin Group, personal communication).
854 G.C. Digby et al.

Table 1 Summary of international physiotherapy societies recommendations

Society Recommendation
...............................................................................................................................................................................
APTA (American Physical Therapy Association)a No comment
APA (Australian Physiotherapy Association) Refer to Robertson et al.28
† Deep heating: SWD C/I within 3 m of inbuilt pulse generator (e.g. pacemaker); MW and US C/I
over pulse generator
† Superficial heating or cold: no C/I
† EStims (electrical stimulation e.g. TENS)—C/I over stimulator
† US (no/low heat)—C/I over pulse generator
† UV/Laser—no C/I
CPA (Canadian Physiotherapy Association)b

Downloaded from https://academic.oup.com/europace/article/11/7/850/498953 by guest on 29 November 2022


No specific recommendations are implemented. A document is in preparation at the time of
writing.
OPA (Ontario Physiotherapy Association)c No specific recommendations are implemented.
International Society of Educators in Physiotherapy Refer to Robertson et al.28
(Australia)
Chartered Society of Physiotherapists (UK)d No comment
Irish Society of Chartered Physiotherapistse No comment

C/I, contraindicated; SWD, short-wave diathermy; MW, microwave diathermy; US, ultrasound.
a
800/999-APTA (2782) (personal communication, 3 July 2008).
b
information@physiotherapy.ca (personal communication, 30 June 2008).
c
physiomail@opa.on.ca (personal communication, 30 June 2008).
d
enquiries@csp.org.uk (personal communication 02 July 2008, 24 July 2008), sarah.bazin@heartofengland.nhs.uk (personal communication, 29 July 2008), þ44 (0)20 7306 6666
(personal communication, 29 July 2008).
e
k.cradock@st-vincents.ie (personal communications, 2 July 2008, 9 July 2008, and 24 July 2008).

Future directions: communication consulting the Arrhythmia Service. Once the Pre-Physiotherapy
between physiotherapists and cardiac Screening Pacemaker/Implantable Cardioverter Defibrillator (ICD)
Assessment Request Form is received in the CRDC, the information
rhythm device clinics will be reviewed by the arrhythmia physician. The second form
Clinical guidelines and recommendations regarding the use of (Appendix 2) will be completed by the arrhythmia physician outlin-
various physical therapy treatment modalities in patients with ing their recommendations. The device manufacturer will be con-
CRD are often lacking, inconsistent, and/or inexplicit. Certainly, tacted at the discretion of the arrhythmia physician. Once the
device patients being considered to receive physical therapy treat- Arrhythmia Service Response to Physiotherapy Clinic Form is com-
ment involving various sources of EMI require careful consider- pleted, it is faxed back to the Physiotherapy Clinic.
ation to ensure their safety. To avoid unnecessary complications Given the logistical difficulties that would be involved in
to the CRD patient, many factors need to be considered including attempting to arrange for the arrhythmia service and phy-
the patient’s dependence on the device, the type of device, the siotherapy clinic to meet to assess these patients together, we
device manufacturer, as well as the treatment method and location. suggest using these forms as a mean of providing an exchange
As previously recommended, when in doubt regarding diagnostic of pertinent patient information between the two treatment
or therapeutic procedures that may result in cardiac device inter- facilities. This collaborative process ensures that patients are
ference, a consultation ought to be performed to the device individually assessed, provided optimal care and are adequately
centre. A collaborative consultation process would ensure an indi- followed.
vidualized approach to treatment involving physiotherapists,
arrhythmia physicians, device manufacturers, and CRD clinic
(CRDC) staff. This individualized approach to treatment would
expedite the transfer of pertinent patient information between
Discussion and conclusion
the Physiotherapy Clinic and the Cardiac Rhythm Device team. Despite years of research, there is no comprehensive policy that
A collaborative and consultative process would identify patients addresses the extent to which physical therapy modalities are
who might truly be at risk of negative consequences with a particu- safe to be used in patients with CRD. The majority of international
lar therapy and would ensure proper follow-up for these patients. societies offer no specific recommendations regarding the use of
Such a collaborative consultative process is being developed for safety of physical therapy modalities in patients with CRD,
CRD patients at our centre based on our previous successful experi- whereas the device manufacturers themselves are not always con-
ence with device patients receiving radiation therapy.29 It has culmi- sistent with regards to their recommendations. In terms of the
nated in the creation of two forms. The first form (Appendix 1) is to local experience, it was found that there was no comprehensive
be completed and faxed to the CRDC by the Physiotherapist who is policy that allowed physiotherapists to determine whether a
Physiotherapy and cardiac rhythm devices 855

Table 2 Cardiac rhythm device manufacturers’ recommendations for use of physical therapy modalities in cardiac
rhythm device patients

Physical therapy modality Recommendations


...............................................................................................................................................................................
Medtronic10
Acupuncture—no electrical stimulus Low risk of affecting pacemaker or ICD
Interferential Electrical Current Therapy Potential pacemaker reversion or magnet rate
Potential ICD shock
Recommend magnet application for ICD and pacemakers
Magnetic Therapy Low risk when following safety precautions
Potential pacemaker reversion or magnet rate

Downloaded from https://academic.oup.com/europace/article/11/7/850/498953 by guest on 29 November 2022


Potential disable of ICD detection circuit
Maintain a 6 in. (15 cm) distance between all therapy magnets and an implanted device
MET (Microcurrent Electrical Therapy) Potential pacemaker reversion or magnet rate
Alpha-Stim 100w
Potential pacemaker inhibition
Potential ICD shock
Pulse Radiation Therapy Low risk when following safety precautions
Potential pacemaker reversion or magnet rate
Potential pacemaker inhibition
Potential ICD shock
Magnet application recommended for pacemaker and ICD. Nerve ablation can damage pacemaker and ICD
circuitry, if it is administered within 6 in. (15 cm) of the device. If the ablation is farther than 6 in. (15 cm),
there is the potential for pacemaker reversion or ICD shock.
TENS Unit (Transcutaneous Electrical Potential pacemaker reversion or magnet rate
Nerve Stimulation)
Potential pacemaker inhibition
Potential ICD shock
There is a lower risk of the pacemaker or ICD detecting TENS when used on extremities
Ultrasound diagnostics/therapeutics Low risk when following safety precautions
Maintain a 6 in. (15 cm) distance between the transducer head and the implanted device
Guidant–Boston Scientific28
Diathermy Direct current interference
Possible inhibition of pacing or asynchronous pacing
Possible device reset
Risk of inappropriate shock
Possible device damage due to heating effects
Recommendations: temporarily deactivate ICD, no direct contact, monitor patient and watch for
interactions, confirm normal device operation and programmed parameters
TENS (Trancutaneous electrical nerve Direct current interference
stimulation)
Possible inhibition of pacing or asynchronous pacing
Risk of inappropriate shock
Generally contraindicated for patients
Recommendations: place TENS electrodes as close together as possible, place TENS electrodes as far from
device/lead system as possible, test TENS settings in clinic before home use (test at maximum output and
high/low frequencies, use lowest clinically effective setting, instruct patient not to change any settings)
Ultrasound No EMI concerns
St. Jude Medical (Hospital EMI and the Pacemaker/Implantable Cardioverter Defibrillator (ICD) Patient, personal communication)
Diathermy Avoid, even if the device is programmed off
TENS Place a magnet over the device, positioned off-centre, to suspend arrhythmia detection. A magnet reversion
will not affect the bradycardia pacing function of the device; therefore, the need for asynchronous pacing
must be evaluated OR program the device to non-tachyarrhythmia configuration, e.g. ‘Defic Off’, ‘All
Functions Off’ or ‘Brady Pacing Only’.
Programming the pacer to the bipolar mode of sensing and decreasing the sensitivity level will minimize the
potential for interaction with TENS stimulation.

Continued
856 G.C. Digby et al.

Table 2 Continued

Physical therapy modality Recommendations


...............................................................................................................................................................................
Burst mode is contraindicated
Recommendations: (i) monitor the patient’s heart rate during the procedure, (ii) electrode placement should
be left to right (lateral not anterior, posterior), (iii) place electrode pads in close proximity to each other,
(iv) maintain a high frequency (.30 Hz) at all times, (v) pacemakers utilizing impedance based sensors
(like minute ventilation or stroke volume) for rate responsive pacing should be programmed to a non-rate
responsive mode to prevent increased pacing rates during TENS therapy, (vi) following the procedure,
evaluate the pacing system by determining the capture and sensing thresholds.
Sorin Group –ELA (Diathermy Warning from User’s Manual & TENS Therapy and ICD, personal communication)
TENS Will not interfere with pacemaker or ICD with proper precautions.

Downloaded from https://academic.oup.com/europace/article/11/7/850/498953 by guest on 29 November 2022


Place the TENS electrodes as close to each other as possible.
Place the TENS electrodes as far from the pulse generator/lead system as possible and monitor cardiac
activity during TENS use.
Diathermy Not recommended. If necessary, the following precautions must be used: keep the current path and ground
plate as far away from the device and the leads as possible (a minimum of 15 cm/6 in.); before the
procedure, deactivate ATP and shock therapies; during the procedure, keep the diathermy device as far as
possible from the cardiac defibrillator, set it at minimum intensity, and use it briefly; after the procedure,
check for proper implant function. The device should never be exposed directly to the diathermy source.
Therapeutic Ultrasound Not recommended

CRD patient would be best treated with regular care or by alterna- patients with bipolar systems, may allow for these modalities to
tive means, although communication between physiotherapists and be safely used in CRD patients. Meanwhile, acupuncture, Laser
arrhythmia services regarding appropriate treatment was found to therapy, manual therapy/traction, and US therapy appear to pose
be practically non-existent. no significant risk to patients with CRD and should be considered
Review of the literature and of the recommendations from CRD as safe treatment in this patient population. Active collaboration
manufacturers lead us to believe that, at the moment, the physical between physiotherapy clinics and Cardiac Rhythm Device Clinic
therapy modalities of TENS, Diathermy, and Interferential Electri- physicians and staff should allow the safe application of most phy-
cal Current Therapy are best avoided in patients with CRD. siotherapy modalities.
However, there is no consensus, and, furthermore, it may be poss-
ible to safely deliver these modalities in a proper setting with Conflict of interest: C.S.S., D.P.R. and A.B. have received honor-
device and patient monitoring, and with specific arrhythmia phys- arium to deliver talks and conferences from Medtronic, St Jude
ician supervision. Further research regarding distance of electrode Medical, Boston Scientific and ELA SORIN group. A.B. has received
placement from CRD, frequencies of stimulation, and sub-selecting unrestricted grants from Boston Scientific and Medtronic. D.P.R.
populations of patients with lower risk, such as non-dependent has received unrestricted grants from St Jude Medical.
Downloaded from https://academic.oup.com/europace/article/11/7/850/498953 by guest on 29 November 2022
857
Physiotherapy and cardiac rhythm devices

Appendix 1
Downloaded from https://academic.oup.com/europace/article/11/7/850/498953 by guest on 29 November 2022
G.C. Digby et al.

Appendix 2
858
Physiotherapy and cardiac rhythm devices 859

References 16. Delpizzo V, Joyner K. On the safe use of microwave and shortwave diathermy
units. Aust J Physiother 1987;33:152 –62.
1. Eriksson M, Schüller H, Sjölund B. Hazards from transcutaneous nerve stimulation
17. Warning on Diathermy and Implanted Leads: FDA Patient Safety News: Show
in patients with pacemakers. Lancet 1978;1:1319.
#13, March 2003. Accessed 08 February 2009, http://www.accessdata.fda.gov/
2. Rasmussen MJ, Hayes DL, Vliestra RE et al. Can transcutaneous electrical nerve
scripts/cdrh/cfdocs/psn/printer.cfm?id=22.
stimulation be safely used in patients with permanent pacemakers? Mayo Clin
18. Beckerman H, de Bie RA, Bouter LM, DeCuyper HJ, Oostendorp RAB. The effi-
Proc 1988;63:443 –5.
cacy of laser therapy for musculoskeletal and skin disorders: a criteria-based
3. Vlay SC. Electromagnetic interference and ICD discharge related to chiropractic
meta-analysis or randomized clinical trials. Phys Ther 1992;72:483 – 9.
treatment. Pacing Clin Electrophysiol 1998;21:1996 –8.
4. Baranchuk A, Morillo CA. Guidelines for the prevention of sudden cardiac death: 19. Volkman H. Interference of implanted pacemaker in medical practice. Herzschritt-
filling the gap. In Gulizia M (ed.), In Emerging Pathologies in Cardiology. Milan: macherther Elektrophysiol 2004;15:65–72.
Springer; 2005. p223 –30. 20. Sluka KA, Walsh D. Transcutaneous electrical nerve stimulation: basic science
5. Anonymous. A comparison of antiarrhythmic-drug therapy with implantable defi- mechanisms and clinical effectiveness. J Pain 2003;4:109 –21.
brillators in patients resuscitated from near-fatal ventricular arrhythmias. The 21. Pinski SL, Trohman RG. Interference in implanted cardiac devices, part II. Pacing
Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. N Engl J Clin Electrophysiol 2002;25:1496 –509.
Med 1997;337:1576 –83. 22. Chen D, Philip M, Philip PA et al. Cardiac pacemaker inhibition by transcutaneous
6. Moss AJ, Zareba W, Hall WJ et al. Prophylactic implantation of a defibrillator in electrical nerve stimulation. Arch Phys Med Rehabil 1990;71:27 –30.

Downloaded from https://academic.oup.com/europace/article/11/7/850/498953 by guest on 29 November 2022


patients with myocardial infarction and reduced ejection fraction. N Engl J Med 23. Philbin DM, Marieb MA, Aithal KH, Schoenfeld MH. Inappropriate shocks deliv-
2002;346:877 –83. ered by an ICD as a result of sensed potentials from a transcutaneous electronic
7. Bardy GH, Lee KL, Mark DB et al. Amiodarone or an implantable cardioverter- nerve stimulation unit. Pacing Clin Electrophysiol 1999;22:400 –1.
defibrillator for congestive heart failure. N Engl J Med 2005;352:225–37. 24. Crevenna R, Stix G, Pleiner J et al. Electromagnetic interference by transcu-
8. Shields N, Gormley J, O’Hare N. Contraindications to continuous and pulsed taneous neuromuscular electrical stimulation in patients with bipolar sensing
short-wave diathermy. Phys Ther Rev 2002;7:133 –43. implantable cardioverter defibrillators: a pilot safety study. Pacing Clin Electrophysiol
9. Holmgren C, Carlsson T, Mannheimer C, Edvardsson N. Risk of interference from 2003;26:626 –9.
transcutaneous electrical nerve stimulation on the sensing function of implantable 25. Siu CW, Tse HF, Lau CP. Inappropriate implantable cardioverter defibrillator
defibrillators. Pacing Clin Electrophysiol 2008;31:151 –8. shock from a transcutaneous muscle stimulation device therapy. J Interv Card Elec-
10. Medtronic Electromagnetic Compatibility Table For Pacemakers and Defibrilla- trophysiol 2005;13:73–5.
tors (ICD). Metronic, Inc.; 2004 Accessed 12 July 2008, at https:// 26. Pyatt JR, Trenbath D, Chester M, Connelly DT. The simultaneous use of a biven-
www.medtronic.com.
tricular implantable cardioverter defibrillator (ICD) and transcutaneous electrival
11. Apaloo F. An acupuncture complication. Acupunct Med 1996;14:41.
nerve stimulation (TENS) unit: implications for device interaction. Europace 2003;
12. Chen FP, Hwang SJ, Lee HP, Yang HY, Chung C. Clinical study of syncope during
5:91 –3.
acupuncture treatment. Acupunct Electrother Res 1990;15:107 – 19.
27. Occhetta E, Bortnik M, Magnani A, Francalacci G, Marino P. Inappropriate implan-
13. Jones SL. Electromagnetic field interference and cardiac pacemakers. Phys Ther
table cardioverter-defibrillator discharges unrelated to supraventricular tachyar-
1976;56:1013 –8.
rhythmias. Europace 2006;8:863 –9.
14. Valtonen EJ, Lilius HG, Tiula E. Disturbances in the function of cardiac pacemakers
28. Robertson VJ, Chipchase LS, Laakso EL, Whelan KM, McKenna LJ. Guidelines for
caused by short wave and microwave diathermies and pulsed high frequency
current. Ann Chir Gynaecol Fenniae 1975;64:284–7. the Clinical Use of Electrophysical Agents. Victoria: Australian Physiotherapy Associ-
15. Vergassola R, Borgioli A, Chiodi L, Rossi D, Fazi A, Lebrun E et al. Changes in ation; 2001.
pacemakers and the wearers of pacemakers as a result of the use of different elec- 29. Campbell D, Bally K, Squires M. Radiation therapy in patients with cardiac rhythm
tromagnetic energy sources. Minerva Cardioangiol 1994;42:27 –32. devices: Minimizing the risk. Can J Cardiol 2007;23:NP14.

You might also like