Field Service Performance Verification Record Form
Forcetriad DC00087444 Revision B Page 1 of 4
Serial Number: _______________________ Date Received: ______________________
Hospital / Organization: ____________________ Software Version Pre___________________ Service Report: __________________________ Software Version Post: _________________
RMA Number: ___________________☐ N/A
Initial Inspection Accept Y/N Power Up Check Accept Y/N
Rear-panel Internal-component inspection N/A Front-panel System self-check Footswitch Verify Audio - Tone audible at high Power-cord and low ranges Low-Voltage Power Supply Voltage (GND TP82) Specification Reading +5 Vdc (TP81) 5.00 to 5.20 Vdc +12 Vdc (TP83) +11.40 to +13.20 Vdc -12 Vdc (TP90) -12.36 to -11.64 Vdc Safety Test in Accordance with IEC 60601-1 (Electrical Safety Testing) Forcetriad Ground Bond Chassis or Earth Chassis or Earth Testing Leakage NC Leakage NO < 0.20 Ω <100µA <500µA
and REM Source and REM Source and REM Sink Current Current Current (Mains on NC <10µA NO <50µA Applied Parts) <50µA Monopolar 1 Center Tine Monopolar 1 Right Tine Monopolar 2 Center Tine Monopolar 2 Right Tine Bipolar Left Tine Bipolar Right Tine Ligasure 1 Left Tine Ligasure 1 Right Tine Ligasure 2 Left Tine Ligasure 2 Right Tine UFP Port REM REM Function Y/N Automatic Mode Y/N Green indicator at 120 Ω RF generation at 20W with 25Ω connected Audible tone at >128Ω & < RF generation stop when disconnected 142Ω Green indicator at 60 Ω RF generation at 20W with 940Ω connected Audible tone >77Ω and <91Ω RF generation stop when disconnected Green indicator at 30Ω RF generation at 20W with 1070Ω connected Green indicator at 10Ω Red indicator before 1Ω This document is electronically controlled M e dt ronic Cont rolle d Inform a t ion Field Service Performance Verification Record Form Forcetriad DC00087444 Revision B Page 2 of 4
Serial Number: ________________________
Verifying Autobipolar Function Y/N Footswitch Mode Y/N
Power at 20W RF generation on/off w/o 45Ω connected Cross Coupling 200Ω Energy Channel Tested Energy Mode/Effect Specifications Reading Mono 2 Cut 0 to 140 mA Monopolar 1 Mono 2 Coag Spray 0 to 140 mA Mono 2 Cut 0 to 140 mA UFP Mono 2 Coag Spray 0 to 140 mA Mono 1 Cut 0 to 140 mA Monopolar 2 Mono 1 Coag Spray 0 to 140 mA Mono 1 Cut 0 to 140 mA Bipolar Mono 1 Coag Spray 0 to 140 mA Mono 1 Cut 0 to 140 mA LigaSure 1 Mono 1 Coag Spray 0 to 140 mA LigaSure 2 Mono 1 Cut 0 to 140 mA Mono 1 Coag Spray 0 to 140 mA Power Delivery at Receptacle Power Energy Mode/Effect Specifications Reading 100 Ω Low at 95W 900 to 1044 mA Bipolar 100 Ω Standard 900 to 1044 mA (Bipolar FTSW) 100 Ω Macro 900 to 1044 mA Monopolar 1 300 Ω Cut Pure at 300W 922 to 1072 mA (2-button HDSW) 500 Ω Coag Spray at 120W 452 to 525 mA Monopolar 2 300 Ω Cut Pure at 300W 922 to 1072 mA (2-button HDSW) 500 Ω Coag Spray at 120W 452 to 525 mA UFP 300 Ω Cut Pure at 300W 922 to 1072 mA (Monopolar FTSW) 500 Ω Coag Spray at 120W 452 to 525 mA 300 Ω Cut 922 to 1072 mA Monopolar 1 (Force TriVerse) 300 Ω VL Mode 753 to 876 mA 500 Ω Coag 452 to 525 mA 300 Ω Cut 922 to 1072 mA Monopolar 2 300 Ω VL Mode 753 to 876 mA (Force TriVerse) 500 Ω Coag 452 to 525 mA
This document is electronically controlled M e dt ronic Cont rolle d Inform a t ion
Field Service Performance Verification Record Form Forcetriad DC00087444 Revision B Page 3 of 4
Serial Number: ________________________
Power Output – Bipolar - 100 Ω
Low 291 to 339 mA 10 Watt Standard 291 to 339 mA Macro 291 to 339 mA Low 504 to 587 mA 30 Watt Standard 504 to 587 mA Macro 504 to 587 mA Low 900 to 1044 mA 95 Watt Standard 900 to 1044 mA Marco 900 to 1044 mA Power Output - Testing Cut Output - 300 Ω Prue 146 to 218 mA 10 Watt Blend 146 to 218 mA Pure 461 to 536 mA 75 Watt Blend 461 to 536 mA 300 Watt Prue 922 to 1072 mA 200 Watt Blend 753 to 876 mA Power Output – Testing Valleylab Mode Output – 300 Ω 10 Watt VL Mode 146 to 218 mA 75 Watt VL Mode 461 to 536 mA 200 Watt VL Mode 753 to 876 mA Power Output – Testing Coag Output 500 Ω Fulgurate 20 to 100mA 1 Watt Spray 20 to 100mA Fulgurate 226 to 263 mA 30 Watt Spray 226 to 263 mA Fulgurate 452 to 525 mA 120 Watt Spray 452 to 525 mA Power Output – Test LigaSure Output – 50 Ω LigaSure 1 LigaSure 2 1A 288 to 355 mA LigaSure 2.7 A 761 to 885 mA 5.5 A 1540 to 1790 mA High-Frequency Leakage 200Ω Power Energy Mode/Effect Specifications Reading Cut Prue 0 to 100 mA Cut Blend 0 to 100 mA UFP Leakage HWD 0 to 100 mA COAG Fulgurate 0 to 100 mA COAG Spray 0 to 100 mA
This document is electronically controlled M e dt ronic Cont rolle d Inform a t ion
Field Service Performance Verification Record Form Forcetriad DC00087444 Revision B Page 4 of 4
Serial Number: ________________________
Cut Prue 0 to 100 mA
Cut Blend 0 to 100 mA Monopolar 1 Leakage HWD 0 to 100 mA COAG Fulgurate 0 to 100 mA COAG Spray 0 to 100 mA Cut Prue 0 to 100 mA Cut Blend 0 to 100 mA Monopolar 2 Leakage HWD 0 to 100 mA COAG Fulgurate 0 to 100 mA COAG Spray 0 to 100 mA Bipolar Standard (left tine) 0 to 69 mA Bipolar Macro (left tine) 0 to 69 mA Bipolar Low (left tine) 0 to 69 mA Bipolar Leakage Bipolar Standard (right tine) 0 to 69 mA Bipolar Macro (right tine) 0 to 69 mA Bipolar Low (right tine) 0 to 69 mA LigaSure (left tine) 0 to 130 mA LigaSure 1 Leakage LigaSure (right tine) 0 to 130 mA LigaSure (left tine) 0 to 130 mA LigaSure 2 Leakage LigaSure (right tine) 0 to 130 mA Cut Prue 0 to 100 mA Cut Blend 0 to 100 mA Patient Return Leakage HWD 0 to 100 mA COAG Fulgurate 0 to 100 mA COAG Spray 0 to 100 mA
Field Service Engineer: Date: