AgraQuant® Histamine Rapid ELISA Test kit – Quick Guide

AgraQuant® Histamine Rapid Test kit includes:


Test procedures for:
● 48 Antibody-coated microwells • Fresh fish
● 1 master block of 48 wells for acylation • Canned fish
● Set of 6 standards • Salted fish
● Set of 2 positive controls • Fish meal
● 1 Bottle of acylation reagent
● 2 Bottles of acylation buffer
● 1 bottle of conjugate
● 1 Bottle of substrate
● 1 Bottle of stop solution
● 1 Bottle of wash buffer

Ordering Information:
● AgraQuant® Histamine Rapid 10002018

Shelf life: 18 months

Also available:
● AgraQuant® Histamine (10002017)
AgraQuant® Histamine Rapid ELISA Test kit – Quick Guide

Limit of detection (LOD): 1 ppm histamine


Fresh fish and other
Fish meal
Limit of quantification (LOQ): 3 ppm histamine fish samples
Range of quantification: 3 – 300 ppm histamine
Sample size 10 g

Extraction
Distiled Water
250 mL, RT
IMPORTANT INFORMATION: Volume

Homogenization Blend - 1 to 2 min Stir - 10 min, RT


The Acylation reagent must be allowed to
Centrifugation 1 mL, 5 min at 3000g
reach room temperature before the assay.
Extract or
Alternatively, it can be always stored at RT Standards
50 µL per well in the master block

(in the Master Block)


(20-25°C).

Acylation
Acylation Buffer 1.5 mL

Acylation
Reagent
50 µL
Wash Buffer Preparation:
Incubation Rotational shake - 5 min, RT at 600 rpm

The wash buffer comes as a 50X concentrate. 1 Add Samples


and Standards
50 µL acylated sample/standard per well

Dilute 1:50 with distiled water before use.


2 Add Conjugate 100 µL conjugate per well - 10 min at RT on a shaker at 600 rpm
1st Incubation

3
20 mL Wash Buffer Concentrate (50X)
Washing Wash 3 times with wash buffer

+
980 mL disitiled water
4 Add Substrate 100 µL substrate per well - 10 min at RT on a shaker at 600 rpm (avoid sunlight)
2nd Incubation

6 Add Stop
solution
100 µL stop solution per well

1000 mL Wash Buffer


7 Read with
ELISA Reader
Read at 450 nm

Romer Labs Inc • Newark, DE, USA • T +1 302 781 6400 • F +1 302 781 6378 • E office@romerlabs.com Romer Labs is part of ERBER Group