Summary-A gravimetric finish has been applied to the semi-micro determination of phosphorus in
organic compounds after fusion with sodium peroxide. It has been shown that the precipitate of
quinoline phosphomolybdate prepared from 3 mg of phosphorus has the theoretical composition
and that, using a special weighing technique, very precise results can be obtained. Two other weighing
techniques are described which give results satisfactory for semi-micro work.
EXPERIMENTAL
Oven-drying conditions: Thermogravimetric examination of the precipitate previously dried at
110” established that no decomposition occurred below a temperature of 370”. It was found that
15 minutes oven drying at 160” was sufficient for routine purposes but for the highest possible accuracy
a drying time of 30 minutes is recommended.
due to the hygroscopic nature of the precipitate. The apparent molecular weights shown in Table I
were determined by multiplying the precipitate weight, in mg, by ten.
TABLEI. APPARENTMOLECULAR
WEIGHT OF THEPRECIPITATE
2214.0 3.100 +2
2212.0 3.097 -1
2214.1 3.100 +2
2212.0 3.097 -1
2213.6 3.099 +1
2211.5 3.096 -2
It is apparent that, under the conditions employed, the precipitate has the theoretical composition
and that the theoretical factor of 0.01400 can be used for calculating the phosphorus content.
I
I
Phos- Phos- Phos- ~
phorus i phorus i phorus Absolute
Compound
found, ’ CA, ( found, error, %
i *g
I *g *g 1 .% , % )
-_. ._ _ / -_-___ I
* The preparation of this compound has not yet been described in the literature.
These precipitates were subsequently dried at 160” and weighed using the two-crucible technique
in order to attempt to establish the composition of the solvent washed precipitate. The results
obtained are shown in Table III.
Weight of precipitate, mg
Phosphorus found (factor 0.013757), mg
Phosphorus found (factor 0.013775), mg 1
The oven-drying check shows that there is no loss of precipitate by the use of the organic solvents.
The factor 0.013757 was derived from the average weight of precipitate which gave an apparent
molecular weight of 2252.05; 0.013775 is the theoretical factor for quinoline phosphomolybdate
dihydrate (molecular weight 224890).
DISCUSSION
The results obtained by the two-crucible weighing technique indicate that, in spite
of the hygroscopicity of the precipitate, very precise results may be obtained by using
108 T. R. F. W. FENNELLand J. R. WEBB
Acknowledgements-We thank Mr. E. J. P. Fear and Dr. I. M. White for provision of samples and
Dr. R. Belcher for useful discussions.