You are on page 1of 5

Accelerated communication

Received: 18 February 2013 Revised: 6 May 2013 Accepted: 7 May 2013 Published online in Wiley Online Library

(wileyonlinelibrary.com) DOI 10.1002/jms.3232

Indirect hydrogen analysis by gas chromatography


coupled to mass spectrometry (GC–MS)
V. Varlet,* F. Smith and M. Augsburger
Gas chromatography (GC) is an analytical tool very useful to investigate the composition of gaseous mixtures. The different
gases are separated by specific columns but, if hydrogen (H2) is present in the sample, its detection can be performed by a
thermal conductivity detector or a helium ionization detector. Indeed, coupled to GC, no other detector can perform this
detection except the expensive atomic emission detector. Based on the detection and analysis of H2 isotopes by low-pressure
chemical ionization mass spectrometry (MS), a new method for H2 detection by GC coupled to MS with an electron ionization
ion source and a quadrupole analyser is presented. The presence of H2 in a gaseous mixture could easily be put in evidence by
the monitoring of the molecular ion of the protonated carrier gas. Copyright © 2013 John Wiley & Sons, Ltd.

Keywords: hydrogen; helium; gas chromatography–mass spectrometry; thermal conductivity detector; carrier gas protonation

Introduction As a result, GC–TCD remains the unique analytical tool to de-


tect H2 in a gaseous mixture. However, another GC detection of
Hydrogen (H2) is one of the main putrefactive gases. It was found in H2 is possible using MS. Based on the detection and analysis of
human bodies in cadaveric alteration in forensic studies in a range H2 isotopes by low-pressure chemical ionization MS, it is
going from 0.5 to 50%.[1,2] A similar range of concentrations was possible to perform an indirect detection of H2 with an electron
also noticed in rabbits[3] and whales cadavers.[4,5] Therefore, the ionization (EI) ion source. Analysis of H2 isotopes by low-pressure
detection of this gas is of high importance in forensic toxicology. chemical ionization involves the detection of protonated reagent
Due to its physicochemical properties (explosive), its detection is gas ions (such as argon, neon, He or N2) rather than the direct
also crucial in environmental and exposure monitorings. detection of 1H+2 and 2H+2 at masses below the scan limits of most
H2 analysis is possible by mass spectrometry (MS). MS was quadrupole GC–MS systems.[9] Gas protonation of H2, He and O2
employed to detect H2 isotopes. Some specific analysers[6] and in a N2 purged atmosphere was already studied with a quadru-
high-resolution systems have given the best results. However, pole ion trap mass spectrometer.[10] It has shown that H2, H2O
when coupled to a prior gas chromatographic separation, H2 and N2 can easily be protonated given ion signals at m/z 3 (H+3 ),
analysis is not possible, especially because no carrier gas lighter 19 (H3O+) and 29 (N2H+). However, O2, and especially He, did
than H2 is available. Indeed, according to the mass spectrometer not exhibit any protonation. Similarly, having an EI ion source
system, it has already been shown that gas chromatography and a quadrupole mass spectrometer, we have put in evidence
coupled to MS (GC–MS) analysis of a lighter gas than the carrier H2 because of the protonation of the carrier gas. A positive MS
gas was not possible.[7] Thus, GC–MS analysis of helium (He) signal appeared at the expected retention time of the gas used
was possible using H2 as carrier gas whereas using nitrogen as carrier gas when analysed as analyte with another carrier gas.
(N2), it displays surprising negative peaks with a quadrupole mass To our knowledge, this is the first time that this phenomenon is
spectrometer. reported with an EI ion source. Protonation of carrier gas by
Consequently, GC analysis of H2 is only possible with a ther- hydrogen in GC–MS analysis is also described for the first time.
mal conductivity detector (TCD), a helium ionization detector
(HID) and an atomic emission detector (AED). For the analysis
of most gases, He is the carrier gas of choice because of the Material and methods
large difference of thermal conductivity between He and
other gases which enhances the TCD sensitivity. However, Reagents
He and H2 thermal conductivity are not very different. Therefore,
All gases (N2, H2 and He) were obtained from Carbagas
N2 can be used as an alternative but if a double TCD and MS
(Lausanne, Switzerland). Putrefaction gases were sampled from
detection is monitored, the MS chromatography with N2 as
carrier gas is less optimal than with He. Nevertheless, H2 TCD
detection with He as carrier gas is possible but complex because
the thermal conductivity of the He–H2 binary mixture is not a * Correspondence to: Vincent Varlet, Forensic Toxicology and Chemistry Unit,
simple monotonic increasing function of H2 concentration and University Center of Legal Medicine Lausanne - Geneva, CH-1011 Lausanne,
Switzerland. E-mail: vincent.varlet@chuv.ch
goes through a positive local minimum at low concentrations
of H2 before increasing (in negative values) to the pure H2 Forensic Toxicology and Chemistry Unit, University Center of Legal Medicine
concentration (100 %).[8] Lausanne - Geneva, CH-1011 Lausanne, Switzerland
914

J. Mass Spectrom. 2013, 48, 914–918 Copyright © 2013 John Wiley & Sons, Ltd.
Indirect hydrogen analysis

human bodies at the University Center of Legal Medicine of separation of permanent gases such as H2, He, argon, oxygen
Lausanne (Switzerland). (O2), N2, methane (CH4) and carbon monoxide (CO) on the
molecular sieve 5 Å PLOT capillary column, and the separation of
Analytical equipment CH4, hydrogen sulfide (H2S), carbon dioxide (CO2) and heavier
volatile hydrocarbons on the Porabond Q (with no separation of
An Agilent 6890N GC (Agilent Technologies, Palo Alto, CA) air constituents) (Fig. 1). Moreover, the double detection (MS and
combined to a headspace gas autosampler and equipped with TCD) allows to get the individual and precise concentration of a
an Agilent Select Permanent Gases column was used. This col- gas in a gaseous mixture (MS quantification) and to get the individ-
umn is specially designed for gas analysis and is constituted of ual proportion of this gas (percentage) in the total gaseous mixture
two capillary columns set in parallel: a molecular sieve 5 Å PLOT (TCD quantification). However, TCD allows the analysis of H2
capillary column (10 m  0.32 mm) and a Porabond Q (50 m  whereas MS cannot theoretically permit this measurement.
0.53 mm). The temperature program was 40  C, held during This chromatographic system was applied to the gas screening
5 min. Injector (splitless mode) was set at 100  C and interface of gaseous samples from human bodies in cadaveric alteration.
MS temperature at 230  C. He and N2 were employed as carrier Figure 2 displays the MS chromatogram result of right heart
gas. The detection was performed by an Agilent 5973 mass spec- ventricule gas mixture sampled from a human body in advanced
trometer (Agilent Technologies, Palo Alto, CA), operating in the EI cadaveric alteration. Important amounts of CO2 were noticed
mode at 70 eV and with a TCD (Agilent Technologies, Palo Alto, followed by N2 and O2 concentrations and finally, traces of CH4,
CA) maintained at 170  C. The end of the Select Permanent Gases ammonia (NH3) and H2S. H2 was not identified on this MS
column was connected to a three-way valve enabling the gases chromatogram where its signal was expected just before O2
to be directed to TCD or MS detectors. The valve was operated signal retention time.
manually to connect the GC effluent to TCD or MS.
GC–MS H2 identification with N2 as carrier gas
Results and discussion Figure 3 displays the MS chromatogram of a pure H2 pulse in GC
injector. Unexpected peaks at the retention times of N2 on
Chromatographic system
Molecular sieve and Porabond Q are noticeable. Blank or N2
The chromatographic system was used to perform a gas screening injections in same chromatographic conditions have shown no
in forensic studies. This equipment allows in the same run the baseline variation. Air injection caused a chromatographic signal

Figure 1. Chromatographic design for hydrogen and other gas analysis.

Figure 2. MS chromatogram (Total Ion Current) of gases in right heart ventricule gas mixture sampled from a human body in advanced cadaveric
915

alteration.

J. Mass Spectrom. 2013, 48, 914–918 Copyright © 2013 John Wiley & Sons, Ltd. wileyonlinelibrary.com/journal/jms
V. Varlet, F. Smith and M. Augsburger

Figure 3. MS chromatogram (Total Ion Current) of pure hydrogen pulse in GC injector (N2 as carrier gas).

only for O2, showing that N2 of air cannot be put in evidence Extract ion chromatograms and mass spectrum have shown a
when N2 is used as carrier gas (Fig. 4). Focusing on the first peak main ion at m/z = 5 (Fig. 6(B) and (C)). This mass spectrum corre-
obtained in Fig. 3, it appears that the positive MS signal (Fig. 5(A)) sponds to the mass spectrum of He except that the molecular ion
corresponds to a negative TCD signal (Fig. 5(B)) specific to H2 is at m/z = 5 instead of m/z = 4. It appears that the carrier gas (He)
when injected in important quantity.[11] The mass spectrum of has been protonated by H2 eluting from the chromatographic
this signal exhibits a molecular ion at m/z = 29, with a second system and has given a signal at the expected retention time of
ion at m/z = 14. This mass spectrum corresponds to the mass the gas when not used as carrier gas, but with a molecular ion
spectrum of N2 except that the molecular ion is at m/z = 29 corresponding to m/z + H+, herein m/z = 4 + H+ = 5.
instead of m/z = 28. It appears that the carrier gas (N2) has been Whatever the carrier gas, the eluting H2 protonates the carrier
protonated by H2 eluting from the chromatographic system gas resulting in a signal at the retention time of the gas when not
and has given a signal at the expected retention time of the used as carrier gas. This unexpected reaction of chemical ioniza-
gas when not used as carrier gas, but with a molecular ion tion of the carrier gas by an eluting compound using an EI ion
corresponding to m/z + H+, herein m/z = 28 + H+ = 29. source allows a reliable detection of H2 by GC–MS. TCD is still
important for quantification, but this result permits to better
investigate MS chromatograms of injected gaseous mixtures.
GC–MS H2 identification with He as carrier gas
Therefore, without TCD detector, H2 cannot be identified in
Figure 6 displays the MS chromatogram of a H2 mixture in air. gaseous samples where its presence is expected such as in a
Unexpected peaks at the retention times of He on Molecular heart ventricule gas mixture sampled from a human body in
sieve and Porabond Q are noticeable. Blank or He injections in advanced cadaveric alteration (Fig. 2). However, knowing the
the same chromatographic conditions have shown no baseline behaviour of H2 with the carrier gas, it is possible to identify H2
variation (Fig. 7). Positive MS signals were obtained (Fig. 6(A)). by GC–MS. Performing an extracted ion chromatogram of the
916

Figure 4. Air injection (N2 + O2) with N2 as carrier gas.

wileyonlinelibrary.com/journal/jms Copyright © 2013 John Wiley & Sons, Ltd. J. Mass Spectrom. 2013, 48, 914–918
Indirect hydrogen analysis

Figure 5. Pulse hydrogen in GC injector (N2 as carrier gas), A : MS signal (zoom from Fig. 5), B : TCD signal (shift of 0.20 min between A and B is due to
the chromatographic design), C : Mass spectrum obtained for H2 peak.

Figure 6. Mixture of hydrogen in air (He as carrier gas), A : MS chromatogram of hydrogen mixture in air, B : Extracted ion chromatogram from MS
917

chromatogram A at m/z = 5, C : Mass spectrum obtained for unexpected peaks.

J. Mass Spectrom. 2013, 48, 914–918 Copyright © 2013 John Wiley & Sons, Ltd. wileyonlinelibrary.com/journal/jms
V. Varlet, F. Smith and M. Augsburger

Figure 7. Blank injection (no injection) MS chromatogram (He as carrier gas).

Figure 8. Extracted ion chromatogram (m/z = 5) from MS chromatogram of Fig. 2 (He as carrier gas).

MS chromatogram at m/z of protonated carrier gas, a positive signal [3] Y. Bernaldo de Quirós, O. González-Díaz, A. Møllerløkken,
of the carrier gas should indicate the presence of H2 as shown in A.O. Brubakk, A. Hjelde, P. Saavedra, A. Fernandez. Differentation
at autopsy between in vivo gas embolism and putrefaction using
Fig. 8. However, unexpected double peaks were identified: proton- gas composition analysis. Int. J. Legal Med. 2013. DOI: 10.1007/
ation of He by H2 is noticeable, but another signal is displayed at s00414-012-0783-6.
O2 retention time. The cause of this phenomenon remains unknown. [4] Y. Bernaldo de Quirós, O. González-Díaz, P. Saavedra, M. Arbelo,
E. Sierra, S. Sacchini, P.D. Jepson, S. Mazzariol, G. Di Guardo,
A. Fernandez. Methodology for in situ gas sampling, transport
and laboratory analysis of gases from stranded cetaceans. Sci.
Conclusion Rep. 2011, 1, 1–10.
[5] Y. Bernaldo de Quirós, O. González-Díaz, M. Arbelo, E. Sierra,
S. Sacchini, A. Fernández. Decompression vs. decomposition: dis-
An unusual reaction of carrier gas protonation by H2 was noticed tribution, amount, and gas composition of bubbles in stranded
during H2 analysis by GC–MS/TCD with an EI ion source. This marine mammals. Front Physiol. 2012, 3. DOI: 10.3389/
reaction could be very useful in GC–MS gas screening and was fphys.2012.00177.
confirmed by two carrier gases (He and N2). Indeed, the presence [6] J. Y. Li, L. Shi, S. L. Hu. Methodological study on measurement of
hydrogen abundance in hydrogen isotopes system by low resolu-
of H2 in a gaseous mixture could be put in evidence easily by the
tion mass spectrometry. Mass Spectrom. Lett. 2011, 2, 1–7.
monitoring of the molecular ion of protonated carrier gas. It gives [7] F. Musshoff, L. Hagemeier, K. Kirschbaum, B. Madea. Two cases of
the opportunity to detect H2 with another GC detector and not to suicide by asphyxiation due to helium and argon. Forensic Sci.
be obliged to be equipped with TCD, HID or AED. Int. 2012. DOI: 10.1016/j.forsciint.2012.08.049.
[8] K. Snavely, B. Subramaniam. Thermal conductivity detector analysis
of hydrogen using helium carrier gas and HayesepW D columns.
Acknowledgement J. Chromatogr. Sci. 1998, 36, 191–196.
The authors greatly thank Kjetil Johnsen (Johnsen Consulting & [9] A. R. Mack, Y. Liu, W. Luo, L. Liu, J. F. Parcher. Detection and analysis
of hydrogen isotopes and spin isomers by low-pressure chemical
Services) for his chromatographic expertise during scientific discussions. ionization mass spectrometry. Anal. Chem. 1994, 66, 115–118.
[10] A. K. Ottens C. Richard Arkin, T. P. Griffin, P. T. Palmer, W.W., Harrison.
References Ion-molecule reactions in quadrupole ion trap mass spectrometry:
implications for lightweight gas analysis. Int. J. Mass Spectrom. 2005,
[1] T. Bajanowski, A. West, B. Brinkmann. Proof of fatal air embolism. Int. 243, 31–39.
J. Legal Med. 1998, 111, 208–211. [11] R. Villalobos, G. R. Nuss. Measurement of hydrogen in pro-
[2] T. Bajanowski, H. Köhler, A. DuChesne, E. Koops, B. Brinkmann. Proof of cess streams by gas chromatography. ISA Trans. 1965, 4,
918

fatal air embolism after exhumation. Int. J. Legal Med. 1998, 112, 2–7. 281–286.

wileyonlinelibrary.com/journal/jms Copyright © 2013 John Wiley & Sons, Ltd. J. Mass Spectrom. 2013, 48, 914–918

You might also like