Professional Documents
Culture Documents
GCMS TCD
GCMS TCD
Received: 18 February 2013 Revised: 6 May 2013 Accepted: 7 May 2013 Published online in Wiley Online Library
Keywords: hydrogen; helium; gas chromatography–mass spectrometry; thermal conductivity detector; carrier gas protonation
J. Mass Spectrom. 2013, 48, 914–918 Copyright © 2013 John Wiley & Sons, Ltd.
Indirect hydrogen analysis
human bodies at the University Center of Legal Medicine of separation of permanent gases such as H2, He, argon, oxygen
Lausanne (Switzerland). (O2), N2, methane (CH4) and carbon monoxide (CO) on the
molecular sieve 5 Å PLOT capillary column, and the separation of
Analytical equipment CH4, hydrogen sulfide (H2S), carbon dioxide (CO2) and heavier
volatile hydrocarbons on the Porabond Q (with no separation of
An Agilent 6890N GC (Agilent Technologies, Palo Alto, CA) air constituents) (Fig. 1). Moreover, the double detection (MS and
combined to a headspace gas autosampler and equipped with TCD) allows to get the individual and precise concentration of a
an Agilent Select Permanent Gases column was used. This col- gas in a gaseous mixture (MS quantification) and to get the individ-
umn is specially designed for gas analysis and is constituted of ual proportion of this gas (percentage) in the total gaseous mixture
two capillary columns set in parallel: a molecular sieve 5 Å PLOT (TCD quantification). However, TCD allows the analysis of H2
capillary column (10 m 0.32 mm) and a Porabond Q (50 m whereas MS cannot theoretically permit this measurement.
0.53 mm). The temperature program was 40 C, held during This chromatographic system was applied to the gas screening
5 min. Injector (splitless mode) was set at 100 C and interface of gaseous samples from human bodies in cadaveric alteration.
MS temperature at 230 C. He and N2 were employed as carrier Figure 2 displays the MS chromatogram result of right heart
gas. The detection was performed by an Agilent 5973 mass spec- ventricule gas mixture sampled from a human body in advanced
trometer (Agilent Technologies, Palo Alto, CA), operating in the EI cadaveric alteration. Important amounts of CO2 were noticed
mode at 70 eV and with a TCD (Agilent Technologies, Palo Alto, followed by N2 and O2 concentrations and finally, traces of CH4,
CA) maintained at 170 C. The end of the Select Permanent Gases ammonia (NH3) and H2S. H2 was not identified on this MS
column was connected to a three-way valve enabling the gases chromatogram where its signal was expected just before O2
to be directed to TCD or MS detectors. The valve was operated signal retention time.
manually to connect the GC effluent to TCD or MS.
GC–MS H2 identification with N2 as carrier gas
Results and discussion Figure 3 displays the MS chromatogram of a pure H2 pulse in GC
injector. Unexpected peaks at the retention times of N2 on
Chromatographic system
Molecular sieve and Porabond Q are noticeable. Blank or N2
The chromatographic system was used to perform a gas screening injections in same chromatographic conditions have shown no
in forensic studies. This equipment allows in the same run the baseline variation. Air injection caused a chromatographic signal
Figure 2. MS chromatogram (Total Ion Current) of gases in right heart ventricule gas mixture sampled from a human body in advanced cadaveric
915
alteration.
J. Mass Spectrom. 2013, 48, 914–918 Copyright © 2013 John Wiley & Sons, Ltd. wileyonlinelibrary.com/journal/jms
V. Varlet, F. Smith and M. Augsburger
Figure 3. MS chromatogram (Total Ion Current) of pure hydrogen pulse in GC injector (N2 as carrier gas).
only for O2, showing that N2 of air cannot be put in evidence Extract ion chromatograms and mass spectrum have shown a
when N2 is used as carrier gas (Fig. 4). Focusing on the first peak main ion at m/z = 5 (Fig. 6(B) and (C)). This mass spectrum corre-
obtained in Fig. 3, it appears that the positive MS signal (Fig. 5(A)) sponds to the mass spectrum of He except that the molecular ion
corresponds to a negative TCD signal (Fig. 5(B)) specific to H2 is at m/z = 5 instead of m/z = 4. It appears that the carrier gas (He)
when injected in important quantity.[11] The mass spectrum of has been protonated by H2 eluting from the chromatographic
this signal exhibits a molecular ion at m/z = 29, with a second system and has given a signal at the expected retention time of
ion at m/z = 14. This mass spectrum corresponds to the mass the gas when not used as carrier gas, but with a molecular ion
spectrum of N2 except that the molecular ion is at m/z = 29 corresponding to m/z + H+, herein m/z = 4 + H+ = 5.
instead of m/z = 28. It appears that the carrier gas (N2) has been Whatever the carrier gas, the eluting H2 protonates the carrier
protonated by H2 eluting from the chromatographic system gas resulting in a signal at the retention time of the gas when not
and has given a signal at the expected retention time of the used as carrier gas. This unexpected reaction of chemical ioniza-
gas when not used as carrier gas, but with a molecular ion tion of the carrier gas by an eluting compound using an EI ion
corresponding to m/z + H+, herein m/z = 28 + H+ = 29. source allows a reliable detection of H2 by GC–MS. TCD is still
important for quantification, but this result permits to better
investigate MS chromatograms of injected gaseous mixtures.
GC–MS H2 identification with He as carrier gas
Therefore, without TCD detector, H2 cannot be identified in
Figure 6 displays the MS chromatogram of a H2 mixture in air. gaseous samples where its presence is expected such as in a
Unexpected peaks at the retention times of He on Molecular heart ventricule gas mixture sampled from a human body in
sieve and Porabond Q are noticeable. Blank or He injections in advanced cadaveric alteration (Fig. 2). However, knowing the
the same chromatographic conditions have shown no baseline behaviour of H2 with the carrier gas, it is possible to identify H2
variation (Fig. 7). Positive MS signals were obtained (Fig. 6(A)). by GC–MS. Performing an extracted ion chromatogram of the
916
wileyonlinelibrary.com/journal/jms Copyright © 2013 John Wiley & Sons, Ltd. J. Mass Spectrom. 2013, 48, 914–918
Indirect hydrogen analysis
Figure 5. Pulse hydrogen in GC injector (N2 as carrier gas), A : MS signal (zoom from Fig. 5), B : TCD signal (shift of 0.20 min between A and B is due to
the chromatographic design), C : Mass spectrum obtained for H2 peak.
Figure 6. Mixture of hydrogen in air (He as carrier gas), A : MS chromatogram of hydrogen mixture in air, B : Extracted ion chromatogram from MS
917
J. Mass Spectrom. 2013, 48, 914–918 Copyright © 2013 John Wiley & Sons, Ltd. wileyonlinelibrary.com/journal/jms
V. Varlet, F. Smith and M. Augsburger
Figure 8. Extracted ion chromatogram (m/z = 5) from MS chromatogram of Fig. 2 (He as carrier gas).
MS chromatogram at m/z of protonated carrier gas, a positive signal [3] Y. Bernaldo de Quirós, O. González-Díaz, A. Møllerløkken,
of the carrier gas should indicate the presence of H2 as shown in A.O. Brubakk, A. Hjelde, P. Saavedra, A. Fernandez. Differentation
at autopsy between in vivo gas embolism and putrefaction using
Fig. 8. However, unexpected double peaks were identified: proton- gas composition analysis. Int. J. Legal Med. 2013. DOI: 10.1007/
ation of He by H2 is noticeable, but another signal is displayed at s00414-012-0783-6.
O2 retention time. The cause of this phenomenon remains unknown. [4] Y. Bernaldo de Quirós, O. González-Díaz, P. Saavedra, M. Arbelo,
E. Sierra, S. Sacchini, P.D. Jepson, S. Mazzariol, G. Di Guardo,
A. Fernandez. Methodology for in situ gas sampling, transport
and laboratory analysis of gases from stranded cetaceans. Sci.
Conclusion Rep. 2011, 1, 1–10.
[5] Y. Bernaldo de Quirós, O. González-Díaz, M. Arbelo, E. Sierra,
S. Sacchini, A. Fernández. Decompression vs. decomposition: dis-
An unusual reaction of carrier gas protonation by H2 was noticed tribution, amount, and gas composition of bubbles in stranded
during H2 analysis by GC–MS/TCD with an EI ion source. This marine mammals. Front Physiol. 2012, 3. DOI: 10.3389/
reaction could be very useful in GC–MS gas screening and was fphys.2012.00177.
confirmed by two carrier gases (He and N2). Indeed, the presence [6] J. Y. Li, L. Shi, S. L. Hu. Methodological study on measurement of
hydrogen abundance in hydrogen isotopes system by low resolu-
of H2 in a gaseous mixture could be put in evidence easily by the
tion mass spectrometry. Mass Spectrom. Lett. 2011, 2, 1–7.
monitoring of the molecular ion of protonated carrier gas. It gives [7] F. Musshoff, L. Hagemeier, K. Kirschbaum, B. Madea. Two cases of
the opportunity to detect H2 with another GC detector and not to suicide by asphyxiation due to helium and argon. Forensic Sci.
be obliged to be equipped with TCD, HID or AED. Int. 2012. DOI: 10.1016/j.forsciint.2012.08.049.
[8] K. Snavely, B. Subramaniam. Thermal conductivity detector analysis
of hydrogen using helium carrier gas and HayesepW D columns.
Acknowledgement J. Chromatogr. Sci. 1998, 36, 191–196.
The authors greatly thank Kjetil Johnsen (Johnsen Consulting & [9] A. R. Mack, Y. Liu, W. Luo, L. Liu, J. F. Parcher. Detection and analysis
of hydrogen isotopes and spin isomers by low-pressure chemical
Services) for his chromatographic expertise during scientific discussions. ionization mass spectrometry. Anal. Chem. 1994, 66, 115–118.
[10] A. K. Ottens C. Richard Arkin, T. P. Griffin, P. T. Palmer, W.W., Harrison.
References Ion-molecule reactions in quadrupole ion trap mass spectrometry:
implications for lightweight gas analysis. Int. J. Mass Spectrom. 2005,
[1] T. Bajanowski, A. West, B. Brinkmann. Proof of fatal air embolism. Int. 243, 31–39.
J. Legal Med. 1998, 111, 208–211. [11] R. Villalobos, G. R. Nuss. Measurement of hydrogen in pro-
[2] T. Bajanowski, H. Köhler, A. DuChesne, E. Koops, B. Brinkmann. Proof of cess streams by gas chromatography. ISA Trans. 1965, 4,
918
fatal air embolism after exhumation. Int. J. Legal Med. 1998, 112, 2–7. 281–286.
wileyonlinelibrary.com/journal/jms Copyright © 2013 John Wiley & Sons, Ltd. J. Mass Spectrom. 2013, 48, 914–918